Professional Documents
Culture Documents
Manual
OPERATION MANUAL
COLOR SERVO-VENTILATOR
This Operation manual contains the information required for the correct use of the COLOR Servo-
Ventilator.
Manufacturer:
K. TAKAOKA IND. E COM. LTDA.
Rua General Izidoro Dias Lopes, 121/141
Bairro Vila Paulicéia - São Bernardo do Campo / SP
Postal Code 09687-100 - Brasil
Phone: + 55 11 4176-3500
Fax: + 55 11 4176-3570
Web site: www.takaoka.com.br
e-mail: kt@takaoka.com.br
CNPJ : 61.489.381/0001-09
I.E.: 103.735.350.115
Suggestions, questions or complaints:
Call Center: + 55 11 4176-3600
Product Classification:
• IEC 60601-2-12/2001 (Medical electrical equipment - Part 2-12: Particular requirements for the
safety of lung ventilators - Critical care ventilators)
• IEC 60601-1-2/2006 (Medical electrical equipment - Part 1-2: General requirements for basic
safety and essential performance - Collateral standard: Electromagnetic compatibility -
Requirements and tests)
• EN 794-1:1997 +amendment 2000- ( Lung ventilators - Part 1: Particular requirements for critical
care ventilators)
• ISO 21647:2004 (Medical electrical equipment — Particular requirements for the basic safety and
essential performance of respiratory gás monitors)
• Class 1 Equipment – Internally Powered
• Type B – IPX1 – Continuous operations
Head Technician:
Alexandre Rodrigues da Silva, Eng.
CREA: Registration No.: 0682082567
3
4
5
6
7
8
9
10
Address: ________________________________________________________________________________
1. Was(ere) the product(s) delivered within the agreed delivery time? Yes No
2. Were the product(s) and the accessory(ies) in accordance with the purchase order? Yes No
GENERAL INDEX
TABLE INDEX
FIGURE INDEX
COMPANY
K. TAKAOKA is a company operating in the hospital equipment manufacturing since 1957, always
cooperating closely with medical professionals. It focuses on the Anesthetics, Intensive Medicine,
Monitoring and Oxygenotherapy areas and is proud to be a market leader, and provides a broad product
line.
One of its priorities is the continued investment in research and development of new ideas and solutions,
standing out for the constant introduction of technological advances and industrial innovations in its
product lines, at the same level as main national and international competitor companies.
Company designs and manufactures with sophisticated equipments most of the components used in its
devices, which explains the strict quality control they are subjected to. Providing a high level support to
all its users is also a concern, by means of its Sales and Technical Assistance Departments.
With distributors among the Brazilian territory and its presence in the international market, K. TAKAOKA
has achieved along the years the loyalty of its clients with its high quality standards and the high efficacy
of its products and services.
Vision:
“To be a global company”.
Mission:
“To be the national leader in the anesthetics and pulmonary ventilators industries, helping to
preserve life, offering high technology and better services to our clients”.
Quality Policy:
“To continuously improve our PRODUCTS, SERVICES and PROCESSES involving our
COLLABORATORS AND PROVIDERS, in order to achieve effectiveness of PRODUCTIVE
PROCESSES, and according to the REGULATORY REQUIREMENTS.”
COLOR SERVO-VENTILADOR 1
Introduction
11. INTRODUCTION
COLOR SERVO-VENTILADOR 2
Safety
22. SAFETY
COLOR SERVO-VENTILADOR 3
Safety
Using the equipment before fully understanding its characteristics and functions
results in risk conditions for the operator, the patient, and the equipment itself.
This product must be periodically checked and shall not be used if any defect is
shown. Preventive inspections and maintenance procedures must be performed
by specialized technicians. If equipment repair is required, the use of original K.
TAKAOKA spare parts is recommended.
COLOR SERVO-VENTILADOR 4
Safety
• The functioning of this equipment is not affected by the use of other equipment
in the proximity, such as high frequency surgery equipment (diathermy),
defibrillators or short wave therapy equipment. Electromagnetic compatibility
tests have been performed by an accredited laboratory.
• This equipment does not send electromagnetic waves which can interfere in the
functioning of other equipment in its proximity. Electromagnetic compatibility tests
have been performed by an accredited laboratory.
•When necessary, open the backup cylinder valve slowly to avoid damage to its
pressure regulating valve.
• The gas from the O2 backup cylinder must only be used to provide a continuous
flow of fresh gases or oxygen directly to the Respiratory System. Don’t use this
gas for other purposes, in order to avoid quick emptying of the cylinder.
• Firstly, check if the on/off switch located in COLOR rear panel is in the off
position.
COLOR SERVO-VENTILADOR 5
Safety
• Use the extensions that are provided with the Servo Ventilator. The use of a
pressure regulating valve of the supply system in the wall is not necessary when
the pressures are within the specified range.
ATTENTION
Oxygen and compressed air feeding pressures must be in the range between 40
and 150 PSI (280 and 1035 kPa).
Connect the air filter that is provided with the device with the Servo Ventilator
compressed air connection and the respective extension. The compressed air
from the source must be free of moisture, oil and impurities.
2.2.3. Important
• The Servo Ventilator must be subjected to an annual revision for a new
calibration, performed by a technician authorized by K. TAKAOKA.
• Check if the Servo Ventilator is correctly configured and if the alarms are
appropriately adjusted before using the equipment.
• When filling up the Nebulizer cup, do not exceed the maximum medication
capacity of the reservoir. Do not keep it connected while not in use.
• Do not press any key using surgical instruments or tools. Use only your
fingertips to press the keys. Pointed or hard objects can damage the keys.
• The COLOR Servo Ventilator is not approved for use with flammable anesthetic
agents.
• Keep the patient under constant observation. Frequently check its pulmonary
expansion and free exhaling.
COLOR SERVO-VENTILADOR 6
Safety
• All parts of the equipment which are in contact with fluids originated from
patients (such as respiratory circuits, flow sensor, exhaling valve diaphragm, etc.)
can be potentially contaminated after use. If classified as semi critical, these
parts should be submitted to a high level of disinfection or sterilization process
before being discarded (at the end of their life cycle) or be discarded as
potentially infected hospital waste.
• All used parts are defibrillation-proof, except for the temperature sensor cable,
which must be removed from the working area (as there may occur burns to the
patient) when a defibrillator is used near the Servo Ventilator.
• All used parts of the Servo Ventilator are made of inert and non-toxic material,
and will not cause irritation or allergy to the patient.
• Any repair required for the Servo Ventilator must only be performed by
specialized technicians, duly authorized by K. TAKAOKA.
• In this manual and the ventilator, the pressure unit is indicated in water
centimeters (cmH2O) and not Pascal (Pa), its corresponding unit in the
International System (IS), where 1 Pa is the same as 0.0102 cmH2O.
In this manual and the ventilator, the respiratory frequency unit is indicated in
respirations per minute (rpm) and not Hertz (Hz), its corresponding unit in the
International System (IS), where 1 Hz is the same as 60 rpm.
• In this manual and the ventilator, the volume unit is indicated in milliliters (ml) or
liters (l) and not in cubic meters (m3), its corresponding unit in the International
System (IS), where 1 E-6 m3 is the same as 1 ml or 0.001 l.
• Upon receipt, check the equipment and component integrity. If there is any
apparent damage to the equipment or its components please contact an
COLOR SERVO-VENTILADOR 7
Safety
• The ventilator should not be covered or kept in places with no air circulation in
order to avoid overheating.
• The ventilator has limited stability and must not be transported or used on
surfaces with an inclination equal to or exceeding 5°, due to risk of toppling over.
COLOR SERVO-VENTILADOR 8
Safety
COLOR SERVO-VENTILADOR 9
Safety
SYMBOLS /
UNIFIED TEXTS PORTUGUESE SPANISH ENGLISH
PONTO DE CONEXÁO P/ CONNECTION POINT
PUNTO DE CONEXIÓN
CONDUTOR NEUTRO EM POR NEUTRL
N EQUIPAMENTO
INSTALADO
PERMANENTE.
PARA CONDUTOR
NEUTRO, EN EQUIPO
INSTALADO PERMANTE
CONDUCTOR, IN
PEFMANENTLY
INSTALLED EQUIPMENT
COLOR SERVO-VENTILADOR 10
Safety
SYMBOLS /
UNIFIED TEXTS PORTUGUESE SPANISH ENGLISH
TREND
TENDÊNCIA TENDENCIA TREND
TREND
COLOR SERVO-VENTILADOR 11
Safety
SYMBOLS /
UNIFIED TEXTS PORTUGUESE SPANISH ENGLISH
ALARME ALARMA
PAUSED ALARM
PAUSADO PAUSADA
CONEXÃO DE
TOMACORRIENTE POWER PLUG
FORÇA
FLUXO DE
FLUJO DE GASES FRESH GAS
F.G.F GASES
FRESCOS FLOW
FRESCOS
LECTURA EM EL
LEITURA NO MEIO READ FROM
MEDIO DE LA
DA ESFERA CENTER OF BALL
ESFERA
CICLO CICLO
MANUAL CYCLE
MANUAL MANUAL
COLOR SERVO-VENTILADOR 12
Safety
SYMBOLS /
UNIFIED TEXTS PORTUGUESE SPANISH ENGLISH
LIGA ON ON
RED DE
REDE DE DADOS NET
COMUNICACIÓN
COLOR SERVO-VENTILADOR 13
Safety
2.4. Abbreviations
Abbreviations Meaning Abbreviations Meaning
COSY Compact respiratory PEEP End Exhalation Positive
System Pressure
FLOW Exhaling Flow PEAK Peak Pressure (airways)
COLOR SERVO-VENTILADOR 14
Safety
2.5. Warnings
Using the equipment before fully understanding its characteristics and functions
results in risk conditions for the operator, the patient and the equipment itself.
ATTENTION
The person responsible for the assembly, operation and maintenance of the Color Servo
Ventilator must be completely familiar with this operation manual.
ATTENTION
The Color Servo Ventilator equipment must not be used stacked on top of another one.
_____________________________________________________________________
ATTENTION
Perform an inspection procedure (check list) before each use of the Color Servo
Ventilator Device. Do not use the equipment if it is not operating perfectly.
COLOR SERVO-VENTILADOR 15
Safety
NOTE:
This Operation Manual is used for all commercially available COLOR Servo Ventilator models.
Other COLOR potentially compatible components are clearly indicated in the text as optional
items.
COLOR SERVO-VENTILADOR 16
Controls and Components
3
3. CONTROLS AND COMPONENTS
COLOR SERVO-VENTILADOR 17
Controls and Components
ATTENTION
Use only components specified as K. Takaoka components. The use of components other
than the ones specified may result in increased electromagnetic emission or decreased
electromagnetic immunity.
6060 Heated
201050003 Humidifier 1
Exhaling Valve
202011152 Diaphragm 1
1
3m O2 extension
202011182 (29/16” threads x
18F Female)
COLOR SERVO-VENTILADOR 18
Controls and Components
Silicone flow
sensor tube with
202011665 400 mm and 1
terminal
K. Takaoka Exclusive
Silicone adult
respiratory circuit – 1
202011669 ICU Mont. Color
3m Air extension (2
female diss
202012068 threads) 1
K. Takaoka Exclusive
K. Takaoka Exclusive
Cell for O2
measurement with
202012085 intermediate and 1
K. Takaoka Exclusive
cable
204010146 1
Operation Manual
COLOR SERVO-VENTILADOR 19
Controls and Components
ATTENTION
Use only components specified as K. Takaoka components. The use of components other
than the ones specified may result in increased electromagnetic emission or decreased
electromagnetic immunity.
1.1.
1.2.
1.3.
1.4.
1.5.
K. Takaoka Exclusive
COLOR SERVO-VENTILADOR 20
Controls and Components
COLOR SERVO-VENTILADOR 21
Controls and Components
Adult/pediatric airway
436010040 adapter 25 unit
The Nebulizer (Figure 1) is an optional COLOR Servo Ventilator component for the
purpose of administering medications to the patient through a gas flow entering the
COLOR SERVO-VENTILADOR 22
Controls and Components
respiratory circuit during the inhaling phase. The Nebulizer reservoir has a medication
capacity of 7 ml.
The monitor in the control area of the Servo Ventilator allows the operator to turn the
nebulization function ON/OFF.
The nebulizer can be mounted next to the endotracheal tube or mask, or on the inhaling
branch.
The nebulizer feeding tube is coupled to the respective NEBULIZER connection tip
located in the Ventilator connection front panel.
NOTE:
The nebulizer can be used by adult, pediatric and neonatal patients, but only in some modes.
The nebulization flow is automatically compensated when determining the tidal volume,
therefore, the exhaled volume is slightly larger than the inhaled volume.
When using of the nebulizer, FiO2 may be influenced.
When filling up the Nebulizer cup, do not exceed the maximum medication capacity of the
reservoir.
ATTENTION
The connection of humidifier filters after the nebulizer when this resource is being
used is not recommended, as these filters may cause a substantial increase in the
respiratory resistance.
Never keep the feeding tube connected to the front panel when the nebulizer
resource is not being used, even if the reservoir is empty.
COLOR SERVO-VENTILADOR 23
Controls and Components
Figure 2 – Monitor LCD Touch Screen / Control Panel / EASY TOUCH Button
Monitor
Color LCD (liquid crystal display) monitor which allows display in 1024 x 728
resolution and touch screen function which allows the view of the following ventilation
curves: pressure x time, flow x time, volume x time, CO2 x time and Auxiliary Pressure
x time, loops of: volume x pressure, flow x volume, in addition to numerical values,
functions, alarms, and navigation display. The monitor has position and inclination
adjustments in addition to image adjustments and configurations.
COLOR SERVO-VENTILADOR 24
Controls and Components
Control Panel
The keys for accessing special functions are located in the control panel, allowing the
access to the parameter values, and highlighting them in the monitor.
NOTE:
The panel keys only highlight parameter values so that they can be adjusted and confirmed
by means of the increase, decrease and confirmation EASY TOUCH button. This safety
system avoids involuntary adjustments.
− Turn button clockwise or counterclockwise until placing the cursor (green circle) over
the parameter to be adjusted.
− Press this button again. The parameter is exhibited in the highlight area of the display
(blue background).
− Turn this button to set the desired numeric value for the parameter. Turning the button
clockwise increases the value, and turning it counterclockwise decreases the value.
Press the button again to confirm the adjustment and to put the new parameter value into
effect. The parameter is removed from the display highlight area.
COLOR SERVO-VENTILADOR 25
Controls and Components
ATTENTION
The two tubes must be coupled in the correct position, according to the
color coding of the ventilator connectors. The tubes have different colors for
a quick identification.
COLOR SERVO-VENTILADOR 26
Controls and Components
COLOR SERVO-VENTILADOR 27
Controls and Components
COLOR SERVO-VENTILADOR 28
Controls and Components
7 - Video
Connector for video cable inlet for the LCD monitor.
NOTE:
The serial cable and external devices are not included with the COLOR Servo Ventilator and
are optional components. For further information about software and hardware
requirements, contact a K. TAKAOKA authorized distributor.
NOTE:
The ventilator can be fed with any voltage between 110 and 220 VAC, as it has an automatic
voltage conversion feature.
COLOR SERVO-VENTILADOR 29
Controls and Components
COLOR SERVO-VENTILADOR 30
Controls and Components
ATTENTION
Oxygen and compressed air feeding pressures must be in the range between 40
and 150 PSI (280 and 1035 kPa).
Connect the air filter that is provided with the device with the Servo Ventilator
compressed air connection and the respective extension. The compressed air form
the source must be free of humidity, oil and impurities.
COLOR SERVO-VENTILADOR 31
Controls and Components
The 6060 Heated Humidifier (Figure 6) was specifically designed for use in respiratory
circuits. It presents electronic temperature control, digital thermometer, pilot lights and
electrical resistance functioning indicator, on/off switch and a transparent chamber with
capacity for 400 ml of sterile water. The gases temperature is measured close to the
patient’s mouth, by means of an electronic temperature sensor.
The COLOR Servo Ventilator main on/off switch also turns on and off the power supply
for the Heated Humidifier.
The Heated Humidifier heats and saturates the inhaling flow that passes through its
chamber with water vapor. The gases then are routed by two corrugated tubes with
drains between the chamber exit and the patient.
Water often condenses inside the patient’s corrugated tube due to the cooling of gases
between the humidifier exit and the "Y" intermediary. The condensed water amount in the
circuit increases with increasing humidifier heating adjustments.
COLOR SERVO-VENTILADOR 32
Controls and Components
1. It is necessary to avoid the condensed water from draining towards the patient, as it
could reach the patient’s lungs. The corrugated tubes must follow an ascending or
horizontal direction up to the patient so that all condensed water in the circuit flows back
to the drains.
2. If the water condensation in the circuit is excessive, the humidifier heating adjustment
must be decreased.
The Heated Humidifier chamber can be easily detached for disinfection, as its cover is
simply threaded to the cup. When assembling the chamber again, make sure the cover
rubber gasket is correctly positioned and well preserved, so that gases leakage does not
occur during ventilation. Close the cover by firmly tightening it.
COLOR SERVO-VENTILADOR 33
Assembly and Preparation
COLOR SERVO-VENTILADOR 34
Assembly and Preparation
The Color is not completely assembled. The LCD monitor, the basic unit, the column
(mobile base) and the articulated arm must be attached, as shown in Figure 7.
COLOR SERVO-VENTILADOR 35
Assembly and Preparation
The mobile base of the COLOR Servo Ventilator has four casters with front brakes to
ensure excellent mobility, hooks for supporting extensions and a handle for easy
transportation.
The Servo Ventilator is attached on the top part of the mobile base with a four screws
set. The Heated Humidifier is attached to the front part of the mobile base by insertion
pins.
Mount the 6060 Heated Humidifier in the column and power it by plugging the power
cable into the outlet located in the back panel of the COLOR Servo Ventilator. The
heated humidifier only operates if the Servo Ventilator is powered by a main power of
110 to 220 VAC.
ATTENTION
The assembly instructions provided must be closely followed. The details in the
following items of this Chapter can also work as guidelines.
COLOR SERVO-VENTILADOR 36
Assembly and Preparation
CONNECTION
1 NEUTRAL PHASE
2 PHASE PHASE
3 GROUND GROUND
Only connect the power cable to a duly grounded plug approved for hospital use, in an
electric wiring compliant with ABNT NBR 13534 standard - "Electrical wiring in health
care facilities - Safety requirements." The female three-pin outlet must be Nema 5-15P
type, where the central round pin is the earth, as shown in Figure 8.
COLOR can be fed by a 110 and 220VAC main power through the power cable
provided with the equipment or by a 12DC 9.0 Ah internal battery.
COLOR SERVO-VENTILADOR 37
Assembly and Preparation
4.2.1. Battery
COLOR Servo Ventilator has an internal battery, which enables its operation in case of
power failure.
• Runtime
The Ventilator internal battery charge duration depends on the mode being used
and the ventilator parameters adjustments.
Example: A fully charged battery can keep the Ventilator operating for
approximately 90 minutes in a medium ventilation condition.
• Recharge
A total battery recharge takes approximately 20 hours, with the Servo Ventilator
turned off and connected to the main power. For longer battery lifetime, keep it at
its maximum charge whenever possible. Constant discharges of the battery
reduce its lifetime.
• Alarm
In the event of lack of power, the Servo Ventilator switches automatically to its
internal battery feeding. When the ventilator is being powered by the internal
battery, the message NO MAIN POWER – OPERATING BY BATTERY will be
shown on the display, the main power LED will be turned off and the power cable
symbol in the lower left corner will be replaced by a battery symbol, indicating
that the ventilator is being powered by the internal battery. To indicate the battery
status, the battery symbol has three states as shown in figure 10.
COLOR SERVO-VENTILADOR 38
Assembly and Preparation
When the battery is above 40% of its capacity, the figure Battery 1 will be shown on the
display. When the battery is between 40% and 20% of its capacity, the figure Battery 2
will be shown on the display, and when the battery capacity is under 20%, the figure
Battery 3 will be shown on the display. When the battery is under 40% of its capacity, an
alarm of medium priority will sound, and when the battery is under 20%, an alarm of
high priority will be shown on the display, indicating that the electrical power must be
restored immediately.
COLOR SERVO-VENTILADOR 39
Assembly and Preparation
ATTENTION
Keep the internal battery always charged. For that, the Servo Ventilator can be left
permanently connected to the main power, even if its on/off switch is in the off
position.
After using the Servo Ventilator for some time without main power feeding, the
battery must be recharged.
If the Servo Ventilator is not used and is kept disconnected from the electric
network for a period of time exceeding 20 (twenty) days, the battery must be fully
recharged.
ATTENTION:
The Heated Humidifier does not operate while the Servo Ventilator is being fed by
its internal battery.
COLOR SERVO-VENTILADOR 40
Assembly and Preparation
COLOR Servo Ventilator features three types of respiratory circuits: adult, pediatric and
neonatal. The neonatal circuit has tubes with smaller diameters than the ones for adults;
therefore, decreasing its complacence for use with low flows.
The respiratory circuits of the COLOR Servo Ventilator include a set of corrugated
tubes, two drains, a “Y” intermediary, and a flow sensor with tubes and lines. The drains
mounted in the inhaling and exhaling branches avoid the accumulation of water in the
circuit.
ADULT CONNECTION
COLOR SERVO-VENTILADOR 41
Assembly and Preparation
ATTENTION
To avoid an accidental disconnection or gas leakage in the respiratory circuit,
attach all connections firmly.
Close the threaded cover of the 6060 Heated Humidifier firmly, making sure its
rubber gasket is present.
The support for corrugated tubes must be positioned close to the Y intermediary
and in a low position in relation to the patient's mouth in order to the condensed
water in the circuit flows back to the drains.
Do not use antistatic or electrically conductive hoses or tubes.
COLOR SERVO-VENTILADOR 42
Assembly and Preparation
COLOR SERVO-VENTILADOR 43
Assembly and Preparation
4.3.1. Drain
The COLOR Servo Ventilator respiratory circuit has 2 (two) drains for collecting
condensed water from inside the corrugated tubes; one drain is for the inhaling branch
and the other for the exhaling branch.
The use of drains is important to avoid issues from the accumulation of water in the
respiratory circuit.
Emptying
To empty the drain, just remove its threaded cup and discharge the accumulated water,
then thread the cup again in the drain body. This operation does not interrupt patient
ventilation, as there is an automatic drain closure system that is activated when the cup
is removed.
Disinfection
The drain is easily disassembled for disinfection by removing its threaded body and
detaching the plunger set (Figure 14).
1. Check if the drain cups are empty. Emptying the drains cups is easily done
without stopping the ventilation, as there is an automatic drain closure system
COLOR SERVO-VENTILADOR 44
Assembly and Preparation
which is activated when the cup is removed. Empty the drains before they are
filled with water.
2. Check if the exhaling valve block is firmly and correctly coupled to the
connection front panel of the Servo Ventilator.
5. Attach the articulated arm to one of the respective lateral supports of the Servo
Ventilator in order to properly accommodate the corrugated tubes. The
articulated arm can be mounted in either of the two sides of the equipment,
depending on the side the patient is.
8. If TGI resource is used, its catheter (optional) must be coupled to the nebulizer
tip located in the connection front panel of the Servo Ventilator.
9. Ensure that the 6060 Humidifier temperature electronic sensor is correctly and
firmly coupled to its respective connection in the “Y” intermediary of the
respiratory circuit.
COLOR SERVO-VENTILADOR 45
Assembly and Preparation
The COLOR Servo Ventilator measures flow, volumes and pressures through a
flow sensor of a “differential pressure” type. The flow sensor must be assembled
in accordance with the procedure described below.
1. Connect the type of flow sensor requested by the screen during the
initialization of the Servo Ventilator, which shall be the most appropriate
type for the patient’s weight informed by the operator. The two available
options for a flow sensor are: adult and pediatric.
2. Couple the appropriate flow sensor for neonatal patients (≤ 6 kg) between
the Y intermediary of the respiratory system and the endotraqueal mask or
tube. For adult and pediatric patients (> 6 kg) the flow sensor must be
coupled to the exhaling branch of the valve block (distal relative to the
patient).
NOTE:
The pediatric flow sensor is appropriate for both neonatal and pediatric patients.
COLOR SERVO-VENTILADOR 46
Assembly and Preparation
NOTE:
The connectors located in the front panel of the Servo Ventilator have a washing flow
system in the flow sensor tubes with the purpose of eliminating the permanence of
secretions and moisture which are the main source of errors during the monitoring of
ventilation parameters. The washing flow is automatically offset when determining current
volume.
ATTENTION:
According to figure 16, NEVER connect the silicone tube to the blue elbow
and the blue tube to the yellow elbow as this would interfere with the sensor
reading.
COLOR SERVO-VENTILADOR 47
Assembly and Preparation
Figure 17 – Connection of the flow sensor tubes on the COLOR Servo Ventilator.
ATTENTION
Check the flow sensor and the tubes color coding for assembling them in the
correct position. The female 15 mm connection (larger) corresponds to the patient
side; the male 15 mm connection (smaller), to the side of the respiratory system
of the ventilation equipment. Both tubes have different colors in order to provide
quick identification.
The flow sensor cleanliness must be frequently checked during the use of the
Servo Ventilator.
COLOR SERVO-VENTILADOR 48
Assembly and Preparation
4.5. O2 sensor
The COLOR Servo Ventilator measures FiO2 by means of an Oxygen Analyzer using an
O2 sensor. Follow the procedures below in order to perform this assembly:
2 3
1 – O2 Sensor
1 2 – Cable
3 – Conector
Figure 19 – O2 Cell/Sensor
COLOR SERVO-VENTILADOR 49
Assembly and Preparation
NOTES:
The request for calibration in 21% or 100% of O2 does not imply in automatic change of
oxygen concentration in the ventilator. Adjust the oxygen concentration in the ventilator
before calibrating and make sure it returns to the desired value.
a) Select the MONITOR key successively, until the screen with the activation
options “O2 Cell Cal 21% or 100%” is shown in the monitoring display.
b) Rotate the EASY TOUCH button until the O2 Cell Cal 21% or 100% item is
placed in the highlight area, and then press the button to select the option.
c) Rotate the EASY TOUCH button to change the status from NCAL (do not
calibrate) to CAL (calibrate).
e) After calibrating, the message “FiO2 xx% Calibrated” will appear in the lower part
of the panel. When the calibration fails, the message “FAILURE Calibrating FiO2
xx%” will appear in the lower part of the panel.
COLOR SERVO-VENTILADOR 50
Assembly and Preparation
ATTENTION
When the calibration fails, check:
• The oxygen cell conditions. The cells have a lifetime of approximately 8,640
hours after being removed from their packing;
• The cable conditions;
• The correct connection of the cable in the oxygen cell and in the ventilator
panel.
a) Place the O2 sensor (in a horizontal position) in contact with the gaseous mixture
to be monitored. For this purpose the intermediary “T” is used, coupled between
the exhaling valve block and the corrugated tube of the inhaling branch (Figure
20).
c) The Servo Ventilator monitoring display then shows the FiO2 concentration in the
gaseous mixture.
ATTENTION
If in doubt about the correct O2 sensor calibration during the use of the ventilator,
repeat the calibration procedure.
COLOR SERVO-VENTILADOR 51
Assembly and Preparation
COLOR SERVO-VENTILADOR 52
Assembly and Preparation
COLOR SERVO-VENTILADOR 53
Assembly and Preparation
a) Fit the female DB9 connector of capnography sensor to male DB9 connector of
adaptor cable (Figure 24) and fit the CO2 inlet (Figure 25) to the other end (male
circular connector), located at the ventilator frontal panel.
COLOR SERVO-VENTILADOR 54
Assembly and Preparation
b) Fit the sensor reader in the airways adaptor until your hear a sound of a “click”
indicating that it is properly fitted (Figure 26).
c) Assemble the CO2 sensor in the respiratory circuit, connecting one airway adapter
end to Y intermediary (Figure 27) and the other end to the mask or endotracheal
tube.
1 – Airways adapter
3- “Y” Intermediary
4- Expiratory branch
trachea.
5- Inspiratory branch
COLOR SERVO-VENTILADOR 55
Assembly and Preparation
ATTENTION
The temperature
sensor should NEVER
be assembled as
indicated in the
picture. That mistake
will result in a false
temperature reading,
due to the room
temperature influence.
Figure 29 – Incorrect Assembly of the temperature sensor in the Y branch of the circuit
COLOR SERVO-VENTILADOR 56
Assembly and Preparation
The temperature sensor should always be removed from the respiratory circuit by the
connector, as shown in figure 30.
ATTENTION
As shown in figure 31, NEVER remove the temperature sensor from the Y branch of the
circuit by pulling the cable, in order to avoid damage to the sensor.
Figure 31 – Incorrect removal of the temperature sensor from the Y branch of the circuit
COLOR SERVO-VENTILADOR 57
Assembly and Preparation
Interconnect O2 inlets and compressed air connections located in COLOR rear panel
with the respective feeding sources of these gases. Use the extensions that are
provided with the Servo Ventilator. The use of a pressure regulating valve for the supply
system in the wall is not necessary should the pressures be within the specified range.
ATTENTION
Compressed air and oxygen feeding sources are adjustable in a range between 40
and 150 PSI (280 and 1035 kPa), through compressed air and O2 internal valves of
the equipment.
COLOR SERVO-VENTILADOR 58
Assembly and Preparation
ATTENTION
Constantly check the filter, noting if it is not saturated with impurities, having a
dark color. Replace it as shown in Figure 34 should that be confirmed.
COLOR SERVO-VENTILADOR 59
Assembly and Preparation
The exhaling valve includes an electromagnetic valve, which defines the inhaling and
exhaling phases of the COLOR Servo Ventilator. During the exhaling phase, the valve is
opened to allow the outflow of exhaling gases.
The exhaling valve block is attached to the connections front panel of the Servo
Ventilator. This allows the respiratory circuit tubes to be directed to either one of the two
sides of the Servo Ventilator, depending on the side where the patient is.
Figure 35 schematically shows the assembly of the exhaling valve block components.
This block has two conic connections in its body for the corrugated tubes of the
respiratory circuit.
COLOR SERVO-VENTILADOR 60
Assembly and Preparation
ATTENTION
Verify the correct assembly of the diaphragm of the exhaling valve (Figure
37), with the perfect groove of the aluminum disc, because the incorrect
assembly ( Figure 38) can cause an inaccurate equipment functioning.
COLOR SERVO-VENTILADOR 61
Assembly and Preparation
ATTENTION
Check periodically the cleanliness and perfect conservation of the exhaling valve
diaphragm. Should a crack or any other irregularity be found in this component, it
must be replaced by a new one. Check periodically the perfect conservation of
the O-ring. The attachment of the block in the connection front panel of the Servo
Ventilator must be firmly performed to avoid gas leakage.
COLOR SERVO-VENTILADOR 62
Adult/Pediatric Ventilation Modes
55. ADULT/PEDIATRIC
VENTILATION MODES
5.1 Introduction
5.2 VCV - Volume Controlled Ventilation
5.3 PCV - Pressure Controlled Ventilation
5.4 PCV/AV® - Pressure Controlled Ventilation with
Assured Volume
COLOR SERVO-VENTILADOR 63
Adult/Pediatric Ventilation Modes
5.1. Introduction
The Table below shows the pediatric/adult ventilation modes available in the COLOR
Servo Ventilator. The modes that require a patient inhaling effort to the breathing trigger
have a protection system against apnea, with automatic change to a backup mode. This
feature results in higher safety for the patient.
COLOR SERVO-VENTILADOR 64
Adult/Pediatric Ventilation Modes
COLOR SERVO-VENTILADOR 65
Adult/Pediatric Ventilation Modes
It provides controlled and assisted cycles. In the controlled cycles, the patient is passive
and the device controls all ventilation. Triggering occurs by time, according to adjusted
breathing frequency.
In assisted cycles, the ventilator responds to the stimuli of the patient respiratory drive
by means of sensibility adjustment by pressure or flow, i.e. the start of every cycle and
the breathing frequency are determined by the patient’s inhaling efforts. An inhaling
pause value can be set where the inhaling flow will remain at zero during the scheduled
time period.
Figure 41 - VCV Mode – Pressure x time and flow x time graphs. The first two cycles correspond to
controlled cycles; the following, to assisted cycles, being the last one with inhaling pause.
Four flow waveform types can be selected: square, descending, ascending, and
sinusoidal.
• Square Wave
The square flow wave has a constant flow characteristic and the flow peak is
equivalent to the calculated flow.
1 Inhaled Volume
2 Inhaling Pause
3 Exhaled Volume
4 Exhaled Time
COLOR SERVO-VENTILADOR 66
Adult/Pediatric Ventilation Modes
• Descending Wave
The descending flow wave reaches a peak flow and decreases linearly.
1 Inhaled Volume
2 Inhaling Pause
3 Exhaled Volume
4 Exhaled Time
• Ascending Wave
The ascending flow wave increases linearly until it reaches the adjusted flow value.
1 Inhaled Volume
2 Inhaling Pause
3 Exhaled Volume
4 Exhaled Time
• Onda Senoidal
1 Inhaled Volume
2 Inhaling Pause
3 Exhaled Volume
4 Exhaled Time
The sinusoidal flow wave starts at zero, increases until the adjusted value is
reached and then returns to zero.
COLOR SERVO-VENTILADOR 67
Adult/Pediatric Ventilation Modes
The Limit Pressure value corresponds to the pressure limit established for delivery of
the adjusted tidal volume. If the “Limited Pressure” is reached, the alarm will sound and
probably the delivered volume will be lower than the adjusted value. If the pressure
exceeds 5 cmH2O of the adjusted limit pressure the equipment will abort the cycle by
opening the exhaling valve as a safety measure.
• Adjustable Parameters
• Base flow: The equipment keeps a continuous flow at exhaling phase set from 4
to 40 l/min. This flow helps in the leakages compensation and in the decrease of
the respiratory circuit resistance.
• TGI: The equipment keeps a continuous flow of 07 l/min at the end of the
exhaling in order to perform the therapy with tracheal gas insufflation.
COLOR SERVO-VENTILADOR 68
Adult/Pediatric Ventilation Modes
• Sigh: Automatically provides a tidal volume 50% higher than the adjusted value,
in each preset cycle number. This resource is activated or deactivated through
the configuration screen of control display.
• Recruitment: This maneuver can perform manual or automatic alveolar
recruitment, where all parameters are adjusted and a continuous pressure on the
airways occurs.
COLOR SERVO-VENTILADOR 69
Adult/Pediatric Ventilation Modes
It provides controlled and assisted cycles. In the controlled cycles, the patient is passive
and the device controls all ventilation. Triggering occurs by time, according to adjusted
breathing frequency. In assisted cycles, the ventilator responds to the stimuli of the
patient respiratory drive by means of sensibility adjustment by pressure or flow, i.e. the
start of every cycle and the breathing frequency are determined by the patient inhaling
efforts.
The Rise Time options allow an inhaling pressurization quickness variation, directly
changing the inhaling flow peak, used for patients experiencing airways obstruction.
Figure 43 – PCV Mode – examples of curves: pressure x time and flow x time
The first cycle is controlled, the second is assisted with pressure sensibility, the third
cycle is assisted with flow sensitivity, with the cycle interrupted due to the pressure
exceeding 5 cmH2O above the set value.
COLOR SERVO-VENTILADOR 70
Adult/Pediatric Ventilation Modes
• Adjustable Parameters
• Base flow: Base flow: The equipment keeps a continuous flow at exhaling phase
set from 4 to 40 l/min. This flow helps in the leakages compensation and in the
decrease of the respiratory circuit resistance.
• Sigh: The equipment automatically provides an inhaling time 50% higher than
the adjusted one, in each cycle number. This resource is activated or deactivated
through the configuration screen of control display.
COLOR SERVO-VENTILADOR 71
Adult/Pediatric Ventilation Modes
COLOR SERVO-VENTILADOR 72
Adult/Pediatric Ventilation Modes
Figure 45 – PCV/AV® Mode: examples of curves: pressure x time and flow x time
Figure 45 – shows pressure and flow curves according to time, in order to assist
PCV/AV® mode understanding.
The start of each inhaling can be operated by the ventilator or by the patient. In the last
case, pressure or flow can trigger the process. At the start of inhaling, the flow reaches
its maximum value and then decreases until the end of the inhaling time. The pressure
is kept constant during each inhaling, and the level of inhaling pressure can be
automatically changed by the ventilator between consecutive cycles to ensure the
delivery of the set tidal volume.
There is a rise time option, which will allow an inhaling pressurization quickness
variation, directly changing the inhaling flow peak, used for patients experiencing
airways obstruction.
COLOR SERVO-VENTILADOR 73
Adult/Pediatric Ventilation Modes
• Adjustable Parameters
COLOR SERVO-VENTILADOR 74
Adult/Pediatric Ventilation Modes
• Base flow: Base flow: The equipment keeps a continuous flow at exhaling phase
set from 4 to 40 l/min. This flow helps in the leakages compensation and in the
decrease of the respiratory circuit resistance.
• Sigh: The equipment automatically provides an inhaling time 50% higher than
the adjusted one, in each cycle number. This resource is activated or deactivated
through the configuration screen of control display.
COLOR SERVO-VENTILADOR 75
Adult/Pediatric Ventilation Modes
Recruitment on/off/manual
Selecting MANUAL, the recruitment will occur just once. Immediately after the
maneuver being concluded in the set time (Recruitment Time), it will return to
controlled mode.
COLOR SERVO-VENTILADOR 76
Adult/Pediatric Ventilation Modes
For spontaneous cycles, the device provides support pressure or CPAP; this phase is
identified by a yellow stripe in the flow x time graph, which corresponds to the
spontaneous interval. In support pressure, cycling can be adjusted, which typically
occurs upon the flow peak 25% fall; there is the possibility to adjust from 5 to 80% of the
initial flow peak for better patient x ventilator synchronism. There is a rise time option,
which will allow an inhaling pressurization quickness variation, directly changing the
inhaling flow peak, used for patients experiencing airways obstruction.
For trigging the spontaneous and assisted cycle (PSV), sensitivity adjustment (trigger)
by pressure or flow is necessary.
Examples of curves: pressure x time and flow x time The first cycle is controlled with a square flow wave;
the second is spontaneous in the CPAP mode. The third indicates a spontaneous cycle with support
pressure, allowing the sensitivity by pressure or flow and the fourth is an assisted cycle triggered int the
synchronism interval with adjustment of inhaling pause.
COLOR SERVO-VENTILADOR 77
Adult/Pediatric Ventilation Modes
• Adjustable Parameters
COLOR SERVO-VENTILADOR 78
Adult/Pediatric Ventilation Modes
In this mode, the ventilator provides controlled, assisted and spontaneous cycles
controlled by pressure.
For the assisted and controlled cycles, the inhaling pressure is kept constant for a
predetermined time (inhaling time), with a flow waveform with a decreasing
characteristic and the tidal volume will depend on the respiratory mechanics of the
patient. This phase is identified in the graph by a red stripe that corresponds to
synchronism interval.
For spontaneous cycles, the device provides support pressure or CPAP; this phase is
identified by a yellow stripe in the flow x time graph, which corresponds to the
spontaneous interval. In support pressure, cycling can be adjusted, which typically
occurs upon the flow peak 25% fall; there is the possibility 5-80% adjustment of the
initial flow peak for better patient x ventilator synchronism. There is a rise time option,
which will allow an inhaling pressurization quickness variation, directly changing the
inhaling flow peak, used for patients experiencing airways obstruction.
For spontaneous cycles, the device provides support pressure; this phase is identified
by a yellow stripe in the flow graph, which corresponds to the spontaneous interval.
For trigging the spontaneous and assisted cycle (PSV), sensitivity adjustment (trigger)
by pressure or flow is necessary.
COLOR SERVO-VENTILADOR 79
Adult/Pediatric Ventilation Modes
Examples of curves: pressure x time and flow x time The first cycle is controlled with decreasing wave
characteristic. The second is spontaneous in the CPAP mode. The third indicates a spontaneous cycle
with support pressure, which indicates the flow peak 25% cycling moment, allowing the sensitivity by
pressure or by flow. The fourth is an assisted cycle, triggered in the synchronism interval.
• Adjustable Parameters
COLOR SERVO-VENTILADOR 80
Adult/Pediatric Ventilation Modes
In this mode, the ventilator provides spontaneous cycles with support pressure assuring
an adjusted minimum minute volume. The start of each cycle and the breathing
frequency are determined by the patient’s inhaling effort. It is necessary to adjust the
sensitivity by pressure or flow. In order to assure a minimum minute volume, the device
changes automatically the support pressure cycle to cycle, using PS value from the
adjusted value, and safety maximum value is the limit pressure.
In support pressure, cycling can be adjusted, which typically occurs upon the flow peak
25% fall; there is the possibility to adjust from 5 to 80% of the initial flow peak for better
patient x ventilator synchronism.
There is a rise time option, which will allow an inhaling pressurization quickness
variation, directly changing the inhaling flow peak, used for patients experiencing
airways obstruction.
Figure 51 - MMV Mode – examples of curves: pressure x time and flow x time.
COLOR SERVO-VENTILADOR 81
Adult/Pediatric Ventilation Modes
In the event of apnea, the ventilator starts to operate in the safety mode with volume
controlled backup mode, where adjustment of breathing frequency and flow is
necessary. The tidal volume will be calculated with the target minute volume divided by
adjusted frequency. A respiratory pause can be set for the Backup.
• Adjustable Parameters
COLOR SERVO-VENTILADOR 82
Adult/Pediatric Ventilation Modes
This mode provides spontaneous cycles with two pressure levels. Spontaneous
breathing with high level of pressure corresponding to Limit Pressure and spontaneous
breathing with low level of pressure corresponding to PEEP.
Set the inhaling time in order to determine the higher time (high level of pressure) and
set the breathing frequency in order to determine the lower time (low level of pressure).
At the lower time, the patient breaths spontaneously upon a continuous pressure
(CPAP), which is determined by the PEEP or it is also possible to set the support
pressure for these breathings of the low level of pressure.
Examples of curves: pressure x time. Note that T1 corresponds to spontaneous breathing of high level of
pressure and T2 corresponds, in the first case, to spontaneous breathing with continuous positive
pressure and, in the second case, with support pressure adjustment.
In support pressure, cycling can be adjusted, which typically occurs upon the flow peak
25% fall; there is the possibility to adjust from 5 to 80% of the initial flow peak for better
patient x ventilator synchronism. Adjustment of flow or pressure sensitivity is necessary.
There is a rise time option, which will allow an inhaling pressurization quickness
variation, directly changing the inhaling flow peak, used for patients experiencing
airways obstruction.
With changes in the higher time adjustment (inhaling time) and in the lower time
(breathing frequency), this mode will have APRV characteristics. Free spontaneous
breathing at a high level of pressure and a low pressure during a short period (relief).
COLOR SERVO-VENTILADOR 83
Adult/Pediatric Ventilation Modes
• Adjustable Parameters
®
Figure 54 – Adjustable parameters for BIPV mode
COLOR SERVO-VENTILADOR 84
Adult/Pediatric Ventilation Modes
In the CPAP mode, the patient breaths spontaneously upon a positive pressure kept
constant in the respiratory circuit. In this mode, the PEEP adjustment, which
corresponds to CPAP (continuous positive pressure) and O2 (FiO2 - Oxygen Inhaled
Faction), is necessary.
On the same screen, there is the PSV mode option, where the patient breaths
spontaneously with the aid of support pressure. Cycling can be adjusted, which typically
occurs upon the flow peak 25% fall; there is the possibility to adjust from 5 to 80% of the
initial flow peak for better patient x ventilator synchronism. Adjustment of flow or
pressure sensitivity is necessary.
There is a rise time option, which will allow an inhaling pressurization quickness
variation, directly changing the inhaling flow peak, used for patients experiencing
airways obstruction.
COLOR SERVO-VENTILADOR 85
Adult/Pediatric Ventilation Modes
ATTENTION
In the event of apnea, the ventilator starts to operate in the safety mode with controlled
pressure backup mode, provided the backup is activate. Control of Controlled Pressure,
breathing frequency and Inhaling Time is necessary.
• Adjustable Parameters
COLOR SERVO-VENTILADOR 86
Adult/Pediatric Ventilation Modes
The cycling (transition from inhaling to exhaling) can occur based on two criteria, i.e., by
adjusting a percentage from 5 to 80% of the peak flow or reaching the preset cycling
time.
There is a rise time option, which will allow an inhaling pressurization quickness
variation, directly changing the inhaling flow peak. Sensitivity adjustment (Trigger) by
pressure or flow is necessary.
ATTENTION
In the event of apnea, the ventilator starts to operate in the safety mode with controlled
pressure backup mode, provided the backup is activated. Adjustment of Controlled
Pressure, breathing frequency and Inhaling Time is possible.
• Adjustable Parameters
COLOR SERVO-VENTILADOR 87
Adult/Pediatric Ventilation Modes
The device provides spontaneous cycles where the patient breaths with the aid of
support pressure, where a target tidal volume is adjusted. The equipment changes
automatically the support pressure to achieve the adjusted tidal volume.
Cycling is can be adjusted, which typically occurs upon the flow peak 25% fall; there is
the possibility to adjust from 5 to 80% of the initial flow peak for better patient x
ventilator synchronism. Adjustment of flow or pressure sensitivity is necessary.
There is a rise time option, which will allow an inspiratory pressurization quickness
variation, directly changing the peak inspiratory flow, used for patients experiencing
airways obstruction.
ATTENTION
In the event of apnea, the ventilator starts to operate in the safety mode with volume
controlled backup mode, where adjustment of tidal volume, breathing frequency and flow
and inhaling pause is necessary.
®
Figure 58 – PSV/AV Mode: examples of curves: pressure x time and flow x time.
Figure 58 shows pressure and flow curves according to time, in order to assist in the PSV/AV®
mode understanding. The patient breaths spontaneously with pressure support cycles. The Servo
Ventilator automatically varies the support pressure value among the consecutive cycles to
COLOR SERVO-VENTILADOR 88
Adult/Pediatric Ventilation Modes
ensure the delivery of adjusted tidal volume by the operator (P1≠P2≠P3). The spontaneous cycles
are triggered by the patient through sensitivity by pressure ou flow. If the patient stops breathing
for a time perior longer than apnea time, the Sevo Ventilator automatically starts to send VCV
cycles.
• Adjustable Parameters
COLOR SERVO-VENTILADOR 89
Adult/Pediatric Ventilation Modes
Cycling can be adjusted, which typically occurs upon the flow peak 25% fall; there is
the possibility to adjust from 5 to 80% of the initial flow peak for better patient x
ventilator synchronism. Adjustment of flow or pressure sensitivity is necessary.
There is a rise time option, which will allow an inspiratory pressurization quickness
variation, directly changing the peak inspiratory flow, used for patients experiencing
airways obstruction.
ATTENTION
In the event of apnea, the ventilator starts to operate in the safety mode with assured
volume controlled pressure backup mode, where adjustment of Controlled Pressure,
breathing frequency and Inhaling Time is necessary. The target volume will correspond
to VSV adjusted volume.
®
Figure 60 - VSV Mode – examples of curves: pressure x time and flow x time.
COLOR SERVO-VENTILADOR 90
Adult/Pediatric Ventilation Modes
Figure 60 shows pressure and flow curves according to time, in order to assist in the VSV® mode
understanding. The patient breaths spontaneously with pressure support cycles. The Servo
Ventilator automatically varies the support pressure value among the consecutive cycles to
ensure the delivery of adjusted tidal volume by the operator (PS1≠PS2≠PS3). The spontaneous
cycles are triggered by the patient through sensitivity by pressure or flow. If the patient stops
breathing for a time period longer than apnea time, the Sevo Ventilator automatically starts to
send PCV/AV cycles, where the support pressure level is also automatically varied by the Servo
Ventilator (PC1≠PC2≠PC3).
• Adjustable Parameters
COLOR SERVO-VENTILADOR 91
Neonatal Ventilation Modes
6 6. NEONATAL VENTILATION
MODES
6.1 Introduction
6.2 PLV – Pressure-Limited Ventilation
6.3 SIMV/P® – Synchronized Intermittent Mandatory
Ventilation with Pressure control
6.4 BIPV® – Bi-level Spontaneous Pressure Ventilation
6.5 CPAP/PS - Continuous Positive Airway Pressure /
Pressure Support Ventilation
6.6 Non-Invasive Ventilation
COLOR SERVO-VENTILADOR 92
Neonatal Ventilation Modes
6.1. Introduction
The Table below shows the neonatal ventilation modes available in the COLOR Servo
Ventilator.
COLOR SERVO-VENTILADOR 93
Neonatal Ventilation Modes
COLOR SERVO-VENTILADOR 94
Neonatal Ventilation Modes
In this mode, the ventilator provides continuous flow, time-cycled with pressure limit.
Exclusive mode for Neonatal ventilation with weight adjustment < or = 6 Kg.
ATTENTION
The flow sensor shall be coupled to the proximal side of the respiratory circuit.
• Adjustable Parameters
COLOR SERVO-VENTILADOR 95
Neonatal Ventilation Modes
For the assisted and controlled cycles, the inhaling pressure is kept constant for a
predetermined time (inhaling time), with a flow waveform with a decreasing
characteristic and the tidal volume will depend on the respiratory mechanics of the
patient. This phase is identified in the graph by a red stripe that corresponds to
synchronism interval.
For spontaneous cycles, the device provides support pressure or CPAP; this phase is
identified by a yellow stripe in the flow x time graph, which corresponds to the
spontaneous interval. In support pressure, cycling can be adjusted, which typically
occurs upon the flow peak 25% fall; there is the possibility to adjust from 5 to 80% of the
initial flow peak for better patient x ventilator synchronism. There is a rise time option,
which will allow an inspiratory pressurization quickness variation, directly changing the
peak inspiratory flow, used for patients experiencing airways obstruction.
For spontaneous cycles, the device provides support pressure; this phase is identified
by a yellow stripe in the flow graph which corresponds to the spontaneous interval.
For trigging the spontaneous and assisted cycle (PSV), sensitivity adjustment (trigger)
by pressure or flow is necessary.
COLOR SERVO-VENTILADOR 96
Neonatal Ventilation Modes
Examples of curves: pressure x time and flow x time. The first cycle is controlled with decreasing wave
characteristic. The second is spontaneous in the CPAP mode. The third indicates a spontaneous cycle
with support pressure, which indicates the flow peak 25% cycling moment, allowing the sensitivity by
pressure or by flow. The fourth cycle is assisted, triggered in the synchronism interval.
• Adjustable Parameters
COLOR SERVO-VENTILADOR 97
Neonatal Ventilation Modes
NOTE:
With changes in the higher time adjustment (inhaling time) and on the lower time
(respiratory frequency), this mode will have APRV characteristics. Free spontaneous
breathing at a high level of pressure and a low pressure during a short period (relief).
• Adjustable Parameters
®
Figure 67 - Adjustable parameters for BIPV - Neonatal mode
COLOR SERVO-VENTILADOR 98
Neonatal Ventilation Modes
• Adjustable Parameters
COLOR SERVO-VENTILADOR 99
Neonatal Ventilation Modes
• Adjustable Parameters
77. OPERATION
2 - MONITOR key
Quick access key for pages selection of the ventilation monitor and alarm
adjustments, which can be shifted and confirmed through the “easy touch” button.
corresponding button on the touch screen. It will also disable the touch screen
keys.
4 – FREEZE key
Key that freezes the graph being presented on the monitor, thus allowing for a
more detailed analysis of the curves. When pressed again, this key unfreezes the
graph. When this key is pressed, a symbol (blue circle) is shown on the left lower
side of the monitor, indicating the freezing status of the graph. However, the data
of the alphanumeric window continue to be presented in real time, and the actual
values of ventilation will continue to be executed. This function can be activated
by pressing the corresponding icon on the touch screen.
NOTE:
5 - VENTILATOR key
Quick access keys for pages selection where the parameters adjustments
options are, which can be shifted and confirmed through the “easy touch” button.
It can be activated by pressing the touch screen.
6 - SAVE LOOP
This key freezes and records a desired flow x volume loop on the graphs screen,
allowing the verification of patient’s performance improvement or reduction in the
next cycles.
7 - MONITOR key
Key that performs the ventilation mode selection, including the options: VCV,
PCV, PCV/AV®, SIMV/V, SIMV/P®, MMV, BIPV®, CPAP, PSV, PSV/AV®, VSV®
and PLV (neonatal). This key must be repeatedly pressed until the desired mode
is highlighted on the monitor, then press the ENTER (easy touch) button to
confirm the selection. The mode change can be performed by touch screen,
pressing the corresponding field mode on LCD screen.
10 – STAND BY key
When this key is pressed for 2 seconds, the ventilator operates on the Stand By
mode, and the respective visual indication on the monitor is displayed. The
ventilator remains inactive; however, adjustments to the ventilation parameters
can be performed. The operator must press the STAND BY key again in order to
cancel this condition,. This mode can be used during the patient preparation or
another special event. This function can be activated by pressing the
corresponding icon on the LCD touch screen.
13 - GRAPH key
This key switches the graphs displayed on the monitor. The available graph
options are Volume X Pressure, Volume X Flow, which will be described below in this
manual.
When the Servo Ventilator is turned on, the monitor displays a system and Servo
Ventilator software initialization screen (this screen is similar to the one in a regular
computer). This information is used for the equipment maintenance.
Then, the Servo Ventilator performs a procedure, verifying the presence of compressed
air and oxygen, adjustment of flow sensors and pressure sensor reference, auto-test of
the internal valves, in the following order: O2 valve; air valve; both valves together.
During this auto-test procedure, the behavior of the electronic components integrating
the system is also evaluated.
Eventual failures in the system (pneumatic, electronic, gas network) found during the
auto-test are indicated through alarm messages and audible sign. The auto-test
procedure takes approximately 20 seconds.
ATTENTION:
Do not use the COLOR Servo Ventilator should the auto-test show any irregularities.
Solve the problem before using the device.
The equipment runs the auto-test, a screen is displayed requesting the operator to
inform the initial adjustments:
Volume / Weight
4 ml/Kg up to 12 ml/Kg
According to entered weight, the equipment will determine in which age-group the initial
modes will be automatically enabled, as follows:
Volume Controlled
Adult/Pediatric
Ventilation
Pressure Controlled
Adult/Pediatric
Ventilation
Synchronized Intermittent
Mandatory Ventilation with
Adult/Pediatric
Volume control
Synchronized Intermittent
Mandatory Ventilation with
Adult/Pediatric
Pressure control
Limited-Pressure
Neonatal
Ventilation
Synchronized Intermittent
Mandatory Ventilation with
Pediatric/Neonatal
Pressure control
The operator can choose the initial mode by placing the EASY TOUCH button on the
desired mode and pressing it to confirm.
2 Mode
3 Parameters
With the previous adjustments performed, their confirmation will be done by pressing
the EASY TOUCH Button on OK. The proper sensor must be coupled according to the
request displayed on the screen.
Figure 74 - Ways to assemble the neonatal, pediatric, and adult respiratory circuits
Press “Enter” for confirming, then the ventilation and monitoring will automatically start.
ATTENTION
The automatic initial adjustment of the ventilation parameters should not be considered by
the operator as the ideal and definitive adjustment for the patient. Before starting the
ventilation, assure the ideal adjustment of each parameter is performed.
Check the correct position of the external flow sensor, using the distal mode sensor for
adult and pediatric patients and proximal mode for neonatal patients. According to the
weight selected in the table above, the circuit assembling is displayed on the next screen
according to the type of patient.
After the definition of the respiratory circuit, the main control screen will be displayed with
the adjustable parameters, monitored parameters and graphs.
Figure 77 – Easy touch button, turning clockwise, counterclockwise and press ENTER.
Turn the easy touch button until the parameter to be chosen; it will be marked with a
green circle. Then press the easy touch button; the parameter value will be
highlighted in blue. Adjust by turning clockwise the easy touch in order to increase
and counterclockwise to decrease the value and confirm the adjustment by pressing
again the easy touch button.
7.7. Messages
The following messages can appear in the lower part of the screen, as Figure 79
shows:
• SPONTANEOUS
That Message momentarily appears in the lower side of the monitor every time the
patient starts a spontaneous breathing cycle with the aid of the support pressure. An
indication of whether the sensitivity was due to FLOW or PRESSURE also appears.
• WINDOW
That message appears in the lower side of the monitor during the period in which the
Servo Ventilator waits for the patient’s inhalation effort for the triggering of a new
breathing.
• AUTO PEEP
Message indicating that there was an air imprisonment larger than the adjusted Auto
peep value.
• STAND BY
Message indicating that Servo Ventilator is on standby mode, which is activated by
pressing the STAND BY key.
• SIGH
Message indicating that Servo Ventilator is in the programmed sigh feature.
• ALVEOLAR RECRUITMENT
That message indicates that Servo Ventilator is performing a recruitment maneuver.
GRAPH UNIT
Flow x time l/min x s
Pressure x Time cmH2O x s
Volume x Time ml x s
Pressure x Flow cmH2O x l/min
Volume x Pressure ml x cmH2O
CO2 x Time mmHg x s
Auxiliary Pressure x Time cmH2O x s
Table 13 – Graph / Unit Table
Press the quick access key in the control panel in order to switch between the
screens with the available graph types.
In the graphs presented, in the shape of a loop, (Flow x Volume and Volume x
Pressure), the recording of a cycle can be performed at any determined moment, for
later comparison with the current cycles. For this purpose, press the key (SAVA
LOOP) of frontal panel at the desired moment:
1. Clears the screen
2. Displays the curve recorded on the memory
3. Displays three cycles in real time, and returns to item 1 above.
Image freezing
The key (FREEZE) freezes the graph being presented on the monitor, thus allowing
a more detailed analysis of the curves.
When pressed again, this key unfreezes the graph. When this key is pressed, the
symbol in red will appear in the left side of the monitor, indicating the graph freezing
status. However, the data of the monitoring screen continue being displayed in real
time.
The Servo Ventilator is provided with a serial port for synchronization with other Color
Servo Ventilator in the ILV function through the proper cable (this is an optional item)
and updating of the equipment software.
7.9. Database
Checking all the values of monitored parameters can be performed in this screen, in
which each value is stored every 30 seconds during 48 hours, i.e., it has a storage
capacity of 5760 records.
When entering the numeric trend (DATABASE), a table is shown where the columns
correspond to the parameters:
1 Date
2 Hour
3 Mode
4 Tidal Volume
5 Frequency
6 Pressure
7 Plateau Pressure
8 Medical Pressure
9 Auxiliary Pressure
10 PEEP
11 Inhaling Flow
12 Exhaling Flow
13 Work
14 Static Compliance
15 Dynamic Compliance
16 Resistance
17 FIO2
18 EtCO2
19 Alarm 1
20 Alarm 2
21 Alarm 3
22 Record
The quick access key “GRAPH” must be pressed in order to show this screen, until
the table is presented.
Figure 80 – Database.
The following procedures must be followed in order to access the database values:
8.1 Introduction
8.2 Alarms Priority Levels
8.3 Alarms Priority Level Audible Differentiation
8.4 Alarms Related to the Display
8.1. Introduction
The Color Ventilator has a complete audiovisual alarm system, which provides high
safety during ventilation. In the event of an alarm condition, the Ventilator provides
Audible Signs and written Messages on the screen in order to assist the operator to
promptly identify the event that has generated the alarm.
NOTE:
The ventilator alarm system complies with ISO 9703-1, ISO 9703-2, and ISO 9703-3
standards.
The intensity and intermittence of the alarm indication are directly related to the priority
level of the occurrence, as specified below.
The COLOR Servo Ventilator Alarm System can be activated by two different control
sources: the ventilator and the ventilation monitor incorporated to the equipment. The
alarms with highest priority, active (which interrupts the ventilation cycle), and
considered safety alarms are related to the ventilator; and the secondary alarms,
considered as warning conditions, are related to the ventilation monitor. The alarms
shown on the ventilator display follow this pattern: Medium Priority Alarms are indicated
with Yellow background and Black letters; while the High Priority alarms are indicated
with Red background color and White letters.
1
Alarms with Inverted Relation and PEEP=Limit Pressure do not present audible signals
At the inferior part of the screen as the picture 81 can appear the alarms that
are informed bellow:
• Fail Comm
Inform a condition of without sent or a fail at the command sent to the main board of the
equipment. In this case, please contact the K. Takaoka Ind. e Com. Ltda technical
assistance attached at the manual.
• Disconnection
It indicates a disconnection in the respiratory circuit or in the flow sensor tubes.
• Apnea
It indicates that the patient has spontaneously stopped breathing, thus being necessary
an immediate action by the operator. Depending on the selected mode, the Ventilator
automatically starts to operate in the backup mode.
• High Pressure
1. It indicates that the current inhaling pressure has exceeded the adjusted value of the
limit inhaling pressure control by 5 cmH2O, due to some abnormal operating condition.
This is an active alarm, which automatically interrupts the inhalation to avoid pressure
excess.
2. It indicates that the inhaling pressure has exceeded the adjusted value of the upper
limit of the ventilation monitor alarm.
• Low Pressure
It indicates that the inhaling pressure is not reaching the lower limit established in the
ventilation monitor alarm menu.
• High PEEP
It indicates that the final exhaling pressure has exceeded the upper limit established in
the ventilation monitor alarm menu.
• Low PEEP
It indicates that the final exhaling pressure has not reached the lower limit established in
the ventilation monitor alarm menu.
• High Frequency
It indicates that the measured breathing frequency has exceeded the upper limit
established in the ventilation monitor alarm menu.
• Low Frequency
It indicates that the measured breathing frequency has not reached the lower limit
established in the ventilation monitor alarm menu.
• High FiO2
It indicates that the measured FiO2 has exceeded the upper limit established in the
ventilation monitor alarm menu.
• Low FiO2
It indicates that the measured FiO2 has not reached the lower limit established in the
ventilation monitor alarm menu.
• Low Battery
It indicates that the internal battery has low charge, requesting an urgent battery
recharge. The visual indication of this alarm is through the BATTERY visual indicator.
This alarm has medium and high priorities.
NOTE
The Ventilator must be immediately connected to the main power when this message is
displayed, in order to recharge the internal battery.
• EXHALING OBSTRUCTION
It indicates that there was an obstruction in the exhaling valve. This may have been
caused by a fold or obstruction (secretions) in the exhaling branch, or even by problems
in the exhaling valve diaphragm (assembling, integrity, and cleaning).
• CLEANING - A process that removes dirtness and organic material from any
surface or object. The cleaning is performed by mechanical friction, immersion,
cleaning and ultrasound machines. It is the most important step of
decontamination; each item must be washed before undergoing any disinfection
or sterilization process. No object should be sterilized if it has organic material on
it (oil, fat, blood, etc.). The cleaning must always be performed with soap and
water. When the immersion method is used, use preferentially enzymatic
detergent. This detergent has specific activity on the organic material; it degrades
and dissolves the material in a few minutes. The objects must remain submerged
during 5 minutes.
NOTE
Hydrogen peroxide (oxygenated water) is a sterilization process that takes place at a
maximum temperature of 45ºC, the materials that cannot be autoclaved can be sterilized
with peroxide, except that ones derived from cellulose.
ATTENTION
Turn the Color Servo Ventilator off before cleaning it. Do not submerge the
ventilator in any liquids.
This cleaning, disinfection, and sterilization procedure for the parts that are in contact
with the patient must be performed before the first use and in the following reutilizations.
1. Use a clean and soft cloth, wet with water and soap (detergent) for cleaning the
external parts of the ventilator. Care must be taken in order to prevent cleaning product
residue accumulation in the device connections. Perform drying with a clean, dry, and
soft cloth after cleaning.
2. Use a clean, soft, and lint free cloth to clean the screen of the Servo Ventilator. Do
not use paper towels or rough cloths to avoid scratching the screen surface.
4. The components of the respiratory circuit, if not used with bacterial filter, must be
disassembled after being used by each patient for disinfection or sterilization, including:
corrugated tubes (plastic or silicone), heated humidifier chamber, drains, nebulizer,
exhaling valve, exhaling valve diaphragm, flow sensor, flow sensor tubes and
intermediaries. Use glutaraldehyde, hydrogen peroxide, or ethylene oxide. The
corrugated plastic tubes, the nebulizer, the intermediaries, and the white flow sensor
cap can undergo thermal-disinfection; however, only in sensitive cycle (chemical
thermal-disinfection), i.e., they support low temperatures up to 60ºC. The other
materials can be sterilized in an autoclave, with a maximum temperature of up to136ºC.
Should the respiratory circuit components be used with a bacterial filter, establish a
daily routine (once a day) for their disinfection or sterilization. The Servo Ventilator
circuit components used in ICU must be reprocessed for each patient, whether filter is
used or not.
5. The silicone parts (tubes and tracheas) have silicone intrinsic characteristics of non
deformation (up to 150ºC, and indefinite lifetime), at room temperature, maintaining the
silicone application and properties. It is suggested the storage conditions to be in a
clean and organized area, free from perforating or contaminant materials, preferentially
following the cleaning instructions, good manufacturing organization practices, non
exposure to sunlight, avoiding exposure to temperature above 40ºC or in areas near
heat-related equipment (such as stoves and sterilizers).
NOTE
After start using, it is suggested the tracheas/tubes to have at most a lifetime of up to 50
sterilizations in a vapor sterilization process (autoclave) in 20-minute cycles with 1
bar/15PSI and 121ºC/250ºF.
ATTENTION
Perform visual and functional inspection for the components after the
cleaning/sterilization procedure in order to detect signals of wearing on them.
Dry tubes and/or with cracks and irregular fittings indicate the need for
replacement.
We should pay special attention to the cleaning of the flow sensor; therefore, the
accumulation of liquid or solid substances may interfere with the values
measured and displayed by the ventilation monitor. Thus, a periodicity for
cleaning must be established by the physician or according to the patient status,
or at least every 2 hours.
When using ethylene oxide, the instructions provided by the sterilization
equipment manufacturer must be followed in order to determine the indicated
temperatures and aeration times.
9.3. FLOWCHART
The processing will be performed following a sequence of steps, illustrated in the
FLOWCHART below.
CLEAN
RINSE
DRY
DEPENDING ON THE
ARTICLE DESTINATION
PACK LIQUID GASEOUS PLASMA MECHANICAL IMMERSE FOR THE IMMERSE THE
(GLUTARALDEHYDE) (ETHYLENE OXIDE) (HYDROGEN FRICTION APPROPRIATED TIME ARTICLE
PEROXIDE) AND TEMPERATURE
STERILIZE FULLY IMMERSE PACK PACK PACK DRY ASSEPTICALLY FILL THE
TUBULATIONS
DRY PACK
ASSEPTICALLY
PACK IN A STORE
STERILIZED BOTTLE
STORE
9.4. GUIDELINES
Methods recommended for processing TAKAOKA Mechanical Ventilation Equipment
Components.
NOTE
* Galvanic cell for FiO2 measurement must be cleaned with a cloth damped in water and
soap. It should not be submerged in a solution.
** Silicone tracheas can be autoclaved and subjected to resistant thermal disinfection.
ATTENTION
The Luer lock cap must NOT be autoclaved. Sterilization with glutaraldehyde is otherwise
recommended.
NOTE
It is recommended the monitor to be always disinfected with alcohol at 70º.
10
10. MAINTENANCE
1.12.
1.13.
10.1 - Preventive Maintenance
10.2 - Emergency Actions
• Check if tubes, sensors, cables and other COLOR’s components are not
damaged, worn or cracked at least once every month. If there is any damage,
replace the component. Don’t use damaged components.
• The exhaling valve diaphragm must be carefully inspected and its integrity
checked at least once a month.
• The exhaling valve diaphragm must be replaced at least once every 6 (six)
months, and whenever necessary.
NOTE
The attachment of the exhaling valve block in the connection front panel of the ventilator
must be firmly performed to avoid gas leakage. Check periodically the perfect
conservation of the O-ring.
• The flow sensor tubes must be replaced at least once every 2 (two) months, and
whenever necessary.
• Check periodically the perfect conservation of the rubber gasket of the Heated
Humidifier cover, in order to achieve a perfect seal when closing the chamber. If
needed, replace this gasket with a new one.
• Check the conditions and periodically replace corrugated tubes of the respiratory
circuit, as they are usual wear components.
• If the maximum inhaling pressure does not reach the expected value, initially
check:
• If the tidal volume does not reach the expected value, initially check:
if there is no leakage in the respiratory circuit;
if there is no leakage in the humidifier cover;
if all connections are tight;
if the pressure control has not a too low regulation;
if the O2 pressures and compressed air supplies are not too low;
if the exhaling valve assembly is correctly assembled, with a clean
diaphragm in perfect conditions;
if the inhaling flow control has not a too low regulation.
• If you cannot normally feed the COLOR with the main power, initially check:
if there is 110 to 220 VAC voltage in the main power outlet;
if the COLOR fuse is not burned.
• The compressed air and oxygen filters must be opened and cleaned two to three
times a year, depending on the Servo Ventilator use rate and the purity of the
medicinal gases. The draining of the filter condensation is automatic.
• Use only the sensors, cables and tubes specified by K. TAKAOKA for the
equipment.
• Do not use the COLOR Servo Ventilator if the self-test shows any irregularity.
Provide a solution for the problem displayed through an authorized K. TAKAOKA
distributor.
• The Servo Ventilator must be submitted to an annual revision for a new
calibration, performed by a technician authorized by K. TAKAOKA.
• For longer battery lifetime, keep the internal battery at its maximum
charge whenever possible. Constant battery discharges reduce its
operating time.
• The internal battery is sealed and does not require maintenance. Should
the battery show any functioning problem, contact the Authorized Technical
Assistance.
• If the battery is not being charged normally, check the voltage in the main
power outlet.
ATTENTION
Use original K. TAKAOKA spare parts only. Using non-original parts may put the
patient’s safety in danger.
Do not perform any internal service to Color and do not open its box. Contact an
authorized Technical Assistance to perform any internal maintenance or a
periodic revision.
NOTE
Additional documentation can be provided in order to performing preventive/corrective
maintenance if previously agreed.
• Contact the manufacturer (Technical Assistance) and inform the device condition.
Do not perform any test or investigation without the presence of an authorized K.
Takaoka technician.
• Record the manufacturer, model and serial number of all devices involved in the
adverse event. Record this information in the patient case report form and/or in an
incident form. If the device is disposable or has disposable components, also record
the classification number and all numbers of all disposable components. It is
important to keep the device and every component involved in the adverse event.
• Do not clean the device, neither subject it to a chemical or physical process or fix
it. These actions may affect its performance and safe use.
• Identify the device, indicating that it has been involved in an adverse event, the
date of event, and name of the person who tagged the device. Indicate in the tag
that the device must not be used, cleaned, fixed or destroyed without consent from
an authority, such as the risk manager. Should the adverse event involve more than
one device, all involved devices should also be tagged and stored.
• Keep the packaging of all disposable components involved in the event and keep it
with the device. The disposable packaging typically includes not only a recorded
device number but also the batch number. Some specifications included in the
package can also be useful for the investigation.
• Before disconnecting the device from the main power or removing the batteries,
check if the device’s memory will not be lost. Many devices have computerized
memories which can be lost if the batteries are removed or if the device is
disconnected from the main power. Experts can use this memory to specifically
determine when and how the adverse events occurred. Read the instructions
manual or contact your clinical engineer in order to determine which devices have
computerized memories and how they must be controlled following an event.
• Store the device and its components in a safe place to avoid subsequent
damages. This will prevent the device to be put into operation again; protected
rooms and devices may need to be used regardless of a previous incident.
11
11. TECHNICAL INFORMATION
9
1.14.
1.15.
11.1 Packaging / Environmental Use Conditions
11.2 Technical Specifications
11.3 Ventilation Monitor Specifications
11.4 Measurement Methods
• Packaging
Temperature from 10ºC to +70ºC, atmospheric pressure from 500 to 1060 hPa
and relative humidity from 10% to 100% (non condensing).
Ventilation Parameters
Nebulizer:
Drug capacity 7 ml
Nebulizing flow Between 5 and 10 l/min, synchronized
with inhalation start.
Table 17 - Nebulizer
TGI
Insufflating flow Between 5 and 10 l/min, synchronized
with exhalation end.
Table 18 - TGI
Special Characteristics
• Reserve modes (backup) for apnea cases;
• LCD display with touch screen function;
• Complete alarm system;
• Curve types: square, decelerated, accelerated and sinusoidal;
• Differential pressure type flow sensor;
• Automatic compensation of respiratory circuit compliance;
• Electronic on/off key;
• Manual, electronic and synchronized inhaling;
• Pressure regulating valves;
• Anti-asphyxia safety valves with resistance between 0.3 and 0.5 (cmH2O/l/min);
• Overpressure electronic valve fixed in the adjusted limit pressure;
• Washing flow in the flow sensor tubes between 2 and 10 (ml/min).
Gases Feeding
NOTE:
The COLOR Servo Ventilator is equipped with internal pressure regulating valves and it
can be connected directly to the outlets of hospital gas supply set up according to current
ISO 7396 standard. The installation of an external regulating valve is not recommended.
The inlet gas connections have internal unidirectional valves which ensure a limitation to
“reverse gas flow”.
CAUTION:
Use of a high pressure inlet to the patient side puts the patient at risk.
Electrical Features
NOTE
The rechargeable internal batteries can only be replaced by authorized K. TAKAOKA
Technical Service.
The hospital electric wiring must be appropriately grounded and according to ABNT NBR
13534 – “Electrical facilities in health care facilities – Safety requirements” standard.
Noncompliance with this recommendation may result in damage to the patient, operator
or equipment, and will void the ventilator warranty.
Others
Conical connections in conformity with ISO 5356 standard
Corrugated tubes in conformity with ISO 5367 standard
Table 21 – Tubes and connections
Height 1445 mm
Width 530 mm
Depth 500 mm
Weight 29 kgf
Table 22 – Dimensions and weight
± (2 cmH2O + 4% of
Plateau inhaling pressure
-30 to 120 1 measure)
(cmH2O)
± (2 cmH2O + 4% of
Exhaling end pressure (PEEP) measure)
-30 to 120 1
(cmH2O)
± (2 cmH2O + 4%of
Average inhaling pressure
-30 to 120 1 measure)
(cmH2O)
Static pulmonary compliance
(Cstat) 1.00 to 100,00 0.01 ± 10%
(mL/cmH2O)
Dynamic pulmonary compliance
(Cdyn)
1.00 to 100.00 0.01 ± 10%
(mL/cmH2O)
Airway resistance (Raw)
1.0 to 100.0 0.1 ± 10%
(cmH2O/l/min)
Inhalation time (s) 0.01 to 20 0.01 ± 0.2 s
I:E Ratio I6:1to 1:10 0.1 ± 30%
FiO2 (%) 14 to 100 1 ± 5%
EtCO2 (optional) (mmHg) 0 to 100 1 ± 5%
Intrinsic exhaling end pressure
(PEEP) ± (2 cmH2O + 4% of
-30 a 120 1
measure)
(ipeep) (cmH2O)
± (2 cmH2O + 4%of
Occlusion pressure (P0.1)
-30 to 120 1 measure)
(cmH2O)
Inhaling effort (WOB) (J/I) 10.00 to 50.00 0.01 ± 10%
Tobin Index (iT) 0 to 200 1 ± 5%
Table 23 - Measured Parameters (Range / resolution / accuracy)
NOTE
The COLOR Servo Ventilator monitors only exhaled tidal volumes above 100ml, due to its
flow sensor sensitivity.
The FiO2 monitor is in accordance with ISO 7767standard.
NOTE
The ventilator alarms system is in compliance with ISO 9703-1, ISO 9703-2, and ISO 9703-3
standards.
NOTE
All calculations are based on flow and distal pressure measurements. The times are taken
from the flow curves, flow integral volumes, pressure curve pressures and compliances,
resistance and work of pressure and flow set.
1.16. Volumes
it
Volumeinh = ∫ Flowinh dt Where “it” is the inhaling time
0
et
Volumeexh = ∫ Flowexh dt Where “et” is the exhaling time
0
8
∑ Volume [n]
n =1
inh
Volumeavg_inh =
8
1.17.
1.18. Frequency
∑ Frequency[n]
n =1
Frequencyavg =
8
It is taken from the time measurement elapsed between two inhalation cycles.
1.22.
1.23. Inhaling and Exhaling Time
∑ Time [n]
n =1
inh
Timeinh_avg =
8
8
∑ Time [n]
n =1
inh
Timeexh_avg =
8
Pressuremax → During the inhaling cycle the largest pressure value read is searched.
Pressureplateau → During the inhaling cycle and after obtaining the largest pressure
value, a pressure stability in a value below Pmax that lasts more than 100 milliseconds
is searched.
PEEP is measured during the exhaling cycle, searching the point where the flow is
smaller than 3 l/min and there is a temporal stability above 200 milliseconds.
∑ Pressure [n]
n =1
Pressureavg =
N
Measure from the start of the inhaling cycle (n=1) until the start of the next inhaling
cycle (N).
1.25. Resistance
Pressureinhaling _ end
Resistance =
Flowmaximum_exhaling
1.26. Complacence
Volumeexp
Complacencestatic =
Pressure plateau − PEEP
Volumeexp
Complacence dinamic =
Pressure max − PEEP
NOTE
The static compliance value is shown in the alphanumeric window, however according to
the measurement method of the plateau pressure previously described, this can be the
same as the maximum pressure, and in this case the dynamic complacency value is
shown.
Static Complacence
Dinamic Complacence
Effort inh = ∫ Pressure_dV Where iv is the initial volume and fv is the final volume
iv
ATTENTION
Excess of noise in the flow measurement line may affect volume calculations.
The Ventilator automatically offsets thermal deviations which result in ± 5 l/min;
above this value, turn the ventilator off and then on.
IEC 61000-3-2
Emissions due to In conformity
voltage fluctuation/
scintillation
IEC 61000-3-3
Immunity Essay ABNT NBR IEC 60601 Conformity Level Electromagnetic Environment -
Essay Level Guidelines
It is recommended that no portable and
mobile RF communication equipment
are used next to any part of the COLOR
Servo Ventilator Model Ventilation
system, including cables, within a
separation distance smaller than
recommended, calculated from the
equation applicable to the transmitter
frequency.
3 .5
d=
E1 P
Radiated RF 3 V/m 3 V/m
IEC 61000-4-3 80 kHz up to 2.5 MHz 7
d=
E1 P
Where P is the transmitter maximum
nominal output power in watts (W),
according to the transmitter
manufacturer, and d is the
recommended separation distance in
meters (m).
NOTE 1 In 80 MHZ and 800 MHZ the highest frequency range is applied.
NOTE 2 These guidelines may not apply to all situations. The electromagnetic propagation is affected by structures, objects and
people absorption and reflection.
a
The Field intensities established by fixed transmitters, such as base radio stations, telephone (cell phone/wireless) and ground
mobile radios, ham radios, AM/FM radio transmissions and TV transmissions cannot be theoretically foreseen with accuracy.
A local electromagnetic inspection is recommended in order to evaluate the electromagnetic environment due to fixed RF
transmitters. Should the field intensity measurement in the local where the COLOR Servo Ventilator Model Ventilation System is
used exceed the above conformity levels, the COLOR Servo Ventilator Ventilation System model should be checked to note
whether its operation is normal. Should an abnormal performance be noted, additional procedures may be required, such as
rearrangement or moving the COLOR Servo Ventilator Model Ventilation System.
b
Above the 150 kHz frequency range up to 80MHz, the field intensity should be less than [V1] V/m.
Table 27 – Manufacturer guidelines and statements – Electromagnetic Immunity (RF frequency
radio)
100 36 36 23
For transmitters with nominal maximum output power not listed above, the recommended separation distance d in meters (m) can
be determined by the equation applicable to the transmitter frequency, where P is the transmitter nominal maximum output power in
watts (W) according to the transmitter manufacturer.
NOTE 1 With 80 MHZ and 800 MHZ the highest frequency range is applied.
NOTE 2 These guidelines may not apply to all situations. The electromagnetic propagation is affected by structures, objects and
people absorption and reflection.
Table 28 - Recommended separation distances among mobile RF communication devices and the
Color Servo Ventilator Model Ventilation System.
13 WARRANTY TERMS
13.
1.28.
The inappropriate use of the equipment and/or its use in non-compliance with the
instructions described in this manual, the use of a voltage different from here specified,
and the use of parts and/or components not homologated by K TAKAOKA IND. E
COM. LTDA. result in warranty void.
Damages caused by accidents or natural agents are not part of the warranty, as well as
batteries, fuses and filters.
The items provided with the equipment are listed below, as well as some optional items
and their respective warranty coverage against “manufacturing defects”.
The COLOR Servo Ventilator life cycle is estimated by 5 years, and it may vary
depending on how it is used and the adequate preventive maintenance.
Name:
Hospital:
Address:
Phone: District:
Defect description:
Name:
Hospital:
Address:
Phone: District:
Defect description: