Professional Documents
Culture Documents
MAINTENANCE MANUAL
GENERAL CONSIDERATION
All operating and maintenance instructions and recommendations described in this manual must be respected.
The Manufacturer recommends regular cleaning and maintenance to keep the device in the best conditions and
achieve optimal results. Training of the user is of particular importance, both as regards the use of the equipment,
and for maintenance and compliance with the operating procedures and all the safety standards indicated in this
manual. In any case, the Manufacturer is always available for any clarification or further information.
I I S. .A. reserves the right to update and improve the products without obligation of notice and notification.
I I S. .A. is not liable for the use of data that may have been modified.
GLOSSARY SYMBOLS
S M
Prohibition
11
1 2 3
4 5 6
7 8 9 10
1.4. WARRANTY
1.4.1. VALIDITY
Devices are guaranteed for twelve months after the delivery date unless otherwise agreed in the purchasing
order.
1.5. CONDITIONS
The warranty covers only manufacturing or manufacturing defects and does not cover damage due to wear,
dirt, modifications or variations of the instrument not authorized by Isoil Impianti S.p.A.. The replacement under
warranty of the defective parts is made ex works.
Some examples of possible hazards caused by non-compliance with these instructions are:
• Failure of the system and/or of its components.
• Damages to people caused by exposition to electrical, mechanical or chemical hazard.
• Pollution of the environment due to the leaking of hazardous substances.
Carefully follow the safety instructions described in this manual; in case of doubt, please contact the Manufacturer.
P P E
Carefully check that the system and its parts are not under pressure before disassembling
the equipment or its accessories for inspection, maintenance or replacement of
components.
Also make sure that all electric or electronic part, if any, are disconnected from their
power supply.
In potentially explosive atmospheres only use explosion-proof components provided with specific identification
and suitable for such application.
According to EU Directive 2014/34/EU (ATEX) the Ex protection is valid only if the device is used in compliance with
the instructions at par.3.1 (intended use).
Never exceed (above or below) the limits indicated on the device name plate.
M A T T C
U C E I 80079-36 D
100 °C T4
70 °C T5
50 °C T6
• Use it within the technical limits indicated on the contract and/or in this manual,
• Follow the procedures mentioned in this manual,
• Do not exceed maximum pressure and flow rate indicated on the label,
• Schedule ordinary maintenance and perform extraordinary maintenance promptly if necessary.
1 Body
2 Diaphragm assembly
3 Cover
Fig. 1
MULTISTEP
The weights and dimensions are purely indicative and refer to basic valve configuration.
D W ( ) F
3” 40 ANSI 150 RF
4” 65 ANSI 300 RF
During shipment, the electrical and process connections are protected by special plugs.
These must not be removed until installation on the pipeline.
• Root valves installation at the inlet and outlet is recommended to ease servicing operations.
The device is not suitable for installation in plants where vibrations are transmitted along
the pipes. Under these conditions, the customer must provide for an adequate system of
staffing of the pipes and, if necessary, a damping system.
Before installation, cleaning of the plant pipings is suggested to remove dirt and foreign
particles, especially in the case of a new plant.
Leave enough space around the device to easily perform commissioning, servicing and
disassembly.
4.3. INSTALLATION
The installation must be performed by qualified personnel who have read and understood
the User and Maintenance manual.
Make sure that the flow direction through the device is the same indicated by the arrow on the
device itself.
The device can be mounted both horizontally and vertically. See the drawings below.
To ensure the seal between flange and pipe a gasket must be inserted between the flanges. This gasket must be
suitable to the product.
• Check that screw connections between the devices are properly tightened.
• Piping should not exercise strain on the device: supports should be provided only on adjacent pipes.
When starting the equipment for the first time, slowly fill in the measuring/transfer line.
N.O N.C
B
A B
P . D D C
1 Valvola base Base valve /
2 Filtro Strainer /
4 Valvola di regolazione flusso Control valve 80VA4039
6 Valvola di ritegno Non return valve 80VA2091
10 Elettrovalvola Normalmente aperta (N.O.) Solenoid valve (N.O) /
11 Elettrovalvola Normalmente chiusa (N.C.) Solenoid valve (N.C.) /
The automatic valve is controlled by two solenoid valves. The electronic counter drives the opening and closing
of the solenoid valves causing a step by step opening or closing of the diaphragm.
4.5. ADJUSTMENT
4.5.1. VALVE CLOSURE
Energizing of the solenoid NC (pos. 11) allows opening of the main valve. The pressure in the chamber above the
diaphragm is completely released and the main valve opens.
The opening speed of the main valve can be increased or decreased by acting on the control valve. Increasing
the opening of the control valve (pos. 5B) increases the amount of product discharged from the chamber above
the diaphragm, speeding up the opening of the main valve. On the contrary, by decreasing the opening of the
control valve (pos. 5B), the amount of product discharged from the chamber above the diaphragm decreases,
therefore slowing down the opening of the main valve.
Adjustment of the opening speed is also possible thought an electronic counter, acting on the closing time of the
solenoid valves.
The closing speed of the main valve can be increased / decreased by acting on the flow control valve.
Increasing the opening of the control valve (pos. 5A) increases the amount of product input of the chamber above
the diaphragm, thus speeding the closure of the main valve. On the contrary, by decreasing the opening of the
control valve (pos. 5A) the amount of product input to the chamber above the diaphragm decreases thus slowing
down the closing of the main valve.
It is possible to adjust the opening speed also by means of an electronic counter acting on the closing time of the
solenoid valves.
The electronic counter acquires the data flow from the pd meter, compares it with the preset, and, by acting
appropriately on the solenoid valves (pos.10 and 11), it regulates the flow rate to the desired value.
Once the desired flow rate is achieved, the solenoid valves are operated to maintain the desired flow rate.
Example of operation diagram controlled by VEGA II
Flow rate
High flow
Low flow
The pumps must be sized properly in order to reach the desired flow.
When in high flow, differential pressure between inlet and outlet of the main valve should be
higher than 40 kPa. Otherwise the valve could close late thus causing a product supply higher
than the preset value.
The non-return function is performed by two non-return valves (pos.6) added to the hydraulic control circuit
(dwg.4571 only).
Maintenance personnel must be well trained and must perform all work carefully, adhering
to all necessary safety measures. Non-authorized personnel must not stay in the working
area during maintenance operations.
All those operations that the operator must carry out, preventively, to ensure
Ordinary maintenance
correct operation over time
Extraordinary maintenance All those operations that the operator must carry out when the device needs it.
Before removing the device from pipeline for any repair, check the Troubleshooting table in this manual.
Maintenance operations can be performed without removing the device from the line.
All substances, lubricants and / or other products used during maintenance and any liquids
discharged from the equipment must be disposed of by observing the relevant standards.
Servicing intervals may greatly vary depending on the product, the environmental conditions, the actual workload,
the operating conditions etc.. Isoil Impianti S.p.A. accepts no responsibility for the breach of the rules of use
mentioned in this manual.
D 1M 6M 12 M 24 M
Base valve CV
Hydraulic circuit CV
For all extraordinary maintenance needed after a failure and/or a fault preventing normal operation of the system,
please contact the Customer Care Department of Isoil Impianti S.p.A.
5.3. DISASSEMBLY
For maintenance operations and spare parts purchase, see attached drawings.
S A I
5.5. STORAGE
If the device and its accessories are not going to be used for a long time, the following precautions must be used:
In case of gaskets storage, remember that they need to be preserved from damages caused by humidity and/or
sunlight.
For correct maintenance use only spare parts recommended by Isoil impianti S.P.A.
Isoil Impianti S.p.A. is not liable for any problem resulting from the use of non original spare parts.
F P C C
Valve does not open; pressure is not Upstream isolation valve closed Open the isolation valve.
registered at valve inlet
Pump is not working / switched off Start the pump or check if it works correctly.
a) Differential pressure across the valve is
Check that pressure is no less than 30 kPa.
Valve does not open although pressure too low
is registered at the valve inlet Check that the electrical connections are made
b) No electrical power at solenoid valves
correctly. If not, connect them properly.
c) NC solenoid valve is defective: it does not
operate upon energizing.
Pressure inside the cover chamber is drained
upon energizing of the NC solenoid valve. Replace the NC solenoid valve.
A faulty NC solenoid valve does not allow
draining and therefore the valve does not
open
d) Diaphragm is ruptured Replace the diaphragm.
b) Stem movement in guide is not flowing If the valve is in vertical position, see § 4.3.
Valve does not close when a) The seal (pos.16) is damaged Replace the seal (pos.16), if damaged.
de-energizing both solenoids and it’s
leaking
b) Non return valve at the outlet of hydraulic
control circuit is damaged. Product flows out Replace the non return valve.
of the cover chamber thus causing leakage
22
27 25
26
19
20 20
7 6
21
21
2 24
12
11
10
8
14
*
9
16
*
3 4
13
18
*
5 15
*
17
28
24 25
23
21
22
19
19
20
6
7 20
12
11
10
8
* 14
* 16
4
*
18 13
5
3
17 15
*
1
27
E B / P N N
E G
C T O ( . .) C ( . .)
Brass and
Ex-db mb M20x 1.5 FKM 230VAC 80EL1174 80EL1213
stainless steel
Brass and
Ex-db mb M20x 1.5 FKM 115 VAC 80EL1219 80EL1220
stainless steel
Brass and
Ex-db mb M20x 1.5 FKM 24 VDC 80EL1176 80EL1214
stainless steel
The electrical connections must be made according to the national regulations of the
installation site.
Cut off the power and release the pressure of the valve before any repair.
S A I
)L[LQJ6FUHZ9LWHGLILVVDJJLR
FUHZ 9LWH
H GL ILVVDJ
Insert the cable in the cable gland components
3 (pos.12).
Take care not to damage the outside diameter
of the cable (zone C) and the inside dia. of the
joint disc (zone D) which build the flameproof =RQH=RQDD&
&
joint when assembled.
Connect the end of the wires 2P+E (max. 2.5 mm2) to the terminals (pos.11). That fuse can be
4
exchanged when needed (s. table 1). In this case, the pins have to be shorted to 12.6 mm.
Push the cable into the housing if necessary, until approx. 5 mm cable sheath has entered in the
5
housing. Tighten the cable gland screw M20x1.5 with a torque of 2-3 Nm.
Refit the cover (pos.9) with correct positioned original sealing ring (pos.10) by help of the 4 screws
(pos.8) and torque of 1.5 to 2Nm.
6
Before doing this, verify zone B of the housing and zone A of the cover on damages or dirts. If
necessary carefully clean the surfaces.
7 Assure the correct mounting of the central O-rings (pos.14 and pos.15).
N : All coils have a built in varistor and a thermal fuse (see table 1 below).
T 1
C C
C . S R . C . S R .
F ( A) F ( A)
80EL1174 495905F4 250 80EL1219 495905E5 400
0$5521(
%52:1
)86,%,/(
)86(
$/,0(17$=,21(
32:(56833/<
%/8
%/8(
0(66$$7(55$
Zd,/E''ZKhE