THICKNESS GAUGE
MG- 411
USER’S MANUAL
USER’S MANUAL
WATERPROOF
THICKNESS GAUGE
MG-411
On the front wall of the meter there is an LCD placed which shows the results
of the thickness measurements in µm or miles.
The unit symbol is displayed next to the readout. The level of the battery
charge is signalised by the symbol.
The probe is inseparably connected with the meter by a 60-meter long cable.
Pic. 1.
-4-
After switching it on by pressing the button (Pic. 2), the meter tests the
memory and the display by displaying all symbols.
Pic. 2
If the test ends successfully, after about 1.5 s the meter switches
automatically to the temperature measuring mode. Displaying the (help)
symbol means that the meter has lost the factory settings and requires
sending to the manufactuter’s repair service.
The meter is switched off by pressing the button. In order to save the
battery, it switches automatically off after every 10 minutes of non-use. This
function is deactivated while working with power adapter and during
calibration.
5. PREPARATION TO WORK
Pic. 3
- press and hold the button until the meter returns to the
measurement mode.
-6-
5.2. Choosing the Auto-Hold
Pic. 4
6. CALIBRATION
The meter is able to store the standard thickness values in range 5 ÷ 1950
µm. In order to enter these values:
- switch the meter on pressing the button;
- in the measurement mode press the button until the symbol
appears in the lower row of the display and the upper row shows the
current standard foli thickness value;
- enter the new standard thickness value by the , buttons;
- back to the measuring mode pressing the button .
The entered value is stored in the meter‘s memory independently from the
power supply.
-8-
Pic. 5
Pic. 6
After entering the calibration mode, it is possible to ommit the zero calibration
and to calibrate only on the standard foil (one-point calibration).
7. MEASUREMENT
When the meter is prepared to work (point 5), the measurement can be
started. The probe should be put and slightly pressed to the measured
surface with its whole circumference. After stabilisation read the result. In
case, when the Auto-Hold function is activated, stabilisation of reading is
signalised by the sound, the HOLD symbol appears and the reading is
freezed on the display for 10 seconds from the moment the probe is taken
from the measured surface. In case of touching the surface with the probe
once again before the 10 seconds have passed, the HOLD symbol and the
freezed reading are erased.
Note: the probe may be put to the surface once again at least 2 seconds
after it has been taken from it. If the break is shorter, the previously
freezed reading may be still displayed.
In order to check the software version number turn the meter off and next,
holding the button, turn the meter on by pressing the button.
Instead of the display test, the screen as in the picture below will appear
(Pic. 7). In the upper row the software version will appear and in the lower
row a type of internal power supply, to which the meter has been adjusted:
- internally powered by 6F22 rechargeable battery;
- internally powered by 6F22 standard battery.
Pic. 7
To view stored readings, in the measuring mode press the button until
the number of the last recorded reading displays in turns with the
corresponding reading (Pic. 8).
Pic. 8
To delete readings:
- in the measuring mode press the button until the last stored number
in turns with corresponding reading will be displayed (Pic. 8).
- with the , buttons choose the number, from which the storage is
to be deleted;
- press and hold the button until the symbol appears, what will
delete the stored readings starting with the chosen number to the last
recorded;
- press and hold the button to return to the measuring mode.
In case of filling up the memory capacity, further readings will not be stored.
To store new readings it is necessary to delete the previous ones acting as it
is described above. In case of necessity to empty the memory, it should be
started from the first number.
- 14 -
THICKNESS MEASUREMENT:
accuracy
range resolution
(±1 digit)
±2 % ±2µm
0 ÷ 1999 µm 1 µm
0 ÷ 19.99 mils 0.01 mils ±2% ±0.08 mils
÷ 80.0 mils 0.1 mils ±2 %
EXTRA FEATURES:
OTHER:
12. EQUIPMENT
Standard set:
1. The meter with thickness sensor;
2. Calibration or controlling foils;
3. Standard plate for calibration;
4. Plastic container for the meter with sensor;
5. User’s manual with warranty.
.........................................
Date of production...........................
Date of sale.....................................
Date of expiry..................................
Correct Disposal of This Product (Waste Electrical & Electronic Equipment)
(Applicable in countries with separate collection systems)
This marking on the product, accessories or literature indicates that the product and its electronic
accessories should not be disposed of with other household waste at the end of their working life. To
prevent possible harm to the environment or human health from uncontrolled waste disposal, please
separate these items from other types of waste and recycle them responsibly to promote the
sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this product, or their local
government office, for details of where and how they can take these items for environmentally safe
recycling.
Business users should contact their supplier and check the terms and conditions of the purchase
contract. This product and its electronic accessories should not be mixed with other commercial wastes
for disposal.