HSB - 0018 Setting Update Version 4.5
HSB - 0018 Setting Update Version 4.5
Brief description
The new Setting Update Version 4.5 includes settings for two (2) new assays and changes/improvements for
existing settings.
Version 4.5 will replace Version 4.4.1.
Please make sure that the HI software is not running before removing method folders!
A) Installation "complete"
Notes
Please be aware that the HI software is currently limited to 50 methods and the complete setting files contain
more than 50 methods.
Since the smr files are listed in alphabetical order in the C:\ProgramData\Hi\Human\Methods folder and are
accessed by the software in the same order, only the first 50 methods will be usable by the software.
Hence, to enable measurement of all necessary but maximum 50 methods, Method folders of a test including
the smr file of unused methods must be deleted from the C:\ProgramData\Hi\Human\Methods folder.
More than 50 methods can cause an interruption of the communication between the analyzer and the PC.
1. Copy the following folders to a different place, or a USB memory stick or any other external storage device:
C:\ProgramData\Hi\Human\Methods
C:\ProgramData\Hi\Human\Application sheets
5. Copy all Assay folders (containing the *.smr-files) into the Methods folder. Each *.smr-file can be found in
a separate folder.
6. After confirming that the new assays are running properly the folders, stored before in “Step 1.” as a back-
up, can be deleted.