PT JAYA BETON INDONESIA

Analisa Kuat Tekan Beton Setelah Kalibrasi


PLANT SADANG

Kalibrasi Alat Umur Pembacaan Alat Strength Test Pembacaan Alat Setelah Kalibrasi
Penunjuk Dial Koreksi (Days) i (Kn) ii (Kn) iii (Kn) I (Kn) ii (Kn) iii (Kn)
No Standard
Xi Yi 7 471 433 405 472.93 434.81 406.72
1 0 0 0 14
2 200 200.1 0.1 28
3 400 400.4 0.4
4 600 601.5 1.5
5 800 797.3 -2.7 Luas Penampang Cyl 10 x 20 = 78.5 Cm2 = 7850 mm2
6 1000 998.4 -1.6 Volume Sample = 1570 Cm 3
= 0.00157 m3
7 1200 1198.3 -1.7
8 1400 1399.6 -0.4 Umur Pembacaan Alat Strength Test Pembacaan Alat Setelah Kalibrasi
persamaan Polinominal (Days) i (Kg/cm2) ii (Kg/cm2) iii (Kg/cm2) i (Kg/cm2) ii (Kg/cm2) iii (Kg/cm2)
Y= 0.000001 X2 - 0.002393x + 0.583333 7 73727 67779 63396 74030 68062 63665
14
28
Grafik Compressive Strength Test Machine
2 Umur Berat Sample Berat Isi Sample
1.5 1.5
(Days) Gr Gr Gr Ton/m3 Ton/m3 Ton/m3
1 7 3956 3955 3926 2.520 2.519 2.501
0.5 0.39999999999997
f(x) = 7.73809523809474E-07 x² − 0.00239285714285713 x + 0.583333333333327
14
0.09999999999999 7
00 43 - 28
0 200 400 600 800 1000 1200 1400 1600
-0.5 0.40000000000009
1
-1
-1.5 -1.60000000000002
-1.70000000000005
-2
-2.5
-2.70000000000005
-3

Column D Polynomial (Column D)

Persamaan Polinominal
Y= 0.000001 X2 - 0.002393x + 0.583333
Keterangan :
Y = Bacaan Alat Setelah Kalibrasi
X = Bacaan Langusng Dial