You are on page 1of 2

LOGIQ E9 R3.1.3 or Later Removable Trackball Cable Compatibility 5749846-100 Rev.

2 1

Removable Trackball and Cables Kit P/N 5737554 Part Number Description
A 5207000-94 Trackball Interface Cable (Hirose version)*
5207000-x family
B 5497858-45 Trackball Interface Cable (Hirose version)*
5208000-x family
C 5393439-11 or Trackball, removable style
5393439-12
n/a 5749846-100 LOGIQ E9 R3.1.3 or Later Removable
Trackball Cable Compatibility Quick Card (this
document)
* 2 straps included, but not shown.
NOTE: BECOME FAMILIAR with the new connectors and compatibility BEFORE
installing this kit or damage to the connectors can occur (see Removable Trackball Cable
Compatibility on page 2 of 2).

Installation Procedures
For replacing the trackball, see section 8-6-21 “Lower Operator Panel (OP) replacement” and 8-6-26 “Trackball Assembly replacement“ located in Chapter 8 of the
• LOGIQ E9 Proprietary Service Manual, Direction 5180483-100, Revision 8, or
• LOGIQ E9 Basic Service Manual, Direction 5573152-100, Revision 3.

3. Use a small flat head screwdriver to completely seat the


connector. The connector will be seated when you hear a click at
each edge of the connector.
Re-install the trackball following the procedures in the
Proprietary Service Manual or the Basic Service Manual.
1. Choose the proper cable for the operator panel and insert as 2. Continue to gently insert the trackball end of the cable into the
shown. trackball. Note the orientation of the cable.

© 2016-2021 by General Electric Co.


LOGIQ E9 R3.1.3 or Later Removable Trackball Cable Compatibility 5749846-100 Rev. 2 2

Removable Trackball Cable Compatibility

Old Removable Trackball New Removable Trackball

You might also like