You are on page 1of 2

GDS Instruments Sdn. Bhd.

124,JalanKapar27l89, TamanAlam Megah, Section 27,40400 ShahAlam, SelangorDarul Ehsan, Malaysia.


(
GDS
CERTIFICATE OF CALIBRAT roN
Page 1 of 2
Job Reference 2022A122 Certificate Reference :
Date of lssue : tlFebruary2O22
GDSt/CEl22BlOA4
CUSTOMER DETAIT Sticker Number : 2A22N22-2
Customer: Prosoil Foundation Consultant
Address: 88/B, lndira Road, Firmgate,
Dhaka-1215, Bangladesh.

Person ln Charge : N/A


Contact Number HP N/A Office : +88 018 19218230

CALIBRATION ITEM DETAILS


lnstrument: Piezocone
Type lndication : Digital Serial Number: N/A
Manufacturer : Geomil Equipment
V Model: AS10CF11P
Serial Number : s18235

Tip Resistance, gc Sleeves Friction,/s Pore Pressure, u


Capacity: 100 kN 15 kN 20 bar
Resolution : 0.01Mpa 0.001Mpa 0.001Mpa
Calibration Sl Unit IN to Mpa kN to Mpa bar to Mpa
Calibration Range 0kN-100kN 0kN-15kN 0 bar - 20 bar
Net Area Ratio : 1000 mm2 15(X)0 mm2 N/A
Date of Calibration : 11 February 2022
Recalibration Date : 11 February 2023
The user should be aware these are many factors may cause this instrument to drift out of calibration
limits prior to the stated recalibration date.

CONDITION OF CALIBRATION ITEM


Date of Received : 27 lanuary 2022
Before Calibration : Good
After Calibration : Condition as Submited by Customer
u Location of calibration : rn-house calibration {control Room}
Method Calibration: GDS-CP03-01

CalibrationEnvironment: Temperature 24.3 24.7 )oc


Relative Humidity 55.0 s5.0 l% RH
REFERENCE STANDARD USED
Calibration Equipment Serial Number Traceable to Certificate Reference
Load Cell with Hydraulic Pump 15319 GDS lnstruments GDS|/LC/21110s1
Hydra ulic Deadweight Tester s80/30188 NMIM NMtM-3560-M-21
Thermohygrometer SYA383 Trescal PSyP-21037949

Calibration by

This
Muhammad
Calibrator
b. Jdmaludin ffi
certifi.ate is issued in accordance with the condition of accreditation by sAMM which assessed the measurement capability of iaboratory
units of measurement realised to corrcsponding nBtionai standard laboratory. Copyright oI this cetificate is owned
Ang
Approve Signatory
An

* Continue next page *


and its traceability to reco8nized national standards and to the
by the issuinB laboratory and may not be reproduced other than in full ex.ept with the prior
written approval from Head of Laboratory.
l Any enquiry please email or contact us as indicate_

Tel: 03-5192 3228 Fax:03-5192 3230 Email: admin@gdsi.com.my www.gdsi.com.my


GDS Instruments Sdn. Bhd. @0474s-
124, Jalan Kapar 27l89,Taman Alam Megah, Section 27,40400 Shah Alam, Selangor Darul Ehsan, Malaysia.

Page 2 of 2
Calibration ltem Piezocone Certificate Reference GAS,lCEl22Bl0f,4
Serial Number: s18235 Date of Calibratisn : 11 2022
Re Result of Calibration

Ti Resistance
Nominal Force Mean Net Ared Ratio { Ac ) LOOO mmz
Error
qc=Qc/Ac Zero Load Error -o.oL %
Qc
,(N Mpa KN Mpa % Maximum Non-LinearitY o.o3 %
0.00 0.00 0,00 0.00 0.00 Baseline Output 229 mV
10.00 10.00 10.00 0.00 0.00 Meosurement Output 8752 mV
20.00 20,0L 20.01 0.01 0.0s Best Fit Curue
30.00 30.02 30.02 0.02 0.07
40.00 44.O2 40.o2 0.02 0.04 y=q(xz)+b{xJ+c
50.00 50,04 50.04 0.03 0.07
60.00 60.04 60.04 0.04 a.o7 a= -2.0979E-0s
70.00 70.o3 70.03 0.03 0.04 b 1.0020E+00
80.00 80.03 80.03 0.03 0.04 L- -1.46508-02
90.00 90.01 90.01 0.01 0.01
100.00 99.96 99.96 -0.05 -0.05

t Expanded measurement of uncertainties at 95% Confident Level = t A.70 Mpa k = 2

Sleeves Friction
2
Nominal Force Mean Net Area Ratio ( As ] 15000 mrn
Error
Fs fs=Fs/As Zero Load Error 0.36 %
KN Mpa KN Mpa % Moximum Non-Linearity a.oz %
0.00 0.000 0.000 0.000 0.00 Baseline output 2L9 mV
1".50 0.100 1.500 0.000 0.00 Meosurement Output 1575 rnV
3.00 0.201 3.015 0.001 0.50 Best Fit Curve
4.50 0.301 4.515 0.001 0.33
6.00 0.400 s.993 -0.001 -0.13 y=atx2)+b(x)+c
7.50 0.498 7.463 -0.002 -0.50
9.00 0.595 8.918 -0.005 -0.92 -
(, 1.4530E-04
10.50 0.694 10.403 -0.006 -0.93 =
fur 6.4727E-Oz
12.00 0.799 11.978 -0.002 -0.19 c= 3.1084E-03
13.50 0.906 L3.583 0.006 0.61
15.00 1.009 15.13s 0.009 0.90

Exponded measurement of uncertainties at 95% Confident Level = ! 0.007 Mpa k = 2

Pore Pressure u
Nominal Zero Load Error -o.o2 %
Mean Error
Pressure Maximum Non-Linearity o.oo %
bar Mpa Mpo % Baseline Output 2OA mV
0 0.000 0.000 0.00 Measurement Output 8A25 mV
2 0.199 -0.001 -0.03 Best Fit Curve
4 0.399 -o.o01 -0.02
5 0.599 -0.001 -0.02 y=a(xzJ+b{x}+c
a 0.799 -0.001 -0.01
10 o.999 -0.001 -0.01 d
g
= 4.3706E-06
'J,2 1.199 -0.001 -0.01 = 9.9890E-02
1,4 1.399 -0.00L -0.01 c= -4.1958E-04
16 1.599 -0.001 -0.01
18 t.799 -0.001 -0.01
20 1.999 -0.001 0.00

Expanded measurement of uncertointies ot 95% Confident Level = t A.007 Mpa k = 2


I'lote - Apply the following equations to calculate force in terms of deflection ( x ) is measurement reading and a,b,c are constant.
- The assessment of calibration item based on method defined by BS EN ISO 22476 and ASTM D5778 standard and the uncertainty stated in this certificate
evaluated in accordance in JCGM 100. e
The reported expanded measurement uncertainty is stated as the standard measurement uncertainty multiplied by the coverage factor k=2 such that
probability corresponds to approximately 95 %.
it
End

written approval from Head of Laboratory. Any enquiry please email or contaat us as indicate.

Tel: 03-5192 3228 Fax: 03-5192 3230 Email: admin@gdsi.com.my www.gdsi.com.my