You are on page 1of 3

UNIT IV ULTRASOUND BASED BLOOD FLOW MEASUREMENT (NON-INVASIVE METHOD)

There are basically two types of ultrasonic blood flow-velocity meters. The first type is the
Doppler-shift and the second is the transit time velocity meter.

(i) DOPPLER-SHIFT FLOW VELOCITY METER


It is a non-invasive technique to measure blood velocity in a particular vessel from the surface of
the body. It is based on the analysis of echo signals from the erythrocytes in the vascular
structures.
Because of the Doppler Effect, the frequency of these echo signals changes relative to the
frequency which the probe transmits.
The Doppler frequency shift is a measure of the size and direction of the flow velocity. The
principle is illustrated in Fig..
The incident ultrasound is scattered by the blood cells and the scattered wave is received by the
second transducer. The frequency shift due to the moving scatterers (blood) is proportional to
the velocity of the scatterers.

The incident frequency in blood, f1 is given by,

The Scattered Frequency in blood, f2 is given by,

The resultant Doppler shifted frequency, ∆f is given by


Velocity of the blood,

Flow rate= Area*Velocity

 The piezo-electric crystal A is electrically excited to generate ultrasonic waves, which enter the
blood.
 Ultrasound scattered from the moving blood cells excites the receiver crystal.
 The electrical signal received at B consists of a large amplitude excitation frequency
component, which is directly coupled from the transmitter to the receiver, plus a very small
amplitude Doppler-shifted component scattered from the blood cells.
 The detector produces a sum of the difference of the frequencies at D. The low-pass filter
selects the difference frequency, resulting in audio frequencies at E. Each time the audio wave
crosses the zero axis, a pulse appears at G.
 The filtered out put level at H will be proportional to the blood velocity. The following two
pitfalls are encountered in Doppler ultrasonic blood flow meters. High frequency response is
usually inadequate which introduces a non-linearity into the input-output calibration curve.
(ii) TRANSIT TIME DIFFERENCE FLOW METER
 A pair (or pairs) of transducers, each having its own transmitter and receiver, are placed
on the artery wall, one (set) on the upstream(A) and the other (set) on the downstream(B).
 The time taken for acoustic waves to travel from Transducer A to Transducer B is shorter.
Because, acoustic wave travel to the same direction of the blood flow.
 The time taken for acoustic waves to travel from Transducer B to Transducer A is Longer.
Because, acoustic wave travel to the opposite direction of the blood flow.
 This time difference is directly proportional to the velocity of blood.
 The larger the difference, the higher the flow velocity.

You might also like