Professional Documents
Culture Documents
Sey-
mour, McGraw-Hill, New York, 1971. Used with permission.)
Typical techniques for molecular weight determination are given in Table 3.4. The
most popular techniques will be considered briefly.
All classic molecular weight determination methods require the polymer to be in
solution. To minimize polymer–polymer interactions, solutions equal to and less than 1
g of polymer to 100 mL of solution are utilized. To further minimize solute interactions,
extrapolation of the measurements to infinite dilution is normally practiced.
When the exponent a in the Mark-Houwink equation is equal to 1, the average
molecular weight obtained by viscosity measurements (Mv) is equal to Mw. However,
since typical values of a are 0.5 to 0.8, the value Mw is usually greater than Mv. Since
viscometry does not yield absolute values of M as is the case with other techniques, one
must plot [] against known values of M and determine the constants K and a in the
Mark-Houwink equation. Some of these values are available in the Polymer Handbook
(Burrell, 1974), and simple comparative effluent times or melt indices are often sufficient
for comparative purposes and quality control where K and a are known.
For polydisperse polymer samples, molecular weight values determined from colli-
gative properties (3.6–3.8), light scattering photometry (3.10), and the appropriate data
treatment of ultracentrifugation (3.11) are referred to as “absolute molecular weights,”
while those determined from gel permeation chromatography (GPC) (3.5) and viscometry
(3.13) are referred to as relative molecular weights. An absolute molecular weight is one
that can be determined experimentally and where the molecular weight can be related,
through basic equations, to the parameter(s) measured. GPC and viscometry require cali-
bration employing polymers of known molecular weight determined from an absolute
molecular weight technique.
Mm
4 10 32 10 25 10 675 10 1.132 10
8 6 6 6 9
106000 106000
Mm 10.6 103 lb / mol
i N i M3i
(iii) Mz
i N i M2i
Mz
4 10000 8 2000 1 5000 3 15000
3 3 3 3
Mz
4 10 64 10 125 10 10125 10 1.4314 10
12 9 9 9 13
Mz 12.64 103 lb / mol Mm 10.6 103 lb / mol Mn 4 103 lb / mol
13.8 MOLECULAR MASS DETERMINATION METHODS
Osmometry, viscometry, light
scattering and sedimentation methods are
commonly used methods for determination of
molar mass of a polymer. Some methods are
given here in detail.
viscosity of solution
0 = viscosity of solvent
t = time of flow of solution
t 0 time of flow of solvent
= density of solution
0 = density of solvent
13.8.2 Sedimentation
Let us consider a particle of polymer in solution which moves under the
influence of gravitational field. The falling particle having mass ms experiences a
centrifugal force which is equal to 2 rm and buoyancy force. The net force acting on
the particle can be written as
846 Modern Physical Chemistry
mv 2
2 rms
r
m r
2
2 rms
r
mr 2 2
2 rms
r
mr 2 2 rms (13.105)
Where, ω is the angular velocity of the rotor in radius per second, r is the
distance from centre of rotation of particle.
mr 2 2 r v
So, net force will be equal to frictional force
dr
f mr 2 2 r v
dt
dr v
f mr 2 2 r m
dt m
dr
f mr 2 2 r mv
dt
dr
f 2 mr 1 v
dt
dr mr 1 v
2
dt f
dr
dt
N Am 1 v
2r NA f
S
M 1 v
NA f
SN A f
M (13.106)
1 v
According to Stoke’s law frictional coefficient of a spherical particle is
f 6 rs
By putting value of f in equation (13.106), we get
SN A 6 rs
M (13.107)
1 v
According to Stoke’s Einstein equation, we know
k BT
D
6 rs
k BT
6 rs
D
Where, kB is the Boltzmann constant, D is the diffusion coefficient and T is
the absolute temperature.
dr
S dt
r 2
1 dr
Sdt
2 r
By integrating above equation within limits
848 Modern Physical Chemistry
t r
1 1
S dt r dr
0
2
r0
1
S t 0
r
ln r r
2 0
1
St ln r ln r0 (13.108)
2
ln r
St 0
r
2
ln r
1
S
t 2
r0
By rearranging equation (13.108), we get
St 2 ln r ln r0
ln r St 2 ln r0 (13.109)
sedimentation coefficient.
13.8.3 Osmometry
Vapour phase and membrane osmometry are two principal types of osmometry
used so far to measure number average molecular weight.
Polymer Chemistry 849
Fig. 3.13 A sketch of a vapor pressure osmometer. Fig. 13.16 Working of vapour phase osmometry Fig. 13.17 Plot of ΔV/C versus C
(Courtesy of Hewlett-Packard Company.)
Problem
Insulin, a hormone that regulates carbohydrate metabolism in the blood, was isolated from a pig. A 0.200-g sample of insulin was
dissolved in 25.0 mL of water, and at 30 oC the osmotic pressure of the solution was found to be 26.1 torr. What is the molecular weight
of the insulin?
Solution: M = 5800
NOTE : This is an “apparent” molecular weight since it is for a single concentration and is not extrapolated to zero concentration.
850 Modern Physical Chemistry
1
RT A 2C A 3C2 A 4C3 ....
C Mn
Fig. 13.18 Functioning of membrane osmometers (a) when both chambers have
solvent, and (b) when one chamber has solvent while other has solution.
1
A 2C (13.111)
RTC Mn
Following results are drawn from equation (13.111)
(i) Equation (13.98) is a linear equation.
(ii) Plot of п/RTC versus C gives an intercept equal to 1/ Mn and slope equal to A2
(Fig. 13.18). Thus inverse of intercept gives value of Mn .
(iii) Value of A2 is measure of polymer-solvent interactions. High value of slope
indicates good polymer-solvent interactions, so that solvent is considered of
good quality.
852 Modern Physical Chemistry
The turbidity which is the total scattering integral over all angles, is often
referred by the Rayleigh ratio R(θ) [ R(θ) R(θ) R(θ)solvent ] which relates the scattered
intensity at angle θ to the incident beam intensity. For particles small compared to ,
the Debye equation is obtained as
Kc Hc 1
2A 2 c ... (13.113)
R(θ) τ M
2
32π3n2 dn
H
3N A λ4 dc
Equation (13.113) is correct only for vertically polarized incident light and for
optically isotropic particles (in which the refractive index is same in all directions are
said to be isotropic e.g. water, glass). The use of unpolarized light requires that τ
multiplied by (1+cos2θ).
For system with particles size less than /20, any observation angle may be
selected since the scattering function is spherically symmetrical. However for
experimental reasons (to avoid interference of primary beam) one usually selects
values of θ not very different from 90º.
Fig. 13.21 (A) Essential parts of scattering instrument and (B) scattering of light by
different parts of large molecule
For polymer molecules, whose coil diameter is larger than /20, the scattering
function is no longer spherically symmetrical. Instead one finds a diminution in
intensity of scattered light due to interference, which is different at different
observation angles. This results from the fact that for larger particles, the different
scattering centers within a particle are so far from one another that the resulting
854 Modern Physical Chemistry
scattered rays have a path difference of the order of 0 to /20. Since the separate
scattering centers within a polymer molecule are activated by one and the same
wavelength of primary light, the scattered radiation resulting from the centers is
coherent, and therefore capable of interference. The degree of diminution resulting
from the interference depends upon the path difference which in turn depends on the
angle θ. The scattering of light by larger particles is shown in Fig. 13.21 (b).
Therefore the intensity of scattered light differs depending on the observation angle
θ. The larger is angle θ, the greater is the reduction in scattering intensity R(θ) . The
factor by which the scattering intensity R(θ) (for a perpendicularly polarized incident
beam diminished by interference) is diminished by interference at an angle θ is called
the scattering function P(θ) [where R(θ) equals the scattered light in the immediate
vicinity of the primary beam, i.e. with θ→0]. Therefore, for the scattering intensity
R(θ) reduced to standard conditions and measured at the angle θ, one can write
Kc 1 1
2A 2 c ... (13.114)
R(θ) P(θ) M
1 16π2 2 θ
1 2
s sin2
P(θ) 3λ 2
Here s 2 is gyration radius which is the average distance from axis of rotation
at which all the masses could be concentrated to produce the same amount of
moment of inertia that the actual distribution of mass possess.
4π θ
s2 sin
λ 2
Kc 1 16π2 2 θ
2A 2 c ... 1 2
s sin2 (13.115)
R(θ) M 3λ 2
Polymer Chemistry 855
Equation (13.115) has two variables c and θ, thus extrapolations can be made
by varying one and kept the other constant and vice versa (Zimm plot Fig. 13.22).
These extrapolations will help to get useful results.
Kc 1 16π2 2 θ
1 2
s sin2
R(θ) c 0 M 3λ 2
The plot of Kc/ R(θ) at c→0 versus sin2 (θ/2) would yield a straight line with
slope and intercept as
1 16π 2 2
Slope s
M 3λ2
1
Intercept
M
Slope 1 16π2 2
2
s (M)
Intercept c 0 M 3λ
Slope 3λ2
s2 2
Intercept c 0 16π
In this way ratio of slope to intercept when c→0 gives value of gyration
radius of polymer coil.
(II) Zimm plot (Fig 13.22) is the representation of light scattering data which can
be used to conduct the following extrapolations of equation (13.115) on a single
graph. In Zimm plot Kc/ R(θ) is plotted against sin2 (θ/2) +kc. k is a constant which is
chosen to give a good display to graph.
Kc 1
(13.116)
R
(θ) c 0,θ0o M
Kc 1 16π2 2 θ
1 2
s sin2 (13.117)
R(θ) c 0 M 3λ 2
Kc 1
2A 2 c (13.118)
R(θ) θ0o M
Fig. 13.22 Zimm plot in which Kc/ R(θ) is plotted versus sin2 (θ/2) +kc
Table 13.1 Applicable weight range and average molecular weights determined from
different methods
Applicable weight
Method Average
range
End group analysis Mn upto 20,000
Membrane osmometry Mn 20,000 to 20,000,00
Vapour phase osmometry Mn upto 40,000
Cryoscopy Mn upto 40,000
Light scattering Mw upto 50,000
Sedimentation Mz upto
3.5 CHROMATOGRAPHY
As will be noted shortly, certain techniques such as colligative methods (Secs. 3.6–3.8),
light scattering photometry, special mass spectral techniques, and ultracentrifugation allow
the calculation of specific or absolute molecular weights. Under certain conditions some
of these allow allow the calculation of the molecular weight distribution (MWD).
These are a wide variety of chromatography techniques including paper and column
techniques. Chromatographic techniques involve passing a solution containing the to-be-
1. tested sample through a medium that shows selective absorption for the different compo-
nents in the solution. Ion exchange chromatography separates molecules on the basis of
their electrical charge. Ion exchange resins are either polyanions or polycations. For a
polycation resin, those particles that are least attracted to the resin will flow more rapidly
through the column and be emitted from the column first. This technique is most useful
for polymers that contain changed moieties.
2. In affinity chromatography, the resin contains molecules that are especially selected
that will interact with the particular polymer(s) under study. Thus, for a particular protein,
the resin may be modified to contain a molecule that interacts with that protein type. The
solution containing the mixture is passed through the column and the modified resin
preferentially associates with the desired protein allowing it to be preferentially removed
from the solution. Later, the protein is washed through the column by addition of a salt
solution and collected for further evaluation.
3. In high-performance liquid chromatography (HPLC), pressure is applied to the col-
umn that causes the solution to rapidly pass through the column allowing procedures to
be completed in a fraction of the time in comparison to regular chromatography.
When an electric field is applied to a solution, polymers containing a charge will
4. move toward either the cathode (positively charged species) or the anode (negatively
charged species). This migration is called electrophoresis. The velocity at which molecules
move is mainly dependent on the electric field and change on the polymer driving the
molecule toward one of the electrodes, and a frictional force dependent on the size and
structure of the macromolecules that opposes the movement. In general, the larger and
more bulky the macromolecule, the greater the resistance to movement, and the greater
the applied field and charge on the molecule the more rapid the movement. While electro-
phoresis can be conducted on solutions it is customary to use a supporting medium of a
paper or gel. For a given system, it is possible to calibrate the rate of flow with the molecular
weight and/or size of the molecule. Here the flow characteristics of the calibration material
must be similar to those of the unknown.
Generally though, electrophoresis is often employed in the separation of complex
molecules such as proteins where the primary factor in the separation is the charge on the
species. Some amino acids such as aspartic acid and glutamic acid contain an “additional”
acid functional group, while amino acids such as lysine, arginine, and histidine contain
“additional” basic groups. The presence of these units will confer to the protein tendencies
to move towards the anode or cathode. The rate of movement is dependent on a number of
factors including the relative abundance and accessability of these acid and base functional
groups.
Figure 3.8 contains an illustration of the basic components of a typical electrophore-
5. sis apparatus. The troughs at either and contain an electrolyte buffer solution. The sample
to be separated is placed in the approximate center of the electrophoresis strip.
Gel permeation chromatography (GPC) is a form of chromatography that is based
on separation by molecular size rather than chemical properties. GPC or size exclusion
chromatography (SEC) is widely used for molecular weight and MWD determination. In
itself, SEC does not give an absolute molecular weight and must be calibrated against
polymer samples whose molecular weight has been determined by a technique that does
give an absolute molecular weight.
Size exclusion chromatography is an HPLC technique whereby the polymer chains
are separated according to differences in hydrodynamic volume. This separation is made
possible by use of special packing material in the column. The packing material is usually
polymeric porous spheres, often composed of polystyrene crosslinked by addition of vary-
ing amounts of divinylbenzene. Retension in the column is mainly governed by the parti-
tioning (or exchanging) of polymer chains between the mobile (or eluent) phase flowing
through the column and the stagnate liquid phase that is present in the interior of the
packing material.
Through control of the amount of crosslinking, nature of the packing material and
specific processing procedures, spheres of widely varying porosity are available. The
motion in and out of the stationary phase is dependent on a number of factors including
Brownian motion, chain size, and conformation. The latter two are related to the polymer
chain’s hydrodynamic volume—the real, excluded volume occupied by the polymer chain.
Since smaller chains preferentially permeate the gel particles, the largest chains are eluted
first. As noted above, the fractions are separated on the basis of size.
The resulting chromatogram is then a molecular size distribution (MSD). The rela-
tionship between molecular size and molecular weight is dependent on the conformation
of the polymer in solution. As long as the polymer conformation remains constant, which
is generally the case, molecular size increases with increase in molecular weight. The
precise relationship between molecular size and molecular weight is conformation-depen-
dent. For random coils, molecular size as measured by the polymer’s radius of gyration,
R, and molecular weight, M, is proportional to Mb, where b is a constant dependent on
the solvent, polymer concentration, and temperature. Such values are known and appear
in the literature for many polymers, allowing the ready conversion of molecular size data
collected by SEC into molecular weight and MWD.
p
f d冢 冣
16 Ve
(3.14)
3.6 OSMOMETRY
A measurement of any of the colligative properties of a polymer solution involves a
counting of solute (polymer) molecules in a given amount of solvent and yields a number-
average. The most common colligative property that is conveniently measured for high
polymers is osmotic pressure. This is based on the use of a semipermeable membrane
through which solvent molecules pass freely but through which polymer molecules are
unable to pass. Existing membranes only approximate ideal semipermeability, the chief
limitation being the passage of low molecular weight polymer chains through the mem-
brane.
There is a thermodynamic drive toward dilution of the polymer-containing solution
with a net flow of solvent toward the cell containing the polymer. This results in an increase
in liquid in that cell causing a rise in the liquid level in the corresponding measuring tube.
The rise in liquid level is opposed and balanced by a hydrostatic pressure resulting in a
difference in the liquid levels of the two measuring tubes—the difference is directly related
to the osmotic pressure of the polymer-containing solution. Thus, solvent molecules tend
to pass through a semipermeable membrane to reach a “static” equilibrium, as illustrated
in Fig. 3.10.
Since osmotic pressure is dependent on colligative properties, i.e., the number of
particles present, the measurement of this pressure (osmometry) may be applied to the
determination of the osmotic pressure of solvents vs. polymer solutions. The difference
in height (⌬h) of the liquids in the columns may be converted to osmotic pressure () by
multiplying the gravity (g) and the density of the solution (), i.e., ⌬hg.
In an automatic membrane osmometer, such as the one shown in Fig. 3.11, the
unrestricted capillary rise in a dilute solution is measured in accordance with the modified
van’t Hoff equation:
RT
C BC2 (3.15)
Mn
As shown in Fig. 3.11, the reciprocal of the number average molecular weight
(M1
n ) is the intercept when data for /RTC vs. C are extrapolated to zero concentration.