You are on page 1of 7

Paraproteins Are a Common Cause of Interferences With

Automated Chemistry Methods


Yusong Yang, MD, PhD; Peter J. Howanitz, MD; Joan H. Howanitz, MD; Helen Gorfajn, MS; Keung Wong, MS

● Context.—Previous studies have shown that paraproteins to 4 of the 6 assays. Paraprotein interferences caused spu-
caused spurious results on individual analytes including to- riously high levels of TBIL in 4 sera measured by the MOD-
tal bilirubin (TBIL), direct bilirubin (DBIL), or HDL-choles- ULAR. In contrast, paraprotein interferences on DBIL were
terol (HDL-C). Studies demonstrating paraprotein interfer- observed by at least 1 method in 44% (39/88) of sera as-
ences with multiple analytes measured by different analyz- sayed, occurring almost exclusively with the AU2700. Para-
ers have not been reported. protein interferences with HDL-C results were present in
Objective.—To systemically investigate interferences of 35% of specimens assayed with the MODULAR and 16%
paraproteins on TBIL, DBIL, and HDL-C measured by the of specimens assayed with the AU2700. In specimens with
Roche MODULAR and the Olympus AU2700. interferences, spuriously low AU2700 DBIL, MODULAR
Design.—Eighty-eight serum specimens with monoclo- HDL-C, and AU2700 HDL-C results occurred with 28%,
nal gammopathies were analyzed using the Roche MOD-
90%, and 91% of specimens, respectively.
ULAR and the Olympus AU2700. Paraprotein interferences
with the MODULAR and AU2700 were identified by ab- Conclusions.—We demonstrated that paraprotein inter-
normal absorbance curves and confirmed by results from ferences with TBIL, DBIL, and HDL-C are relatively com-
the Ortho Vitros 950 or inconsistent laboratory informa- mon and provided explanations why these interferences
tion. occurred. Although it is difficult to predict which speci-
Results.—Spurious results occurred in 89 of 528 mea- mens cause interferences, spurious results appeared meth-
surements; 29 specimens did not demonstrate any inter- od and concentration dependent.
ferences whereas 26 specimens gave spurious results in 2 (Arch Pathol Lab Med. 2008;132:217–223)

M onoclonal immunoglobulins, also known as parapro-


teins, have been reported to interfere with a num-
ber of routine chemistry methods including analyses for
88 serum specimens collected from patients with mono-
clonal gammopathies and measured using 2 commonly
used automated chemistry analyzers.
creatinine,1,2 C-reactive protein,3 glucose,4 iron,5,6 inorganic
phosphate,7–9 and uric acid.10 Recent studies have shown MATERIALS AND METHODS
that paraproteins caused false-positive results for total bil- This study was conducted as part of a laboratory performance
irubin (TBIL)11–13 and direct bilirubin (DBIL)14 and false- improvement program for patient result reporting. Serum spec-
negative results for HDL-cholesterol (HDL-C).12,15–19 The imens were collected from patients with monoclonal gammopa-
frequency of paraprotein interferences with these analytes thies determined by serum protein electrophoretic and immu-
on automated chemistry analyzers was reported to be very nofixation studies during a 1-year period. The sera were free of
rare,12–14 and comparative studies of paraprotein interfer- hemolysis, icterus, and lipemia by routine index checks on au-
ences with TBIL, DBIL, and HDL-C have not been report- tomated analyzers. None of the specimens were duplicates or
ed with major automated chemistry analyzers. These in- triplicates; if they were from the same patients, they were ob-
terferences may cause incorrect diagnoses, inappropriate tained at different times.
Serum protein electrophoretic and immunofixation studies
treatments, longer hospital stays, and increased morbidity were performed using the SPIFE 3000 (Helena Laboratories,
in patients who already may be gravely ill. We systemi- Beaumont, Tex), and serum total protein concentrations were de-
cally investigated the potential interference of parapro- termined with the Roche P800 MODULAR (Roche Diagnostics,
teins on measurements of TBIL, DBIL, and HDL-C using Indianapolis, Ind). Paraprotein concentrations were quantitated
by the Quickscan 2000 (Helena Laboratories).
Serum levels of TBIL, DBIL, and HDL-C were measured in-
Accepted for publication October 12, 2007.
dependently using the Roche P800 MODULAR, Olympus
From the Department of Pathology, State University of New York
AU2700 (Olympus Diagnostics, Melville, NY), and Vitros 950 (Or-
Downstate Medical Center, Brooklyn, NY (Drs Yang, P. J. Howanitz,
and J. H. Howanitz, and Ms Gorfajn); and Institute of Pathology, Clin- tho-Clinical Diagnostics, Rochester, NY) automated chemistry an-
ical Chemistry Laboratory, Kings County Hospital Center, Brooklyn, NY alyzers.
(Mr Wong). Paraprotein interferences with the Roche MODULAR and
The authors have no relevant financial interest in the products or Olympus AU2700 were identified by an abnormal reaction ab-
companies described in this article. sorbance curve obtained directly from the reaction monitor in-
Reprints: Yusong Yang, MD, PhD, Department of Pathology, Box 24, tegrated into the automated analyzers and confirmed by at least
State University of New York Downstate Medical Center, 450 Clarkson 1 additional finding. Absorbance curve data from the Olympus
Ave, Brooklyn, NY 11203 (e-mail: yusong.yang@downstate.edu). AU2700 printouts were plotted manually, whereas similar data
Arch Pathol Lab Med—Vol 132, February 2008 Paraprotein Interferences With Automated Methods—Yang et al 217
Figure 1. Scatter plot of total bilirubin (TBIL)
concentrations measured by 2 automated
chemistry analyzers. 䢇, Specimens without
interferences by paraproteins; 䉭, specimens
displaying interferences by paraproteins mea-
sured by the Roche MODULAR. * To convert
to SI units in micromoles per liter, multiply
values in milligrams per deciliter by 17.104.

from the Roche MODULAR were used directly. Because it had mined by the absorbance after the addition of R2 subtracted from
been shown that the Vitros 950 did not provide spurious results the baseline absorbance in the presence of R1.
for DBIL and TBIL, it was used for confirmation of interferenc-
es.12,14,16 The Ektachem slide method used in the Vitros 950 was RESULTS
interference free because the specimens had to penetrate through The present study included 88 serum specimens from
several layers before reaching the reaction layer, and the interfer- 52 patients with monoclonal gammopathies treated at a
ing compounds were probably filtered out.20 Therefore, when po-
tential interferences were identified and an adequate specimen
major municipal hospital or a university hospital. The
remained, confirmation of the spurious results was made by ad- paraprotein concentrations of these specimens ranged
ditional measurements using the Vitros 950, except for HDL-C, from 0.5 to 8.9 g/dL with a mean of 3.67 g/dL. The ma-
as our instrument did not have this method included in its test jority (60%) of the paraproteins were immunoglobulin (Ig)
menu. Paraprotein interferences also were confirmed by speci- G ␬, 18% were IgG ␭, 11% were IgA ␭, and 7% were IgM
mens with negative DBIL or negative HDL-C results, specimens ␬. Specimens with paraprotein concentrations less than 3.0
with abnormally high TBIL levels from nonicteric patients with- g/dL had spurious results in 7.2% of assays; when para-
out evidence of liver diseases and hemolysis, or specimens with protein concentrations were 3.0 to 5.0 g/dL, spurious re-
DBIL levels higher than TBIL levels. sults were seen in 22.4% of assays, and when paraprotein
The Roche MODULAR TBIL assay is a chromogenic 2-point
concentrations were greater than 5.0 g/dL, spurious re-
end point assay modified from the diazo method21 that is per-
formed in 1 cuvette. First the patient’s serum is incubated with sults were seen in 31.3% of assays.
reagent 1 (R1), which contains surfactant to accelerate the reac- Spurious results occurred in 89 of 528 measurements;
tion and to solubilize protein. Then, reagent 2 (R2) containing a 29 specimens did not demonstrate any interferences,
diazonium salt and HCl is added to produce azobilirubin, which whereas 26 specimens gave spurious results in 2 to 4 of
is red under acidic conditions. Before R1 is added, the serum the 6 assays. Of 88 serum specimens, 4 specimens (4.3%)
specimen should not be turbid. In contrast, the Olympus AU2700 from 3 patients with monoclonal gammopathies displayed
TBIL assay is a 2-cuvette, 1-point end point assay that uses sim- spuriously high Roche MODULAR TBIL concentrations.
ilar reagents. Both the reaction cuvette and the blank cuvette The interferences in 3 specimens initially were suggested
contain surfactant, whereas the chromogen is not added to the by nonicteric sera with abnormally high TBIL values of
blank cuvette.
The Roche MODULAR and Olympus AU2700 DBIL assays re-
greater than 25 mg/dL (428 ␮mol/L), (Figure 1) and sub-
semble their manufacturers’ TBIL assays, other than that a solu- sequently were identified and confirmed along with a
bilizing agent such as surfactant is not added. When paraprotein fourth specimen by abnormal reaction absorbance curves
interferences with DBIL initially were suspected in the specimens (Figure 2, C and D) compared with those obtained from
with either negative DBIL values or DBIL concentrations higher normal control serum (Figure 2, A), measurements by the
than TBIL concentrations, they subsequently were confirmed by Vitros 950, and icteric serum without measurable para-
abnormal reaction absorbance curves. proteins (Figure 2, B). In our study, the interferences were
The Roche HDL-C Plus second-generation assay and the Olym- found exclusively in specimens from patients with IgG ␬
pus HDL-C assay are 2-reagent homogenous systems for the se- or IgG ␭ monoclonal gammopathies, and in these 4 spec-
lective measurement of serum HDL-C in the presence of other
imens, the paraprotein concentrations (mean, 5.45 g/dL)
lipoprotein particles. Reagent 1 (R1) selectively forms colorless
water-soluble complexes with non-HDL lipoprotein, which is pro- were higher than the overall mean of the 88 monoclonal
tected from subsequent enzymatic reactions. The HDL-C is then proteins studied.
determined enzymatically by the addition of R2 containing cho- When we manually performed the Roche MODULAR
lesterol esterase, cholesterol oxidase, and a chromogen system TBIL assay in test tubes according to manufacturer’s sug-
into the reaction mixture. The HDL-C concentration is deter- gested protocol using a specimen with the suspected in-
218 Arch Pathol Lab Med—Vol 132, February 2008 Paraprotein Interferences With Automated Methods—Yang et al
Figure 2. Reaction absorbance curves of the
Roche MODULAR total bilirubin (TBIL) assay
monitored. A, Normal control serum. B, An
icteric patient serum without paraprotein. C
and D, Nonicteric, nonhemolytic serum spec-
imens containing paraproteins. The gray ver-
tical bars indicate the points at which mea-
surements are taken to calculate results. IgG
indicates immunoglobulin G.

terference, the specimen was slightly turbid visually be- DBIL when measured by the Roche MODULAR, and this
fore adding R1 but became clear after the addition of R1. result was confirmed as spurious by the Vitros 950.
However, the addition of R2 caused the reaction mixture We found that these interferences occurred when the
to become clear instead of the red translucent appearance reaction absorbance curve of the blank cuvette (DBIL-B)
observed with the control patient’s specimens. When the did not parallel the absorbance curve of the reaction cu-
TBIL concentrations of these sera were measured indepen- vette (DBIL-C) (Figure 4). The Olympus instrument was
dently by the Olympus AU2700 and Vitros 950 automated programmed to subtract the absorbance of the blank cu-
chemistry analyzers, no interferences were identified with vette from the absorbance of the reaction cuvette at the
these 2 instruments. end point, then tally the difference and convert this dif-
Using the criteria we established for interference, 39 ference into a numerical result. If the absorbance of the
(44%) of 88 specimens were found to display interferences blank cuvette was lower than the end point absorbance of
with DBIL on the Olympus AU2700 (Figure 3). However, the reaction cuvette, the instrument reported a positive
only 1 specimen was found to have interferences with value (Figure 4, A); in contrast, if the absorbance of the
Arch Pathol Lab Med—Vol 132, February 2008 Paraprotein Interferences With Automated Methods—Yang et al 219
Figure 3. Scatter plot of direct bilirubin
(DBIL) measurements made by 2 automated
chemistry analyzers. 䢇, Specimens without
evidence of paraprotein interferences; 䉭,
specimens displaying interferences by para-
proteins measured by the Olympus AU2700;
䡵, specimens displaying interferences by
paraproteins measured by the Olympus
AU2700 and the Roche MODULAR. * For SI
units in micromoles per liter, multiply results
in milligrams per deciliter by 17.104.

blank cuvette was higher than the end point absorbance


of the reaction cuvette, the instrument reported a negative
value (Figure 4, B).
Among the 39 specimens with Olympus AU2700 DBIL
interferences, 14 specimens had spuriously high values
and 25 specimens had spuriously low values. However, the
specimens displaying interferences with the Olympus
AU2700 DBIL method tended to give irreproducible DBIL
values, as demonstrated by a specimen that varied from
falsely positive (13.0 mg/dL [222 ␮mol/L]) to falsely neg-
ative (⫺17.3 mg/dL [296 ␮mol/L]) values when measured
repeatedly (data not shown).
Figure 5 depicts 31 (35%) of 88 specimens that demon-
strated HDL-C interferences with the Roche MODULAR,
and of these, 90% were spuriously low. In contrast, 14
(16%) of 88 specimens demonstrated HDL-C interferences
when measured by the Olympus AU2700. Of these inter-
fering sera, 7 demonstrated interferences with both meth-
ods.
Examples of absorbance curves demonstrating interfer-
ences with HDL-C as identified by abnormal reaction ab-
sorbance curves compared with those from control spec-
imens and confirmed by negative HDL-C values are
shown in Figure 6. In Figure 6, B and C, the Roche MOD-
ULAR absorbance values of the blank obtained just before
the addition of R2 were higher than after the addition of
R2, thereby giving spurious results.
There was a much higher baseline absorbance before the
addition of R2 in the specimens displaying interferences
with the Olympus AU2700 HDL-C method compared


Figure 4. Absorbance curves from the Olympus AU2700 direct bili-
rubin (DBIL) assay. A and B, Abnormal reaction absorbance curves of
the Olympus AU2700 DBIL assay from 2 patient specimens containing
paraproteins. The DBIL assay is a 2-cuvette, end point assay utilizing
a sample blank; the sample blank cuvette (DBIL-B) absorbance is sub-
tracted from the color cuvette (DBIL-C) absorbance to determine the
net reaction absorbance. The arrows indicate reaction end point.

220 Arch Pathol Lab Med—Vol 132, February 2008 Paraprotein Interferences With Automated Methods—Yang et al
Figure 5. Scatter plot of high-density lipo-
protein cholesterol (HDL-C) measurements
made by 2 automated chemistry analyzers.
䢇, Specimens without evidence of parapro-
tein interferences; 䉱, specimens displaying
interferences by paraproteins measured by
the Roche MODULAR; 䡵, specimens dis-
playing interferences by paraproteins mea-
sured by the Olympus AU2700; #, speci-
mens displaying interferences by paraproteins
measured by the Roche MODULAR and the
Olympus AU2700. * For SI units in millimoles
per liter, multiply results in milligrams per
deciliter by 0.0259.

with that in the control specimens; after the addition of A recent study showed that paraprotein caused only
R2, absorbance values did not increase as much as ex- spuriously high levels with Olympus AU2700 DBIL mea-
pected (Figure 6, D), thereby accounting for the spuriously surements.14 However, in contrast, our study demonstrat-
low values of HDL-C. Visually checking cuvettes contain- ed that most DBIL interferences (64%) were spuriously
ing specimens with the interference also demonstrated low when measured by the Olympus AU2700 DBIL. Be-
white precipitates that were formed after the addition of cause the Olympus DBIL assay gave exceedingly variable
R1. repetitive results ranging from a markedly positive to
The paraprotein interferences appear to be concentra- markedly negative interferences in specimens containing
tion dependent, as dilution of patient sera and treatment a monoclonal protein, we believe that describing an inter-
of the patient by therapeutic plasmapheresis or chemo- ference as negative or positive is unsatisfactory.
therapeutic agents causing decreasing monoclonal pro- The paraprotein interferences with TBIL, DBIL, and
teins minimized paraprotein interferences (data not HDL-C appeared to be method dependent. The interfer-
shown). ences with TBIL and HDL-C mainly were observed with
the Roche MODULAR, while DBIL interferences occurred
COMMENT most frequently with the Olympus AU2700.
Our studies demonstrate that the interferences with The Roche TBIL assay is a 1-cuvette, 2-point end point
TBIL, DBIL, and HDL-C are relatively common in the assay. Because the addition of R2 to the reaction mixture
presence of paraproteins. Comparing results among dif- changed the assay conditions to more acidic (pH 1–2),
ferent analyzers and monitoring absorbance curves led us paraproteins could form insoluble complexes with R2,
to conclude that paraprotein interferences are more prev- which cause the interferences. On the other hand, the
alent than previously reported. Pantanowitz and cowork- Olympus TBIL assay is a 2-cuvette end point assay. Both
ers11 found that artifactual TBIL occurred in 2 (1.2%) of cuvettes, the sample blank cuvette and the reaction cu-
115 patient sera, whereas Nauti et al14 reported that para- vette, contain the same reagents, except the chromogen is
protein interferences with the Olympus AU2700 DBIL only in the reaction cuvette. Therefore, the condition in the
were observed in only 3 (1.5%) of 200 serum specimens. assay is not changed as there is no addition of R2.
Spuriously low levels of HDL-C on the Roche 917 analyzer The chromogenic reaction of the DBIL assay occurs at a
caused by paraprotein were found in 2 (13%) of 15 spec- very low pH (⬍1). A ‘‘protein stabilizing agent’’ is includ-
imens at a major medical center.12 To our knowledge, para- ed to avoid protein precipitation. It appears that the sol-
protein interferences with HDL-C measured by the Olym- ubilization capacity of the Olympus AU2700 protein sta-
pus automated chemistry analyzers have not been report- bilizing agent is not sufficient to prevent the proteins in
ed. Our study revealed that the incidence of paraprotein solution from precipitating at such an acidic pH, which
interference with the Olympus AU2700 and Roche MOD- may cause more interferences with DBIL on the Olympus
ULAR DBIL assays was 16% and 35%, respectively, in 88 AU2700.
serum specimens. We also found that paraproteins inter- As mentioned in the ‘‘Results,’’ spuriously low HDL-C
fere with Olympus AU2700 HDL-C, although the frequen- results observed in the patients with monoclonal gam-
cy (16%) was not as high as that on the Roche MODULAR mopathies on the Roche MODULAR and the Olympus
(35%). Specimens that gave spurious results in one assay AU2700 may be due to the high baseline absorbance in
had a high likelihood of yielding a spurious result in other the presence of R1 reagent. The R1 reagent contains a
assays. buffer system to selectively form a water-soluble complex
Arch Pathol Lab Med—Vol 132, February 2008 Paraprotein Interferences With Automated Methods—Yang et al 221
Figure 6. Comparison of high-density lipo-
protein cholesterol (HDL-C) measurements
made by 2 automated chemistry analyzers;
the Roche MODULAR (A through C) and the
Olympus AU2700 (D) HDL-C assay absor-
bance curves. A, A control patient serum. B
and C, Serum specimens containing parapro-
teins. The gray vertical bars indicate the
points at which measurements were taken to
calculate results. D, Reaction absorbance
curve of the Olympus AU2700 HDL-C assay.
The arrows indicate photometric points at
which R1 (point 0) and R2 (point 10) reagents
were added. The reaction stops at photomet-
ric point 27. IgG indicates immunoglobu-
lin G.

with low-density lipoprotein, very low-density lipopro- creased paraprotein concentrations caused interferences to
tein, and chylomicrons. It may be that the differences of disappear, while disease progression and increasing
the components or concentrations of the components of monoclonal protein concentrations caused interferences to
the buffer system in the Roche MODULAR and Olympus increase. It is important to appreciate the occurrence of
AU2700 HDL-C assays caused more interferences with these interferences to avoid misinterpretation of the re-
HDL-C to be observed on the Roche MODULAR. sults, especially because low HDL-C (⬍40 mg/dL [684
Our studies demonstrate that paraprotein interferences ␮mol/L]) is considered to be an independent risk factor
with DBIL and HDL-C are relatively common, and spu- for coronary heart disease.22 To automatically detect the
rious results of TBIL, DBIL, and HDL-C appear to be interference, Smogorzewska et al12 suggested manually
methodology and concentration dependent. In a few pa- programming the analyzer to trigger a flag when the dif-
tients followed for a period of time, treatment that de- ference of absorbance between certain points was above a
222 Arch Pathol Lab Med—Vol 132, February 2008 Paraprotein Interferences With Automated Methods—Yang et al
preset value for the Roche MODULAR TBIL assay or the 5. Dorizzi R, Battaglia P, Lora A. Iron measurement in patients with monoclo-
nal immunoglobulin: a further caution. Clin Chem. 1991;37:589–590.
absorbance immediately before the addition of the R2 6. Bakker AJ. Influence of monoclonal immunoglobulins in direct determina-
agent was above a designated value for the Roche AU2700 tions of iron in serum. Clin Chem. 1991;37:690–694.
HDL-C assay. This may help when the abnormal absor- 7. Sinclair D, Smith H, Woodhead P. Spurious hyperphosphataemia caused by
an IgA paraprotein: a topic revisited. Ann Clin Biochem. 2004;41:119–124.
bance reaction curves follow a similar kinetic pattern. 8. Barutcuoglu B, Parildar Z, Mutaf I, Habif S, Bayindir O. Spuriously elevated
However, our studies showed that the abnormal reaction inorganic phosphate level in a multiple myeloma patient. Clin Lab Haematol.
curves of the Roche MODULAR TBIL assay and the Olym- 2003;25:271–274.
9. Savory DJ, Pearce CJ. Paraprotein interference causing pseudohyperphos-
pus AU2700 DBIL assay have different patterns (Figures 2 phataemia: evaluation of an improved methodology. Ann Clin Biochem. 1995;
and 4), which may lead to missing some interferences if a 32:498–501.
programmable rule is used. An approach that we favor for 10. Langman LJ, Allen LC, Romaschin AD. Interference of IgM paraproteins in
bilirubin assays is to program the analyzer to automati- the Olympus AU800 uric acid assay. Clin Biochem. 1998;31:517–521.
11. Pantanowitz L, Horowitz GL, Upalakalin JN, Beckwith BA. Artifactual hy-
cally flag results for confirmation when specimens have perbilirubinemia due to paraprotein interference. Arch Pathol Lab Med. 2003;
TBIL results greater than 20 mg/dL (342 ␮mol/L) and the 127:55–59.
icteric flag is not markedly positive.13 To prevent results 12. Smogorzewska A, Flood JG, Long WH, Dighe AS. Paraprotein interference
in automated chemistry analyzers. Clin Chem. 2004;50:1691–1693.
preceded by a negative sign from being reported, we rec- 13. Sheppard CA, Allen RC, Austin GE, et al. Paraprotein interference in au-
ommend programming the laboratory information system tomated chemistry analyzers. Clin Chem. 2005;51:1077–1078.
to reject patient DBIL and HDL-C results that are preced- 14. Nauti A, Barassi A, Merlini G, d’Eril GV. Paraprotein interference in an
assay of conjugated bilirubin. Clin Chem. 2005;51:1076–1077.
ed by a minus sign. 15. Bakker AJ, Zijlstra A, Leemhuis MP. False negative results in the Roche
In conclusion, we found that paraprotein interferences assay for HDL-cholesterol. Ann Clin Biochem. 2003;40:572–575.
with clinical laboratory measurements are far more fre- 16. Kadri N, Douville P, Lachance P. Monoclonal paraprotein may interfere
with the Roche Direct HDL-C Plus assay. Clin Chem. 2002;48:964.
quent than previously reported. If an interference occurs 17. Tsai LY, Tsai SM, Lee SC, Liu SF. Falsely low LDL-cholesterol concentra-
with one assay, it is likely that other assays also will be tions and artifactual undetectable HDL-cholesterol measured by direct methods
affected. These types of interferences appear in part to be in a patient with monoclonal paraprotein. Clin Chim Acta. 2005;358:192–195.
18. Berth M, Delangehe J. Protein precipitation as a possible important pitfall
concentration dependent. Specimens with paraproteins in the clinical chemistry analysis of blood samples containing monoclonal im-
cause spurious results by causing precipitation in TBIL, munoglobins: 2 case reports and a review of the literature. Acta Clin Belg. 2004;
DBIL, and HDL-C assays. 59:263–273.
19. Baca A, Haber RJ, Susishi K, Frost PH, Ng VN. Artifactual undetectable
References HDL-cholesterol with the Beckman Synchron LX and Vitros 950 assays temporally
1. Datta P, Graham GA, Schoen I. Interference by IgG paraproteins in the Jaffe associated with a paraprotein. Clin Chem. 2004;50:255–256.
method for creatinine determination. Am J Clin Pathol. 1986;85:463–468. 20. Zaman Z, Sneyers L, Van Orshoven A, Blanckaert N, Marien G. Elimination
2. Hummel KM, von Ahsen N, Kuhn RB, et al. Pseudohypercreatininemia due of paraprotein interference in determination of plasma inorganic phosphate by
to positive interference in enzymatic creatinine measurements caused by mono- ammonium molybdate method. Clin Chem. 1995;41:609–614.
clonal IgM in patients with Waldenstrom’s macroglobulinemia. Nephron. 2000; 21. Westwood A. The analysis of bilirubin in serum. Ann Clin Biochem. 1991;
86:188–189. 28:119–130.
3. Yu A, Pira U. False increase in serum C-reactive protein caused by mono- 22. Expert Panel on Detection, Evaluation, and Treatment of High Blood Cho-
clonal IgM-lambda: a case report. Clin Chem Lab Med. 2001;39:983–987. lesterol in Adults. Executive Summary of The Third Report of The National Cho-
4. Dimeski G, Carter A. Rare IgM interference with Roche/Hitachi modular lesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and
glucose and gamma-glutamyltransferase methods in heparin samples. Clin Chem. Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III). JAMA.
2005;51:2202–2204. 2001;285:2486–2497.

Arch Pathol Lab Med—Vol 132, February 2008 Paraprotein Interferences With Automated Methods—Yang et al 223

You might also like