4.3.

3 HbA1c Correlation (Sebia Capillarys 3 Octa (994) VS Sebia Capillarys 3 Octa (1016) )

Capillarys 3 Octa 994

Capillarys 3 Octa 1016

GRAPH 2: HbA1c Correlation between Sebia Capillarys 3 Octa (994) VS Sebia Capillarys 3 Octa
(1016)
y = -0.0319 + 1.0043 x, slope = 1.000, R2= 0.9991, n =44
Bias of clinical decision limit of 6.5%.
[(y-x)/x) *100], ‘x’: clinical decision limit
=0.15%

Bias clinical decision limit of 7.5%.


[(y-x)/x) *100], ‘x’: clinical decision limit
=0%

Medical Y -0.0319 + Allowable


Decision Level 1.0043 x Bias %Bias Performance Conclusion
(X) NGSP Level 1
6.5 6.49 0.01 0.15 1.5 Acceptable
7.5 7.5 0 0 Acceptable
Capillarys 3 Octa 994 & 1016

Mean of Capillarys 3 Octa 994 & 1016

FIGURE 1: Bland-Altman Plot for Sebia Capillarys 3 Octa (994) VS Sebia Capillarys 3 Octa (1016)
There is no pattern bias observed, all data scattered evenly across the concentration. The mean bias for
Sebia C3O 994 and and Sebia C3O 1016 analyzer for Hemoglobin A1c measurement is 0 with 95%
confidence interval (-0.15 to 0.15).