You are on page 1of 68

AlphaRudder

Alphaline Repeater Display


Installation and Operation Manual
alphatronmarine.com
Contents

I Preface.......................................................................................................... 4
I.i Glossary......................................................................................................................................................................4
I.i.i Definitions........................................................................................................................................................... 4
I.i.ii Abbreviations..................................................................................................................................................... 6
I.i.iii Norms and Standards.......................................................................................................................................7

II Caution........................................................................................................ 8
II.i Warranty.................................................................................................................................................................... 9
II.ii Storage..................................................................................................................................................................... 9

III Introduction.............................................................................................. 10

1 Installation Instructions........................................................................... 11
1.1 Mechanical............................................................................................................................................................. 11
1.1.1 Dimensions.................................................................................................................................................... 11
1.1.2 Mounting........................................................................................................................................................ 11
1.1.3 Fit Mounting Frame.......................................................................................................................................12
1.1.4 Fit Water Seal............................................................................................................................................... 12
1.1.5 MF Instrument electric connections.............................................................................................................. 13
1.1.6 Cable............................................................................................................................................................. 13
1.1.7 Cable preparation..........................................................................................................................................13
1.1.8 Grounding...................................................................................................................................................... 13
1.1.9 Power supply.................................................................................................................................................14
1.1.10 Serial connection.........................................................................................................................................14
1.1.11 Relay........................................................................................................................................................... 15
1.1.12 Connect NMEA............................................................................................................................................15
1.1.13 Connect Rudder Feedback and Rudder Interface......................................................................................17
1.1.13.1 Technical Specifications Interface......................................................................................................19
1.2 Software................................................................................................................................................................. 20
1.2.1 Software updates MF....................................................................................................................................20

2 Operation................................................................................................... 21
2.1 Power..................................................................................................................................................................... 21
2.2 Main screen............................................................................................................................................................21
2.3 Menu handling........................................................................................................................................................22
2.3.1 Default Values............................................................................................................................................... 23
2.4 Alert Handling.........................................................................................................................................................25
2.5 Settings...................................................................................................................................................................27
2.5.1 Generic.......................................................................................................................................................... 28
2.5.1.1 Touch screen calibration (TOUCH CAL)............................................................................................. 28
2.5.1.2 Touch screen sensitivity (TOUCH SET).............................................................................................. 29
2.5.1.3 Change Date and Time (DATE TIME).................................................................................................29
2.5.1.4 Change contrast theme (THEME)....................................................................................................... 30
2.5.1.5 Change Range Settings (RANGE SCALE)..........................................................................................31
2.5.1.6 About the MF instrument (ABOUT)..................................................................................................... 31
2.5.2 Advanced.......................................................................................................................................................32
2.5.2.1 Advanced settings password (ADV SET)............................................................................................ 32

2 | Contents
2.5.2.2 Advanced settings menu......................................................................................................................32
2.5.2.3 Central dimming (CNTRL DIM)............................................................................................................33
2.5.2.4 Ethernet configuration (ETH CONFIG)................................................................................................ 34
2.5.2.5 Serial port (UART) configuration (UART CONFIG)............................................................................. 35
2.5.2.6 Serial port monitor (SERIAL MON)......................................................................................................36
2.5.2.7 NMEA Talker........................................................................................................................................ 38
2.5.2.8 Factory reset (RESET).........................................................................................................................38
2.5.2.9 Log (LOG)............................................................................................................................................ 39
2.5.2.10 Calibrate Rudder (RUDDER SET)..................................................................................................... 40
2.5.2.11 Range Settings...................................................................................................................................41

3 Specifications............................................................................................43
3.1 Electrical................................................................................................................................................................. 43
3.2 Environmental.........................................................................................................................................................43
3.3 Mechanical............................................................................................................................................................. 43
3.4 Maintenance........................................................................................................................................................... 44

4 Appendix A................................................................................................45
4.1 Mechanical Drawings............................................................................................................................................. 45
4.1.1 MFM.............................................................................................................................................................. 46
4.1.2 Rudder Feedback MD...................................................................................................................................47
4.1.3 Rudder Feedback HD................................................................................................................................... 48
4.1.4 RFU Linkage Transmission...........................................................................................................................49
4.1.5 Analog Interface Mk2.................................................................................................................................... 50
4.2 Connection Diagrams.............................................................................................................................................51
4.2.1 MF Instrument............................................................................................................................................... 51
4.2.2 Rudder Feedback..........................................................................................................................................53
4.2.3 AlphaRudder..................................................................................................................................................55
4.2.4 Analog Interface Mk2.................................................................................................................................... 57
4.3 Data Sheets........................................................................................................................................................... 59
4.3.1 Alphaline MF................................................................................................................................................. 59
4.3.2 Rudder Feedback MD/HD.............................................................................................................................59
4.3.3 Analog Interface Mk2.................................................................................................................................... 59
4.4 Type Examination Certificates............................................................................................................................... 59
4.4.1 Alphaline MF................................................................................................................................................. 59
4.4.2 Analog Interface Mk2.................................................................................................................................... 59
4.4.3 Rudder Feedback..........................................................................................................................................59
4.5 MED Certificates Alphaline MF..............................................................................................................................59
4.5.1 Rudder Indicator............................................................................................................................................59
4.6 Thales Certificates................................................................................................................................................. 59
4.6.1 MFM.............................................................................................................................................................. 60
4.6.2 Rudder Feedback MD...................................................................................................................................61
4.6.3 Rudder Feedback HD................................................................................................................................... 62
4.6.4 Analog Module.............................................................................................................................................. 63

5 Appendix B................................................................................................64
5.1 ISO 9001 certificate Alphatron Marine R&D..........................................................................................................64

6 Appendix C................................................................................................65
6.1 EC Declaration of Conformity................................................................................................................................ 65

3 | Contents
I Preface
The Alphatron Marine MF range of instruments was designed for navigation and control of ships and is based on generic
hard- and software, allowing for many different applications.
• Thoroughly read this instruction manual before installation and operation of the equipment.
• We recommend to keep this manual nearby the equipment to ensure ready access to it.

I.i Glossary
The glossary contains a list of Definitions and a list of Abbreviations.

I.i.i Definitions
The meaning of standard definitions as used in this manual are explained in the table of Definitions.
See Table 1: Definitions on page 5

4 | Preface
Redundant A device that is equipped with multiple part of the same type, for example a double power
supply. This equipment will continue to function when one of the redundant part fails.
Heading users Navigation equipment that uses heading/course information for functioning.
Hardware The physical parts of the Alphaline MF system.
LED's Light Emitting Diodes, these are used for signaling statuses of hardware and software signals to
the user.
Central alarm System that is connected to all vital systems on a ship and that is able to give a centralized
system/Bridge watch indication of the (alarm)status of all connected systems.
monitoring
NMEA protocol Protocol standard for transmitting and receiving of asynchronous serial data sentences.
Talker Device which transmits data. This is usually called transmitter or TX.
Listener Device which receives data. This is usually called receiver or RX.
ISO GND Isolated Ground. This is a ground connection to be used for reference signal. It is different from
EARTH and should normally not be connected to EARTH.
Grounding point/ Point on the chassis of the Alphaline MF system which should be connected to the ship's mass.
stud
Printed Circuit Board A printed circuit board, or PCB, is used to mechanically support and electrically connect
electronic components using conductive pathways, or traces, etched from copper sheets,
laminated onto a non-conductive substrate.
(Galvanic)isolated Electrical separation of two circuits. There is no current flowing directly from one circuit to
another. Electrical energy and/or information can still be exchanged between the sections by
other means, such as by induction or by optical means(like a transformers or opto couplers).
CAN bus Controller Area Network. This is a network based serial bus system used for exchanging
information. It is the advanced version of RS485/422 serial busses.
Reverse polarity This is a part of the power supply hardware that prevents any damage to the equipment when
protection the power supply is connected to the wrong polarity.
ROT signal - Rate Of Turn signal indicates the course change of a ship in degrees per minute. This signal can
be analog using voltage or current, or can be an NMEA data signal.
Heading/bearing Navigation type of instruments displaying the heading/course of a ship.
repeaters
Baud rate This is the transmission speed of serial interfaces in characters per second.
Transmitting interval The frequency at which complete NMEA sentences are being transmitted in number of times per
second.
Factory setting Instrument setting for backlight color, language, number of connected apparatus, etc. as
configured as a new instrument by the factory.
Flash memory Non volatile type of memory. This type of memory retains its contents even when the instrument
is turned off. All the settings of the inter switch such as number of connected instruments, the
selected instrument, language, backlight color are stored in the flash memory of the inter switch
and the display unit.
Firmware (embedded) software inside the processors of the Alphaline MF system.
Compass safe The minimum distances to equipment that will not cause an unacceptable deviation of the ship's
distance standard and steering compasses.

Table 1: Definitions

5 | Preface
I.i.ii Abbreviations
Abbreviations as used in this manual are explained in the table of Abbreviations.Table 2: Abbreviations on page 6
A Ampere
CAN Controller Area Network
DC Direct Current
DP Dynamic Position
ECDIS Electronic Chart Display Information System
GPS Global Positioning System
I/O Inputs and Outputs
I.S. Inter Switch
LED Light Emitting Diode
mA milliamperes
MF Multi Function
mm Millimeter
NC Normally Closed
NMEA National Marine Electronics Association
NO Normally Open
OA Operational Alarm
TAP Type Approval Program
PCB Printed Circuit Board
RCU Remote Control Unit
ROT Rate Of Turn
VAC Volts Alternating Current
VDC Volts Direct Current
VDR Voyage Data Recorder
W Watt

Table 2: Abbreviations

6 | Preface
I.i.iii Norms and Standards
The whole of the Alphaline range of products comply with the applicable standards, norms and regulations.
See Figure 1: Norms and Standards on page 7

NORMS & STANDARDS

Alphaline Repeater Display MFS-H


Alphaline Repeater Display MFS-V

Alphaline Repeater Display MFM


Alphaline Repeater Display MFL

AlphaRemoteControlInterface
NMEA Distribution Interface

AlphaRudderInterface
AlphaTurnInterface

AlphaTouchPad
AlphaHeading
AlphaTMC(+)

AlphaRudder
AlphaTurn

numbers correspond with Matrix 2 6 8 7 19 20 21 22 23


IEC 60945 (2002) including IEC 60945
1 Corrigendum 1(2008) √ √ √ √ √ √ √ √ √ √ √ √ √
2 Standard DNV 2.4 √ √ √ √ √ √ √ √ √ √ √ √
3 IEC 61162 series √ √ √ √ √ √ √ √ √ √ √ √ √
4 IEC 62288 (2014) √ √ √ √ √ √ √ √
ISO 22090-2 (2004) including Corrigendum
5 2005 √
6 ISO 20673 (2007) √
7 ISO 20672 (2007) √
8 ISO 8728 (1997) √
Figure 1: Norms and Standards

7 | Preface
II Caution
To safely install and operate this instrument, so as not to adversely affect the warranty, the WARNINGS and CAUTIONS
must be adhered to.

• WARNING
• Indicates potential risk of injury or death to users of the product.
• WARNING
• Do not disassemble or modify the equipment. Failure to observe this instruction may cause a fire, electric
shock, or equipment failure.
• WARNING
• Do not insert or remove the power cord or operate switches with a wet hand. Otherwise, you may suffer an
electrical shock.
• WARNING
• Operate the equipment only at the power supply voltage of DC 12V or 24V. Failure to observe this instruction
can cause a fire, electric shock, or equipment failure
• WARNING
• Do not scratch, damage, modify, heat, pull, excessively bend, or heavily load the power supply cable. It may
cause a fire, or electric shock.
• WARNING
• Immediately turn off the power and disconnect the power supply cable if the equipment is generating any
smoke or odor, or is overheated. Immediately inform our local service agent of the symptom to have it
repaired. Prolonged equipment operation under such a condition can cause a fire or electric shock.
• WARNING
• Do not place a vessel containing liquid on the equipment. It may cause a fire, electrical shock, or a failure to
the equipment if knocked over.
• CAUTION
• Indicates potential risk of damage to equipment.
• CAUTION
• Any modification to this equipment without prior written permission from ALPHATRON MARINE will void the
warranty.
• CAUTION
• Installation of this product shall only be done by a certified installation company approved by either
ALPHATRON MARINE or by an official ALPHATRON MARINE distributor. Acting otherwise will void the
warranty.
• CAUTION
• This product must be installed in accordance with the installation methods described in this manual. Acting
otherwise will void the warranty.
• CAUTION
• This product contains no operator serviceable parts. Service and repair shall only be carried out by personnel
trained and certified by ALPHATRON MARINE.
• CAUTION
• Do not allow the Display Unit to fall or immerse into water. The equipment can be damaged.
• CAUTION
• When removing the power, be sure to remove the cord terminal correctly. If the cord is pulled, the cord may
get damaged resulting in a fire or an electrical shock.
• CAUTION
• If the instruments are not stored as described it will void the warranty.

8 | Caution
• CAUTION
• When cleaning the surface, do not use any organic solvent such as thinner or benzine. Otherwise, the paint
and markings on the surface may get damaged. For cleaning the surface, remove the dust and debris and
wipe with a clean dry cloth.

II.i Warranty
Non-compliance with the installation, operation and maintenance requirements may void the warranty.
Read the Caution requirements.
Contact the Alphatron dealer regarding the terms of the warranty.

II.ii Storage
The Alphaline range of instruments are sensitive to humidity, temperature fluctuations and aggressive substances. Store
them appropriately.
• CAUTION
• If the instruments are not stored as described it will void the warranty.

9 | Caution
III Introduction
Each type in this navigation and control instrument product range consists of a display unit and, if applicable, one or
more external remote I/O modules.
The following 6 versions are available in the Alphatron Marine MF range of instruments.

Alphaline MFS-H 5.0 inch display in ½ DIN mounting LCD orientation horizontal
Alphaline MFS-V 5.0 inch display in ½ DIN mounting LCD orientation vertical
Alphaline MFS-VJ 5.0 inch display in ½ DIN mounting LCD orientation vertical, joystick
with rotary encoder and push button
for advanced control of third-party
systems
Alphaline MFS-VR 5.0 inch display in ½ DIN mounting LCD orientation vertical, rotary
encoder and push button for
advanced control of third-party
systems
Alphaline MFM 6.5 inch display in DIN mounting LCD orientation vertical
Alphaline MFL 8.4 inch display in 1.5 DIN mounting LCD orientation horizontal

The Seagoing AlphaRudder Indicator Display is an MED Type Approved system for showing rudder angle information.
Depending on requirements, this system is expandable with a rudder feedback unit which can be displayed in the
repeater.

10 | Introduction
1 Installation Instructions
Installation follows a generic method and is applicable to the complete range of MF instruments.
This chapter describes the installation into a console.

1.1 Mechanical
Strictly follow the prescribed installation method.
• CAUTION
• This product must be installed in accordance with the installation methods described in this manual. Acting
otherwise will void the warranty.
The Location Class/Category of the Alphaline MF instruments is: EXPOSED (May be used outside)
The Alphaline MF display is supplied complete with the following parts.
• Display unit (5.0, 6.5 or 8.4 inch)
• Mounting frame with 2 sets of screws for fixing to either steel or wood.
• 3 off Phoenix connectors for power supply and signals.
• USB stick with manual.
• Mounting template.
• Grounding lug.

1.1.1 Dimensions
Carefully check the applicable drawing(s) of the Alphaline MF instrument. See Mechanical Drawings on page 45.

1.1.2 Mounting
The display unit can only be flush mounted. Carefully consider the location and angle of the display unit for maximum
visibility. Make sure that there is enough space to connect cables. The display unit can be installed horizontally, vertically
or under an angle, see Figure 2: Flush mounting on page 12
Note For outdoor fitting use correct sealing arrangement.

11 | Installation Instructions
Figure 2: Flush mounting

1.1.3 Fit Mounting Frame


Prior to fitting the display unit, install the mounting frame.

1. Make a square hole in the (overhead) console. For dimensions see drawing and use template provided. MFM on
page 46
2. Push the mounting frame into the hole and attach it with four screws.
3. Push the display unit into the mounting frame.
INFO:
Note The instrument is locked into position by a spring system.

Note Use the MF Display Overhead Mounting Kit for securing the display unit to an overhead console, to
prevent the unit from falling out.

1.1.4 Fit Water Seal


The instrument can also be positioned outside.
Care must be taken in applying the seal.

12 | Installation Instructions
1.1.5 MF Instrument electric connections
All six MF instrument versions share the same electronics with identical connections.
For pinouts see andTable 4: Serial connections 8 pin on page 15 and Table 5: Serial connections 12 pin on page
15.

1.1.6 Cable
Use the following connection cables:
Name Specification Shield Norm
[Y/N}
Power LIYCY 1x2x0.5mm² XAI Y IEC 60092-352
USB USB Y
Serial 3x2x0.5mm² Y IEC 61162-2
Serial 2x2x0.5mm² Y IEC 61162-1
Ethernet Ethernet CAT 5e S/FTP Y IEC 61162-450

Table 3: Connection Cables

1.1.7 Cable preparation


Cable preparation and cable connections as described in this manual are essential for the correct functioning of the
instrument.
1. Remove approx. 80 mm of the plastic cable sheath, but leave the grounding shield untouched.
2. Cut away approx. half the length of the now visible cable shield and fold the remaining shield back over the cable
sheath.
3. Wrap insulating tape over half of the visible grounding shield.
4. Attach the remaining visible grounding shield to the metal saddle on the rear of the instrument. See Figure 3: Cable
preparation on page 13
Note Always check the drawing for the correct shielding of signals. See chapter Connect NMEA on page 15
for description of how to apply shielding.

Figure 3: Cable preparation

1.1.8 Grounding
To function properly, the instrument must be grounded to the ship’s mass.
For this purpose the instrument has a grounding bolt. Connect the grounding bolt to the ship’s mass with a low
impedance connection.

13 | Installation Instructions
Figure 4: Grounding Bolt

Note The grounding strap must be as short as possible. If wire is used, use a minimum of 6 mm2 copper wire.
See Figure 4: Grounding Bolt on page 14
Note Always check the drawing for the correct shielding of signals. See chapter Connect NMEA on page 15
for description of how to apply shielding.

1.1.9 Power supply


The display has one 24 VDC (nominal) power input. Power consumption is approx. 500 mA. Connect to a power supply
> 1 A.
Note The power supply input is protected against connecting to the wrong polarity.

Note In rush current ∼ 4A pf approximately 4A.

1.1.10 Serial connection


The MF instrument has four IEC 61162 serial ports with Transmit (Tx) and Receive (Rx) connections. Three ports are of
type IEC 61162-1, one port is of type IEC 61162-2.
Note for explanation of correct connectionof IEC61162-1 and IEC61162-2 signals refer chapter Connect NMEA
on page 15
The instrument has two serial connectors, connector 1 with 8 connections and connector 2 with 12 connections.
For pin connections see Figure 5: Serial pin connections on page 14, Table 4: Serial connections 8 pin on page 15
and Table 5: Serial connections 12 pin on page 15

Figure 5: Serial pin connections

14 | Installation Instructions
1 COM0 IEC61162-2 Tx+ 1 COM1 IEC61162-1 Tx+
2 COM0 IEC61162-2 Tx- 2 COM1 IEC61162-1 Tx-
3 COM0 IEC61162-2 GND 3 COM1 IEC61162-1 Rx+
4 COM0 IEC61162-2 Rx+ 4 COM1 IEC61162-1 Rx-
5 COM0 IEC61162-2 Rx- 5 COM2 IEC61162-1 Tx+
6 Relay Contact NO 6 COM2 IEC61162-1 Tx-
7 Relay Contact C 7 COM2 IEC61162-1 Rx+
8 Relay Contact NC 8 COM2 IEC61162-1 Rx-

Table 4: Serial connections 8 pin 9 COM3 IEC61162-1 Tx+


10 COM3 IEC61162-1 Tx-
11 COM3 IEC61162-1 Rx+
12 COM3 IEC61162-1 Rx-

Table 5: Serial connections 12 pin

Note The instrument should always be connected according to the connection diagram. See Figure 47:
Connection Diagram MF Instrument on page 52. Ensure connection to the correct COM port.

1.1.11 Relay
One relay output is available for legacy alarm monitoring systems without serial ALR connection. This MF instrument
relay output is located on the 8-pin connector on the following pins. Use the NC (Normally Closed), or NO (Normally
Open) connection depending on the application.
For pin lay out, see Table 6: Relais connection on page 15.

6 Relais Normally Open


7 Relais Common
8 Relais Normally Closed

Table 6: Relais connection

1.1.12 Connect NMEA


Serial data from sensors such as GPS, Speedlog and others is commonly known as NMEA. NMEA is an American
standard and the naming is commonly used.
However common and well known, NMEA is the American standard. In the regulations is referred to the IEC standard for
the correct protocol description. These are:
IEC61162-1 and IEC61162-2. The manual will use both these standards as there is a difference between them. This
chapter describes both standards to clarify the differences and the correct application of them.
This chapter explains some basic knowledge about the serial interfaces used in the Marine Electronics.
For detailed information on both standards, please read the IEC-61162-1 and IEC61161-2 documents.
• IEC 61162-1
This standard is the most commonly used. In the standard, the sender (Tx side) and receiver (Rx side) are referred to
as Talker and Listener.
It has the following characteristics:
• Communication speed: BAUD rate of 4800, 8 databits, 1 stop bit, no parity
• One talker (TX side, commonly a sensor such as GPS) may be connected to one or more listeners (RX side, e.g.
displays or computer systems such as ECDIS). The circuit looks as shown in Figure 6: IEC 61162-1 Circuits on
page 16

15 | Installation Instructions
Figure 6: IEC 61162-1 Circuits
There are some requirements to the wiring of the signal.
• Tx+ and Tx- are connected to Rx+ and Rx- respectively.
• On the transmitting side the shield of the cable is connected to earth , and is NOT connected on the listener side.
Note In case the NMEA signal is connected both ways, two cables should be used with proper shielding
as described above. (It is just above situation times two for the opposite direction).
Only one talker is connected to the bus, and multiple listeners may be connected to the same bus without the need
for a signal isolator/multiplier. However, it is not forbidden and good practice to use a multiplier since it eliminates
possible faults caused by short circuiting of the signal by a faulty listener. Alphatron is able to supply such a multiplier
if needed (NMEA distribution module mk.2)
• IEC-61162-2
This standard is applied to provide better characteristics when the data transmission speed is higher.
It has the following characteristics:
• Communication speed: BAUD rate of 38400, 8 databits, 1 stop bit, no parity.
• One talker (TX side, commonly a sensor such as GPS) may be connected to one or more listeners (RX side, e.g.
displays or computer systems such as ECDIS). The circuit looks as shown in Figure 7: IEC 61162-2 Circuits on
page 16

Figure 7: IEC 61162-2 Circuits


The main difference with the IEC61162-1 standard is that a COMMON signal is added here for a good reference to
isolated ground.
• Tx+ and Tx- are connected to Rx+ and Rx- respectively.
• The shield of the cable is connected to the earth on the transmitting side, and is NOT connected on the listener.

16 | Installation Instructions
Note In case the NMEA signal is connected both ways, two cables should be used with proper shielding
as described above. (It is just above situation times two for the opposite direction).
Only one talker is connected to the bus, and multiple listeners may be connected to the same bus without the need
for a signal isolator/multiplier. However, it is not forbidden and good practice to use a multiplier since it eliminates the
possible faults caused by short circuiting of the signal by a faulty listener. Alphatron is able to supply such a multiplier
if needed (NMEA distribution module mk.2)
The cable should be of a shielded type, and the shield connected to the earth on the Talker side and NOT to earth on the
Listener side.
The IEC61162 specifies a number of different cables which can be used. Of these cables, Alphatron recommends to use
single shielded cable, as shown in Figure 8: Single Shielded Cable on page 17

Figure 8: Single Shielded Cable

So, for a two way connection two times 3 wire cable is used with shield connected on the Talker side.
Note For MODBUS we recommend to use the IEC61162-2 connections on equipment. IEC61162-1 ports are
also usable and tested to work with MODBUS on speeds up to 38400 BAUD, however the IEC61162-2 ports are
a little better suited because of the C wire for signal ground reference.
To make connections, proceed as follows:
1. Connect data signal to the COM port as drawn in the Connection Diagram of the AlphaRudder. See Figure 49:
Connection Diagram on page 55

1.1.13 Connect Rudder Feedback and Rudder Interface


The Rudder Feedback unit consists of a potentiometer output, combined with limit switches (HD version), that can be
connected to the display unit via an interface.
Rudder may be connected to the MF instrument using a serial signal ($xxRSA sentence, first rudder field), or by using a
Rudder Feedback.
1. Set the micro switches on the front as explained below, in order for the potentiometer to be connected and configured
correctly.
INFO:
Note The position of the switches A to F are as follows:

17 | Installation Instructions
SWITCH FUNCTION
CONTACT
A Setting analog channel 1
B Setting analog channel 1
C Setting analog channel 2
D Setting analog channel 2
E Spare
F Slave address setting

Table 7: Switch Meaning

Figure 9: Micro Switch


Locations

Location of the switches are shown in Fig.Figure 9: Micro Switch Locations on page 18

SWA SWB FUNCTION


OFF OFF +/- 5V & +/- 25mA input
OFF ON +/- 10V input
ON OFF +/- 20V input
ON ON Rudder input

Table 8: Explanation Switches A and B

SWC SWD FUNCTION


OFF OFF +/- 5V & +/- 25mA input
OFF ON +/- 10V input
ON OFF +/- 20V input
ON ON Rudder input

Table 9: Explanation Switches C and D


Note For alternative settings, i.e. when other signal sources are used, refer to tables:Table 7: Switch
Meaning on page 18, Table 8: Explanation Switches A and B on page 18 and Table 9: Explanation
Switches C and D on page 18
Note Set the switches A and B to the ON position when a Rudder Feedback Unit is connected to analog
input 1.

18 | Installation Instructions
2. Connect the Rudder Feedback Unit to the Analog Interface Mk2. See Connection Diagram Figure 49: Connection
Diagram on page 55
INFO:
Note To prevent signal error, the cable with analog signal from the Rudder Feedback Unit to the Analog
Interface Mk.2 should be as short as possible. Position the Analog Interface Mk.2 as close as possible to the
Rudder Feedback Unit. The analog signal cable must not exceed 20 meters. Ensure the analog signal cable
is separated from high voltage and high current cables, by placing it in a different trunk.
Note Position the Analog Interface Mk.2 in a cabinet where there can be no interference from electronics
such as frequency drives. (use a different / separate cabinet)
Note The analog signal cable must be securely earthed to the Rudder Feedback Unit side, using the EMC
cable gland. The other side of the cable should remain unearthed.
INFO:
• CAUTION
• Failure to observe these guidelines could lead to serious system malfunction or wrong rudder information.
• WARNING
• The analog signal cable must be securely earthed to the Rudder Feedback Unit side, using the EMC
cable gland. On the side of the Analog Interface Mk.2, the shielding MUST be connected to the GND
connection of the Analog Input of the channel used.
3. Check functionality by checking the LED's
• Power LED - Steady green
• Modbus Rx LED - Blinking green
• Modbus Tx LED - Blinking red
• NMEA LED - Blinking red
• Status LED - Shows Functionality: OK - Steady green, ERROR - Steady Red
4. Use IEC61162-2 Interface to connect module to the display unit

1.1.13.1 Technical Specifications Interface


The Interface has a Power Supply Input and signals input and output.

1.1.13.1.1 Power Supply Input


Power Supply Input is as follows:
1. Single 24VDC input range
2. Power: appx. 2.4W (24VDC@100mA)
3. Polarity protection: yes, via serial diode

1.1.13.1.2 Inputs and Outputs


Inputs and Outputs are as follows:
1. 1 x IEC61162-1 connection for NMEA signals
2. 1 x IEC61162-2 connection for Modbus connection to MFx display
3. 2 x Analog input
The Analog Interface has 2 analog signal input channels and each channel can have the following signals defined in the
terminals, depending on switch position, see Table 10: Analog Signal Output Channel 1 on page 20 and Table 11:
Analog Signal Output Channel 2 on page 20

19 | Installation Instructions
Switch A Switch B Terminal 11 Terminal 12 Terminal 13 Terminal 14 Terminal 17 Terminal 18
OFF OFF ch1 +/- 5V X ch1 0V X X X
OFF OFF X X X X ch1 =/- ch1 0V
20mA
OFF ON ch1 +/- 10V X ch1 0V X X X
ON OFF ch1 +/- 20V X ch1 0V X X X
ON ON X Pot. Wiper ch1 0V Rudder X X
Signal Power

Table 10: Analog Signal Output Channel 1

Switch C Switch D Terminal 19 Terminal 20 Terminal 21 Terminal 22 Terminal 15 Terminal 16


OFF OFF ch2 +/- 5V X ch2 0V X X X
OFF OFF X X X X ch2 0V ch2 +/-
20mA
OFF ON ch2 +/- 10V X ch2 0V X X X
ON OFF ch2 +/- 20V X ch2 0V X X X
ON ON X Pot. Wiper ch2 0V Rudder X X
Signal Power

Table 11: Analog Signal Output Channel 2

Note Inputs of channel 1 and 2 are electrically isolated. Connect the inputs seperately perchannel and do not
combine wiring, such as ground.

1.1.13.1.3 Weight
The Analog Interface Mk2 weighs 200 gr.

1.2 Software
The software version for this range of instruments is 1.X

1.2.1 Software updates MF


Alphatron Marine is constantly improving and updating its products by developing new functionalities and improving
usability and performance.
Visit our support website www.jrc.am/support for the newest manuals and to check that your product is still running the
latest software. Due to the nature of our products and solutions, software and relevant instructions will be available to
authorized distributors and dealers only.
Software update files are made available by Alphatron and can be acquired from the Alphatron Service Desk
www.jrc.am/support. The update procedure consists of adding a zip file containing the new software/firmware to the root
directory on the SD card.
On start-up (boot) of the MF, the root directory of the SD card will be checked for the presence of a *.zip file. If this file is
found, the MF application will make a backup of the current software/firmware. Next it will try to unzip the new software/
firmware. If this is successful, the MF application will restart and try to run the new software/firmware. If no problem
is found during the new boot, then the old software/firmware will be deleted and the new application will run normally.
Otherwise, the existing software/firmware will automatically be reverted to.
Note DO NOT shut down the device during an update procedure, because this can cause irreversible damage
to the files on the SD card.

20 | Installation Instructions
2 Operation
All MF instruments are operated in a similar way because of a common user interface.
Being familiar with one instrument in the MF range makes operation of other instruments in the range easier.

2.1 Power
The unit must be connected to the power at all times.
• Use the Power button in the front panel of the instrument to switch the power ON and OFF.
Note In the OFF position, the Power button is still dimly lit for easy identification in the dark. This only applies
when the instrument is connected to the power supply.
• In a blackout situation the instrument will always return to the last power state. This means that if the instrument
was switched ON before a blackout, it will be ON after a blackout. The same applies to the OFF status. When
the instrument is switched ON, the Power button is lit in accordance with the DIM setting of the LCD backlight, so
dimming together with the LCD panel.

2.2 Main screen


When the MF instrument is switched ON, it will boot into the MF application. The MF application will show the instrument
specific data applicable to the particular instrument.
During the initialization time, while the instrument is starting up, this picture will be displayed on the screen Figure 10:
Splash Screen on page 21. Depending on the size of the instrument, the letters in the lower half of the screen may
show MFS-V, MFS-H, MFM, or MFL.

Figure 10: Splash Screen

After the initialization time has been completed, the main screen appears. The Main Screen displays four standard
buttons. See Figure 11: Main Screen AlphaRudder on page 22 and Table 12: Main screen buttons on page 22

21 | Operation
Figure 11: Main Screen AlphaRudder

Functionality of the buttons is as described in table below.

Shows the instrument is functioning normally. If there's a malfunction a colored alert symbol will
show here.

Opens the SETTINGS MENU, where settings can be adjusted and shows the functionality of the
instrument by rotating. If this stops rotating, the software of the instrument is very busy or has
crashed.

Touching this button will set the LCD dim level brighter.

Touching this button will set the LCD dim level one step less bright.

Table 12: Main screen buttons

2.3 Menu handling


Menu handling is standardised for all different menu screens and for all different instruments. Example: see Figure 12:
Menu Example on page 23 and Table 13: Explanation of on-screen buttons and icons on page 23.

22 | Operation
Figure 12: Menu Example

Go to Main Screen, touch again to go to MENU SETTINGS

Go back to previous screen, without saving settings.

Undo the last value change

Accept current settings and applied Settings will be saved.

Undo all settings and exit Menu

Normal Condition, meaning there is no alarm.

Scroll between standard settings using + or - button

Navigate through Menu pages using < or > button. Not visible
on this screen, but visible on subsequent menu screens.

Table 13: Explanation of on-screen buttons and icons

2.3.1 Default Values


When the ADR is reset, it will return to the start-up screen, where a new instrument can be selected. The newly selected
instrument will contain the DEFAULT VALUES.

23 | Operation
The Default Values for the whole range of Alphaline instruments is shown in Figure 13: Default Values all MF instruments

DEFAULT VALUES ALPHALINE MF SYSTEMS

SteeringRepeater_8,4_LS_SEA
DepthSingle_6,5_PT_INL

RudderROT_6,5_PT_INL

RudderROT_8,4_LS_INL

MeteoRel_6,5_PT_SEA

SpeedLog_6,5_PT_SEA

SpeedLog_8,4_LS_SEA
WindRel_6,5_PT_SEA
ShipSpeed_5_PT_SEA
Alphanav_8,4_LS_INL

ShipSpeed_5_PT_SEA

Heading_6,5_PT_SEA

Heading_8,4_LS_SEA
ShipSpeed_5_LS_SEA
ShipSpeed_5_PT_INL

Rudder_6,5_PT_SEA
Meteo_6,5_PT_SEA
Rudder_6,5_PT_INL

Meteo_8,4_LS_SEA
Heading_5_PT_SEA

Depth_6,5_PT_SEA
Heading_5_LS_SEA
Meteo_6,5_PT_INL

Remote_5_PT_SEA
Depth_6,5_PT_INL

Wind_6,5_PT_SEA
Clock_6,5_PT_SEA

Clock_8,4_LS_SEA

Wind_8,4_LS_SEA
Wind_6,5_PT_INL
Rudder_5_PT_INL

ROT_6,5_PT_SEA
Depth_5_LS_SEA
ROT_6,5_PT_INL
Depth_5_PT_INL

Clock_5_PT_SEA
Clock_5_LS_SEA

TMC_5_PT_SEA
Wind_5_PT_INL

MFS-V MFS-V MFS-V MFS-V MFM MFM MFM MFM MFM MFM MFL MFL MFS-H MFS-H MFS-H MFS-H MFS-V MFS-V MFS-V MFS-V MFS-V MFM MFM MFM MFM MFM MFM MFM MFM MFM MFL MFL MFL MFL MFL MFL

AlphaRudderTurn (River)
AlphaRudderTurn (River)
AlphaRudder (River)
AlphaRudder (River)

AlphaMeteo (River)
AlphaSpeed (River)

AlphaDepth (River)
AlphaDepth (River)

AlphaWind (River)
AlphaWind (River)

AlphaTurn (River)

AlphaNav (River)

AlphaSpeedLog
AlphaSpeedLog

AlphaHeading+
AlphaHeading
AlphaHeading
AlphaHeading
AlphaHeading

AlphaRemote

AlphaRudder
AlphaTMC(+)

AlphaMeteo
AlphaMeteo

AlphaSpeed
AlphaSpeed

AlphaDepth
AlphaSpeed
AlphaDepth

AlphaWind
AlphaWind

AlphaTime
AlphaTime
AlphaTime
AlphaTime

AlphaTurn
numbers correspond with Matrix 11 10 13 12 11 16a 4 10 5 12 14 5 18 15 7 13a 18 7 3 13a 2 18 15 7 16 6 8 13a 17 12a 18 7 16 17 9 12a
Menu Touch Screen Sensitivity
Sensitivity 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
Menu Theme
Theme DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY DAY
Language ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH - - - - - - - - - - - - - - - - - - - - - - - -
Menu Date/Time [1/2]
Year Compulsory Input = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Month Compulsory Input = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Day Compulsory Input = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Menu Date/Time [2/2]
Hour Compulsory Input = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Minutes Compulsory Input = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Time Zone Compulsory Input = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =

Figure 13: Default Values all MF instruments


Menu Range/Scale Settings
SCALE 10 - - - 10 - 90 - - - 10 (Depth) - - 10 - 40 - - - 40 - - 10 - - - - 40 - - - - - - - -
ALARM DISABLED DISABLED - DISABLED DISABLED DISABLED - DISABLED DISABLED DISABLED - DISABLED - DISABLED - - - - - - - - DISABLED - DISABLED - DISABLED - - DISABLED - - DISABLED - - DISABLED
(Rudder) (Rudder)
Menu Units
DEPTH - - - - - - - - - - - - - metre - - - - - - - - metre - - - - - - - - - - - - -
SPEED (sail) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
SPEED (wind) - - - bft - bft - - - bft - - - - - - - - - - - - - - bft - - - - bft - - bft - - bft
Menu Ship Settings
SPEED - - - - - - - - - - - - - - - - - - - - - - - - - - - - GPS SOG - - - - GPS SOG - -
Menu Wind Settings
WIND FILTER - - - DISABLED - DISABLED - - - DISABLED - - - - - - - - - - - - - - DISABLED - - - - DISABLED - - DISABLED - - DISABLED
Menu ROT Settings
ROT Damping - - - - - - DISABLED - DISABLED - - DISABLED - - - - - - - - - - - - - DISABLED - - - - - - - - - -
Damping Time (ms) - - - - - - 1000 - 1000 - - 1000 - - - - - - - - - - - - - 1000 - - - - - - - - - -
Menu ADV_SET Central Dimmer
Selection ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON ON
Menu ADV_SET Ethernet
IP Address: 192.168.31.xx 32 32 32 32 31 31 31 31 31 31 33 33 32 32 32 32 32 32 32 32 32 31 31 31 31 31 31 32 31 31 33 33 33 33 33 33
Subnet Mask 255.255.255.0 = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Gateway 192.168.31.1 = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Menu ADV_SET UART [1/2]
UART UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0 UART 0

24 | Operation
Parity NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE NONE
Stop Bits 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
Menu ADV_SET UART [2/2]
Data Bits 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
Baud Rate 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800 4800
Protocol NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA NMEA
Menu ADV_SET NMEA FILTER
ID 1: Sentence DPT - VTG MWV DPT MWV - - - MWV DPT - - DPT HDT VTG - HDT - VTG GGA - DPT HDT MWV - - VTG HDT MWV - HDT MWV HDT HDT MWV
ID 2: Sentence - - - - - - - - - - HDT - - - - - - - - - - - - - VTG - - - - VTG - - VTG - - VTG
ID 3: Sentence - - - - - - - - - - VTG - - - - - - - - - - - - - HDT - - - - HDT - - HDT - - HDT
Menu ADV_SET NMEA TALKER

on page 24
TALKER ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** ** **
USED / UNUSED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED USED
Menu ADV_SET Heading Settings
ANGLE OFFSET - - - - - - - - - - - - - - 0 - - 0 - - - - - 0 - - - - - - - 0 - - - -
Menu ADV_SET Ship Settings
SHIP LENGTH [M] - - - - - - - - - - - - - - - - - - - - - - - - - - - - Compuls. Inp. - - - - Compuls. Inp. - -
BOW GPS [M] - - - - - - - - - - - - - - - - - - - - - - - - - - - - Compuls. Inp. - - - - Compuls. Inp. - -
STERN GPS [M] - - - - - - - - - - - - - - - - - - - - - - - - - - - - Compuls. Inp. - - - - Compuls. Inp. - -
Menu ADV_SET Wind Settings
FILTER (ms) - - - 1000 - 1000 - - - 1000 - - - - - - - - - - - - - - 1000 - - - - 1000 - - 1000 - - 1000
2.4 Alert Handling
If a malfunction occurs, an audible and visible alert will be displayed.
An alert will be shown as a sequence of events on the display, see Figure 14: Alert Handling Main Screen on page
25.

Alert Push the Alarm Push the Alarm


occurrence button bar to
to show alarm acknowledge and
info show next alert
info

Push the Push the Push the Warning


Alarm bar to Warning bar to bar to
acknowledge acknowledge and acknowledge and
and show show next alert show next alert
next alert info info info

Push the icon Push the icon for


for acknowledge
acknowledge alarm And alert
alarm and alert bar to disappear
bar to
disappear

Figure 14: Alert Handling Main Screen

25 | Operation
Alert handling while operating the Menu, see Figure 15: Alert Handling in Menu Settings: on page 26
Alert Push the Alarm Push the Alarm
occurrence button bar to
to show alarm acknowledge
info and show next
alert info

Push the Alarm Push the Push the


bar to Warning bar to Warning bar to
acknowledge acknowledge and acknowledge
and show next show next alert and show next
alert info info alert info

Push the icon Push the icon for


for acknowledge
acknowledge alarm And alert
alarm and alert bar to disappear
bar to disappear

Figure 15: Alert Handling in Menu Settings:

26 | Operation
Icon
Icon Name Icon Description Icon Graphic
Number
0 Normal Condition Stationary circle with grey tick mark

Active -
1 Flashing red triangle with loudspeaker and alert text
Unacknowledged Alarm

2 Active - Silenced Alarm Flashing red triangle with loudspeaker with prominent
line through it and alert text

Active - Acknowledged
3 Red triangle with exclamation mark and alert text
Alarm

4 Active - Responsibility Red triangle with arrow pointing to the right and alert
transferred alarm text

5 Rectified - Flashing red triangle with tick mark and alert text
Unacknowledged Alarm

6 Active - Flashing yellowish orange circle with loudspeaker and


Unacknowledged alert text
Warning
7 Active - Silenced Flashing yellowish orange circle with loudspeaker with
Warning prominent line through it and alert text

8 Active - Acknowledged Yellowish orange circle with exclamation mark and


Warning alert text

9 Active - Responsibility Yellowish orange circle with arrow pointing to the right
transferred Warning and alert text

10 Rectified - Flashing yellowish orange circle with tick mark and


Unacknowledged alert text
Warning
11 Caution Yellow square with exclamation mark and alert text

Plus sign to be presented together with icon number 1


12 Aggregation +
- 11
13 Acknowledgement not Red triangle with cross to be presented together with
allowed for Alarm¹ icon numbers 1, 2 and 5

14 Acknowledgement not Yellowish orange circle with cross to be presented


allowed for Warning¹ together with icon numbers 6, 7 and 10

¹ "Acknowledge not allowed" icon is used when a Category A alert cannot be acknowledged in a task situation.

Table 14: Alert Management Icons

2.5 Settings
All instruments have Generic and Advanced Settings dependant on who needs access to it.
All users are authorised to use the Generic Settings. Only the commissioning engineer has access to the Advanced
Settings, which are only needed during commissioning or troubleshooting.

27 | Operation
2.5.1 Generic
Every user of the MF instrument can access the generic settings.
Touch the MENU button in the instrument screen to enter the SETTINGS menu page, see Figure 16: Settings Menu on
page 28

Figure 16: Settings Menu

2.5.1.1 Touch screen calibration (TOUCH CAL)


Screens can be calibrated by drawing lines on the screen with your fingers.
The screen can be calibrated by following the instructions that appear on the screen.

Figure 17: Touchscreen Calibration Figure 18: Calibration Reference Figure 19: Calibration Lines
Points

1. In the SETTINGS menu, press the TOUCH CAL button .


The touchscreen calibration menu page appears, see Figure 17: Touchscreen Calibration on page 28.
2. Calibrate the screen by touching the CONFIRM CALIBRATE THE TOUCH SCREEN button.
The calibration screen appears with text requesting to touch the four numbered reference points that appear on the
screen, see Figure 18: Calibration Reference Points on page 28.
3. Touch the four reference points in sequence 1 to 4, as they light up.
Text appears VALID or INVALID

28 | Operation
4. When valid, test calibration for accuracy by dragging a finger across the screen, see Figure 19: Calibration Lines on
page 28
The line that appears on the screen should follow the finger.
5. Touch the ACCEPT button if calibration appears to be correct, or press the AGAIN button to repeat the calibration
process.
6. When the text INVALID appears press the AGAIN button to repeat the calibration process.

2.5.1.2 Touch screen sensitivity (TOUCH SET)


Screen sensitivity can be adjusted.
Setting the desired sensitivity is achieved by touching the appropriate buttons on the screen.

Figure 20: Touchscreen sensitivity

1. In the SETTINGS menu, press the TOUCH SET button . The touchscreen sensitivity menu page appears, see Figure
20: Touchscreen sensitivity on page 29.
2. Touch the + or - buttons to select the desired sensitivity value.
3. Touch the Accept √ button if sensitivity level appears to be correct.
INFO: There are 10 different options for touchscreen sensitivity, from level 1 for low touchscreen sensitivity, up to
level 10 for very high touchscreen sensitivity.
4. Touch the ← button to return to SETTINGS screen.

2.5.1.3 Change Date and Time (DATE TIME)


Date and Time settings can be easily adjusted.
Setting the desired time is achieved by touching the appropriate buttons on the screen.

29 | Operation
Figure 21: Date & Time page 1 Figure 22: Date & Time page 2

1. In the SETTINGS menu, press the DATE TIME button .


The Date menu page appears, see Figure 21: Date & Time page 1 on page 30
2. Touch the + and - buttons to set the correct DATE values.
3. Touch the > button to select the TIME page, see Figure 22: Date & Time page 2 on page 30
4. Touch the + and - buttons to set the correct TIME values.
5. Touch the Accept √ button to activate the chosen values and return to the SETTINGS screen.
6. To abort, press the ← button to return to the SETTINGS screen.

2.5.1.4 Change contrast theme (THEME)


Contrast colors can be easily adjusted to Day, Dusk and Night settings.
Set the contrast colours by touching the appropriate buttons on the screen.

Figure 23: Theme Menu

30 | Operation
1. In the SETTINGS menu, press the THEME button.
The THEME menu page appears, see Figure 23: Theme Menu on page 30 and
2. Touch the + or - buttons to select the desired Mode values.
INFO: Three different Mode values can be selected, in accordance with ambient light; Day, Dusk and Night.
3. Touch the Accept √ button to confirm the chosen Mode.

2.5.1.5 Change Range Settings (RANGE SCALE)


There are various options for Range Settings depending on which instrument it is being used.
With the Menu Range Setting the valid working range is determined. The Scale setting determines at which precise
setting the alarm will be activated, both audible and visual, when the valid working range is exceeded.

Figure 24: Range Settings Rudder


Figure 25: Range Settings Rudder Alarm

1. In the SETTINGS menu, press the RANGE SCALE button.


The RANGE SETTINGS menu page appears and the ALARM bar reads DISABLED, see Figure 24: Range Settings
Rudder on page 31
2. Touch the + button to change ALARM value to the maximum working rudder angle. This value is the Rudder Limit
Alarm. When the rudder angle is beyond this angle, say 45˚, there will be an audible and visible alarm, see Figure 25:
Range Settings Rudder Alarm on page 31
INFO:
Note To disable the Rudder Limit Alarm, set the value to DISABLED.

3. Touch the √ button to confirm.


The MAIN MENU appears again.

2.5.1.6 About the MF instrument (ABOUT)


The About screen contains details of the version of the software, type approval and when it was built.
Touch the MENU button on the screen to return to the Menu.

31 | Operation
Figure 26: About the MF instrument

1. In the SETTINGS menu, press the ABOUT button .


The menu page appears, see Figure 26: About the MF instrument on page 32.
Note When asking for manufacturer's support this information will be useful.

2.5.2 Advanced
The Advanced Settings are password protected and are for use by commissioning engineers only.
In the SETTINGS menu, press the ADV SET button.

2.5.2.1 Advanced settings password (ADV SET)


A password is required to open the menu .

Figure 27: Password dialog

1. In the SETTINGS menu, touch the ADV SET button .


The Password dialog screen appears, see Figure 27: Password dialog on page 32.
2. Key in the password and touch the Accept √ button.

2.5.2.2 Advanced settings menu


In this menu various settings may be adjusted.

32 | Operation
After touching the Accept √ button in the password screen, the ADVANCED SETTINGS menu screen appears, see
Figure 28: Advanced Settings Menu on page 33

Figure 28: Advanced Settings Menu

1. Select the button to configure a specific setting, or touch the ← button to return to the Main menu.

2.5.2.3 Central dimming (CNTRL DIM)


The MF instrument supports Central dimming as a listener (slave) from a standard (IEC 61162-1) dimming sentence.
This dimming signal should be connected to serial port COM3, see Table 4: Serial connections 8 pin on page 15 and
Table 5: Serial connections 12 pin on page 15. Also see Figure 5: Serial pin connections on page 14.
If a central dimmer is connected, it can be enabled through the Central dimming menu, see Figure 29: Central dimming
on page 34. There are no other configuration options for central dimming.

33 | Operation
Figure 29: Central dimming

1. In the ADVANCED SETTINGS menu touch the CNTRL DIM button.


The CENTRAL DIMMER SCREEN appears, see Figure 29: Central dimming on page 34.
2. Touch the + or - button to select CENTRAL DIMMING on or off and confirm with the Accept √ button.

2.5.2.4 Ethernet configuration (ETH CONFIG)


The MF instrument has one network interface to connect to a ship's network.
This can be useful when the instrument is connected to a remote interface or a PLC which uses MODBUS/TCP. The
ethernet interface can also be used for IEC611612-450 signals.
Note This option has been included for future use.

In the Ethernet menu pages, touch a setting to enter the setting dialog. Use the numeric pad to enter the required setting.

34 | Operation
Figure 30: Ethernet Configuration Figure 31: Numeric Pad

1. In the ADVANCED SETTINGS menu, touch the ETH CONFIG button.


The Ethernet Configuration menu page appears, see Figure 30: Ethernet Configuration on page 35.
2. Touch the bar below IP ADDRESS, SUBNET MASK, or GATEWAY to change these.
The numeric pad appears every time one of the bars is touched, see Figure 31: Numeric Pad on page 35
3. Touch the numbers to insert new numbers and Touch the X button to delete numbers.
4. Touch the Accept √ button when numbers are correct.

2.5.2.5 Serial port (UART) configuration (UART CONFIG)


The MF instrument is equipped with 4 serial ports (also called ’UART’ or COM).
The settings of the serial ports are divided over two screens as shown in Figure 32: UART screen 1 on page 36 and
Figure 33: UART screen 2 on page 36. Use the < or > button to toggle between the pages.
The default setting for NMEA IEC61162-1) data is 4800 BAUD, 8 databits, 1 stopbit. The default setting for MODBUS
over serial port is 19200 BAUD, 8 databits, 1 stopbit and EVEN parity. See also Figure 13: Default Values all MF
instruments on page 24
Select and configure NMEA when a sensor is connected with the standard NMEA signals. Select ModbusMaster when
an interface (such as Rudder Interface or ROT interface) is used.

35 | Operation
Figure 32: UART screen 1 Figure 33: UART screen 2

1. In the ADVANCED SETTINGS menu, touch the UART CONFIG button .


The UART screen 1 dialog screen appears, see Figure 32: UART screen 1 on page 36
2.
• Touch the + button, or the - button to set UART from 0 to 3
• Touch the + button, or the - button to set PARITY to ODD, EVEN, FORCED 0, FORCED 1, NONE.
• Touch the + button, or the - button to set STOP BITS to 1or 2
3. Touch button > to toggle to screen 2, see Figure 33: UART screen 2 on page 36 and follow the same principle as
for screen 1. Touch button < to return to screen 1.
4. Touch the Accept √ button when settings are correct.

2.5.2.6 Serial port monitor (SERIAL MON)


It is possible to show serial data transmitted and received over the serial ports.
Usually, NMEA signals should be viewed as ASCII data and MODBUS data should be viewed as HEX data. Received
data as well as sent data can be viewed (but not simultaneously). Select the appropriate button (Tx / Rx). For NMEA, Rx
is normally used.

36 | Operation
Figure 34: Serial Port 1 Figure 35: Serial Port 3

1. In the ADVANCED SETTINGS menu, press the SERIAL MON button.


The Serial Port Monitor 1 appears, see Figure 34: Serial Port 1 on page 37.
2. Touch the < or > buttons to select next port, see Figure 35: Serial Port 3 on page 37.
3. Touch Rx/Tx buttons to select desired setting.
4. Touch ASCII/HEX BUTTONS to select desired setting.
5. Touch the ← button to return to ADVANCED SETTINGS menu.
INFO:
Note All serial ports are configured the same way, the serial port setup screen has the following options:
• Baud rate: - Baud rate is a setting for Modbus devices, NMEA talkers, and other devices. It is also known
as symbol rate and modulation rate. The term roughly means the speed that data is transmitted. It is a
derived value based on the number of symbols transmitted per second. Valid values are: 1200, 2400,
4800, 9600, 19200, 38400, 57600 or 115200. The default value for NMEA is 4800. The default value for
remote I/O devices is 19200.
• Data bits: - The numbers of data bits in each character can be 7 (for true ASCII), 8 (for any kind of data,
as this matches the size of a byte) 8 data bits are almost universally used in newer applications. 7 bits
is only used on special occasions. Valid values are: 7 or 8. The default value for NMEA and remote I/O
devices is 8.
• Stop bits: - Stop bits sent at the end of every character allow the receiving signal hardware to detect
the end of a character and to resynchronize with the character stream. Electronic devices usually use
one stop bit. Valid values are: 1 or 2. The default value for NMEA is 1. The default value for remote I/O
devices is 1.
• Parity: - Parity is a method of detecting errors in transmission. When parity is used with a serial port, an
extra data bit is sent with each data character, arranged so that the number of 1 bits in each character,
including the parity bit, is always odd or always even. If a byte is received with the wrong number of 1's,
then it must have been corrupted. However, an even number of errors can pass the parity check. Valid
values are: N ( None ), O ( Odd ) or E ( Even ). The default value for NMEA is N (None). The default value
for remote I/O devices is E (Even)
Note The dark gray frame is used to monitor serial data. If a remote I/O device is connected, it will display
hexadecimal Modbus data, when an NMEA device is connected it will show the NMEA serial data strings.
Note Settings are only activated if the ‘Apply’ button is pressed.

37 | Operation
2.5.2.7 NMEA Talker
With this menu, some additional NMEA sentence filters can be applied. This is depending on the configuration of the
software.
• TALKER - If this is set to **, the NMEA (IEC61162-1) talker ID is not checked and always accepted. This is the
default behaviour, normally it cannot be changed to another value. For example, if this is set to: ** -> All sentences
accepted ($**HDT,...) HE -> Only HE is accepted as valid data ($HEHDT,...) - All other sentences i.e. ($GPHDT,...)
are ignored.
• USED/UNUSED - This setting defines if a sentence is used at all. If this is set to UNUSED, all sentences are ignored.
This makes is possible for example to filter specific messages. For example when it is set as follows: HDT -> Unused,
HDG -> Used
Now only messages of the type HDG are used in the system.
Setting the desired selection is achieved by touching the appropriate buttons on the screen as follows: See Figure 36:
NMEA Talker on page 38

Figure 36: NMEA Talker

1. Select the correct sentence type by touching the < or > buttons.
2. The TALKER (manufacturer) is usually indicated as a "Wildcard" by **
3. Touch the + or - buttons to select USED or UNUSED

2.5.2.8 Factory reset (RESET)


All menu settings can be returned to the factory default setting

38 | Operation
Figure 37: Factory reset Figure 38: First Start-Up Screen

1. In the ADVANCED SETTINGS menu, touch the RESET button.


The EXECUTE FACTORY RESET dialog page appears, see Figure 37: Factory reset on page 39.
2. Touch the CONFIRM RESET ALL SETTINGS button for a full reset to factory default values.
INFO:
Note See Figure 13: Default Values all MF instruments on page 24

The ARD reverts to the "First Start-Up" screen, see Figure 38: First Start-Up Screen on page 39
3. To return to the previous menu page, touch the ← button.

2.5.2.9 Log (LOG)


The log register shows information useful for fault finding.
Touch the buttons on the screen to see the log content.

39 | Operation
Figure 39: Log Screen

1. In the ADVANCED SETTINGS menu, press the LOG button.


The LOG screen appears, see Figure 39: Log Screen on page 40
2. Touch the ˅ or ˄ buttons to scroll through the readings.
3. Take a picture of the log screen and contact the Alphatron Service Desk www.jrc.am/support about errors.
INFO:
Note E.g. use mobile phone for taking picture.

4. Touch the ← button return to the Advanced Settings Menu.


INFO: Touching the MAIN button takes you back to the MAIN screen.

2.5.2.10 Calibrate Rudder (RUDDER SET)


A sensor on the rudder provides information to the instrument, where the rudder angle can be calibrated to an accurate
setting.
Prior to calibrating rudder ensure that:
• The ship is not sailing and autopilot is not activated.
• If the rudder is set to starboard, the value in the ADC: field must be higher than if the rudder is set to port. If this is not
correct, the potentiometer connected to the interface must be reversed.
When the Rudder signal is connected using a Analog Interface, the analog interface needs to be calibrated. Follow
the procedure in this chapter to calibrate. If the rudder signal is connected using standard NMEA sentences, this is not
applicable.

40 | Operation
Figure 40: Rudder Settings

1. In the menu, press the ADV SET button.


The PASSWORD dialog appears.
2. Key in the password and confirm with the Accept √ button.
The ADVANCED SETTINGS menu appears.
3. Touch the RUDDER SET button.
The RUDDER SETTINGS screen appears, see Figure 40: Rudder Settings on page 41
4. Set the rudder to the centre position, then press the CENTRE button.
The value will change to around 0˚.
5. Move the rudder to the SB angle identical to the Rudder Range value. (E.g. 90˚) Then press the STBD button.
6. Move the rudder to the PS angle identical to the Rudder Range value. Then press the PORT button.
7. Touch the Accept √ button to store the settings in the Remote I/O module.
Check the Value by moving the rudder, it should now display the correctly calibrated values.

2.5.2.11 Range Settings


This setting allows to setup multiple types of Rudder Scale.
SCALE and TYPE require to be determined:
• SCALE - Sets the visible rudder scale. It is best to select the scale which corresponds with the maximum rudder
angle of the ship, or the nearest higher value if the same value is not available. (i.e. select 90˚, if the ship's rudder
angle is 85˚).
• TYPE - Defines the orientation of the scale. There are 4 options:
1. FORE - This is the default setting, it shows a rudder scale with 0 value on top.
2. AFT - This shows a rudder scale with 0 value on top, but PS and SB are switched. This setting is particularly
useful on aft looking bridges for example on car ferries with two bridges.
3. FORE INVERSED - This setting shows a rudder scale with 0 value at the bottom.
4. AFT INVERSED - This setting shows a rudder scale with 0 value at the bottom, and with PS and SB switched.
This setting is particularly useful on aft looking bridges for example on car ferries with two bridges.

41 | Operation
Figure 41: ADVANCED RANGE SCALE MENU

To execute the settings proceed as follows:


1. Touch the ADV SET button in the SETTINGS MENU, see Figure 16: Settings Menu on page 28
The PASSWORD DIALOG appears.
2. Key in the password and confirm with the √ button.
The ADVANCED SETTINGS MENU appears, see Figure 28: Advanced Settings Menu on page 33
3. Touch the RANGE SCALE button.
The ADVANCED RANGE SCALE MENU appears, see Figure 41: ADVANCED RANGE SCALE MENU on page
42
4. Touch the + or - button to select the required SCALE: 90˚ (default), 45˚, 60˚, or 70˚.
5. Touch the + or - button to select the required option and confirm with the √ button.
6. The ADVANCED SETTINGS screen appears again.

42 | Operation
3 Specifications
The whole range of the MF instruments have the same electrical, environmental and mechanical specifications.

3.1 Electrical
Power supply, Data ports, Display and Additional Connections make up the electrical specifications.

Power supply
Single source 24VDC input.
Rated power usage 12 Watt. (24VDC@500mA)
Reverse polarity protection

Data ports
1 x IEC61162-2 serial port for fast NMEA signals (rated at typical 38400 kbps).
3 x IEC61162-1 serial port for NMEA signals (rated at typical 4800 kbps).
1 x IEC61162-450 network port for ethernet based signals. (10/100 mbps).
All data ports are galvanic isolated, full duplex and support NMEA signals according IEC61162, and MODBUS for
communication with industrial controllers such as PLCs.

Display
Alphaline MFS - Screen resolution 800 x 480 pixels, brightness 450cd/m2
Alphaline MFM - Screen resolution 640 x 480 pixels, brightness 450cd/m2
Alphaline MFL - Screen resolution 800 x 600 pixels, brightness 450cd/m2
All versions have a 4-wire resistive touch screen, single touch, glove operable.

Additional connections
1 x USB port for software update and maintenance.
1 x Relais contact with Normally Open and Normally Closed contacts.

Alphaline MFS-VJ only: Joystick with X and Y movement, push button and rotary encoder.
Alphaline MFS-VR only: Rotary encoder with push button.

3.2 Environmental
Environmental classification of the Alphaline MF is in accordance with both DNV 2.4 and IEC60945 international norms.
The Alphaline MF is Type Approved according the following classes (DNV 2.4 table 2.1): see Table 15: DNV 2.4 table
2.1 on page 43

Temperature: Class D Operating -25˚C - +55˚C


Humidity: Class B Storage -25˚C - +70˚C
Vibration: Class A Humidity < 95% (at 40˚C)
EMC compatibility: Class B Water resistance IP56 front
Enclosure: Class C IP22 back

Table 15: DNV 2.4 table 2.1 Compass Safe Distance 0.1 Meter

Table 16: Environmental

3.3 Mechanical
Mechanical specifications differ only in terms of size and weight.
Refer to Table 17: Dimensions on page 44 and to Mechanical Drawings in Appendix A . Dimensions are as shown on
drawings frontview WxHxD overall.

43 | Specifications
Product Width (mm) Height Depth Weight (kg)
(mm) (mm)
Alphaline MFS-H 180 96 80 0,754
Alphaline MFS-V 96 180 80 0,754
Alphaline MFS-VJ 96 180 102 0,768
Alphaline MFS-VR 96 180 102 0,768
MFM 160 180 80 1,225
MFL 240 180 80 1,825

Table 17: Dimensions

3.4 Maintenance
This product contains no operator serviceable parts. Service and repair shall only be carried out by personnel trained and
certified by ALPHATRON MARINE.
• CAUTION
• When cleaning the surface, do not use any organic solvent such as thinner or benzine. Otherwise, the paint
and markings on the surface may get damaged. For cleaning the surface, remove the dust and debris and
wipe with a clean dry cloth.

44 | Specifications
4 Appendix A
Appendix A contains the follwing subjects:
1. Mechanical Drawings
2. Connection Diagrams
3. Data Sheets
4. Type Examination Certificates
5. MED Certificates Alphaline MF
6. Thales Certificates
7. WSV Certificates Alphaline MF

4.1 Mechanical Drawings


The following pages show the mechanical drawings of this instrument and associated interfaces.

45 | Appendix A
6 5 4 3 2 1

Sideview Frontview Sideview Aftview


D D
4.1.1 MFM

Figure 42: MFM

46 | Appendix A
180
180
180
180
C C

72 8 160 8 72 160

11 138 11
Bottomview
4

8
30
4x
M4

72
B Panel Cutout B
138 x 172 mm
180
120
172

30
4

8 144 8
160
A Designed by Checked by Approved by Date Date
A
h.koppe 23-2-2015

Edition Sheet
MFM_030215 1/1
6 5 4 3 2 1
6 5 4 3 2 1

Zero Mark

9
D D

Figure 43: RFU MD


 170

47 | Appendix A
 19
0
4.1.2 Rudder Feedback MD

C C
Earth bolt M6

Cablegland M25 x 1,5


Cable diameter: 9-16 mm

+0,00
Ø20 - 0,05

40
10
B B

133
110
130
P1
P2

10
190

Designed by Checked by Approved by Date Date


A A
Material: Black Anodized Aluminum h.koppe 24-3-2015
Weight: 4,2 kg Rudder Feedback Unit MD
Environmental: IP 65
Edition Sheet
Mounting: 4x M8 Bolt ( Not included ) 3109.0196 1/1
6 5 4 3 2 1
6 5 4 3 2 1

Rudder Feedback Unit HD


3109.0194
D D
1
70

PS 1

48 | Appendix A
 19
0 SB 1
PS 2
SB 2

Figure 44: Rudder Feedback HD


P1
4.1.3 Rudder Feedback HD

P2
C C

Ø20

40
10
Earth bolt M6
Cablegland M25 x 1,5
B Cable diameter: 9-16 mm B

243
240
220
10
190

Designed by Checked by Approved by Date Date


A A
Material: Black Anodized Aluminum h.koppe 16-3-2015
Weight: 6,0 kg
Environmental: IP 65
Edition Sheet
Mounting: 4x M8 Bolt ( Not included ) Rudder Feedback Unit HD 1/1
6 5 4 3 2 1
6 5 4 3 2 1

M10 L
Topview M10 R Example with RFU HD
Welded on Tube Welded on Tube
D D

Tube Ø10mm inside


Yard Supply

49 | Appendix A
263

240

150 - 350 mm

Figure 45: RFU Linkage Transmission


C 2 C
0 Example with RFU MD
4.1.4 RFU Linkage Transmission

Aftview
153

130

40
20
PARTS LIST
11 B
B ITEM QTY PART NUMBER TITLE 8 10
14 11
1 1 3109.0198-01 RFU LT Base 14
11
2 2 3109.0198-02 RFU LT Rod 12 14 15 11
3 1 3109.0198-03 RFU LT End 13 9
4 1 3109.0198-04 RFU LT Din 6912 M6 x 30 SS A4 6 13
13
5 1 3109.0198-05 RFU LT DIN 127 - A 6 SS A4 6 11
6 4 3109.0198-06 RFU LT DIN 916 - M6 x 10 SS A4 7
2
7 1 3109.0198-07 RFU LT Din 71802 M10 R 3 12
8 1 3109.0198-08 RFU LT Din 71802 M10 L 2
9 1 3109.0198-09 RFU LT Stud M10 R 11
10 1 3109.0198-10 RFU LT Stud M10 L 6 4 13
A 11 6 3109.0198-11 RFU LT Din 127 A 10 Designed by Checked by Approved by Date Date
A
12 2 3109.0198-12 RFU LT DIN 125 - A 10,5 5 h.koppe 25-3-2015
1
13 4 3109.0198-13 RFU LT DIN 934 - M10 R 6 RFU Linkage Transmission
14 3 3109.0198-14 RFU LT DIN 934 - M10 L Edition Sheet
15 1 3109.0198-15 RFU LT Tube YardSupply 3109.0198 1/1
6 5 4 3 2 1
6 5 4 3 2 1

D D

Topview

50 | Appendix A
22,6
114

Figure 46: Analog Interface Mk2


4.1.5 Analog Interface Mk2

Frontview Sideview Aftview


C C

114

114

114
B B

22,6 114 22,6

Designed by Checked by Approved by Date Date


A A
h.koppe 24-3-2015

Analog Interface Mk2


Edition Sheet
3109.0192 1/1
6 5 4 3 2 1
4.2 Connection Diagrams
The Connection Diagrams illustrate the connections to other equipment, hardware and to power.

4.2.1 MF Instrument
Connect the instrument according to the Connection Diagram. See Figure 47: Connection Diagram MF Instrument on
page 52

51 | Appendix A
Figure 47: Connection Diagram MF Instrument

52 | Appendix A
4.2.2 Rudder Feedback
Connect the instrument according to the Connection Diagram. See Figure 48: Connection Diagram Rudder Feedback on
page 54

53 | Appendix A
Figure 48: Connection Diagram Rudder Feedback

54 | Appendix A
4.2.3 AlphaRudder
Connect the instrument according to the Connection Diagram. See Figure 49: Connection Diagram on page 55
Figure 49: Connection Diagram

55 | Appendix A
56 | Appendix A
4.2.4 Analog Interface Mk2
Connect the instrument according to the Connection Diagram. See Figure 50: Connection Diagram on page 57
Figure 50: Connection Diagram

57 | Appendix A
58 | Appendix A
4.3 Data Sheets
The Data Sheets provide information regarding the Electrical and Data Connections, the Display and Environmental
Classifications.
A data sheet has been generated for each instrument type and the various interfaces that can be connected to them.

4.3.1 Alphaline MF
The Alphaline MF range consists of six different display sizes, each with their dedicated data sheet.
In outline the Alphaline MF range has the base properties as shown in this Table 17: Dimensions on page 44

4.3.2 Rudder Feedback MD/HD


Data sheet Rudder Feedback MD/HD
Figure 51: Data Sheet Rudder Feedback MD/HD

4.3.3 Analog Interface Mk2


Data sheet Analog Interface Mk2
Figure 52: Data Sheet Analog Interface Mk2

4.4 Type Examination Certificates

4.4.1 Alphaline MF
4.4.2 Analog Interface Mk2
Figure 53: Type Examination Certificate Analog Module

4.4.3 Rudder Feedback


Figure 54: Type Examination Certificate Rudder Feedback

4.5 MED Certificates Alphaline MF

4.5.1 Rudder Indicator


Figure 55: MED Certificate Rudder Indicator

4.6 Thales Certificates


The following pages show the Thales Certificates of this instrument and associated interfaces.

59 | Appendix A
4.6.1 MFM

Figure 56: MFM Certificate

60 | Appendix A
4.6.2 Rudder Feedback MD

Figure 57: Rudder Feedback MD Certificate

61 | Appendix A
4.6.3 Rudder Feedback HD

Figure 58: Rudder Feedback HD Certificate

62 | Appendix A
4.6.4 Analog Module

Figure 59: Analog Module Certificate

63 | Appendix A
5 Appendix B
ISO Certificates

5.1 ISO 9001 certificate Alphatron Marine R&D

Figure 60: ISO 9001 Certificate Alphatron Marine R&D

64 | Appendix B
6 Appendix C
6.1 EC Declaration of Conformity
Declarations of Conformity relates to standards or other normative documents.

65 | Appendix C
Figure 61: Alphaline Repeater Display MFx (in all variants)

66 | Appendix C
Figure 62: Rudder feedback unit MD / HD

67 | Appendix C
All over the world,
close to the customer

JRC/Alphatron Marine
Schaardijk 23 (harbor 115) The information in this document is subject to change without
3063 NH Rotterdam notice and does not represent a commitment on the part of
The Netherlands Alphatron B.V.
T +31 10 453 4000 Document : Installation and Operation Manual
F +31 10 453 4010 Project nr. : AlphaRudder
info@alphatronmarine.com Version : V1.0
alphatronmarine.com © All rights reserved Alphatron Marine

Centers of Excellence
Houston, Rotterdam, Singapore, Tokyo

You might also like