HUMAN SERVICE BULLETIN

HumaStar 100/200 HSB No 0018

Subject Setting Update Version 4.5


Issue date 20.04.2018
Author Gabriele Moos, Christof Dembny
Instrument(s) HumaStar 100/200 (Ref 16890/16895)
Serial number  All
range  From: …
 To: …
Implementation  Immediately
 Next service
 Upon problem occurrence
 Information
Part(s) recall  Yes
 No
 Not applicable

Brief description
The new Setting Update Version 4.5 includes settings for two (2) new assays and changes/improvements for
existing settings.
Version 4.5 will replace Version 4.4.1.

Details about the new Setting Update Version 4.5


 Settings for two (2) new assays
CHOLINESTERASE liquicolor, REF 12007, 1 x 120 ml
URINARY TOTAL PROTEIN liquicolor, REF 10590, 3 x 50 tests
 Changes and improvements for existing settings
HbA1c-2: Revised setting to correct the limitation in reporting HbA1c results < 6% or
< 42 mmol/mol Hb.
Protein assays: Automated standard dilutions were added to the existing settings;
information on suitable diluents are now included in the setting PDF file.

Summary of changes and improvements:

Kit name REF Analyzer Changes / improvements


APOLIPOPROTEIN A1 11101 HumaStar 100 & 200 Standard dilution rate and diluent
APOLIPOPROTEIN B 11102 HumaStar 100 & 200 Standard dilution rate and diluent
CHOLINESTERASE liquicolor 12007 HumaStar 100 & 200 New !
COMPLEMENT C3 11110 HumaStar 100 & 200 Standard dilution rate and diluent
COMPLEMENT C4 11113 HumaStar 100 & 200 Standard dilution rate and diluent
CYSTATIN-C liquidirect 11150 HumaStar 200 Std 0 – saline/NaCl added to PDF file
HbA1c liquidirect 10770 HumaStar 100 & 200 Reaction type Endpoint; Sample and
reagent volumes, incubation and reading
time, blank correction
MICROALBUMIN 11120 HumaStar 100 & 200 Standard dilution rate and diluent
RHEUMATOID FACTORS 11261P HumaStar 100 & 200 Standard dilution rate and diluent
TRANSFERRIN 11115 HumaStar 100 & 200 Standard dilution rate and diluent
URINARY TOTAL PROTEIN liquicolor 10590 HumaStar 100 & 200 New !

16890_16895_HSB_0018 Setting Update Version 4.5 MKA_CDE.docx Form 7.2-02.06-01


Vers. 01
1/2 Valid from 04.05.2017
Instruction / Procedure
The new Setting Update Version 4.5 can be downloaded from the login section of the HUMAN website:
 Either Service > Technical Downloads > Instruments > HumaStar 100.
 Or Service > Technical Downloads > Instruments > HumaStar 200.

There are two update version 4.5 download files available:


A) V4.5 (complete) containing the complete set of methods.
B) V4.5 (changed only) containing only methods that have changed.

Please make sure that the HI software is not running before removing method folders!

A) Installation "complete"

1. Download the file hs100_settings_v4.5.zip or hs200_settings_v4.5.zip.

2. Extract the file. The folders hs100_settings_v4.5 or hs200_settings_v4.5 will be created.

3. Open the folder and run the start_HS100.bat or start_HS200.bat file.


A backup of the current Human folder will be created (e. g. Human_2018_5_4_10_9_17).
A new Human folder with the version 4.5 settings will be created.

Notes
Please be aware that the HI software is currently limited to 50 methods and the complete setting files contain
more than 50 methods.
Since the smr files are listed in alphabetical order in the C:\ProgramData\Hi\Human\Methods folder and are
accessed by the software in the same order, only the first 50 methods will be usable by the software.
Hence, to enable measurement of all necessary but maximum 50 methods, Method folders of a test including
the smr file of unused methods must be deleted from the C:\ProgramData\Hi\Human\Methods folder.

More than 50 methods can cause an interruption of the communication between the analyzer and the PC.

B) Installation "changed only"

1. Copy the following folders to a different place, or a USB memory stick or any other external storage device:
C:\ProgramData\Hi\Human\Methods
C:\ProgramData\Hi\Human\Application sheets

2. Download the file hs100_settings_v4.5_changed_only.zip or hs200_settings_v4.5_changed_only.zip.

3. Extract the file.

4. Copy the *.pdf files into the Application sheets folder.

5. Copy all Assay folders (containing the *.smr-files) into the Methods folder. Each *.smr-file can be found in
a separate folder.

6. After confirming that the new assays are running properly the folders, stored before in “Step 1.” as a back-
up, can be deleted.

16890_16895_HSB_0018 Setting Update Version 4.5 MKA_CDE.docx Form 7.2-02.06-01


Vers. 01
2/2 Valid from 04.05.2017