AS901-005 Rev.

09/16

SC-CAL PLUS
HEMATOLOGY CALIBRATOR
CALIBRATOR

LOT PLUS1122
2022-12-05

The U.S. Units The International Units


Instrument Parameter Calibration Values Acceptable Range Calibration Values Acceptable Range
3 9
BC-3200 WBC 9.8 x 10 /uL 9.6 - 10.0 9.8 x 10 /L 9.6 - 10.0
6
RBC 4.60 x 10 /uL 4.52 - 4.68 4.60 x 1012/L 4.52 - 4.68

HGB 13.5 g/dL 13.3 - 13.7 135 g/L 133 - 137

MCV 87.1 fL 85.1 - 89.1 87.1 fL 85.1 - 89.1


3
PLT 253 x 10 /uL 241 - 265 253 x 109/L 241 - 265
3 9
BC-3600 WBC 9.2 x 10 /uL 9.0 - 9.4 9.2 x 10 /L 9.0 - 9.4
CD version 1.8(FDA) or 6
RBC 4.53 x 10 /uL 4.45 - 4.61 4.53 x 1012/L 4.45 - 4.61
greater
HGB 13.4 g/dL 13.2 - 13.6 134 g/L 132 - 136

MCV 86.0 fL 84.0 - 88.0 86.0 fL 84.0 - 88.0


3
PLT 249 x 10 /uL 237 - 261 249 x 109/L 237 - 261
3 9
BC-3600 WBC 9.5 x 10 /uL 9.3 - 9.7 9.5 x 10 /L 9.3 - 9.7
CD version lower than 6
RBC 4.64 x 10 /uL 4.56 - 4.72 4.64 x 1012/L 4.56 - 4.72
1.8 and/or manufacture
date before March of 2015 HGB 13.5 g/dL 13.3 - 13.7 135 g/L 133 - 137
THIS ANALYZER  MCV 85.4 fL 83.4 - 87.4 85.4 fL 83.4 - 87.4
NOT APPROVED 
3
FOR USE IN THE US PLT 257 x 10 /uL 245 - 269 257 x 109/L 245 - 269
3
BC-5390 WBC 9.70 x 10 /uL 9.50 - 9.90 9.70 x 109/L 9.50 - 9.90
6
RBC 4.53 x 10 /uL 4.45 - 4.61 4.53 x 1012/L 4.45 - 4.61

HGB 13.4 g/dL 13.2 - 13.6 134 g/L 132 - 136

MCV 85.0 fL 83.0 - 87.0 85.0 fL 83.0 - 87.0


3
PLT 251 x 10 /uL 239 - 263 251 x 109/L 239 - 263

Before using, refer to the instruction sheet for mixing directions.

Shenzhen Mindray Bio-Medical Electronics Co., Ltd.


Mindray Building, Keji 12th Road South, High-tech Industrial Park, Nanshan, ShenZhen 518057, P.R.China
Tel: +86 755 26582479 26582888
Fax: +86 755 26582934 26582500

U. S. Contact: Mindray North America


Address: 800 MacArthur Blvd., Mahwah, NJ 07430-0619,USA
Toll Free: +1 (800) 288 2121 AS901-005 Rev. 09/16