You are on page 1of 5

BIOMEDICAL ENGINEERING-

APPLICATIONS, BASIS & COMMUNICATIONS 121

NUMERICAL INVESTIGATION OF SPAN LENGTHS


AFFECTING MECHANICAL RESPONSES IN ANTERIOR
RESIN-BONDED FIXED PARTIAL DENTURE
CHUN-LI LIN1,2, JIAN-JIA CHEN2
1
Department of Mechanical Engineering, 2Graduate Institute of Medical Mechatronic,
Chang Gung University, Tao-yuan, Taiwan
Biomed. Eng. Appl. Basis Commun. 2005.17:121-125. Downloaded from www.worldscientific.com

ABSTRACT
by HOWARD UNIVERSITY on 01/26/15. For personal use only.

The geometric shape and mechanical structure of RBFPD compared to conventional FPD are
relatively complex and unstable. The low retention rate between the retainer and abutment affects the
prosthesis/abutment interface de-bonding, and closely relates to the design of the prosthesis and
varied occlusion status. This study used reverse engineering (RE) and computer-assisted design
(CAD) to construct two solid models of anterior RBFPD with different span lengths. After mesh
generation, biomechanical interactions of span length in RB prosthesis with two loading conditions
(axial and lateral) were performed by FE analysis. The simulated results indicated that lateral
occlusal force increased significantly 2-3 times maximum stress than that of axial occlusal force. For
different span lengths simulation, the analysis on static movement finds that longer pontic would lead
to high stress to the prosthesis. Thus, the length of the pontic has significant effect on the overall
intensity of the prosthesis under static clenching loading, and the effect of lateral loading exceeds
that of axial loading.

Biomed Eng Appl Basis Comm, 2005(June); 17: 121-125.


Keywords: finite element analysis, RBFPD, biomechanics, span.

1. INTRODUCTION restoration of the space was necessary and important.


Taking affordability and long-term usage into account,
dentists usually would adopt traditional fixed partial
Lost of anterior incisor would change the dentures (FPDs) to treatment [1-2]. However, dentin
occlusion and mastication states and influences the might be injured owning to the little thickness of
aesthetics. Under normal occlusal loading, the adjacent enamel during tooth preparation [2]. The pulp chamber
teeth would move forward and backward to the space, may be also threatened in the process of abutment
and opposite teeth would extrude inducing preparation for young patients. Based on the above
malocclusion and further damage the occlusion reasons and the increasing recognition on the dentition
balance, resulting in overloading partial dentition and aesthetics, In 1973, Rochette [3] proposed the design
consequently periodontal disease. Hence, the conception of resin-bonded fixed partial denture
(RBFPD) to ameliorate the weakness of traditional
Received: Oct 28, 2004; Accepted: April 30, 2005 appliance. However, the geometrical shape and
Correspondence: Chun-Li Lin, PhD mechanical structure of RBFPD was relatively
Department of Mechanical Engineering, complex and unstable when compared with traditional
Graduate Institute of Medical Mechatronic, FPD. The RBFPD fixed on the abutment only with
Chang Gung University, Tao-yuan, Taiwan. connector and mesial/distal retainers, lower retention
E-mail: cllin@mail.cgu.edu.tw rate of retainer and abutment usually induces interface
de-bonding and causes treatment to fail. Recent studies

11
122 Vol. 17 No. 3 June 2005

indicated that lower retention rate not only relatives 2. MATERIALS AND METHODS
with interface bonding technique of retainer/abutment
but influences by the structure design of denture,
preparation of abutment and occlusal loading 2.1 Finite Element Model Generation
conditions. Tooth plastic model of upper incisors (left and
Recently, many studies have focused on the right) and left upper lateral incisor (B9-500D, Dental
improvement of bonding material and augmentation of study model, Nissin dental product Inc., Japan) were
the bonding strength for retainer/abutment interface. selected as the investigated samples. Point clouds of
However, the clinical success rate could not effectively the teeth surfaces were scanned and saved as closed
rise yet, mainly owing to the prosthesis design, curve surface data using four axis laser scanning
abutment preparation and occlusal loading conditions. system (LSH800, 3D-family, Taiwan). The geometric
Literatures indicated that the added slot and groove maxillary bone was obtained from computer
designs could efficiently transfer the occlusion force tomography (CT) images and the solid model was
Biomed. Eng. Appl. Basis Commun. 2005.17:121-125. Downloaded from www.worldscientific.com

and increase the retention rate of retainer/abutment reconstructed by stacking a series of slices after
interface when prepared abutment teeth [5-10]. On the detecting contours of cortical and cancellous bone in
structure design of retainer framework, length of span image processing system (Mimics, Materialise,
was the most important factor to dominate the Belgium). Then the three teeth and maxillary bony
biomechanical aspects of the RBFPD design. In solid models were assembled in CAD system
by HOWARD UNIVERSITY on 01/26/15. For personal use only.

addition, different occlusions were mentioned also (Freeform, SensAble Technologies, USA), the root of
affecting RB prostheses clinical retention [4, 11-12]. left incisor (below CEJ) needed to cut and smooth to
Hansson and Bergstrom observed para-functional adopt the geometric shape of prosthesis pontic.
activity and bruxism as significant stress factors in Prosthesis design parameters, such as connectors, rest
premature debonding in RB prostheses [13]. However, seats and grooves were also generated in the CAD
there have been insufficient studies to illustrate the system according to standard detail dimensions in
relationship between length of span design and clinic. On the basis of the contours of the abutment
occlusal forces in RBFPD. This is because traditional teeth, enamel, pulp and periodontal ligaments (PDL)
experiments and clinical observations cannot provide volumes (0.3mm thickness) were scaled to create
enough information to understand the internal according to literature data [17, 24, 25]. In order to
biomechanics of complicated RB prosthesis systems understand the influences of different span lengths
[14-16]. Consequently, computer simulations based on under different loading states within the RBFPD
the FE method must be used as a complementary tool system. Two solid models with 8.5mm and 7mm span
for evaluating the detailed mechanical responses in RB lengths must be constructed according to previously
prosthesis applications [16]. The aim of this procedures and could be exported into finite element
investigation was to evaluate the biomechanical package (ANSYS, Swanson Analysis Inc., Houston,
interactions of span length in RB prosthesis with two PA. USA) in standard IGES format. Due to irregularity
loading conditions (axial and lateral) using three- and complexity of the geometric couture of
dimensional FE analysis. retainer/abutment system, tetrahetral quadratic element
(solid 92) was adopted to mesh the solid models by
smart mesh approach providing in ANSYS. Linear and
elastic enamel, dentin, pulp, prosthesis, PDL and
alveolar bone (cortical and cancellous) material
properties were assigned according to the volume
definition from literatures (see Table 1).

Table 1. Material properties assigned to dental tissues, alveolar bone (cortical


and cancellous), prosthesis, and epoxy resin

Young's Modules(MPa) Poisson's Ratio Reference


Gold Alloy 99960 0.33 27,28
Ceramic 60000 0.19 27,28
Enamel 84100 0.3 27,28
Dentin 18300 0.31 27,28
Cortical Bone 14700 0.3 27,28
Cancelloud Bone 490 0.3 27,28
PDL 50 0.45 26

12
BIOMEDICAL ENGINEERING-
APPLICATIONS, BASIS & COMMUNICATIONS 123

2.2 Parameter Analysis 3. RESULTS


After the FE models generation, analyses would
be carried out on the different denture structure design The three dimensional of two RBFPD FE models
(length of span) of anterior RBFPD with subjecting to were shown in figure 1 and consisted of 44789, 42831
different occlusal loadings (axial and lateral). Static nodes and 179156, 171323 elements for 7 and 8 mm
analysis would simulate the loading state when anterior span length models, respectively. Due to the different
region was biting tightly. Therefore, the exterior nodes material characteristics of gold alloy retainer, ceramics
at the mesial and distal surfaces of the alveolar bone coping and remaining biological teeth tissue, von
were fixed in all directions as the boundary conditions Mises and maximum principle stresses should be
for all models. Axial and lateral loadings (50N) from chosen for metal frames, ceramics and abutment teeth
lingual side to the buccal side were applied of their brittle properties, respectively. When whole
respectively on the third of crown height. The system was subjected to axial loading, the results
maximum principal and von Mises stresses in
Biomed. Eng. Appl. Basis Commun. 2005.17:121-125. Downloaded from www.worldscientific.com

indicated that maximal stress value occurred on the


biological tissue (remaining tooth) and prosthesis, ceramic structure part of the system and stress
respectively, were computed and recorded for each concentration were found at the bottom area of left
model as a result of their different material connector/abutment interface (Fig. 2) regardless of
characteristics, i.e., brittle tooth structure and metal whether different span length designs. For
prosthesis. There were six RBFPD FE simulation
by HOWARD UNIVERSITY on 01/26/15. For personal use only.

investigating the influence of length of pontic span, not


models (two different span length models with two only the maximal von Mises stress of metal frame but
loading conditions). also the maximal principle stress of ceramics and
abutment part of 8.5mm design were higher than those
of 7mm design under axial occlusion loading. Under

(a) (b)
(a) (b)

(c) (d) (c) (d)


Fig. 2. (a) and (b) Locations of maximum stress on
Fig. 1. FE models of anterior RBFPDs (a) 8.5 mm, denture under axial loading for 7 and 8.5 mm. (c)
(b) 7 mm, (c) and (d) inner structures of and (d) Locations of maximum stress on denture
retainer/abutment system. under lateral loading for 7 and 8.5 mm.

13
124 Vol. 17 No. 3 June 2005

lateral loading, the position of maximal stress and The simulated results showed that when the
stress tendency were same as those under axial loading position was on the center of bridge influence
loading, and the stress value on denture system was of horizontal loading from lingual side to the buccal
higher than that under axial loading. The stress value side was higher than the vertical loading from the
under lateral loading was about 2-3 times as that under occlusal side to the bone side on the denture system
axial loading (Fig. 3), which indicated that lateral strength. When the center of bridge was subjected to
loading did more harm to the denture system. occlusal loading, vertical force would make the
abutment move downwards due to the flexibility of the
PDL. After a distance of downwards movement, the
4. DISCUSSION tooth roots would contact the bottom alveolar bone
owing to the compress of PDL and stress was
Compared with conventional FPD, an RBFPD has transferred to the alveolar bone which would
the ability to preserve dentin substance, reduce the effectively bear the stress of retainer/abutment bonding
Biomed. Eng. Appl. Basis Commun. 2005.17:121-125. Downloaded from www.worldscientific.com

probability of pulp damage, and accommodate system and decrease the stress value. When the
dentition aesthetics. However, several clinical reports RBFPD system was subjected to horizontal loading,
have documented a low survival rate of 75% for although the tooth cervix would also contact the lateral
RBFPDs over periods of 2 to 11 years [18-20]. alveolar bone due to the compress of PDL, loading on
Continued adhesion between the retainer and the the lingual 2/3 crown height of the pontic would
by HOWARD UNIVERSITY on 01/26/15. For personal use only.

abutment tooth has been identified as the most induce bending moment effect on the lateral abutment
important factor influencing long-term RBFPD which would increase the stress values.
success. Consequently, many studies have focused on When the RBFPD was under different loading
adhesive systems and new prosthesis materials to direction, results indicated that 8.5mm length of span
improve the bonded strength of the retainer/abutment design would induce higher stress distribution on the
interface. Other studies have found that retainer/abutment bonding system. This was because
retainer/abutment retention is influenced by the the loading acted on the centre of pontic, and the arm
adhesive system and prosthesis material and dominated of force from loading position to the lateral abutment
by the framework design (retainer thickness or shape) fixation of 8.5mm design was longer comparing with
and occlusal contact [19]. Therefore, the RBFPD 7mm design. Higher bending moment on the RBFPD
biomechanical aspects regarding to prosthesis design system caused the retainer/abutment interface
concepts must be investigated further. However, it was receiving higher stress value.
difficult to understand and get the real biomechanical Based on the limitations of this numerical
aspects and information inside the tooth by traditional investigation, some of the assumptions such as the
experiment and clinical examining methods. This study PDL material properties and loading conditions might
constructed computer 3D FE model of anterior RBFPD determine the accuracy of the mechanical responses
using revise engineering and CAD technology and and stress states obtained in this study. Linear elastic
performed the analysis to understand the influences of (homogeneous and isotropic) PDL characterization
different span lengths of RBFPD under axial and was used due to numerical convergence considerations
lateral occlusion loading. and a larger variation for the PDL physical properties
in the literatures [21-23]. The Young's modulus of the
PDL (70.3 MPa) assumed in this study was considered
a static normal occlusal loading condition. In addition,
another limitation was the resin cement lute not
included in our mapping FE models because of we
only focused on investigating the biomechanical
performance of RBFPD with different span lengths and
their interactions. All simulations were performed
under perfect bonded conditions between the retainer
and the abutment teeth as a result of bonding strength
were assumed reaching an acceptable level. Therefore,
the results from limitations of the modeling procedure
thus give only a general insight into the tendencies of
biomechanical aspects under average conditions,
without attempting to simulate individual clinical
situations. The mechanical responses obtained from all
Fig. 3.Analysis results of different span length simulations are a first approximation and need to be
under axial and lateral loadings. validated with clinical trials.

14
BIOMEDICAL ENGINEERING-
APPLICATIONS, BASIS & COMMUNICATIONS 125

REFERENCES Element Approach. Comput Methods Programs


Biomed 2003; 72: 55-64.
17. Eshleman JR, Janus CE and Jones CR: Tooth
1. Shillingburg HT, Jacobi R, Brackette SE: preparation designs for resin-bonded fixed partial
Fundamentals of tooth preparations. Quintessence, dentures related to enamel thickness. J Prosthet
Chicago. 37-39, 1987. Dent 1988; 60: 18-22.
2. Shillingburg HT, Hobos, Whitsett LD: 18. Creugers NH and Kayser AF: An analysis of
Fundamentals of fixed prosthodontics. multiple failures of resin-bonded bridges. J Dent
Quintessence, Chicago. 27-30, 1978. 1992; 20: 348-351.
3. Rochette AL: Attachment of a splint to enamel of 19. Priest G: An eleven-year reevaluation of resin-
lower anterior teeth. J Prosthet Dent 1973; 30: 418- bonded fixed partial dentures. Int J Periodontics
423. Restorative Dent 1995; 15: 239-247.
4. Botelho M: Resin-bonded prostheses: The current 20. Dimashjieh MR and Al-Shammery AR: Long-term
Biomed. Eng. Appl. Basis Commun. 2005.17:121-125. Downloaded from www.worldscientific.com

state of development. Quintessence Int 1999; 30: survival of sleeve-designed fixed partial dentures: a
525-534. clinical report. J Prosthet Dent 2000; 84: 591-593.
5. McLaughlin G: Composite bonding of etched 21. Provatidis CG: A comparative FEM-study of tooth
metal anterior splint. Compend Contin Educ Dent mobility using isotropic and anisotropic models of
1981; 2: 279-283. the periodontal ligament. Finite Element Method.
by HOWARD UNIVERSITY on 01/26/15. For personal use only.

6. Meiers JC and Meetz HKL: Design modifications Med Eng Phys 2000; 22: 359-370.
for etched metal, acid etched, resin-bonded 22. Toms SR, Dakin GJ, Lemons JE and Eberhardt
retainers. Gen Dent 1985; 33: 41-44. AW: Quasi-linear viscoelastic behavior of the
7. Ziada HM, Benington IC and Orr JF: Photoelastic human periodontal ligament. J Biomech 2002;
stress analysis in resin bonded bridge design. Eur J 35:1411-1415.
Prosthodont Rest Dent 1995; 3: 217-222. 23. Yoshida N, Koga Y, Peng CL, Tanaka E and
8. El Salam Shakal MA, Pfeiffer P and Hilgers RD: Kobayashi K: In vivo measurement of the elastic
Effect of tooth preparation design on bond strength modulus of the human periodontal ligament. Med
of resin-bonded prostheses: A pilot study. J Eng Phys 2001; 23: 567-572.
Prosthet Dent 1997; 77: 243-249. 24. Eahleman JR, Janus CE and r Jones: Tooth
9. Saad AA, Claffey N, Byrne D and Hussey D: preparation designs for resin-bonded fixed partial
Effect of groove placement on retention/resistance dentured to enamel thickness. J Prosthet Dent
of maxillary anterior resin-bonded retainer. J 1988; 60; 1-12.
Prosthet Dent 1995; 74: 133-139. 25. Farah, JW, Craig, RG, and Meroueh KA: Finite
10. Barrack G and Bretz WA: A long-term prospective element analysis of three- and four-unit bridges. J.
study of etched-cast restoration. Int J Prosthodont Oral Rehabil 1989; 16:603-611, 1989.
1993; 6: 428-434. 26. Rees JS and Jacobsen PH: Elastic modulus of the
11. Creugers NH, Snoek PA, Van t Hof MA and periodontal ligament. Biomaterials 1997; 18: 995-
Kayser AF: Clinical Performance of Resin-bonded 999.
Bridge: A 5 Year Prospective Study. II. The 27. Fenner DN, Robinson PB and Cheung P M-Y: 3-D
Influence of Patient-Dependent Variables. J Oral finite element analysis of thermal shock in a
Rehabil 1989; 16: 521-527. premolar with a composite resin MOD restoration.
12. Crispin BJ: A Longitudinal Clinical Study of Med Eng Phys 1998; 20: 269-275.
Bonded Fixed Partial Dentures: The First Five 28. Yang HS, Lang LA, Albert D, Guckes and Felton
Years. J Prosthet Dent 1991; 66: 336-342. DA: The effect of thermal change on various
13. Hannson O and Bergstrom B: A Longitudinal dowel-and-core restorative materials. J Prosthet
Study of Resin-bonded Prostheses. J Prosthet Dent Dent 2001; 86: 74-80.
1996; 76: 132-139.
14. Ko CC, Chu CS, Chung KH and Lee MC: Effect of
Posts on Dentin Stress Distribution in Pulpless
Teeth. J Prosthet Dent 1992; 68: 421-427.
15. Lin CL, Chang CH and Ko CC: Multifactorial
Analysis of an MOD Restored Human Premolar
Using Auto-mesh Finite Element Approach. J Oral
Rehabil 2001; 28: 576-585.
16. Lin CL, Lee HE, Wang CH and Chang KH:
Interfacial Stress Analysis between Posterior
Resin-Bonded Bridge and Abutments by Finite

15

You might also like