Professional Documents
Culture Documents
1
8/17/2011
2
8/17/2011
Implementation
Volumetric Modulated Arc Therapy
(VMAT) • The VMAT Report uses information from two early reports:
– Bedford and Warrington, "Commissioning of volumetric
• This report includes information for Acceptance modulated arc therapy,” Int J Radiat Oncol Biol Phys. 73, 537-45
(2009)
Testing, Commissioning, and Routine QA for
– Ling, Zhang, Archambault, Bocanek, Tang, and Losasso,
VMAT "Commissioning and quality assurance of RapidArc radiotherapy
• The Routine QA description includes a delivery system," Int J Radiat Oncol Biol Phys. 72, 575-81 (2008)
recommendation for patient-specific QA • These reports touch on the different aspects mentioned in the
measurements last slide with varying emphasis
• The aim of the TETAWG report is to provide needed details for
overall safe use of this new technology
Implementation Implementation
The TETAWG report deals with Acceptance Testing, Commissioning and
• The TETAWG report follows the two-step procedure suggested in Routine QA for MAT. This report separates, as recommended in the
the publication by Ezzell, G. A., Galvin, J. M. et al. (2003) Ezzell paper, the testing into the components listed below
"Guidance document on delivery, treatment planning, and clinical
implementation of IMRT: report of the IMRT Subcommittee of the
AAPM Radiation Therapy Committee." Med Phys 30(8) pp 2089- •Specific Component QA
115: – Follows TG142 as much as possible
(1) a group of tests that evaluate the performance of individual – Describes the testing procedures
components of the overall delivery system; and •End-to-end testing
(2) tests that evaluate the end result of the IMRT (in this case – Includes recommendations for benchmark end-to-end testing
VMAT) dose delivery. – Includes recommendations for patient-specific QA measurements
3
8/17/2011
End-to-end patient- Verify each patient’s Apply patient’s plan to generic phantom with ion Before start of all
standard dose rate.
95% points in agreement to 5% and 5 mm
specific check treatment chamber and film new plans
Note: This information is intended as an example only! Values can change during review of this report.
• Perform the asynchronous sliding window. (See the leaf Notice that the leaves hold at tra jec to ry
used for VMAT delivery, but does not include the gantry point in a detector (usually film)
motion. The tolerance values are extracted from both TG 142 sees the unobstructed beam is
(the ±1 value) and the Bedford et al reference (the ±5% intensity
constant throughout the irradiation
process. However, the holding of
change). the leaf positions at positions that
• Tolerance: ±5% intensity change relative to gantry zero baseline are spaced by 2.0 cm tests the
profile and ±1 mm from average abutment line for gantry zero ability of the leading and following
leaves to stop at the exact same
position. place.
L ea f P o sitio n (c m )
4
8/17/2011
Implementation
Specific Tests:
5
8/17/2011
Implementation
There are other ways for performing this test!
Implementation Implementation
There are other ways for performing this test!
Specific Tests:
• There are many devices available for this
type of testing for VMAT • End-to-end test for benchmark cases
• These devices will not necessarily give you
(AAPM Task Group 119)
the same information • Tolerance: 95% of points in agreement
• They must provide equivalent information to 4% and 4 mm
6
8/17/2011
Implementation Implementation
Specific Tests:
Specific Tests:
• Treatment interruption test. Use benchmark end-to-end test
• Perform patient-specific end-to-end QA that includes measurement of dose distribution and absolute
dose at a point. Interrupt beam in middle of delivery and
measurements prior to the start of treatment and for continue treatment to completion. The tolerance levels set
any plan change for this test are for comparison to test results for the same
prescription when an interruption was not introduced. Thus,
• Tolerance: 95% points in agreement to 5% and 5 the overall deviation from the calculated dose distribution is
mm on the order of 6% and 6 mm.
• Tolerance: 98% of points in agreement to 2% and 2 mm
compare with reference uninterrupted delivery
Dynamic dose
calibration test
Absolute dose accuracy for
dynamic delivery
Sliding window dose test
±4% relative to static abutted field
irradiation
Com • The synchronization of subsystems and beam
MLC calibration with
dynamic prescription
Slide-and-shoot field abutment test
±5% from baseline beam profile and
AT, Com, Annual parameters during VMAT delivery requires extra
VMAT component-
MLC calibration with
moving window
Asynchronous sliding window field abutment tests/four
±1mm from average line and for
the gantry zero position
AT, Com, Monthly
QA steps and procedures relative to standard IMRT
cardinal gantry angles, various dose rates
specific tests prescription
Field flatness/symmetry various gantry angles (four ±1% relative to baseline (defined relative
dose delivery
Beam profile stability as
function of dose rate
cardinal angles) and dose rates (including highest
and lowest used for VMAT)2
to profile at gantry zero position
and standard dose rate)
AT, Com, Annual
• The TETAWG report on VMAT Acceptance
VMAT rotational
accuracy test
Test of VMAT component
synchronization with
gantry rotation
Rotational Accuracy Test ±10% local dose at periphery AT, Com, Monthly Testing, Commissioning and Routine QA describes
End-to-end data Data transfer among various
Benchmark end-to-end test including dose measurement
95% of points in agreement to 3% and 3
AT, Com, Monthly
the additional tests needed for this new treatment
transfer check systems mm
End-to-end patient- Verify each patient’s Apply patient’s plan to generic phantom with ion Before start of all
95% points in agreement to 5% and 5 mm
specific check treatment chamber and film new plans
Note: This information is intended as an example only! Values can change during review of this report.
7
8/17/2011
Thank You