OPERATION MANUAL
2015/08
Rev. 1.08
1
ID510 WEIGHING CONTROLER
Contents
PROCUREMENT MODEL SELECTION .................................................................................................................................... 4
CHAPTER 1.0 INTRODUCTION .................................................................................................................................... 5
1.1 OVERVIEW ................................................................................................................................................................... 5
CHAPTER 2.0 KEYPAD AND SETUP MENU ............................................................................................................... 6
2.1 KEYPAD OPERATION .................................................................................................................................................... 6
2.2 OPERATOR MENU......................................................................................................................................................... 6
2.2.1 Top Menu ............................................................................................................................................................. 6
2.2.2 Scale Configuration.(Scale Calibration Config.) .............................................................................................. 7
2.2.3 Scale Cal.(Scale Calibration) .............................................................................................................................. 8
2.2.5 Batching Control.(Batching Controller Configuration) .................................................................................... 10
2.2.6 Serial Port(Serial Port Configuration) ............................................................................................................... 11
2.2.7 I/O (I/O Option Assignment).............................................................................................................................. 13
2.2.8 PLC Fieldbus (PLC Field Bus Configuration) .................................................................................................. 15
2.2.9 LAN (TCP/IP) .................................................................................................................................................... 15
2.2.10 Maintenance .................................................................................................................................................... 16
2.2.11 Total (Scale Print Total & Fill Total) ............................................................................................................... 17
2.3FUNCTION KEY OPERATION INSTRUCTIONS ................................................................................................................. 18
2.3.1 MEM ............................................................................................................................................................... 18
2.3.2 STRT ............................................................................................................................................................... 18
2.3.3 EMPTY ........................................................................................................................................................... 18
2.3.4 PAUS ............................................................................................................................................................... 18
2.3.5 ABORT ............................................................................................................................................................ 19
2.3.6 RSUM ............................................................................................................................................................. 19
2.3.6 NEXT .............................................................................................................................................................. 19
2.3.6 RST ................................................................................................................................................................. 19
CHAPTER 3.0 INGREDIENTS USED........................................................................................................................... 19
3.1 TARGET SETTING ........................................................................................................................................................ 19
3.2 THE PROCESS INTRODUCTION .................................................................................................................................... 20
3.2.1 SEQUENCE BATCHING PROCESS(MSEQUEN OR ASEQUEN) .............................................................................. 20
3.2.2 RANDOM FEED PROCESS(MRANDOM OR ARANDOM)...................................................................................... 21
3.2.3 THE PROCESS OF RANDOM FEED WITH FORMULA (PFRAND) ............................................................................. 22
3.3 FILLING MATERIALS ................................................................................................................................................... 22
3.4 AUXILIARY COMPARATOR .......................................................................................................................................... 22
CHAPTER 4.0 SERIAL COMMUNICATION APPLICATION .................................................................................. 23
4.1 CONTINUOUS OUTPUT ................................................................................................................................................ 23
4.2 COMMAND INPUT/OUTPUT ......................................................................................................................................... 25
4.3 PRINT OUT (KEYBOARD PRINT OR COMMAND PRINT) .................................................................................................. 25
4.3.1 Single line printing shows the weight output format.......................................................................................... 25
4.3.2 Single - line printing of wool, leather, net weight output format ....................................................................... 25
4.3.3 Multi-line printing of wool, leather, net weight output format ........................................................................... 26
4.4 MODBUS-RTU ........................................................................................................................................................ 27
4.6 BATCH RECORD PRINTING(FILLPRT) .......................................................................................................................... 32
4.7 THE CUMULATIVE PRINT(TOTALPRT) ......................................................................................................................... 32
CHAPTER 5.0 BUS OPTIONS APPLICATION ........................................................................................................... 33
5.1 PROFIBUS-DP ......................................................................................................................................................... 33
5.1.1 DESCRIPTION OF PARAMETER PREPARATION............................................................................................................ 33
5.1.2 STEP7 THE CONFIGURATION WIZARD ..................................................................................................................... 33
5.1.2.1 Install ID510 GSD file to STEP7 .................................................................................................................... 33
5.1.2.2 Configure configuration word length.............................................................................................................. 34
5.1.2.3 Monitor reading weight display ...................................................................................................................... 34
5.1.2 DATA FORMAT DEFINITION ...................................................................................................................................... 35
5.1.2.1 Integer form (configuration length is 2 words) ............................................................................................... 35
5.1.2.2 Floating point form (configuration length is 6 words) ................................................................................... 36
5.1.2.4 Table of variables corresponding to S0~S7 in floating point form ................................................................. 39
2
ID510 WEIGHING CONTROLER
CHAPTER 6.0 SD MEMORY CARD APPLICATION................................................................................................. 43
6.1PRINT STORAGE FUNCTION ......................................................................................................................................... 43
6.2BATCH RECORD PRINTING AND STORAGE FUNCTION ................................................................................................... 43
CHAPTER 7.0 FREE WEIGHT CALIBRATION ........................................................................................................ 44
7.1 FREE WEIGHT CALIBRATION PRINCIPLE ...................................................................................................................... 44
7.2 FREE WEIGHT CALIBRATION PROCESS ........................................................................................................................ 44
7.3 EXAMPLE OF ZERO WEIGHT CALIBRATION .................................................................................................................. 44
CHAPTER 8.0 EMERGENCY RECOVERY CORRECTION RESULTS ................................................................. 45
CHAPTER 9.0 HARDWARE DESCRIPTION .............................................................................................................. 45
9.1 DIP SWITCH ............................................................................................................................................................... 45
9.2 WIRING DEFINITION ................................................................................................................................................... 46
9.2.1 THE MAIN BOARD .................................................................................................................................................... 46
9.2.2 I/O CHOOSE PLATE ............................................................................................................................................... 46
CHAPTER 10.0 MACHINE SIZE .................................................................................................................................. 47
APPENDIX 1 SHOWS & ERROR.................................................................................................................................. 47
3
ID510 WEIGHING CONTROLER
4
ID510 WEIGHING CONTROLER
1.1 Overview
One analog load cell I/F, support up to ten 350 Ohm(350 Ohm x 10) load cells
Up to 150,000 display Division
128X64 dot matrix OLED display
RTC(Date & Time)
Two serial ports
COM1 - RS232
COM2 - RS232/485
10/100M Ethernet port
Continuous Output Weight ( TCP/UDP)
Demand Input and Output (TCP/UDP)
MODBUS-TCP Server (TCP)
8-Input/12-Output OC type
Remote configuration, calibration and lock keypad via COM1,COM2 and PLC I/F
200Hz weight updating speed
200Hz target comparison speed
100Hz PLC I/F (PROFIBUS-DP) weight updating speed
50Hz MODBUS-RTU, MODBUS_TCP updating speed
Option Board
PROFIBUS-DP
4~20mA analog output option : x1 or x2
SD memory card
Operating Environment Conditions
Operating Temperature: -100C 400C Humidity: 10% 95% No Condensation
Stroage Temperature: -400C 600C Humidity: 10% 95% No Condensation
5
ID510 WEIGHING CONTROLER
1) Page 1
Scale Configuration :
Enter Scale Calibration Configuration Page, See 2.2.2
Scale Cal. :
Enter Scale Calibration Page, See2.2.3
Scale Appl. :
Enter Scale Application Configuration Page. See 2.2.4
Blend Fill :
Enter Blend Fill Configuration Page. See 2.2.5
Serial Port :
Enter Serial Port Configuraton Page, See 2.2.6
6
ID510 WEIGHING CONTROLER
2) Page 2
PLC Fieldbus :
Enter PLC Field Bus Configuration Page, See 2.2.8
LAN (Ethernet) :
Parameters set into LAN interface, See 2.2.9
Maintenance :
Enter Maintenance Page, See 2.2.9.10
X10 Mode OFF/ON : X10 Mode Switch
Calibration Mode :
7
ID510 WEIGHING CONTROLER
2.2.3 Scale Cal.(Scale Calibration)
2) Range Correction
Filter :
L 0– Lightest filter L1,L2…..L9-Heaviest filter
Power Up Zero:
0 – Disable power up zero, pre-zero point will be used after power on
1~50% - The range of enable power up zero
※ Not more than ±10% automatically in W&M sealing mode
Pushbutton Zero :
0 – Disable Pushbutton Zero
1~50% - Enable pushbutton zero range
※ Not more than ±2% automatically in W&M sealing mode
2) Page 2
Pushbutton Tare :
Disable – Disable Pushbutton Tare
Enable – Enable Pushbutton Tare
3) Page 3
Present Tare :
4) Page 4
5) Page 5
Motion Checking :
Rate Unit :
S – Second, M – Minute, H – Hour
6) Page 6
9
ID510 WEIGHING CONTROLER
1) Page 1
Wt—In Materials :The number of feeding ingredients is set to support 1~8 materials. If it is
configured with 0, the feeding is prohibited
Fill Mode : MSequen – Manual sequential batching , ASequen – Automatic sequential batching
MRandom – Manual random batching ARandom – Automaticl random batching
PFRand – With the recipe to manage the random mix pattern
Wt—In Enable Checking :
Disable – Check before starting feed to disable signal function,
Enable - Check before starting feed to enable signal function
2) Page 2
Start Delay: Delay time before starting each material feeding: 0~9999 x0.1 second
3) Page 3
Auto Spill Adjust Mode: Disable – Self-learning is prohibited, Enable – Allow self-learning
Auto Spill Factor: 0 – Self-learning, 1~100 – Revise the advance according to the learning factor
Drain Time: Air time, 0~9999 x0.1s
4) Page 4
Auto JOG Mode: Disable – Banned feeding
Manual – Manual feeding is allowed
Auto – Automatic filling material
JOG ON Time: 0~99 x0.1s,Open the slow feed valve time when filling
JOG OFF Time: 0~99 x0.1s,Shut off the slow feed valve time when filling
10
ID510 WEIGHING CONTROLER
5) Page 5
Weight—Out Start Delay Mode: 0~9999 x0.1 s, When starting discharge (blanking and discharging),
delay the setting time before starting discharge processing
6) Page 6
7) Page 7
1) Page 1 (COM1)
Assignment
None:
CRPT: 17 bytes continuous output weight , Refer to 4.1
CPRT-C: 18 bytes continuous output weight, with CHK(Checksum),Refer to 4.1
DPRT-1: Demand Input and 1-line to print display weight,Refer to 4.2, 4.3
DPRT-1-C: Demand Input and 1-line to print display weight with CHK(Checksum),
Refer to 4.2, 4.3
DPRT-2: Demand input and 1-line print gross weight, tare weight and
net weight output, Refer to 4.2, 4.3
DPRT-2-C: Demand input and 1-line to print gross weight, tare weight and
11
ID510 WEIGHING CONTROLER
net weight output , with CHK(Checksum), Refer to 4.2, 4.3
DPRT-3: Demand input and 3-line to print gross weight, tare weight and
net weight output, Refer to 4.2, 4.3
DPRT-3-C: Demand input and 3-line to print gross weight, tare weight and
net weight output with CHK(Checksum),Refer to 4.2, 4.3
DPRT-5: Demand input and print display weight, date and time in lines
Refer to 4.3.4
Baud Rate : 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
2) Page 2 (COM2)
12
ID510 WEIGHING CONTROLER
2.2.7 I/O (I/O Option Assignment)
13
ID510 WEIGHING CONTROLER
Output Assignment
Motion ON – Motion Flag of Scale Status
Net ON – Net Flag of Scale Status
Center of Zero ON – Center of Zero of Scale Status
Zero Tol. OK ON – Zero Tolerance OK
Fast Feed Fast feed output of feed in batching process
Fine Feed Fine feed output of feed in batching process
Running The batching process is in operation
Suspended Suspension of batching process
Feeding time delay Start feeding
Tolerance Burden overshoot detection output
Feed complete ON - end of feed for ingredients
Discharge delay ON – The meter is in the delay before starting discharge (emptying and discharging)
Empty feeding Empty feed output
Rapid feeding Quick discharge output with constant value
Slow feeding Constant output of blanking at slow speed
Discharging End of discharge (empty discharge or constant discharge)
finished
Assist 1 Auxiliary one (Under) output
Auxiliary two Auxiliary two (Over) output
Ready Batch ready output
Alarm Charge alarm output
Material 1 Material 1 Output
Material 2 Material 2 Output
Material 3 Material 3 Output
Material 4 Material 4 Output
Material 5 Material 5 Output
Material 6 Material 6 Output
Material 1 Fast Fast feed output of material no. 1
Material 1 Slow Slow feed output of material no. 1
Material 2 Fast Fast feed output of material no. 2
Material 2 Slow Slow feed output of material no. 2
Material 3 Fast Fast feed output of material no. 3
Material 3 Slow Slow feed output of material no. 3
Material 4 Fast Fast feed output of material no. 4
Material 4 Slow Slow feed output of material no. 4
Material 5 Fast Fast feed output of material no. 5
Material 5 Slow Slow feed output of material no. 5
Material 6 Fast Fast feed output of material no. 6
Material 6 Slow Slow feed output of material no.6
Material 7 Fast Fast feed output of material no. 7
Material 7 Slow Slow feed output of material no. 7
Material 8 Fast Fast feed output of material no. 8
Material 8 Slow Slow feed output of material no. 8
Vibration 1 Feed shock 1 output
Vibration 2 Feed shock 2 output
Vibration 3 Feed shock 3 output
Vibration 4 Feed shock 4 output
NO.1 Ready Material no. 1 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
- finish
NO.2 Ready Material no. 2 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
14
ID510 WEIGHING CONTROLER
- finish
NO.3 Ready Material no. 3 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
- finish
NO.4 Ready Material no. 4 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
- finish
NO.5 Ready Material no. 5 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
- finish
NO.6 Ready Material no. 6 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
- finish
NO.7 Ready Material no. 7 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
- finish
NO.8 Ready Material no. 8 ready state. In PFRand mode, flash - ready, slow flash - feed, long light
- finish
PROFIBUS Option
Words :
Continuous Output :
Continuous output weight and receive demand through TCP connection of TCP1025 port and 2025
port.
None : Disable Continuous output connection
TCP1 :Continuous output port 1025, 17 byte without checking
TCP2 : Continuous output port 1025, 18 bytes
UDP1 :2025 Continuous output port, 17 bytes without parity
UDP2 : Continuous output port 2025, 18 bytes
15
ID510 WEIGHING CONTROLER
Demand Input/output :
Connection via TCP 1024 port or UDP 2024 port, the instrument will receiveCommand input, and print
the weight data in the appropriate format,
None : Disable Demand Input/Output connection
DPRT-1: Demand Input and 1-line print display weight output,Refer to 4.2, 4.3
DPRT-1-C: Demand Input and 1-line to print display weight output with
CHK(Checksum) Refer to 4.2, 4.3
DPRT-2: Demand input and 1-line to print gross weight, tare weight and
net weight output Refer to 4.2, 4.3
DPRT-2-C: Demand input and 1-line to print gross weight, tare weight and
net weight output, with CHK(Checksum), Refer to 4.2, 4.3
DPRT-3: Demand input and 3-line to print gross weight, tare weight and
net weight output, Refer to 4.2, 4.3
DPRT-3-C: Demand input and 3-line to print gross weight, tare weight and
net weight output with CHK(Checksum),Refer to 4.2, 4.3
DPRT-5: Demand input and print display weight, date and time in lines
Refer to 4.3.4
MODBUS-TCP Server :
MODBUS-TCP Server is supported through TCP connection of port 502.
Refer to 4.5 MODEBUS-TCP
2.2.10 Maintenance
1) Scale interface: check the working status of sensor through internal code, check the
overload record
2) I/O
Input status can be viewed by displaying IN1~ IN8
View Inputs status, Light – ON, Dark – OFF
Control Output, press key: 1 to 9 and F2(10), F3(11), F4(12) to control output ON/OFF
16
ID510 WEIGHING CONTROLER
1) Page 1
Total Mode :
Disable: Turn off the accumulative function
Print : Print command accumulation allowed
Wt In : Feed accumulation of ingredients
Wt Out : Aggregate output of ingredients
In&Out : Feed/discharge accumulation of ingredients
Total Weight:
In Print command accumulative (Print) mode, the accumulative weight is the
accumulation of the displayed weight.
In Fill mode, the cumulative weight is the accumulation of the feeding weight.
Print : Print the accumulated information, refer to 4.6 ingredients record printing(FillPrt)
and 4.7 ingredients cumulative printing (TotalPrt)
Clear Total : Accumulated times and accumulated weights can be cleared by keyboard
operation, modbus-rtu and profibus-dp communication
2) Page 2 (cumulative)
17
ID510 WEIGHING CONTROLER
3) Page 3 (cumulative)
2.3.1 MEM
Input/see preset point target, zero tolerance, and auxiliary output threshold
2.3.2 STRT
Start the sequential batching only if the sequential batching mode and the batching number is greater than 1.
Start the recipe with the recipe management random ingredient mode (PFRand), and only allow to start a
certain ingredient after starting the recipe.
2.3.3 EMPTY
Start emptying
2.3.4 PAUS
18
ID510 WEIGHING CONTROLER
2.3.5 ABORT
The process of batching shall be suspended and the batching shall be completely terminated
2.3.6 RSUM
2.3.6 NEXT
In the process of manual sequence batching, when one kind of material feeding is finished, wait for
confirmation to proceed to the next material feeding
2.3.6 RST
Under PFRand, when the recipe is finished, the ingredient flow must be reset through the RST function key or
the stop cancel command of external input, and all completed output signals are cleared before starting the next
recipe.
19
ID510 WEIGHING CONTROLER
3.2 The process introduction
Support automatic zero clearing, automatic keyboard skinning and preset skinning before feeding of the
first material
Support before feeding 2, 3, 4, 5, 6, 7 and 8 kinds of materials, automatic peeling (net weight batching
mode)
In overshoot detection, overload is found and manual and automatic feeding is supported
After the completion of feeding, if the feeding mode is configured as a-dump or a-dose, the feeding will be
performed automatically
(M-dump or A-dump) must set the zero tolerance. When the weight is reduced to the zero tolerance range,
the emptying and discharging will be finished after the discharge delay
In the process of feeding, the PAUSA button or external PAUSA can be used to cancel the input of PAUSA
to suspend the ingredient. When the ingredient is suspended, the external start continues signal can be
input to resume the ingredient
20
ID510 WEIGHING CONTROLER
3.2.2 Random feed process(MRandom or ARandom)
Starting material no. 1, starting material no. 2, starting material no. 3, starting material no. 4, starting
material no. 5, starting material no. 6, starting material no. 7, starting material no. 8 signals to trigger the
addition of various materials
The feeding mode will not be detected after the end of feeding. The empty and the constant feeding will
always work in the manual mode (M-dump and M-dose).
In the process of feeding, the PAUSA button or external PAUSA can be used to cancel the input of PAUSA
to suspend the ingredient. When the ingredient is suspended, the input of the external corresponding
material starting signal will resume the ingredient. If the input is suspended again, the ingredient will be
completely terminated
21
ID510 WEIGHING CONTROLER
Automatic and manual blanking is supported for both incoming and fixed-value blanking.
Automatic feeding mode: if the feeding result is less than the set qualification range, the feeding can be realized
automatically
Manual feeding mode: if the feeding result is less than the set qualification range, the feeding can be realized
manually
When filling materials, by controlling the slow feeding output (ON/OFF) of various materials, the opening and
closing time must be set
Zero tolerance comparator, when the gross weight is less than or equal to set the upper limit value, the
output conduction
Upper limit comparator, the current gross weight is greater than or equal to the set upper limit value, and
the output is on
Lower limit comparator, current gross weight less than or equal to set upper limit value, output on
22
ID510 WEIGHING CONTROLER
Chapter 4.0 Serial communication application
COM1, COM2 and Ethernet all support continuous output. COM1 and COM2 baud rate are recommended to be no
less than 9600, otherwise it will affect the real-time performance of instrument weighing data.
Ethernet's continuous output ports are 1025COM1 and COM2, which support continuous output.
COM1 and COM2 baud rate are recommended to be no less than 9600, otherwise it will affect the
real-time performance of instrument weighing data.17- Byte Continuous output
Chara 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
cter
Data S S S S W0 W W2 W3 W4 W5 T T T T T T CR
T T T T 1 W0 W1 W2 W3 W4 W
X A B C 5
Instru
ctions A B- State C- Weight D - Tare weight E
Instructions:
A – The starting character STX,(ASCII 0x02)
STA, STB,STC – Please turn over B - State table
C – Display weight, net weight in the state of net weight;The gross weight is the gross weight
D – Tare Weight
** The weight in C and D has no symbol and decimal point. For the decimal point, please refer to the decimal
position definition section in the STA. For the symbol of weight, please refer to the symbol bit in STB
E – A carriage return(CR,ASCII 0x0D)
F – CHK, Is the sum of the first 17 characters low 7 bits, supported only when 18-byte continuous output
23
ID510 WEIGHING CONTROLER
B – State
Bit 0~2 Point position Bit 0,1,2 Point position
2 1 0 2 1 0
0 0 0 XXXX00 1 0 0 XXXX.XX
0 0 1 XXXXX0 1 0 1 XXX.XXX
0 1 0 XXXXXX 1 1 0 XX.XXXX
0 1 1 XXXXX.X 1 1 1 X.XXXXX
STA
Bit3,4
4 3 Dividing factor
0 1 X1
1 0 X2
1 1 X5
Bit-5 Always 1
Bit-6 Always 0
Bit-7 EVEN/ODD Check digit
Bit-0 0 – Gross state, 1- Net state
Bit-1 0 – Weight positive, 1 – Negative weight
Bit-2 0 – No overload, 1 – Upper or lower overloads
STB
Bit-3 0 – The steady state, 1 – Dynamic
Bit-4 0 – X10 Disable, 1- X10 Enable
Bit-5 0 – It's not in the range of zero tolerance, 1- Within the range of zero
tolerance
Bit-6 Always 1
Bit-7 EVEN/ODD Check digit
Bit-0 0
Bit-1 0
Bit-2 0
Bit-3 0
STC Bit-4 0
Bit-5 Always 1
Bit-6 0
Bit-7 EVEN/ODD Check digit
24
ID510 WEIGHING CONTROLER
Both COM1, COM2 and Ethernet support even command input/output, and Ethernet has a command input/output
port of 1024.
Ethernet also supports the following command inputs when working in continuous output mode, and
both COM1 and COM2 support even command input/output.Command input/output definition
Command Definition Response
C Clear None
T Tare None
D Preset Tare None
P Print Refer to 4.3 Print Output
Z Zero None
A Print the accumulated report See cumulative print reports
Character 1 2~9 10 11 ~ 12 13 14 15 16 17
Data STX DWT SP kg SP G/N CR CHK LF
Character 1~8 9 10 ~ 11 12 13 14 15
Data DWT SP kg SP G/N CR LF
4.3.2 Single - line printing of wool, leather, net weight output format
26
ID510 WEIGHING CONTROLER
4.4 MODBUS-RTU
Modbus-rtu protocol is supported in both COM1/COM2 and modbus-tcp Server is supported in Ethernet port with
port 502. Through modbus-rtu /TCP, the host can realize the following functions:
Support 03, 06, 16 command, does not support the bit operation, 16 command is only valid for 32
digits
32-bit integer and 32-bit floating point decoding order as follows: 3412
Read instrument displays weight, tare and instrument status
Parameter setting and remote scale calibration
MODBUS sampling frequency 50 Hz is the highest, it is suggested that the configuration when less than 30 Hz,
communication latency minimum 30 ms serial port baud rate must be above 9600, COM1 or COM2 support baud
rate is: 9600,19200,38400,57600, 115200(RS485 mode is not supported)
MODBUS-RTU / MODBUS-TCP
Address Instructions Action
attribute
40001 Display weight scale(1), signed 16-bit integer R(Only)
40002 Current gross weight scale(1) Signed 16-bit integers R(Only)
0 0 – Off center, 1 – In the center of the zero
1 0 – Gross, 1 – Net
2 0 – Steady state, 1 – Dynamic
40003 3 0 – No overloading, 1- Overload
4 0 – No underload, 1- Underload R(Only)
5 0 – Zero was not caught on startup,1-Boot to zero
6 0 – Weight data is invalid(2), 1 – Weight data available
7 0 – Not in Setup mode, 1 – In the Setup mode
8 0 – M7 Fast feed OFF, 1 – M7 Fast feedON
9~15 Dividing value Index 0-15
0 0 - Zero tolerance outside,1 – Within the range of zero
tolerance
1 0 - Auxiliary output one OFF, 1 - Auxiliary output one
ON
40004 2 0 - Auxiliary output two OFF, 1 - Auxiliary output two R(Only)
ON
3 0 - Run OFF, 1 - Run ON
4 0 – Suspended OFF, 1 - Suspended ON
5 0 – M1 Fast feed OFF, 1 – M1 Fast feed ON
6 0 – M1 Fine feed OFF, 1 – M1 Fine feed ON
7 0 – M2 Fast feed OFF, 1 – M2 Fast feed ON
8 0 – M2 Fine feed OFF, 1 – M2 Fine feed ON
9 0 – M3 Fast feed OFF, 1 – M3 Fast feed ON
10 0 – M3 Fine feed OFF 1 – M3 Fine feed ON
11 0 – M4 Fast feed OFF, 1 – M4 Fast feed ON
12 0 – M4 Fine feed OFF 1 – M4 Fine feed ON
13 0 – M5 Fast feed OFF, 1 – M5 Fast feed ON
14 0 – M5 Fine feed OFF 1 – M5 Fine feed ON
15 0 – M6 Fast feed OFF, 1 – M5 Fast feed ON
Next page
27
ID510 WEIGHING CONTROLER
MODBUS-RTU / MODBUS-TCP
Address Instructions Action
attribute
0 0 – M6 Fine Feed OFF 1 – M6 Fine Feed ON
1 0 – Error detection OFF, 1 – Error detection ON
2 0 – Feed end OFF, 1 – Feed end ON
40005 3 0 – Discharge delay OFF, 1 –Discharge delay ON R(Only)
4 0 – Empty feeding OFF, 1 – Empty feeding ON
5 0 – Quick discharge at fixed value OFF,1 –Quick discharge
at fixed value ON
6 0 – Slow feeding at set value OFF,1 –Slow feeding at set
value ON
7 0 – Discharging finished OFF, 1 –Discharging finished ON
8 0 – Alarm output OFF, 1 – Alarm output ON
9 0 - Super poor Under OFF, 1 –Super poor Under ON
10 0 - Super poor Over OFF, 1 –Super poor Over ON
11 0 – Ready OFF, 1 – Ready ON
12 0 – M7 Fast feed OFF, 1 – M7 Fast feed ON
13 0 – M7 Fine feed OFF 1 – M7 Fine feed ON
14 0 – M8 Fine feed OFF 1 – M8 Fine feed ON
15 0 – Keyboard unlock, 1 – Keyboard lock
0~7 Input 1 (IN1) ~ Input 8 (IN8)
40006 8~15 Output 1 (OUT1) ~ Output 8 (OUT8) R/W
0 0 1 Manual filling material
1 0 1 Keyboard reset
2 0 1 Tare
3 0 1 Clear
4 0 1 Print
5 0 1 Preset peeled
6 1 - The keyboard is prohibited, 0 – Keyboard allows
40007 R/W
7 0 1 Start/continue
8 0 1 Suspend/cancel
9 0 1 Empty feeding
10 0 1 Emptying or setting of blanks
11 0 1 The alarm reset
12
13
14
15 0 1 Active configuration parameter
40008 Zero tolerance difference degree(1) R/W
40009 Auxiliary comparator threshold of 1 degree(1) R/W
40010 Auxiliary comparator threshold of 2 degree(1) R/W
Next page
28
ID510 WEIGHING CONTROLER
MODBUS-RTU / MODBUS-TCP
Address Instructions Action
attribute
R/W
29
ID510 WEIGHING CONTROLER
MODBUS-RTU / MODBUS-TCP
Address Instructions Action
attribute
40046/47 Print the cumulative weight fraction or the total cumulative weight of R/W
feed ingredients (1) (3)
40048/49 Printing cumulative times or batching cumulative times (1) (3) R/W
40050/51 Aggregate degree of M1 (1) (3) (4) R(only)
40052/53 Aggregate degree of M2 (1) (3) (4) R(only)
40054/55 Aggregate degree of M3 (1) (3) (4) R(only)
40056/57 Aggregate degree of M4 (1) (3) (4) R(only)
40058/59 Aggregate degree of M5 (1) (3) (4) R(only)
40060/61 Aggregate degree of M6 (1) (3) (4) R(only)
40062/63 Cumulative weight fraction of batching output (1) (3) R/W
40064/65 Cumulative number of batching discharge (1) (3) R/W
40066 Start batching command R/W
0 1 Start M1 batching 0 2 Start M2 batching
0 3 Start M3 batching 0 4 Start M4 batching
0 5 Start M5 batching 0 6 Start M6 batching
0 7 Start M7 batching 0 8 Start M8 batching
40067 M1 This time feed ingredients weight point degree (1) R(only)
40068 M2 This time feed ingredients weight point degree (1) R(only)
40069 M3 This time feed ingredients weight point degree (1) R(only)
40070 M4 This time feed ingredients weight point degree (1) R(only)
40071 M5 This time feed ingredients weight point degree (1) R(only)
40072 M6 This time feed ingredients weight point degree (1) R(only)
40073 M7 Target value degree (1) R/W
40074 M7 Fine feeding point degree (1) R/W
40075 M7 Take the lead point degree (1) R/W
40076 M7 Down tolerance point degree (1) R/W
40077 M7 Up tolerance point degree (1) R/W
40078 M8 Target value degree (1) R/W
40079 M8 Fine feeding point degree (1) R/W
40080 M8 Take the lead point degree (1) R/W
40081 M8 Down tolerance point degree (1) R/W
40082 M8 Up tolerance point degree (1) R/W
40083/84 Aggregate degree of M7 (1) (3) (4) R(only)
40085/86 Aggregate degree of M8 (1) (3) (4) R(only)
40087 M7 This time feed ingredients weight point degree (1) R(only)
40088 M8 This time feed ingredients weight point degree (1) R(only)
Only 40001 is a 16-bit signed integer, the others are unsigned 16-bit integers
(2) – When in the configuration menu, the startup fails to capture the zero, the upper overload, and the lower
overload state, the weight data is invalid.
Where 40001 is 0, the host needs to test this bit to ensure the system safety and reliability
※ After the host sets the new dividing value through MODBUS, it must be
re-initialized through Bit15 in 40007
(3) – The accumulative weight and accumulative number are all 32-bit integers. When writing, it must be written
by 16 command. Only 0 is supported to clear the accumulative weight and cumulative number
(4) – The accumulations of materials including M1, M2, M3, M4, M5 and M6 can only be read and cannot be
written. When 40054/55(cumulative weight of ingredients) or 40056/57 (cumulative weight of ingredients) is
cleared, 40058/59(cumulative weight of M1), 40060/61(cumulative weight of M2), 40062/63(cumulative weight of
M3), 40064/65(cumulative weight of M4), 40066/67(cumulative weight of M5) will be cleared automatically.
40068/69(M6 accumulated weight)
31
ID510 WEIGHING CONTROLER
When the application of COM1, COM2 and Ethernet is configured as "FillPrt", the recording of batching is
supported, and the Ethernet printing output port is 1024.
The specific printing format is as follows:
The cumulative result will be automatically printed when the cumulative print command (A) is received
When the application of COM1, COM2 and Ethernet is configured as "TotalPrt", the accumulative result is
supported. When receiving the accumulative print command (A), the accumulative result will be automatically
printed out, and the Ethernet print outlet is 1024 port. The specific printing format is as follows:
Print Wt In Wt Out In&Out
Scale ID: 001 Scale ID: 001 Scale ID: 001 Scale ID: 001
Total Wt: 161306kg Total Wt In: 566.32kg Total Wt: 161306kg Total Wt In: 566.32kg
Total Count: 00020 Total In Count: 00010 Total Count: 00020 Total In Count: 00010
Date: 2011-06-13 M1 total weight: 51.56kg Date: 2011-06-13 M1 total weight: 51.56kg
Time: 09:26:38 M2 total weight: 100.56kg Time: 09:26:38 M2 total weight: 100.56kg
M3 total weight: 101.56kg M3 total weight: 101.56kg
M4 total weight: 70.56kg M4 total weight: 70.56kg
M5 total weight: 81.56kg M5 total weight: 81.56kg
M6 total weight: 101.56kg M6 total weight: 101.56kg
Date: 2011-06-13 Date: 2011-06-13
Time: 09:29:14 Time: 09:29:14
32
ID510 WEIGHING CONTROLER
5.1 PROFIBUS-DP
Integer format: in the menu, the configuration word length is set to 2-w. In this mode, the configuration
word length is 2W input / 2W output.
The meter weight data read by the host is converted to integer format data, such as:
Suppose the meter score is 0.1
PLC operation The actual weight PLC Read the PLC Write
data data
PLC Read display weight 300.5 3005
PLC Read the preset point 100.7 1007
target value
PLC Set the target value of 100.7 1007
the preset point
Floating point number format: in the menu, the configuration word length is set to 6-w. In this mode, the
configuration word length is 6W input / 6W output.
The meter weight data read by the host is the actual weight in floating point
format.
33
ID510 WEIGHING CONTROLER
Select floating point number format, configuration word length is 6 words (6W)
34
ID510 WEIGHING CONTROLER
35
ID510 WEIGHING CONTROLER
15 14 1 12 11 10 9 8 7 6 5 4 3 2 1 0
3
W0 Sto Start Lo Em Sta Re
p the Pr ck pty rt set S S S S S S S S R/
can Cont in Ke the the 7 6 5 4 3 2 1 0 W
cel inue t yb dis ala
to oar cha rm
d rgi
ng
W1 Value_0
W2 Value_1
The Pre
O O O O O O O O O O O cum C set Ta Zer
W3 12 11 1 9 8 7 6 5 4 3 2 ulati l leat re o
0 ve e her
prin a for
t r peel
ing
W4
W5
The batch state is output through combination of W5 and batch output state
2
Batch output state 2
Bit0 M2 Fast Feed
Bit1 M2 Fine Feed
Bit2 M3 Fast Feed
Bit3 M3 Fine Feed
Bit4 M4 Fast Feed
Bit5 M4 Fine Feed
Bit6 M5 Fast Feed
Bit7 M5 Fine Feed
Bit8 M6 Fast Feed
Bit9 M6 Fine Feed
Bit10 Run
Bit11 Suspend
Bit12 Out-of-tolerance alarm
Bit13 Under the super poor
Bit14 The super poor
Bit15 Material 1
Bit16 Material 2
Bit17 Material 3
Bit18 Material 4
Bit19 Material 5
Bit20 Material 6
Bit21 Ready
Bit22 Feed end
Bit23 Discharging finished
Bit24 Alarm output
Bit25 M7 Fast Feed
Bit26 M7 Fine Feed
Bit27 M8 Fast Feed
Bit28 M8 Fine Feed
Bit29 Material 7
Bit30 Material 8
38
ID510 WEIGHING CONTROLER
39
ID510 WEIGHING CONTROLER
40
ID510 WEIGHING CONTROLER
41
ID510 WEIGHING CONTROLER
PROFIBUS-DP Note
When you are in the configuration menu, the machine does not catch zero, upper overload, and lower overload, the
weight data is invalid at this time, such as the display weight, gross weight and net weight will be 0, and the host
needs to detect this bit to ensure the system safety and reliability
(1): You can only clear by writing 0
(2): M1-m6 accumulative weight will be automatically cleared when cleaning the cumulative
number of ingredients or accumulated weight
42
ID510 WEIGHING CONTROLER
When the scale is applied, the printing storage is configured as "Enable" and loaded with SD memory card. When
the meter prints weight data, it will automatically save the printing Record to the high-capacity SD card. The file
name is fixed as Record.
The specific format is as follows:
Record Scale Date Time Gross Tare Net Unit
1 1 11/8/2010 8:05:58 5.04 0.04 5.00 Kg
2 1 11/8/2010 8:15:03 6.06 0.06 6.00 kg
3 1 11/8/2010 8:20:11 9.04 0.04 9.00 kg
4 1 11/8/2010 8:26:27 7.94 0.06 7.88 kg
5 1 11/8/2010 9:06:30 10.94 0.90 10.04 kg
6 1 11/8/2010 9:18:38 9.94 0.80 9.14 kg
7 1 11/8/2010 9:34:44 8.92 0.08 8.84 kg
When the records in the ingredients are configured as "SaveSD" or "Prt&Save" and loaded into the SD card, the
specific records of the ingredients will be automatically saved to the SD card in the process of ingredients. The file
name is fixed as Fill.XLS, which can be opened directly under EXCEL.
Scale Action Date Time Target +Tol -Tol Actual Unit
1 M1 11/8/2010 8:05:58 5.00 0.05 0.05 5.02 Kg
1 M2 11/8/2010 8:15:03 10.00 0.05 0.05 9.98 kg
1 M3 11/8/2010 8:20:11 9.00 0.05 0.05 9.02 kg
1 Empty 11/8/2010 10:34:44 49.96 kg
1 M1Tol 11/8/2010 20:34:44 51.56 kg
1 M2Tol 11/8/2010 20:34:44 100.56 kg
1 M3Tol 11/8/2010 20:34:44 101.56 kg
1 InTol 11/8/2010 20:34:44 566.32 kg
1 OutTol 11/8/2010 20:34:44 566.32 kg
1 M1 12/8/2010 11:05:58 5.00 0.05 0.05 5.02 Kg
1 M2 12/8/2010 11:15:03 10.00 0.05 0.05 9.98 kg
1 M3 12/8/2010 11:20:11 9.00 0.05 0.05 9.02 kg
1 Dose 12/8/2010 11:21:11 3.00 0.04 0.04 3.01 Kg
1 Dose 12/8/2010 11:30:11 3.00 0.04 0.04 2.99 Kg
Attention:
After the SD card is inserted, the meter must be reenergized to ensure normal operation, and the SD card
writes a protection mark.If the SD card has no write protection, but "SD Fail!"The SD card is damaged
43
ID510 WEIGHING CONTROLER
Chapter 7.0 Free weight calibration
ID510 supports the calibration function by entering the sensitivity of the sensor, thus eliminating the difficulty
that cannot be calibrated without weight, or all kinds of trouble caused by requiring large weight. The following
calibration process:
The instrument is equipped with weight - free calibration
The actual zero sensitivity of the input sensor (the attenuation coefficient of the sensor after it passes
through the junction box should be considered)
Zero calibration
At the end of calibration, place the object with known weight to test the weight. If necessary, adjust
the attenuation coefficient according to the test results.
Zero calibration
44
ID510 WEIGHING CONTROLER
When the system fails and the instrument needs to be replaced, and the calibration is not available, it can be
handled through the emergency recovery correction result.
On the scale interface maintenance interface, page down to the correction result maintenance interface. If the
correction mode is "2-p", only the range correction maintenance can be seen.If it is "3-p", you can see the range
correction maintenance and linear correction maintenance;The result of emergency recovery is achieved by input
zero internal code, range correction weight, range correction result, linear correction weight and linear correction
result.
Attention: the emergency recovery results can only be used in the case that there is no way to calibrate,
and the scale is not guaranteed to be absolutely correct. Once the conditions are mature, it must be
re-calibrated before use.
There are four switches on the motherboard, and the specific functions are defined as follows
1 2 3 4
ON - No calibration ON – Allow key lock ON – Load factory ON – Enter the update
(lead seal) OFF – Disable key defaults program mode, which
OFF – Allow the lock must be OFF for
calibration normal use
45
ID510 WEIGHING CONTROLER
* Attention: that the output current of each OUT should not exceed 100mA!
A 24VDC can be 12 to 30VDC, with a typical reliable connection of 12VDC or
24VDC
46
ID510 WEIGHING CONTROLER
Wrong starting zero, too heavy Wrong zero, too small weight Over Capacity
47