user manual
low temperature circulating water tank
Page One
Page2
Thank you for purchasing a product from J.O Tech. We, J.O Tech Co., Ltd.
This device is a new product from Jeotech Co., Ltd., which is focusing on the localization of physical and chemical devices.
This user manual describes the functions and handling precautions of the purchased device.
are describing. All users should read this user manual before using the device.
Please read carefully to the end. To use this unit effectively and correctly
Before using this device, be sure to read the user's manual carefully and completely
Safety Precautions
WarningIf you operate this unit incorrectly,
Page3
[ Customer Support Contact ]
Seoul Branch
[Tel] 02-2627-3800
[Fax] 02-325-2335
Yeongnam Branch
[Tel] 053-381-9591
[Fax] 053-381-9594
www.jeiotech.com
e-mail : sales@jeiotech.com
If you have any questions while using the device, please contact the contact information above or the place of purchase.
please give it to me. Fast and accurate answers to faithfully utilize your devices
Page4
index
Page5
1. Purpose and characteristics of the device
It is a multi-purpose low-temperature circulating water tank used in various fields such as pharmacy,
medicine, chemistry, and biology.CLS(Custom Logical Safe)-Control Systemhas been implemented
(One)CLS-Control Systemmeans “a control device equipped with a logical safety mechanism dedicated to a specific device”
Thermal safety is perfect like in a laboratory where various flammable chemicals exist.
Research by JEO Tech for the safe use of heating-type devices operated in a required environment
Logic and composition developed by JinThe highest safety level control deviceis.
(2)CLS-Control SystemThe temperature control and heating amount of the silver device are capable of sophisticated PID calculation.
It is performed by the main CPU, and all operations for safety are separately built-in Logic of the preceding function.
The device is designed to perform in the IC so that it can withstand any electrical or electronic disturbances applied to the device.
(3)CLS-Control SystemWhen the safety device applied to the silver device operates, power is supplied to various parts2
All phases are immediately completely blocked, and the activated unsafe element is presented to the user as a visual and auditory device.
In a safe state, de-energized and until restarted after the unsafe element has been removed
will keep.
(4)CLS-Control Systemwill continue to operate the device when power is restored after a power outage
You want to stop the operation of the device even if power is restored due to other factors nearby.
(5)CLS-Control SystemThere is a weak current within only 5V and 10mA at the contact point of the safety device used.
It is constructed to flow, so there is no damage to the contact point even after a long time of use, so its durability is very long.
and an electronic device that controls the current flow of a heater that uses a large current when the safety device is activated
After stopping the control of (thyristor) first, through the contact separation of the power switch (Magnetic Switch)
By cutting off the power, damage to the contacts of the power switch and occurrence of noise were prevented.
Prevent damage to the contact point of the power switch and the occurrence of noise by performing in the reverse order when starting.
I did.
Page6
2. Equipment installation and precautions
(1) After opening the package, check the contents of the device.
Device body (1set), user manual (1 copy), glass tube fuse (2EA), silicone hose, communication
(2) This device is designed to operate normally under the specified power supply. Before installing the device
Check if the power on the ID Plate (name plate attached to the back of the product) matches the power used.
Power must be level 2 or higherUse a grounded power supplyand the Power Cord must be
(3) A suitable thermal liquid should be as non-flammable as possible, and in the case of flammability, the flash point (Fp) is 40°C
with more than 20 liquids°Liquids that can be circulated and pumped even below C should be used.
5 above the flash point of the flammable fruit°It is safe to use at a temperature lower than C (Fp-5).
- Be sure to use it under the power supply condition that can be grounded.
- Fruit oil has a flash point of 5°C low (Fp -5°C) Use it as a condition.
(4) When transporting the product, use an appropriate tool or carry it with two or more people.
(5) The outlet should be installed near the device and should be easy to use.
(6) Install the device on a flat and hard floor that is not subject to vibration or impact.
(7) Avoid heat sources or direct sunlight, and ambient temperature 5°C to 40°C, relative humidity
(8) Where there is an organic solvent such as Acetone, Methylene Chloride, etc. or strong high-frequency noise
(9) Prevent moisture, organic solvents, dust, and corrosive gas from entering the control box of the device.
do it.
Page7
3. Name and operation of each part
1) Display
Equipped with Main CPU equipped with proven software capable of Digital PID-Auto tune, Pt-
100ΩEquipped with the highest safety level control device such as temperature compensation function and heating amount control function for the sensor.
I am saying
Page8
(1) RUN LED
This LED indicates whether the device is running or stopped. When the device is in normal operation
The LED lights up, and the LED turns off when the instrument is stationary.
This is the LED that displays the timer with the function of starting the device after the set time has elapsed. LED
When is blinking, the timer is running; when the LED is fully lit, the timer is running.
A timer with a function to stop the device after the set time elapses after PV and SV match
LED to indicate. When the LED is blinking, the timer is running and the LED is fully lit.
It is a device that detects the low water level of the thermal liquid (distilled water, alcohol, oil, etc.) in the bath.
Because sus-ball is used, it can be used at high temperatures as well as durability, and when low water level is detected,
The machine will stop and the buzzer will continue to sound. In this case, after replenishing the solution, the Start/Stop
The temperature of the heating liquid in the bath becomes higher than the set value of the mechanical overheat prevention device installed in the front part.
In this case, the machine stops, a buzzer sounds and the overheat warning LED blinks. device is insecure
Since it is in a stopped state due to an element (overheating), take an appropriate action and press the Start/Stop button 1
Press twice to remove the buzzer and LED flickering before using the device.
It is used to set the timer, auto run function, and adjust the pump circulation amount.
Page9
(10) Up button
After changing the setting value, press this button to save the setting value.
A button that locks the buttons on the controller so they cannot be used.
When executing Auto Tuning, press this button for 3 seconds to activate the function.
When operating or stopping the device, or when the operation of the device is interrupted due to an unsafe factor
In this case, it is used to remove the warning sound and LED blinking after removing the risk factor.
Displays the set temperature or the remaining time when the timer is running.
2) Over-Temp. protector
It is a safety device with an independent circuit to protect the device and test subject from overheating.
The function of the controller is unstable or the temperature is higher than the set temperature due to the user's inexperienced operation.
When it heats up, it automatically stops the machine and sounds an alarm so the user can recognize overheating.
3) Communication Port
You can monitor the device status and change settings through the program. Also, the device
The control status can be saved and output through the printer is also possible.
Page10
4) Main Power Switch
5) Fuse
It is used to drain the solution in the bath with a specially designed valve structure. counterclockwise valve
The solution in the bath is drained by turning it 2 or 3 times in the opposite direction.
Please note that if you turn the valve more than 5 times, the valve may come off. Drainage after high temperature test
When doing this, please turn the valve after wearing gloves.
7) Frozen UNIT
It is a place where the optimal refrigeration system is configured, and the better the ventilation is, the more efficient it is.
8) CONDENSER COVER
In order to facilitate the cleaning of the condenser, it is a removable screen that is used to keep the refrigerator at all times.
Page11
4. Using the device
1) Basic use
(1) Before connecting the power, fill the bath with an amount of thermal fluid (distilled water, oil) suitable for the purpose of the experiment.
(2) In case of internal circulation, connect the silicone hose to IN/OUT LINE.
(3) Over Temp. 10% higher than the set value. Position the Protector Knob.
(4) Connect the power, turn on the Main Power Switch, and input the desired setting value.
The number on the window displaying the set temperature blinks during
I write.
(2) When the number on the window displaying the set temperature is flickering
(3) When there is no key input for about 10 seconds during temperature setting
In this case, the change is canceled and the status returns to the original state.
is.
Page12
3) additional function of the button
go Press key to finish setting. The set temperature is stored in the memory.
is a function. key with Press the key to set the temperature and
0000Press key to finish setting. The set temperature is stored in the memory.
is a function. key with Press the key to set the temperature and
0000Press key to finish setting. The set temperature is stored in the memory.
You can change between ℃ and ℉. Click the button to complete the setting
I'm done.
- It compensates the temperature error caused by the error of the temperature sensor and other causes.
function.
- The BAIS correction value is displayed on the current temperature display window (PV) and the set temperature table
Page13
The temperature displayed on the set temperature display window (SV) is
to the same value as the temperature on the thermometer. button and button
to finish.
Page14
③ The function of Timer is as follows.
Tem pe ra tu re
T im e
W a it on T im er W a it o ff T im er
☞ Wait on Timer
- The device operates after the time set in On Timer has elapsed.
- The maximum settable value is 99 hours 59 minutes, and the minimum value is 1 minute.
- Off Timer operates from the moment PV (current temperature) and SV (set temperature) coincide.
After the set time has elapsed, the device will stop.
- If both timers are set together, the device operates as shown in the figure above.
is.
To cancel Timer setting When the button is pressed, the LED turns off and the
Or, if you want to cancel only one of the two timers, set the time of the timer you want to cancel.
Page15
5) additional function of the button
If the device is stopped due to a power outage, or if the power is accidentally turned off,
This is a feature that automatically enables the device to work once it is restored.
If set to Yes, the device will run as soon as the power is restored.
- It is a function to control the circulation amount by adjusting the speed of the motor speed.
Finish.
When setting the motor speed to 4 or less, the viscosity characteristics of the solution according to the temperature change should be considered.
In the case of low temperature, the viscosity of the solution increases and the motor becomes overloaded, which causes malfunctions and does not
It is saved as
② If you press and hold the button (about 3 seconds), the next
In the window that displays the temperature, such as the rim, Auto
It hangs.
The current temperature and the set temperature become the same.
Page16
7) Lock function
(1) This button locks the buttons on the controller so that they cannot be used.
① When the Lock button is pressed for about 3 seconds, the Lock function is set with the Lock indicator and a beep sound.
② Press the Lock button again for about 3 seconds to release it.
③ Unnecessary elements caused by accidentally pressing a button during temperature control are removed in advance.
Page17
5. MAINTENANCE AND MANAGEMENT OF THE DEVICE
1) Device body
(1) Turn off the Main Power Switch and unplug the power plug from the power outlet.
(3) Clean the bath with a soft cloth moistened with neutral detergent.
(4) Clean the bath with a soft cloth moistened with distilled water.
(7) If the device is cleaned or decontaminated by other methods not specified in the manual, the user must
Appropriate methods not to damage the equipment should be checked with the manufacturer prior to implementation.
(8) Toxic chemicals that damage human skin or substances that generate dangerous gas overflow
When it spills, wear gloves that are safe for toxic chemicals and a mask that is safe for toxic gas.
2) Accessory
(1) Soak Spring Wire Rack and Test Tube Rack in neutral detergent.
RW-0525G/40G - -
Page18
6. Measures to be taken in case of abnormality
(3) Check that the Run LED on the display is off. Start/Stop button if off
(1) When “Err.S” is displayed on the window displaying the current temperature Input sensor disconnection or sensor
- If the device does not operate normally, turn off the power immediately.
- If you need A/S, you must contact JEO Tech or the place of purchase to receive repairs.
- When handling electrical parts inside the device, only qualified personnel must handle it.
Page19
7. Product Warranty Standards
It is one year from the date of purchase, and the free warranty repair service is not available after one year.
is. After the free warranty repair service period, you must pay the purchase price for the necessary parts.
During the free warranty repair service period, repair, replacement and
Refund service is available. In the case of export products, Jeio Tech is responsible for all services.
The product may be damaged due to accidents, contamination, or neglect of precautions due to natural disasters such as fire or water disaster.
Even if it is, you cannot receive the service. In addition, due to overvoltage ignoring the rated voltage,
Operation and misuse under abnormal conditions, including damage to the product, or resulting
Depreciation and damage to parts may also prevent you from receiving this service.
Postage by entering the model name, date of purchase, and serial number of the damaged product along with a detailed description
and insurance premiums payable to the payee. Of course, during the service period, the return fee is also
In principle, Jeio Tech bears the cost. Request product repair after the service period has elapsed
Please time.
This warranty repair service covers only the free repair, replacement and refund of the Product, and for any other specific purpose.
does not apply to And in case of data loss due to other causes,
and Jeio Tech is not responsible for any indirect accidental damage or loss.
As some of the above provisions do not make exceptions, the consumer who purchased this product
Page20
8. Specification 1
Model RW-0525G RW-1025G RW-2025G RW-3025G
Temperature StabilityOne)
0.05℃ 0.05℃ 0.05℃ 0.08℃
At -10℃ (bath fluid: ethanol)
Sensor Pt 100Ω
Bath
Step One 2 3 4 5
Flow Rate 5.6 (l/min) 8 (l/min) 12.8 (l/min) 18.8 (l/min) 28 (l/min)
Max. pressure,
5.9 PSI, (28.0 liters/minute)
max.flow rate
Max. suction
10.0 (liters/minute)
flow rate2)
Page21
at 0℃ 230W 250W 410W 570W
Electric Requirement
6.8A 8.9A 12.5A 13.5A
220VAC, 60Hz
Page22
8. Specifications 2
Model RW-0540G RW-1040G RW-2040G RW-3040G
Temperature StabilityOne)
0.05℃ 0.05℃ 0.05℃ 0.08℃
At -10℃ (bath fluid: ethanol)
Sensor Pt 100Ω
Bath
Step One 2 3 4 5
Flow Rate 5.6 (l/min) 8 (l/min) 12.8 (l/min) 18.8 (l/min) 28 (l/min)
Max. pressure,
5.9 PSI, (28.0 liters/minute)
max.flow rate
Max. suction
10.0 (liters/minute)
flow rate2)
Page23
at 0℃ 150W 330W 660W 700W
Electric Requirement
8.3A 10.4A 13.5A 14.5A
220VAC, 60Hz
Page24
9. Monitoring Program installation
(1) Insert the installation CD into the CD-ROM drive and the installation program will start automatically.
(2) If you click the “Next” button, you can specify the path to install the program.
(3) Click the “Start Installation” button to start installing the program.
(4) After the installation is completed, the Lab Tracer icon is created on the desktop.
caution: If you use Windows 95, 98 series OS, there may be an error in measurement time due to OS structure.
Page25
10. How to use Monitoring Program
(One)Comm → ConnectClick to connect the PC and the device through RS-232 communication.
(If the connection is notComm → Port settingClick to set the port number)
If the communication connection is successful, it is displayed at the bottom of the program window (circled).On Linemark appears.
Displays the current temperature, and the window below shows the output of the heater as a graph.
(2)View menu
View → ParameterIf you select , the current temperature, set temperature, and heater output are displayed along with the graph.
View → StatusIf you select , another window appears at the bottom of the current temperature and set temperature display window.
is created. In this window, items indicating the operating status of the device are displayed. Operating,
Page26
The next picture isViewof the menuStatusWowParameterAfter selecting , monitor the device status
One screen.
Through three windows (Graph window, Parameter window, Status window), the operating status of the device is accurately monitored.
The upper window among the graph windows indicates the current temperature (red line) and the set temperature (blue line).
When the blue line is Hi, it means On (operation), and when it is Low, it means Off (not operation).
indicates.
When the door of the device is opened, the blue line indicates Hi, and when it is closed, it indicates Low.
- PV indicates the current temperature, SV indicates the set temperature, and Heat indicates the output of the heater.
- Run Time indicates the time the device has been operated.
- This is the time from the moment the Run LED turns on by pressing the button to start the device.
- Wait On Timer, Wait Off Timer is the remaining time of the set Timer. in other words,
It is the time minus the elapsed time from the set time.
- Power Frequency indicates the frequency of the power currently being used.
Page27
(3) Menu Icon Description
ⓐⓑ ⓒⓓⓔ ⓕⓖ ⓗⓘ ⓙⓚⓛⓜ
ⓒ Connect (Ctrl + C)
It is used when connecting the experimental device and PC through RS-232 communication.
ⓓ Disconnect (Ctrl + D)
ⓔ Exit (Ctrl + X)
End program button. When this icon is executed, the graph so far
ⓕ Print (Ctrl + P)
ⓖ Preview
ⓗ ⓘ Scroll button
Auto Trace if you want the end point of the graph (the current temperature point of the device) to always be in the center of the window
On.
Set the range of Y-axis (temperature range) of graph automatically and manually. set manually
Page28
ⓝⓞⓟ ⓠⓡ ⓢⓣⓤⓥ ⓦⓧ ⓨ
ⓟ Panel View
The same screen as the front of the control unit's control panel is displayed on the PC screen.
ⓠ This is the function to set the pattern of program run and run the program.
- Program function can only be controlled using PC and cannot be set in Main.
This function is available only when Program Run is set on the device.
To change the temperature unit of the device, it must be changed in the control unit of the device.
Page29
① Print range
(If Graph, Status, Parameter Frame is on the main screen, it is printed, if not
· Print Current Page: Prints the page of the current main screen.
② Number of copies
If you press the Up and Down buttons, you can set from 1 to 100 sheets.
③ PV print interval
If Print at equal intervals is checked, PV and SV are converted to text according to the set print interval.
It will be printed.
④ Memo
This function allows you to make simple notes on the printed page.
⑤ Select Print
Page30
If the Print at equal intervals checkbox is checked on the print screen, the above
As shown in the figure, PV and SV are printed at regular intervals during preview and printing.
Also, if there is a particular location that the user wants to check, move the mouse cursor to the desired location.
When clicked, PV and SV are displayed along with the green line as shown in the picture above.
③ Zoom in/out
(6) Display
All operations of the display window are configured the same as those of the device.
If the PC and the device are connected, all control operations can be performed equally with the PC.
Page31
Click the PRG icon or select Pattern -> Pattern settings from the menu to
As shown in Figure 3 below, if you click the mouse on the SV Pattern Programming window, the set temperature and time,
If you want to edit the set value, drag & drop the selected Step (blue) to change the value.
If you change the temperature and time using the shortcut key, the temperature is changed by 1℃ and the time is changed in units of 1 minute.
Convenient.
① Shortcut
↑ : raise the temperature by 1℃ Alt + ↑ : Move the edit point to the right (next step)
↓ : Decrease the temperature by 1℃ Alt + ↓ : Move the position of the edit point to the left (previous step)
← : Decrease time by 1 minute Alt + ← : Move the position of the edit point to the left (previous step)
→ : Round up the time by 1 minute Alt + → : Move the edit point to the right (next step)
Page32
② Last step delete
④ Pattern save
Save the set pattern. The extension of the saved file is PIT.
Select the folder you want to save, name the file, and click the Save button to save it.
⑤ Pattern open
Select the file you want to open and press Open to display the saved pattern.
⑥ Start
After pattern setting is completed, press this button to operate the device. At this time, the low water level, Over
Temp. If there is something wrong with the device, such as, it will not work.
If the number of repetitions of the pattern is specified, the entire set pattern is repeated as many times as the specified number of times.
“Deleting the previous data”When you start after checking the checkbox, the previously tested data is
Page33
☞ Caution
If 99 hours is exceeded by increasing the number of repetitions, it does not run in Program Mode.
When setting the pattern, time and temperature must be set according to the performance of the device.
If you forcefully set the pattern beyond the maximum performance of the device, you will not be able to perform the desired experiment.
Page34
Page35