Professional Documents
Culture Documents
research-article2020
JVA0010.1177/1129729820909024The Journal of Vascular AccessElli et al.
Abstract
Introduction: In the literature, the change of a syringe pump is described as a dangerous situation, especially in the
case of vasoactive drug administration.
Methods: Different variables have been studied (central venous pressure, pump displacement in relation to the patient
position, utilization of a stopcock, or a neutral displacement needle-free connector between the syringe and the infusion
tubing) to understand their influence on medication administration in terms of backflow or bolus creation when changing
the syringe.
Results: We performed 576 measurements with different combinations. With respect to all the observations, in
comparison with “time zero,” we found the following differences expressed in microliters: 0 (±1) at the plunger opening;
0 (±3) at the syringe extraction from the pump; 0 (±7) at the syringe disconnection from the infusion tubing; 0 (±11)
at the syringe reconnection to the infusion tubing; 1 (±7) at the syringe insertion in the pump; 3 (±23) at the plunger
closing; 8 (±33) at the stabilization at the maneuver end.
Conclusion: The syringe change can be a very critical moment given different influencing variables. Syringe pump
position, displaced higher than the patient level, always generates a medication bolus that is higher at the lowering of
the central venous pressure value. The presence of a neutral displacement needle-free connector reduces the incidence
of boluses. When the pump is placed at the patient level, the presence of neutral displacement needle-free connector
reduces the establishment of boluses, even in a central venous pressure of −5 mmHg simulations.
Keywords
Backflow, bolus, changeover, infusion, inotrope infusion, needle-free connector, syringe pump vasoactive drug
pumping, vasoactive drug administration
Introduction 1
eneral Intensive Care Unit, Emergency Department – ASST Monza,
G
San Gerardo Hospital, University of Milano-Bicocca, Monza, Italy
Background and rationale 2
School of Nursing, Department of Medical and Biological Sciences,
Udine University, Udine, Italy
Intensive care unit (ICU) patients often require adminis- 3
Medical & Surgical Intensive Care Unit, Careggi University Hospital,
tration of intravenous therapies. A concentrated continu- Florence, Italy
ous infusion of medications, rather than a diluted 4
General Intensive Care Unit, Azienda Ospedaliera Universitaria di
concentration, is frequently used because of fluid restric- Parma, Parma, Italy
5
tions in critically ill patients.1 In particular, vasoactive General Intensive Care Unit, Istituto Auxologico, Milano, Italy
6
Department of Anesthesia and Intensive Care Medicine, Niguarda Ca’
drugs are the mainstay of hemodynamic management of Granda, University Hospital of Milano-Bicocca, Milano, Italy
critically ill patients.2,3 Medical devices, such as syringe
pumps, are commonly used for the continuous infusion of Corresponding author:
Alberto Lucchini, General Intensive Care Unit, Emergency Department
highly concentrated drugs. The syringe pumps, thanks to – ASST Monza, San Gerardo Hospital, University of Milano-Bicocca, Via
technological innovation, are able to deliver very precise Pergolesi 33, 20900 Monza, Italy.
drug doses, in the order of 0.1 mL/h. However, the current Emails: a.lucchini@asst-monza.it; alberto.lucchini@unimib.it
2 The Journal of Vascular Access 00(0)
literature identifies several factors that can contribute to pressure (CVP). CVP values can vary considerably depend-
reducing the stability of the administration: device’s tech- ing on the type of ventilation, blood volume, and many
nical features, catheter patency, start-up delay, mechani- other factors; differences of 10–15 mmHg (15–20 cmH2O)
cal occlusion of the infusion tubing, siphoning, users’ can be found between a patient in mechanical ventilation
knowledge, and so on.1 These factors are capable of modi- and one in spontaneous breathing with severe respiratory
fying the stability of the infusion and determining hemo- distress.8,9 If the pressure of the infusion tubing is higher
dynamic imbalances. Various precautions are suggested to than the “blood pressure,” the solution in the tubing will be
avoid changes in the current flow during infusion: avoid pushed toward the patient, thus causing a bolus of the medi-
vertical displacements of the pump, use of needle-free cation with an immediate overdose. If the pressure in the
connectors (NFC), use of two parallel pumps, and avoid infusion tubing is lower than the “blood pressure,” blood
infusion of different drugs in the same tubing.1,4,5 The will flow back into the infusion tubing, thus stopping the
greatest risk of altering the stability of the infusion arises infusion of the medication and, perhaps, a delay in stabiliz-
from the replacement of syringe. In every ICU, syringe ing the infusion. It would be desirable to reduce these risks
replacement is a common practice performed by many with an optimal use of the available technology and the use
people. The risk of hemodynamic imbalances or other of standardized change procedures.
problems is always present.6,7 During the change, the con-
tinuity and integrity of infusion tubing and protection
against the interactions between blood pressure and tub- Objective
ing pressure are lost. Figure 1 shows a modification The purpose of this experimental study is to analyze the
induced by syringe replacing in a patient with norepineph- effects of syringe replacement on the continuity of drug
rine infusion. In this case, the infusion of 0.10 µg/kg/min infusion, based on the absence/presence of different
(infusion rate of 5 mL/h) of norepinephrine required 8 min syringe infusion set-ups.
to recover the previous arterial pressure values.
We can call “blood pressure” the pressure of the blood-
stream at the tip of the catheter, which causes resistance to Methods
infusion. We can call “tubing pressure” the pressure
induced by the height of the liquid column in the infusion
Study design and setting
tubing in relation to the patient. In ICU wards, drugs are We conducted an “in vitro experimental study” during
generally administrated via a central venous catheter, so the August/September 2017 in a laboratory of an ICU in a
“blood pressure” is generally similar to central venous North Italian University Hospital. We assembled a
Elli et al. 3
simulation system for measuring fluid displacement into the variables studied are explained in Figure 2. An air bub-
the infusion tubing during syringe replacement. The simu- ble was inserted in the tubing at the “displacement measur-
lation system was assembled with the following devices: ing area” and used as reference for measurements10. The
two different types of syringe pump (Pilot A2®; Fresenius displacement of the air bubble was measured during each
Kabi AG, Bad Homburg, Germany; Injectomat Agilia®, of the six steps of the syringe change after stopping of infu-
Fresenius Kabi AG), a 50-mL syringe (BD Perfusion 50 mL sion: plunger opening (S0), syringe extraction from the
Syringe; Becton Dickinson, Franklin Lakes, NJ, USA), a pump (S1), disconnection from infusion tubing (S2), recon-
250-cm-long infusion tubing (ICU Medical, San Clemente, nection to the infusion tubing (S3), syringe insertion in the
CA, USA), three-way stopcock (Bicakcilar, Istanbul, pump (S4), plunger closing (S5), and stabilization at the
Turkey), hemodynamic pressure transducer (TruWave; end of maneuver (S6). Four different operators (nurses with
Edwards Lifesciences, Irvine CA, USA), multiparametric up to 5 years of experience in ICU) carried out the “six
monitor (Draeger SC 9000; Draeger Medical, Telford, steps” of standardized change procedure in all possible
Pennsylvania USA), two different types of neutral dis- combinations of the studied variables. During each step, the
placement NFC (Neutraclear®; Becton Dickinson, movements of the air bubble in the “displacement measur-
Bionecteur®; Vygon, Ecouen, France) with different perfor- ing area” were manually noted, as reported in Figure 3. The
mances,10,11 and 500 mL of colored normal saline connected measures, expressed in centimeters, were then converted
to an Infusion Set (Rays SpA, Osimo, Italy). Variables stud- into milliliter of fluid displacement in the infusion tubing.
ied were the following: type of pump used, pump highness A positive displacement was identified as “bolus” and a
in relation to the simulated patient,12,13 three different CVP negative displacement was identified as “backflow” (reflux
simulated values,8,9 connection between the syringe and of blood into the catheter).
the infusion tubing, and operator. The pump positions
considered in relation to the patient were 50 cm above the
Statistical analysis
position of the heart, patient in supine position (heart level)
and 32 cm below the patient, with respect to heart level. Statistical analysis was performed using Statistical
Simulated CVP values analyzed were +15, 0, and Package for the Social Sciences, version 22.0, (SPSS Inc.,
−5 mmHg. How the simulation system was assembled and Chicago, IL, USA). Continuous variables were expressed
4 The Journal of Vascular Access 00(0)
Table 1. Fluid displacement with different devices and set-up combinations.
CVP: central venous pressure; S0: plunger opening; S1: syringe extraction from the pump; S2: disconnection from infusion line; S3: reconnection to
the infusion line; S4: syringe insertion in the pump; S5: plunger closing; S6: stabilization at the end of maneuver.
All results are expressed in microliters.
CVP (–5 mmHg—example: a patient with adult respiratory our study, we tested two different devices of neutral pres-
distress syndrome undergoing non-invasive ventilation)16 sure with different performances. In two recent reports,10,11
and a displacement of the pump higher than the patient the Bionecteur demonstrated high performances in main-
level. In this case, there will be a bolus of the medication taining a neutral pressure in the intravascular tubing,
equal to 56 ± 37 µL corresponding to a 0.028-µg infusion. achieving similar performances to those created for periph-
The air enters into the administration set when in the dis- erally inserted central venous catheters (PICC). The appli-
connection no occlusive system is applied, excluding the cation of the device reduces in an important way all the
manual procedure done by the operator. Air entrance will situations of bolus or backflow in all the different situations
further determine a stop in the medication infusion when tested. The other tested device, the Neutraclear, given the
the air will reach the catheter tip. All of these varying situ- lower performances compared to the Bionecteur, reduces
ations can be limited, as observed in our study, when a the complications in a limited way. The use of neutral dis-
device such as a stopcock or a neutral displacement NFC is placement NFC reduces the backflow or the bolus compli-
placed between the syringe and the administration set. In cations especially when the CVP is higher than 0 mmHg
6 The Journal of Vascular Access 00(0)
Figure 4. Mean value of bolus/backflow, step: stabilization at the end of maneuver.