CAT.NO.E13111-T2112
Compact design for personal use,
to network with your computer.
The bench-top type temperature and humidity chamber series feature environmental testing
performance in a compact design.
Available in 20 liters and 60 liters capacities, these models offer a temperature range from
−20℃, −40℃, or − 60℃, to +150℃ or +180℃, while achieving excellent performance.
The chamber comes with user-friendly touch panel display, allows three-way access to
the chamber, and offering a broader range of options for superior expandability.
1
To minimize our chambers potential environmental impact
R-449A is the best alternative to R-404A
64%
Reduction
R-404A
R-449A
GWP
GWP
3920 1397
*R-449A is available on request
* Shown are equipped with options. (Standard: φ 50mm cable port on the right (not shown)) 2
Characteristics
440mm 440mm
Wide Variations
440mm
With three temperature range patterns
(−60℃, −40℃, or −20℃ ), two-volume
奥行
selections (22.5L or 64L), and with or
696mm
奥行 696mm w it hout hu m id it y, t h e B e n ch - t o p
696mm
chamber offers a total of 12 models to
690mm
690mm
3
Characteristics
4
Characteristics
Instrumentation Interlock
I/O Terminals
The chamber comes with I/O terminal
as a standard, which allows the chamber
t o wo r k i n s y n c h r o n i z a t io n w it h
external measuring instrumentation
such as a multimeter.
Three-way access
Chamber comes with a φ 50mm cable
port on the right side as a standard, but
you can enlarge it and or add more cable
ports on the right, left and the top.
International standards
Complies with the following standards:
ISO 12100 Safety of Machinery
IEC 60204-1 Low Voltage
IEC 61000-6-2 EMC
EN 55011 EMC
Custom-made Equipment
5
Controller P-200
Trend-graph display
Information notification
The INFO icon will blink when chamber
information is requiring attention. By ● Trend-graph
Email alert
p r e s s i n g t h e i c o n , yo u w i l l f i n d
notifications such as “Check Humidity
Tray” and “Check Wet Bulb Wick.” Ethernet PC
Intranet
(company LAN)
6
SH −20/−40/− 60 to +150℃
(+180℃)・30 to 95%rh
Refrigeration unit
7
TEMPERATURE & HUMIDITY CONTROL RANGE (SH) ■ Low GWP Refrigerant
100
95%rh
90
Relative Humidity (%rh)
80
70
60 R-449A is available on request.
50 (SH/SU-222, 262, 642, 662 only)
40
30
20
10
0
10 20 30 40 50 60 70 80 90 100
15℃ Temperature (℃) 85℃
unit:mm unit:mm
● SH-222・242 ● SU-222・242
440 17 696 34
440 17 696 34
86
86
300
300
690
●
625
250
504
250
100
95%rh
90
100
35
80
Relative Humidity (%rh)
70
● SH-642・662
60
● SU-642・662
540 17 921 5
50
540 17 921 5
40
86
86
30
20
10
400
544
730
400
665
544
0 400
10 20 30 40 50 60 70 80 90 100
400
15℃ Temperature (℃) 85℃
100
35
● SH-242-5・262 ● SU-242-5・262
440 17 786 31
440 17 786 31
86
86
300
690
504
300
250
504
625
250
100
35
8
SU −20/−40/− 60 to +150℃
(+180℃)
9
MODEL FITTINGS
S - 2 ●
Temperature (Humidity) recorder terminal
●
Specimen power supply control terminal
Temperature range 2: −20 to +150℃/180℃
4: −40 to +150℃/180℃
●
Alarm output terminal
6: −60 to +150℃/180℃ ●
External output terminal
Capacity 2: 22.5L
●
Cable port (φ 50 mm ×1)
6: 64.0L ●
Power cable (except 222, 242 model)
H: Temperature & humidity chamber
●
Water supply tank (SH only)
U: Temperature chamber ●
Humidifying tray drain plug (SH only)
●
Drain hose
●
Drain socket for water sensor box (SH only)
Ethernet port (LAN)
SAFETY DEVICES
●
●
USB memory port
●
Instrumentation interlock output terminal
●
Control circuit overcurrent protection
(except SH/SU-222, 242)
●
Instrumentation interlock input terminal
●
Cartridge fuse for control circuit short-circuit protection
●
System error (Error)
●
Room temperature compensation burnout detection circuit ACCESSORIES
●
Dry bulb temperature burnout detection circuit
●
Absolute upper/lower temperature limit alarm ●
Shelf (Stainless steel)
(w/ built-in T/H controller) Load capacity (evenly distributed)
●
Expansion analog board sensor burnout detection circuit SH/SU-222, 242, 262, 242-5 0.5kg
(SH/SU-242-5 only) SH/SU-642, 662 5kg
●
Temperature switch for air circulator Max. number of shelves
●
Thermal fuse SH/SU-222, 242, 262, 242-5 5 stages (pitch 35mm)
SH/SU-642, 662 5 stages (pitch 50mm)
●
Temperature switch for condenser fan ●
Connector (For temperature/humidity recorder terminals)
●
Overheat protector SH: 2/ SU: 1
●
Wet bulb temperature burnout detection circuit (SH only) ●
Cable port plug (rubber) 1 (φ 50 mm)
●
Refrigerator-1 error detection ●
Cartridge fuse
●
Refrigrator-2 error detection (except SH/SU-222, 242) SH/SU-222, 242, 262 (B type, 250V 7A) 1
●
Humidifier dry heat protector (SH only) SH/SU-642, 662, 242-5 (B type, 250V 7A, 6A) 1 of each
●
Humidifier water level detection (SH only) ●
Wet-bulb wick 1 box (SH only)
●
Temperature upper limit deviation alarm ●
Humidifying tray drain hose 2m 1 (SH only)
(w/ built-in T/H controller) ●
Drain hose for water sensor box (0.3m) 1
●
Absolute upper/lower humidity limit alarm (SH only) ●
Stylus pen (For touch panel operation) 1
(w/ built-in T/H controller) ●
Operation manual 1 set
●
System error (Alarm)
●
Water tank drought switch (SH only)
●
Chamber door switch
●
Water tank low-level switch (SH only)
●
Specimen power supply control terminal
Safety precautions
● Do not use specimens which are explosive or inflammable,
or which contain such substances. To do so could be
hazardous, as this may lead to fire or explosion.
● Do not place corrosive substances in the chamber. If
corrosive substances are generated by the specimen, the
life of the chamber may be significantly shortened
specifically because of the corrosion of stainless steel
and copper and because of the deterioration of resin and
silicon.
● Do not place life forms or substances that exceed
10
Options
Utility Observation
Power plug (220V AC only) Door with viewing window Inner glass door
W:352
64L type: W215×H290 mm
169
* Standard performance may not be met under
The water supply certain conditions. Inquire for details.
unit is installed [Example]
227
on the back of
・ SH/SU-242 Temp. extreme achievement time
the chamber.
(Pull down time)
From + 20 to −35℃ (Setting:− 40℃ )
Within 60 min.
* A water purifier (sold separately) is also ・ SH/SU-242-5 Temp. rate of change
available. (Heat up rate)
From − 21 to +131℃ 4.0℃ /min.
(Pull down rate)
Roof top water tank From +131 to − 21℃ 4.0℃ /min.
Hand-in port
An additional tank that supplements
the volume of the standard cartridge Roof top viewing window Equip the chamber with hand-in port
tank is provided to carry out continuous to operate specimen under test.
Effective view: W181×D107 mm <For inner glass door>
operation.
Effective water volume: 5L
* Not available on SH/SU-242-5 φ130mm ×1, at the center of the inner
* Standard performance may not be met under
Location: Chamber ceiling glass door.
certain conditions. Inquire for details.
* The connection port without pressure-reducing [Example] <For chamber side wall>
valve is required when selecting this option. ・ SH/SU-242 Temp. extreme achievement time φ130mm ×1 (select left or right side)
(Pull down time) Any hand-in port selected comes with
From + 20 to −35℃ (Setting:− 40℃ ) radial rubber seal.
Automatic water refill Within 60 min.
periodically.
Wet-bulb wick
46
*To prevent damage in the event of water leakage when installing the following optional
products, a dew tray (page 15) and other preventive measures can be prepared.
• Continuous water supply
11
Options
Specimen setting
107
Available dimensions: 181 Spec.: AC 6 to 24V 0.1 to 3A
●
φ 25 mm DC 1.5 to 60V 0.1 to 3A
●
φ 50 mm Connector Type: Block 10P (+5P, −5P)
●
φ100 mm Enclosure: Magnetized box with isolator
●
flat cable port (W100×H25 mm) Temperature Range:
* Comes with a rubber plug and a cap. −70 to +180℃ 30 to 95%rh
* Standard performance may not be met under Left side φ100 Cable port (option)
certain conditions. Inquire for details.
〈SH/SU-222・242・262・242-5〉 〈SH/SU-642・662〉
* Top not available on SH/SU-242-5
Cable port
50
Connector
58 50
178
Magnetic External
128
108
Connector Terminal
50 100 50 100
150 200
Top side Top side
Internal
178 Connector Terminal
108
128 50
58 50
100
200
100
150
50
50
Standard equipped
℃
非常停止
EMERGENCY STOP
● ●
●
for φ 50 mm For flat cable port
●
●
for φ100 mm For φ 50 mm with slits
●
for φ50mm spiral-wrapped plug for flat cable port with slits 12
Options
Same as standard shelf. The kit includes: Communication ports to connect the
18-8 Cr-Ni Stainless steel cable ties chamber to a PC.
magnetic cable cover ●
RS-485
dew tray ●
RS-232C
●
GPIB
Communication cables
Specimen basket
13
Options
Measurement
A temperature & humidity recorder The following contact signals are Capable of applying voltage to the
that utilizes a liquid-crystal display installed on the option box, or stand specimen, used for bias testing. The
fitted with a touch-panel. with option box. DC power supply unit synchronizes
Display: 5.7inch color touch panel ●
Time up output terminal with constant program operations, and
Scan interval: 5 sec. (default) E n a b l e s p o w e r s u p p l y a n d /o r can be set for each temperature and
Internal recording media: temperature measurement of the humidity program step.
Flash memory 8MB specimen synchronised with the ●
5V
External recording media: timer. ●
12V
CF memory card port ●
Time signal terminal ●
15V
(Includes a 256 MB CF card) Add up to 10 signal terminals to the ●
24V
USB memory port 1 equipped as standard. ●
48V
< Temperature & humidity type >
●
Temp. & humid. SP attainment output * Not available on SH/SU-242-5
Portable type
Recording method: Dot
Recording paper: Effective width 100 mm
No. of inputs:
< Temperature & humidity type >
Temperature 5, Humidity 1 Option box B (stand-embedded paperless recorder)
●
−50 to +150°C/0 to 100% rh
●
−100 to +150°C/0 to 100% rh
●
−100 to +200°C/0 to 100% rh
< Temperature type >
Temperature 6
●
−100 to +200°C
107
181 181
Options
Performance Safety
〈SH/SU-222・242・262・242-5〉 〈SH/SU-222・242・262・242-5〉
〈SH/SU-642・662〉 〈SH/SU-642・662〉
Sensors are attached to the specimen If SH/SU-242-5
* Top not available on the temperature inside the
* Top not chamber
available Please select lighted or blinking, and
on SH/SU-242-5
to allow exposure tests that provide decreases excessively, the chamber requirement of buzzer sound.
temperature stress to the specimen. stops operating to prevent the specimens Displayed levels: up to level 4
50
50
Insulated type
●
from being damaged. Pole length: 226mm
58 50
58 50
* Not available on SH/SU-242-5 * The length can be reduced by 10mm (up to
178
178
128
128
108
108
56mm)
External device alarm * Wit h st a nd or st a ck i ng cha mbe r, plea se
intput terminal contact us.
50 100 50
10050 100 50 100
150 150 200 200
Equips
Top side the chamber with a terminal
Top sidethat
Top side Top side
is used to stop operation of the chamber
178 178
108 108
in the event that an external device128 50 128 50
58 50 58 50
linked to the chamber malfunctions.
100
100
200
200
100
100
150
50
50
50
Attached in place of the wet bulb wick. Equips the chamber with a terminal
Measurement range: 0 to 100%rh that outputs the door open status.
Accuracy: ± 2%rh Capable of controlling an external
(−20 to +40℃ and 0 to 90%rh) Right side
device that operates along Right
withside
door
Right side Right side
* Same location points for left side * Same location
* Samepoints
location
for left
pointsEmergency
sidefor left side stop* pushbutton
Same location points for left side
op e r at ion a nd re cord s t he t e mp -
Standard equipped Standardhistory.
erature disturbance equipped
Stops the chamber immediately.
Available with or without guard.
温度過昇防止器 非常停止
OVERHEAT EMERGENCY STOP
PROTECTOR
℃
MODE
Document
●
CD ●
Testing and inspection report
●
Booklet ●
Test data
●
Temperature (& humidity) uniformity
measurement
●
Calibration report
●
Calibration certificate
●
Traceability system chart
●
Traceability certificate
15
Options
Stand variation
Stand equipped with casters for easy transfer or transportation. (leveling feet provided)
Choose among 3 sizes: C (Dolly type), L (Low type) or H (High type)
Dimensions : mm For SH/SU-642・662 (For 222・242・262・242-5)
H type
The C type stand fits on the lower part.
・With shelf ・With option box ・With 19 inch rack
Chamber with L stand Move the shelf to install instrumentation Following options can be installed. 19 inch size instrumentation or
can fit under the H stand. or measurement devices. ・Paperless recorder controller can be set to the rack.
・Output terminal set
・Specimen temperature control
・Program-synched DC power supply
210
1140
975
726
770
745
670 (570)
Shelf size : W577×D900
Depth : 925 (815) (W473×D790)
M type L type
・With shelf ・With water tank ・With shelf ・With water tank
shelf : adjustable 3 pitches With water capacity : 20L (10L) shelf : adjustable 3 pitches Capacity : 18L
Drainage capacity : 20L (10L)
55 (75)
360 (400)
50 (50)
50 (50)
235
95 (115)
50
700
700
50
540 (440) * The connection port without
pressure-reducing valve is
255
540 (440)
Depth : 860 (750)
* For safety reasons, make sure to secure the chamber and the stand together
with anchoring fixtures. We also recommend to fix the stand itself to the floor.
* Please inquire for 2-stage stacked chambers.
16
Options
● Retrofit is not available. ○ Retrofit is available.
SH SU
Page OPTION
222 242 262 642 662 242-5 222 242 262 642 662 242-5
※3 ※3 ※3 ※3 ※3 ※3
Power plug (220V AC) ● ● ● ● ● ● ● ● ● ● ● ●
Continuous water supply ※1
○ ○ ○ ○ ○ ○ ─ ─ ─ ─ ─ ─
Roof top water tank ※1
○ ○ ○ ○ ○ ─ ─ ─ ─ ─ ─ ─
Automatic water refill ※1 ○ ○ ○ ○ ○ ○ ─ ─ ─ ─ ─ ─
11 Wet-bulb wick ○ ○ ○ ○ ○ ○ ─ ─ ─ ─ ─ ─
Door with viewing window ※1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Roof top viewing window ※1 ● ● ● ● ● ─ ● ● ● ● ● ─
Inner glass door ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Hand-in port Inquire for details.
Additional cable port Inquire for details.
12 Cable port rubber plug ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
EZ connect cable port plug ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Shelf ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Specimen basket ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Cable organizer kit ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
13
Noise reduction rear cover ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Interface ※1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Communication cables ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Paperless recorder ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Temperature (humidity) recorder ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
External output terminal set (×3) ※1 ※2
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Temperature recorder output
14 ○ ○ ○ ○ ○ ○ ─ ─ ─ ─ ─ ─
terminal
Thermocouple ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Applying DC power supply ※1 ※2 ○ ○ ○ ○ ○ ─ ○ ○ ○ ○ ○ ─
Option box ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Specimen temperature control ※1 ※2 ○ ○ ○ ○ ○ ─ ○ ○ ○ ○ ○ ─
Electrostatic capacitance-type
● ● ● ● ● ● ─ ─ ─ ─ ─ ─
humidity sensor control
Airflow adjuster ※1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Overcool protector ※1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
External device alarm
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
intput terminal ※1
15
Door opening signal
○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
output terminal ※1
Status indicator light ● ● ● ● ● ● ● ● ● ● ● ●
Emergency stop pushbutton ※1 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Chamber dew tray ● ● ● ● ● ● ● ● ● ● ● ●
Operation maual ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
Reports & certificates ● ● ● ● ● ● ● ● ● ● ● ●
16 Stand variation Inquire for details.
*1 except 115V AC NEC, 200V AC NEC
*2 The option box is required when selecting these options.
*3 C type only
Chamber performance may not satisfy specifications depending on the combination of options.
17
Network
* Requires an intranet Supported browser: Internet Explorer 11
Login privileges
Displaying data in trend-graph Screen Constant/
Chamber monitor Run/Stop Configuration
Privileges Program setup
Settings and measured data saved in the
chamber can be displayed in graphs on Administrator
User
インターネット
エスペック
E-mail オンラインコア(別売)
notifications
試験室
A
Mail server
OnlineCore Test
area
B
Offices Network
camera
Edit Measuring
Test chambers of instrument
other manufacturers
18
Network
https://www.espec.co.jp/english
Head Office
3-5-6, Tenjinbashi, Kita-ku, Osaka 530-8550, Japan
Tel : 81-6-6358-4741 Fax : 81-6-6358-5500