Operators Manual

CoaLyser

For in-vitro diagnostic use only!


Liability Disclaimer

DIALAB GmbH makes no express or implied warranty regarding this manual, its quality, performance, or
appropriate use regarding any type of specific procedure. Furthermore, this manual may be modified by
DIALAB GmbH without notice and without implying any obligation or liability on the part of the company.

Software Copyrights

All software for the CoaLyser (in the following Analyser-Software) is the user and instrument control Software
for the CoaLyser and is the intellectual property of the initial manufacturer. Intellectual property rights shall
remain with the initial manufacturer. DiaLAB GmbH is entitled to use the Analyser-Software and the printed
accompanying material. Any violation of property rights or copyright or trademark or using conditions may be
subject to legal action. The initial manufacturer reserves the rights to modify the software, hardware and
documentation without any prior written notice.

Intellectual Property

DIALAB GmbH CoaLyser is registered trademark of DIALAB GmbH.

No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or
translation into any language (human or computer) in any form, or by any means whatsoever, without the
prior written permission of DIALAB GmbH in behalf of the initial manufacturer.

Any reference to products, documentation, articles within this publication not expressly described as property
of the initial manufacturer and DIALAB GmbH shall be considered as intellectual property of the publishing
company of this certain product, documentation and or article.

Revision History

Analyser
Version Manual Date Software-Version -
(mm/dd/yy)
Release -

1.0 12/16/09 -- Initial version

1.1 04/16/10 V01.01 SW-Update

1.2 03/23/11 V01.04 SW-Update

1.3 08/30/12 V01.09 SW-Update

Dialab Produktion und Vertrieb von chemisch-technischen Produkten


und Laborinstrumenten Gesellschaft m.b.H.
A-2351 Wiener Neudorf, IZ-NÖ Süd, Hondastraße, Objekt M55
Tel.: +43-2236-660910-0; Fax.: +43-2236-660910-30
E-Mail.: office@dialab.at
Warranty

DIALAB GmbH warrants to the original purchaser for a period of one (1) year after the date of supply against
defects in material and workmanship and defects arising out from failure to conform to specification
applicable on the date of supply, provided they are properly installed, used and with accessories and
consumables in accordance with these operating instructions.

DIALAB GmbH further agrees to correct, either by repair, or, at its election, by replacement, any such defect
found on examination to have occurred, under normal use and service, during such one-year period,
provided DIALAB GmbH is properly notified in writing upon discovery of such defect.

DIALAB GmbH shall not be liable under this warranty for any defect arising from abuse of the system, failure
to operate and maintain the system in accordance with the documentation included with the instrument,
including repair service, alternation or modification of the system by any person other than authorized
personnel, or use of modified, changed, or previously used disposables, defects caused by
transportation/shipping and accidents, storms, lightening, fire, water/other liquids, other natural catastrophes,
theft, riots, plundering, the effects of war or other instances of acts of god.

The warranty of DIALAB GmbH set forth above and the obligations and liabilities of DIALAB GmbH there
under are exclusive and in lieu of all other remedies or warranties, express or implied, arising by law or
otherwise, with respect to the system delivered hereunder (including without limitation any obligation of
DIALAB GmbH with respect to merchantability, fitness for particular purpose and non-infringement). In no
event shall DIALAB GmbH be liable for incidental or consequential damages, however, arising and whatever
or not occasioned by DIALAB GmbH negligence.

This warranty shall not be extended or altered except by written instrument signed by DIALAB GmbH.

Exclusions of manufacturers warranty:

 The use of none permitted peripheral devices, e.g. printer.


 Improper use/operating faults and non-adherence to the user instructions.

 Attempted repairs by the customer or third-parties without authorization by manufacturer.


 Defective maintenance by third parties without authorization by manufacturer.
 Device defects due to power failure, heat losses or similar reasons.
 Accidents, storms, lightening, fire, water/other liquids, other natural catastrophes, theft, riots, plundering,
the effects of war or other instances of acts of god.
 Defects caused by transportation/shipping.
 The use of third party components, unauthorized consumables or accessories which do not comply to
manufacturers specifications.
 Non-authorised system changes.
 Resetting safety functions, deletion of passwords etc.
 Loss of customer data or software from repair and installation processes.
 Improper use of machine capacity or output.
 Inappropriate customer operating environment.
 Products from which the category plates, serial numbers, part numbers on the machine or machine parts
have been removed or changed.
 Wear and tear appearances at components like for example LCD.
 Breaks and scratches, dirtying on e.g. LCD displays or dirtying of the measuring channels.
Certifications

Specifications for Safety

This instrument conforms to the appropriate European Directives.

IVD The instrument type described in this manual bears an IVD and CE mark,
which confirms the compliance with essential requirements of the following:

European Directive: 98/79/EC In-vitro Diagnostic directive.

For further information see chapter 7.5 Safety Specifications


CONTENTS

TABLE OF CONTENTS
LIST OF FIGURES --------------------------------------------------------------------------------- 4 
1  INTRODUCTION ------------------------------------------------------------------------------- 8 
1.1  Intended use----------------------------------------------------------------------------------------- 8 
1.2  Purpose of this instruction manual ------------------------------------------------------------- 8 
1.3  Validity of this instruction manual -------------------------------------------------------------- 8 
2  HAZARD AND PRECAUTIONS ----------------------------------------------------------- 9 
2.1  Hazard Warnings and Safety Notices --------------------------------------------------------- 9 
2.2  Standard Symbols ------------------------------------------------------------------------------- 12 
3  SYSTEM DESCRIPTION ------------------------------------------------------------------ 13 
3.1  Analyser description ----------------------------------------------------------------------------- 13 
3.1.1  Touch Screen ----------------------------------------------------------------------------- 14 
3.1.2  Working Plate ----------------------------------------------------------------------------- 16 
3.1.3  Cuvette Ring Position ------------------------------------------------------------------- 17 
3.1.4  Sample Area------------------------------------------------------------------------------- 18 
3.1.5  Dilutor ---------------------------------------------------------------------------------- 20 
3.1.6  Pipettor ---------------------------------------------------------------------------------- 20 
3.1.7  Rinsing System --------------------------------------------------------------------------- 21 
3.1.8  Peripheral equipment and mains connection -------------------------------------- 23 
3.2  Software -------------------------------------------------------------------------------------------- 24 
3.2.1  Software control via touch screen ---------------------------------------------------- 25 
3.2.2  Software Overview ----------------------------------------------------------------------- 30 
3.2.3  Program Menus--------------------------------------------------------------------------- 31 
3.2.3.1  Main Menu ------------------------------------------------------------------------- 31 
3.2.3.2  Setup Menu ------------------------------------------------------------------------ 37 
3.3  Measuring Principle ------------------------------------------------------------------------------ 38 
3.4  Reagents ------------------------------------------------------------------------------------------- 39 
3.5  Load Cuvette Rings to analyzer -------------------------------------------------------------- 40 
4  INSTALLATION ------------------------------------------------------------------------------ 44 
4.1  General Setup ------------------------------------------------------------------------------------- 44 
4.2  Removing the Transport Protections -------------------------------------------------------- 46 
4.3  Assembly of the analyser----------------------------------------------------------------------- 46 
4.3.1  Mount the diluter syringe --------------------------------------------------------------- 46 
4.3.2  Connect the Rinsing System ---------------------------------------------------------- 47 
4.3.3  Connect an UBS Memory Device ---------------------------------------------------- 49 
4.3.4  Connect an external printer ------------------------------------------------------------ 49 
4.3.5  Connect an external Barcode Scanner (option) ---------------------------------- 51 
4.3.6  Connect a Personal Computer (RS232C for Service only) -------------------- 51 
4.3.7  Connect a HOST Computer (LAN interface) -------------------------------------- 51 
5  OPERATION ---------------------------------------------------------------------------------- 52 
5.1  System Start-up ---------------------------------------------------------------------------------- 52 
5.1.1  Switch On ---------------------------------------------------------------------------------- 52 
5.1.2  System Liquids---------------------------------------------------------------------------- 55 
5.1.3  Fill liquid system and flush system --------------------------------------------------- 55 
5.2  Routine Menu ------------------------------------------------------------------------------------- 57 
5.2.1  Load/Replace Cuvettes ----------------------------------------------------------------- 57 
5.2.2  Load Reagents---------------------------------------------------------------------------- 59 
5.2.3  Load Samples (Routine & STAT) ---------------------------------------------------- 68 
5.2.4  Load Controls (QC-Control) ----------------------------------------------------------- 77 
5.2.5  Start Measurements (Routine, STAT & Controls) -------------------------------- 81 
5.2.5.1  STAT Program -------------------------------------------------------------------- 83 

CoaLyser – Operators Manual 1.3 Page 1


CONTENTS

5.2.5.2  Continuous Loading ------------------------------------------------------------- 85 


5.2.5.3  Status ------------------------------------------------------------------------------- 88 
5.2.5.4  STOP -------------------------------------------------------------------------------- 89 
5.2.6  Show/print results ------------------------------------------------------------------------ 91 
5.2.7  Results Memory -------------------------------------------------------------------------- 93 
5.2.7.1  Show/print reaction curves ----------------------------------------------------- 94 
5.2.8  Analyser Status --------------------------------------------------------------------------- 97 
5.3  Quality Control (QC Program) ----------------------------------------------------------------- 98 
5.3.1  Control Setup----------------------------------------------------------------------------- 100 
5.3.1.1  Define additional QC-Control Parameter ---------------------------------- 104 
5.3.2  QC-Periods ------------------------------------------------------------------------------- 110 
5.3.2.1  View a QC-Period (statistics) ------------------------------------------------- 111 
5.3.2.2  View a QC-Period (chart)------------------------------------------------------ 113 
5.3.2.3  Print a QC-Period --------------------------------------------------------------- 115 
5.3.2.4  Finish a QC-Period before end of a month ------------------------------- 116 
5.3.2.5  Delete a regularly ended QC-Period --------------------------------------- 118 
5.3.3  View daily results ------------------------------------------------------------------------ 120 
5.3.4  Print all QC-Periods at a time -------------------------------------------------------- 121 
5.3.5  Delete all QC-Periods at a time ------------------------------------------------------ 122 
5.4  Setup Menu --------------------------------------------------------------------------------------- 125 
5.4.1  Im/Export test & reagent parameter via USB-Stick ----------------------------- 126 
5.4.1.1  Mount an USB-Stick ------------------------------------------------------------ 126 
5.4.1.2  Import data from USB-Stick--------------------------------------------------- 126 
5.4.1.3  Export data to USB-Stick ------------------------------------------------------ 129 
5.4.2  Select Working Modes ----------------------------------------------------------------- 131 
5.4.3  Setup Reagent definition -------------------------------------------------------------- 132 
5.4.3.1  Parameter Page 1--------------------------------------------------------------- 133 
5.4.3.2  Parameter Page 2--------------------------------------------------------------- 136 
5.4.4  Setup Test Definition ------------------------------------------------------------------- 140 
5.4.4.1  Liquids ----------------------------------------------------------------------------- 142 
5.4.4.2  Evaluation ------------------------------------------------------------------------- 144 
5.4.4.3  Determination -------------------------------------------------------------------- 150 
5.4.4.4  Show Flags (Define Results/Run list) -------------------------------------- 153 
5.4.4.5  Define QC-Control Parameter------------------------------------------------ 154 
5.4.5  Calibration--------------------------------------------------------------------------------- 160 
5.4.5.1  Edit Calibration for Clotting (manual calibration) ------------------------ 162 
5.4.5.2  Edit Calibration for Kinetic / Diff. / 1st deri (manual calibration) ------ 169 
5.4.5.3  Auto Calibration for Clotting (automated calibration) ------------------- 175 
5.4.6  Delete Results --------------------------------------------------------------------------- 192 
5.4.7  Setup Language ------------------------------------------------------------------------- 193 
5.4.8  Setup Date & Time ---------------------------------------------------------------------- 194 
5.4.9  Change Password ---------------------------------------------------------------------- 195 
5.4.10  Setup Beeper----------------------------------------------------------------------------- 196 
5.4.11  Setup STAT Autoprint ------------------------------------------------------------------ 197 
5.4.12  Define Headline-------------------------------------------------------------------------- 197 
5.4.13  Initiate Analyser Selftest --------------------------------------------------------------- 198 
5.5  Switch off analyser ------------------------------------------------------------------------------ 200 
6  SAFETY IN OPERATION -----------------------------------------------------------------201 
6.1  Maintenance and Hygiene -------------------------------------------------------------------- 201 
6.1.1  Maintenance frequency ---------------------------------------------------------------- 202 
6.1.2  Cleaning of the housing---------------------------------------------------------------- 203 
6.1.3  Clean Pipettor ---------------------------------------------------------------------------- 203 

CoaLyser – Operators Manual 1.3 Page 2


CONTENTS

6.1.4  Clean Containers/Sensors Replace Filter ----------------------------------------- 204 


6.1.5  Perform cleaning cycle ----------------------------------------------------------------- 204 
6.1.6  Perform washing cycle ----------------------------------------------------------------- 206 
6.1.7  Protocol of Maintenance Check ----------------------------------------------------- 207 
6.1.8  How to dispose of used disposables ----------------------------------------------- 207 
6.1.9  Disposal of the Analyser --------------------------------------------------------------- 207 
6.2  Troubleshooting --------------------------------------------------------------------------------- 208 
6.2.1  Application errors ------------------------------------------------------------------------ 208 
6.2.2  Error Flags in the Results/Run or Results/Memory List ------------------------ 209 
6.2.3  Errors during operation ---------------------------------------------------------------- 209 
6.2.4  How to change fuses ------------------------------------------------------------------- 213 
7  APPENDIX ------------------------------------------------------------------------------------214 
7.1  Materials supplied ------------------------------------------------------------------------------- 214 
7.2  Disposables/Consumables-------------------------------------------------------------------- 214 
7.3  Maintenance Kit --------------------------------------------------------------------------------- 215 
7.4  Technical Data ----------------------------------------------------------------------------------- 216 
7.5  Safety Specifications --------------------------------------------------------------------------- 218 
7.6  Mathematics -------------------------------------------------------------------------------------- 219 

CoaLyser – Operators Manual 1.3 Page 3


LIST OF FIGURES

LIST OF FIGURES
Figure 1 CoaLyser system ....................................................................................................................... 13
Figure 2 Touch Screen............................................................................................................................. 14
Figure 3 Reagent Area ............................................................................................................................. 16
Figure 4 Measuring Rotor ........................................................................................................................ 17
Figure 5 Cuvette Ring .............................................................................................................................. 17
Figure 6 Sample Area............................................................................................................................... 18
Figure 7 Internal Barcode Scanner .......................................................................................................... 19
Figure 8 Dilutor Module ............................................................................................................................ 20
Figure 9 Pipettor ....................................................................................................................................... 20
Figure 10 Container and level sensors for distilled water (green) and waste water (red) .............................. 21
Figure 11 Rinsing System .......................................................................................................................... 21
Figure 12 External interfaces (rear view) ................................................................................................... 23
Figure 13 Main Menu functions .................................................................................................................. 25
Figure 14 Area frame, scroll buttons, select window .................................................................................. 26
Figure 15 Select box, Check Box ............................................................................................................... 27
Figure 16 Movable Window........................................................................................................................ 28
Figure 17 Movable Window (Graph) .......................................................................................................... 29
Figure 18 Message Box ............................................................................................................................. 29
Figure 19 MAIN MENU .............................................................................................................................. 31
Figure 20 Run Preparation ......................................................................................................................... 32
Figure 21 Required Reagents .................................................................................................................... 33
Figure 22 Short form Results ..................................................................................................................... 34
Figure 23 Result Selection ......................................................................................................................... 35
Figure 24 Result/Memory ........................................................................................................................... 35
Figure 25 Status Menu ............................................................................................................................... 36
Figure 26 Setup Menu ............................................................................................................................... 37
Figure 27 Measuring Principle ................................................................................................................... 38
Figure 28 CuvCARD/ChipCARD Reader ................................................................................................... 41
Figure 29 Message: Load CuvCARD ......................................................................................................... 41
Figure 30 Message: Overwrite loaded balance .......................................................................................... 42
Figure 31 Message: No C-Rings loaded .................................................................................................... 42
Figure 32 Message: Overview of loaded C-Rings ...................................................................................... 43
Figure 33 Carry handles............................................................................................................................. 44
Figure 34 Mount of Syringe in Dilutor V1 .................................................................................................. 46
Figure 35 Mount of Syringe in Dilutor V2 ................................................................................................... 47
Figure 36 Connect the level sensor tubes to the analyser ......................................................................... 48
Figure 37 Container for distilled water. Container and tubes are coded green. ......................................... 48
Figure 38 Container for waste water. Container and tubes are coded red. ................................................ 48
Figure 39 USB-Stick Interface.................................................................................................................... 49
Figure 40 Status Line with active print ....................................................................................................... 50
Figure 41 Analyser Status Message .......................................................................................................... 52
Figure 42 Message: Insert Cuvette Ring .................................................................................................... 53
Figure 43 Pump Test ................................................................................................................................. 53
Figure 44 Pump Test failed ........................................................................................................................ 54
Figure 45 Pump Test OK ........................................................................................................................... 54
Figure 46 MAIN MENU .............................................................................................................................. 55
Figure 47 Run Preparation Menu ............................................................................................................... 56
Figure 48 Load C-Ring ............................................................................................................................... 57
Figure 49 Remaining C-Ring balance ........................................................................................................ 58
Figure 50 Load Reagents........................................................................................................................... 59
Figure 51 Test Select List .......................................................................................................................... 59
Figure 52 Reagent/Washer Table .............................................................................................................. 60
Figure 53 Renew all reagents .................................................................................................................... 61
Figure 54 Load Reagents........................................................................................................................... 61
Figure 55 Tools in Load Reagents ............................................................................................................. 62
Figure 56 Vials Menu ................................................................................................................................. 63
Figure 57 Vials Menu Example: PT ............................................................................................................ 63
Figure 58 Message: Renew Reagent ......................................................................................................... 64
Figure 59 Change reagent position ............................................................................................................ 64
Figure 60 Target Position ........................................................................................................................... 65
Figure 61 Change vial size......................................................................................................................... 65
Figure 62 Tools in Vials Menu.................................................................................................................... 66
Figure 63 Exit Load Reagents Menu .......................................................................................................... 67
Figure 64 Load Samples ............................................................................................................................ 68
Figure 65 Edit Sample ID ........................................................................................................................... 68
Figure 66 Entered Sample ID..................................................................................................................... 69
Figure 67 Using internal barcode scanner ................................................................................................. 70
Figure 68 Sample/STAT Positions ............................................................................................................. 71

CoaLyser – Operators Manual 1.3 Page 4


LIST OF FIGURES

Figure 69 Sample Rack 1..11..................................................................................................................... 71


Figure 70 Sample Rack 12..22................................................................................................................... 72
Figure 71 STAT Rack 1..3.......................................................................................................................... 72
Figure 72 Vial Frame ................................................................................................................................. 73
Figure 73 Sample with assigned vials ........................................................................................................ 74
Figure 74 Select test/assay ........................................................................................................................ 74
Figure 75 Clone Mode ............................................................................................................................... 75
Figure 76 Tools in Load Samples .............................................................................................................. 75
Figure 77 Run Preparation Menu ............................................................................................................... 76
Figure 78 Load Controls............................................................................................................................. 77
Figure 79 Control Positions ........................................................................................................................ 78
Figure 80 Vial Frame ................................................................................................................................. 78
Figure 81 Select test/assay ........................................................................................................................ 79
Figure 82 Tools in Load Samples .............................................................................................................. 80
Figure 83 Run Preparation Menu ............................................................................................................... 80
Figure 84 Menu Measurement ................................................................................................................... 81
Figure 85 Tools in Measurement Menu ..................................................................................................... 82
Figure 86 Running Measurement ............................................................................................................... 82
Figure 87 Prepare STAT run ...................................................................................................................... 83
Figure 88 Load STAT Samples .................................................................................................................. 83
Figure 89 Warning Message before STAT ................................................................................................. 84
Figure 90 Load STAT Samples .................................................................................................................. 84
Figure 91 Prepare continuous load ............................................................................................................ 85
Figure 92 Load additional samples ............................................................................................................ 85
Figure 93 Result list of finfished samples ................................................................................................... 86
Figure 94 Load additional samples ............................................................................................................ 86
Figure 95 Using internal barcode scanner ................................................................................................. 87
Figure 96 Menu Measurement ................................................................................................................... 88
Figure 97 Status Window ........................................................................................................................... 88
Figure 98 Immediate STOP ....................................................................................................................... 89
Figure 99 Warning user interrupt ............................................................................................................... 89
Figure 100 Pending Measurements ............................................................................................................. 90
Figure 101 Short Form Results .................................................................................................................... 90
Figure 102 Tools in Running Measurement Screen ..................................................................................... 91
Figure 103 Results/Run ............................................................................................................................... 91
Figure 104 Results/Memory 1 ...................................................................................................................... 93
Figure 105 Results/Memory 2 ...................................................................................................................... 94
Figure 106 Curve Input ................................................................................................................................ 94
Figure 107 Curve Input keyboard ................................................................................................................. 94
Figure 108 Tools in Results/Memory ........................................................................................................... 95
Figure 109 Reaction Curves ........................................................................................................................ 95
Figure 110 Tools in Reaction Curve Diagram .............................................................................................. 96
Figure 111 Status Menu ............................................................................................................................... 97
Figure 112 QC Main Page ........................................................................................................................... 99
Figure 113 Select Control Position ............................................................................................................. 100
Figure 114 Define Control Plasma ............................................................................................................. 101
Figure 115 Edit Control Plasma ................................................................................................................. 102
Figure 116 Delete Control Plasma ............................................................................................................. 102
Figure 117 QC Main Page ......................................................................................................................... 103
Figure 118 Main Menu ............................................................................................................................... 103
Figure 119 Setup Menu ............................................................................................................................. 104
Figure 120 Select Test Screen................................................................................................................... 104
Figure 121 Test Menu Main Page .............................................................................................................. 105
Figure 122 Test Menu Select Control ........................................................................................................ 105
Figure 123 Test Menu QC.......................................................................................................................... 106
Figure 124 Test Menu Select Control List .................................................................................................. 107
Figure 125 Test Menu selected Control ..................................................................................................... 107
Figure 126 Test Menu Main Screen ........................................................................................................... 108
Figure 127 Tools in Test Menu Main Page ................................................................................................ 109
Figure 128 QC Main Page ......................................................................................................................... 110
Figure 129 Select active QC-Period List .................................................................................................... 110
Figure 130 QC Result Table ...................................................................................................................... 111
Figure 131 QC-Period overview 1 .............................................................................................................. 112
Figure 132 QC-Period overview 2 .............................................................................................................. 112
Figure 133 Levey-Jennings Chart .............................................................................................................. 113
Figure 134 Tools in Edit Reagent Page 1 .................................................................................................. 114
Figure 135 Print QC-Period........................................................................................................................ 115
Figure 136 Tools in QC Main-Page ............................................................................................................ 115
Figure 137 Finishing QC-Period................................................................................................................. 116

CoaLyser – Operators Manual 1.3 Page 5


LIST OF FIGURES

Figure 138 Info Box printed QC-Period ...................................................................................................... 116


Figure 139 Delete QC-Period..................................................................................................................... 117
Figure 140 Deleted QC-Period Info ............................................................................................................ 117
Figure 141 Print/Finish Info ........................................................................................................................ 118
Figure 142 QC-Result Functions ................................................................................................................ 119
Figure 143 Delete QC-Period..................................................................................................................... 119
Figure 144 Deleted QC-Period confirmation .............................................................................................. 120
Figure 145 Daily results table..................................................................................................................... 120
Figure 146 QC Main Page ......................................................................................................................... 121
Figure 147 Print all QC-Periods ................................................................................................................. 121
Figure 148 Printing all QC-Periods ............................................................................................................ 122
Figure 149 QC Main Page ......................................................................................................................... 122
Figure 150 Delete all QC-Periods 1 ........................................................................................................... 123
Figure 151 Delete all QC-Periods 2 ........................................................................................................... 123
Figure 152 QC Main Page ......................................................................................................................... 123
Figure 153 Delete QC file........................................................................................................................... 124
Figure 154 Setup Menu ............................................................................................................................. 125
Figure 155 Stop System............................................................................................................................. 126
Figure 156 Im/Export Test & Reagent Parameters .................................................................................... 127
Figure 157 Import Test & Reagent Parameters ......................................................................................... 127
Figure 158 Importing Parameters .............................................................................................................. 128
Figure 159 Parameter Import OK ............................................................................................................... 128
Figure 160 Im/Export Test & Reagent Parameters .................................................................................... 129
Figure 161 Export Test & Reagent Parameters ......................................................................................... 130
Figure 162 Remove USB-Stick .................................................................................................................. 130
Figure 163 Mounting another USB-Stick .................................................................................................... 130
Figure 164 Setup Working Modes .............................................................................................................. 131
Figure 165 Select Reagent Screen ............................................................................................................ 132
Figure 166 Edit Reagent Main Screen ....................................................................................................... 133
Figure 167 Edit Reagent Page 1 ................................................................................................................ 133
Figure 168 Tools in Edit Reagent Page 1 .................................................................................................. 135
Figure 169 Edit Reagent Main Screen ....................................................................................................... 135
Figure 170 Edit Reagent Page 2 ................................................................................................................ 136
Figure 171 Tools in Edit Reagent Page 2 .................................................................................................. 138
Figure 172 Tools in Edit Reagents Main Screen ........................................................................................ 139
Figure 173 Select Test Screen................................................................................................................... 140
Figure 174 Test Menu Main Page .............................................................................................................. 140
Figure 175 Test Menu Liquids.................................................................................................................... 142
Figure 176 Tools in Test Menu Liquids ...................................................................................................... 143
Figure 177 Test Menu Evaluation .............................................................................................................. 144
Figure 178 Test Menu Tech Clotting .......................................................................................................... 145
Figure 179 Test Menu Evaluation/Kinetic .................................................................................................. 147
Figure 180 Test Menu Tech Kinetic ........................................................................................................... 148
Figure 181 Test Menu Evaluation/Kinetic .................................................................................................. 149
Figure 182 Test Menu Main Page .............................................................................................................. 150
Figure 183 Test Menu Determin. ............................................................................................................... 150
Figure 184 Test Menu Misc. Smoothing .................................................................................................... 152
Figure 185 Test Menu Main Page .............................................................................................................. 153
Figure 186 Test Menu Show-Flags ............................................................................................................ 153
Figure 187 Test Menu Main Page .............................................................................................................. 154
Figure 188 Test Menu Select Control ........................................................................................................ 155
Figure 189 Test Menu QC.......................................................................................................................... 155
Figure 190 Test Menu Select Control List .................................................................................................. 156
Figure 191 Test Menu selected Control ..................................................................................................... 157
Figure 192 Test Menu Main Screen ........................................................................................................... 158
Figure 193 Tools in Test Menu Main Page ................................................................................................ 158
Figure 194 Setup Menu ............................................................................................................................. 160
Figure 195 Select test Calibration .............................................................................................................. 161
Figure 196 Calibration Menu Main Page .................................................................................................... 161
Figure 197 Calibration Menu Clotting ......................................................................................................... 162
Figure 198 Calibration Menu Main Page .................................................................................................... 165
Figure 199 Calibration Values 1st evaluation .............................................................................................. 165
Figure 200 Calibration Curve Graph .......................................................................................................... 167
Figure 201 Tools in Calibration Graph Screen ........................................................................................... 168
Figure 202 Calibration Menu Main Page .................................................................................................... 169
Figure 203 Calibration Menu Kin/Diff/1st deri ............................................................................................. 169
Figure 204 Calibration Menu Main Page .................................................................................................... 171
nd
Figure 205 Calibration Values 2 evaluation ............................................................................................. 172
Figure 206 Tools in Calibration Value Screen ............................................................................................ 173

CoaLyser – Operators Manual 1.3 Page 6


LIST OF FIGURES

Figure 207 Setup Menu ............................................................................................................................. 175


Figure 208 Select test Calibration .............................................................................................................. 175
Figure 209 Calibration Menu Main Page .................................................................................................... 176
Figure 210 Calibration Menu Clotting ......................................................................................................... 177
Figure 211 Calibration Menu Main Page .................................................................................................... 179
Figure 212 Calibration Values (concentration) ........................................................................................... 179
Figure 213 Calibration Menu Main Page .................................................................................................... 181
Figure 214 Control & Calibrator Positions C1-C8....................................................................................... 182
Figure 215 Place Calibrator Position 1 ....................................................................................................... 183
Figure 216 Reagent Status List Calibration................................................................................................ 183
Figure 217 Tools in Calibration Measurement Menu ................................................................................. 184
Figure 218 Running Measurement Calibration........................................................................................... 185
Figure 219 Tools in Calibration Measurement Menu ................................................................................. 185
Figure 220 Reaction Curves Calibration .................................................................................................... 186
Figure 221 Running Measurement Calibration........................................................................................... 186
Figure 222 Results/Run List Calibration ..................................................................................................... 187
Figure 223 Warning Message Calibration .................................................................................................. 187
Figure 224 Warning Message Measuring Status ....................................................................................... 188
Figure 225 Auto Calibration finished .......................................................................................................... 188
Figure 226 Calibration Menu Main Page .................................................................................................... 189
Figure 227 Calibration data ........................................................................................................................ 189
Figure 228 Calibration Curve ..................................................................................................................... 190
Figure 229 Tools in Calibration Graph Screen ........................................................................................... 190
Figure 230 Delete Results.......................................................................................................................... 192
Figure 231 Delete Results Message Box ................................................................................................... 192
Figure 232 Setup Language....................................................................................................................... 193
Figure 233 Select Language ...................................................................................................................... 193
Figure 234 Setup Date & Time................................................................................................................... 194
Figure 235 Date & Time Settings ............................................................................................................... 194
Figure 236 Change Password.................................................................................................................... 195
Figure 237 Setup Password ....................................................................................................................... 195
Figure 238 Setup Beeper ........................................................................................................................... 196
Figure 239 Setup STAT autoprint .............................................................................................................. 197
Figure 240 Define Headline ....................................................................................................................... 197
Figure 241 Selftest ..................................................................................................................................... 198
Figure 242 Selftest Transmission ............................................................................................................... 198
Figure 243 Selftest finished........................................................................................................................ 199
Figure 244 Setup Menu ............................................................................................................................. 199
Figure 245 Shutdown Analyzer .................................................................................................................. 200
Figure 246 Shutdown Selection ................................................................................................................. 200
Figure 247 Perform Cleaning Cycle ........................................................................................................... 205
Figure 248 Perform Washing Cycle ........................................................................................................... 206
Figure 249 Change fuses, voltage selector ................................................................................................ 213 

CoaLyser – Operators Manual 1.3 Page 7


INTRODUCTION

1 INTRODUCTION
This chapter provides a global introduction to the CoaLyser system. It is
recommended to read this chapter first.

NOTE
Read this manual carefully before you attempt to operate the CoaLyser
system.

1.1 Intended use

The CoaLyser named hereafter as analyser, is a fully automated photo-


optical blood plasma hemostasis instrument (for human samples) capable
of performing clotting, chromogenic* and immunoassays* simultaneously in
random access or batch mode from primary and secondary tubes.

This IVD product is intended for professional use only. As IVD this
device is intended for human use and not for veterinary or general
laboratory applications.

For in-vitro diagnostic use only!

* Depending on the analyser model the system is capable to perform a wide


range of coagulometric, chromogenic and immunologic coagulation tests
such as Prothrombin time, activated and partial Thromboplastin time,
Fibrinogen, derived Fibrinogen and special tests such as single factor
assays, Anti-Thrombin III, Protein C, Protein S, C-Reactive Protein, D-
Dimer and others based on the wavelength available in different analyser
types.

However; the manual will indicate any difference between the available
model types if required. The analyser is intended for use by trained
operators in coagulation labs for in-vitro diagnostics.

1.2 Purpose of this instruction manual

The instruction manual describes the analyser and is directed to all users of
the system.

This instruction manual gives all information necessary for the installation,
operation, maintenance, storing, packing, and transport of the analyser.

Service information is handled separately.

Careful observation of all information, especially of hazards and precautions


will ensure the correct and safe operation of the analyzer. Therefore it is
absolutely necessary to read the instruction manual completely.

This instruction manual was prepared with greatest care. Should you have
any questions please refer to your distributor or to the manufacturer.

1.3 Validity of this instruction manual

All information in this instruction manual refers to this analyser.


For all requests please indicate type and serial number of your system
which can be found on the type plate on the back of the instrument.

CoaLyser – Operators Manual 1.3 Page 8


HAZARD AND PRECAUTIONS

2 HAZARD AND PRECAUTIONS

This chapter contains important hazard and precaution information.

NOTE Read chapter before you attempt to operate the analyser.

2.1 Hazard Warnings and Safety Notices

The cautions and safety regulations in this instruction manual meet


international classifications; refer to chapter 2.2 Standard Symbols.

This manual uses different types of symbols to alert you to important


information. Symbols and their associated information are labelled in text
where they occur and set off from surrounding paragraphs as shown in the
following examples.

Symbol warns of a risk of injury or of a risk to life (for example by electrical


shock).
Danger!

Symbol warns of a risk of injury or of the instrument being severely


damaged.
Caution!

NOTE Symbol introduces rules to be observed.

The following caution and safety regulations must be observed at all times:

1. Electrical safety
Check that the operating voltage is set correctly before you connect the
device to the main power supply.

To connect the device to the power supply, use only sockets which are
Danger!
grounded in order to keep the risk of an electrical shock as low as possible.

Use only grounded extension cables.

Never remove protective guards or secured components since you


could expose electrically live parts in this way.

Electrical connection contacts (plugs, sockets, etc.) can be electrically live.

Even after a device has been switched off, components (e.g.


capacitates) can be under voltage as the result of an electrical charge.

All current carrying parts are sources of danger for an electrical shock.
Never use the instrument on a moist surfaces (floor, worktable or
countertop), or place containers of liquids on top of the instrument. Liquids
spilled into the instrument may cause an electrical shock.

Carry out only the maintenance work and/or the replacement of parts
described in these operating instructions. Switch off the instrument
and disconnect it from all power sources before performing
maintenance or repair work.

CoaLyser – Operators Manual 1.3 Page 9


HAZARD AND PRECAUTIONS

Unauthorized work on the analyser can lead to the guarantee obligation


becoming null and void with necessary expensive service work to correct it.

All work which requires the analyzer to be opened may only be carried
out by a technician who is familiar with the risks related thereto.

Use only replacement fuses of the stated type and with the stated nominal
current.
- Never use fuses which have been "repaired".
- Never short-circuit the fuse holder.
2. Fire and explosion
hazards
Do not place any flammable or hazardous explosive material in the
proximity of the analyzer.

Caution!

3. Mechanical safety
(Analyzer is operating)
Never open screw-attached housing part! There is a risk of injury.

Avoid touching the pipettor and other moving parts during analyser
Caution! operation. Moving parts (fan, motor drives and pipettor) can cause injuries.

Do not open any covers before a measuring series has been terminated or
interrupted. Opening covers can damage the pipettor.

Ensure the pipettor does not stop over the sample positions or the
measuring rotor when the instrument is switched off.
- Use the protection cap for the pipettor if the system will not be operated
for a long time, to avoid damage and contamination.

Put the instrument out of operation and secure it against inadvertent use if
you come to the conclusion that it can no longer be operated safely.

This is to be assumed in the following cases:


- the instrument shows visible damage,
- the instrument no longer functions,
- the instrument has been stored or transported for a prolonged period
under unfavourable conditions.

4. Samples/Reagents Risk of infections


Avoid any contact between samples and/or test reagents with skin or
mucous membranes and any contact with parts of the instrument that come
into contact with samples/reagents. All consumables, e.g. cuvettes,
Caution! reagent vessels, pipette tips that may be used to transfer liquid
samples and reagents should be considered to be potentially infectious. If
specimens are spilled on the system, wipe them off at once and disinfect
the instrument (see Chapter 6.1). Reagents may cause irritation to the skin
and mucous membranes.

Always follow the manufacturer’s instructions for correct use given in


the literature accompanying the reagent.

Dispose of spent samples and waste reagents and all consumables that
have come into contact with them after completion of the measurements
strictly in accordance with statutory directives and laboratory guidelines.

NOTE Wear safety gloves! There is the risk of infection.

CoaLyser – Operators Manual 1.3 Page 10


HAZARD AND PRECAUTIONS

The correct Washing Solution must be used to avoid carry-over after using
thrombin containing reagents.
Caution!
The analysers Washing Solution has an open vial stability of 7 days at 20°C
to 30°C.

For cleaning the system waste tubing and pipettor the correct analyser
Cleaner must be used to avoid contamination and grow of alga.
Keep the Cleaner bottle closed if not used. Keep away from cold, heat and
sunlight. Store the Cleaner at room temperature.

5. Waste Liquids

The waste liquid in the waste container must be disposed in compliance


with applicable regulations.
Caution!
The liquid waste of the analyser is potentially infectious. To reduce the
possibility of exposure, use of personal safety gear (eye goggles, gloves,
etc.) is necessary. Any liquid waste material is to be emptied into approved
biohazard containers.

The liquid system tubing’s are not guaranteed to be resistant against


organic solvents. For this reason, use the analyser Cleaner only.

6. Accuracy and precision


of the measured results
In order to ensure a flawless operation of the analyser measure control
samples and/or standards and watch the function of the instrument closely.

Vorsicht!
Caution! False results may lead to an incorrect diagnosis and, therefore, risk for the
patient.

7. Cuvette Ring handling


If a new cuvette ring is loaded the analyser performs an automatic cuvette
recognition check.

Caution! Always place a new cuvette ring matching the marking on the cuvette ring
rotor and cuvette ring, see chapter 3.1.3 Cuvette Ring Position.

The software can distinguish between an empty and a full (used) cuvette
ring. However, there may be circumstances to those the liquid may not be
recognised, for example water (clear liquids, dried residues). Therefore
when loading new cuvette rings ensure an empty cuvette rings is always
being used.

If a new cuvette ring is replaced against a used cuvette ring after the
cuvette recognition has been performed, the following pipetting may lead to
cuvette overflow.

8. Restrictions for samples


and reagents
Resistance of cuvettes to organic solvents cannot be guaranteed. For this
reason, organic solvents should not be used unless they have been
expressly permitted. Only use the original manufacturer’s cuvettes and
Caution! mixers! Use cuvettes and mixers only once! Prior of each measurement
ensure a mixer is added to each cuvette.

9. User qualification
The analyser should only be operated by trained personnel. Ask you local
dealer or distributor for further information on the availability of user
trainings.
Caution!

CoaLyser – Operators Manual 1.3 Page 11


HAZARD AND PRECAUTIONS

10. Barcode Scanner


Be careful when using the barcode scanner laser light.

- Avoid direct eye contact with the laser beam.


- Follow the manufacturer’s instructions.
Danger! - Ensure the scanner was tested and authorized for use by the
manufacturer.
CAUTION
LASER LIGHT
DO NOT STARE INTO BEAM

670nm LASER DIODE


1.0 MILLIWATT MAX
CLASS II LASER PRODUCT

VORSICHT
LASERSTRAHLUNG, NICHT IN DEN LICHTSTRAHL BLICKEN,
SICHTBARE LASERSTRAHLUNG, LASERKLASSE 2

2.2 Standard Symbols

A large number of symbols are used in this Operator’s Manual, on the


instrument itself, and on available accessories and consumables. Their
meanings are given below:

Manufacturer Order Number

In Vitro Diagnostic Medical Device Serial Number

CE Conformity Batch Code

Refer to Operators Manual Do Not Reuse

Protect From Sunlight Caution

Temperature Limitation Danger

Humidity Limitation NOTE Important Information

Recyclable Corrosive

Electric Shock Warning Biological Risk

Fuse This side up

Separate collection, handling and


Switch off Analyser and disconnect
disposal for waste electrical and
all power sources before performing
electronic equipment and its
maintenance or repair work.
components

CoaLyser – Operators Manual 1.3 Page 12


SYSTEM DESCRIPTION

3 SYSTEM DESCRIPTION
This chapter contains detailed analyser description, its components and
provides detailed information on how to unpack and install the analyser the
first time.

3.1 Analyser description

The analyser represents a new the state of the art development and is a
microprocessor-controlled, two-channel photo-optical fully automated blood
plasma hemostasis analyser. The analyser is a modular construction.

Figure 1 CoaLyser system

1- Pipettor
2- Safety Shield
3- Cup Array for Sample Cups (Controls/Calibrators)
4- Reagent Block (removable)
5- Washer / Cleaner Position
6- Rinsing Station
7- Measuring / Cuvette Ring Rotor
8- Internal Barcode Scanner (depending on analyzer type)
9- STAT Positions
10 - Sample Positions
11 - USB Stick Interface (underneath) + handles to carry analyzer
12 - Touch sensitive colour display

CoaLyser – Operators Manual 1.3 Page 13


SYSTEM DESCRIPTION

Handling of safety shield:

It is recommended to always close the safety shield during analyser


Caution! operation to avoid injuries and damage of moving parts. Open the
safety shield only when the analyzer is not in operation.

The safety shield must be opened all the way to the top.

3.1.1 Touch Screen


The touch screen control panel offers several functions. On one hand, it is
designed for the visual display of the entire instrument control software. On
the other hand, it is used for inputting information, acknowledgment of
software messages, display of application and test ranges and for
communication with the user.

The LCD colour touch screen offers a visible area of 11.5 x 8.7cm and
provides a display resolution of 320 x 240 pixels, ¼ VGA).

Its touch sensitivity makes it possible to navigate your way round the
software simply by pressing the surface of the user control panel. Use either
your finger tip or a blunt object, e.g. a touch screen pen designed for this
purpose.

Never use a sharp object (finger nail, knife, letter opener) to press on
NOTE the control panel as it might cause irreparable damage to the control
panel surface.

Figure 2 Touch Screen

CoaLyser – Operators Manual 1.3 Page 14


SYSTEM DESCRIPTION

Function and Navigation Keys

The touch screen display of the analyser offer certain

Scroll-up by one window/page

Scroll-up by one line

Scroll-down by one window/page

Scroll down by one line

ESC key to cancel and leave a display/action

OK key to select an action

ESC& Exit key to leave a display without confirming


and saving input

Save&Exit to leave a display and save input

Tools key to enter into a selection menu offering


further actions

Keyboard (alphanumeric / numeric)


Alphanumeric keys, some with special symbols, for
text and numerical information input

Activates upper or lower case

Activates numeric keys and special symbols

Space bar, or overwriting input

Numeric keys, for numerical information input

Exponent key

Short dash key, e.g. to separate two numbers

Decimal or comma key

Arrow keys to move the cursor around the input field

For deleting the input to the left of the cursor

Confirm and save input

Leave input field (keyboard window) without


confirming and saving input

CoaLyser – Operators Manual 1.3 Page 15


SYSTEM DESCRIPTION

3.1.2 Working Plate


The analyser is equipped with a working plate which offers all positions to
load routine & STAT samples, reagents, controls & calibrator, washer,
cleaner and cuvette rings. This Working Plate is actually divided into three
single areas:

1- Reagent Area, consists of :


Washer & Cleaner Positions,
Reagent Block (removable),
Control & Calibrator Positions

2- Cuvette Ring Position, consists of :


Cuvette Ring / Measuring Rotor,
Wash / Rinsing Position

3- Sample Area, consists of :


Routine Sample Cup Positions,
STAT Positions

Reagent Area
The Reagent Area is a combination of different liquid positions at room
temperature. The Reagent Area offers positions for Washer and Cleaner,
reagent positions, diluents as well as positions for controls & calibrators.

Figure 3 Reagent Area

W1-W2 2x Washer & Cleaner Positions.

Reagent Block with 15 positions.


- (R1-R3) for 15ml reagent vials (left and right position stirred)
R1-R15
- (R4-R15) for 5ml reagent vials. The reagent block can be
removed from the working plate for external cooling.

C1-C8 8x Control & Calibrator Positions.

To place control and calibrator vessels, the Analyser uses the


controls/calibrator positions (C1-C8).

CoaLyser – Operators Manual 1.3 Page 16


SYSTEM DESCRIPTION

The control and calibrator positions accommodate the following tubes:

Type Volume Size Tube

Sample-Cup 4ml 61mm x 12.1mm diameter Secondary

Hitachi Cups Secondary

3.1.3 Cuvette Ring Position


The measuring module consists of:

- The Measuring rotor which has 32 cuvette positions. The temperature of


the measuring rotor is set to 37°C +/-0.4°C.
- The measuring rotor contains a triangular slot on the rotor surface for
proper ring placement (triangular arrow).
- The measuring rotor is equipped with 2 measuring channels, each
carrying a high-resolution LED photometer and reference channels, the
photometers are able to measure at a wavelength of 405nm, 570mn* and
740nm* (* depending on model type).

Figure 4 Measuring Rotor

Figure 5 Cuvette Ring

CoaLyser – Operators Manual 1.3 Page 17


SYSTEM DESCRIPTION

When placing the cuvette ring into the measuring rotor make sure the
NOTE
triangular markings of the cuvette ring match the same direction as
the triangular slot on the measuring rotor and the cuvette ring is
pushed all the way down.

Each well of the cuvette ring contains a 1 x 4 mm stir bar. These stir bars
are necessary to obtain accurate measurements due to generating a
homogeneous distribution of the plasma and the reagent. These stir bars
are driven magnetically by motors underneath the measuring rotor.

Only use the original cuvette bars and stir bars provided by the
NOTE
manufacturer to operate the system, otherwise you may risk wrong
measuring results and the warranty becoming null and void.

3.1.4 Sample Area


For placing samples into the analyser use the sample positions on the
working plate. Primary tubes and sample cups can be placed into the STAT
and sample positions.

Figure 6 Sample Area

1-3 3x STAT Positions.

S1 - S11 1x Sample Rack contains sample positions 1-11


S12 - S22 1x Sample Rack contains sample positions 12-22

CoaLyser – Operators Manual 1.3 Page 18


SYSTEM DESCRIPTION

The STAT and Sample Positions accommodate the following sample tubes:

Type Volume Size Tube

Sample-Cup 4ml 61mm x 12.1mm diameter Secondary

3ml 66mm x 11.5mm diameter


Sarstedt  Monovette  Primary
5ml 92mm x 11.5mm diameter

2.7ml 65mm x 10.3mm diameter


Terumo  Venoject  Primary
5ml 99mm x 12.4mm diameter

4,5ml 75mm x 12.3mm diameter


Beckton Dickinson Vacutainer Primary
5ml 98mm x 12.3mm diameter

Hitachi 2ml Primary

Sample IDs can be introduced to the analyser either by using an


NOTE
external barcode scanner (optional available) or by internal barcode
scanning (depending on analyzer type). To read how to connect and
operate a barcode scanner, see chapter 4.3.5 Connect an external
Barcode Scanner (option) and 5.2.3 Load Samples.

Figure 7 Internal Barcode Scanner

CoaLyser – Operators Manual 1.3 Page 19


SYSTEM DESCRIPTION

3.1.5 Dilutor
The analyser is equipped with a dilutor module fitted with a 250µl syringe and
a selection valve, which switches between dispensing and rinsing mode.

Figure 8 Dilutor Module

3.1.6 Pipettor
The Pipettor automatically executes all test preparations. It aspirates sample
plasma (as well as standards and controls) from the sample positions and
pipettes it into a measuring cuvette. It then aspirates one or more reagents
from reagent vials, placed in the reagent block or controls/calibrator area and
dispenses these into the measuring cuvette.

To prevent contamination, the pipettor is rinsed in the rinsing station after


each pipetting step.

The pipettor is a capacitive pipetting device heated at 37°C which is capable


to move into three directions:

- X-Direction: left and right


- Y-Direction: up and down
- Z-direction: backward and forward

Figure 9 Pipettor

CoaLyser – Operators Manual 1.3 Page 20


SYSTEM DESCRIPTION

3.1.7 Rinsing System


The rinsing system is used to clean the pipettor and the tube system. It
contains of a rinsing station, two pumps (distilled and waste pump), a dilutor
and two containers (distilled water, waste water). The containers can be
placed behind or beside the analyser

The distilled water container is marked with a green label, whereas the waste
container is marked with a red label.

dist waste
water water

Figure 10 Container and level sensors for distilled water (green)


and waste water (red)

Both containers are fitted with level sensors. When distilled water is empty or
when the waste water container is full, an appropriate message displays on
the screen.

The levels of the liquids will be checked:

- upon leaving the run preparation menu


- prior to the start of a run
- during the run
- prior to the start of a calibration

The distilled water tube is internally connected via the distilled water pump to
the dilutor valve. Distilled water is used as the system liquid and for
washing/rinsing of the pipettor. The rinsing solution is supplied to the rinsing
station and, after several rinsing cycles, pumped into the waste water
container by means of the waste water pump.

P
P

Pump
Distilled water Diluter with distilled Rinsing- Waste water
container water pump station container

Figure 11 Rinsing System

CoaLyser – Operators Manual 1.3 Page 21


SYSTEM DESCRIPTION

Besides the rinsing solution the analyser offers two positions within the
reagent area, see section Reagent Area, used to place vials with Cleaner
and or Washing Solution.

The Washing Solution is used during intensive rinsing to clean the pipettor
from thrombin containing reagents (thrombin in activator).

The Cleaner on the other hand is used to clean the system to prevent the
growth of algae and bacteria and to remove possible contamination (mainly in
the waste water system and pipettor).

There are different types of rinsing programs available:

Type Cycle description

Flush 10s System water, 10 sec. rinsing

System water, 1 sec. rinsing


Rinsing
(only available in Reagent Setup)
Washer 1 (Pos. W1)
- 3 sec. rinsing with system water
Intensive 1 - 150µl Washer 1
- 3 strokes up to 250µl
- 3 sec. rinsing with system water
Washer 2 (Pos. W2)
- 5 sec. rinsing with system water
Intensive 2 - 150µl Washer 2
- 3 strokes up to 250µl
- 5 sec. rinsing with system water
Washer 1 (Pos. W1)
- 1.5 sec. rinsing with system water
Intensive 3 - 100µl Washer 1
- 5 strokes up to 250µl
- 1.5 sec. rinsing with system water

Intensive 4 4 sec. rinsing with system water

Cleaner ( Pos. W2)


- 1 sec. rinsing with system water
Cleaning Cycle
- 200µl Cleaner
(replace Washer 2)
- 3 strokes up to 100µl
- 20 sec. rinsing with system water

CoaLyser – Operators Manual 1.3 Page 22


SYSTEM DESCRIPTION

3.1.8 Peripheral equipment and mains


connection
In general no additional equipment is necessary to operate the analyser.

The integrated module and system software controls the system and
manages all data handling. However, for external data transfer and storage
of parameters and results the analyser is equipped with several interfaces.

Using these interfaces the analyser can be connected to an external


computer using either the RS232C interfaces or the integrated LAN
interface. Measured results, reagent and test-specific data, software
screens can even be printed by means of an external USB printer device
connectable to the analyser or stored by on an USB memory device.

The backplane of the analyser carries the mains supply connection. The
instrument is designed for mains supply voltages between 100V and 240V,
at 50/60Hz.

The analyser mains isolation switch is above the power supply connection.
This is accommodated in a housing that also carries the fuse holder and the
power supply selector.

The analyser is delivered without power supply cable. The cable


NOTE must meet the local regulations (e.g. IEC 320, VDE, CSA-C22.2, No.
21 and No. 49).
The recommended length is 1.5 m, and the minimum cross section 3x
0.75 mm2. A cold device plug is used at the system’s end.
Use only fuses recommended by the manufacturer. Observe the
instructions on the plate located on the back side of the analyser (see
picture below). The analyser is set to 230V by the manufacturer,
change fuses (supplied in accessories) in power filter, if analyzer will
be operated in 115V, see chapter 6.2.4 How to change fuses.

Figure 12 External interfaces (rear view)

CoaLyser – Operators Manual 1.3 Page 23


SYSTEM DESCRIPTION

Power Filter, contains:


- Fuse holder
1-
- Power Switch I = ON, O = OFF
- Mains Connection
2- DC BC-Scanner (Power connection external Barcode Scanner)
3- Extern. BC-Scanner (Serial interface Barcode Scanner)
4- USB Printer (External printer interface)
5- LAN Interface
6- 2x RS 232C interfaces (Com 1 / Com 2)
7- Waste Water (Connection for Level Sensor red)
8- Dist. Water (Connection for Level Sensor green)

3.2 Software

The operating system software interface allows the user to operate the
analyser as an independent measuring instrument without any peripheral
equipment.

The analyser software is stored in a ROM and is started after the instrument
is switched on. It controls the analyser through the start-up initialization and
controls all functions for analyser setup and the analysis programme.

Some analyser software application areas are protected with an additional


PIN code (password). After inputting a pre-defined PIN code, other software
levels licensed to the authorised user may be accessed; these enable the
setting and adjustment of additional instrument and test/reagent specific data or
parameters.

Without access to the requisite PIN code, these software levels remain locked.
The next software level (expert level) has been set to default level 12345 by
the manufacturer. This may be amended at any time, see Chapter 5.4.9
Change Password.

User Interface
The dialogue between analyser and user takes place via the built-in touch
screen; refer also to chapter 3.1.1 Touch Screen

The menus and menu items are selected by pressing the surface of the
user control panel either by a finger tip or a blunt object, like a touch screen
pen designed for this purpose.

Never use a sharp object (finger nail, knife, letter opener) to press on
NOTE the control panel as it might cause irreparable damage to the control
panel surface.

To operate the analyser software different kind of menus, buttons, boxes


and displays are available to select certain menus and functions, to enter
information and parameters and to display results, analyser information and
to view graphics.

The following pages provide further information on available menus,


software areas, functions and defines the functionality of the analyser
software by using the touch sensitive colour display.

CoaLyser – Operators Manual 1.3 Page 24


SYSTEM DESCRIPTION

3.2.1 Software control via touch screen

Figure 13 Main Menu functions

Displays associated functions combined in sub-


1- Area Frame
menus or titles specific software areas.

Are used to access additional menu / software


2- Buttons
areas, to confirm entry or to exit a screen.

Displays active or inactive functions, date & time


3- Status Line information, individual messages, irregularities
and problems occurred during measurement.

To enter information using the virtual keyboard.


By pressing onto the white input fields a
4- Input Field
keyboard opens, allowing either alphanumeric or
only numeric entries.

CoaLyser – Operators Manual 1.3 Page 25


SYSTEM DESCRIPTION

Figure 14 Area frame, scroll buttons, select window

Displays associated functions combined in sub-


1- Area Frame
menus or titles specific software areas.

Are used to scroll up and down within a screen to


choose between certain available selections.
2- Scroll buttons
Single arrows allow movement by one line
Double arrows allow movement by window/page

Are used to access additional menu / software


3- Buttons
areas, to confirm entry or to exit a screen.

4- Select Window Displays available selections to choose.

CoaLyser – Operators Manual 1.3 Page 26


SYSTEM DESCRIPTION

Figure 15 Select box, Check Box

Select between individual options by pressing the


1- Select Box
arrow buttons.

To set individual parameters On/Off press onto


the check box field.
2- Check Box
OFF= no set 
ON = set 

To enter information using the virtual keyboard.


By pressing onto the white input fields a
3- Input Field
keyboard opens, allowing either alphanumeric or
only numeric entries.

CoaLyser – Operators Manual 1.3 Page 27


SYSTEM DESCRIPTION

Figure 16 Movable Window

Information on windows not being displayed in


total may be moved by keeping the finger onto
Moveable
1- the screen and then moving up/down and
Window
left/right. The window will move according to the
fingers direction.

Are used to scroll up and down within a screen to


choose between certain available selections.
2- Scroll buttons
Single arrows allow movement by one line
Double arrows allow movement by window/page

To enter information using the virtual keyboard.


By pressing onto the white input fields a
3- Input Fields
keyboard opens, allowing either alphanumeric or
only numeric entries.

CoaLyser – Operators Manual 1.3 Page 28


SYSTEM DESCRIPTION

Figure 17 Movable Window (Graph)

Graphs of calibration and reaction curves not


Moveable being displayed in total on the screen may be
1- Window moved by keeping the finger onto the screen and
(Graph) then moving up/down and left/right. The window
will move according to the fingers direction.

Figure 18 Message Box

Provide information, Messages on Errors,


1- Message Box
Warnings and further available functions.

CoaLyser – Operators Manual 1.3 Page 29


3.2.2
MAIN MENU

Run Preparation Measurement Results / Run Results / Memory Status Shutdown Setup CuvCARD
press press press press press press press press
SYSTEM DESCRIPTION

button button button button button button button button


Software Overview

Load C-Ring Start the run Results of last run Results Storage Analyzer status Switch off analyzer Passwords Load C-Ring balance
- Load new cuvettes - Start measurement - Show joblist results - Show all results - View analyzer status - Switch off - Edit/print Password/PIN
- View reagent list - Print joblist results - Show results by sample - Print analyzer status - Restart
- View cuvette info - Show results by test - Show version info (quit) Reagent Setup
Load Reagents
- View sample info - Display curve graphs - Define reagent set up
- Load tests - Show joblist - Print results / graphs QC
- Print reagent set up
- Set reagent vial size - Print joblist
- Re-new reagents Date & Time press
- Remove reagents ... during run button

- Define positions - Set up date & time


- Add STAT runs
- Load Samples during runs Im / Export data Control Setup
Load Samples - Show analyzer status - Im/Export parameters - Define Controls
- Load patient samples - Stop/interrupt run - Update Software - Select positions
- Edit /enter patient ID - Show reaction curves
- Select sample vial - Print reaction curves Selftest QC-Periods
- Clone mode - Show results - Initiate analyzer selftest - View QC-Periods
- Prepare joblist - View QC-Results
Test Setup - Show Charts

CoaLyser – Operators Manual 1.3


Load Controls - Define test set up Daily Results
- Print test set up
- Select/Load Controls - View daily QC-results
- Set control vial size Calibration
- Renew Controls All QC-Periods
- Select QC-reagent - Set STAT Autoprint - Edit calibration of test
- Define Working Modes - Enter calibration curves - Print QC-Periods
- Set Beeper - Auto calibration - Delete QC-Periods
Rinsing Modes
- Define Headline - Print calibration data
- Flush
- Intensive wash Delete Results
- Cleaning - Delete results storage
- Maintenance
Language
- Set analyzer language

Page 30
SYSTEM DESCRIPTION

3.2.3 Program Menus


This section gives an overview about the available software menus and
software areas to provide a quick guideline thought the overall analyser
software. See chapter 5.2 Routine Menu and 5.4 Setup Menu for an
accurate description of the detailed software functions.

3.2.3.1 Main Menu


The Main Menu automatically appears after the analyser has been switched
ON and the initialization process of all analyser modules has successfully
been finished. Sometimes you must confirm a message, e.g. about the
Shield Control being disabled first.

Figure 19 MAIN MENU

The Main Menu is separated into two main sections

Menu Button Function

Within this area all submenus related to run


Measurement Area preparation, measurement, result organization
and status information are located.
Within this area all submenus related to general
analyser setup, reagent and test definition,
Setup Area
quality control (only available if software is
installed) and calibration settings are located.
By pressing into the input field a numerical
keypad opens allowing the entry of a PIN Code
Password/PIN
enabling access to other available software
levels.

By pressing the reset button you will reset the


Reset
PIN Code to the lowest available access level.

Switching off the Analyser or reboot the


Shutdown
system.

Submenus of the Main Menu are described in the following sections.

CoaLyser – Operators Manual 1.3 Page 31


SYSTEM DESCRIPTION

Measurement Area

Run Preparation

Figure 20 Run Preparation

Menu Button Function

Load C-Ring Load of cuvette ring into the measuring rotor.

Displays the reagent load window, the positioning of


Load Reagents reagents and washers in the reagent area and
selection of usable reagent/washer vials.
Displays the sample and STAT load window,
Load Samples positioning of samples/STAT and selection of
sample tubes usable.
Load Controls
Displays the controls load window, positioning of
(only available if QC
controls and selection of control tubes usable.
is installed)

Selection of available rinsing, washing and cleaning


Rinsing Modes
cycles.

OK Return to MAIN MENU

CoaLyser – Operators Manual 1.3 Page 32


SYSTEM DESCRIPTION

Measurement

Figure 21 Required Reagents

Menu Button Function

Reagent Status Displays joblist information according to reagents


List and washers available and needed.
Information about the cuvette ring, start cuvette,
Cuvettes required, cuvettes to run measurement and
available cuvettes in total.

Information about the samples and/or controls to be


Samples+Controls
measured, fully and partially processed samples.

Selection of additional functions, like displaying


Tools
joblist and timeline, printing the display

Start the Run Button to initiate the measurement

Exit Return to MAIN MENU

CoaLyser – Operators Manual 1.3 Page 33


SYSTEM DESCRIPTION

Results

Figure 22 Short form Results

This window displays results of the last run in an abstract form only. This list
will be deleted as soon as new results are available out of a next run. To
define the result information shown in this window; see chapter 5.4.4.4
Show Flags (Define Results/Run list).

Menu Button Function

Print To print short form result list via an external printer.

OK Return to MAIN MENU

CoaLyser – Operators Manual 1.3 Page 34


SYSTEM DESCRIPTION

Ring Storage

Figure 23 Result Selection

Menu Button Function

Shows all available results of the result storage in a


All
list
Shows all available results of the result storage
Sample
sorted by sample.
Shows all available results of the result storage
Test
sorted by test.

Figure 24 Result/Memory

CoaLyser – Operators Manual 1.3 Page 35


SYSTEM DESCRIPTION

Status

Figure 25 Status Menu

Menu Button Function


Displays the analyser status concerning
temperatures of rotor and pipettor, Container Status,
Display cuvette ring information, irregularities in the results
Information measured, QC information as well as other errors
appeared during measurement. Any problems
occurring may be marked coloured.

Print To print the Status Screen

OK Return to MAIN MENU

CoaLyser – Operators Manual 1.3 Page 36


SYSTEM DESCRIPTION

3.2.3.2 Setup Menu

Setup

Figure 26 Setup Menu

Certain areas within the Setup may only be accessed by entering a PIN
Code in the Main Menu. The software indicates the level of password
required accessing into another level, see chapter 5.4.9 Change
Password.

Menu Button Function Software Level

Passwords To set passwords to access different software level. Expert

Reagent Setup To define and edit reagent specific parameters Anyone

Date & Time To edit analyser date & time. Anyone

Im/Export To im/export test & reagent parameters via USB Stick. Anyone
Initiates an automatic self test procedure of the
Selftest Anyone
analyser.
To define and edit test specific parameters. To
Test Setup Anyone
combine individual reagents.
Perform calibration and management of calibration
Calibration Anyone
data, curves.
Delete Results To delete all results of the ring memory Expert
Selection of different available languages for user
Language Anyone
interface
Check Box to activate/deactivate automated print of
STAT Autoprint Anyone
STAT parameters
Select Box to select whether the measuring run shall
Working Modes Anyone
be by sample or by test
Check Box to activate/deactivate acoustic signals
Beeper Anyone
performed by the analyser.
Headline Defines print headline Anyone

Tools Access to other functions Anyone

OK Return to MAIN MENU Anyone

CoaLyser – Operators Manual 1.3 Page 37


SYSTEM DESCRIPTION

3.3 Measuring Principle

The analyser uses a photo-optical detection principle. Clotting, chromogenic


and immunologic assays can be run simultaneously in random access,
depending on the available wavelength installed in the analyser.

The measuring module consists of a two channel LED photometer, attached


to the measuring rotor. Depending on the analyser there are either

- 2 channel LED photometer with a wavelength 405nm only or


- 2 channel LED photometer with wavelengths at 405nm/570nm/740nm

Test cuvette

LED Photodiode

perm. magnet

Stirrer - Motor
Photometer
Electronics
Display

Figure 27 Measuring Principle

A light beam passes through the cuvette and is received by the photodiode.
As soon as the starting reagent is added to the plasma, the measuring time
starts.

Any change in the light transmittance is detected and converted into an


electrical signal by the photodiode. The signal is converted into a digital
value which is transformed to the Extinction value.

When the assay is complete, the raw data (seconds for coagulation assays
and mE/sec (Extinction) for chromogenic assays) are processed and
reported.

CoaLyser – Operators Manual 1.3 Page 38


SYSTEM DESCRIPTION

3.4 Reagents

For proper coagulation analysis we recommend the use of Dialab adapted


lines of reagents, controls and buffers or other recognized manufacturers.
Always read the information leaflet in the pack and observe the instructions
given by the reagent manufacturer.

Utilize reagents and controls only according to directions as provided


NOTE
by the reagent manufacturer to avoid incorrect measuring results or
malfunctions of the analyser.

Contamination
With the application of different reagents, and here especially reagents
containing thrombin, there is a danger of reagent carry-over.

NOTE When adding reagents and samples the pipettor is exposed to


reagents and hence a point of contamination.

If liquid residue or dried-on remnants can be seen on the workplate of the


analyser, remove them with laboratory disinfectant solution and cotton buds.

Other things you may need to or


must take into consideration:

Only use the analyser according to the required ambient conditions.


NOTE Protect the measuring channels from direct sunlight or other light
sources.

Always place a cuvette ring in the measuring rotor before pipetting.


Ensure each cuvette is equipped with a mixer. Pipetting reagent or
sample into the measuring vials and or tubes can significantly
contaminate the analyser and could even render it faulty, so that
expensive cleaning or repair might become necessary.

Only use the original manufacturer’s cuvette rings and mixers as they
are subject to strict quality control. The use of other makes of cuvette
rings and any associated instrument problems as a result, will
invalidate your guarantee.

Use cuvette rings once only. Multiple use of cuvette rings may
produce incorrect results; this in turn can become an indirect hazard
for the patient.

Perform regular quality control checks. Please refer to the guidelines


of use provided by the reagent manufacturer.

CoaLyser – Operators Manual 1.3 Page 39


SYSTEM DESCRIPTION

3.5 Load Cuvette Rings to analyzer


(depending on model)

To guarantee safe, correct and precise coagulation analysis the


manufacturer has executed extensive trails and studies concerning the
analyser and its related original cuvettes and mixers. To ensure proper
coagulation analysis of the system even in the routine work the use of
original cuvettes and mixers, as provided by the manufacturer, is required.

Depending on the model type the analysers are equipped with a cuvette
management system, which allows loading of a cuvette ring credit balance
of original cuvette rings to the analyser, by means of a cuvette card, so-
called CuvCARD. The individually loaded cuvette ring credit balance
enables the analyser to perform coagulation measurements in the same
quantity of cuvette rings as loaded on the analyser.

At first delivery the analyser is preloaded with a cuvette ring balance


NOTE
by the manufacturer corresponding to the quantity of cuvette rings as
delivered in the accessory box, i.e. 10 cuvette rings (320 single
cuvettes).

Each cardboard box of original cuvette rings (containing 6 boxes of 10


cuvette rings), suitable for your analyser, is also equipped with a
corresponding CuvCARD. This CuvCARD, as you will find in the cardboard
box, carries a quantity of 60 cuvette rings. By using this CuvCARD you can
load the cuvette rings, 10 at a time, to the analyser.

Once the cuvette ring credit balance is fully used, meaning if measurements
of the same quantity of the loaded cuvette ring balance have been
performed the analyser will automatically request for re-load respectively
reminds you to order new cuvette rings.

As far there is still a cuvette ring credit balance on the CuvCARD,


NOTE
keep the CuvCARD safe. Loss or damage of the CuvCARD would
waste the cuvette ring balance on the Card.

The cuvette ring balance once loaded to the analyser will only be
NOTE
deducted when the sample plasma has been added to the cuvette. If
the run is stopped before this time the cuvette is not deducted.
However, once sample plasma is added to a cuvette the cuvette is
considered being used and the counter will deduct it from the loaded
cuvette ring balance.

CoaLyser – Operators Manual 1.3 Page 40


SYSTEM DESCRIPTION

The following describes how to load cuvette ring balance from a CuvCARD
to the analyser.

 At first insert the CuvCARD into the CuvCARD reader slot with the
memory chip facing down and in the direction of insertion. Leave the
CuvCARD in the reader, see picture below.

Figure 28 CuvCARD/ChipCARD Reader

 From the Main Menu, press the button CuvCARD to access cuvette
ring loading procedure.

The following display appears:

Figure 29 Message: Load CuvCARD

CoaLyser – Operators Manual 1.3 Page 41


SYSTEM DESCRIPTION

 Confirm this message by pressing the button Yes if you intent to load a
cuvette ring balance of 10 new cuvette rings to the analyser.

 By pressing the button No you will leave this message box without
loading any cuvette ring balance.

After confirming above message box by pressing the Yes button and in
case the analyser still carries a cuvette ring balance the following
ATTENTION message box appears:

Figure 30 Message: Overwrite loaded balance

When loading a cuvette ring balance from the CuvCARD to the


NOTE
analyser, the remaining balance eventually still on the analyser will be
overwritten. Make sure the balance on the analyser is fully used prior
of loading a new balance from the CuvCARD to avoid any waste of
cuvette rings.

 By pressing the button No you will leave this message box without
loading any cuvette ring balance. The existing balance will be kept
available the following message appears:

Figure 31 Message: No C-Rings loaded

CoaLyser – Operators Manual 1.3 Page 42


SYSTEM DESCRIPTION

 Confirm this message by pressing the button YES if you intent to load a
cuvette ring balance of 10 new cuvette rings to the analyser.

The analyser starts the loading procedure from the CuvCARD; make sure
the CuvCARD is in the CuvCARD reader during the complete loading
process.

Once the cuvette ring balance is loaded, the following overview appears:

Figure 32 Message: Overview of loaded C-Rings

The overview table provides certain information on the loading status of the
cuvette rings loaded to the analyser as well as status information of the
CuvCARD balance remaining.

Print
 Press the Print button to print the Overview of loaded C-Rings window.

OK
 Press the OK button to leave this window.

Disposal of the CuvCARD may be done in the domestic waste after


NOTE the cuvette balance on the CuvCARD is fully used and the CuvCARD
is devalued.

Please refer to chapter 6.2.3 Errors during operation further


NOTE information on status messages generated by the CuvCARD handling.

Refer to chapter 5 OPERATION, to continue with measurements.

CoaLyser – Operators Manual 1.3 Page 43


INSTALLATION

4 INSTALLATION
This chapter provides instructions on how to unpack, install and prepare the
analyser for use.

Before performing any tasks in this chapter, review the Hazard and
NOTE
Precautions listed in chapter 2 of this manual.

4.1 General Setup

Remove the analyser from its packing and verify that the instrument has not
been damaged during transport and the accessories are complete, see
chapter 7.1 Materials supplied. Please notify your distributor or the
manufacturer immediately in the event that the shipment was damaged or
incomplete.

 For the safety of the instrument and yourself the analyzer should be
carried or moved by two persons.

To carry the analyzer use the handles located on the left and right bottom
side of the analyzer.

Figure 33 Carry handles

 Make sure a stable and level area for installation was selected and the
installation area is free from dust, liquids, solvents, gas and acid vapors.

 There should be free space of min. 15 cm around the analyser having free
access to the interface panel and power switch in the rear of the analyser
and to install the liquid containers.

 Lift the system from the box (with assistant), and place on the installation
area.

 Remove all foam inserts and set aside.

Pay attention to the pipettor!


Be careful not to damage the pipettor, flat cable and steel tubing; any
damaged or defective parts must be replaced by a service technician before
operating.

CoaLyser – Operators Manual 1.3 Page 44


INSTALLATION

 Place the system in the correct position on the table.

 Remove all tapes and transport protection devices.

 Make sure the power cable will reach the outlet without using any
extension cords.

 There should be enough light for proper operation, but no direct sunlight
and artificial light, e.g. lamps to avoid premature formation of algae in the
tubing system.

 The ambient temperature must range from +15°C to a maximum of +


30°C. Avoid significant temperature changes, humidity and voltage
variations, see chapter 7.4 Technical Data.

The analyser should be placed in an area that allows access to the


NOTE main switch at all times. In case of any emergency, it may be
Ensure there necessary to immediately switch off the analyser.

NOTE
The analyser power cord must comply with the national standards in
the country the equipment is supplied.

Do not damage any packaging material, and keep them for possible
NOTE later use, including the transport protection devices. Store them
away from heat and humidity.

CoaLyser – Operators Manual 1.3 Page 45


INSTALLATION

4.2 Removing the Transport


Protections

The analyser is secured with several transport protection parts to avoid any
damage of the analyser or components during its transportation.

Prior of use the analyser has to be released from any of these transport
protection devices.

Transport protection devices are:

1. Cardboard insert inside transfer arm: Protecting the pipettor from


moving in Z-direction.
2. Cardboard insert backside: Protecting the pipettor transfer arm from
moving in X-direction.
3. Red plastic plate with cable ties: Protects the pipettor from moving in Y-
direction.
4. Empty red cuvette ring in the measuring rotor to protect from
contamination of the measuring cells.

4.3 Assembly of the analyser

Before connecting the analyser to the main power supply, complete all
instructions in this section.

4.3.1 Mount the diluter syringe


Mount the 250 µl dilutor syringe to the dilutor of the system according, see
figure below. When properly mounted, the syringe should not be wedged,
and must be watertight.

For Installation proceed as follows:


Dilutor version 1
 Fill in distilled water in the syringe
 Open the black piston holder coverage by turning it ccw
 Push down the piston holder with two fingers (instrument is powered off).
 Screw the cylinder into the diluter valve.
 Insert the syringe with the piston in the holder.
 Take care that the syringe is free of air and completely filled with DI
water
 Move the piston holder upwards until the piston hits the piston holder
bottom
 close the piston holder coverage by turning it cw.

Figure 34 Mount of Syringe in Dilutor V1

CoaLyser – Operators Manual 1.3 Page 46


INSTALLATION

Dilutor version 2
 Fill in distilled water in the syringe
 Unscrew the black piston holder coverage
 Push down the piston holder with two fingers (instrument is powered off).
 Insert the syringe with the piston in the holder.
 Screw the cylinder into the diluter valve.
 Take care the syringe is free of air and completely filled with DI water
 Attach the piston holder coverage

Figure 35 Mount of Syringe in Dilutor V2

4.3.2 Connect the Rinsing System


The analyser is equipped with two water containers (distilled water and
waste water). All containers should be located to the left side, behind the
analyser. The distilled water temperature may not exceed the specified
temperature of the analyser (+15°C to + 30°C)

 Place the distilled water container (with the green label) and the empty
waste water container (with the red label) to the left rear of the analyser.
Fill the distilled water container with at least 1 Liter of distilled water.

 Place the colour-coded sensors into the appropriate containers (green


for dist. water, red for waste water). Ensure there are no air bubbles in
the tubing of the distilled water container, or in the container itself.

 Plug the green and red colour-coded couplings into their matching
sockets on the back of the system (see figure 11). A „click“ indicates the
proper seating of the connectors.

The system will only function if the distilled and waste water tubes have
been connected properly to the system.

NOTE The system must be switched OFF.

CoaLyser – Operators Manual 1.3 Page 47


INSTALLATION

The couplings are color-coded:


red for
waste water

green for Re
d
dist water

Gr
ee
n

< NOTE >


Decoupling of the tube adapter

Press on the metal shield to


release the tube adapter.

Figure 36 Connect the level sensor tubes to the analyser

The distilled water container must have at least 1 Liter of distilled


NOTE water. The waste water container should be emptied every day. Be
careful not to have air bubbles in the distilled water container as
welll as in the tube. The distilled water container may not be placed
in direct sunlight.

Container

dist
water

Figure 37 Container for distilled water. Container and tubes are coded green.

waste
water

Figure 38 Container for waste water. Container and tubes are coded red.

CoaLyser – Operators Manual 1.3 Page 48


INSTALLATION

4.3.3 Connect an UBS Memory Device


The analyser is equipped with two USB interfaces. While one USB interface
is located at the rear side of the analyser utilized for connecting an external
printing device the second USB interface can be found at the right side of
the analyser within the carrying handle, see figure below.

Figure 39 USB-Stick Interface

This interface offers the option to connect a USB memory device to record
and save measuring data, i.e. results, reaction curve and patient data and to
import and export test and reagent specific data using a commercially
available USB stick. These data can be viewed later on the display of the
analyser. To record measuring data first insert a USB Stick into the USB
interface located as shown in the figure above. Leave the USB Stick in the
interface, the USB Stick will me mounted once the analyser is switched on.

The USB Stick should have a memory of at least 1GB.

4.3.4 Connect an external printer


The analyser offers the possibility for connecting an external USB printer
device using the USB interface at the rear side of the analyser; this allows
print put of results, reaction curves, instrument- and test parameters.

 To connect the USB printer device use the USB cable provided in the
accessories of the printer. Use the corresponding USB interface in the
rear of the analyser to connect the printer to the analyser.

 Connect the power supply plug to the printer.

 The analyser will automatically detect the USB printer once connected
to the analyser. To verify the connection is set correctly the symbol print
in the status line of the analyser must turn green.

Recommended printer types / interface parameter settings:

CoaLyser – Operators Manual 1.3 Page 49


INSTALLATION

Interface Manufacturer Type/model


USB Hewlett For example:
Packard Deskjet 6940, Deskjet D5560, Officejet
6000 and later models

Specifications:
- USB interface
- Support of PCL 3 printer protocol

Observe the printer manufacturer’s operating instructions,


NOTE
including the default interface parameter settings.

Figure 40 Status Line with active print

CoaLyser – Operators Manual 1.3 Page 50


INSTALLATION

4.3.5 Connect an external Barcode


Scanner (option)
 Connect the optional barcode scanner to the analyser. Use the
recommended type only.

 Connect the barcode scanner data cable to the RS232C socket (marked
Extern. BC-Scanner) on the analyser (see figure 12 No.2/3).

 Plug the power supply into the socket (marked DC BC-Scanner) next to
it.

4.3.6 Connect a Personal Computer


(RS232C for Service only)
The analyser is equipped with two serial RS 232C interface ports on the
rear side of the analyser.

The RS 232C interfaces are specified as follows:

Com 1: This interface is used for manufacturer operation only.

Com 2: This interface is used for service operation only.

 Before switching on the computer, the PC must have been connected to


the external periphery instruments. Use the data cable recommended
by manufacturer or dealer only. Then connect the PC to the power
supply. Do not switch the computer on yet!

 Now connect the data cable to the socket (marked Com1 or Com2) at
the back of the analyser and to the interface port of your computer.

4.3.7 Connect a HOST Computer


(LAN interface)
Additionally the analyser disposes of a LAN (Local Access Network)
interface on the rear side of the analyser which is supposed to be used for
connecting external PC and HOST Systems for data transfer.

The LAN interface can be used as follows:

Uni-Directional: To transfer analyzer parameters and patient results to a


connected Personal Computer or HOST System.

Bi-Directional: To receive sample joblists from a HOST computer and to


transfer patient results and parameter setting to a
connected HOST or Personal Computer. The Bi-
Directional HOST communication requires additional
software available to the system. Please contact the
manufacturer or your local distributor to get more
information on the software and how to set up a HOST
communication.

CoaLyser – Operators Manual 1.3 Page 51


OPERATION

5 OPERATION
This chapter provides a step-by-step introduction to routine operation of the
analyser.

NOTE
Before performing any tasks in this chapter, review the Hazard and
Precautions listed in chapter 2 of this manual.

5.1 System Start-up

To keep the information on routine operation as simple as possible, this


chapter contains basic information. Detailed information on the individual
menus and screens is located in 5.2 Routine Menu.

After installation use the following procedure to start up the analyser.

5.1.1 Switch On
Set the power switch on the back side of the analyser to the position I = ON,
see also figure 12.

The analyser now runs through a full initialization procedure checking all
internal modules automatically. The following describes the sequence of the
initialization procedure:

Figure 41 Analyser Status Message

Above window provides information about the number of wavelength


available in this analyzer model, customer ID and analyzer software installed
as well as the release date.

In case the system still carries a cuvette ring in the measuring rotor the
following box appears. This message indicates the number of used cuvettes
in the cuvette ring.

CoaLyser – Operators Manual 1.3 Page 52


OPERATION

Figure 42 Message: Insert Cuvette Ring

By pressing the button YES a new unused cuvette ring must be inserted into
the measuring rotor first.

By pressing the button No the cuvette ring still in the measuring rotor will be
further used.

 Confirm this message box by pressing Yes or No according to the


corresponding action.

During the initialization procedure the analyser beside other modules also
checks the status of the liquid system, meaning the function of the DI pump,
waste pump and level detection.

Figure 43 Pump Test

This procedure will be repeated three times if any liquid levels deviate from
the expected values, e.g. due to any malfunction of the pumps and level
detection or if the analyser is taken into operation the first time after initial
setup when the liquid system is not yet filled up with system water.

In seldom cases if the liquid system can not be filled up during this
procedure or due to any malfunction of the pumps and level detection the
following message appears:

CoaLyser – Operators Manual 1.3 Page 53


OPERATION

Figure 44 Pump Test failed

By pressing the button OK the analyser allows access to all menus but locks
all areas required to perform any measurements due to safety reason.

In this case confirm this message by pressing the button OK and repeat the
initialization procedure once again.

 Switch off analyser by pressing the button Shutdown and press Restart
to again start the initialization procedure.

If the problem persists check chapter 6.2 Troubleshooting for further


information.

After successful pump and level detection test the following message
appears:

Figure 45 Pump Test OK

Once the initialization process is finished the MAIN MENU appears in the
display.

CoaLyser – Operators Manual 1.3 Page 54


OPERATION

Figure 46 MAIN MENU

5.1.2 System Liquids


Before starting a routine run operation:

 Check the level of the distilled water container.


 Empty the waste water container.

When the distilled water container is empty, use the following procedure:

 Fill the distilled water container with distilled water at room temperature,
see also chapter 4.3.2 Connect the Rinsing System

 Place the two level sensors in the containers making sure that the colour
of each level sensor matches its corresponding container.

5.1.3 Fill liquid system and flush system


For transport safety reasons the analyser has been delivered in dry
condition, meaning without any system liquids. Before using the analyser it
is necessary to perform a rinsing process to fill all system modules with
liquid (pumps, dilutor and tubing’s). This procedure shall also be performed
to clean and remove any air bubbles in the tubing. Further you can check
whether all tube and screw connections have been installed properly.

Should you notice any leaking, tighten the connections to the point
NOTE
that they are watertight.

 From the Main Menu, press the button Run Preparation the following
display appears.

CoaLyser – Operators Manual 1.3 Page 55


OPERATION

Figure 47 Run Preparation Menu

 Use the Select Box arrow buttons in the “Do Washing Cycle” frame until
Flush 10s appears in the select box.

 By pressing the button Perform the flush/rinsing procedure starts. The


analyser will rinse the system with distilled water.

 After the initial installation this procedure should be repeated at least 5


times until the tubing is completely filled with system liquid and the
tubing is free of any air bubbles inside the tubing.

 Check the system again for any leakage’s and re-mount, if necessary,
all connections and tubes before repeating the rinsing cycle.

Daily Maintenance
On a daily maintenance basis it is recommended using the above described
procedure even to clean and remove possible air bubbles in the tubing (if
system has already been used).

CoaLyser – Operators Manual 1.3 Page 56


OPERATION

5.2 Routine Menu

Following system start-up, the analyser is now ready for routine operation.
The next chapters describe how to:

- Load and replace a cuvette ring


- Load Reagents & Samples
- Start Measurements
- Show/print results
- Handle the result ring storage
- Show/print reaction curves
- Display the analyser status

5.2.1 Load/Replace Cuvettes


This chapter describes how to load and replace a new cuvette ring and
handling of a used cuvette ring.

 From the Main Menu, press the button Run Preparation to access the
run preparation area.

 From the Run Preparation Menu, press the button Load C-Ring the
following display appears.

Figure 48 Load C-Ring

Perform one of the following:

Option 1:
1. If applicable first remove any used cuvette ring from the measuring rotor
if this ring can not or shall not be further used.

2. Afterwards or in case there is no cuvette ring, insert a new cuvette ring


into the measuring rotor. Check chapter 3.1.3 Cuvette Ring Position to
ensure correct insertion of the cuvette ring and ensure a cuvette ring
balance is loaded to the analyser, see chapter 3.5 Load Cuvette Rings
to analyzer.

3. Press Done to confirm the cuvette ring has been correctly inserted.

CoaLyser – Operators Manual 1.3 Page 57


OPERATION

After pressing the button Done an Info message box appears showing the
remaining balance of cuvette rings available on the system. When loading a
new cuvette ring and confirming this procedure with Done will deduct a
loaded cuvette ring from the cuvette ring balance available on the analyzer,
see chapter 3.5 Load Cuvette Rings to analyzer.

Figure 49 Remaining C-Ring balance

Each load of a new cuvette ring will deduct one ring from the loaded
NOTE
cuvette ring balance after confirming with Done.

Make sure first that all cuvettes of a cuvette ring already in the
measuring rotor have been fully used before loading a new cuvette
ring. Otherwise you may risk wasting unused cuvettes.

Option 2:
1. If there is a partly used cuvette ring inside the measuring rotor you may
decide to further use this cuvette ring until all cuvettes are consumed. In
this case press Quit to continue.

CoaLyser – Operators Manual 1.3 Page 58


OPERATION

5.2.2 Load Reagents


 From the Main Menu, press the button Run Preparation to access the
run preparation area.
and/or
 From the Run Preparation Menu, press the button Load Reagents the
following display appears.

Figure 50 Load Reagents

This display offers information on currently loaded reagents and required


washers and provides access to certain other menus needed to set up and
load reagents.

Load Test
 By pressing onto the not used buttons (---) a list of selectable tests will
be opened.

Figure 51 Test Select List

 Use the scroll-up and down buttons to move up and down inside the
displayed list. The selected test will be highlighted, example APTT.

 Once a test (here APTT) is selected press OK to confirm selection. By


pressing ESC you will leave this window without any test selection.

CoaLyser – Operators Manual 1.3 Page 59


OPERATION

Required Reagents
The selected test will be shown in the Load Test list as well as in the
Reagent Position section, see example APTT.

Figure 52 Reagent/Washer Table

The Reagent Position Table is an exact reflection from the reagent area of
the analyser. It assigns the positions to where the reagents have to be
placed on the analysers, see chapter 3.1.2 Working Plate.

 Place the reagents in the assigned position on the analyser


corresponding to the positions as shown in the display, example APTT
reagent = R4, CaCl2 = R5.

Make sure to always place the reagents exactly on the same position
NOTE
as shown in the display to avoid any failure in your analysis.

Crossed out and grey fields, e.g. R1 show free not used positions on
the Reagent Area.

None crossed out and grey fields, e.g. R2 (PT) show already used
positions and the reagent assigned to it.

Required Washer
It is mandatory to always load washer on the analyser! The Washer
position W1 will be checked prior to each run. If a washer is detected during
the reagent check procedure the corresponding position will be displayed in
the Washer Area on the screen. Not used Washer positions will be crossed
out.

 If not yet done place a vial containing washing solution on the Reagent
Area.

CoaLyser – Operators Manual 1.3 Page 60


OPERATION

Renew all loaded reagents and


washers

 By pressing the button renew all all loaded reagents and washers will
be renewed and the stability timer re-starts. The following window
appears:

Figure 53 Renew all reagents

By pressing the button YES renew all procedure will be started once the
load reagent screen (Safe&Exit) is being pressed.

By pressing the button No renew all procedure won´t be started.

 Confirm this message box by pressing Yes or No according to the


corresponding action, the following window appears:

Tools

Figure 54 Load Reagents

The Tools button in the Load Reagent Menu offers several options and
access to other menus and actions.

 Press the Tools button to access further additional areas the following
Select Window appears.

CoaLyser – Operators Manual 1.3 Page 61


OPERATION

Figure 55 Tools in Load Reagents

The following Tools are available:

Available Tools Description


By pressing this button the vial menu screen will
Hardcopy be printed out, provided a printer is connected
to the analyser.
By pressing this button the vial menu opens
vials allowing change of reagent positions and vial
sizes.

By pressing this button you will leave this


Quit
window without any changes.

By pressing this button a previously selected


Remove test
test in the Load Test list will be removed.

By pressing this button all test in the Load Test


Remove all list incl. all run-settings for each test will be
removed.

CoaLyser – Operators Manual 1.3 Page 62


OPERATION

Change reagent positions and


vial size (vials button)

 By pressing the vials button the following window appears:

Figure 56 Vials Menu

The following options are available:

Renew Reload or refill reagents or washer by pressing on the reagent or washer


position you intend to reload. Crossed out fields can not be accessed.

The selected reagent or washer position will be highlighted, example PT.

Figure 57 Vials Menu Example: PT

 If you intend to reload or refill a reagent or re-set the reagent stability


times press the Renew button and a message box to renew the reagent
appears.

 If you intend to reload or refill a washer only press the W1 (not crossed
out) button and a message box to renew the washer appears.

CoaLyser – Operators Manual 1.3 Page 63


OPERATION

Figure 58 Message: Renew Reagent

 Replace or refill the assigned reagent or washer in the Reagent Area


and confirm this message by pressing the OK button.

 The levels of the replaced/refilled reagents or washers will be detected


again by leaving the Load Reagent Menu with the Save&Exit button.

Make sure to place the reagents and washers always in the assigned
NOTE
position on the analyser. Otherwise you risk failure in your analysis.

Change reagent position To change a certain reagent position in the Reagent Area press on the
reagent you intend to change. Crossed out fields can not be accessed.

The selected reagent will be highlighted, example PT as shown in figure 57.

 Press the Change pos. button and the Target Position Menu appears.

Figure 59 Change reagent position

This window provides information about the reagent position still available
for changing. All none visible positions are already used by loaded reagents
or the selected vial size does not fit into the hidden target position.

CoaLyser – Operators Manual 1.3 Page 64


OPERATION

 Select one of the available reagent positions by pressing on the


corresponding button, example R1. The reagent will be moved to the
selected position, the following display appears.

Figure 60 Target Position

 PT has now changed to position R1. Move the PT vial in the Reagent
Area to the new assigned position.

Make sure to place the reagents and washers always in the assigned
NOTE
position on the analyser. Otherwise you risk failure in your analysis.

Change reagent vial size To change another size of a reagent vial, for example if reagent bottles of
certain reagent manufacturers are used press again on the reagent you
intend to change the vial size. Crossed out fields can not be accessed.

The selected reagent will be highlighted, example APTT.

Figure 61 Change vial size

CoaLyser – Operators Manual 1.3 Page 65


OPERATION

The loaded reagent positions are framed by different colours. Each colour of
a frame indicates the vial size selected corresponding to the frame Vials on
the right side of the display. In the Vials frame only the available vial sizes
can be selected.

The following vial sizes are available:

Reagent positions Selectable vial sizes Colour

15ml Glass Magenta


12ml Glass Green
R1 - R3
5ml Glass Yellow
4ml Plastic Blue

5ml Glass Yellow


R4 - R15 4ml Plastic Blue
Cup Orange

 Select one of the available reagent vial sizes by pressing on the


corresponding button in the Vials frame, example 4ml Plastic. The
coloured frame of this reagent will be changed according to the vial size
selected.

Tools In certain menus the Tools button offers several options and access to other
menus and actions

 Press the Tools button in the Vials Menu to access additional areas the
following Select Window appears.

Figure 62 Tools in Vials Menu

CoaLyser – Operators Manual 1.3 Page 66


OPERATION

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

By pressing this button you will be asked to


ESC confirm whether all changes done in the vial
menu shall be confirmed or waived.
By pressing this button the vial menu screen will
Hardcopy be printed out, provided a printer is connected
to the analyser.

Save&Exit (Vials Menu) By pressing the Save&Exit button all changes in the Vial Menu will be
confirmed and returns back to the Load Reagents Menu.

Figure 63 Exit Load Reagents Menu

Save&Exit By pressing the Save&Exit button in the Load Reagent Menu confirms all
changes in this menu and starts level check of all reagents and washers on
board.

CoaLyser – Operators Manual 1.3 Page 67


OPERATION

5.2.3 Load Samples


(Routine & STAT)
 From the Main Menu, press the button Run Preparation to access the
Run Preparation Menu.
and/or
 From the Run Preparation Menu, press the button Load Samples the
following display appears.

Figure 64 Load Samples

Within this menu the complete sample handling can be controlled and set. It
further provides information on currently loaded samples, samples to load,
selectable sample cups/tubes and tests available for each sample.

Enter/Edit Sample ID
(manually)
 Select a sample by pressing onto the sample button once, e.g. (sample
01). The sample box will be highlighted.

 By double-clicking onto the sample buttons, e.g. (sample 01) an


alphanumeric keyboard opens allowing the entry of sample/patient
names/numbers.

Figure 65 Edit Sample ID

 Use the keyboard to enter the sample/patient ID required, see chapter


3.1.1 Touch Screen for further information. For sample/patient ID entry
15 alphanumeric characters are available.

CoaLyser – Operators Manual 1.3 Page 68


OPERATION

When entering a sample/patient ID certain characters are not allowed


NOTE to be used since these may create conflicts when transmitted to a
connected HOST computer later on. The following characters are
permitted to be used if samples/patient results are entered:

ABCDEFGHIJKLMNOPQRSTUVWXYZabcdefghijklmnopqrstuvwxyz012
3456789!&()[]{}=#~+-,^@

Avoid spaces between symbols, words & numbers.

 Press the Enter button once the sample/patient ID is entered.

Figure 66 Entered Sample ID

 Entered sample/patient ID´s appear in the sample box as shown above.


If no sample/patient ID is entered sample 01…sample 22 will remain as
shown.

Enter/Edit Sample ID (with


barcode scanner)
Instead of entering sample ID´s manually you may also load sample ID´s by
using either an external barcode scanner, see first chapter 4.3.5 Connect
an external Barcode Scanner (option) or an internal barcode scanner
(availability depends on analyzer type).

 Select a sample by pressing onto the sample button once, e.g. (sample
01). The sample box will be highlighted.

 Do NOT enter a sample ID manually.


Samples are identified by the sample ID on the barcode label.
Depending on the type of barcode, the sample ID can be numeric,
alphanumeric or alphabetic.

 Using an external barcode scanner direct the scanner window to the


barcode label of the vial/tube.

 Press the button on the barcode scanner. Continue pressing the button
until a beep confirms entry of ID.

 The scanned sample ID should now appear in the selected sample box.

CoaLyser – Operators Manual 1.3 Page 69


OPERATION

 Using the internal barcode scanner hold the sample vial/tube on the
upper end with your finger tips and guide the vial/tube into the scanning
position located the working plate, see picture below. Keep the barcode
in direction of the scanner light beam.

 The barcode will automatically detect and scan the sample ID.

 The scanned sample ID should now appear in the selected sample box.

 Place the sample vial/tube into the sample rack using the same position
as shown in the sample select box.

Figure 67 Using internal barcode scanner

A large number of barcodes can be used. If the system does not


NOTE
accept a certain barcode contact the manufacturer or enter the patient
ID manually.

 Repeat described steps to enter more samples.

Select Sample Rack


The analyser is equipped with 22 sample positions divided in two racks of 11
samples plus one rack of 3 STAT positions, see chapter 3.1.4 Sample Area.

CoaLyser – Operators Manual 1.3 Page 70


OPERATION

Figure 68 Sample/STAT Positions

1-3 3x STAT Positions.


S1 - S11 1x Sample Rack contains sample positions 1-11
S12 - S22 1x Sample Rack contains sample positions 12-22

 To select between the sample and STAT racks use the arrow keys as
described as follows:

Figure 69 Sample Rack 1..11

CoaLyser – Operators Manual 1.3 Page 71


OPERATION

Figure 70 Sample Rack 12..22

Figure 71 STAT Rack 1..3

The Load STAT Samples window allows entry and definition of STAT
(Emergency loading) samples. These samples will be handled with highest
priority in the measurement sequence. 3 sample positions are continuously
available to initiate immediate STAT measurement. The loading of samples
as well as the loading of reagents follows the normal procedure for loading
of the samples and reagents see chapters 5.2.2 Load Reagents and 5.2.3
Load Samples.

 To enter sample/patient ID´s precede as previously described.

 To load additional reagents for a STAT sample, refer to chapter 5.2.2


Load Reagents.

CoaLyser – Operators Manual 1.3 Page 72


OPERATION

Select Sample Vial


A various selection of primary and secondary tubes/vials may be assigned to
each sample/patient, see chapter 3.1.4 Sample Area for further information.

 Press onto the sample/patient you intend to assign a sample vial/tube.


The sample/patient button will be highlighted.

 Select a vial/tube from the list as shown in the vial frame of the display,
see next figure.

Figure 72 Vial Frame

Below table shows the vials/tubes placeable in the sample area. Based on
the vial/tube used select the corresponding colour in the vial frame of the
display.

The STAT and Sample Positions accommodate the following sample tubes:

Type Volume Size Tube

Sample-Cup 4ml 61mm x 12.1mm diameter Secondary

Sarstedt Monovette 3ml 66mm x 11.5mm diameter


Primary
Sarstedt Monovette 5ml 92mm x 11.5mm diameter

Terumo Venoject 2.7ml 65mm x 10.3mm diameter


Primary
Terumo Venoject 5ml 99mm x 12.4mm diameter

Beckton Dickinson Vacutainer 4,5ml 75mm x 12.3mm diameter


Primary
Beckton Dickinson Vacutainer 5ml 98mm x 12.3mm diameter

Hitachi 2ml Secondary

Make sure to place the same sized (coloured) vial/tube into the
NOTE
sample area as selected in the vial frame. Only one type can be
selected at a time. Otherwise you risk failure in your analysis or
damage of the pipettor.

CoaLyser – Operators Manual 1.3 Page 73


OPERATION

Once the vial/tube has been selected in the vial frame the same coloured
dot beside the sample will appear on all samples.

Figure 73 Sample with assigned vials

Assign test(s) to a sample


(joblist preparation)
To assign tests (assays) to a sample you may select out of the tests/assays
as listed in the Load Sample Menu. Within this list all tests/assays will be
shown as previously loaded, see chapter 5.2.2 Load Reagents.

 Press onto the sample/patient if you intend to assign a test/assay. The


sample/patient button will be highlighted.

 Select the test/assay from the list as shown on the right side of the
display; see figure below example PT and APTT by pressing onto the
test/assay buttons.

Figure 74 Select test/assay

 Selected tests/assays will be highlighted. Repeat this procedure for all


samples you intent to assign tests/assays for.

CoaLyser – Operators Manual 1.3 Page 74


OPERATION

Clone mode To make creating a joblist easier the Clone mode offers the possibility to
copy/clone certain joblist information from one sample already defined to
another undefined sample.

 If you intend to clone a sample press first the sample already set. The
sample will be highlighted.
 Press the Clone mode button.
 Now press onto the sample buttons you intent to assign the joblist
information for, e.g. sample (02).
 The test/assay information is now copied to the sample (02). Press
again the Clone mode button to deactivate.

Figure 75 Clone Mode

The Clone mode will only copy the test/assay information to another
NOTE
sample. Due to safety reasons sample ID and vial type information
won’t be copied and must be entered manually by sample.

Tools The Tools button in the Load Sample Menu offers several options and
access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 76 Tools in Load Samples

CoaLyser – Operators Manual 1.3 Page 75


OPERATION

The following Tools are available:

Available Tools Description


Opens the timeline window showing each step
Timeline of the joblist entered in a diagram and provides
run time information.
By pressing this button the current screen will
Hardcopy be printed out, provided a printer is connected
to the analyser.
Get request By pressing this button the analyzer verifies
(Only available with Bi- whether there is a sample request provided by a
Directional HOST software) connected HOST.

By pressing this button you will leave this


Quit
window without any changes.

Pressing this button will erase all sample


Erase Jobl.
information (Joblist) of the selected sample.

Show Jobl. Opens and shows the joblist currently entered.

OK
 Pressing the OK button in the Load Sample Menu confirms all changes
in this menu. The following display appears:

Figure 77 Run Preparation Menu

 In the Run Preparation Menu press OK again to return back to the MAIN
MENU. You are now prepared to start the measurements.

CoaLyser – Operators Manual 1.3 Page 76


OPERATION

5.2.4 Load Controls (QC-


Control)
This chapter describes how to load and measure control plasmas inside a
quality control period. To proceed with this section requires previous
definition and setup of control plasma, see chapter 5.3 Quality Control (QC
Program).

The following information is important prior processing a quality control run:

 Up to 8 controls can be defined and placed into positions C1 to C8, see


chapter 3.1.2 Working Plate, Reagent Area.
 The names and positions of the controls can not be changed in this
menu, see chapter 5.3.1 Control Setup.
 Up to 3 controls out of max. 8 can be assigned to a test at a time, see
chapter 5.3.1.1 Define additional QC-Control Parameter.
 Control Plasmas already assigned to a test and part of an active QC-
Period cannot be modified before QC-Period is finished and printed, see
chapter
 Once a control is assigned to a test, the control period automatically
starts.

To load controls proceed as follows:

 From the Main Menu, press the button Run Preparation to access the
Run Preparation Menu.
and/or
 From the Run Preparation Menu, press the button Load Controls the
following display appears.

Figure 78 Load Controls

Within this menu the complete control plasma handling can be controlled
and set. It further provides information on currently loaded controls, controls
to load, selectable control cups/tubes and reagents available for each
control.

 Select a control by pressing onto the control button once, e.g. (Control
1). The control box will be highlighted.

 The number shown on the left side of the selected control indicates the
predefined position (C3) where the vial/tube must be placed in the
reagent area.

CoaLyser – Operators Manual 1.3 Page 77


OPERATION

Figure 79 Control Positions

W1-W2 2x Washer & Cleaner Positions.

Reagent Block with 15 positions.


- (R1-R3) for 15ml reagent vials (left and right position stirred)
R1-R15
- (R4-R15) for 5ml reagent vials. The reagent block can be
removed from the working plate for external cooling.

C1-C8 8x Control & Calibrator Positions.

To place control and calibrator vessels, the Analyser uses the


controls/calibrator positions (C1-C8).

Select Control Vial


A selection of two secondary tubes/vials may be assigned to each control.

 Select a vial/tube from the list as shown in the vial frame of the display,
see next figure.

Figure 80 Vial Frame

Below table shows the vials/tubes placeable in the control area. Based on
the vial/tube used select the corresponding colour in the vial frame of the
display.

CoaLyser – Operators Manual 1.3 Page 78


OPERATION

The Control Positions accommodate the following vials/tubes:

Type Volume Size Tube

Sample-Cup 4ml 61mm x 12.1mm diameter Secondary

Hitachi 2ml Secondary

Make sure to place the same sized (coloured) vial/tube into the
NOTE
control area as selected in the vial frame. Only one type can be
selected at a time. Otherwise you risk failure in your analysis or
damage of the pipettor.

Renew
The accuracy and precision of control plasma depends on its stability time.
To define the stability time, see chapter 5.3.1 Control Setup.

Press the renew button to re-start the defined stability time but ensure the
control used is still inside this stability.

Assign test(s) to a control


To assign tests (assays) to a control you may select out of the tests/assays
as listed in the Load Controls Menu. Within this list all tests/assays will be
shown as previously loaded, see chapter 5.2.2 Load Reagents.

 Select the test/assay from the list as shown on the right side of the
display, see figure below example PT by pressing onto the test/assay
buttons.

Figure 81 Select test/assay

 Selected tests/assays will be highlighted. Repeat this procedure for all


controls you intent to assign tests/assays for.

Tools The Tools button in the Load Control Menu offers several options and
access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

CoaLyser – Operators Manual 1.3 Page 79


OPERATION

Figure 82 Tools in Load Samples

The following Tools are available:

Available Tools Description


Opens the timeline window showing each step
Timeline of the joblist entered in a diagram and provides
run time information.
By pressing this button the current screen will
Hardcopy be printed out, provided a printer is connected
to the analyser.

By pressing this button you will leave this


Quit
window without any changes.

Pressing this button will erase all control


Erase Jobl.
information (Joblist) of the selected control.

Show Jobl. Opens and shows the joblist currently entered.

OK
 Pressing the OK button in the Load Control Menu confirms all changes
in this menu. The following display appears:

Figure 83 Run Preparation Menu

CoaLyser – Operators Manual 1.3 Page 80


OPERATION

In the Run Preparation Menu press OK again to return back to the MAIN
MENU. You are now prepared to start the control measurements.

5.2.5 Start Measurements


(Routine, STAT & Controls)
 From the Main Menu, press the button Measurement to access the
Measurement Menu.

Figure 84 Menu Measurement

The following window provides information related to the current available


joblist.

The following information is available:

Available information Description

Provides information on the loaded reagents


Reagent Status List and washers, its required volumes, loaded
volumes, stability and lot numbers.

Provides information on the cuvette ring loaded:


- Start Cuvette No.
Cuvettes
- Required cuvettes for the run
- Free cuvettes available on the loaded ring
Provides information on the samples/controls
loaded:
- Number of loaded samples/controls
Samples & Controls
- Number of samples/controls to be processed
- Number of samples/controls partially
processed

Tools The Tools button in the Measurement Menu offers several options and
access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

CoaLyser – Operators Manual 1.3 Page 81


OPERATION

Figure 85 Tools in Measurement Menu

The following Tools are available:

Available Tools Description

By pressing this button the screen will be printed,


Hardcopy
provided a printer is connected to the analyser.
By pressing this button you will leave this window
Quit
without any changes.

Joblist Opens the joblist currently entered.

Opens the timeline window showing each step of


Timeline the joblist entered in a diagram and provides run
time information.

Exit Select Exit to return to the MAIN MENU.

Start the Run Select Start the Run to begin measurement.

The analyser starts preparing the measurement and starts the run if there is
no pending problem. Results and/or related run information are shown in this
window once available.

Figure 86 Running Measurement

CoaLyser – Operators Manual 1.3 Page 82


OPERATION

5.2.5.1 STAT Program

The STAT button accesses the Emergency (Load STAT) Menu during a
routine run. This menu command can be used to interrupt the current
measurement in order to load up to 3 emergency samples.

After interrupting the routine run by pressing the STAT button the program
waits until all current measurements have been finished. All routine samples
which have not been worked off will remain in the routine joblist and will be
measured after the STAT tests have been finished.

If you intend to initiate a STAT run by interrupting the routine run now follow
the instructions as provided in the next appearing screens until below
display appears:

Figure 87 Prepare STAT run

The following message appears. If you intend to interrupt the routine run
confirm this message by pressing the Yes button. To continue the routine
runs press the No button.

Press the Yes button to initiate a STAT run the following window appears:

Figure 88 Load STAT Samples

CoaLyser – Operators Manual 1.3 Page 83


OPERATION

All current running and prepared routine measurements will be finished first.
The window indicates the time until the STAT measurement can be
performed. Once the time counter has ran down the following message
appears:

Figure 89 Warning Message before STAT

Confirm this message and the following by pressing the OK button until next
window appears:

Figure 90 Load STAT Samples

To load STAT samples and start a STAT measurement continue the same
way as described in chapter 5.2.2 Load Reagents 5.2.3 Load Samples
and 5.2.5 Start Measurements.

CoaLyser – Operators Manual 1.3 Page 84


OPERATION

5.2.5.2 Continuous Loading


The Load button allows additional samples to be continuously loaded during
a routine run. This menu command can be used to interrupt the current
measurement in order to scan and load additional samples as soon as
current measurements are finished.

After interrupting the routine run by pressing the Load button the program
waits until the current measurements have been finished. All routine
samples which have not been worked off will remain in the routine joblist and
will be measured once the additional samples have been loaded and the
measurement has been re-started.

If you intend to additionally load samples by interrupting the routine run


follow the instructions as provided in the next appearing screens. After
pressing the Load button the following display appears:

Figure 91 Prepare continuous load

If you intend to interrupt the routine run confirm this message by pressing
the Yes button. To continue the routine runs press the No button.

Press the Yes button to proceed loading additional samples. The following
window appears:

Figure 92 Load additional samples

CoaLyser – Operators Manual 1.3 Page 85


OPERATION

All current running and prepared routine measurements will be finished first.
The window indicates the time until the continuous loading can be
performed. Once the counter has ran down the following message appears:

Figure 93 Result list of finfished samples

This window displays the current available results from the previously
finished rountine samples. The following can be done:
resend
 Press the resend button to send results to a HOST computer. Only
possible if an active connection to a HOST PC is available.
Print
 Press the Print button to print the short form result list.
OK
 Press the OK button to leave the results list to access the load sample
menu.

Confirm this message by pressing the OK button to access the load sample
menu. The next window appears:

Figure 94 Load additional samples

The load samples Menu appears. To load further load samples proceed as
described in the following.

Additional samples can now be loaded either by selecting existing samples


or by introducing additional samples. The samples can be added in different
ways:

CoaLyser – Operators Manual 1.3 Page 86


OPERATION

 To manually add samples select a sample position by pressing onto


the sample button once. The sample box will be highlighted.

 By double-clicking onto the sample button an alphanumeric keyboard


opens allowing the entry of sample/patient names/numbers.

 Use the keyboard to enter the sample/patient ID required, see chapter


5.2.3 Load Samples (Routine & STAT) for further information.

Instead of entering sample ID´s manually you may also load sample ID´s by
using either an external barcode scanner, see first chapter 4.3.5 Connect
an external Barcode Scanner (option) or an internal barcode scanner
(availability depends on analyzer type).

 Select a sample by pressing onto the sample button once. The sample
box will be highlighted. Do NOT enter the sample ID manually.

 Using an external barcode scanner direct the scanner window to the


barcode label of the vial/tube.

 Press the button on the barcode scanner. Continue pressing the button
until a beep confirms entry of ID.

 The scanned sample ID should now appear in the selected sample box.

 Using the internal barcode scanner hold the sample vial/tube on the
upper end with your finger tips and guide the vial/tube into the scanning
position located the working plate, see picture below. Keep the barcode
in direction of the scanner light beam.

 The barcode will automatically detect and scan the sample ID.

 The scanned sample ID should now appear in the selected sample box.

 Place the sample vial/tube into the sample rack using the same position
as shown in the sample select box.

Figure 95 Using internal barcode scanner

 Repeat described steps to enter more samples.

CoaLyser – Operators Manual 1.3 Page 87


OPERATION

 Once all samples are loaded pressing the OK botton in the Load
Samples Menu will guide you back to the Main Menu.

 From the Main Menu, press the button Measurement to access the
Measurement Menu.

Figure 96 Menu Measurement

Re-Start the Run Select Start the Run to continue the measurements. The analyser starts
preparing the measurement and re-starts the run if there is no pending
problem.

5.2.5.3 Status
The Status button provides access to an information window, see figure
below. For further information on the Status button see chapter 5.2.8
Analyser Status.

Figure 97 Status Window

CoaLyser – Operators Manual 1.3 Page 88


OPERATION

5.2.5.4 STOP
Press this button to immediately interrupt a current run, in case there are
any pending problems or other reasons. The current run will be stopped by
keeping all pending measurements in the joblist.

 Press the STOP button and follow the directions as provided on the
display.

Figure 98 Immediate STOP

Figure 99 Warning user interrupt

After confirming the Warning Message, see figure above the following
window appears providing information of the still pending measurements.
Confirm this message by pressing the OK button, the following window
appears.

CoaLyser – Operators Manual 1.3 Page 89


OPERATION

Figure 100 Pending Measurements

 Return to the Run Preparation Menu to check/modify the joblist and


repeat the measuring procedure as described. When a run is finished or
interrupted the Results of last run list appears:

Figure 101 Short Form Results

The following procedure is performed after a run is finished.

resend
 Press the resend button to send results to a HOST computer. Only
possible if an active connection to a HOST PC is available.
Print
 Press the Print button to print the short form result list.
OK
 Press the OK button to leave this window.

Tools The Tools button in the Running Measurement screen offers several
options and access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

CoaLyser – Operators Manual 1.3 Page 90


OPERATION

Figure 102 Tools in Running Measurement Screen

The following Tools are available:

Available Tools Description

By pressing this button you enter into the result


Results
list.
By pressing this button you will interrupt the run
Interrupt after the current measurements have been
finished.
By pressing this button you will print the screen,
Print
provided a printer is connected to the analyser.

By pressing this button you will leave this


Quit
window without any changes.

By pressing this button you will see the reaction


R-Curve
curve graph of the current run in real time.

Opens the timeline window showing each step


Timeline
of the joblist entered in a diagram.

5.2.6 Show/print results


 From the Main Menu, press the Results/Run button to access the short
form result list.

Figure 103 Results/Run

CoaLyser – Operators Manual 1.3 Page 91


OPERATION

This window displays all results of the last run sorted by sample. In case of
any measuring problem the results involved will be highlighted in orange
colour indicating the failure occurred.

The following failure codes are available:

Error Flags Description

State: OK The measurement has been finished with no problems.

No clot detection:
State: none No clot has been found within the set measuring time
and valid clotting range.
Deviation Error:
State: D The specified mean value deviation set in the test setup
is exceeded.
Temperature Error:
State: T
Temperature problems occurred at pipettor or rotor
Mixer Error:
State: M
The mixer motor(s) do not properly work.
Reagent Mixer Error:
State: S
The reagent mixer motor does not properly work.
Photometer Error:
State: F A photometer error was detected during a
measurement.
Liquid Error:
State: L
Not enough reagent or sample available during the run.
QC Error:
A test under Quality Control must be measured within
State: Q the last 24 hours using the assigned control plasma
(requires QC- before a routine measurement can be done. Additionally
Software installed) this test may not be in violation to the Westgard-Rules.
The Q Error will appear in case these requirements are
not fulfilled.

resend To resend the short form results to a HOST computer press the resend
button. Make sure a HOST computer is connected to the analyser, prior
starting to send results, see chapter 4.3.7 Connect a HOST Computer
(LAN interface).

Print To print the short form result list press the Print button. Make sure a printer
is connected to the analyser, prior starting a print-out, see chapter 4.3.4
Connect an external printer.

OK Press the OK button to return to MAIN MENU.

CoaLyser – Operators Manual 1.3 Page 92


OPERATION

5.2.7 Results Memory


 From the Main Menu, press the button Results/Memory to access the
result memory of the analyser.

Figure 104 Results/Memory 1

 From the Main Menu, press the button Results/Memory to access the
result memory of the analyser.

The internal analyzer memory keeps a maximum of 100 results in memory.


Once the memory contains 100 results the next following results measured
will delete automatically the oldest result in the ring storage.

The following result selection is available:

Result Selection Description

Displays all results available in chronologic


All
order.

Sample Displays the results sorted by sample/patient.

Test Displays the results sorted by test/assay.

CoaLyser – Operators Manual 1.3 Page 93


OPERATION

5.2.7.1 Show/print reaction curves


In the ring result storage you may show and print the reaction curves of the
results available. Up to six reaction curves may be displayed in the same
diagram.

Figure 105 Results/Memory 2

 From the Results/Memory window select an input field visible at the


bottom of this window (only available if USB Stick is attached to system).

Figure 106 Curve Input

 Press into an input field to open the keyboard entry. The following
window appears.

Figure 107 Curve Input keyboard

 Enter the Result No. as shown on the left side of the results ring storage
list, example 23 using the numerical keypad. Confirm by pressing the
Enter button.

 Follow the same procedure to select more reaction curves to display.

 When all result nos. are entered press the Tools button in the result ring
storage window.

CoaLyser – Operators Manual 1.3 Page 94


OPERATION

Tools The Tools button in the ring result ring storage offers several options and
access to other menus and features.

Figure 108 Tools in Results/Memory

The following Tools are available:

Available Tools Description


By pressing this button selected results will be
Print printed including all result data and curve
information.

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

Opens the curve diagram window to show the


Graph
reaction curves selected.

 Press the Graph button in the Tools window to display the selected
reaction curves.

Figure 109 Reaction Curves

CoaLyser – Operators Manual 1.3 Page 95


OPERATION

Tools The Tools button in the reaction curve diagram window offers additional
options and access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 110 Tools in Reaction Curve Diagram

The following Tools are available:

Available Tools Description


Increases the scale of the displayed curve. By
Zoom IN pressing into the curve screen and holding you
may move the curve into all directions.

Reduces the scale of the displayed curve. Only


Zoom OUT
available after using Zoom IN first

Resets the scale of the curve diagram to the


Zoom RESET
initial setting.

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to leave this menu.

CoaLyser – Operators Manual 1.3 Page 96


OPERATION

5.2.8 Analyser Status


At any time the analyser offers information on the current status of the
system. Depending on the current state of the analyzer the Status button
may turn into different colours in case any problem occurs.

 From the Main Menu, press the button Status to access the Status
Menu.

Figure 111 Status Menu

The Status Menu provides information on the current state of container level
sensors, occurred measuring failures, still available cuvettes of the cuvette
ring, cuvette ring credits available, temperature of the measuring rotor and
pipettor, the state of motors as well as information on any QC related
information, memory allocation and displays any irregularities and errors
which may appear during a run.

The Status button displays the following states of the analyzer:

A green Status button means that no particular problems have been


discovered which could in any way cause problems to the analyzer
hardware, measurements, QC Status and memory space available.
However, problems can still occur, e.g. requests of tests with invalid
calibration.

An orange Status button means that there is a general problem or other


information available, e.g. reminder to run the daily quality control or a
limited number of cuvette ring balance available. In general measuring
series are not affected.

A red Status button means that measuring series cannot start or have been
stopped due to an error.

The Status window can be moved in all directing by touching onto the
screen with your finger tip.

 Additionally this status information can be printed by pressing the Print


button.

 Leave this menu by pressing the OK button.

CoaLyser – Operators Manual 1.3 Page 97


OPERATION

5.3 Quality Control (QC Program)


(Only available if QC software is installed on analyzer)

Laboratory quality control (QC) is designed to detect, reduce, and correct


deficiencies in a laboratory's internal analytical process prior to the release
of patient results and improve the quality of the results reported by the
laboratory. Quality control is a measure of precision or how well the
measurement system reproduces the same result over time and under
varying operating conditions. Laboratory quality control material is usually
run at the beginning of each shift; after an instrument is serviced; when
reagent lots are changed; after calibration, and when patient results seem
inappropriate.

For such control measurements either specific control plasmas or patient


plasmas will be used. Interpretation of quality control data involves both
graphical and statistical methods. Quality control data is visualized using a
Levey-Jennings Chart. The dates of analyses are plotted along the X-axis
and control values are plotted on the Y-axis. The distance from the mean is
measured in one, two, and three standard deviations (SD). The mean and
also the standard deviations are marked on the Y-axis. Inspecting the
pattern of plotted points provides a simple way to detect increased random
error and shifts or trends in calibration.

The formulation of Westgard Rules is used to analyse data in the statistical


control charts. These rules can be applied to see whether the results from
the samples when the control was done can be released, or if they need to
be rerun. Westgard Rules are used to define specific performance limits for
a particular assay and can be used to detect both random and systematic
errors. Westgard Rules are programmed in to analyzers software to
determine when an analytical run should be rejected. These rules need to be
applied carefully so that true errors are detected while false rejections are
minimized.

The QC program can only be fully accessed by authorized users in the


expert level, see chapter 3.2 Software. Without access to this level only
limited areas of the QC program can be used.

To set up controls and a QC-Period for a specific test the following must be
considered:

 Up to 8 controls can be defined


 Up to 3 different controls can be assigned to a test at a time
 Each control assigned to a test is analyzed separately for the control
period
 A QC-Period is limited to a calendar month
 Within a QC-Period days without results are allowed
 The day change time is fixed at 12:00pm (midnight)
 Controls are loaded to the 8 control positions C1 to C8, see chapter
3.1.2 Working Plate, Reagent Area.
 A control measurement will be handled on a higher priority in the run
than a stat measurement
 A control result is treated like a sample within a run but control
measurements are always measured at first in a run.
 Control measurements run in single determination only.
 If a test is set to Auto-repeat the control measurement uses the
measuring time for Auto-repeat, see chapter 5.4.4.2 Evaluation, Auto-
repeat but the control measurement won’t be repeated automatically.

CoaLyser – Operators Manual 1.3 Page 98


OPERATION

 If a control measurement for a test is run a second time a day the


previous measured QC result will be overruled by the newest control
result available.
 If converted results are set to calculate control results the test used for
this control measurement must have a valid calibration, see chapter
5.4.5 Calibration.

A calibration is not possible if reagents are not set to calibrate, see


NOTE
section Reagent Setup/Page 1 / Calibrate (check box) in chapter
5.4.3.1 Parameter Page 1.

 From the Main Menu, press the button QC to access the Quality Control
Program. The following window appears:

Figure 112 QC Main Page

The QC Main Page allows you to define whether the QC Program shall be
activated, how to setup controls, view running QC-Periods, statistics and
graphs and daily results measured as well as to define specific evaluation
rules according to Westguard regulations and to print/finish and delete QC-
Periods.

Quality Control on
By using this check box button you may define or modify whether the QC
Program will be set to on or off. Set individual parameters On/Off by pressing
onto the field, e.g. ON (QC Control on) = .

During an active running QC period the QC Program cannot be set


NOTE
to off. Finish and print the running QC-Period first before
deactivating the the QC Program.

CoaLyser – Operators Manual 1.3 Page 99


OPERATION

Westguard Rules
The check box buttons in the Westgard Rules frame allow you to define and
select the available Westgard Rules regulations.

By using this check box button you may define or modify whether certain
rules will be set to on or off. Set individual parameters On/Off by pressing
onto the field, e.g. ON (1_3S on) = .

The following Westgard Rules are available:

Rules Description
Refers to a control rule that is commonly used with a Levey-
Jennings Chart when the control limits are set as the mean
13s plus 3s and the mean minus 3s. A run is rejected when a
single control measurement exceeds the mean plus 3s or the
mean minus 3s control limit.
A run is rejected when 2 consecutive control measurements
22s exceed the same mean plus 2s or the same mean minus 2s
control limit.
A run is rejected when 1 control measurement in a group
R4s exceeds the mean plus 2s and another exceeds the mean
minus 2s.
A run is rejected when 4 consecutive control measurements
41s exceed the same mean plus 1s or the same mean minus 1s
control limit.

A run is rejected when 8 consecutive control measurements


8x fall on one side of the mean

5.3.1 Control Setup


From the QC Main Page Screen press the Control Setup button to define
or modify the control plasma.

The following display appears:

Figure 113 Select Control Position

CoaLyser – Operators Manual 1.3 Page 100


OPERATION

Until no Controls are defined, the position buttons remain empty. By


pressing onto the blue field you define the position to place the Control
Plasma on the analyzer. There are up to 8 positions on board available.

Position C1&C2 can also be used for Control Plasma. Please


NOTE
consider that these positions are also used for diluents/calibrators
required for calibration. The use of positions C1 and C2 for controls
and calibrators requires the cups to be replaced manually. To avoid
errors it is recommended to use positions C3-C8 for control plasmas
only.

To define or modify a Control Plasma select the position and press the
suitable button, e.g. Pos 3.

The following display appears:

Figure 114 Define Control Plasma

Name
By pressing into the Name input field a keypad opens allowing the entry or
modification of the name of a control plasma. The control plasma name may
have up to 15 alphanumerical characters. E.g. Control1

Abbrev.
By pressing into the Abbrev. (abbreviation) input field a keypad opens
allowing the entry or modification of the abbreviation of a control plasma
name. The control plasma abbreviation may have up to 5 alphanumerical
characters. This abbreviation is used in other menu screens for identification
of the control plasma. E.g. Con1

Lot-No. By pressing into the Lot-No. input field a keypad opens allowing the entry
or modification of the lot number of a control plasma. The lot number may
have up to 15 alphanumerical characters, e.g. ABC1234

A new lot number can only be used if a control plasma is not


NOTE
assigned to a running QC-Period.

Stability By pressing into the Stability input field a keypad opens allowing the entry
or modification of the stability time of a control plasma. The stability time
may have up to 2 numerical characters. Input range: 1 - 99 (h).

CoaLyser – Operators Manual 1.3 Page 101


OPERATION

Figure 115 Edit Control Plasma

Delete
By pressing the Delete button in the Edit Control Plasma Page will delete
the Control Plasma setup.

A Control Plasma can not be deleted if assigned to a running QC-


NOTE
Period. The QC-Period must be finished and printed first.

The following display appears:

Figure 116 Delete Control Plasma

Pressing the Yes button will delete the Control Plasma setup. No will keep
the entry as previously done.

Tools
The Tools button in the Edit Control Plasma Screen offer several options
and access to other menus and actions.

 Press the Tools button to access further additional areas the following
Select Window appears.

CoaLyser – Operators Manual 1.3 Page 102


OPERATION

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

ESC&Exit By pressing the ESC&Exit button you will leave the Edit Control Plasma
Screen without saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Edit Control Plasma Screen confirms
all changes and modifications done.

Proceed the same way as described in this chapter if you intent to setup or
modify additional Control Plasmas.

After pressing the Save&Exit button you will return back to the QC Main
Page.

Figure 117 QC Main Page

OK Press the OK button to return back to the Main Menu.

Figure 118 Main Menu

CoaLyser – Operators Manual 1.3 Page 103


OPERATION

5.3.1.1 Define additional QC-Control Parameter


(only available if QC software is installed on analyzer)

This section describes how assign previously defined Control Plasma to a


test and how to enter additional QC-Control Parameter required for result
calculation and statistical/graphical evaluation.

 From the Main Menu Screen press the Setup button.

The following display appears:

Figure 119 Setup Menu

 In the Setup Menu press the Test Setup button.

The following display appears:

Figure 120 Select Test Screen

The Select Test Screen contains a list of all tests saved.

 To select a test for quality control definition use the arrow keys to first
select the test and then press the OK button to continue.

CoaLyser – Operators Manual 1.3 Page 104


OPERATION

The following example provides description on how to define a PT test.


Proceed the same way if you intent to assign other test for quality control
definition.

 In the Select Test Screen chose PT and confirm by pressing the OK


button.

Figure 121 Test Menu Main Page

 To assign control plasmas and or modify or add parameters and result


calculations press the QC button.

The following display appears:

Figure 122 Test Menu Select Control

Up to three different Control Plasmas can be assigned individually to a test.


By pressing onto the individual buttons you enter into the QC Control
parameter page. Repeat the procedure for each Control Plasma to be added
to the test as described as follows.

CoaLyser – Operators Manual 1.3 Page 105


OPERATION

Figure 123 Test Menu QC

Select result type By pressing the Select Box arrow buttons you can select between 6 different
type of results, meaning whether the QC result calculation will be based on
raw values or converted values.

Converted values can only be selected and memorized if the test is


NOTE subject to a calibration and has a valid calibration, see chapter 5.4.5
Calibration. Otherwise only result types are selectable which are
applicable to the test selected.

The following result types are available:

Available result types Description


Sets the result calculation to clotting, raw values
Clotting
will be calculated in [sec].
Sets the result calculation to clotting conversion,
Clotting 1st conversion
1st conversion values will be calculated in [%].
Sets the result calculation to clotting conversion,
Clotting 2nd Conversion
2nd conversion values will be calculated in [INR].
Sets the evaluation algorithm to Kinetic,
Difference or 1st Derivation (depending on test
Kinetic/Difference/1st
setup), raw values will be calculated in [mE/s]
derivation raw
for Kinetic, [mE] for Difference, [Delta mE/sec]
for 1st derivation.
Sets the evaluation algorithm to Kinetic,
Kinetic/Difference/1st Difference or 1st Derivation (depending on test
derivation conversion setup), converted values will be calculated in
the unit defined in calibration setup.
Sets the evaluation algorithm to the maximum
Max. Time time of the measuring signal in 1st Derivation
mode. Values will be calculated in [sec]
Off No result calculation

Control Plasma By pressing onto the blue field a window opens allowing you to select
Control Plasma from a list of previously defined Controls, see chapter 5.3.1
Control Setup

The following window appears.

CoaLyser – Operators Manual 1.3 Page 106


OPERATION

Figure 124 Test Menu Select Control List

The Select Control Plasma Screen contains a list of all previous Controls
memorized. A total of 8 Controls can be defined and memorized.

 To select a Control use the arrow keys to first select the Control Plasma
and then presses the OK button to continue.

The selected Control Plasma appears in the following window:

Figure 125 Test Menu selected Control

Target value By pressing into the Target value input field a numerical keypad opens
allowing the entry or modification of the target value expected to be
achieved from the control plasma and test measurement. Depending on the
test and result type selected the target value must be entered either as raw
value, e.g. in seconds or as converted value, e.g. % (PT).

1S-deviation By pressing into the 1S-deviation input field a numerical keypad opens
allowing the entry or modification of the 1 sigma deviation based on the
target value entry. Input Range: ≥5% and ≤ 20% of the target value in
seconds.

Example:
Target Value = 12 sec.
1S-Deviation = min. 0.6 sec. / max. 2.3 sec.

Based on the 1S-deviation entry the ±1S, ±2S and ±3S thresholds are
calculated.

CoaLyser – Operators Manual 1.3 Page 107


OPERATION

1S-deviation definition:

 The single standard deviation corresponds to the ± 1S Range


 The double standard deviation corresponds to the ± 2S Range
 Three times the standard deviation corresponds to the ± 3S Range.

Delete By pressing the Delete button in the Test Menu QC Screen your will delete
the parameter setting and remove the Control Plasma from the test setup
specifications.

Control Plasma cannot be deleted if Control Plasma is used in an


NOTE
active QC-Period. Active QC-Periods must be finished and printed
first.

Tools The Tools button in the Test Menu QC Screen offers additional options and
access to other menus and features.

The following Tools are available:

Available Tools Description


By pressing this button you will leave this
Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Main Screen.

Figure 126 Test Menu Main Screen

Tools
The Tools button in the Test Menu Main Screen offer several options and
access to other menus and actions.

 Press the Tools button to access further additional areas the following
Select Window appears.

CoaLyser – Operators Manual 1.3 Page 108


OPERATION

Figure 127 Tools in Test Menu Main Page

The following Tools are available:

Available Tools Description


By pressing this button the selected test will be
Delete fully deleted from the test list. Reagents assigned
to certain tests require the test to be deleted first.
By pressing this button the selected test may be
Save as copied and saved with the same parameters as
another test in the test list.
By pressing this button you will leave this window
Quit
without any changes.
By pressing this button the test parameters will be
Print
printed via the external connected printer.
By pressing this button the Test Menu Main
Hardcopy screen will be printed, provided a printer is
connected to the analyser.

ESC&Exit By pressing the ESC&Exit button you will leave the Test Menu without
saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Test Menu confirms all changes and
modifications done in this menu. Prior of acceptance the software will check
conformity on all edited parameters. In case of any invalid inputs the
software will indicate the corrections required.

All necessary information to start a quality control period has now been
entered into the system. To load controls and run a QC measurement,
proceed as described in chapter, load controls.

To view daily results, statistical information and show graphical information


(Levey Jennings Chart) or get further information on how to print, finish and
delete QC-Period proceed with chapter

CoaLyser – Operators Manual 1.3 Page 109


OPERATION

5.3.2 QC-Periods
The following section describes how to view all available quality control data
such as graphical and statistical of a running QC-Period.

 From the Main Menu Screen press the QC button to enter the QC Main
Page.

The following display appears:

Figure 128 QC Main Page

 From the QC Main Page press the QC-Periods button.

The following display appears:

Figure 129 Select active QC-Period List

The following overview shows all currently active QC-Periods of the actual
year and month and related Test and assigned control.

 To select a specific QC-Period use the arrow keys to move up and down
the screen. The selected QC-Period will be highlighted. Press the OK
button to view further information.

CoaLyser – Operators Manual 1.3 Page 110


OPERATION

The following display appears:

Figure 130 QC Result Table

From the QC-Result Table you can select out of a variation of available
information related to the selected QC-Period.

The following is available:

Available Information Description


By pressing this button the selected QC-Period
will be deleted. However, a QC-Period must
Delete
always be finished and printed first prior deletion
otherwise this action will not be allowed.
Shows the statistical calculations, results and
Table predefined setup information for this specific QC
test and control combination.
Shows the graphical information (Levey-
Chart Jennings Chart) for this specific QC test and
control combination.

By pressing this button you will leave this


Quit
window without any changes.

Pressing this button will automatically finish and


Finish
print a selected QC-Period.

Pressing this button will print the current


Print available results and charts of a selected QC-
Period.

5.3.2.1 View a QC-Period (statistics)

 By pressing the Table button you will enter into the statistical overview
of the selected QC-Period.

The following overview shows all currently available information related to


the selected QC-Period.

CoaLyser – Operators Manual 1.3 Page 111


OPERATION

Figure 131 QC-Period overview 1

The statistical QC-Results and Setting list provides information on the


current settings of a selected QC-Period, the test and control assigned to
this QC-Period as well as status information and currently done
measurements, standard deviations and mean values. It further provides
information on the Status of a QC-Period, meaning whether it has been
printed or finished already.

The window can be moved by touching the finger tip onto the screen and
moving it to all sides or additionally by using the arrows (up/down, left/right)
available.

Print By pressing the Print button will initiate a print-out of this list.

OK By pressing the OK button you will leave this window and enter into an
additional table providing further information on each result measured during
this QC-Period and Westgard-Rules violations according to the selected
rules for this QC-period.

Figure 132 QC-Period overview 2

Print By pressing the Print button will initiate a print-out of this list.

OK By pressing the OK button you will return to the QC-Result Table.

CoaLyser – Operators Manual 1.3 Page 112


OPERATION

5.3.2.2 View a QC-Period (chart)

 By pressing the Chart button you will enter into the chart overview of the
selected QC-Period.

This menu command calls up the graphic display of the results of the
selected QC-Period in a Levey-Jennings Chart.

Figure 133 Levey-Jennings Chart

The selected QC-period is displayed as a Levey-Jennings Chart.

X-Axis:
The X-Axis comprises a maximum of 31 days so that the complete month
can be displayed. The respective date is shown on the X-Axis in the format
Day of Month. In the upper line of the chart the respective year and month
are displayed including the test and control used.

Y-Axis:
The Y-Axis is scaled so that all results can be displayed. The result range is
either displayed as raw value in (sec.) or as converted value, e.g. mg/dl
depending on the Control Setup, see chapter 5.3.1 Control Setup. The
following labelled horizontal lines are displayed:

Red line + 3s / -3s:


Three times the standard deviation of the mean value (target value) as
entered in chapter 5.3.1.1 Define additional QC-Control Parameter.

Orange line + 2s / -2s:


Two times the standard deviation of the mean value (target value) as
entered in chapter 5.3.1.1 Define additional QC-Control Parameter.

Yellow line + 1s / -1s:


Upper and lower limit of the confidence range for accuracy controls, see
chapter 5.3.1.1 Define additional QC-Control Parameter, 1S-Deviation.

Green Line norm:


Nominal value as entered in Target Value, see chapter 5.3.1.1 Define
additional QC-Control Parameter.

/:
QC-Results are above or below the + 3s / -3s deviation.

CoaLyser – Operators Manual 1.3 Page 113


OPERATION

Tools The Tools button in the Levey-Jennings Chart window offers additional
options and access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 134 Tools in Edit Reagent Page 1

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Performs a print of the Levey-Jennings Chart


Hardcopy
screen only.

OK By pressing the OK button you will return to the QC-Result Table.

CoaLyser – Operators Manual 1.3 Page 114


OPERATION

5.3.2.3 Print a QC-Period


A QC-Period can be printed including all current available information
related to this QC-Period. If a QC-Period has been printed the period will be
assigned with the status printed in the statistic settings and result list, see
chapter 5.3.2.1 View a QC-Period (statistics).

Figure 135 Print QC-Period

 By pressing the Print button in the QC-Result Table you will start
printing the selected QC-Period.

Keep the print-out of all QC-Periods at a safe place as they are the
NOTE
final document of the quality control done. After printing and
finishing a QC-Period all QC-Results will be deleted from the system
and won’t be available for later use.

Once the QC-Period is printed you will return back to the QC-Result Table.

Figure 136 Tools in QC Main-Page

CoaLyser – Operators Manual 1.3 Page 115


OPERATION

5.3.2.4 Finish a QC-Period before


end of a month
A regular QC-Period is limited to a calendar month. QC-Periods regularly
end at the last day of the month and will be finished automatically the first
day of the following month (at 0:00) if no user interaction took place.
However, a selected QC-Period can be finished before end of a month if
requested.

 By pressing the Finish button you will start finishing a current running
QC-Period.

The following display appears:

Figure 137 Finishing QC-Period

 By pressing No you will leave this window and return back to the QC-
Result Table. Press Yes to finish the QC-Period selected. In case the
QC-Period has not been printed until date an automated print-out of the
complete QC-Period (statistics/charts) starts. Otherwise the following
display appears:

Figure 138 Info Box printed QC-Period

 Confirm this message by pressing the OK button.

CoaLyser – Operators Manual 1.3 Page 116


OPERATION

Figure 139 Delete QC-Period

Check again the print-out available for this particular QC-Period to ensure
adequate documentation of your QC-Period results.

 By pressing No the QC-Period will be continued. Press Yes to finally


confirm the QC-Period to be finished and deleted.

Figure 140 Deleted QC-Period Info

 Confirm this message by pressing the OK button.

CoaLyser – Operators Manual 1.3 Page 117


OPERATION

5.3.2.5 Delete a regularly ended


QC-Period
A QC-Period which regularly ends after the end of a month will be finished
automatically on the first day of the following month and kept in the result file
storage unless the decision has been made to delete finally from the
system. To delete such QC-Periods follow the instructions as follows:

Once a QC-Period is deleted all results and charts related to this QC-
NOTE Period are deleted from the system.

 To delete a regularly ended QC-Period press the QC button in the Main


Menu to enter the QC Main Page.

 In the QC Main Page press the QC-Periods button and select a QC-
Period out of the QC-Period list. Press OK to access the QC-Result
Table window.

 By pressing the Delete button in the QC-Result Table you will start the
deletion process for the selected QC-Period.

The following display appears if the selected QC-Period has neither been
printed nor finished yet:

Figure 141 Print/Finish Info

 Confirm message by pressing the OK button. Follow the description as


provided in chapter 5.3.2.3 Print a QC-Period, 5.3.2.4 Finish a QC-
Period before end of a month before proceeding with the deletion of
the selected QC-Period.

If the QC-Period ended regularly it will be automatically finished the first day
of the following month.

 To finally delete a regularly ended QC-Period press the QC button in the


Main Menu to enter the QC Main Page.

 In the QC Main Page press the QC-Periods button and select a QC-
Period out of the QC-Period list. Press OK to access the QC-Result
Table window.

CoaLyser – Operators Manual 1.3 Page 118


OPERATION

The following display appears:

Figure 142 QC-Result Functions

Above window shows the QC-Period selected has been finished


automatically by the system as the Finish button is not available for use.
However, before being able to delete the selected QC-Period it is necessary
to print first.

 Press the Print button to start printing the QC-Period. Make sure a
printer is up and connected to the analyzer and enough paper is
available for the print-out.

As soon as the print-out is done the QC-Period can be deleted from the
system.

Press the Delete button in the QC-Result Table and the following display
appears:

Figure 143 Delete QC-Period

Confirm above message by pressing either Yes to delete the selected QC-
Period or No if you intent not to delete the selected QC-Period. By pressing
Yes the following window appears.

CoaLyser – Operators Manual 1.3 Page 119


OPERATION

Figure 144 Deleted QC-Period confirmation

 Confirm this message by pressing the OK button.

Keep the print-out of all QC-Periods at a safe place as they are the
NOTE
final document of the quality control done. After printing and
finishing a QC-Period all QC-Results will be deleted from the system
and won’t be available for later use.

5.3.3 View daily results


This menu in the QC Main Page allows you to view the current QC results of
the day.

 From the Main Menu Screen press the QC button to enter the QC Main
Page.

 By pressing the Daily Results button the following display appears:

Figure 145 Daily results table

 From the Daily Result List press either the Print button to print the table
for your records or the OK button to return back to the QC Main Page
window.

CoaLyser – Operators Manual 1.3 Page 120


OPERATION

5.3.4 Print all QC-Periods at a time


This menu in the QC Main Page allows you to print all available QC-Periods
at a time.

 From the Main Menu Screen press the QC button to enter the QC Main
Page.

The following display appears:

Figure 146 QC Main Page

 From the QC Main Page press the Print button in the All QC-Periods
frame. Make sure a printer is up and connected to the analyzer and
enough paper is available for the print-out.

The following display appears:

Figure 147 Print all QC-Periods

Confirm above message by pressing either Yes to print all finished QC-
Periods or No if you intent not to print. By pressing Yes the following window
appears.

CoaLyser – Operators Manual 1.3 Page 121


OPERATION

Figure 148 Printing all QC-Periods

5.3.5 Delete all QC-Periods at a time


This menu in the QC Main Page allows you to delete all available already
printed and finished QC-Periods at a time.

 From the Main Menu Screen press the QC button to enter the QC Main
Page.

The following display appears:

Figure 149 QC Main Page

 From the QC Main Page press the Delete button in the All QC-Periods
frame. Make sure all QC-Periods are printed and finished first.

The following display appears:

CoaLyser – Operators Manual 1.3 Page 122


OPERATION

Figure 150 Delete all QC-Periods 1

Confirm above message by pressing either Yes to delete all finished and
printed QC-Periods or No if you intent not to delete. By pressing Yes the
following window appears.

Figure 151 Delete all QC-Periods 2

It is necessary to confirm deletion of all QC-Periods separately. Confirm by


pressing the Yes button, No will cancel this decision. The following display
appears:

Figure 152 QC Main Page

CoaLyser – Operators Manual 1.3 Page 123


OPERATION

Now all available QC-Periods have been deleted from the system. To finally
delete the QC file used press again the QC-Periods button in the QC Main
Page and confirm the following QC-Periods select window with OK.

The following message appears:

Figure 153 Delete QC file

Confirm above message by pressing either Yes to delete the QC-data file or
No if you intent not to delete.

All available QC-data has now been successfully removed from the system.

CoaLyser – Operators Manual 1.3 Page 124


SETUP

5.4 Setup Menu

The Setup Menu allows you to define all operational analyser, test and
reagent parameters either manually or loading the same using an USB-
Stick. Certain areas within the Setup Menu may only be accessed by
entering a PIN Code in the Main Menu. The software indicates the level of
password required accessing into another level.

Figure 154 Setup Menu

Menu Button Function Software Level

Passwords To set passwords to access different software level. Expert

Reagent Setup To define and edit reagent specific parameters Anyone

Date & Time To edit analyser date & time. Anyone

Im/Export To im/export test & reagent parameters via USB Stick. Anyone
Initiates an automatic self test procedure of the
Selftest Anyone
analyser.
To define and edit test specific parameters. To
Test Setup Anyone
combine individual reagents.
Perform calibration and management of calibration
Calibration Anyone
data, curves.
Delete Results To delete all results of the ring memory Expert
Selection of different available languages for user
Language Anyone
interface
Check Box to activate/deactivate automated print of
STAT Autoprint Anyone
STAT parameters
Select Box to select whether the working mode of a
Working Modes Anyone
measurement shall be done by sample or by test
Check Box to activate/deactivate acoustic signals
Beeper Anyone
performed by the anlyser.
Headline Defines print headline Anyone

Tools Access to other functions Anyone

OK Return to MAIN MENU Anyone

CoaLyser – Operators Manual 1.3 Page 125


SETUP

5.4.1 Im/Export test & reagent


parameter via USB-Stick
This chapter describes how to Import and Export test and reagent specific
parameters by using a USB-Stick.

5.4.1.1 Mount an USB-Stick


Prior using a USB-Stick to import or export test and reagent parameters it is
necessary to introduce (mount) the USB device to the analyser software
first. If not already attached to the system the following description explains
how to mount a USB-Stick.

 From the Main Menu, press the button Shutdown the following display
appears.

Figure 155 Stop System

 By pressing the button Switch Off the software allows you to switch off
the power switch located at the back side of the analyser. Switch off
analyser by setting power switch to position O = OFF.

 After switching off the analyser add the USB-Stick to the USB interface
located at the right side of the analyser within the carrying handle, see
Figure 39 USB-Stick Interface.

 Re-start the analyser again by setting the power switch to position I =


ON and follow the initialization sequence of the analyser. The USB-Stick
has now been mounted by the system software.

5.4.1.2 Import data from USB-Stick


Once an USB-Stick is mounted by the system software parameters can be
imported from an USB-Stick.

 From the Main Menu, press into the window underneath Password/PIN
to enter a PIN Code.

 A numerical keypad opens allowing entry of the Pin Code level. Enter
the PIN Code 12345 to allow access to the expert level of the software
and confirm with ENTER.

CoaLyser – Operators Manual 1.3 Page 126


SETUP

 From the Main Menu, press the button Setup to access the Setup
Menu.

 From the Setup Menu, press the button Im/Export Test&Reagent


Parameters via USB-Stick.

Figure 156 Im/Export Test & Reagent Parameters

The following Features are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

By pressing this button you will start the export


Export sequence to load data from the analyser to the
USB-Stick.
By pressing this button you will start the import
Import sequence to load data from the USB-Stick to the
analyser.

 Press the button Import to load test and reagent parameters from the
USB-Stick to the analyser.

Figure 157 Import Test & Reagent Parameters

CoaLyser – Operators Manual 1.3 Page 127


SETUP

By importing test and reagent parameters to the analyser all current


NOTE available parameters on the analyser will be overwritten. Export
such data first to the USB-Stick in case you want to safe the current
system parameters, see chapter 5.4.1.3 Export data to USB-Stick.

 Press Yes to start the import sequence. By pressing No you will leave
this window. While importing the parameters the following window
appears.

Figure 158 Importing Parameters

Figure 159 Parameter Import OK

 Once the import sequence is finished confirm by pressing the OK


button. The parameters have successfully being brought to the analyser.

 Back in the Setup Menu press the OK button to get back to the Main
Menu.

 Shutdown the system, refer to chapter 5.4.1.1 Mount an USB-Stick and


remove the USB-Stick carrying the test and reagent parameters. You
may now attach another USB-Stick used to record results data.

CoaLyser – Operators Manual 1.3 Page 128


SETUP

5.4.1.3 Export data to USB-Stick


To transfer test and reagent parameters to an USB-Stick you may proceed
the same way as described in the chapters before. Make sure an USB-Stick
is mounted by the system software, see chapter 5.4.1.1 Mount an USB-
Stick prior you start exporting parameters.

 From the Main Menu, press into the window underneath Password/PIN
to enter a PIN Code.

 A numerical keypad opens allowing entry of the Pin Code level. Enter
the PIN Code 12345 to allow access to the expert level of the software
and confirm with ENTER.

 From the Main Menu, press the button Setup to access the Setup
Menu.

 From the Setup Menu, press the button Im/Export Test&Reagent


Parameters via USB-Stick.

Figure 160 Im/Export Test & Reagent Parameters

The following Features are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

By pressing this button you will start the export


Export sequence to load data from the analyser to the
USB-Stick.
By pressing this button you will start the import
Import sequence to load data from the USB-Stick to the
analyser.

 Press the button Export to transfer test and reagent parameters from
the analyser to the USB-Stick.

CoaLyser – Operators Manual 1.3 Page 129


SETUP

Figure 161 Export Test & Reagent Parameters

Figure 162 Remove USB-Stick

 Once the export sequence is finished confirm above message by


pressing either Yes to remove the USB-Stick or No if you intent to leave
the USB-Stick mounted to the analyser. By pressing Yes the following
window appears.

Figure 163 Mounting another USB-Stick

CoaLyser – Operators Manual 1.3 Page 130


SETUP

 You may decide whether a new USB-Stick, for example to safe result
data shall be mounted or the current USB-Stick shall remain mounted.
Press Yes to mount another USB-Stick after you have replaced the USB
Sticks.

 By pressing the OK button you get back to the Main Menu.

5.4.2 Select Working Modes

In the Setup Menu you may further select whether the working mode shall
be done by sample or by test.

By pressing the corresponding radio button you select the desired working
mode.

By sample = All tests/reagents assigned to a sample/patient will be


measured patient oriented, sample by sample.

By test = All tests/reagents assigned to a sample/patient will be


measured test oriented, test by test.

 Press the radio button in the Working Modes frame to select the
desired measuring sequence.

Figure 164 Setup Working Modes

CoaLyser – Operators Manual 1.3 Page 131


SETUP

5.4.3 Setup Reagent definition


This chapter describes how define a new reagent or modify existing reagent
data.

 From the Main Menu, press the Setup button to access the Setup Menu.

 From the Setup Menu, press the Reagent Setup button to access the
Select Reagent Screen.

Figure 165 Select Reagent Screen

The Select Reagent Screen contains a list of all reagents saved, including
buffers and diluents.

The list contains up to 50 reagent definitions available. Free reagent


definition files can be programmed by the user.

1) To display or modify the data of a reagent use the arrow keys to first
select the reagent and then press the OK button to continue.

2) To define a new reagent use the arrow key and move upward until a free
position appears and also press the OK button to continue.

The following example provides description on how to program a PT


reagent. Proceed the same way if you intent to enter new reagent data or to
modify existing reagents.

 In the Select Reagent Screen chose PT and confirm by pressing the OK


button.

CoaLyser – Operators Manual 1.3 Page 132


SETUP

Figure 166 Edit Reagent Main Screen

Name
By pressing into the Name input field a keypad opens allowing the entry or
modification of the name of a reagent. The reagent name may have up to 15
alphanumerical characters.

Abbrev.
By pressing into the Abbrev. (abbreviation) input field a keypad opens
allowing the entry or modification of the abbreviation of a reagent name. The
reagent abbreviation may have up to 5 alphanumerical characters. This
abbreviation is used in other menu screens for identification of the reagent.

Parameters
The parameter frame allows access to other sections required to define or
modify a reagent.

5.4.3.1 Parameter Page 1


By pressing the Page 1 button the following window opens:

Figure 167 Edit Reagent Page 1

Within this screen the following general reagent data can be defined or
modified:

CoaLyser – Operators Manual 1.3 Page 133


SETUP

Lot No. By pressing into the Lot-No. input field a keypad opens allowing the entry
or modification of the lot number of a reagent. The lot number may have up
to 15 alphanumerical characters.

If you enter a new lot number and leave the keypad window, all tests
NOTE
using this reagent must be recalibrated.

Stability By pressing into the Stability input field a keypad opens allowing the entry
or modification of the stability time of a reagent. The stability time may have
up to 5 numerical characters. Input range: 1 - 99 (h).

Vial By using the select box arrows you may define or modify the vial size of the
reagent bottle used.

The following vial sizes are available:

Reagent positions Selectable vial sizes

15ml Glass
12ml Glass
R1 - R3
5ml Glass
4ml Plastic

5ml Glass
R4 - R15 4ml Plastic
Cup

Calibrate By using this check box button you may define or modify whether a reagent
requires calibration, see chapter 5.4.5 Calibration. Set individual parameters
On/Off by pressing onto the field, ON = .

Once a reagent is set to calibrate all tests using this reagent must be
NOTE
calibrated, e.g. when changing the reagent lot. Test having an
invalid calibration may not be used for a run.

Stirring Parameters This parameter frame allows definition of stirring parameters used for reagents
which require stirring due to the risk of sedimentation of particles. Such
reagents must be placed in the reagent area either on position R1 and/or R3,
see chapter 3.1.2 Working Plate. The following settings are available:

Stirred By pressing into the Stirred input field a keypad opens allowing the entry or
modification of the stirrer rpm (revolutions per minute). 0=off indicates the
reagent does not require stirring. Input range: 0 - 500 (rpm).

Cycle time By pressing into the Cycle time input field a keypad opens allowing the
entry or modification of the total duration (interval) of the stirring cycle in
seconds. After a stirring interval ended a new interval will be started.
Input range: 0, 10 - 500 (sec.).

Cycle Pause By pressing into the Cycle pause input field a keypad opens allowing the
entry or modification of the duration of the pause time until a new cycle time
starts again. Input range: 0 - 499 (sec.). Input must be 5 sec. less the
entered Cycle time, e.g. Cycle time = 200, entry range at Cycle Pause =
0…195 sec.

CoaLyser – Operators Manual 1.3 Page 134


SETUP

The difference between Cycle time and Cycle Pause results to the
NOTE
actual stirring time, e.g. Cycle time = 60 sec. and Cycle Pause = 40
sec. the actual stirring time is 20 sec. (60s-40s).

Tools The Tools button in the Edit Reagent No.1 Page 1 window offers additional
options and access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 168 Tools in Edit Reagent Page 1

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Edit Reagent Main Screen.

Figure 169 Edit Reagent Main Screen

CoaLyser – Operators Manual 1.3 Page 135


SETUP

5.4.3.2 Parameter Page 2


To enter further parameters in this section continue by pressing the Page 2
button allowing entry or modification of the following parameters:

Figure 170 Edit Reagent Page 2

Aspiration By pressing the Select Box arrow buttons you may select between 4
predefined liquid aspiration profiles. These profiles contain parameters
related to the speed the liquids will be aspirated by the probe of the
analyser.

The following aspiration profiles are available:

Available profiles Description

max_speed: 6500 ms
Profile 1 (Default) ramp_speed: 250 ms
ramp_break: 100 ms

max_speed: 4000 ms
Profile 2 ramp_speed: 200 ms
ramp_break: 60 ms

max_speed: 4000 ms
Profile 3 ramp_speed: 300 ms
ramp_break: 35 ms

max_speed: 5000 ms
Profile 4 ramp_speed: 160 ms
ramp_break: 30 ms

The manufacturer recommends leaving the profiles always set to


default. Before changing the profiles consult your local dealer or the
NOTE
manufacturer!

CoaLyser – Operators Manual 1.3 Page 136


SETUP

Dispensing By pressing the Select Box arrow buttons you may select between 4
predefined liquid dispensing profiles. These profiles contain parameters
related to the speed the liquids will be distributed by the probe of the
analyser.

The following dispensing profiles are available:

Available profiles Description

max_speed: 6500 ms
Profile 1 (Default) ramp_speed: 250 ms
ramp_break: 100 ms

max_speed: 4000 ms
Profile 2 ramp_speed: 200 ms
ramp_break: 60 ms

max_speed: 4000 ms
Profile 3 ramp_speed: 300 ms
ramp_break: 35 ms

max_speed: 5000 ms
Profile 4 ramp_speed: 160 ms
ramp_break: 30 ms

The manufacturer recommends leaving the profiles always set to


default. Before changing the profiles consult your local dealer or the
NOTE
manufacturer!

Washing By pressing the Select Box arrow buttons you may select between 5
predefined washing profiles. These profiles contain parameters related to
the rinse and wash duration times and volumes of liquids required.

Type Cycle description

Normal System water, 1 sec. rinsing

Washer 1 (Pos. W1)


- 3 sec. rinsing with system water
Intensive 1 - 150µl Washer 1
- 3 strokes up to 250µl
- 3 sec. rinsing with system water
Washer 2 (Pos. W2)
- 5 sec. rinsing with system water
Intensive 2 - 150µl Washer 2
- 3 strokes up to 250µl
- 5 sec. rinsing with system water
Washer 1 (Pos. W1)
- 1.5 sec. rinsing with system water
Intensive 3 - 100µl Washer 1
- 5 strokes up to 250µl
- 1.5 sec. rinsing with system water

Intensive 4 4 sec. rinsing with system water.

CoaLyser – Operators Manual 1.3 Page 137


SETUP

Waste Vol. By pressing into the Waste vol. input field a keypad opens allowing the
entry or modification of the waste volume in µl. The waste volume is taken in
addition to the dispensing volume in order to obtain a better pipetting
accuracy and will be discarded later in the rinsing station. Default = 5µl.
Input range: 2 - 20 (µl).

Tests, Reagent/Buffer is
assigned to This window provides information on the tests to which a reagent/buffer as
edited or modified in this section is assigned to.

Tools The Tools button in the Edit Reagent No.1 Page 2 window offers additional
options and access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 171 Tools in Edit Reagent Page 2

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Edit Reagent Main Screen.

CoaLyser – Operators Manual 1.3 Page 138


SETUP

Tools
Tools button in the Edit Reagent Main Screen offer several options and
access to other menus and actions.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 172 Tools in Edit Reagents Main Screen

The following Tools are available:

Available Tools Description

By pressing this button the selected reagent will


be fully deleted from the reagent list. Reagents
Delete
assigned to certain tests require the test to be
deleted first.
By pressing this button the selected reagent
may be copied and saved with the same
Save as
parameters as another reagent in the reagent
list.

By pressing this button you will leave this


Quit
window without any changes.

By pressing this button the reagent parameters


Print will be printed via the external connected
printer.

By pressing this button the edit reagent screen


Hardcopy will be printed, provided a printer is connected
to the analyser.

ESC&Exit By pressing the ESC&Exit button you will leave the Edit Reagent Menu
without saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Edit Reagent Menu confirms all
changes and modifications done in this menu and starts level check of all
reagents and washers on board.

CoaLyser – Operators Manual 1.3 Page 139


SETUP

5.4.4 Setup Test Definition


This chapter describes how to define a new test or to modify an existing test.

 From the Main Menu, press the Setup button to access the Setup Menu.

 From the Setup Menu, press the Test button to access the Select Test
Screen.

Figure 173 Select Test Screen

The Select Test Screen contains a list of all tests saved. A total of 50 tests
can be saved. Free test definition files can be programmed by the user.

1) To display or modify the data of a test use the arrow keys to first select
the test and then press the OK button to continue.

2) To define a new test use the arrow keys and move upward until a free
position appears and also press the OK button to continue.

The following example provides description on how to program a PT test.


Proceed the same way if you intent to enter new test data or to modify
existing test.

 In the Select Test Screen chose PT and confirm by pressing the OK


button.

Figure 174 Test Menu Main Page

CoaLyser – Operators Manual 1.3 Page 140


SETUP

Name
By pressing into the Name input field a keypad opens allowing the entry or
modification of the test name. The test name may have up to 15
alphanumerical characters.

Abbrev.
By pressing into the Abbreviation (abbrev.) input field a keypad opens
allowing the entry or modification of the abbreviation to be assigned to a
test. The test abbreviation may have up to 5 alphanumerical characters.
This abbreviation is used in other menu screens for identification of the test.

Parameters
The parameter frame allows access to other sections required to define or
modify a test. The following sections are available:

The following Parameters are available:

Parameters Description
To define all liquid related parameters, such as:
- Sample Volume
Liquids - Buffer/Diluent dilution
- Reagent Volumes
- Incubation times

To define evaluation and calculation modes, such as:


- 1st Evaluation
Evaluation - 2nd Evaluation
- Wavelength setting
- Max. Measuring Time

To define additional test parameters, such as:


- No of determination
Determin.
- Max. Deviation & Lag Phase
(Determination)
- Mixing parameters
- Raw Value Parameters

To define the parameters to be displayed or printed in


Show flags
the short from result list.
To assign certain control plasma parameters to
QC defined controls, such as:
(only available if QC - Type of result
Software is installed - Control Plasma
to the analyzer) - Target value in seconds
- 1S-Deviation

CoaLyser – Operators Manual 1.3 Page 141


SETUP

5.4.4.1 Liquids
From the Test Menu Main Screen press the Liquids button to display or
modify liquid related parameters for a test, like sample and reagent volumes.

The following display appears:

Figure 175 Test Menu Liquids

Sample By pressing into the Sample input field a numerical keypad opens allowing
the entry or modification of the total Sample Volume in µl. For diluted
samples (undiluted sample + buffer) or for undiluted samples, this
corresponds to the required Sample Volume. Input range: 2.0 - 250.0 (µl).

Sample Start By using this check box button you may define or modify whether the
measurement will be started either with Sample or Reagent. Set individual
parameters On/Off by pressing onto the field, e.g. ON (Sample Start set) = .

Buffer/Diluent By pressing into the Buffer/Diluent input field a numerical keypad opens
allowing the entry or modification of the Dilution Value. After entering a
Dilution Value > 0µl (0= no dilution) a selection window appears allowing
you to select a Buffer/Diluent. Input range: 0.0 - 49.0 (at 100µl Sample Vol.).

 Enter the Dilution Value


 Select the Buffer/Diluent

Example: 50µl Sample Volume total


Pipetting: 2µl Sample + 48µl Buffer/Diluent
Entry: 1+24

The range depends on the total Sample Volume, whereas the minimum
pipetting volume is 2µl.

Incubation time By pressing into the Incubation (incubat.) input field a numerical keypad
opens allowing the entry or modification of the Incubation Time in sec.
Input range: 0.0 - 999.0 (sec.).

Reagent 1 + 2 Up to 2 additional reagents can be selected from the Select Reagent List. By
pressing into the Reagent 1 + 2 input fields a numerical keypad opens
allowing the entry or modification of the Reagent 1 + 2 volumes in µl. After
entering a Reagent Volume > 0µl (0= no reagent) a selection window
appears allowing you to select a reagent and to define additional incubation
times for each reagent selected. Input range: 0.0, 2.0 - 250.0 (µl).

CoaLyser – Operators Manual 1.3 Page 142


SETUP

 Enter the Reagent Volume


 Select the reagent
 Enter Incubation Times for reagent 1+2

The reagents will be added to the cuvette in order listed.

Start Reagent By pressing into the Start Reagent input field a numerical keypad opens
allowing the entry or modification of the Start Reagent volume in µl. After
entering a Start Reagent volume > 0µl (0= no reagent) a selection window
appears allowing you to select a Start Reagent. Input range: 0.0, 2.0 - 250.0
(µl).

 Enter the Start Reagent volume


 Select the Start Reagent

NOTE The total volume in the cuvette must be between 150µl and 300µl.

Tools The Tools button in the Test Menu Liquids Screen offers additional options
and access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 176 Tools in Test Menu Liquids

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Main Screen.

CoaLyser – Operators Manual 1.3 Page 143


SETUP

5.4.4.2 Evaluation
From the Test Menu Main Screen press the Evaluation button to display or
modify evaluation mode parameters for a test. The following display
appears:

Figure 177 Test Menu Evaluation

1st Evaluation By pressing the Select Box arrow buttons you may assign clotting detection
with seconds as raw values. This evaluation mode is used to analyze the
measuring signal (= reaction curve) to be applied for test using a first
evaluation.

The following selection is available:

Available methods Description


Sets the evaluation method to clotting and
allows entry of additional parameters by using
the Tech button.
Clotting (sec.) Calibration parameters for this test can only be
entered if clotting is selected, see chapter
5.4.5.1 Edit Calibration for Clotting (manual
calibration).
No evaluation method is set. No calibration
off parameters can be assigned to tests using 1st
evaluation.

After setting the evaluation method to clotting [sec.] an additional button


Tech appears allowing the entry of additional parameters.

 Press the Tech button.

CoaLyser – Operators Manual 1.3 Page 144


SETUP

Figure 178 Test Menu Tech Clotting

Valid Clotting By pressing into the Valid Clotting input fields a numerical keypad opens
allowing the entry or modification of the min. /max. Valid Clotting Range in
sec. The Valid Clotting Range defines the range within a measured clotting
time will be detected and the range of raw values which may be entered as
calibration values, refer to chapter 5.4.5.1 Edit Calibration for Clotting
(manual calibration). Measured clotting times or calibration points outside
this range will be marked with an error flag “none”.

Valid Clotting Input Ranges and dependency


Times
Input range: 1 - 3.600 (sec.)
Min. Valid Lowest value: >/= Lag Phase
Clotting >/= Learn Phase
Highest value: < max. Valid Clotting

Input range: up to 3.600 (sec.)


Max. Valid Lowest value: > min. Valid Clotting
Clotting Highest value: </= Measuring time (auto repeat off)
Highest value </= 2x Measuring time (auto repeat on)

Auto-repeat By using this check box button you may define or modify whether the
measurement will be repeated automatically if a measurement can not be
detected in the first run, e.g. if the max. measuring time has been passed
over and no clotting could be found. Set individual parameters On/Off by
pressing onto the field, e.g. ON =  (Auto-repeat active).

By setting this button a measurement without clot detection will be


automatically repeated after the initial run has been finished. These
measurements will be repeated with the double measuring time, see chapter
5.4.4.2 Evaluation / Measuring Time.

NOTE It is mandatory to set the max. valid clotting time to 2x the set
measuring time if auto repeat is active otherwise the clotting times
found outside the valid clotting time will not be calculated and the
result would be shown as “none”.

CoaLyser – Operators Manual 1.3 Page 145


SETUP

Example: Measuring Time = 30 sec.


There is no result detected within the 30 sec. Run will be
repeated automatically with a Measuring Time of 60 sec. Max.
valid clotting time must be set to max. 60sec.

Learn Phase By pressing into the Learn Phase input field a numerical keypad opens
allowing the entry or modification of the learn phase in sec. The Learn
Phase defines the time range of the threshold width used for the clotting
evaluation.

The defined learn phase time is also depending on the Lag phase setting,
see section Determin. The learn phase may not be higher than the entered
min. Valid Clotting time.
Input range: 1 - 60 (sec.).

Example:
Lag Phase = 8 sec.
Learn Phase = 1 sec.
That means: Lag Phase (8) - Learn Phase (1) = 7 seconds after start of a
measurement the evaluation for the clotting detection begins. Whereas the
min. Valid Clotting Time may not be lower than 8 sec. because during the
Learn Phase the thresholds for clot detection are adjusted by the software.

Algorithm of clot By pressing the Select Box arrow you may select out of 12 predefined
detection evaluation algorithms having different influence on the clotting detection.
Once selected a short description will be provided below the Select Box
window.

The manufacturer recommends leaving the algorithm as set to


default. Before changing the algorithm consult your local dealer or
NOTE
the manufacturer!

Scaling By pressing into the Scaling input field a numerical keypad opens allowing
the entry or modification of the Scaling Factor. The Scaling Factor is used to
adjust the photometrical measuring value as a multiplier for the extinction
values respectively the transmission values or digits depending on the
parameterized raw values. Input range: 0.1 - 99999. Default = 20.

Tools The Tools button in the Tech Clotting Screen offers additional options and
access to other menus and features.

The following Tools are available:

Available Tools Description


By pressing this button you will leave this
Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Evaluation Screen.

CoaLyser – Operators Manual 1.3 Page 146


SETUP

Figure 179 Test Menu Evaluation/Kinetic

2nd Evaluation By pressing the Select Box arrow buttons you may select between 3
predefined evaluation modes. These methods are used to analyze the
measuring signal (= reaction curve) to be applied for tests using a second
evaluation.

The following selection is available:

Available modes Description

No evaluation method is set. No calibration


Off parameters can be assigned to tests using
2ndevaluation.

Sets the evaluation method to Kinetic and


allows entry of additional parameters by using
Kinetic [mE/min] the Tech button.
Calibration parameters for this test can be
entered if kinetic is selected, see chapter 5.4.5.2

Sets the evaluation method to Difference and


allows entry of additional parameters by using
the Tech button.
Difference [mE]
Calibration parameters for this test can be
entered if difference is selected, see chapter
5.4.5.2

Sets the evaluation method to 1st derivation and


allows entry of additional parameters by using
the Tech button.
1st derivation [mE/s]
Calibration parameters for this test can be
entered if 1st derivation is selected, see chapter
5.4.5.2

After setting the evaluation method to Kinetic [mE/min], Difference [mE] or


1st derivation [mE/s] an additional button Tech appears allowing the entry
of additional parameters.

 Press the Tech button and the following window appears.

CoaLyser – Operators Manual 1.3 Page 147


SETUP

Figure 180 Test Menu Tech Kinetic

Evaluation Points By pressing into the Evaluation Points input fields numerical keypads
open allowing the entry or modification of the T1/T2 Evaluation Points in sec.

Evaluation Points Input Ranges

Input range: 1 - 3.600 (sec.)


T1 Lowest value: Lag Phase + 1 sec.
Highest value: T2 - 1sec.

Input range: up to 3.600 (sec.)


T2 Lowest value: T1 + 1sec.
Highest value: Measuring time - 2 sec.

The settings of the T1/T2 Evaluation Points have different effects on the
evaluation method selected:

Selection Description

The evaluation of the slope takes place in the


Kinetic [mE/min] defined time range between T1 and T2.

The evaluation of the extinction difference takes


place by means of measuring values Ext. [T2] -
Ext. [T1] if rising is set active.
Difference [mE]
The evaluation of the extinction difference takes
place by means of measuring values Ext. [T1] -
Ext. [T2] if rising is set inactive.

The evaluation of the maximum slope [mE/sec.]


1st derivation [mE/s] and the time [sec.] takes place in the time range
up from T1 and until T2.

Rising Values By using this check box button you may define or modify whether the
expected raw values of the measuring results are rising or falling. Set
individual parameters On/Off by pressing onto the field, e.g. ON =  (rising
values set).

CoaLyser – Operators Manual 1.3 Page 148


SETUP

Correlation Coefficient By pressing into the Correlation Coefficient input field a numerical keypad
opens allowing the entry or modification of the Correlation Coefficient used
for the evaluation of the linearity of a calibration curve.
Input range: 0 - 1,000.

Assuming the measured values between T1 and T2 rise or fall


approximately linear the slope is calculated by linear regression over all
measurements between T1 and T2. This is the searched raw result in
mE/min. In addition, the correlation coefficient is calculated. Is this
coefficient close to 1, the measured values corresponding well to this line, if
close to 0 they doing not - e.g. because the values are extremely noisy, or
because the curve has too much curvature. Therefore a limit on the
correlation coefficients is set in order to decide whether a curve can be
evaluated or not.

This parameter is only for a Kinetic evaluation method.

Tools The Tools button in the Tech Kinetic Screen offers additional options and
access to other menus and features.

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Evaluation Screen.

Figure 181 Test Menu Evaluation/Kinetic

Wavelength By pressing the Select Box arrow buttons you can select between the
available wavelengths depending on the test requirements.

Depending on the analyser model the following wavelength are available:

 405 nm (only for analyzers equipped with 1wavelenght)


 570 nm
 740 nm

CoaLyser – Operators Manual 1.3 Page 149


SETUP

Measuring Time By pressing into the Measuring Time input field a numerical keypad opens
allowing the entry or modification of the Measuring Time in sec. The reaction
curve is evaluated between the end of the lag phase and this time
(Measuring Time). Input range: 20 - 3.600 (sec.).

Tools The Tools button in the Test Menu Evaluation Screen offers additional
options and access to other menus and features.

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Main Screen.

Figure 182 Test Menu Main Page

5.4.4.3 Determination
From the Test Menu Main Screen press the Determin. button to display or
modify test related parameters, like no. of determination, mixing and raw
value parameters. The following display appears:

Figure 183 Test Menu Determin.

CoaLyser – Operators Manual 1.3 Page 150


SETUP

No. of Determination By pressing into the No. of Determination input field a numerical keypad
opens allowing the entry or modification of the number of determination
(single or multiple) assigned to a test.
Input range: 1 - 32 (Default = 1).

Is the number of determination set to multiple, meaning > 1 the maximum


deviation must be defined to specify to what extent each single value may
deviate, see Max. Deviation.

Max. Deviation By pressing into the Max. Deviation (CV) input field a numerical keypad
opens allowing the entry or modification of the maximum deviation in %. The
maximum deviation indicates the permitted deviation of the measured values
for multiple determinations. This field only appears if the No. of
Determination is set to > 1. Input range: 1 - 20 (%).

Once the single values are outside of this range the single values will be
marked by the software.

Lag Phase By pressing into the Lag Phase input field a numerical keypad opens
allowing the entry or modification of the Lag Phase in sec. The reaction
curve is evaluated after the end of this time (after start of measurement) and
the defined measuring time. Input range: 0 - 60 (sec.).

Mixing rpm 1 By pressing into the rpm1 input field a numerical keypad opens allowing
the entry or modification of the mixer (stirrer) start speed at start of
measurement. Input range: 0, 200 - 999 (rpm).

Mixing rpm 2 By pressing into the rpm2 input field a numerical keypad opens allowing
the entry or modification of the mixer (stirrer) end speed after termination of
the mixing time (mixing ramp). Input range: 0, 200-999 (rpm).

Example: rpm1 = 500; rpm 2 = 800, ramp = 5 sec.


The measurement will start with 500 rpm mixer revolutions. The final
revolutions of 800 rpm will be reached after 5 seconds until the end of the
entered measuring time.

Mixing ramp By pressing into the ramp input field a numerical keypad opens allowing
the entry or modification of the time until rpm2 starts. Input range: 2-300
(sec.).

Raw Value Parameters The Raw Value Parameter section allows parameterization of the raw value
settings used for the evaluation algorithm to calculate the measuring result.
Additionally the raw value ranges within the raw values will be calculated
can be defined.

Raw Values By pressing the Select Box arrow buttons you can select between 4
available evaluation algorithms.

The following evaluation algorithms are available:

Available algorithms Description


Sets the evaluation algorithm to Extinction, raw
Extinction [mE]
values will be calculated in [mE].
Sets the evaluation algorithm to 1st Derivation,
1st derivation [mE/s]
raw values will be calculated in [mE/s].
Sets the evaluation algorithm to Digits, raw
Digits
values will be calculated in digits.
Sets the evaluation algorithm to Transmission,
Transmission [%]
raw values will be calculated in [%].

CoaLyser – Operators Manual 1.3 Page 151


SETUP

Valid By pressing into the Valid input field numerical keypads open allowing the
entry or modification of the range within the raw values will be evaluated.
Raw values outside this range will be marked. Input range: see below table.

Evaluation algorithm Input Ranges

Extinction [mE] - 500 - + 3200

- 200 - + 200
Exception: To evaluate this algorithm, 1st
1st deviation [mE/s] deviation must be set under Raw Value as well
as in the Evaluation section as 2nd Evaluation,
see chapter 5.4.4.2 Evaluation

Digits 0 - + 8388607

Transmission [%] - 50 - + 150

Smoothing The Smoothing button allows you to select between a numbers of digital
filters available to even (smooth) the measuring signals.

 Press the Smoothing button and the following window appears.

Figure 184 Test Menu Misc. Smoothing

The Select type of digital lowpass screen contains a list of digital filter
settings available. To select a digital filter for a test use the arrow keys to
first select the filter and then press the OK button to continue. By pressing
the ESC button you will leave this screen without any selection.

It is recommended to consult your local dealer or the manufacturer


NOTE
first before changing the digital filter setting!

Tools The Tools button in the Test Menu Miscellaneous Screen offers additional
options and access to other menus and features.

CoaLyser – Operators Manual 1.3 Page 152


SETUP

The following Tools are available:

Available Tools Description


By pressing this button you will leave this
Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Main Screen.

Figure 185 Test Menu Main Page

5.4.4.4 Show Flags (Define Results/Run list)


Within this section you may define the contents respectively results to be
shown in the Results per run list, see chapter 5.2.6 Show/print results.

From the Test Menu Main Screen press the Show-Flags button to display or
modify the contents of the short form result list.

The following display appears:

Figure 186 Test Menu Show-Flags

CoaLyser – Operators Manual 1.3 Page 153


SETUP

Set flags By using the check box buttons you may define or modify the results
separated by evaluation mode to be displayed and or printed within the
results per run list. The availability of the result flags depends on the
selection done in 1st and 2nd evaluation.

Set individual parameters On/Off by pressing onto the field, ON =  (active).

Tools The Tools button in the Test Menu Show-Flags Screen offers additional
options and access to other menus and features.

The following Tools are available:

Available Tools Description


By pressing this button you will leave this
Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Main Screen.

Figure 187 Test Menu Main Page

5.4.4.5 Define QC-Control Parameter


(only available if QC software is installed on analyzer)

Within this section you may assign up to three control plasmas to the
selected test if part of a Quality Control Period. See also chapter 5.3 Quality
Control (QC Program).

 From the Test Menu Main Screen press the QC button to assign control
plasmas and or modify the parameters and result calculations.

The following display appears:

CoaLyser – Operators Manual 1.3 Page 154


SETUP

Figure 188 Test Menu Select Control

Up to three different Control Plasmas can be assigned individually to a test.


By pressing onto the individual buttons you enter into the QC Control
parameter page. Repeat the procedure for each Control Plasma to be added
to the test as described as follows.

Figure 189 Test Menu QC

Select result type By pressing the Select Box arrow buttons you can select between 6 different
types of results, meaning whether the QC result calculation will be based on
raw values or converted values.

Converted values can only be selected and memorized if the test is


NOTE
subject to a calibration and has a valid calibration, see chapter 5.4.5
Calibration. Otherwise only result types are selectable which are
applicable to the test selected.

CoaLyser – Operators Manual 1.3 Page 155


SETUP

The following result types are available:

Available result types Description


Sets the result calculation to clotting, raw values
Clotting
will be calculated in [sec].
Sets the result calculation to clotting conversion,
Clotting 1st conversion
1st conversion values will be calculated in [%].
Sets the result calculation to clotting conversion,
Clotting 2nd Conversion
2nd conversion values will be calculated in [INR].
Sets the evaluation algorithm to Kinetic,
Difference or 1st Derivation (depending on test
Kinetic/Difference/1st
setup), raw values will be calculated in [mE/s]
derivation raw
for Kinetic, [mE] for Difference, [Delta mE/sec]
for 1st derivation.
Sets the evaluation algorithm to Kinetic,
Kinetic/Difference/1st Difference or 1st Derivation (depending on test
derivation conversion setup), converted values will be calculated in
the unit defined in calibration setup.
Sets the evaluation algorithm to the maximum
Max. Time time of the measuring signal in 1st Derivation
mode. Values will be calculated in [sec]
Off No result calculation

Control Plasma By pressing onto the blue field a window opens allowing you to select
Control Plasma from a list of previously defined Controls, see 5.3.1 Control
Setup.

The following window appears.

Figure 190 Test Menu Select Control List

The Select Control Plasma Screen contains a list of all previous Controls
memorized. A total of 8 Controls can be defined and memorized, refer to
chapter XYZ to define Control Plasmas and start a QC period.

 To select a Control use the arrow keys to first select the Control Plasma
and then presses the OK button to continue.

CoaLyser – Operators Manual 1.3 Page 156


SETUP

The selected Control Plasma appears in the following window:

Figure 191 Test Menu selected Control

Target value By pressing into the Target value input field a numerical keypad opens
allowing the entry or modification of the target value expected to be
achieved from the control plasma and test measurement. Depending on the
test and result type selected the target value must be entered either as raw
value, e.g. in seconds or as converted value, e.g. % (PT).

1S-deviation By pressing into the 1S-deviation input field a numerical keypad opens
allowing the entry or modification of the 1 sigma deviation based on the
target value entry. Input Range: ≤5% and ≥20% of the target value.

Example:
Target Value = 12 sec.
1S-Deviation = min. 0.6 sec. / max. 2.3 sec.

Based on the 1S-deviation entry the ±1S, ±2S and ±3S thresholds are
calculated.

1S-deviation definition:
 The single standard deviation corresponds to the ± 1S Range
 The double standard deviation corresponds to the ± 2S Range
 Three times the standard deviation corresponds to the ± 3S Range.

Delete By pressing the Delete button in the Test Menu QC Screen your will delete
the parameter setting and remove the Control Plasma from the test setup
specifications.

Control Plasma cannot be deleted if Control Plasma is used in an


NOTE
active QC-Period. Active QC-Periods must be finished and printed
first.

Tools The Tools button in the Test Menu QC Screen offers additional options and
access to other menus and features.

CoaLyser – Operators Manual 1.3 Page 157


SETUP

The following Tools are available:

Available Tools Description


By pressing this button you will leave this
Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Test Menu Main Screen.

Figure 192 Test Menu Main Screen

Tools
The Tools button in the Test Menu Main Screen offer several options and
access to other menus and actions.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 193 Tools in Test Menu Main Page

CoaLyser – Operators Manual 1.3 Page 158


SETUP

The following Tools are available:

Available Tools Description


By pressing this button the selected test will be
Delete fully deleted from the test list. Reagents assigned
to certain tests require the test to be deleted first.
By pressing this button the selected test may be
Save as copied and saved with the same parameters as
another test in the test list.
By pressing this button you will leave this window
Quit
without any changes.
By pressing this button the test parameters will be
Print
printed via the external connected printer.
By pressing this button the Test Menu Main
Hardcopy screen will be printed, provided a printer is
connected to the analyser.

ESC&Exit By pressing the ESC&Exit button you will leave the Test Menu without
saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Test Menu confirms all changes and
modifications done in this menu. Prior of acceptance the software will check
conformity on all edited parameters. In case of any invalid inputs the
software will indicate the corrections required.

CoaLyser – Operators Manual 1.3 Page 159


SETUP

5.4.5 Calibration
Use this menu to calibrate a test manually or by automated calibration, to
define or change the calibration parameters for a test. A calibration is
required for test to which the raw values need to be converted into
concentration units/activities.

The test must/can be calibrated if:

 If the evaluation mode for 1st and 2nd evaluation is not set to OFF, refer
to chapter 5.4.4.2 Evaluation.

 A Lot Number was added or changed in a reagent set to calibrate, see


chapter 5.4.3.1 Parameter Page 1.

 If reagent parameters are changed using a reagent set to calibrate, see


chapter 5.4.3.1 Parameter Page 1.

A calibration is not possible if reagents are not set to calibrate, see


NOTE
section Reagent Setup/Page 1/Calibrate (check box) in chapter
5.4.3.1 Parameter Page 1.

 From the Main Menu, press the button Setup to access the Setup
Menu.

Figure 194 Setup Menu

 From the Setup Menu, press the button Calibration to access the
Calibration Menu.

CoaLyser – Operators Manual 1.3 Page 160


SETUP

Figure 195 Select test Calibration

The Select Test Screen for calibration contains a list of all tests available.

 Select a test for calibration by using the arrow keys first to select the test
and then press the OK button to continue.

The calibration for test screen appears.

Figure 196 Calibration Menu Main Page

General Settings The analyzer offers two different ways to calibrate a test.

- Under Edit Calibration a test can be calibrated manually by editing the


calibration data previously measured. Enter manually the evaluated
results achieved by a normal measurement with dilutions or if available
with standard plasmas or those which have been provided in the
package inserts of the reagent supplier used.

- Under Auto Calibration a test will be automatically diluted from a


standard and the measured calibration data will be automatically added
into a calibration curve and memorized for the test.

The following chapters will describe how to enter or change parameters for
each way of calibration:

CoaLyser – Operators Manual 1.3 Page 161


SETUP

Chose mode of calibration By pressing the Select Box arrow button you can select between 2 available
calibration modes. The different modes define how the raw value of a
calibration curve are converted and further allows entry of calibration
parameters for tests requiring 2 different calibration curves, for example
derived Fibrinogen.

The following calibration modes are available:

Calibration modes Description

Converts raw values [sec.] into a defined


Clotting
unit/activity.

Kinetic: Converts raw value [mE/min.] into a


defined unit/activity.

Kinetic / Difference / Difference: Converts raw value [delta mE] into a


1st derivation defined unit/activity.

1st deri.: Converts raw value [mE/sec.] into a


defined unit/activity.

Before selecting the calibration mode in this screen the


NOTE corresponding evaluation mode must be assigned to the test first,
see chapter 5.4.4.2 Evaluation to assign the 1st and or 2nd
evaluation to a test also defining whether the test allows entry of 2
calibration curve parameters.

5.4.5.1 Edit Calibration for Clotting


(manual calibration)

 Select Clotting by using the arrow keys to enter calibration parameters


for tests based on clotting or to define parameters for a test using 1st
evaluation.

Calculation Press the Calculation button in Edit Calibration to access the Calibration
Menu for Clotting, the following window appears. Contents of the boxes that
appear on the screen vary depending on the units selected in the test.

Figure 197 Calibration Menu Clotting

CoaLyser – Operators Manual 1.3 Page 162


SETUP

1st conversion By pressing the Select Box arrow button you can select between 2 available
conversion modes to define the 1st conversion.

The following 1st conversion modes are available:

Conversion modes Description


No conversion of raw values, results will be shown
in seconds only. By selecting none, no further
None
entries on this page are required. No calibration
curve is measured
Allows conversion of raw values into
units/activities. By selecting this conversion mode
Interpolation further entries are required to parameterize the
test and calibration curve. Values are interpolated
from point to point.
Allows conversion of raw values into
units/activities. By selecting this conversion mode
Regression further entries are required to parameterize the
test and calibration curve. Regression is the best
fit over all measured calibration points.

Unit conversion By pressing into the Unit conversion input field a alphanumerical keypad
opens allowing the entry or modification of the unit/activity for conversion of
the raw value, meaning from sec. into unit/activity, e.g. %, mg/dl, g/l.

These conversion units will be used if QC Controls are set to result type
conversion, see chapter 5.4.4.5 Define QC-Control Parameter, Select
result type.

Mode in/out By pressing the Select Box arrow button of the left select box you can define
the scaling of the x-axis of the calibration curve graph whereas the right
select box will define the y-axis setting of the calibration curve graph. The
following selections are available for both axis:

 log - logarithmic calculations are applied to values before graphing.


 rec - the reciprocal (1 divided by the value) is used on the graph.
 lin - Linear value is used directly on the graph.

The mode in/out selection lin/lin is the most common setting used. Other
combinations, e.g. log/rec may require certain modifications in the
conversion range values, see Conversion range input field to display the
calibration curve graph.

Output format By pressing the Select Box arrow button you can define the decimal point
setting for the measuring result display. There are 4 alternatives available:

 <xxx.x> Input range: 0.1 - 999.9 (units)


 <xx.xx.> Input range: 0.01 - 99.99 (units)
 <x.xxx> Input range: 0.001 - 9.999 (units)
 <xxxx.> Input range: 0 - 9999 (units)

Rising By using this check box button you may define or modify whether the
expected converted values (unit/activity) of the measuring results are rising
or falling. Set individual parameters On/Off by pressing onto the field, e.g. ON
=  (converted values rising).

CoaLyser – Operators Manual 1.3 Page 163


SETUP

Calibration curves must be entered in such a way that you start with the
highest % values and a shortest measuring time (raw value) in seconds for
falling curves (coagulation factors). For rising curves (inhibitors or INR-Cal)
you have to start with the lowest % / INR values and the shortest measuring
times (raw value) in seconds. The software will support this procedure only.

The arrow marking in the Calib Value page indicate whether the
NOTE values must rise or fall. Converted values can be set to rising or
falling. Raw values must always rise.

Convers. range By pressing into the Conversion range input field a numerical keypad
opens allowing the entry or modification of the extrapolated result limits
within the results will be calculated and displayed in the calibration curve
graph. Input range: 0.1 - 9999 (depending on defined output format).

2nd conversion By pressing the Select Box arrow button you can select between 3 available
conversion modes to define the 2nd conversion.

The following 2nd conversion modes are available:

Conversion modes Description


No 2nd conversion of raw values, results will be
shown in sec. and defined activity only. By
None
selecting none, no further entries for MNPT or ISI
are possible.
Allows conversion of the INR (International
Normalized Ratio). Selecting this conversion
INR mode requires further entry of the ISI
(International Sensitivity Index) as provided by the
reagent package insert.

Allows conversion of the Ratio (Sample / MNPT =


Ratio
mean normal PT or aPTT).

MNPT By pressing into the MNPT input field a numerical keypad opens allowing
the entry or modification of the Mean Normal PT time at 100% activity in
sec. The same input field shall be used for setting the Mean Normal aPTT
time for evaluating the Ratio. Input range: 1 - 300 (sec.).

ISI By pressing into the ISI (International Sensitivity Index) input field a
numerical keypad opens allowing the entry or modification of the ISI value
according to reagent package insert. Input range: 0.01 - 99.99.

The ISI input field only appears if INR is selected in 2nd conversion.

Tools The Tools button in the Test Menu Show-Flags Screen offers additional
options and access to other menus and features.

The following Tools are available:

Available Tools Description


By pressing this button you will leave this
Quit
window without any changes.

Hardcopy Performs a print of the screen only.

CoaLyser – Operators Manual 1.3 Page 164


SETUP

OK Press the OK button to return back to the Calibration Menu Main Page.

Figure 198 Calibration Menu Main Page

Enter Calibration Values Press the Calib Values button in Edit Calibration to access the Calibration
Menu for Clotting, the following window appears. Contents of the boxes that
appear on the screen vary depending on the units selected in the test.

Figure 199 Calibration Values 1st evaluation

To enter calibration curve points manually perform one of the following:

Option 1:

1. If available use the values as provided in the package insert of the


reagent manufacturer.

Option 2:

 Enter manually the evaluated results achieved by a normal


measurement with dilutions or if available with standard plasmas.

Prior entering the calibration curve values make sure the values correspond
to the following settings:

CoaLyser – Operators Manual 1.3 Page 165


SETUP

 Raw values entered must be inside the defined Valid Clotting range,
see chapter 5.4.4.2 Evaluation.

 Converted values entered must be inside the defined conversion


range, see chapter 5.4.5.1 Edit Calibration for Clotting (manual
calibration), Convers. range

 Values entered must correspond to the rising/falling definition, see


chapter 5.4.5.1 Edit Calibration for Clotting (manual calibration),
Rising

Unit [%] By pressing into the unit, here % input field a numerical keypad opens
allowing the entry or modification of the converted value (unit/activity).

Points on the calibration curve that are not to be used must have the
NOTE entry 0.0%. A minimum of 2 points must be defined. It is
recommended defining at least 3 points. In maximum 7 calibrations
curve points can be defined.

Raw Value [sec.] By pressing into the Raw Value, here sec. input field a numerical keypad
opens allowing the entry or modification of the raw values in sec.

Dilution Within this section the dilution based on the entered values (converted and
raw values and the defined concentration value will be calculated
automatically by the software.

Example: 1+ 9.0 = 1 part standard plasma will be diluted with 9 parts


buffer or a 1:10 dilution.

invalid = The converted value entered deviates from the


following conditions:

1. Value is outside the defined conversion range, see chapter


5.4.5.1 Edit Calibration for Clotting (manual calibration),
Convers. range.

2. Value is higher as the set concentration value, see


Standard / Concentration in this chapter.

Standard This section allows entry of further information on Standard Plasmas used to
verify a calibration curve.

Concentration By pressing into the Concentration input field a numerical keypad opens
allowing the entry or modification of the concentration value of the undiluted
Standard used. The Input range depends on the definition set in chapter
5.4.5.1 Edit Calibration for Clotting (manual calibration), Convers.
range.

Name By pressing into the Name input field an alphanumeric keypad opens
allowing the entry or modification of the name of the Standard used. For
Name entry 15 alphanumeric characters are available.

Lot No. By pressing into the Lot No. input field an alphanumeric keypad opens
allowing the entry or modification of the name of the Standard used. For Lot
No. entry 15 alphanumeric characters are available.

CoaLyser – Operators Manual 1.3 Page 166


SETUP

Tools
The Tools button in the Calibration Value Screen offers several options and
access to other menus and actions.

 Press the Tools button to access further additional areas.

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

By pressing this button the entered calibration


curve points will be displayed as graph, see
Graph chapter 5.4.5.1 Edit Calibration for Clotting
(manual calibration), Display Calibration
Graph.
By pressing this button the calibration value
Hardcopy screen will be printed, provided a printer is
connected to the analyser.

Display Calibration Graph Press the Graph button to open the graph for the entered calibration curve.
Contents of the boxes that appear on the screen vary depending on the
units selected in the Calibration Menu.

Figure 200 Calibration Curve Graph

A graph with all entered curve points and calibration date & time information
appears.

Tools
The Tools button in the Calibration Graph Screen offers additional options
allowing the graph to be modified, printed and displayed in other scaling
rates.

 Press the Tools button to access further additional areas the following
Select Window appears.

CoaLyser – Operators Manual 1.3 Page 167


SETUP

Figure 201 Tools in Calibration Graph Screen

The following Tools are available:

Available Tools Description


Increases the scale of the displayed graph. By
Zoom IN pressing into the curve screen and holding you
may move the curve into all directions.
Zoom OUT
(appears only when graph Reduces the scale of the displayed graph.
was zoomed in first)

Resets the scale of the curve diagram to the


Zoom RESET
initial setting.

Flip Interchanges the x-/y-axis definition

By pressing this button you will leave this


Quit
window without any changes.

To print the calibration curve graph as currently


Hardcopy
shown in the screen.

OK Press the OK button twice to return back to the Calibration Menu Main
Screen.

ESC&Exit By pressing the ESC&Exit button you will leave the Calibration Menu
without saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Calibration Menu confirms all changes
and modifications done in this menu. Prior of acceptance the software will
check conformity on all edited parameters. In case of any invalid inputs the
software will indicate the corrections required.

CoaLyser – Operators Manual 1.3 Page 168


SETUP

5.4.5.2 Edit Calibration for Kinetic / Diff. /


1st deri (manual calibration)

Figure 202 Calibration Menu Main Page

 Select Kinetic/Diff./1st deri. by using the arrow keys to enter calibration


parameters for tests not using seconds as raw values or to define
parameters for a test using 2nd evaluation.

Basically this section is structured in the same way as already described in


chapter 5.4.5.1 Edit Calibration for Clotting (manual calibration). For
further entry of calibration parameters based on the 2nd evaluation only the
differences will be outlined in the following.

Calculation Press the Calculation button to access the Calibration Menu for Clotting,
the following window appears. Contents of the boxes that appear on the
screen vary depending on the units selected in the test.

Figure 203 Calibration Menu Kin/Diff/1st deri

CoaLyser – Operators Manual 1.3 Page 169


SETUP

1st conversion By pressing the Select Box arrow button you can select between 2 available
conversion modes to define the 1st conversion.

The following 1st conversion modes are available:

Conversion modes Description

No conversion of raw values, results will be shown


in mE (milliExtinction) only. By selecting none, no
None
further entries on this page are required. No
calibration curve is measured.
Allows conversion of raw values into
units/activities. By selecting this conversion mode
Interpolation further entries are required to parameterize the
test and calibration curve. Values are interpolated
from point to point.
Allows conversion of raw values into
units/activities. By selecting this conversion mode
Regression further entries are required to parameterize the
test and calibration curve. Regression is the best
fit over all measured calibration points.

Unit conversion By pressing into the Unit conversion input field a alphanumerical keypad
opens allowing the entry or modification of the unit/activity for conversion of
the raw value, meaning from mE into unit/activity, e.g. mg/dl, mg/l.

Mode in/out By pressing the Select Box arrow button of the left select box you can define
the scaling of the x-axis of the calibration curve graph whereas the right
select box will define the y-axis setting of the calibration curve graph. The
following selections are available for both axis:

 log - logarithmic calculations are applied to values before graphing.


 rec - the reciprocal (1 divided by the value) is used on the graph.
 lin - Linear value is used directly on the graph.

The mode in/out selection lin/lin is the most common setting used. Other
combinations, e.g. log/rec may require certain modifications in the
conversion range values, see Conversion range input field to display the
calibration curve graph.

Output format By pressing the Select Box arrow button you can define the decimal point
setting for the measuring result display. There are 4 alternatives available:

 <xxx.x> Input range: 0.1 - 999.9 (units)


 <xx.xx.> Input range: 0.01 - 99.99 (units)
 <x.xxx> Input range: 0.001 - 9.999 (units)
 <xxxx.> Input range: 0 - 9999 (units)

Rising By using this check box button you may define or modify whether the
expected converted values (unit/activity) of the measuring results are rising
or falling. Set individual parameters On/Off by pressing onto the field, e.g. ON
=  (converted values rising).

Calibration curves must be entered in such a way that you start with the
highest converted values and a highest measuring time (raw value) in mE
for falling curves. For rising curves you have to start with the lowest
converted values and the highest measuring times (raw value) in mE. The
software will support this procedure only.

CoaLyser – Operators Manual 1.3 Page 170


SETUP

The arrow marking in the Calib Value page indicate whether the
NOTE values must rise or fall. Converted values can be set to rising or
falling. Raw values must always rise.

Convers. range By pressing into the Conversion range input field a numerical keypad
opens allowing the entry or modification of the extrapolated result limits
within the results will be calculated and displayed in the calibration curve
graph. Input range: 0.1 - 9999 (depending on defined output format).

Tools The Tools button in the Test Menu Show-Flags Screen offers additional
options and access to other menus and features.

The following Tools are available:

Available Tools Description


By pressing this button you will leave this
Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Calibration Menu Main Page.

Figure 204 Calibration Menu Main Page

Enter Calibration Values Press the Calib Values button to access the Calibration Menu for Kinetic,
Difference or 1st Derivation, the following window appears. Contents of the
boxes that appear on the screen vary depending on the units selected in the
test.

CoaLyser – Operators Manual 1.3 Page 171


SETUP

Figure 205 Calibration Values 2nd evaluation

To enter calibration curve points manually perform one of the following:

Option 1:

2. If available use the values as provided in the package insert of the


reagent manufacturer.

Option 2:

 Enter manually the evaluated results achieved by a normal


measurement with dilutions or if available with standard plasmas.

Prior entering the calibration curve values make sure the values correspond
to the following settings:

 Converted values entered must be inside the defined conversion


range, see chapter 5.4.5.2 Edit Calibration for Kinetic / Diff. / 1st deri
(manual calibration), Convers. range.

 Values entered must correspond to the rising/falling definition, see


chapter 5.4.5.2 Edit Calibration for Kinetic / Diff. / 1st deri (manual
calibration), Rising

Unit [mg/dl] By pressing into the Unit, here mg/dl input field a numerical keypad opens
allowing the entry or modification of the converted value (unit/activity).

Points on the calibration curve that are not to be used must have the
NOTE entry 0.0%. A minimum of 2 points must be defined. It is
recommended defining at least 3 points. In maximum 7 calibrations
curve points can be defined.

Raw Value [mE/min] By pressing into the Raw Value, here mE/min. input field a numerical
keypad opens allowing the entry or modification of the raw values in sec.

Dilution Within this section the dilution based on the entered values (converted and
raw values and the defined concentration value will be calculated
automatically by the software.

CoaLyser – Operators Manual 1.3 Page 172


SETUP

Example: 1+ 9.0 = 1 part standard plasma will be diluted with 9 parts


buffer or a 1:10 dilution.

invalid = The converted value entered deviates from the


following conditions:

1. Value is outside the defined conversion range, see chapter


5.4.5.2 Edit Calibration for Kinetic / Diff. / 1st deri,
Convers. range.

2. Value is higher as the set concentration value, see


Standard / Concentration in this chapter.

Standard This section allows entry of further information on Standard Plasmas used to
verify a calibration curve.

Concentration By pressing into the Concentration input field a numerical keypad opens
allowing the entry or modification of the concentration value of the undiluted
Standard used. The Input range depends on the definition set in chapter
5.4.5.2 Edit Calibration for Kinetic / Diff. / 1st deri, Convers. range.

Name By pressing into the Name input field an alphanumeric keypad opens
allowing the entry or modification of the name of the Standard used. For
Name entry 15 alphanumeric characters are available.

Lot No. By pressing into the Lot No. input field an alphanumeric keypad opens
allowing the entry or modification of the name of the Standard used. For Lot
No. entry 15 alphanumeric characters are available.

Tools
The Tools button in the Calibration Value Screen offers several options and
access to other menus and actions.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 206 Tools in Calibration Value Screen

CoaLyser – Operators Manual 1.3 Page 173


SETUP

The following Tools are available:

Available Tools Description

By pressing this button you will leave this window


Quit
without any changes.

By pressing this button the entered calibration curve


points will be displayed as graph, 5.4.5.1 Edit
Graph
Calibration for Clotting (manual calibration),
Display Calibration Graph.

By pressing this button the calibration value screen


Hardcopy will be printed, provided a printer is connected to the
analyser.

OK Press the OK button to return back to the Calibration Menu Main Screen.

ESC&Exit By pressing the ESC&Exit button you will leave the Calibration Menu
without saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Calibration Menu confirms all changes
and modifications done in this menu. Prior of acceptance the software will
check conformity on all edited parameters. In case of any invalid inputs the
software will indicate the corrections required.

CoaLyser – Operators Manual 1.3 Page 174


SETUP

5.4.5.3 Auto Calibration for Clotting


(automated calibration)

The following chapters will describe how to perform an automated


calibration on the system. The following procedures will be described using
Fibrinogen as an example. Perform all other tests in the same way.

 From the Main Menu, press the button Setup to access the Setup
Menu.

Figure 207 Setup Menu

 From the Setup Menu, press the button Calibration to access the
Calibration Menu.

Figure 208 Select test Calibration

The Select Test Screen for calibration contains a list of all tests available.

 Select a test for calibration, here Fibrinogen by using the arrow keys
first to select the test and then press the OK button to continue.

CoaLyser – Operators Manual 1.3 Page 175


SETUP

The calibration for test screen appears.

Figure 209 Calibration Menu Main Page

Chose mode of calibration First select between the 2 available calibration modes by pressing the Select
Box arrow button. The different modes define how the raw value of a
calibration curve are converted and further allows entry of calibration
parameters for tests requiring 2 different calibration curves, for example
derived Fibrinogen.

The following calibration modes are available:

Calibration modes Description

Converts raw values [sec.] into a defined


Clotting
unit/activity.

Kinetic: Converts raw value [mE/min.] into a


defined unit/activity.

Kinetic / Difference / Difference: Converts raw value [delta mE] into a


1st derivation defined unit/activity.

1st deri.: Converts raw value [mE/sec.] into a


defined unit/activity.

Before selecting the calibration mode in this screen the


NOTE corresponding evaluation mode must be assigned to the test first,
see chapter 5.4.4.2 Evaluation to assign the 1st and or 2nd
evaluation to a test also defining whether the test allows entry of 2
calibration curve parameters.

 For Fibrinogen select Clotting by using the arrow keys to enter


calibration parameters.

Before starting an Auto Calibration make sure the following has been done:

 The reagent supposed to be calibrated must be set to calibrate, see


chapter 5.4.3.1 Parameter Page 1, Calibrate.

 The reagents and diluents required for the test calibration must be
loaded on the analyzer and available in reagent area of the system. See
chapter 5.2.2 Load Reagents.

CoaLyser – Operators Manual 1.3 Page 176


SETUP

Calculation Press the Calculation button in Edit Calibration to access the Calibration
Menu for Clotting, the following window appears. Contents of the boxes that
appear on the screen vary depending on the units selected in the test.

Figure 210 Calibration Menu Clotting

1st conversion By pressing the Select Box arrow button you can select between 2 available
conversion modes to define the 1st conversion.

The following 1st conversion modes are available:

Conversion modes Description


No conversion of raw values, results will be shown
in seconds only. By selecting none, no further
None
entries on this page are required. No calibration
curve is measured
Allows conversion of raw values into
units/activities. By selecting this conversion mode
Interpolation further entries are required to parameterize the
test and calibration curve. Values are interpolated
from point to point.
Allows conversion of raw values into
units/activities. By selecting this conversion mode
Regression further entries are required to parameterize the
test and calibration curve. Regression is the best
fit over all measured calibration points.

Unit conversion By pressing into the Unit conversion input field a alphanumerical keypad
opens allowing the entry or modification of the unit/activity for conversion of
the raw value, meaning from sec. into unit/activity, e.g. %, mg/dl, g/l.

Mode in/out By pressing the Select Box arrow button of the left select box you can define
the scaling of the x-axis of the calibration curve graph whereas the right
select box will define the y-axis setting of the calibration curve graph. The
following selections are available for both axis:

 log - logarithmic calculations are applied to values before graphing.


 rec - the reciprocal (1 divided by the value) is used on the graph.
 lin - Linear value is used directly on the graph.

CoaLyser – Operators Manual 1.3 Page 177


SETUP

The mode in/out selection lin/lin is the most common setting used. Other
combinations, e.g. log/rec may require certain modifications in the
conversion range values, see Conversion range input field to display the
calibration curve graph.

Output format By pressing the Select Box arrow button you can define the decimal point
setting for the measuring result display. There are 4 alternatives available:

 <xxx.x> Input range: 0.1 - 999.9 (units)


 <xx.xx.> Input range: 0.01 - 99.99 (units)
 <x.xxx> Input range: 0.001 - 9.999 (units)
 <xxxx.> Input range: 0 - 9999 (units)

Rising By using this check box button you may define or modify whether the
expected converted values (unit/activity) of the measuring results are rising
or falling. Set individual parameters On/Off by pressing onto the field, e.g. ON
=  dark blue (converted values rising).

Calibration curves must be entered in such a way that you start with the
highest % values and a shortest measuring time (raw value) in seconds for
falling curves (coagulation factors). For rising curves (inhibitors or INR-Cal)
you have to start with the lowest % / INR values and the shortest measuring
times (raw value) in seconds. The software will support this procedure only.

The arrow marking in the Calib Value page indicate whether the
NOTE values must rise or fall. Converted values can be set to rising or
falling. Raw values must always rise.

Convers. range By pressing into the Conversion range input field a numerical keypad
opens allowing the entry or modification of the extrapolated result limits
within the results will be calculated and displayed in the calibration curve
graph. Input range: 0.1 - 9999 (depending on defined output format).

2nd conversion By pressing the Select Box arrow button you can select between 3 available
conversion modes to define the 2nd conversion.

The following 2nd conversion modes are available:

Conversion modes Description


nd
No 2 conversion of raw values, results will be
shown in sec. and defined activity only. By
None
selecting none, no further entries on this page are
required.
Allows conversion of the INR (International
Normalized Ratio). Selecting this conversion
mode requires further entry of the ISI
INR
(International Sensitivity Index) and the MNPT
(mean normal PT) as provided by the reagent
package insert.

Allows conversion of the Ratio (Sample / MNPT =


Ratio
mean normal PT or aPTT).

For Fibrinogen a 2nd conversion is not required and should be set to none.

CoaLyser – Operators Manual 1.3 Page 178


SETUP

Tools The Tools button in the Test Menu Show-Flags Screen offers additional
options and access to other menus and features.

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK Press the OK button to return back to the Calibration Menu Main Page.

Figure 211 Calibration Menu Main Page

Enter Calibration Values Press the Calib Values button in Edit Calibration to access the Calibration
Menu for Clotting, the following window appears. Contents of the boxes that
appear on the screen vary depending on the units selected in the test.

Figure 212 Calibration Values (concentration)

CoaLyser – Operators Manual 1.3 Page 179


SETUP

Within this window certain entries are required manually.

Unit [g/l] First enter the concentration/converted values steps for the standard used.
By pressing into the Unit, here g/l input field a numerical keypad opens
allowing the entry or modification of the converted value (unit/activity).

Points on the calibration curve that are not to be used must have the
NOTE entry 0.0%. A minimum of 2 points must be defined. It is
recommended defining at least 3 points. In maximum 7 calibrations
curve points can be defined.

Prior entering the calibration curve values make sure the values correspond
to the following settings:

 Converted values entered must be inside the defined conversion


range, see chapter 5.4.5.1 Edit Calibration for Clotting (manual
calibration), Convers. range

 Values entered must correspond to the rising/falling definition, see


chapter 5.4.5.1 Edit Calibration for Clotting (manual calibration),
Rising

Raw Value [sec.] When performing an automated calibration the entry of raw values is not
required. There may be raw values available already which have been
calculated from the last calibration.

Dilution Within this section the dilution based on the entered values (converted and
raw values and the defined concentration value will be calculated
automatically by the software.

Example: 1+ 9.0 = 1 part standard plasma will be diluted with 9 parts


buffer or a 1:10 dilution.

invalid = The converted value entered deviates from the following


conditions:

1. Value is outside the defined conversion range, see chapter


5.4.5.1 Edit Calibration for Clotting (manual calibration),
Convers. range

2. Value is higher as the set concentration value, see Standard /


Concentration in this chapter.

Standard This section allows entry of further information on Standard Plasmas used to
verify a calibration curve.

Concentration By pressing into the Concentration input field a numerical keypad opens
allowing the entry or modification of the concentration value of the undiluted
Standard used. The Input depends on the definition set in chapter 5.4.4.1
Liquids, Buffer/Diluent., please refer to explanation as provided below.

CoaLyser – Operators Manual 1.3 Page 180


SETUP

The concentration value entered for automated calibration depends


NOTE on the entry of the dilution ratio as defined in the Test Setup for the
particular test, see chapter 5.4.4.1 Liquids. The dilution value
Buffer/Diluent as defined in Test Setup/Liquids must be multiplied
with the value of concentration required for this test.

Example:
Test Setup/Liquids is set to a 1+9 dilution for Fibrinogen.
The concentration value required is 2,8 g/l.

Multiply 2.8g/l with the dilution ratio of 1+9, i.e. 10 = 28.00 g/l

Enter 28.00 g/l into the concentration field by using the numerical
keypad. This is important to ensure correct measurement of your
calibration values in order to avoid invalid dilutions.

The lowest pipetting volume is 2µl that means when having a


NOTE sample volume of for example 100µl the max. dilution may not
exceed 1+49 ( 2µl + 98µl).

Name By pressing into the Name input field an alphanumeric keypad opens
allowing the entry or modification of the name of the Standard used. For
Name entry 15 alphanumeric characters are available.

Lot No. By pressing into the Lot No. input field an alphanumeric keypad opens
allowing the entry or modification of the name of the Standard used. For Lot
No. entry 15 alphanumeric characters are available.

OK Once these data have been entered leave this menu by pressing the OK
button to return back to the Calibration Menu Main Screen.

Figure 213 Calibration Menu Main Page

CoaLyser – Operators Manual 1.3 Page 181


SETUP

Auto Calibration In section Auto Calibration the following settings can be done prior to finally
start the automated calibration.

Diluent A diluent can be selected by pressing onto the button. The selection of a
diluent may only be possible if a reagent is used for calibration which does
normally not require a buffer or a sample does not require dilution, for
example PT.

Tests to which a buffer is assigned to already do not require the selection of


a diluent in this case the button will be displayed grey. The reagent buffer is
used s diluent for calibration.

Cup Select the type of cup which shall be assigned to the control & calibrator
position by pressing the assigned radio button. Control & calibrator cups
may be placed in the controls/calibrator positions (C1-C8), see below
picture. During the auto calibration run the software will provide information
on what position must be used.

Figure 214 Control & Calibrator Positions C1-C8

No. of Determination By pressing into the No. of Determ. input field a numerical keypad opens
allowing the entry or modification of the number of determination (single or
multiple) assigned to a test for calibration.
Input range: 1 - 2 (Default = 1).

Is the number of determination set to multiple, meaning > 1 the maximum


deviation must be defined to specify to what extent each single value may
deviate, see Max. Deviation.

Make sure the number of measurements for the automated


NOTE
calibration always corresponds to the available number of cuvettes
available in the cuvette ring used for this run. Auto calibration runs
can not be divided into two cuvette rings.

Max. Deviation By pressing into the Max. Deviation (CV) input field a numerical keypad
opens allowing the entry or modification of the maximum deviation in %. The
maximum deviation indicates the permitted deviation of the measured
values for multiple determinations. This field only appears if the No. of
Determination is set to > 1. Input range: 1 - 20 (%).

Once the single values are outside of this range the single values will be
marked by the software.

CoaLyser – Operators Manual 1.3 Page 182


SETUP

Start After all data has been entered and all reagents/diluents have been placed
on the system he auto calibration can be started.

Press the START button to perform the automated calibration run and the
following Info Message box appears.

Below message appears. Place the control/calibrator cup into position C1


within the reagent area of the analyzer and confirm message by pressing the
OK button.

Figure 215 Place Calibrator Position 1

If an assigned test buffer is used as diluent a corresponding message


appears for a moment.

If the diluent must be provided separately, a message appears requesting to


place a calibrator in position ‘control 2’.

After confirming this message by confirming this message with OK, the level
sensor checks if liquids are available.

Volume quantity is not calculated for control position 1 and 2. Make


NOTE
sure enough calibrator and diluent is available.

The following window provides information related to the current available


joblist.

Figure 216 Reagent Status List Calibration

CoaLyser – Operators Manual 1.3 Page 183


SETUP

The following information is available:

Available information Description

Provides information on the loaded reagents


Reagent Status List and washers, its required volumes, loaded
volumes, stability and lot numbers.

Provides information on the cuvette ring loaded:


- Start Cuvette No.
Cuvettes
- Required cuvettes for the run
- Free cuvettes available on the loaded ring

Tools The Tools button in the Measurement Menu offers several options and
access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 217 Tools in Calibration Measurement Menu

The following Tools are available:

Available Tools Description

By pressing this button the screen will be printed,


Hardcopy
provided a printer is connected to the analyser.
By pressing this button you will leave this window
Quit
without any changes.
Opens the timeline window showing each step of
Timeline the joblist entered in a diagram and provides run
time information.

Exit Select Exit to return to the MAIN MENU.

Start the Run Select Start the Run to begin measurement.

CoaLyser – Operators Manual 1.3 Page 184


SETUP

The analyser starts preparing the measurement and starts the run if there is
no pending problem. Results and/or related run information are shown in this
window once available.

Figure 218 Running Measurement Calibration

Tools The Tools button in Running Measurement offers several options and
access to other menus and features.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 219 Tools in Calibration Measurement Menu

The following Tools are available:

Available Tools Description

By pressing this button you enter into the result


Results
list.

By pressing this button you will print the screen,


Print
provided a printer is connected to the analyser.

By pressing this button you will leave this


Quit
window without any changes.

CoaLyser – Operators Manual 1.3 Page 185


SETUP

By pressing this button you will see the reaction


R-Curve
curve graph of the current run in real time.

Opens the timeline window showing each step


Timeline
of the joblist entered in a diagram.

Show/print reaction curves To display reaction curves of the current auto calibration run press the
button R-Curve, the following window appears:

Figure 220 Reaction Curves Calibration

 From the R-Curve window press the OK button to return to the Running
Measurement screen or press Hardcopy to print the current screen.

Figure 221 Running Measurement Calibration

Once the auto calibration run is finished all available results will be displayed
in the Results of last run list, the following display appears:

CoaLyser – Operators Manual 1.3 Page 186


SETUP

Figure 222 Results/Run List Calibration

In the Results of last run list all results of the last auto calibration run will be
displayed as follows:

 If double determination has been selected prior to the auto


calibration run the results listed are the mean values from both
determinations.

 When single determination was selected the results for each


measurement will be displayed.

 Results which have a coloured background indicate a problem


occurred during measurement, for example a deviation error. Such
results marked as “none” in the list will be entered into the calibration
curve raw value data as 0.0 sec.

Print By pressing the Print button the Results of last run list will be printed on your
external printer. Once these data have been entered leave this menu by
pressing the OK button to return back to the Calibration Menu Main Screen.

OK Leave this screen by pressing the OK button to return back to the


Calibration Menu Main Screen.

Figure 223 Warning Message Calibration

CoaLyser – Operators Manual 1.3 Page 187


SETUP

The following message appears. If you intend to overwrite the current


calibration values with the latest values measured press the Yes button. To
keep the current calibration values press the No button.

The following message appears in case any problems occurred during the
last calibration run. See chapter 6.2.2 Error Flags in the Results/Run or
Results/Memory List to identify error codes in this screen.

Figure 224 Warning Message Measuring Status

Confirm this message by pressing the OK button, the following display


appears:

Figure 225 Auto Calibration finished

The auto calibration run is finished now, all measured data have been added
to the calibration curve/data. Confirm this message by pressing the OK
button, the following display appears:

CoaLyser – Operators Manual 1.3 Page 188


SETUP

Figure 226 Calibration Menu Main Page

Calib. Values To verify and view the calibration data and curve press the Calib Value
button. The following display appears:

Figure 227 Calibration data

In case no raw value [0.0 sec] could be measured for 0.60 g/l concentration as
an example due to low signals for this high dilution. There are 2 possibilities to
fix the problem:

a. Enter “0” into the g/l field by overwriting the “0.60” g/l entry. A “0” in the
concentration will not be used for calibration curve data

b. Enter a value greater than 20.3 [sec] for this calibration step.

CoaLyser – Operators Manual 1.3 Page 189


SETUP

Show calibration curve By pressing the Tools button followed by the Graph button the latest
calibration curve will be displayed.

Figure 228 Calibration Curve

Tools
The Tools button in the Calibration Graph Screen offers additional options
allowing the graph to be modified, printed and displayed in other scaling
rates.

 Press the Tools button to access further additional areas the following
Select Window appears.

Figure 229 Tools in Calibration Graph Screen

CoaLyser – Operators Manual 1.3 Page 190


SETUP

The following Tools are available:

Available Tools Description


Increases the scale of the displayed graph. By
Zoom IN pressing into the curve screen and holding you
may move the curve into all directions.
Zoom OUT
(appears only when graph Reduces the scale of the displayed graph.
was zoomed in first)

Resets the scale of the curve diagram to the


Zoom RESET
initial setting.

Flip Interchanges the x-/y-axis definition

By pressing this button you will leave this


Quit
window without any changes.

To print the calibration curve graph as currently


Hardcopy
shown in the screen.

OK Press the OK button twice to return back to the Calibration Menu Main
Screen.

Print By pressing the Print button the complete calibration data, results and curve
information will be printed.

ESC&Exit By pressing the ESC&Exit button you will leave the Calibration Menu
without saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Calibration Menu confirms all changes
and modifications done in this menu. Prior of acceptance the software will
check conformity on all edited parameters. In case of any invalid inputs the
software will indicate the corrections required.

CoaLyser – Operators Manual 1.3 Page 191


SETUP

5.4.6 Delete Results


The button Delete Results in the Setup Menu allows you to clear the result
ring memory list.

By confirming the delete results message with Yes will delete all
NOTE available results from the ring result memory of the analyser. There
will be no option to recover these results once deleted.

Figure 230 Delete Results

 Press the Delete Results button, the following message appears.

Figure 231 Delete Results Message Box

Press the Yes button if you intent to delete all results from the result ring
memory list. By pressing No all available results will be kept safe in the
result list.

CoaLyser – Operators Manual 1.3 Page 192


SETUP

5.4.7 Setup Language


The button Language in the Setup Menu allows you to install further
software languages or to set the software language to factory default
(English).

Figure 232 Setup Language

 Press the Language button, the following message appears.

Figure 233 Select Language

The Select Language Setup screen contains a list of software languages


available.

Select a language using the arrow keys first to select the language and then
press the OK button to continue.

Pressing the ESC buttom will leave the screen without any changes to the
previous settings.

CoaLyser – Operators Manual 1.3 Page 193


SETUP

5.4.8 Setup Date & Time


The button Date & Time in the Setup Menu allows you to set the actual date
& time information of the analyser.

Figure 234 Setup Date & Time

 Press the Date & Time button, the following message appears.

Figure 235 Date & Time Settings

 By pressing into the Date input field a numerical keypad opens allowing
the entry or modification of the date. Input format: dd.mm.yyyy

 By pressing into the Time input field a numerical keypad opens


allowing the entry or modification of the date. Input format: hh:mm:ss

ESC&Exit By pressing the ESC&Exit button you will leave the Date & Time Menu
without saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Date & Time Menu confirms changes
and modifications done in this menu. . The system will re-start to set the new
entry of Date & Time.

CoaLyser – Operators Manual 1.3 Page 194


SETUP

5.4.9 Change Password


The button Passwords in the Setup Menu allows you to set or change
certain PIN Codes (5 digits) required to access different levels of the
software.

The Passwords Menu can only be entered in the software level expert, see
chapter 3.2 Software.

Figure 236 Change Password

 Press the Passwords button, the following message appears.

Figure 237 Setup Password

 By pressing into the Expert input field a numerical keypad opens


allowing the entry or modification of the current PIN Code for the expert
level. Input format: 0 - 99999.

Tools The Tools button in the Password Menu screen offers additional options
and access to other menus and features.

CoaLyser – Operators Manual 1.3 Page 195


SETUP

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

ESC&Exit By pressing the ESC&Exit button you will leave the Password Menu without
saving the previous done changes and modifications.

Save&Exit Pressing the Save&Exit button in the Password Menu confirms changes
and modifications done in this menu.

5.4.10 Setup Beeper


The check box button Beeper in the Setup Menu allows you to activate
acoustical response performed by the analyser software in case of warning
messages or problems occur.

Figure 238 Setup Beeper

 By using this check box button you may define whether the beeper shall
be active or set inactive. Set individual parameters On/Off by pressing
onto the field, e.g. ON =  (beeper active).

CoaLyser – Operators Manual 1.3 Page 196


SETUP

5.4.11 Setup STAT Autoprint


The check box button STAT Autoprint in the Setup Menu allows you to
activate an automated result print function when STAT results are available.

Figure 239 Setup STAT autoprint

 By using this check box button you may define whether the results of a
STAT measurement will be automatically printed once available. Set
individual parameters On/Off by pressing onto the field, e.g. e.g. ON = 
(STAT autoprint active).

5.4.12 Define Headline


The input field Headline in the Setup Menu allows you to enter a head line
title used on every printout performed by the analyser. The headline may be
used to define the name of the analyser or to identify the respective
department the analyser is used.

Figure 240 Define Headline

 By pressing into the headline input field an alphanumeric keypad opens


allowing the entry or modification of the headline. For headline entry 15
alphanumeric characters are available.

CoaLyser – Operators Manual 1.3 Page 197


SETUP

5.4.13 Initiate Analyser Selftest


The button Selftest in the Setup Menu allows you to check the general
analyser functions at any time. This system test will initiate a check up of all
positions, pumps and level detection to determine any failure possibly
existing.

Figure 241 Selftest

 Press the Selftest button to start the self test sequence of the analyser.
At first the photometer transmissions will be checked.

Figure 242 Selftest Transmission

 Confirm this message by pressing the Yes button. Now the analyser
starts checking all available positions, modules and guides you through
a full selftest procedure. Any failures will be outlined and require
confirmation. Each successful check must be confirmed by pressing the
OK button. After successful testing procedure the following message
appears:

CoaLyser – Operators Manual 1.3 Page 198


SETUP

Figure 243 Selftest finished

 After successful finalization of the Selftest sequence confirm this


message by pressing the OK button to return back to the Setup Menu
Main Screen.

Figure 244 Setup Menu

Tools The Tools button in the Setup Menu offers additional options and access to
other menus and features.

The following Tools are available:

Available Tools Description

By pressing this button you will leave this


Quit
window without any changes.

Hardcopy Performs a print of the screen only.

OK By pressing the OK button you will return to the Main Menu.

CoaLyser – Operators Manual 1.3 Page 199


SETUP

5.5 Switch off analyser

To switch off the analyser follow the next steps as described in below
section.

 From the Main Menu, press the button Shutdown.

Figure 245 Shutdown Analyzer

The following selection appears:

Figure 246 Shutdown Selection

The following Selection is available:

Available Tools Description


By pressing this button you will leave this
Quit window without any changes. An analyzer info
window will be displayed shortly.
By pressing this button you will shut down the
Switch Off software. Switch off mains to finally turning off
analyzer.
By pressing this button the analyzer will initiate
Restart
a system restart.

CoaLyser – Operators Manual 1.3 Page 200


SAFETY IN OPERATION

6 SAFETY IN OPERATION

6.1 Maintenance and Hygiene

Refer to this chapter for the recommended maintenance schedule and


procedures for maintaining the analyzer.

To obtain accurate and precise results the manufacturer recommends


cleaning and maintaining the system on a regular basis. Disregarding these
instructions may cause contamination and operational failures of the system
and may also lead to the manufacturer’s warranty becoming null and void.

Carefully observe the following cleaning and hygiene instructions:

Do not use organic acids for cleaning the instrument. Always use the
surface cleaning agents recommended by the specialist trade for this
purpose. Always use a moist cloth; do not spray on or pour on any liquids,
as they can affect the correct function of the instrument or damage it.
Caution!
Keep the instrument free from dust and liquid spillages. When not used for a
prolonged period, place a dust cover over the instrument or place it in a
cabinet.

If liquid has been spilled on the instrument, remove the contamination with a
clean, absorbent non-woven cloth considering all applicable hygienic
requirements.

If liquid has accidentally entered, or been pipetted into a measuring


channel, remove the liquid with a pipette and then clean the measuring
channel with a lint-free cloth. Please consider all applicable hygienic
requirements.

Contact our technical service if subsequent control measurements do not


produce the expected result.

The Analyser is fitted with a lithium battery type Li-Mn CR 2430 (life
approx. 5 years). It should be replaced by an authorised service operator
after 5 years at the latest. Otherwise a faultless operation can not be
guaranteed.

Before performing any task in this chapter, review the hazard and
NOTE precautions listed in chapter 2 HAZARD AND PRECAUTIONS of this
manual.

Power down the analyzer and disconnect the power cord from the facility
NOTE power for all cleaning tasks.

CoaLyser – Operators Manual 1.3 Page 201


SAFETY IN OPERATION

6.1.1 Maintenance frequency

Step Cycle Service needed


daily weekly monthly yearly yes no

Clean working plate X X

Run a flush cycle X X

Run a washing cycle using Washing


X X
Solution (Intensive 1)
Clean waste water tubing using Cleaning
X X
Solution
Fill wash station with Cleaning Solution,
X X
wait 2 minutes & run flush cycle

Clean pipettor tip X X

Clean distilled Water Container and


2x X
sensor (70% alcohol solution)

Renew DI water in distilled water container X X

Clean Waste Water Container and


X X
sensor (70% alcohol solution)

Clean the housing (70% alcohol solution) X X

Replace distilled + waste water container X X

Replace fresh water filter at distilled water


X X
sensor (green)

Replace 250 µl syringe X X

Replace DI water tubing (Maintenance Kit) X X

Check/replace sample positions insert in


X X
working plate incl. single insert segments
Reverse flush of Pipettor with hot distilled
X X
water
Recalibration of photometer (automated by
X X
service software)

Check/Replace pipettor (recommended) X X

Check/Replace dilutor valve


X X
(recommended)
Optical inspection, cleaning & lubrication of
X X
X/Y/Z cogwheel and toothed rack
Check of Measuring rotor and Pipettor
X X
temperatures

Check motors and pumps for noise X X

Check of USB-port connectors for printer


X X
and USB-stick for wear out / function

CoaLyser – Operators Manual 1.3 Page 202


SAFETY IN OPERATION

6.1.2 Cleaning of the housing


The housing and working plate of the analyzer should be cleaned in regular
intervals to avoid any contamination.

NOTE Do not use organic acids, bleach, Sodium Hypocloride or


Glutaraldehyde.

Only use the Cleaner according to the instructions provided or 70% alcohol
for cleaning. Use a dampened cloth. Never spray or pour any liquids onto or
into the system which may negatively impact the analyzer’s functions.

If liquids have accidentally run into the system contact the customer service
of the distributor or the manufacturer.

Avoid any direct skin contact with samples and/or test reagents as
well as those parts of the instrument which might be contaminated
NOTE
with samples and/or test reagents.
These parts are potentially infectious. If samples were spilled onto
the system wipe them off immediately and disinfect the
instrument. Reagents might irritate the skin or mucous
membranes. Observe the manufacturer’s instructions in the
package insert how to handle the reagents properly.
Make sure to dispose of the sample and test material in compliance
with biohazard and legal provisions.

6.1.3 Clean Pipettor


The pipettor must be free of damages and deposits. To clean the end of the
pipettor:

 Shut down software.

 Power down the analyzer and remove the power cord.

 Open the safety shield.

 Clean the bottom end of the pipettor (steel tubing) with a lint-free cloth
(or tissue) soaked with a 70% alcohol solution. Use care to avoid
bending of the probe.

NOTE Never use bleach to clean the pipettor.

CoaLyser – Operators Manual 1.3 Page 203


SAFETY IN OPERATION

6.1.4 Clean Containers/Sensors


Replace Filter
The containers and level sensors must be cleaned in regular intervals to
avoid contamination and the risk of inaccuracy of the measuring results.
Additionally it is recommended to replace the white filter at the end of the
distilled level sensor (green) on a monthly basis.

To clean the containers/sensors and replace the filter:

 Shut down software.

 Power down the analyzer and remove the power cord.

 Remove the distilled water (green) and waste water (red) level sensors
from the containers and disconnect them from the analyzer, see chapter
3.1.7 Rinsing System and 4.3.2 Connect the Rinsing System

 Remove the white fresh water filter from the end of the Distilled Water
Sensor.

 Clean both sensors and the end of the sensors using a 70% alcohol
solution and rinse them with distilled water to remove any traces of the
alcohol solution.

 Place a new fresh water filter at the end of the distilled water sensor
after cleaning.

 Empty the distilled water and waste water container and clean the inside
using a 70% alcohol solution.

 Rinse both containers thoroughly with distilled water to remove any


traces of the alcohol solution.

 Refill the distilled water container using fresh distilled water. The
container should have a minimum of 1 litre. Make sure no air bubbles
remain in the container. Use Distilled Water (aqua dest.) only!

 Connect the level sensors back to the analyser, see chapter 4.3.2
Connect the Rinsing System
 and place back the sensors again into the corresponding containers.

6.1.5 Perform cleaning cycle


It is necessary to clean the liquid system, tubing’s at regular intervals to
avoid reagent or sample residues, possibly causing inaccurate results or
improper functioning of the analyzer. Therefore, regular cleaning cycles
must be run to guarantee proper functioning of the system.

Carefully read the package insert supplied within the Cleaner box.
NOTE Also take note to the information as provided in the package insert to
ensure proper use and handling of the cleaner.

To clean the liquid system:

CoaLyser – Operators Manual 1.3 Page 204


SAFETY IN OPERATION

 Place a 12ml bottle of Cleaner (supplied materials) into the reagent area
position W2, see chapter 3.1.2 Working Plate. Make sure to remove
the lid prior of placing the bottle into its position.

 From the Main Menu, press the button Run Preparation the following
display appears.

Figure 247 Perform Cleaning Cycle

 Use the Select Box arrow buttons in the “Rinsing Modes” frame until
cleaning cycle appears in the select box, see above picture.

 By pressing the button Perform the cleaning procedure starts. The


analyser will rinse the system using the Cleaner provided.

Cleaning procedure:
- 1 sec. rinsing with system water
- 200µl Cleaner
- 3 strokes up to 100µl
- 20 sec. rinsing with system water

 This procedure should be repeated at least 3 times until the tubing’s


and liquid system is completely cleaned.

 After final cleaning cycle perform another flush cycle. Use the Select
Box arrow buttons in the “Rinsing Modes” frame until Flush 10s
appears in the select box.

 By pressing the button Perform the flush cycle starts. The analyser will
rinse the system with system water (distilled water). Repeat this
procedure 3 times to ensure no Cleaner remains in the liquid system.

If necessary repeat the flush procedure until no foam residues of


NOTE Cleaner will remain in the liquid system, waste tubing (red marked).
Check if waste water runs through the waste tubing into the Waste
Water Container and no foam remains in the waste tubing.

After finalization of the Cleaning cycle make sure to remove the


NOTE Cleaner bottle again from position W2 in the reagent area.

CoaLyser – Operators Manual 1.3 Page 205


SAFETY IN OPERATION

6.1.6 Perform washing cycle


Especially after the use of reagents containing thrombin, like Fibrinogen it is
recommended to perform washing cycles to avoid carry over.

To perform a washing cycle:

 Place a 15ml bottle of Washing Solution (supplied materials) into the


reagent area position W1, see chapter 3.1.2 Working Plate. Make sure
to remove the lid prior of placing the bottle into its position.

 From the Main Menu, press the button Run Preparation the following
display appears.

Figure 248 Perform Washing Cycle

 Use the Select Box arrow buttons in the “Rinsing Modes” frame until
Intensive 1 appears in the select box, see above picture.

 By pressing the button Perform the washing procedure starts. The


analyser will rinse the system with washing solution.

Washing procedure:
- 3 sec. rinsing with system water
- 150µl Washing Solution (W1)
- 3 strokes up to 250µl
- 3 sec. rinsing with system water

 This procedure should be repeated 2 times until the pipettor is


disinfected.

 After final washing cycle perform another flush cycle. Use the Select
Box arrow buttons in the “Rinsing Modes” frame until Flush 10s
appears in the select box.

 By pressing the button Perform the flush cycle starts. The analyser will
rinse the system with system water (distilled water). Repeat this
procedure 2 times to ensure no washing solution remains in the liquid
system.

CoaLyser – Operators Manual 1.3 Page 206


SAFETY IN OPERATION

6.1.7 Protocol of Maintenance Check


The manufacturer recommends recording all maintenance check-ups of the
system in which all service, maintenance and repair, replacement of parts,
cleaning and disinfection cycles will be documented.

If the customer service has to repair or replace parts on the system you
should make the maintenance check protocols available. This might simplify
the repair.

6.1.8 How to dispose of used


disposables
Remember to dispose of used cuvette rings, primary/secondary tubes, and
reagent vials in compliance with legal provisions.

All disposables used with the system must be treated as


NOTE
potentially infectious. Avoid any skin contact with sample and
reagent material which might be on the disposables.
Dispose of used disposables in compliance with legal provisions or
laboratory regulations. Do not put used disposables, samples and
reagents into the usual garbage.

Cuvette rings might not be resistant to organic acids. Therefore do


not use organic acids unless they are expressly permitted.

6.1.9 Disposal of the Analyser

The following must be considered when disposing of the Analyser:

 The upper and lower part of the housing is made of polyurethane foam.

 Mechanical parts are predominantly made of aluminium and precious


metals.

 Electronic parts must be disposed of in accordance with national


directives for the disposal of electrical parts.

 For the safety of the operating personnel make sure the Analyser has
been disinfected before disposal.

CoaLyser – Operators Manual 1.3 Page 207


SAFETY IN OPERATION

6.2 Troubleshooting

Faults may caused by actions during use and/or by the system. The
analyzer displays error messages and warnings in the display and outputs
these via the printer, provided an external printer is connected.

Additionally irregularities and errors may appear in the Status

6.2.1 Application errors


Application errors may cause incorrect results. Possible causes are:

 Air bubbles are in the cuvette.

 The temperature of the start reagent deviates from 37°C.

 The reagent has been placed incorrectly.

 The sample or control is too old.

 No stir bar has been placed into the cuvette.

 Reagents have been carried over (PT or Fibrinogen reagent).

 A reagent with the wrong lot number has been used.

 The reagent has not been used according to the package insert.

 The reagent used does not correspond to the test selected.

 No or an incorrect calibration curve is applied to the reagent lot.

 Errors appeared during the sample collection or centrifugation.

 No stirrer is placed into the reagent vial.

 Method parameters relevant for measuring are incorrect.

 Reagent vial size placed in the reagent area does not correspond to the
reagent vial parameterized in Load Reagent.

 Sample vial size placed in the sample area does not correspond to the
sample vial parameterized in Load Samples.

Should any of these errors occur and they are recognized in time, they must
be remedied immediately.

Certain of these errors can only be recognized when determining control


plasmas.

As a result we recommend running control plasma on a daily basis prior to


running routine determinations.

CoaLyser – Operators Manual 1.3 Page 208


SAFETY IN OPERATION

6.2.2 Error Flags in the Results/Run or Results/Memory List

Error Cause Action


State: OK The measurement has been finished with None
no problems.

none No clot has been found within the set Check test settings, see chapter
measuring time and valid clotting range. 5.4.4.2, valid clotting, auto repeat
and measuring time

State: D Deviation Error: Check reagent and test settings


The specified mean value deviation set in are correct, widen max. deviation
the Test Setup is exceeded. range. Repeat test.

State: T Temperature Error: Call customer service


Temperature problems occurred at pipettor
or rotor

State: M Mixer Error: Call customer service


The mixer motor(s) do not properly work.

State: S Reagent Mixer Error: Call customer service


The reagent mixer motor does not properly
work.

State: F Photometer Error: Call customer service


A photometer error was detected during the
measurement.

State: L Liquid Error: Check liquid levels of all used


Not enough reagent or sample available liquids and refill if necessary.
during the run.

State: Q QC Error: Run QC measurements with test


(requires QC-Software installed) A test under Quality Control must be using the assigned control
measured within the last 24 hours using the plasmas prior starting a routine
assigned control plasma before a routine run. Ensure test is verified and
measurement can be done. Additionally this not in violation to the Westgard
test may not be in violation to the Westgard- rules.
Rules. The Q Error will appear in case these
requirements are not fulfilled.

6.2.3 Errors during operation

Error Cause Action


No function shown after Power supply o.k.? Fuse o.k.? Check if correct power supplied,
switching on the device. Power cable connected correctly? check fuses

Analyzer fails during Power supply failure, fuse, power cable Check if correct power supplied,
operation loose? check fuses

There are air bubbles in the Analyzer was not filled with system water Check the screw connections of
pipettor tubing correctly. Dilutor tubing is not tight. the pipettor tube. Rinse with
system DI water, see chapter
5.1.3.

CoaLyser – Operators Manual 1.3 Page 209


SAFETY IN OPERATION

Error Cause Action


No free reagent position, The reagent area is fully loaded, no free Remove reagents from the
test could not be loaded. positions remain available to load more reagent area which may not be
reagents used for the next run.
No enough cuvettes The current joblist requires more cuvettes Cuvette ring carries 32 cuvettes.
as provided by the inserted cuvette ring. Run same number of samples
first and continue additional
samples using a new cuvette ring.
Test has no valid calibration Test applied for a run has no valid Enter calibration data for this test
calibration. first, see chapter 5.4.5

Reagent has no valid The stability of a reagent has timed out. Renew reagent stability, see
stability chapter 5.2.2, 5.4.3.1

Not enough liquids, out of The liquid volume required for a run does Refill the liquid missing according
liquid (reagent, washer, not correspond to the volume loaded. to the quantity required.
sample)

System-Water is EMPTY The dist. water container is empty Refill the system water container,
see chapter 4.3.2, 6.1.4

Waste-Water is FULL The waste water container is full Empty and clean the waste
container, see chapter 6.1.4

System-Water is N.C. System water plug (green level sensor) is Correctly connect the system
not connected water connector, see chapter
4.3.2
Waste-Water is N.C. Waste water plug (red level sensor) is not Correctly connect the waste water
connected connector, see chapter 4.3.2

Value out of range, reset to The value entered for a certain parameter is Value will be set to the max.
max. value. out of the maximum input range. allowed input value automatically,
check correctness of parameter.

Value out of range, reset to The value entered for a certain parameter is Value will be set to the min.
min. value. out of the minimum input range. allowed input value automatically,
check correctness of parameter.

Temperature errors Errors occurred at rotor, pipettor which Call customer service
prevent components to reach correct
operating temperature.

No level found in No liquid level could be found at the Check is reagent/washer was
position…. reagent/washer placed. placed correctly and liquid is
available.

LOCKED: Rotor out of The measuring and C-Ring rotor does not Check STATUS Menu for more
Temperature! reach its required temperature. information. If problem persists
contact technical service

Level in Wash Station too The pipettor could not detect system water Check if enough system water is
low for Dip-In in the wash station. This may be caused by in the container (green) and the
a DI pump malfunction of defective pipettor level sensor is connected
level detection. correctly. If problem persists
contact technical service.

CoaLyser – Operators Manual 1.3 Page 210


SAFETY IN OPERATION

Error Cause Action


Still wet in Wash Station Wash Station is full of system water. This Check if waste container (red) is
may be caused by a waste pump failure empty and level sensor is
and/or a level detection error. connected correctly. If problem
persists contact technical service.

Pumps + Level detection During initialization/test a problem related to Check pumps and correct
test failed! Measuring etc. is the pumps (DI/waste) and/or level detection connection of level sensors.
locked! (sensors) was discovered. Restart system. If problem
persists contact technical service.

Error 1: a) Level at pos. During initialization/test a problem related to Check pump and correct
xxx. Check for: Waste the waste pump and/or level detection was connection of level sensors.
pump failure! discovered. Restart system. If problem
persists contact technical service.

Error 1: Hardware_o.c, line Pipettor moving error in Z-direction or Check if there is any blockage of
xxx; Z is blocked at xxxx pipettor malfunction, wiring of pipettor the pipettor, remove and restart
system. Check cable connection.
Otherwise contact customer
service.
Error 2: Servos.c, line xxx, Pipettor moving error in X-direction or Check if there is any blockage of
Position_Error, mot=XMOV pipettor malfunction, wiring of pipettor the pipettor, remove and restart
system. Check cable connection.
Otherwise contact customer
service.
Error 1: Stepper.c, line xxx, Dilutor moving error (motor) or malfunction Check if there is any blockage of
StepRun timeout. of motor. the dilutor, remove and restart
system. Check cable connection
and motors. Otherwise contact
customer service.
Error 1: Stepper.c, line xxx, Syringe can not move as expected. Check if there is any blockage of
StepGotoPos INVALID the syringe, remove and restart
system. Check for correct
installation of syringe. Otherwise
contact customer service.
No printer found! External printer is not connected of not Check printer is correctly
powered on. connected to the USB interface,
power on printer.

Wrong license code. The entered license code to update the Check license code was entered
Software remains locked software is incorrect or entered wrong correctly, otherwise contact
customer service.

New Cuvette Ring not Cuvette Ring inserted incorrect. Insert Cuvette Ring and check
found correct insertion.

Not all cuvettes are clean Some cuvettes have found to be dirty/used. Insert new cuvette ring, check if
any cuvettes are used.

Cuvette Ring missing No Cuvette Ring was placed in rotor Place Cuvette Ring in rotor prior
start of measurements.

Invalid Customer ID Customer ID number on CuvCARD does Contact technical service


not correspond to cuvette type used.

CoaLyser – Operators Manual 1.3 Page 211


SAFETY IN OPERATION

Error Cause Action


Invalid Cuvette ID Cuvette type on cuvette card does not Contact technical service
correspond to cuvette type on analyser.

CuvCARD not compatible Utilized CuvCARD may not carry cuvette Try other cuvette card or contact
with analyzer asset or is a wrong card. technical service.

CuvCARD write error. No Inserted card type may be defect. Re-insert card and repeat
booking done. process, otherwise contact
technical service
Balance and reserve fully The cuvette ring balance and reserve Insert new CuvCARD and load
used quantity is now fully used, no further new balance to analyser, once
measurements are possible. the cuvette balance and reserve
quantity is used.
CuvCARD devalued! Cuvette ring balance on CuvCARD is fully Use new CuvCARD delivered
used, CuvCARD is empty! with cuvettes boxes.

Only reserve Message provides information that only Prepare to load new CuvCARD
measurements are certain reserve measurements are available and order new cuvette rings.
possible. for measurements.

After current C-Ring further The measuring channels will be locked Load new cuvette balance to
measurements are locked. once the current C-Ring is fully used. analyser

CoaLyser – Operators Manual 1.3 Page 212


SAFETY IN OPERATION

6.2.4 How to change fuses


The analyser has a Power Supply Unit (PSU) for a wide voltage range so
that it can be used in a power supply range from 100-240V, 50/60Hz. Fuses
may need to be changed if the analyzer does not operate even though it is
properly connected.

Only replace the analyzer fuses with fuse types specified by the
manufacturer. Observe the information on the serial number plate on the
analyser backplane. The system is supplied set to 240V.

Only use fuses that meet the following requirements:

5A/250 V time lag (T5A L250V)


IEC 60127-2/III

To change fuses follow the procedure below:

 Switch off analyzer.


 Disconnect analyzer from power supply and remove cable.
 Carefully lift the fuse holder with a screw driver. Refer to figure below.

Tab to open the fuse carrier

Fuse carrier

Main switch - ON or OFF

Pivot of cover

Power plug

Figure 249 Change fuses, voltage selector

A red fuse carrier (= voltage selector) becomes visible.

 Set the screwdriver in the upper slot of the fuse carrier and drive
carrier out of the housing.
 Check fuses to see if wires have melted and replace accordingly.
 Insert fuse carrier into the housing again.
 Close cover of the fuse carrier.
 Reconnect analyzer to power supply and check proper operation.

NOTE
Contact authorized customer service in case the exchange of fuses
has been unsuccessful.

CoaLyser – Operators Manual 1.3 Page 213


APPENDIX

7 APPENDIX

7.1 Materials supplied

Material

Analyzer 1 x CoaLyser (405nm) or


1 x CoaLyser (405/570/740nm)

Accessories 1x Distilled Water Container, incl. screw lid


1x Waste Water Container, incl. screw lid
1x Distilled Water Sensor
1x Waste Water Sensor
1x 12 Filter for Distilled Water Sensor
1x Syringe 250µl
1x 10 Cuvette Rings
1x 11 Sample Cups 4ml
1x Bottle Washing Solution
1x Bottle Cleaner
1x 3 Reducer Rings
1x USB Stick
1x Set of 2 fuses (230V)
1x CD-ROM with Operators Manuals
1x Packing carton and foam inserts

7.2 Disposables/Consumables

Material

Cuvette Rings Cuvette Rings, incl. Stir Bars


6x10 rings (6x320 measurements)

Sample Cups Sample Cups 4ml


2x250 Cups

Washing Solution Washing Solution 15ml (thrombin in activator)


5x15ml bottles

Cleaner Cleaner (pipettor/tubing)


6x12ml bottles

Teflon Stir bars Stir bars for reagent mixing


10x13mm

Reducer Rings Reducer rings for reagent vials 5ml


1x3 rings

Barcode Scanner 1 x external Barcode Scanner, incl. cable

CoaLyser – Operators Manual 1.3 Page 214


APPENDIX

7.3 Maintenance Kit

Kit consists of: Quantity

Tubings Liquid System Tubing 1

Container Distilled Water Container, incl. screw lid 1


Distilled Waste Container, incl. screw lid 1

Level Sensors Distilled Water Sensor (green) 1


Waste Water Sensor (red) 1

Filters Filter for Distilled Water Sensor 12

Syringe 250µl Syringe 1

CoaLyser – Operators Manual 1.3 Page 215


APPENDIX

7.4 Technical Data

General Description
Type of instrument Fully automated coagulation analyzer

Application Coagulation, Chromogenic, Immunologic tests

Working methods Sample oriented


Test selective
STAT (emergency)

Measuring principle Photometric; with magnetic stirrers to homogenize


the test mixture and increase the sensitivity

Light source LED, light emitting diode

Cuvette Volume 150µl - max. 300µl

Test throughput PT 108/h, aPTT 60/h, +/- 20 tests/h

Cuvette volume min. 150 µl, max. 300 µl (test suspension)

Measuring channels Two measuring channels at


- one wavelength 405nm or
- three wavelength 405/570/740nm

Calibration - Manual input of calibration points (max. 7 points),


two calibration curves for each test or
- Automatic calibration

STATS STAT loading is possible at any time in up to three


assigned positions. Routine testing can be interrupted
for STAT loading and then automatically giving
priority to the STAT sample

Analyzer data
Reagent positions 15 positions at room temperature
- (R1-R3) for 15ml reagent vials (left/right pos. stirred)
- (R4-R15) for 5ml reagent vials.
Reagent block can be removed from working area
for external cooling.

Washer positions 2 positions at room temperature


- (W1-W2) for Washer & Cleaner

Calibration/Dilution 8 positions at room temperature


- (C1-C8) for Controls, Calibrators & Dilutions

Sample positions 2x11 positions for primary & secondary tubes/cups


1x3 positions for STAT

Measuring rotor One heated measuring rotor at 37°C +/-0.4°C with


two measuring channels, designed for cuvette rings

Cuvette rings Cuvette ring with 32 cuvettes

Wash Station One wash station for pipettor cleaning

Pipettor Capacitive, movable in three directions (x, y, z),


heated at 37°C

CoaLyser – Operators Manual 1.3 Page 216


APPENDIX

Dilutor 250µl syringe


- minimal dosing 2µl
- maximal dosing 250µl
- Resolution: 12 half steps for 1 µl (for 250 µl syringe)
- Level sensors for distilled and waste water container

Rinsing Two modes


- Regular with distilled water
- 4 intensive with washing solution for probe cleaning
(Thrombin inactivator)

Cleaning Cleaning of pumps and tubing with Cleaner

Display Touch sensitive colour display


Visible area: 11.5x8.7cm
Display resolution: 320 x 240 pixels, ¼ VGA).

Software Operating system with Graphic User


Interface (GUI)

Programmable 50 reagents/tests can be freely programmed


Methods

Interfaces 1x USB for external printer device


1x USB for USB Memory device
1x LAN for external PC connection
2x RS 232C serial interfaces
1x external Barcode Scanner connection

Physical data
Voltage: 100 - 240 V
Frequency: 50 - 60 Hz
Capacity: 300 VA
5A/250 V time lag (T5A L250V)
Fuses:
IEC 60127-2/III
Noise emission < 56 dB (A)

Ambient Conditions
Usage: Inner rooms only, dry environment
Altitude:  2.000 meters
Mains voltage fluctuation: Max.  10% of rated voltage
IP Code (IP20): 2 = < 12.5mm, Fingers or similar objects
accord. IEC 60529 0 = Not protected against harmful ingress of water
Pollution degree: Class 1
Temperature: during operation +15°C - +30°C
during transport + 2°C - +50°C
Humidity: 10% - < 85%, non-condensing

Dimensions
Instrument (WxHxD): 78 x 58 x 50 cm 30.7x 22.8 x 19.7 in.
Shipping:

Weight
Instrument: 25 kg 55 lb
Shipping: 30 kg 66 lb

CoaLyser – Operators Manual 1.3 Page 217


APPENDIX

7.5 Safety Specifications

This instrument conforms to the appropriate European Directives.


The instrument type described in this manual bears an IVD and CE mark,
IVD which certifies conformity with the requirements of the following European
directives:

98/79/EC in-vitro Diagnostics directive

EN55011:2007 + A2:2007, EN61326-2-6:2006, EN61000-3-2:2006,


EN61000-3-3:1995 +A1:2001 +A2:2005, EN60601-1-2:2007, EN591:2001

EN61010-1:2001, EN61010-2-101:2002, EN61010-2-081/A1:2003

CoaLyser – Operators Manual 1.3 Page 218


APPENDIX

7.6 Mathematics

Computation of PT %
1
(( meas. time-100 % time)/factor)+0.01) = PT %

Range for conversions:

Method Unit from to exceeded

PT: % 2,0 250 error


Ratio --- ---
INR --- 25 error

Fibrinogen: g/l 0,4 9,999 error


mg/dl 40 999,9 error

Another limit is the lag phase of the method.

Extrapolation
PT >100 % linear extrapolation over the last two higher points.
PT <10 % linear extrapolation over the last two lower points.

Fibrinogen: linear extrapolation respectively over the last two points.

Calibration curve axes


PT: reciprocal/linear
Fibrinogen: log/log
Extr. Factor log/log
Intr. Factor log/log

Conversion to ratio
and INR:
RATIO = measured clotting time / normal value

INR = RATIO ISI (International Normalized Ratio)

ISI = International Sensitivity Index according to package insert

CoaLyser – Operators Manual 1.3 Page 219