You are on page 1of 8

Medium-Term Outcomes After Implantation

of Expanded Polytetrafluoroethylene
Valved Conduit
Yoshio Ootaki, MD, PhD, Allison S. Welch, PA, Michael J. Walsh, MD,
Michael D. Quartermain, MD, Derek A. Williams, DO, and Ross M. Ungerleider, MD
Divisions of Cardiothoracic Surgery and Pediatric Cardiology, Wake Forest Baptist Health, Winston-Salem, North Carolina

Background. Pulmonary valve replacement is 18 patients) and 0.1 mm thick ePTFE membrane for the
becoming the most frequent congenital heart surgery leaflet material. Valve function was assessed by echo-
performed on adolescents and young adults. Numerous cardiogram after the implantation.
surgical options are available, including autologous Results. There were no postoperative hospital deaths.
pericardium, mechanical valves, allografts, and bio- Mean follow-up was 2.6 years (range, 4 months to 4.7
prosthetic valves. Each option has limitations with dura- years), and there have been no reoperations for the ePTFE
bility and freedom from reintervention for stenosis or VC. There was one reintervention for stenosis at the distal
insufficiency, particularly in the pediatric population. anastomosis. Pulmonary insufficiency was mild or less in
The purpose of this study was to analyze our uniquely 24 patients (92%). The average peak instantaneous pres-
designed, expanded polytetrafluoroethylene (ePTFE) sure gradient between the right ventricle and the pulmo-
valved conduit (VC) for right ventricular outflow tract nary artery was 22.4 ± 15.1 mm Hg at latest follow-up.
reconstruction. Conclusions. Compared with historical data for other
Methods. Beginning in 2012, ePTFE VC were implan- pulmonary valve replacement options, our ePTFE VC
ted in 26 patients with a median age of 9.8 years (range, shows superior medium-term performance, with less
1.0 to 15.9). Bicuspid VC were used in the first 3 patients, reintervention or significant valve dysfunction.
and tricuspid VC were used in 23 patients. The ePTFE VC
is fashioned from commercially available ePTFE tube (Ann Thorac Surg 2017;-:-–-)
graft (16 mm in 1 patient, 20 mm in 7 patients, 24 mm in Ó 2017 by The Society of Thoracic Surgeons

P ulmonary valve replacement is becoming the most


frequent congenital heart surgical procedure
performed on adolescents and young adults. Numerous
Expanded polytetrafluoroethylene (ePTFE) has been
reported for use in pulmonary valve replacement [9–18].
Monocusp valves were reported in 2002 [9]; and bicuspid
surgical options are available and included autologous valves have been created in situ, and more recently, in
pericardium [1], mechanical valves [2] (although these are conduits [13, 15]. Intermediate term follow-up with
not commonly used in the pulmonary position), allografts respect to freedom from significant valve deterioration
[3], and bioprosthetic valves (both as stented and freestyle has been excellent. Tricuspid valves have been created
porcine valves [4, 5] and as bovine jugular vein conduits primarily in Japan where allografts have not been widely
[6]). More recently, there is growing enthusiasm for available [10–12]. The follow-up now exceeds 10 years
percutaneously inserted bioprosthetic valves [7]. How- with excellent results. Construction of these valves, as
ever, percutaneously inserted valves have demonstrated a described in the literature, has seemed somewhat
disturbingly high incidence of subacute bacterial endo- complicated, however. In 2013, we described our
carditis in addition to stent fracture and risks for late innovative technique of simplified tricuspid ePTFE valved
valve dysfunction [8]. None of these options is ideal, and conduit (VC) construction and implantation, which
each has problems with durability and freedom from consists of commercially available ePTFE graft and 0.1
stenosis or insufficiency requiring reoperation, mm thick ePTFE membrane (W.L. Gore & Associates,
particularly in the pediatric population. Flagstaff, AZ) [19]. The purpose of this study is to review
the medium-term outcomes of our ePTFE VC.

Accepted for publication July 10, 2017.


Patients and Methods
Presented at the Sixty-third Annual Meeting of the Southern Thoracic
Surgical Association, Naples, FL, Nov 9–12, 2016. Patients
Address correspondence to Dr Ootaki, Wake Forest Baptist Health, Pe-
From January 2012 to May 2016, we implanted a bicuspid
diatric Cardiothoracic Surgery, Medical Center Blvd, Winston-Salem, ePTFE VC in 3 patients and a tricuspid ePTFE VC in 23
NC 27157; email: yotaki@wakehealth.edu. patients. The median age at the time of implantation was

Ó 2017 by The Society of Thoracic Surgeons 0003-4975/$36.00


Published by Elsevier Inc. http://dx.doi.org/10.1016/j.athoracsur.2017.07.013
2 OOTAKI ET AL Ann Thorac Surg
TRICUSPID EPTFE VALVE 2017;-:-–-

9.9 years (range, 1.0 to 15.9). Patients’ median body weight This research study was approved by the Institutional
was 28.9 kg (range, 7.8 to 64.9 kg). Three sizes of VC were Review Board at Wake Forest School of Medicine.
used depending on the patient’s body size and preoper-
ative right ventricular outflow tract measurement: 16 mm Construction of the Conduit
in 1 patient, 20 mm in 7 patients, and 24 mm in 18 We construct this VC using commercially available PTFE
patients. Primary diagnosis included tetralogy of Fallot in conduits (16 mm, 20 mm, or 24 mm) and 0.1 mm thick
15 patients, truncus arteriosus in 3 patients, aortic PTFE membrane for the leaflet material. After the patient
valvular disease in 4 patients, and other pulmonary valve is intubated, the surgeon creates the conduit on a sterile
disease in 4 patients. Twenty-four patients (92%) had a table. A bicuspid VC is created using the technique
history of median sternotomy. Concomitant procedure at described by Quintessenza and associates [13], and a
the time of ePTFE VC implantation included Ross tricuspid VC is created using our unique technique,
operation in 4 patients, closure of atrial septal defect in 3 which has been previously described (Fig 1) [19].
patients, closure of ventricular septal defect in 1 patient, Depending on the size of the VC, we create a PTFE
and mitral valve replacement in 1 patient. Currently, membrane with three equal-sized leaflets, according to
patients who require pulmonary valve conduits smaller the formula in Figure 2.
than 16 mm or larger than 24 mm receive other options A 24-mm PTFE conduit is then cut to about 6 cm in
such as allografts or bioprosthetic valves. length. The conduit does not need to be longer, and making

Fig 1. Bicuspid (left) and tricuspid (right) expanded polytetrafluoroethylene valved conduit in open (top) and closed (bottom) position.
Ann Thorac Surg OOTAKI ET AL 3
2017;-:-–- TRICUSPID EPTFE VALVE

once the conduit is reinverted. Once the leaflet material is


wrapped around the conduit, we fix each of the other
commissural apposition areas with different sutures in
symmetrical positions. All three leaflets are sewn using
continuous sutures. After sewing the three leaflets to the
surface of the conduit, the tube graft can be reinverted,
bringing the leaflets to the inside. Commissure sutures
using 6-0 monofilament sutures are placed to create
sinuses. We place these in a vertical mattress fashion at the
opposing edges of the valves. We then confirm that the VC
can hold saline. An arrow (drawn with a sterile marking
pen) is used to mark the direction of flow and is placed on
the top of the conduit in a location that leaves a commissure
directly posterior. It usually takes about 60 minutes to
construct a tricuspid VC.

Echocardiographic Evaluation
The implanted VC were assessed using transesophageal
echocardiography in the operating room immediately
after surgery and transthoracic echocardiogram for
follow-up imaging. The pulmonary regurgitation was
classified as trivial (0), mild (1), moderate (2), and severe
(3). Peak instantaneous pressure gradient was assessed by
measuring peak velocity through the conduit.

Statistical Analysis
Univariate analysis for survival, freedom from reinter-
vention, and freedom from replacement of ePTFE VC
were performed using the Kaplan-Meier method. All
analysis was carried out using Microsoft Excel (Microsoft,
Redmond, WA).

Results
Of the 26 patients in our cohort, there were no hospital
deaths after implantation. Mean follow-up was 2.6 years
(range, 4 months to 4.7 years). One patient who had a
Ross operation and implantation of an ePTFE VC in the
pulmonary position with an uneventful hospital course
died 15 months after the surgery for unknown reasons
(Fig 3). This patient showed no regurgitation and minimal
pressure gradient (peak of 20 mm Hg) in the ePTFE VC at
14-month follow-up. Overall patient survival was 95.5% at
4 years. There was one reintervention during the follow-
up period for moderate stenosis at the distal ePTFE VC
suture line (Fig 4). Freedom from death and reinterven-
tion was 90.7% at 4 years. There were no reoperations for
Fig 2. Creating trileaflet expanded polytetrafluoroethylene (ePTFE)
valved conduit. (A) Create a piece of 0.1-mm ePTFE membrane that is failed ePTFE VC. There was no significant difference in
18 mm in height by 78 mm long for 24 mm valved conduit. (B) Fold survival or reintervention rate between groups of leaflet
the ePTFE membrane into thirds. (C) Leave 5 mm on the top, and cut design or size of the ePTFE VC.
the bottom into an appropriate sinus shape. (R ¼ radius.) Figure 5 shows the echocardiographic peak instanta-
neous pressure gradient through the VC. The average
peak instantaneous pressure gradient between the right
it too long hampers valve construction. The conduit is ventricle and the pulmonary artery measured by echo-
inverted so that the valve leaflets can be created on the cardiogram was 22.4  15.1 mm Hg in latest follow-up
outside. The leaflet material is placed 1.5 cm down from the echocardiogram. One patient had a 65 mm Hg pressure
top of the conduit. The commissures of each leading edge gradient and trivial pulmonary regurgitation by
are brought together and connected using CV-6 PTFE echocardiogram 1 day after the surgery. Cardiac cathe-
suture (W.L. Gore & Associates). We choose to tie the knots terization the next day showed a 28 mm Hg peak-to-peak
on the “inside” of the conduit, which will be the outside pressure gradient between the right ventricle and
4 OOTAKI ET AL Ann Thorac Surg
TRICUSPID EPTFE VALVE 2017;-:-–-

Fig 3. Kaplan-Meier curve of survival.

Fig 5. Pressure gradient through the conduit measured by echocardio-


pulmonary artery. His echocardiogram pressure gradient gram (peak).
decreased during the subsequent follow-up period;
however, the peak pressure gradient by echocardiogram
with allografts have been excellent, especially in neonates
increased to 65 mm Hg at 19-month follow-up. Cardiac
and infants; however, valve deterioration over time has
catheterization showed a 36 mm Hg pressure gradient at
been an issue, especially in smaller allografts [20].
the distal anastomosis. The pressure gradient decreased
Freedom from explantation for age less than 10 years at
to 12 mm Hg after balloon dilation. Pulmonary valve
implantation was 78.9% in 5 years, 59.6% in 10 years, and
function was excellent, and there was no pulmonary valve
32.5% in 15 years. In some countries, like Japan, allografts
regurgitation before or after the reintervention. Figure 6
are not widely available. Bovine jugular VC has been an
demonstrates the pulmonary regurgitation of the ePTFE
alternative to allografts since 1999. Early results are also
VC assessed by echocardiogram. Two patients (7.7%)
excellent although valve deterioration or endocarditis has
showed moderate pulmonary regurgitation, although
been an issue in long-term use [6]. Freedom from catheter
neither had any clinical symptoms. Three patients (11.5%)
intervention and conduit replacement in 5 years was
showed mild pulmonary regurgitation, and 21 patients
82.7% and 82.1%, respectively. Stented or nonstented
(80.8%) showed trivial pulmonary regurgitation. One
bioprosthetic valves have also been a preferred option in
patient had suspected infective endocarditis due to fever
some centers; however, valve deterioration is an issue for
and thickened anterior ePTFE leaflet. The patient was
long-term use of these valves [4, 5]. Lee and associates
treated successfully with antibiotics. No other patients
[21] compared the performance of three types of bio-
demonstrated any signs of late endocarditis.
prosthesis (stented porcine, stented bovine pericardial,
and stentless porcine valves). They reported freedom
Comment from repeat pulmonary valve replacement at 10 years was

Many materials have been used for pulmonary valve


replacement in the pediatric population. Allografts have
been widely used for more than 50 years. Early outcomes

Fig 6. Pulmonary regurgitation assessed by echocardiogram: 0 ¼


trivial, 1 ¼ mild, 2 ¼ moderate, 3 ¼ severe. Latest echocardiogram
showed trivial pulmonary regurgitation in 21 patients (80.8%), mild
Fig 4. Kaplan-Meier curve of freedom from death or reintervention. in 3 patients (11.5%), and moderate in 2 patients (7.7%).
Ann Thorac Surg OOTAKI ET AL 5
2017;-:-–- TRICUSPID EPTFE VALVE

84.6% for stented bovine pericardial valves, 48.5% for malfunction, valve function was well maintained in 75%
stented porcine valves, and 31.2% for stentless porcine of patients at 5 years [18], which is also slightly lower than
valves [21]. Recently, there has been growing enthusiasm our experience. Because we only utilized pulmonary
for percutaneously inserted bioprosthetic valves. valve conduits of 16 mm, 20 mm, and 24 mm, the size of
However, endocarditis and early deterioration has been the conduit might contribute to the reintervention rate.
seen in large cohort studies [7, 8]. None of the valves There has been no standardized description for design
discussed provide consistently reliable and excellent of an ePTFE VC although the anatomy of the normal
outcomes, especially in the pediatric population. heart valve suggests that a tricuspid geometric construc-
Prosthetic VC using ePTFE were first described by a tion would be an appropriate model for an ePTFE VC. To
Japanese group in 2002 [10]. Valve function was well appreciate the measurements that guide the creation of
preserved in these patients for as long as 10 years. our valve, it is essential to understand the geometry of
Techniques described in the literature have utilized normal tricuspid valves. Although there are some differ-
creation of bulging sinuses, but creation of these valves is ences between aortic and pulmonary valves, the data on
cumbersome [10], and there have been no defined normal aortic leaflet anatomy are well described and
benefits for this technique. Several other investigators easily translated to the tricuspid configuration for our
reported techniques to create bicuspid [13, 15] or pulmonary VC. In an elegant manuscript describing
tricuspid [14, 16] ePTFE VC, but other techniques for aortic valve anatomy in 100 healthy human hearts, the
tricuspid valves are difficult to recreate. Our tricuspid VC height of each cusp is 50% of leaflet length [22]. Therefore,
is simple to create and easy to implant compared with the length of each leaflet is a critical factor in determining
bicuspid VC because of its shorter length. It takes less the height of the leaflet. The study also showed that there
than 60 minutes to create our tricuspid ePTFE VC without is a 5-mm coaptation area for each leaflet.
difficulty. Review of the literature demonstrates that When constructing a PTFE valve, the height of the
freedom from reintervention or valve replacement seems leaflet (and thus the length of the conduit) can differ
to be substantially better for ePTFE VC compared with all substantially depending on whether a bicuspid or
other types of valves. Shinkawa and associates [18] tricuspid construction is employed. In a bicuspid con-
reported the freedom from reintervention rate for ePTFE struction, the leaflet length is multiplication of 3 radius
VC was 83.7% at 5 years, which is slightly lower than our plus 2 mm (Fig 7). In a tricuspid valve, each leaflet length
experience, but their technique with valve creation was is multiplication of 2 radius plus 2 mm, which calculates to
also more cumbersome. When severe conduit insuffi- a leaflet length shorter than what is needed for a bicuspid
ciency or an instantaneous pressure gradient greater than valve. The importance of leaflet length in determining
50 mm Hg through the conduit was considered conduit height is illustrated in Figure 7, which demonstrates

Fig 7. Leaflet design for


bicuspid (left) and tricuspid
(right) expanded polytetra-
fluoroethylene valved
conduit. (R ¼ radius.)
6 OOTAKI ET AL Ann Thorac Surg
TRICUSPID EPTFE VALVE 2017;-:-–-

construction of a 24-mm VC. For a bicuspid valve, the compare our outcomes with other types of VC. Our
length of each leaflet is 3 radius, or 36 mm. Because of the patient population is relatively older, and we did not
relatively less pliable features of ePTFE compared with implant VC size smaller than 16 mm. However, our
human valve tissue, we have been using a leaflet height freedom from reintervention or reoperation seems to be
that is 70% of the leaflet length. For this example, 70% of lower compared with other types of VC from numerous
36 mm calculates to a leaflet height of 25.2 mm, and we published reports. Our study is also limited by a relatively
add an additional 1 mm to account for material needed for short follow-up period. Early and intermediate results are
holding sutures, giving a leaflet height of 26.2 mm. For a excellent for our ePTFE VC. However, long-term data will
tricuspid valve, the leaflet length is 2 R, or 24 mm, and be important as valve function in most other types of VC
70% of this size calculate a leaflet height of 16.8 mm, but deteriorates significantly after 10 years. We are
again we add an additional 1 mm to allow for suturing encouraged by the Japanese data that have followed these
(17.8 mm). Therefore, the tricuspid VC can be shorter than patients for more than 10 years, and we are optimistic that
a bicuspid VC—in the case shown, by nearly 1 cm—and our outcomes, in light of current data, will be similar and
that can provide a substantial advantage during surgical excellent. We will continue to follow up the valve function
implantation. This consistently “shorter” height for of our ePTFE VC. Finally, echocardiographic assessment
tricuspid versus bicuspid valves was validated in the pa- of pulmonary insufficiency is difficult to quantify, and we
per describing leaflet height in bicuspid and tricuspid must rely on qualitative assessments for this study.
aortic valves [23]. Bicuspid aortic valves have greater In conclusion, a tricuspid ePTFE VC can be easily
leaflet heights compared with leaflets in tricuspid aortic constructed in a size appropriate for pediatric patients.
valves. We do not have detailed in vitro valve perfor- Compared with historical data for other pulmonary valve
mance data in different configuration. These data would replacement options, our ePTFE VC shows superior
help for designing best ePTFE valved conduit in the medium-term performance, with less reintervention or
future. significant valve dysfunction. Compared with previously
Another possible advantage of tricuspid VC is closing described bicuspid conduits, our tricuspid VC has (1)
position of the leaflet itself. To have full opening of the shorter cusp height in each sinus—allowing for the entire
bicuspid VC, one half circumflex length of VC is neces- valve length to be shorter; (2) more physiologic coaptation
sary for each leaflet length. Conversely, one third of the valve leaflets (linear versus nonlinear); and (3)
circumflex length of VC is necessary for each leaflet potentially improved hemodynamics at high flow rates.
length to have full opening of the tricuspid VC (Fig 7). The
leaflets are positioned in a nonlinear shape at the closing
phase in the bicuspid configuration, but the leaflets face References
straight into each other at the closing phase in the 1. Pande S, Agarwal SK, Majumdar G, Narwaley M,
tricuspid configuration. We believe that this physiologic Shukla RK, Arora M. Reconstruction of trileaflet pulmo-
valve anatomy in tricuspid configuration would have an nary valve using autologous pericardium. Heart Lung Circ
2011;20:325–8.
advantage over bicuspid configuration in terms of valve
2. Freling HG, van Slooten YJ, van Melle JP, et al. Pulmonary
function, especially when the leaflet becomes thicker over valve replacement: twenty-six years of experience with
time [24]. mechanical valvar prostheses. Ann Thorac Surg 2015;99:
Long-term data are available after ePTFE VC 905–10.
placement from Japan [12] and percutaneously inserted 3. Tavakkol Z, Gelehrter S, Goldberg CS, Bove EL, Devaney EJ,
Ohye RG. Superior durability of SynerGraft pulmonary
bioprosthetic valves from a multicenter study in the allografts compared with standard cryopreserved allografts.
United States [7]. Although we need information more Ann Thorac Surg 2005;80:1610–4.
than 10 years after percutaneously inserted bioprosthetic 4. Alfieris GM, Swartz MF, Lehoux J, Bove EL. Long-term
valves, the ePTFE VC has advantages in terms of freedom survival and freedom from reoperation after placement of a
from reintervention. Percutaneously inserted bio- pulmonary xenograft valved conduit. Ann Thorac Surg
2016;102:602–7.
prosthetic valves also have a limitation in patients who 5. Dunne B, Suthers E, Xiao P, Xiao J, Litton E, Andrews D.
have large right ventricular outflow tracts. Therefore, our Medium-term outcomes after pulmonary valve replacement
current recommendation for replacing right ventricular with the Freestyle valve for congenital heart disease: a case
outflow tract is using ePTFE VC. If the size of the right series. Eur J Cardiothorac Surg 2016;49:e105–11.
6. Dave H, Mueggler O, Comber M, et al. Risk factor analysis of
ventricular outflow tract is appropriate, percutaneously 170 single-institutional Contegra implantations in
implanted bioprosthetic valves would be an option. pulmonary position. Ann Thorac Surg 2011;91:195–303.
Although we have not surgically replaced any ePTFE VC, 7. Cheatham JP, Hellenbrand WE, Zahn EM, et al. Clinical and
a percutaneously implanted bioprosthetic valves would hemodynamic outcomes up to 7 years after transcatheter
be an option, as we believe the ePTFE VC to be a perfect pulmonary valve replacement in the US melody valve
investigational device exemption trial. Circulation 2015;131:
“landing zone.” We did not consider smaller-sized ePTFE 1960–70.
VC (less than 16 mm) for smaller children, but that could 8. O’Donnell C, Holloway R, Tilton E, Stirling J, Finucane K,
be an option because the ePTFE VC works well even with Wilson N. Infective endocarditis following Melody valve
smaller diameter [24]. implantation: comparison with a surgical cohort. Cardiol
Young 2016;10:1–8.
There are several limitations to this study. This study 9. Turrentine MW, McCarthy RP, Vijay P, McConnell KW,
was not randomized to assess various types of right Brown JW. PTFE monocusp valve reconstruction of the right
ventricular outflow tract VC, and we could not directly ventricular outflow tract. Ann Thorac Surg 2002;73:871–80.
Ann Thorac Surg OOTAKI ET AL 7
2017;-:-–- TRICUSPID EPTFE VALVE

10. Yamagishi M, Kurosawa H, Nomura K, Kitamura N. Fan- 17. Chang TI, Chang CI. An efficient way to make a trileaflet
shaped expanded polytetrafluoroethylene valve in the conduit for pulmonary valve replacement. Ann Thorac Surg
pulmonary position. J Cardiovasc Surg (Torino) 2002;43: 2013;96:e163–5.
779–86. 18. Shinkawa T, Tang X, Gossett JM, et al. Valved polytetra-
11. Miyazaki T, Yamagishi M, Nakashima A, et al. Expanded fluoroethylene conduits for right ventricular outflow tract
polytetrafluoroethylene valved conduit and patch with reconstruction. Ann Thorac Surg 2015;100:129–37.
bulging sinuses in right ventricular outflow tract recon- 19. Ootaki (Otaki) Y, Quintessenza JA, Hopkins JA, et al. Im-
struction. J Thorac Cardiovasc Surg 2007;134:327–32. plantation of tricuspid expanded-polytetrafluoroethylene
12. Miyazaki T, Yamagishi M, Maeda Y, et al. Expanded (ePTFE) valved conduit for right ventricular outflow tract.
polytetrafluoroethylene conduits and patches with bulging Presented at the 60th annual meeting of the Southern
sinuses and fan-shaped valves in right ventricular outflow Thoracic Surgical Association, Scottsdale, AZ, Oct 30–Nov 2,
tract reconstruction: multicenter study in Japan. J Thorac 2013.
Cardiovasc Surg 2011;142:1122–9. 20. Boethig D, Goerler H, Westhoff-Bleck M, et al. Evaluation of
13. Quintessenza JA, Jacobs JP, Morell VO, Giroud JM, 188 consecutive homografts implanted in pulmonary
Boucek RJ. Initial experience with a bicuspid polytetra- position after 20 years. Eur J Cardiothorac Surg 2007;32:
fluoroethylene pulmonary valve in 41 children and adults: a 133–42.
new option for right ventricular outflow tract reconstruction. 21. Lee C, Park CS, Lee CH, et al. Durability of bioprosthetic
Ann Thorac Surg 2005;79:924–31. valves in the pulmonary position: long-term follow-up of 181
14. Ando M, Takahashi Y. Ten-year experience with handmade implants in patients with congenital heart disease. J Thorac
trileaflet polytetrafluoroethylene valved conduit used for Cardiovasc Surg 2011;142:351–8.
pulmonary reconstruction. J Thorac Cardiovasc Surg 22. Jatene MB, Monteiro R, Guimar~ aes MH, et al. Aortic valve
2009;137:124–31. assessment. Anatomical study of 100 healthy human hearts.
15. Yoshida M, Wearden PD, Dur O, Pekkan K, Morell VO. Right Arq Bras Cardiol 1999;73:75–86.
ventricular outflow tract reconstruction with bicuspid valved 23. Sch€afers HJ, Schmied W, Marom G, Aicher D. Cusp height in
polytetrafluoroethylene conduit. Ann Thorac Surg 2011;91: aortic valves. J Thorac Cardiovasc Surg 2013;146:269–74.
1235–9. 24. Yamashita E, Yamagishi M, Miyazaki T, et al. Smaller-sized
16. Kim H, Sung SC, Chang YH, Lee HD, Park JA. A new expanded polytetrafluoroethylene conduits with a
simplified technique for making tricuspid expanded poly- fan-shaped valve and bulging sinuses for right ventricular
tetrafluoroethylene valved conduit for right ventricular outflow tract reconstruction. Ann Thorac Surg 2016;102:
outflow reconstruction. Ann Thorac Surg 2013;95:e131–3. 1336–44.

DISCUSSION
DR JAMES A. QUINTESSENZA (Lexington, KY): Great work, Our idea is totally different from the valve (in Japan), and for
Yoshio and Ross. The need for better solutions for our ever the aorta, yes, we need a sinus to protect the coronary artery. I do
growing population of patients for right ventricular outflow tract not know the exact answer for the pulmonary valve. And we are
continues, and there are many options available for these pa- creating some kind of small sinus after putting a stitch into each
tients, which underscores the fact that we really do not have a commissure. So maybe this can help to prevent the open stuck
great solution. An ideal valve might be one that is viable, able to position, I think.
repair itself, nonthrombogenic, will grow with the patient, be
easy to implant, and inexpensive—and clearly, that does not DR QUINTESSENZA: I noticed that thrombosis does not seem
exist. So most patients now receive a bioprosthesis with really to be a problem. I thought these patients were on aspirin. Can
good intermediate results, but all of these we know will ulti- you comment in terms of thrombogenicity or what you are doing
mately fail along a biodegenerative curve, which looks really to avoid thrombosis if anything?
good for about 5 years and then they start having a decay.
Your group at Wake Forest has continued to present on this DR OOTAKI: Actually we did not give anything for the patient,
series of patients, which actually was presented at the 2013 because I believe the Japanese group did not give anything
meeting here, your initial series. This is a continuation of that. after surgery, and they had nice results. I really respect their
And it is really about an expanded polytetrafluoroethylene results. So we did not. This is a cheap drug, and to prevent such
(ePTFE) valve, a synthetic, nonbiologic valve, which really has kind of thing maybe it is beneficial. Also at this time I do not
favorable hemodynamics, is nonthrombogenic, reasonably du- have any explantation of the valve, so I do not know. If I see at
rable, and durability, especially if we refer to the literature from least 1 or 2 patients, then I may start giving some aspirin, I
Japan, which has some reports up to 10 years with 95% freedom think.
from failure. So it really has the potential for great durability.
This valve that you present seems to be a bit simpler than some DR QUINTESSENZA: There are some data that suggests aspirin
of the Japanese reports, and specifically it avoids the need for inhibits the thickness of neointimal hyperplasia on ePTFE, so it
bulging sinuses, and there are none in your valve. And this may be of value.
brings me to my first question. What are your thoughts about the And just a final question. Do you have any plans to proceed
bulging sinuses for these types of prostheses? What function do with a transcatheter version of this valve? Interestingly, I saw a
they provide and do you do anything in your technique that report last year from Shanghai demonstrating a transcatheter
obviates the need for creating these bulging sinuses? ePTFE valve as a proof of concept model in a sheep. So this may
be an interesting avenue to pursue with these types of valves,
DR OOTAKI: First of all, thank you so much for teaching me that might provide improved durability compared with our
how to make a bicuspid valve. It is excellent. But after implan- available Melody or Sapien valves. Also, the ability to compress
tation in 3 patients, actually I noticed that the height is longer, so the ePTFE material into a smaller diameter delivery system may
that is why I started a tricuspid leaflet. be another advantage over thicker biologic tissue.
8 OOTAKI ET AL Ann Thorac Surg
TRICUSPID EPTFE VALVE 2017;-:-–-

DR OOTAKI: I did not see that paper. It was 3 years ago, and I regurgitant flow or the diastolic flow that fills up with blood,
did a same comment. If I have time, I really would like to make essentially, and drives the leaflet toward the center. I presume
some kind of percutaneous version, but I could not. This valve is that is the mechanism or part of the mechanism for closure and
the best valve so far, I think, but nothing came out for the need for those little commissural plication stitches.
commercialization. Gore-Tex has a patent that they filed in 2014,
and nothing comes up in the market. The idea is there, but DR TAKESHI SHINKAWA (Little Rock, AR): We presented our
creating such kind of materials in the market is pretty tough, I ePTFE tricuspid valved conduit result 3 years back at the STSA
think. meeting. The outcome was great. And I just want to comment
that we believe that the sinus of Valsalva did affect the longevity.
DR KRISTINE GULESERIAN (Miami, FL): Jim, one comment We started our Gore-Tex conduit in 2008, and we started with
just to maybe answer your question about having the sinuses. bicuspid, changed it to tricuspid and changed the sinus of Val-
When I was in John Mayer’s laboratory trying to bioengineer salva. So probably we can report a difference in the future.
pulmonary valves, of course we went back to look at the history, My other comment is about stitching the Arantius node in the
and Leonardo Da Vinci has described the sinuses of Valsalva middle leaflet, I was also part of the Japanese group, and we
beautifully. And the other piece that is sort of missing is the node made a nodule in the middle leaflet in the first series; however,
of Arantius to help with the central coaptation. And in compu- we saw more calcification, so the Japanese group stopped doing
tational flow dynamic studies, at least in bioengineered valves, it. That is for information. Thank you.
without the sinuses, the leaflets when they open hit the wall, and
that can affect their long-term durability in terms of creating DR OOTAKI: Thank you so much. It is good to have sinuses if
shear stress at the mid leaflet. With PTFE, one would think you we can, but honestly when I see the first paper from Japan, oh, no
would not need to have that concern, but I do not know that we way I can make it, right? I believe all of the surgeons here, you
really know the answer to that question. can make this valve, and I think this is the easiest way and the
simplest way and still have a nice result, I believe. I believe just a
DR QUINTESSENZA: There is that little plication stitch that is few surgeons are using these ePTFE valves, but you should
right there at the commissure, so that kind of reduces the change your mind and you should try. I have several patients
excursion of the leaflet all the way to the wall, that would limit it a who are waiting for conduit repair after their homograft (im-
little bit, and maintain that potential space so that during the plantation). I think homograft is not a good solution.

You might also like