CODE SYMBOLS SCOPE OF DELIVERY
1 2 3 4 5 6 7 8 9 10 11 12
ZA J 5 - Description Main unit: Analyzer unit x 1, 2 of fuses, one of in-
Measuring range struction manual, one of accessory kit
4 Y 2 range selectable (one of restrictor, six of packings, two of
Minimum range 2%O2 filters, two of spanners)
Maximum range 100%O2
Items to be prepared separately:
Range setting before delivery 5/25 O2
range
Standard gas for calibration
(Refer to "Measuring range" in SPECIFI- Code symbol ZBM SY04-01
CATION) Reference gas
Response time • Measuring range 0 to 2 ... 100%O2
2 Standard type (2sec/90% of full scale) N2 : Code symbol ZBM NY04-01
(5% or more O2 of full scale) • Measuring range 21 to 23 ... 100%O2
Power supply and pipe connection 21 to 19 ... 0%O2
dimension Air : Code symbol ZBM RY04-01
5 85 to 264V AC 50Hz/Rc1/4 or instrumentation air
6 85 to 264V AC 60Hz/Rc1/4 • Measuring range 100 to 98 ... 0%O2
7 85 to 264V AC 50Hz/NPT1/4
O2 : Code symbol ZBM SY04-01
8 85 to 264V AC 60Hz/NPT1/4
Buffer tank ZBH410 3
Use
0 For general service
Note
Display
N Japanese (1) When using this analyzer, be sure to provide a buffer
E English
tank in the exhaust path immediately after the analyzer.
Option This is required for preventing output fluctuation due to
Y Without
disturbances such as large noise and strong wind.
A Automatic calibration
B Alarm output, remote range, range
(2) In case of air-suppletion range, span gas concentration
identification should be ordered by specifying 80 to 100% of the full
C Transmission function (RS-232C) scale. At this time, clearly describe "manufacture ac-
D A+B cording to weighing method and accuracy within O2 con-
E A+C centration corresponding to ± 1% of full scale."
F B+C Otherwise, concentration error will increase.
G A+B+C
[Example] In case of 21 to 19%O2 range:
Construction Span gas concentration
A Desk-top type 19.0 to 19.4%O2/N2, manufacture accord-
B 19 inches rack mount type
ing to weighing method, accuracy within
C Panel flush mount type
± 0.02%O2.
(3) In case of oxygen-suppletion range, use O2 gas of purity
99.99% or more for zero gas in ranges of 10% or less O2
full scale, and that of purity 99.9% or more in other ranges.
(4) Use a copper pipe (6/4 copper pipe) for the reference
line. It should be as short as possible from the gas cyl-
inder. If it is long, it takes time until it is stabilized after
CONSTITUTION power ON. Do not use a plastic pipe, as it causes errors
in reading due to permeantion of oxygen.
Sampling system diagram (Example)
Buffer tank
Flow checker
Gas sampler Membrane
filter
Atmosphere
Pump
Electromagnetic
valve
Peltier gas
Safety drain cooler Span Zero Reference
trap gas gas gas
cylinder cylinder cylinder
Regulation valve ZBM ZBM ZBM
Drain
Note) Polyethylene tude
Copper pipe Ø6/Ø4
Exhaust
3
ZAJ
ZAJ
CO2
220
146
Supporter
Power switch Operation panel (Rack mounting plate) • Supporter more than 70% of the instrument mass
at the bottom of the case.
< Front view > < Side view >
Terminal block 2
436 +20
RS232C connector
Power supply
FG terminal