You are on page 1of 3

Registration Number ECN-000794

Revision Number B

Subject Release of DRYPIX 6000 Application Software V10.3

Date 2018/07/27

Department Medical Systems Business Division

Originator SHIGEO MATSUMOTO

Edited by NAOKO SHIMOKAWA

Approved by MASAHIRO MATSUSHITA

Affected Products
DRYPIX6000

Scope of Application

Content

Applicable To (Serial Nos. Etc.)

Design Change Reason


Serviceability
Content

Action
Not Required(infomation only)
Content

Time Frame
Not Applicable
Content

Completion Deadline
N/A

Notes

Estimated Work Time

ECN-000794 1/ 3
Implementation H
Completion H

Charge
Free of Charge

Affected Documents
No
Content

Description Of Change

Content
DRYPIX 6000 application software V10.3 is released.
Name : DRYPIX 6000 (FF) application software V10.3
No.  : 114Y2298100A03
■ For the units having the following serial number, V10.3 is installed at the factory.
S/N 86539319~86539321,86639513 or later
CN: S/N xxxxxxxx
■ Installation/check procedure
See the Service Manual.
This software cannot be installed onto devices having the CPU and PRN boards (V1.0 to V2.4).
Install this software onto devices having the CPR board (V10.0 or later).

Supply Of The Parts


No
When
with this ECN

Name/Code
DRYPIX 6000 (FF) application software V10.3
114Y2298100A03
Quantity

Return Of The Replaced Parts To FUJI


No

Special Tools / Instruments


No
Content

Details

ECN-000794 2/ 3
This software provides the following changes:
No. 1:
More robust CAN communication (countermeasure against suspension of Linux OS due to noise and other
problems)
No. 2:
Correction of processing time for outputting images by the current model
No. 3:
Countermeasure against the phenomenon in which the image for the last test printing is printed on an
output cleaning film

*For details, see the attachment.

(For US Only)C&R Assessment

SIAS Published
WW
Content

ECN-000794 3/ 3