True-RMS Digital
Multimeter
Instruction Manual
TABLE OF CONTENTS A. INTRODUCTION
A. INTRODUCTION
1. Congratulations!!
1. Congratulations . . . . . . . . . . . . . . . . . .3
Thank you for purchasing Snap-on products. The meter is
2. Product Description . . . . . . . . . . . . . . .3
easy to use and is built to last. It is backed by a 7 year
3. EC Declaration of Conformity . . . . . . . .4 limited warranty. Please remember to complete and return
B. SAFETY CONSIDERATIONS . . . . . . . . . . . . .5 your product warranty registration card.
C. TECHNICAL DATA
1. Features and Benefits . . . . . . . . . . . . . .6 2. Product Description
2. Product Applications . . . . . . . . . . . . . .7 The EEDM525D is a hand held, True-RMS autoranging
3. Specifications . . . . . . . . . . . . . . . . .8-11 DMM. The EEDM525D measures ACV, DCV, ACA, DCA,
Resistance, Frequency, Duty Cycle, Diodes, Capacitance,
D. MEASUREMENT TECHNIQUES Temperature and Continuity.
1. Controls and Functions . . . . . . . . . . .12
a) Push Buttons . . . . . . . . . . . .12 The EEDM525D also features:
b) Rotary Switch . . . . . . . . . . . .13
c) Input Jacks . . . . . . . . . . . . . .13 • REC Records Min/Max readings during
d) Disable Auto Power Off . . . .13 specified measurement intervals.
2. Step by Step Procedures:
• RANGE Allows the user to manually range the
a) Measuring DC Volts . . . . . . .14
EEDM525D instead of autoranging.
b) Measuring AC Volts . . . . . . .15
c) Measuring DC Amps . . . . . .16 • HOLD Holds the reading on the display for
d) Measuring AC Amps . . . . . .17 easy viewing.
e) Measuring Resistance . . . . .18 • REL Displays the value as a difference to
f) Measuring Diodes . . . . . . . .19 a reference value
g) Continuity Buzzer . . . . . . . . .20
h) Measuring Capacitance . . . .21 • AUTO OFF Preserves battery life.
i) Measuring Frequency . . . . . .22
The EEDM525D comes complete with the following
j) Measuring Temperature . . . .23
accessories:
k) Record Mode . . . . . . . . . . . .24
l) Relative Mode . . . . . . . . . . .24 EEDM525D Instrument
F. MAINTENANCE Soft Pouch
1. Testing Fuses in Circuit . . . . . . . . . . . .25 Rubber Boot
2. Fuse Replacement . . . . . . . . . . . . . . .25 Test Lead Set
Instruction Manual
3. Battery Replacement . . . . . . . . . . . . . .26
4. Cleaning Your Meter . . . . . . . . . . . . . .27
G. TROUBLE SHOOTING GUIDE . . . . . . . . . . .27
2 3
3. EC Declaration of Conformity B. SAFETY CONSIDERATIONS
This is to certify that model EEDM525D conforms to the
protection requirements of the council directive WARNING: Please follow manufacturers test
89/336/EEC, in the approximation of laws of the member procedures whenever possible. Do not attempt to
states relating to Electromagnetic compatibility and measure unknown voltages or components until a
73/23/EEC, The Low Voltage Directive by application of the complete understanding of the circuit is obtained.
following standards:
Read instructions before operating:
EN61326 : 1997 + A1 + A2 : 2001 Be sure these instructions accompany the tool when
EN61010-1 :2001 Safety Standard passed from one user to a new or inexperienced user.
To ensure conformity with these standards, this instrument Equipment is used in a manner not specified by the
must be operated in accordance with the instructions and manufacturer, the protection provided by the
specifications given in this manual. equipment may be impaired.
CAUTION:
Even though this instrument complies with the GENERAL GUIDELINES
immunity standards, the accuracy can be affect-
ed by strong radio emissions not covered in the ALWAYS
above standards. Sources such as hand held • Test the EEDM525D before using it to make sure
radio transceivers, radio and TV transmitters, it is operating properly.
vehicle radios and cellular phones generate • Inspect the test leads before using to make sure
electromagnetic radiation that could be induced
there are no breaks or shorts.
into the test leads of this instrument. Care
• Double check all connections before testing.
should be taken to avoid such situations or
alternatively, check to make sure that the • Have someone check on you periodically if
instrument is not being influenced by these working alone.
emissions. • Have complete understanding of circuit being
measured.
• Disconnect power to circuit then, connect test
leads to the EEDM525D, then to circuit being
measured.
NEVER
• Attempt to measure unknown high voltages.
• Attempt to measure current with the meter in
parallel to the circuit.
• Connect the test leads to a live circuit before
setting up the instrument.
• Touch any exposed metal part of the test lead
assembly.
4 5
INTERNATIONAL SYMBOLS 2. Product Applications
6 7
3. Specifications f. ACV (True-RMS, 1kHz - 20kHz)
IEC 61010-1 Over Voltage: Range Resolution Accuracy Impedance
CAT II - 1000V
CAT III - 600V
4V 0.001V
Pollution Degree 2 40V 0.01V ±(2.0% + 3 digits) 10MW
400V 0.1V
a. DCmV 750V 1V Unspecified
Range Resolution Accuracy Impedance
40mV 0.01mV ±(0.5% + 2 digits) 10MW g. DCA
400mV 0.1mV Range Resolution Accuracy Overload Protection
40µA 0.01µA ±(0.8% +2 digits)
b. DCV 400µA 0.1µA Fuse
Range Resolution Accuracy Impedance 4000µA 1µA ±(0.8% +5 digits) 0.5Amp/600V
400mV 0.1mV 40mA 0.01mA
4V 0.001V 400mA 0.1mA
40V 0.01V ±(0.5% + 2 digits) 10MW 4A 0.001A ±(1.2% +5 digits) Fuse
400V 0.1V 10A 0.01A 10Amp/600V
1000V 1V
h. ACA (True-RMS)
c. ACmV (True-RMS, 40 - 400Hz)
Range Resolution Accuracy Overload Protection
Range Resolution Accuracy Impedance 400µA 0.1µA Fuse
40mV 0.01mV ±(1.5% + 3 digits) 10MW 4000µA 1µA ±(1.2% +5 digits) 0.5Amp/600V
400mV 0.1mV 40mA 0.01mA
400mA 0.1mA
d. ACV (True-RMS, 40 - 400Hz) 4A 0.001A ±(1.5% +10 digits) Fuse
Range Resolution Accuracy Impedance
10A 0.01A 10Amp/600V
400mV 0.1mV ±(1.5% + 3 digits)
4V 0.001V *Warning: Use only correct size, voltage and current rated fuses.
Test Leads: Use only correct type and overvoltage category rating.
40V 0.01V ±(0.75% + 3 digits) 10MW
400V 0.1V
750V 1V ±(1.0% + 5 digits) i. OHM (Resistance,W)
Range Resolution Accuracy Overload Protection
e. ACV (True-RMS, 400Hz - 1kHz) 400W 0.1W ±(0.8% +5 digits) 600V DC or
Range Resolution Accuracy Impedance 4kW 0.001kW AC Peak
4V 0.001V 40kW 0.01kW ±(0.5% +3 digits)
40V 0.01V ±(1.5% +5 digits) 10MW 400kW 0.1kW
400V 0.1V 4MW 0.001MW ±(1.0% +5 digits)
750V 1V ±(2.0% + 5 digits) 40MW 0.01MW ±(1.5% +10 digits)
8 9
j. Duty Cycle / Hz o. Temperature (K-Type)
Range Range Resolution Accuracy
0.1 ~99.9% (0.5Hz to 500kHz, Width > 2uS) CENTIGRADE
Accuracy -40° to 10°C 0.1° ±(3.0% + 5°C)
((0.1% + 0.05% / kHz) +1 Count 10° to 200°C 0.1° ±(1.0% + 3°C)
200° to 400°C 0.1° ±(2.0% + 5°C)
k. Diode Test 400° to 1300°C 1°C ±(3.0% + 7°C)
Test Voltage Max Test Current Over Load Protection FAHRENHEIT
2.7V Approx. 1mA 600 V DC or Peak AC -40° to 50°F 0.1° ±(3.0% + 2.5°F)
50° to 392°F 0.1° ±(1.0% + 1.5°F)
l. Continuity Buzzer 392° to 752°F 0.1° ±(2.0% + 5°C)
Test Voltage Threshold Over Load Protection
752° to 2372°F 1°F ±(2.0% + 5°C)
0.6V < 30W 600 V DC or Peak AC
p. General Specifications
m. Capacitance Max. Volt. between 1000V
any Input and Ground
Range Resolution Accuracy Overload Protection Fuse Protection mA: 0.5Amp/600VAC A: 10Amp/600VAC
40nF 0.01nF Display Type 4,000 Count, 2 times per second update
Operating Temp. 0° to 40°C (32° to 104°F)
400nF 0.1nF Storage Temp. -10° to 50°C (14° to 122°F)
4uF 0.001nF ±(3.0% +10 digits) 600V DC or Relative Humidity 0% to 80%
40uF 0.01uF AC Peak Power Supply 2 Each 1.5 Volt “AA” Batteries
Battery Life 200 hrs. Typical
400uF 0.1uF Size (H x L x W) 33mm x 86mm x 187mm (1.3in x 3.4in x 7.4in)
4000uF 1uF ±(7.0% +10 digits) Weight 340g (12oz)
n. Frequency (Hz)
Range Resolution Accuracy Overload Protection
10Hz 0.01Hz
100Hz 0.1Hz
1KHz 0.001KHz ±(0.05% +3 digits) 600V DC or
10KHz 0.01KHz AC Peak
100KHz 0.1KHz
1MHz 0.001MHz
10MHz 0.01MHz
10 11
D. MEASUREMENT TECHNIQUES
b. Rotary Switch
1. Controls and Functions:
OFF Turns the EEDM525D off.
Function for measuring AC voltage (ACV).
Function for measuring DC voltage (DCV).
Function for measuring AC/DC millivolts (mV).
LCD Funtion for measuring resistance, diode testing
and continuity buzzer
Display
Push 40 Function for measuring up to 40 DC micoamps
Buttons (uA)
Function for measuring up to 4000 AC/DC
microamps.
Function for measuring up to 400 AC/DC
Rotary milliamps.
Switch Function for measuring up to 10 AC/DC amps
Hz Function for measuring Frequency
CAP Function for measuring Capacitance
Input
c. Input Jacks
Jacks
VWHz Red test lead connection for all Volt, Ohm,
Frequency and Capacitance functions.
a. Push Buttons COM Black test lead connection for all functions.
RANGE Activates manual ranging. Hold in for 3 seconds uAmA Red test lead connection for current
to return to autorange. measurements on the uA and mA functions.
REC Activates the Min/Max mode.
A Red test lead connection for current
Hz/DUTY Toggles the ACV or ACA measurement mode to measurements on the A function.
Hz or Duty Cycle mode.
REL Displays value as a difference of reference value. d. Disable Auto Power Off (APO)
With the rotary switch in the OFF position Press and
FUNCTION Activates optonal fuctions highlighted in orange hold down the FUNCTION push button while turning
around dial. APO (Auto Power Off) is disabled in the instrument on.
this function. Hold in for 3 seconds to deactivate.
LPF Activates Low Pass Filter.
12 13
2. Step by Step Procedures:
b. MEASURING AC VOLTS
a. MEASURING DC VOLTS
CAUTION!
CAUTION! Do not attempt to make a voltage measure
Do not attempt to make a voltage mea ment if a test lead is plugged in the A or
surement if a test lead is plugged in the µmA input jack. Instrument damage and/or
A or µmA input jack. Instrument damage personal injury may result.
and/or personal injury may result.
WARNING!
WARNING! Do not attempt to make a voltage measure
Do not attempt to make a voltage ment of more than 750V or of a voltage l
measurement of more than 1000V or of a level that is unknown.
voltage level that is unknown.
Instrument set-up:
FUNCTION BLACK RED MINIMUM MAXIMUM
Instrument set-up: TEST LEAD TEST LEAD READING READING
FUNCTION BLACK RED MINIMUM MAXIMUM COM VWHz 0.001V 750V
TEST LEAD TEST LEAD READING READING
mV COM VWHz 0.1mV 400.0mV Measurement Procedure:
V COM VWHz 0.001V 1000V
1. Disconnect power to the circuit to be measured.
Measurement Procedure: 2. Plug the black test lead into the COM input jack.
1. Disconnect power to the circuit to be measured. 3. Plug the red test lead into the VWHz input jack.
2. Plug the black test lead into the COM input jack. 4. Set the rotary switch to the function depending on
the voltage to be measured.
3 Plug red test lead into the VWHz input jack.
5. Connect the test leads to the circuit to be measured.
4. Set rotary switch to either the or range,
depending on the voltage to be measured. 6. Reconnect power to the circuit to be measured.
5. Connect the test leads to the circuit to be measured. 7. Read the voltage on the EEDM525D.
6. Reconnect power to the circuit to be measured. Optional Functions:
7. Read the voltage on the EEDM525D.
RANGE Manually select the appropriate range.
Optional Functions: REC Activate MIN/MAX record mode (page 24).
RANGE Manually select the appropriate range.
Hz/DUTY Scroll between Hz, Duty, and Volts.
REC Activate MIN/MAX record mode (page 24).
REL Activate REL mode (page 24).
REL Activate REL mode (page 24).
HOLD Freeze the reading on the LCD.
HOLD Freeze the reading on the LCD.
14 15
c. MEASURING DC AMPS d. MEASURING AC AMPS
CAUTION!
CAUTION! Do not attempt to make a current measure
Do not attempt to make a current measurement with ment with the test leads connected in
the test leads connected in parallel with circuit to be parallel with the circuit to be tested. Test
tested. Test leads must be connected in series with leads must be connected in series with the
the circuit. circuit.
WARNING! WARNING!
Do not attempt to make a current measurement of Do not attempt to make a current measure
circuits with more than 600V present. Instrument ment of circuits with more than 600V
damage and /or personal injury may result. present. Instrument damage and /or
personal injury may result.
Instrument set-up:
FUNCTION BLACK RED MINIMUM MAXIMUM Instrument set-up:
TEST LEAD TEST LEAD READING READING FUNCTION BLACK RED MINIMUM MAXIMUM
40µA COM µAmA 0.01µA 40µA TEST LEAD TEST LEAD READING READING
µA COM µAmA 0.1µA 4000µA µA COM µAmA 0.1µA 4000µA
mA COM µAmA 0.01mA 400mA mA COM µAmA 0.01mA 400mA
10A COM A 0.001A 10.00A 10A COM A 0.001A 10.00A
WARNING! CAUTION!
Do not attempt to make resistance measurements Do not attempt to make diode measurements with
with circuit energized. For best results, remove circuit energized. The only way to accurately test
the resistor completely from the circuit before a diode is to remove it completely from the circuit
attempting to measure it. before attempting to measure it.
1. Disconnect power to the circuit to be measured. 5. Connect black test lead to the banded end of the diode
and the red test lead to the non-banded end of the
2. Plug the black test lead into the COM input jack.
diode.
3. Plug the red test lead into the VWHz input jack.
6. Reading on the display should be between
4. Set the rotary switch on the EEDM525D to the W
0.5 and 0.8 volts.
function.
5. Connect the test leads to the circuit to be measured. 7. Reverse test lead connections in 5 above.
6. Read the resistance value on the EEDM525D. 8. Reading on the display should be OFL (Overload).
18 19
g. CONTINUITY BUZZER h. MEASURING CAPACITANCE
WARNING! CAUTION!
Do not attempt to make continuity measurements Disconnect power to the circuit(s) to be
with circuit energized. measured. Discharge the capacitorto be
measured completely before attempting to
Instrument set-up: measure.
FUNCTION BLACK RED
TEST LEAD TEST LEAD
COM VWHz
Instrument set-up:
Measurement Procedure: FUNCTION BLACK RED MINIMUM MAXIMUM
TEST LEAD TEST LEAD READING READING
1. Disconnect power to the circuit to be measured. CAP COM VWHz 0.01nF 4000uF
2. Plug the black test lead into the COM input jack. Measurement Procedure:
3. Plug the red test lead into the VWHz input jack.
4. Set the rotary switch to the function. 1. Disconnect power and discharge the capacitor to be
measured.
5. Press yellow push button to activate continuity buzzer. 2. Plug the black test lead into the COM input jack.
6. Connect the test leads to the circuit to be measured. 3. Plug the red test lead into the VWHz input jack.
7. Listen for the buzzer to confirm continuity. 4. Set the rotary switch to the CAP function.
5. Connect the test leads to the capacitor to be
measured.
6. Read the capacitance on the EEDM525D.
Optional Functions:
20 21
i. MEASURING FREQUENCY j. MEASURING TEMPERATURE
CAUTION! CAUTION!
Do not attempt to make a frequency Disconnect power to the circuit(s) to be
measurement if a test lead is plugged in the measured. Discharge the capacitorto be
A or µmA input jack. Instrument damage measured completely before attempting to
and/or personal injury may result. measure.
WARNING!
Never attempt a frequency measurement
with a voltage source reater than 600V Instrument set-up:
FUNCTION TEMP. TEMP. MINIMUM MAXIMUM
Instrument set-up: INPUT ADPT. INPUT ADPT. READING READING
FUNCTION BLACK RED MINIMUM MAXIMUM TEMP COM µmATEMP 0.1°F 2462°F
TEST LEAD TEST LEAD READING READING 0.1°C 1350°C
Hz COM VWHz 0.001Hz 10MHz
Measurement Procedure:
Hz/DUTY Scroll between Hz and Duty %. REL Activate REL mode (page 24).
HOLD Freeze the reading on the LCD. FUNCTION Toggle between °C or °F.
22 23
k. RECORD MODE F. MAINTENANCE
The record mode saves minimum (MIN) and maximum 1. Testing Fuses In Circuit: Both the A and mAµA input
(MAX) values measured for a series of reading. Activate jacks are fuse protected. The fuses can be tested
the function as follows: without removing them from the instrument as
1. Depress the REC button on the EEDM525D. follows:
2. The EEDM525D will immediately start to record a. Set the EEDM525D to the diode test function.
MIN/MAX values. REC will be on the LCD to b. Insert the red test lead into the V input jack.
show record mode has been activated. The
reading on the LCD will be the actual reading. c. Touch the tip of the red test lead into the A
The EEDM525D will give a confirmation beep input jack making sure you contact the metal.
every time a new value is recorded. d. If the display reads any number, the fuse is good.
3. Press the REC button a second time and the If the display reads .OL, the fuse is open and
MIN reading will be displayed. must be replaced.
4. Press the REC button a third time and the MAX e. Repeat the same procedure for the uAmA input
reading will be displayed on the LCD. jack.
l. RELATIVE MODE a. Disconnect and remove all test leads from live
circuits and from the EEDM525D.
The Relative mode compares readings to a known value b. Remove EEDM525D from protective boot.
and displays it as a difference to that value on the LCD.
c. Remove the three screws from the lower back of
1. Measure the known value on the EEDM525D housing holding the compartment cover in place.
and press the REL button, the LCD will display
zero. d. Remove the compartment cover.
2. Measure next device for comparison. e. Remove the old fuse(s) and replace it with new
fuse(s).
3. The LCD will display the difference between the
new device and the stored reference value. f. Reassemble the instrument in reverse order
from above.
5. To terminate the Relative mode, hold the REL
button down for approximately 2 seconds or
turn the rotary switch to a different function.
24 25
3. Cleaning Your Meter: The EEDM525D can be wiped
3. Battery Replacement: The EEDM525D will display a clean with a damp cloth and mild detergent. Do not
battery symbol in the upper left corner of the LCD when submerse in water.
the two internal 1.5 Volt “AAA” batteries need
replacement. Batteries are replaced as follows:
a. Disconnect and remove all test leads from live
circuits and from the EEDM525D.
G. TROUBLE SHOOTING GUIDE
b. Remove EEDM525D from protective boot.
c. Remove the three screws from the lower back of Problem Probable Causes
housing holding the compartment cover in place.
Does not power up
d. Remove the compartment cover. • Dead or defective battery
e. Remove old batteries and replace with new • Broken wire from battery
batteries. snap to PCB
f. Reassemble instrument in reverse order from
above. Won’t display current readings
• Open fuse
• Open test lead
• Improperly connected to
circuit under test
Battery/Fuse Compartment
10 Amp
Fuse
0.5 Amp
Fuse