You are on page 1of 7

No.

2A701-626EN*B

SPECIFICATIONS
FOR
ENDOCAVITARY TRANSDUCER

MODEL PVT-781VTE
(2A701-626EN*B)

©TOSHIBA MEDICAL SYSTEMS CORPORATION 2015-2017


ALL RIGHTS RESERVED

Applicable System
Production Schedule : 2017-03
2A701-626EN*B

Trademarks
CIV-FlexTM is a trademark of CIVCO Medical Solutions.

IMPORTANT!
1. No part of this manual may be copied or reprinted, in whole or in part,
without written permission.
2. The contents of this manual are subject to change without prior notice
and without our legal obligation.

2
2A701-626EN*B

1. OVERVIEW
1.1 Scope
These specifications describe the endocavitary transducer model PVT-781VTE for use in combination
with Toshiba diagnostic ultrasound systems.

1.2 Purpose and Application


This transducer is designed for transvaginal and transrectal examinations.

Intended use : Imaging of organs and localization of lesions, flow


velocity measurement and mapping
Indications for use : Transvaginal examination
Transrectal examination
Target organs : Uterus, ovaries, fetus, and pelvic region
Prostate gland, seminal vesicles, bladder
Method of application : Insertion into the vagina
Insertion into the rectum

Exercise extreme care when performing Doppler or color Doppler


obstetrical and gynecological examinations so as to avoid any exposure to
the fetus.

In vitro fertilization is considered to be an investigational application.

1.3 Composition

(1) Transducer : 1
(2) Carrying case : 1
(3) Cable clamp : 1
(4) Operation manual : 1
(5) Guidelines for cleaning, : 1
disinfection, and sterilization

3
2A701-626EN*B

2. SPECIFICATIONS
2.1 Performance

(1) Frequency : 7.0 MHz


(2) Range of frequency : 3.0 MHz to 10.0 MHz
(3) Focal length : 35 mm
(4) Viewing angle : Approximately 180°
(5) Number of elements : 150 elements

2.2 External Dimensions and Mass

(1) Insertion tip


Length : 140 mm
Maximum diameter : φ23.8 mm
(2) Transducer length : 300 mm
(3) External dimensions of other components : See the outline drawing.
(4) Mass : Approximately 970 g

2.3 Compatible Systems

• TUS-AI900
• TUS-AI800
• TUS-AI700
• TUS-AI600
• TUS-A500
• TUS-A400
• TUS-A300

2.4 Operating Conditions

(1) Ambient temperature : 5°C to 35°C


(2) Relative humidity : 30% to 85% (no condensation)
(3) Atmospheric pressure : 700 hPa to 1060 hPa

2.5 Storage and Transportation Conditions

(1) Ambient temperature : -10°C to 50°C


(2) Relative humidity : 30% to 95% (no condensation)
(3) Atmospheric pressure : 700 hPa to 1060 hPa

4
2A701-626EN*B

2.6 Conformance Standards

(1) General* : IEC 60601-1 (1988), Amd. 1 (1991), Amd. 2 (1995)


IEC 60601-1 (2005)
(2) Particular* : IEC 60601-2-37 (2001), Amd. 1 (2004), Amd. 2 (2005)
IEC 60601-2-37 (2007)

* This product is compliant with at least one of the general standards above, and with at least one
of the particular standards above depending on the date of shipment of the diagnostic
ultrasound system used in combination.

2.7 Available Biopsy Adaptor

UAGV-035A
Manufacturer : Toshiba medical systems corporation

Disposable Endocavity Needle Guide


Manufacturer : CIVCO Medical Solutions
Part number : 680-121

2.8 Available Transducer Sheath


Latex-Free CIV-FlexTM Covers

Manufacturer : CIVCO Medical Solutions


Part number : 610-006

5
2A701-626EN*B

3. OUTLINE DRAWING

PVT-781VTE

(Unit: mm)

Permissible dimensional tolerance: ±10% unless otherwise specified in the dimensional outline
drawings.

6
LEGAL MANUFACTURER

1385, SHIMOISHIGAMI, OTAWARA-SHI, TOCHIGI-KEN 324-8550, JAPAN