● This unit is not constructed to be waterproof or dust Measuring: mils/µm button

OPERATING INSTRUCTIONS proof, so do not use it in a very dusty environment or in 1. Turn on the power. 1. Press “mils/µm” key to switch between mils and µm.
TG 900 one where it will get wet. 2. Pull the trigger to the measuring the thickness of the (1 mils = 25.4 µm)
● Please make sure there is no air bubbles underneath. foil. 2. Selected unit is shown in LCD above the reading.
COATING THICKNESS GAUGE ● One point calibration: must implement this proce- 3. Releasing trigger to stop measuring and automatically
dure for each time hold the display reading.
● Two point calibration: suggest implementing for 4. During the trigger being pulled, reading is refreshed OPERATION
frequent testing points to increase measuring accu- every second. 1. Pull the trigger to turn on the meter.
racy. 5. During the trigger being pulled, APO is inactive. 2. Press the sensor on the object.
● This meter can only measure ferrous material. 3. Pull the trigger to the measuring the thickness, releasing
trigger to stop measuring the thickness.
CALIBRATION
SPECIFICATIONS Self-calibration
GENERAL One point calibration: (4.0mils/102µm) MAINTENANCE
Display: 3½ digit liquid crystal display (LCD) wit maxi- 1. Select unit: (mils/µm) Battery Replacement
mum reading of 1999. 2. Press “CAL” and “▼” about four seconds till display 1. Power is supplied by a 9 volt “transistor” battery
Low battery indication: The “ ” is displayed when the shows “CAL”, “HOLD”, “1 −1”, and “1 −1” will flash. (NEDA 1604, IEC 6F22).
battery voltage drops below the operating level. 3. Put 4.0mils (102µm) calibration sheet on the metal plate, 2. Pull off battery cover “ ”.
Measurement rate: 1 second, nominal. and then put the sensor tip on calibration sheet holding 3. Remove the battery cover by gently sliding it onwards
Operating Environment: 32°F to 122°F (0°C to 50°C) at the trigger to measure at the same time. And release it to the bottom of the meter.
< 75% R.H. hold readings, press “CAL” button the display will 4. Remove and disconnect the old battery from the meter
Storage Temperature: -4°F to 140°F (-20°C to 60°C), show “− − −” flashing three time and shows “1 − 2”. and replace with a new unit. Wind the excess lead
0 to 80% R.H. with battery removed from meter. And hold “CAL” for four seconds to exit calibration length and put the top of battery beneath the battery
Auto power off: 15 seconds. program. chamber. Install the battery and put the battery cover.
Standby consuming current: <6µA. 4. One point calibration must use 4.0mils/102µm standard
Battery: Standard 9V battery (NEDA 1604, IEC 6F22 sheet to proceed. Cleaning
006P). Periodically wipe the case with a damp cloth and deter-
INTRODUCTION Battery Life: 9 hours (continuity) typical (contain Back- Two point calibration: gent, do not use abrasives or solvents.
This instrument is a portable easy to use 3½ digit, com- lit).
pact-sized digital coating thickness gauge sighting de- 1. Turn on the power.
Dimensions: 148mm (H) x 105mm(W) x 42mm(D). 2. Select unit (mils or µm).
signed for simple one hand operation. Meter comes with Weight: Approx. 157g (including battery).
Backlit LCD display, Auto-Hold function and auto power 3. Press and hold “CAL” key (for about 4 sec) until “2-1”
down (15 seconds approx.) after releasing trigger to extend flashes on the LCD and “CAL”, “HOLD” indicated at
ELECTRICAL middle-lower of LCD. These mean the gauge is in cali-
battery life.
Thickness Range: 0 to 40.0mils (0 to 1000µm). bration mode. Press and hold “CAL” key for about 4
Display Resolution: 0.1mils/1µm. seconds. To exit the calibration mode.
CAUTION Accuracy: 4. Press the sensor onto the metal calibration plate, hold
● Do not use the unit near any device which generates ±4dgts on 0 to 7.8mils the trigger and wait unit there is a measured reading on
strong electromagnetic radiation or near a static electri- ±7dgts on 0 to 199µm the LCD, and adjust the reading to “00.0” (for mils,
cal charge, as these may cause errors. ±(3%+4dgts) on 7.9mils to 40mils(200µm to 1000µm) “000” for µm) with “▲” and “▼” keys. Then press
● Do not use the unit where it may be exposed to corro- Temperature Coefficient: ±0.1% of reading, whichever is “CAL” key, LCD shows “---” 3 times, and LCD
sive or explosive gases. The unit may be damaged, or greater, change in accuracy per °F/°C change in ambient shows “2-2”. Press and hold “CAL” key for about 4
explosion may occur. operating temperature above 82.4°F/28°C or below
seconds. To abort the input data and exit the calibration
● Do not keep or use this unit in an environment where it 64.4°F/18°C.
mode.
Response Time: 1 second.
will be directly illuminated by sunshine, or where it 5. Place the foil with known thickness on the metal cali- 
condensation. If you do, it may be deformed, its insula- bration plate, press the sensor onto the foil, hold the Please Attention
tion may be damaged, or it may no longer function ac-
cording to specification.
OPERATING INSTRUCTIONS trigger and wait unit there is a measured reading on the
Turn on and off power: LCD, and adjust the reading to the thickness of the foil
● Do not place the meter on or around hot objects with “▲” and “▼” keys. Then press “CAL” key, LCD
(70°C/158°F). It may cause damage to the case. 1. Pull the trigger to turn on power, LCD shows “run”.
2. Auto Power off (APO) function: Leave the gauge with- shows “---” 3 times, and LCD shows “2-3”.
● If the meter is exposed to significant changes in ambient 6. Remove the gauge from the foil and calibration plate.
temperature (hot to cold or cold to hot). Allow 30 out operation for 15 seconds, power turns off automati-
cally. 7. Press and hold the “CAL” key for 4 seconds to exit the
minutes for temperature stabilization, before taking calibration mode, the power will go off and the calibra-
measurement. tion is completed.
● If the meter continues use over one minute, the accuracy 8. If it exits the calibration mode, the procedure is not
of the measurement of the higher thickness will become completed. The power does not go off, and the calibra-
degraded. But the meter is still within its specified ac- tion data kept the previous calibration settings.
curacy.
● Condensation may form on the sensor when going from
“ ” button
a cold to hot environment-wait 10 minutes for conden-
Use “ ” button to select turn on or off the Back-Light. Ver 2. 09/09/08
sation to dissipate before taking measurements. 140-00113-C/17008-01131