NC 59 of 64 - HbA1c
NC 59 of 64 - HbA1c
3 HbA1c Correlation (Sebia Capillarys 3 Octa (994) VS Sebia Capillarys 3 Octa (1016) )
GRAPH 2: HbA1c Correlation between Sebia Capillarys 3 Octa (994) VS Sebia Capillarys 3 Octa
(1016)
y = -0.0319 + 1.0043 x, slope = 1.000, R2= 0.9991, n =44
Bias of clinical decision limit of 6.5%.
[(y-x)/x) *100], ‘x’: clinical decision limit
=0.15%
FIGURE 1: Bland-Altman Plot for Sebia Capillarys 3 Octa (994) VS Sebia Capillarys 3 Octa (1016)
There is no pattern bias observed, all data scattered evenly across the concentration. The mean bias for
Sebia C3O 994 and and Sebia C3O 1016 analyzer for Hemoglobin A1c measurement is 0 with 95%
confidence interval (-0.15 to 0.15).