You are on page 1of 7

909024

research-article2020
JVA0010.1177/1129729820909024The Journal of Vascular AccessElli et al.

Original research article


JVA The Journal of
Vascular Access

The Journal of Vascular Access

Changing the syringe pump: A challenging


1­–7
© The Author(s) 2020
Article reuse guidelines:
procedure in critically ill patients sagepub.com/journals-permissions
https://doi.org/10.1177/1129729820909024
DOI: 10.1177/1129729820909024
journals.sagepub.com/home/jva

Stefano Elli1 , Elisa Mattiussi2, Stefano Bambi3 ,


Serena Tupputi1, Salvatore San Fratello4, Angela De Nunzio4,
Salvatore D’Auria5, Roberto Rona1, Roberto Fumagalli6
and Alberto Lucchini1

Abstract
Introduction: In the literature, the change of a syringe pump is described as a dangerous situation, especially in the
case of vasoactive drug administration.
Methods: Different variables have been studied (central venous pressure, pump displacement in relation to the patient
position, utilization of a stopcock, or a neutral displacement needle-free connector between the syringe and the infusion
tubing) to understand their influence on medication administration in terms of backflow or bolus creation when changing
the syringe.
Results: We performed 576 measurements with different combinations. With respect to all the observations, in
comparison with “time zero,” we found the following differences expressed in microliters: 0 (±1) at the plunger opening;
0 (±3) at the syringe extraction from the pump; 0 (±7) at the syringe disconnection from the infusion tubing; 0 (±11)
at the syringe reconnection to the infusion tubing; 1 (±7) at the syringe insertion in the pump; 3 (±23) at the plunger
closing; 8 (±33) at the stabilization at the maneuver end.
Conclusion: The syringe change can be a very critical moment given different influencing variables. Syringe pump
position, displaced higher than the patient level, always generates a medication bolus that is higher at the lowering of
the central venous pressure value. The presence of a neutral displacement needle-free connector reduces the incidence
of boluses. When the pump is placed at the patient level, the presence of neutral displacement needle-free connector
reduces the establishment of boluses, even in a central venous pressure of −5 mmHg simulations.

Keywords
Backflow, bolus, changeover, infusion, inotrope infusion, needle-free connector, syringe pump vasoactive drug
pumping, vasoactive drug administration

Date received: 1 July 2019; accepted: 25 January 2020

Introduction 1
 eneral Intensive Care Unit, Emergency Department – ASST Monza,
G
San Gerardo Hospital, University of Milano-Bicocca, Monza, Italy
Background and rationale 2
School of Nursing, Department of Medical and Biological Sciences,
Udine University, Udine, Italy
Intensive care unit (ICU) patients often require adminis- 3
Medical & Surgical Intensive Care Unit, Careggi University Hospital,
tration of intravenous therapies. A concentrated continu- Florence, Italy
ous infusion of medications, rather than a diluted 4
General Intensive Care Unit, Azienda Ospedaliera Universitaria di
concentration, is frequently used because of fluid restric- Parma, Parma, Italy
5
tions in critically ill patients.1 In particular, vasoactive General Intensive Care Unit, Istituto Auxologico, Milano, Italy
6
Department of Anesthesia and Intensive Care Medicine, Niguarda Ca’
drugs are the mainstay of hemodynamic management of Granda, University Hospital of Milano-Bicocca, Milano, Italy
critically ill patients.2,3 Medical devices, such as syringe
pumps, are commonly used for the continuous infusion of Corresponding author:
Alberto Lucchini, General Intensive Care Unit, Emergency Department
highly concentrated drugs. The syringe pumps, thanks to – ASST Monza, San Gerardo Hospital, University of Milano-Bicocca, Via
technological innovation, are able to deliver very precise Pergolesi 33, 20900 Monza, Italy.
drug doses, in the order of 0.1 mL/h. However, the current Emails: a.lucchini@asst-monza.it; alberto.lucchini@unimib.it
2 The Journal of Vascular Access 00(0)

Figure 1.  Change of syringe in a patient with continuous norepinephrine infusion.


sAP: systolic arterial blood pressure; mAP: mean arterial blood pressure; dAP: diastolic arterial blood pressure.

literature identifies several factors that can contribute to pressure (CVP). CVP values can vary considerably depend-
reducing the stability of the administration: device’s tech- ing on the type of ventilation, blood volume, and many
nical features, catheter patency, start-up delay, mechani- other factors; differences of 10–15 mmHg (15–20 cmH2O)
cal occlusion of the infusion tubing, siphoning, users’ can be found between a patient in mechanical ventilation
knowledge, and so on.1 These factors are capable of modi- and one in spontaneous breathing with severe respiratory
fying the stability of the infusion and determining hemo- distress.8,9 If the pressure of the infusion tubing is higher
dynamic imbalances. Various precautions are suggested to than the “blood pressure,” the solution in the tubing will be
avoid changes in the current flow during infusion: avoid pushed toward the patient, thus causing a bolus of the medi-
vertical displacements of the pump, use of needle-free cation with an immediate overdose. If the pressure in the
connectors (NFC), use of two parallel pumps, and avoid infusion tubing is lower than the “blood pressure,” blood
infusion of different drugs in the same tubing.1,4,5 The will flow back into the infusion tubing, thus stopping the
greatest risk of altering the stability of the infusion arises infusion of the medication and, perhaps, a delay in stabiliz-
from the replacement of syringe. In every ICU, syringe ing the infusion. It would be desirable to reduce these risks
replacement is a common practice performed by many with an optimal use of the available technology and the use
people. The risk of hemodynamic imbalances or other of standardized change procedures.
problems is always present.6,7 During the change, the con-
tinuity and integrity of infusion tubing and protection
against the interactions between blood pressure and tub- Objective
ing pressure are lost. Figure 1 shows a modification The purpose of this experimental study is to analyze the
induced by syringe replacing in a patient with norepineph- effects of syringe replacement on the continuity of drug
rine infusion. In this case, the infusion of 0.10 µg/kg/min infusion, based on the absence/presence of different
(infusion rate of 5 mL/h) of norepinephrine required 8 min syringe infusion set-ups.
to recover the previous arterial pressure values.
We can call “blood pressure” the pressure of the blood-
stream at the tip of the catheter, which causes resistance to Methods
infusion. We can call “tubing pressure” the pressure
induced by the height of the liquid column in the infusion
Study design and setting
tubing in relation to the patient. In ICU wards, drugs are We conducted an “in vitro experimental study” during
generally administrated via a central venous catheter, so the August/September 2017 in a laboratory of an ICU in a
“blood pressure” is generally similar to central venous North Italian University Hospital. We assembled a
Elli et al. 3

Figure 2.  Experimental set-up.

simulation system for measuring fluid displacement into the variables studied are explained in Figure 2. An air bub-
the infusion tubing during syringe replacement. The simu- ble was inserted in the tubing at the “displacement measur-
lation system was assembled with the following devices: ing area” and used as reference for measurements10. The
two different types of syringe pump (Pilot A2®; Fresenius displacement of the air bubble was measured during each
Kabi AG, Bad Homburg, Germany; Injectomat Agilia®, of the six steps of the syringe change after stopping of infu-
Fresenius Kabi AG), a 50-mL syringe (BD Perfusion 50 mL sion: plunger opening (S0), syringe extraction from the
Syringe; Becton Dickinson, Franklin Lakes, NJ, USA), a pump (S1), disconnection from infusion tubing (S2), recon-
250-cm-long infusion tubing (ICU Medical, San Clemente, nection to the infusion tubing (S3), syringe insertion in the
CA, USA), three-way stopcock (Bicakcilar, Istanbul, pump (S4), plunger closing (S5), and stabilization at the
Turkey), hemodynamic pressure transducer (TruWave; end of maneuver (S6). Four different operators (nurses with
Edwards Lifesciences, Irvine CA, USA), multiparametric up to 5 years of experience in ICU) carried out the “six
monitor (Draeger SC 9000; Draeger Medical, Telford, steps” of standardized change procedure in all possible
Pennsylvania USA), two different types of neutral dis- combinations of the studied variables. During each step, the
placement NFC (Neutraclear®; Becton Dickinson, movements of the air bubble in the “displacement measur-
Bionecteur®; Vygon, Ecouen, France) with different perfor- ing area” were manually noted, as reported in Figure 3. The
mances,10,11 and 500 mL of colored normal saline connected measures, expressed in centimeters, were then converted
to an Infusion Set (Rays SpA, Osimo, Italy). Variables stud- into milliliter of fluid displacement in the infusion tubing.
ied were the following: type of pump used, pump highness A positive displacement was identified as “bolus” and a
in relation to the simulated patient,12,13 three different CVP negative displacement was identified as “backflow” (reflux
simulated values,8,9 connection between the syringe and of blood into the catheter).
the infusion tubing, and operator. The pump positions
considered in relation to the patient were 50 cm above the
Statistical analysis
position of the heart, patient in supine position (heart level)
and 32 cm below the patient, with respect to heart level. Statistical analysis was performed using Statistical
Simulated CVP values analyzed were +15, 0, and Package for the Social Sciences, version 22.0, (SPSS Inc.,
−5 mmHg. How the simulation system was assembled and Chicago, IL, USA). Continuous variables were expressed
4 The Journal of Vascular Access 00(0)

situation, the presence of a neutral connector (Bionecteur)


reduces the incidence of boluses. When the pump is dis-
placed at the patient level, the presence of neutral displace-
ment NFC reduces the establishment of boluses, even in
simulations considering a CVP of −5 mmHg. Finally, if the
pump is displaced lower than the patient level, a backflow
is established, within average volumes of 50 µL in the
absence of a stopcock or a neutral displacement NFC and
with a CVP value of 10 mmHg. In this case, the device
that mainly reduces the occurrence of backflow is the
stopcock. The influence given by the variable “operator”
results in statistical significance only at the reconnection to
the infusion tubing (p = 0.044) and at the plunger closing
(p = 0.034). In Figure 5, it is possible to appreciate that the
average backflow never results in statistical significance in
Figure 3.  Displacement measuring area. relation to the pump used, even if there are different values
for each step.
as mean ± SD or median (interquartile range). One-way
repeated-measures analysis of variance (rmANOVA) was Discussion
used to evaluate the differences at the six steps of the
syringe change (plunger opening, syringe extraction from In the past research, intravascular drug administration
the pump, disconnection from infusion tubing, reconnec- focused mainly on medication concentration and methods
tion to the infusion tubing, syringe insertion in the pump, of administration. In the literature, the major data concern
plunger closing, and stabilization at the end of maneuver), the administration of vasoactive drugs.3,4 However, there is
with only possible combinations of the six variables inves- a scarce discussion about syringe exchange events and the
tigated (NFC, CVP values, syringe pump position, opera- subsequent effects on patient’s vital signs. In effect, many
tor, type of syringe pump). A value of p < 0.05 was intensive care practitioners have experienced the storm that
considered statistically significant. can develop following the procedure.14,15 The main results
of our study emphasize that exchange of the syringe in the
pump, especially during vasoactive drug handling, can be
Results
dangerous for the stability of patients’ vital signs. Nurses
We performed 576 measurements with different combina- should always consider the existing correlation between the
tions of the investigated variables. With respect to all the patient’s CVP and the vertical displacement of the pump to
observations, in comparison with “time zero,” we found choose the better pump position, when deciding definite
the following differences expressed in microliters: 0 (±1) pump displacement. The side effects due to the association
at the plunger opening; 0 (±3) at the syringe extraction of the two considered variables are maximized when any
from the pump; 0 (±7) at the syringe disconnection from dispositive is placed between the infusive pump and the
the infusion tubing; 0 (±11) at the syringe reconnection to infusion tubing. Our data suggest that, without a disposi-
the infusion tubing; 1 (±7) at the syringe insertion in the tive, a maximal backflow can occur in the intraluminal
pump; 3 (±23) at the plunger closing; 8 (±33) at the stabi- catheter tract (thus not visible for the operator) when we
lization at the maneuver end. All these maneuvers sum- consider a high CVP, and the pump is lower than the patient
marize all the modifications given by the 576 analyzed position. The registered backflow can vary between
variables. If data are analyzed considering the simulated −54 ± 82 µL. This situation can be represented by the fol-
values of CVP, the absence/presence of a stopcock between lowing clinical scenario: an ambulance transfer of an 80-kg
the syringe and the infusion tubing, and the vertical posi- patient, the trunk of whom is mobilized at 30° with the
tion of the pump, they can demonstrate the influence of syringe pumps displaced at the leg level; continuous nor-
each single variable in determining boluses or backflow. In epinephrine infusion within a dilution of 4 mg in 50 mL for
Table 1, data are represented in relation to the seven inves- a 4 mL/h speed (0.004 µg/kg/min). In the case of a −54 µL
tigated phases combined with the previous considered backflow, a 7-min time is necessary for the infusion to
variables. In Figure 4, the average value at the end of “wash” the backflow until the medication administration is
maneuver is graphically represented, in relation to the dif- restored. Alternatively, a manual bolus equal to 0.028 µg
ferent devices and the different experimental set-ups. It is (0.004 µg/kg/min × 7 min) is necessary to restore the medi-
appreciable how a position of the pump, displaced higher cation administration. If we still consider any dispositive
than the patient level, always generates a medication bolus between the syringe and the infusion tubing, an opposite
that is higher at the lowering of the CVP value. In this situation can happen in the considered patient with a low
Elli et al. 5

Table 1.  Fluid displacement with different devices and set-up combinations.

Syringe pump Simulated Connector/ S0 S1 S2 S3 S4 S5 S6


position CVP, mmHg stopcock
High (+50 cm 10 No 0 ± 1 1 ± 2 1 ± 2 3 ± 4 14 ± 10 20 ± 15 26 ± 20
up to Neutraclear 0 ± 0 0 ± 2 0 ± 2 3 ± 3 8 ± 4 14 ± 9 21 ± 17
simulated Bionecteur 0 ± 0 0 ± 1 0 ± 1 0 ± 1 0 ± 1 1 ± 2 2 ± 4
patient) Stopcock 0 ± 0 1 ± 2 1 ± 2 1 ± 2 1 ± 2 4 ± 8 33 ± 24
0 No 0 ± 0 1 ± 1 1 ± 2 2 ± 4 14 ± 14 27 ± 31 33 ± 34
Neutraclear 0 ± 1 1 ± 2 0 ± 3 7 ± 4 17 ± 8 28 ± 13 42 ± 28
Bionecteur 0 ± 1 1 ± 1 1 ± 2 3 ± 4 6 ± 7 11 ± 15 16 ± 27
Stopcock 0 ± 2 2 ± 3 2 ± 3 2 ± 3 2 ± 3 4 ± 7 28 ± 22
–5 No 0 ± 1 3 ± 5 4 ± 7 7 ± 10 19 ± 14 32 ± 22 56 ± 37
Neutraclear 0 ± 0 1 ± 1 1 ± 2 6 ± 10 14 ± 17 22 ± 22 29 ± 24
Bionecteur 0 ± 0 2 ± 2 2 ± 3 5 ± 4 10 ± 8 17 ± 14 23 ± 18
Stopcock 1 ± 1 1 ± 1 1 ± 2 1 ± 2 2 ± 2 4 ± 6 31 ± 22
Equal (some 10 No 0 ± 0 0 ± 0 0 ± 1 1 ± 2 2 ± 3 6 ± 6 9 ± 10
level to Neutraclear 0 ± 1 –1 ± 1 –1 ± 2 –1 ± 1 –1 ± 3 0 ± 4 2 ± 7
simulated Bionecteur 0 ± 0 0 ± 0 0 ± 1 0 ± 0 0 ± 1 0 ± 1 1 ± 3
patient) Stopcock 0 ± 0 0 ± 1 1 ± 1 1 ± 1 1 ± 1 1 ± 1 3 ± 7
0 No 0 ± 0 0 ± 1 0 ± 1 1 ± 3 5 ± 6 10 ± 18 13 ± 21
Neutraclear 0 ± 0 1 ± 1 0 ± 1 1 ± 4 1 ± 4 3 ± 9 9 ± 19
Bionecteur 0 ± 0 0 ± 0 1 ± 2 2 ± 3 2 ± 2 3 ± 4 9 ± 23
Stopcock 0 ± 0 0 ± 1 1 ± 2 1 ± 2 1 ± 2 2 ± 2 6 ± 9
–5 No 0 ± 0 0 ± 1 1 ± 2 3 ± 4 6 ± 5 9 ± 9 13 ± 12
Neutraclear 0 ± 0 0 ± 3 0 ± 5 10 ± 10 13 ± 9 21 ± 11 34 ± 16
Bionecteur 0 ± 0 1 ± 3 2 ± 2 5 ± 6 6 ± 6 9 ± 10 21 ± 21
Stopcock 0 ± 0 1 ± 1 1 ± 2 1 ± 2 1 ± 2 3 ± 6 17 ± 19
Lower 10 No 0 ± 0 –4 ± –4 –12 ± 22 –27 ± 50 –40 ± 65 –46 ± 72 –54 ± 82
(−32 cm down Neutraclear 0 ± 0 –4 ± 7 –7 ± 9 –10 ± 9 –19 ± 20 –23 ± 25 –31 ± 38
to simulated Bionecteur 0 ± 0 –3 ± 3 –4 ± 5 –6 ± 6 –16 ± 17 –21 ± 20 –22 ± 20
patient) Stopcock 0 ± 0 0 ± 1 1 ± 1 1 ± 2 1 ± 2 1 ± 2 –2 ± 7
0 No 0 ± 0 –1 ± 3 –2 ± 4 –3 ± 6 –5 ± 8 –6 ± 9 –10 ± 15
Neutraclear 0 ± 0 –1 ± 2 –2 ± 4 –1 ± 4 –3 ± 4 –4 ± 5 –8 ± 13
Bionecteur 0 ± 0 –2 ± 3 –3 ± 4 –5 ± 7 –13 ± 15 –14 ± 17 –20 ± 32
Stopcock 0 ± 0 1 ± 2 1 ± 2 1 ± 2 1 ± 2 1 ± 2 –1 ± 14
–5 No 0 ± 0 0 ± 1 –1 ± 2 –1 ± 2 –1 ± 3 –1 ± 4 0 ± 7
Neutraclear 0 ± 0 0 ± 1 –1 ± 1 –1 ± 1 –1 ± 2 –2 ± 4 –3 ± 8
Bionecteur 0 ± 0 –4 ± 9 –5 ± 11 –8 ± 14 –15 ± 20 –17 ± 25 –28 ± 47
Stopcock 0 ± 1 1 ± 1 1 ± 1 1 ± 2 1 ± 2 1 ± 2 4 ± 11

CVP: central venous pressure; S0: plunger opening; S1: syringe extraction from the pump; S2: disconnection from infusion line; S3: reconnection to
the infusion line; S4: syringe insertion in the pump; S5: plunger closing; S6: stabilization at the end of maneuver.
All results are expressed in microliters.

CVP (–5 mmHg—example: a patient with adult respiratory our study, we tested two different devices of neutral pres-
distress syndrome undergoing non-invasive ventilation)16 sure with different performances. In two recent reports,10,11
and a displacement of the pump higher than the patient the Bionecteur demonstrated high performances in main-
level. In this case, there will be a bolus of the medication taining a neutral pressure in the intravascular tubing,
equal to 56 ± 37 µL corresponding to a 0.028-µg infusion. achieving similar performances to those created for periph-
The air enters into the administration set when in the dis- erally inserted central venous catheters (PICC). The appli-
connection no occlusive system is applied, excluding the cation of the device reduces in an important way all the
manual procedure done by the operator. Air entrance will situations of bolus or backflow in all the different situations
further determine a stop in the medication infusion when tested. The other tested device, the Neutraclear, given the
the air will reach the catheter tip. All of these varying situ- lower performances compared to the Bionecteur, reduces
ations can be limited, as observed in our study, when a the complications in a limited way. The use of neutral dis-
device such as a stopcock or a neutral displacement NFC is placement NFC reduces the backflow or the bolus compli-
placed between the syringe and the administration set. In cations especially when the CVP is higher than 0 mmHg
6 The Journal of Vascular Access 00(0)

Figure 4.  Mean value of bolus/backflow, step: stabilization at the end of maneuver.

performances represents the easiest and safer implementing


solution.10,11 From the results of our study, we can limit all
the possible complications given by the change of syringe
pumps with the application of simple nursing interventions
as the following:

Utilization of neutral displacement NFC to limit bolus


and backflow events;
Displacement of the pumps infusing vasoactive drugs
higher than the patient level (always considering the
entry point of the catheter if the patient is not in supine
position);
Reduce the medication concentration to diminish the
total amount of medication eventually contained in the
bolus or in the backflow.
Figure 5.  Difference in the six investigated steps between
two different syringe pumps investigated.
Study limitations
and when the syringe pumps are displaced matching to the This is a single-center study. The results are achieved by
patient level or higher. When the syringe pumps are dis- the utilization of only two NFC whose performances can
placed lower than the patient level, the use of a stopcock differ from other available devices. The two NFC used
between the syringe and the administration set seems to were purchased by our hospital at the time of the study.
represent the preferable solution to reduce the backflow Furthermore, we did not experiment the use of the new-
events. The application of a stopcock is simple, but can generation NFC designed for PICC,10,11 given their high
generate a stop of the medication’s administration if inac- cost and the presence of a double valve system. When con-
curately left closed until the pressure alarm of the pump sidering the CVP system, it was unable to reproduce dif-
rings.17 Applying a neutral displacement NFC with high ferent CVP values or physiological fluctuation produced
Elli et al. 7

in humans by heart and lung interactions. This CVP fluc- References


tuation could be very pronounced, especially when consid- 1. Trim JC and Roe J. Practical considerations in the admin-
ering a patient with spontaneous breathing due to acute or istration of intravenous vasoactive drugs in the critical care
chronic respiratory insufficiency, thus influencing the setting: the double pumping or piggyback technique-part
achieved results, and all data were simulated considering a one. Intensive Crit Care Nurs 2004; 20(3): 153–160.
patient in supine position. 2. Holmes CL. Vasoactive drugs in the intensive care unit.
Finally, we did not address the problem related to sur- Curr Opin Crit Care 2005; 11(5): 413–417.
face disinfection of the NFC. In clinical practice, the disin- 3. Annane D, Ouanes-Besbes L, de Backer D, et al. A global
fection of NFC could increase the problems related to the perspective on vasoactive agents in shock. Intensive Care
Med 2018; 44(6): 833–846.
interruption of the infusion, amplifying the results high-
4. Arino M, Barrington JP, Morrison AL, et al. Management of
lighted in our study. Moreover, in our study, when discon- the changeover of inotrope infusions in children. Intensive
nected, the NFC would remain exposed to air for a few Crit Care Nurs 2004; 20(5): 275–280.
seconds, reducing the need for disinfection. The manufac- 5. de Barbieri I, Frigo AC and Zampieron A. Quick change
turers recommend the devices to be replaced after about versus double pump while changing the infusion of ino-
350 connections and/or 7 days of use. This would mean a tropes: an experimental study. Nurs Crit Care 2009; 14(4):
daily average of connections equal to 50. When syringe 200–206.
changes become frequent, nurses usually increase the drug 6. Rakza T, Richard A, Lelieur AC, et al. Factors altering low-
concentration. It is unlikely that NFC, used on pump flow drug delivery using syringe pumps: Consequences on
syringes, will reach 50 connections/day. vasoactive drug infusion in preterm infant. Arch Pediatr
2005; 12(5): 548–554.
7. Achuff BJ, Achuff JC, Park HH, et al. Epinephrine syringe
Conclusion exchange events in a paediatric cardiovascular ICU: analys-
ing the storm. Cardiol Young 2018; 28(3): 409–415.
The syringe change can be a very critical moment given 8. Li DK, Wang XT and Liu DW. Association between ele-
different influencing variables. Syringe pump position, vated central venous pressure and outcomes in critically ill
displaced higher than the patient level, always generates a patients. Ann Intensive Care 2017; 7(1): 83.
medication bolus that is higher at the lowering of the CVP 9. Zhou X, Zhang Z, Li J, et al. Targeted end point of CVP
value. In this situation, the presence of a neutral displace- of 15mm Hg would be better than that of 12mm Hg in the
patients with mechanical ventilation. Am J Emerg Med
ment NFC reduces the incidence of boluses. When the
2018; 36(2): 331–332.
pump is displaced at the patient level, the presence of a 10. Elli S, Abbruzzese C, Cannizzo L, et al. In vitro evaluation
neutral displacement NFC reduces the establishment of of fluid reflux after flushing different types of needleless
boluses, even in simulations considering a CVP of connectors. J Vasc Access 2016; 17(5): 429–434.
−5 mmHg. A low CVP, or a negative one, can increase the 11. Hull GJ, Moureau NL and Sengupta S. Quantitative assess-
risks. Bolus or backflow is not completely preventable but ment of reflux in commercially available needle-free IV
can be limited and controlled. More studies are awaited to connectors. J Vasc Access 2018; 19(1): 12–22.
investigate the problem. 12. Igarashi H, Obata Y, Nakajima Y, et al. Syringe pump
displacement alters line internal pressure and flow. Can J
Anaesth 2005; 52(7): 685–691.
Declaration of conflicting interests 13. Donald AI, Chinthamuneedi MP and Spearritt D. Effect of
The author(s) declared no potential conflicts of interest with changes in syringe driver height on flow: a small quantita-
respect to the research, authorship, and/or publication of this tive study. Crit Care Resusc 2007; 9(2): 143–147.
article. 14. Giraud T, Dhainaut JF, Vaxelaire JF, et al. Iatrogenic com-
plications in adult intensive care units: a prospective two-
Funding center study. Crit Care Med 1993; 21(1): 40–51.
15. Larsen GY, Parker HB, Cash J, et al. Standard drug con-
The author(s) received no financial support for the research, centrations and smart-pump technology reduce continuous-
authorship, and/or publication of this article. medication-infusion errors in pediatric patients. Pediatrics
2005; 116(1): e21–e25.
ORCID iDs 16. Neff T, Fischer J, Fehr S, et al. Start-up delays of infusion
syringe pumps. Paediatr Anaesth 2001; 11(5): 561–565.
Stefano Elli https://orcid.org/0000-0002-8785-824X
17. Deckert D, Buerkle C, Neurauter A, et al. The effects of mul-
Stefano Bambi https://orcid.org/0000-0002-0360-4470 tiple infusion line extensions on occlusion alarm function of
Alberto Lucchini https://orcid.org/0000-0002-7475-277X an infusion pump. Anesth Analg 2009; 108(2): 518–520.

You might also like