You are on page 1of 204

Progeny, Inc.

EUREKA 1407 Barclay Blvd.


Buffalo Grove, IL 60089

Tel. (847) 850-3800


Fax (847) 850-3801

LINEAR IV ™

Automatic Collimation System

Installation
Operation
and Maintenance

00-02-1517 Rev. Q
Copyright © 1993 Progeny, Inc. Corp.
November 2005 ECN: P1095
PROGENY, INC.
Manufacturers of Eureka Collimators and Cables

INSTALLATION ADVISORY

TO: INSTALLERS, SERVICE PERSONNEL, AND USERS OF X-RAY SYSTEMS

COLLIMATOR MOUNTING INFORMATION

In order to ensure a safe and secure mounting of this collimator to the x-ray tube housing, the
following installation guidelines must be followed:

1. Two different length screws are provided in the cloth bag containing the spacers.
Determine the correct length of screw to use, taking into account the collimator
spacing requirements and/or peculiarities of the tube housing port boss.
2. Clean the screws and housing port boss with alcohol and, if necessary, remove any
debris which may be present in the tube housing mounting holes.
3. Securely fasten the upper mounting ring and spacers to the mounting surface
located on the tube housing port. As a precaution, a medium strength thread locking
compound such as Loctite #242, should be applied to the screws before fastening
the collimator mounting ring to the tube housing. The screws provided have a Nylok
patch, as vibration resistant mounting screws are strongly recommended.
Verify that the collimator mounting screws engage the housing by at least five (5)
threads when used with the any required collimator spacer plate(s).
4. In order to fasten the Collimator to the Tube Housing, it is necessary that the four (4)
collimator detent ball plungers (located on the top of the collimator) are aligned with
the detent holes located on the collimator tube mounting plate (i.e. collimator is
mounted in either the 0, -90, or +90 degree swivel position).
NOTE: It is much easier to mount the collimator when the tube is inverted (upside-
down) or if the collimator is placed on the table top and the tube is lowered onto it.
5. Carefully support the collimator in place and attach the clamping ring. The hinge of
the clamping ring must line up with the pin in the lower mounting ring. Securely
fasten the #6-32 socket head cap screw locking the collar halves in place. Use only
the provided Collar Locking Screw (26-00752), do not replace with other
hardware. In addition install two (2) Collar Locks (70-10038), which provides a
fail-safe for the Collar/Screw Assembly.
6. After mounting the collimator and/or performing any service to it or the tube housing,
inspect the fit of the collimator and the tube housing. Grasp and attempt to move
the collimator and then the tube housing assembly while inspecting for loose joints
or gaps between the tube/collimator assembly, as well as other tube mounting
areas. If a problem is found, consult factory personnel.
7. It is strongly recommended that a periodic inspection (at least every 12 months)
should be made to ensure mounting integrity.

WARNING

Failure to adhere to the above guidelines may result in loosening, damaged screws or mount
failure which could result in heavy components falling during use. Incidents of loose system
components should be reported immediately to x-ray service personnel for repair.
TABLE OF CONTENTS

SECTION PAGE

1.0 INTRODUCTION ........................................................................... 1-1


1.1 You Have Legal Obligations………………………………………………… 1-3
1.2 Background……………………………………………………………………. 1-3
1.3 Linear IV Standard Features………………………………………………… 1-4
1.4 Specifications – Linear IV……………………………………………………. 1-6
1.5 Remote Control Unit Interface………………………………………………. 1-7
1.6 Radiation and Mechanical/Electrical Warning…………………………….. 1-7
1.7 Compatibility………………………………………………………………… 1-8
1.8 Maintenance………………………………………………………………….. 1-9
1.9 European Representative……………………………………………………. 1-10
2.0 INSTALLATION……………………………………………………… 2-1
2.1 Unpacking……………………………………………………………………... 2-3
2.2 Equipment Supplied…………………………………………………………. 2-3
2.3 Collimator Mounting………………………………………………………….. 2-3
2.4 Power Supply Chassis Mounting…………………………………………… 2-5
2.5 Interconnect Wiring………………………………………………………….. 2-6
3.0 ALIGNMENT PROCEDURES……………………………………….. 3-1
3.1 Light Field / X-ray Congruence Test……………………………………….. 3-2
3.2 Field Projection Lamp and Mirror Adjustment…………………………….. 3-5
3.3 Cross Hair Window Adjustment……………………………………………. 3-6
3.4 Bucky Centering Light-Line Adjustment…………………………………… 3-6
4.0 CALIBRATION…………………………………………………………. 4-1
4.1 Calibration Procedure……………………………………………………….. 4-3
4.2 Disabling Hardware Calibration Interlock………………………………….. 4-16
4.3 Enabling Hardware Calibration Interlock…………………………………… 4-17
5.0 COLLIMATOR FUNCTION CHECK………………………………… 5-1
5.1 Collimator Function Check Procedures……………………………………. 5-3
5.2 Horizontal Collimator Function Check……………………………………… 5-6
5.3 Preparation For Function Check Procedures…………………………….. 5-6
5.4 Collimator Tilt Switch Checkout……………………………………………. 5-6
5.5 Table Tilt Switch Checkout (Tilting Tables Only)…………………………. 5-6
November 2005 (i) Linear IV Collimator
TABLE OF CONTENTS (Continued)
SECTION PAGE

5.6 Continuous SID Monitor Checkout…………………………………………. 5-7


5.7 External SID Signal Test (Horizontal)……………………………………… 5-7
5.8 Tilting Table Receptor SID Signal Checkout………………………………. 5-8
5.9 External Radiographic IR Signals Checkout………………………………. 5-8
5.10 External Fluoroscopic IR Signals Checkout……………………………….. 5-9
5.11 Tilting Table External Radiographic IR Signal Checkout………………… 5-10
5.12 Spotfilm IR Signals Checkout……………………………………………….. 5-11
5.13 Stereo/Tomo Bypass Signal Checkout…………………………………….. 5-11
5.14 Notes on Cassette Tray Adjustment……………………………………….. 5-11
6.0 COMPLIANCE VERIFICATION……………………………………… 6-1
6.1 Verification of Tests to be Performed………………………………………. 6-3
XR 8-2.14 Visual Definition (Radiographic) of X-ray Light Field……………………... 6-5
XR 8-2.15 Intensity of Light Field Illumination…………………………………………. 6-10
XR 8-2.16 Minimum Field Size………………………………………………………….. 6-12
XR 8-2.17 X-ray Field/Receptor Center Alignment…………………………………… 6-15
XR 8-2.18 Indication of X-ray Field Size………………………………………………. 6-16
XR 8-2.19 Positive Beam Limitation…………………………………………………….. 6-17
XR 8-2.20 X-ray Field Limitation and Alignment……………………………………….. 6-18
XR 8-2.21 Return to Positive-Beam Limitation with Image Receptor Change……… 6-21
XR 8-2.09 Beam Quality (Half-Value Layer (HVL))……………………………………. 6-22
XR 8-2.24 Fluoroscopic X-ray Field Limitation…………………………………………. 6-28
XR 8-2.25 Fluoroscopic Source-to-Skin Distance (SID)………………………………. 6-34
XR 8-2.27 Primary Barrier – X-ray Lockout…………………………………………….. 6-37
XR 8-2.29 Alignment of X-ray Field Edges to Selected Portions of the Image
Receptor……………………………………………………………………….. 6-37
XR 8-2.30 Alignment of Center of X-ray Field to Center of Selected Portion of
Image Receptor………………………………………………………………. 6-39
7.0 CONTINUOUS SID ADJUSTMENT…………………………………. 7-1
7.1 General………………………………………………………………………… 7-2
7.2 SID Monitor Installation Guidelines…………………………………………. 7-2
7.3 SID Monitor Potentiometer Adjustment……………………………………. 7-3
7.4 (Option) Differential (Table) SID Monitor Potentiometer Verification…… 7-4
7.5 SID Circuit Verification………………………………………………………. 7-4

November 2005 ( ii ) Linear IV Collimator


TABLE OF CONTENTS (Continued)
SECTION PAGE

8.0 OPERATING INSTRUCTIONS………………………………………. 8-1


8.1 General Instructions…………………………………………………………. 8-3
8.2 Front Panel Description…………………………………………………….. 8-3
8.3 List of Messages……………………………………………………………. 8-5
8.4 Radiographic (Automatic) Operation………………………………………. 8-6
8.5 Radiographic (Table Top/Manual) Operation……………………………… 8-8
8.6 Continuous and Differential SID Operation……………………………… 8-8
8.7 Horizontal (Wall) Operation………………………………………………… 8-9
8.8 Fluoroscopic Operation…………………………………………………….. 8-9
8.9 Spotfilm Operation…………………………………………………………… 8-10
8.10 Tiliting Operation and Tilting Table Operation…………………………….. 8-11
8.11 Stereo/Tomo Operation…………………………………………………… 8-12
9.0 REPLACEMENT PARTS LIST………………………………………. 9-1
10.0 APPENDIX……………………………………………………………… 10-1
Definitions……………………………………………………………………... 10-2
Linear IV Collimator Pausch URO Table Interface……………………….. 10-6
11.0 THEORY OF OPERATION………………………………………….. 11-1
11.1 Mechanical Operation………………………………………………………. 11-3
11.2 Electrical Operation…………………………………………………………. 11-3
11.3 Power Supplies……………………………………………………………… 11-6
11.4 Continuous SID Circuits…………………………………………………….. 11-6
11.5 Image Receptor Size Circuit………………………………………………… 11-6
11.6 Collimator Feedback Circuit………………………………………………… 11-6
11.7 A/D Converter Circuit………………………………………………………… 11-6
11.8 CPU Core Section (Schematic 70-08313 Sheet 1, Board 70-08314)….. 11-7
11.9 CPU Display/Key Board Section (70-08313 Sheet 2)……………………. 11-7
11.10 CPU Input/Output Section (70-08313 Sheet 3)…………………………… 11-7
11.11 Master Board (Schematic 70-08319, Board 70-08320)…………………. 11-8
11.12 Driver Board (Schematic 70-08315, Board 70-08316)…………………… 11-9
11.13 Display Board (Schematic 70-08317, Board 70-08318)………………….. 11-9
11.14 Front Panel 70-08152……………………………………………………….. 11-9
11.15 Interconnect Cable 70-08166………………………………………………. 11-10
12.0 SCHEMATICS AND BILLS OF MATERIAL………………………... 12-1

November 2005 ( iii ) Linear IV Collimator


LIST OF ILLUSTRATIONS

SECTION FIGURE PAGE

2.0 2.1 Component Identification………………………………. 2-2


2.2 Collimator Dimensions…………………………………. 2-4
2.3 Mounting Holes…………………………………………… 2-6
2.4 Wiring Block Diagram…………………………………… 2-8
2.5 Exposure Hold Circuit………………………………….. 2-9
2.6 Tomo/Stereo Bypass Circuit……………………………. 2-9
2.7 Image Receptor Circuit………………………………….. 2-10
2.8 Vertical SID Monitor Circuit……………………………... 2-12
2.9 Differential Vertical SID Monitor Circuit………………... 2-12
2.10 Horizontal SID Switch Circuit…………………………… 2-13
2.11 Table Tilt Monitor Orientation…………………………… 2-14
2.12 Table Tilt Monitor Circuit………………………………… 2-14
2.13 Power Supply Circuit…………………………………….. 2-15
2.14 Spotfilm Alignment Switch………………………………. 2-15
2.15 RAD/Fluoro Mode Switch………………………………. 2-16
2.16 Fluoro/Spot Film Switch………………………………… 2-17
2.17 Image Intensifier Mode Control…………………………. 2-17
2.18 Tube Select Switch……………………………………… 2-18
3.0 3.1 Light Field Congruence…………………………………. 3-3
3.2 Light Field Adjustment………………………………….. 3-4
6.0 6.1 BRH/FDA Compliance Test Stand…………………….. 6-2
6.2 Compliance Stand Detail……………………………….. 6-2
6.3 Light Field vs. X-ray Field Error Measurements………. 6-4
6.4 Calculation Example.…………………………………….. 6-7
6.5 Metal Marker Method…………………………………… 6-7
6.6 Half-Value Layer Determination Graphs………………. 6-27
6.7 Fluoroscopic X-ray Field Limitation Test Stand
Method……………………………………………………. 6-29
6.8 Fluoroscopic X-ray Field Limitation – Direct Method…. 6-31
6.9 Image Intensified Fluoroscope, Test Set-Up………….. 6-33
6.10 Image Intensified Fluoroscope, Metal Marker Method
Inside Edge……………………………………………….. 6-34
6.11 Focal Spot Geometry………………………………….. 6-36

November 2005 ( iv ) Linear IV Collimator


LIST OF ILLUSTRATIONS (Continued)

SECTION FIGURE PAGE

6.12 Calculation Example…………………………………….. 6-41


6.13 Spotfilm, X-ray Field Limitation, Test Stand Method,
Reference………………………………………………… 6-42
6.14 Spotfilm X-ray Field Limitation, Test Stand Method,
Actual……………………………………………………… 6-42
8.0 8.1 Linear IV Front Panel……………………………………. 8-2
10.0 10.1 Linear IV Tilt Switch Wiring (70-10147)……………….. 10-4
10.2 Linear IV Feedback Pots Wiring………………………... 10-4
10.3 Linear IV Motor Wiring…………………………………... 10-5
11.0 11.1 Linear IV Collimator……………………………………… 11-2
11.2a Collimator Block Diagram – General………………….. 11-4
11.2b Collimator Block Diagram – Logic…………………….. 11-5
11.3 Interconnect Cable (70-08166)……………………….. 11-11

LIST OF TABLES

SECTION FIGURE PAGE

5.0 5-1 Linearity Test Record................................................... 5-13


6.0 6-1 Minimum Beam Quality Requirements......................... 6-23
6-2 Aluminum Equivalent of Primary Beam Total Filtration 6-23
6-3 Highest Design Operating Range................................ 6-26
6-4 Half-Value Layers as a Function of Filtration and
Tube Potential for Diagnostic Units............................. 6-27
10.1 Collimator / Tube mOunting Information……………. 10-3

November 2005 (v) Linear IV Collimator


SECTION 1.0

INTRODUCTION

November 2005 (1-1) Linear IV Collimator


This product has been tested by Underwriter’s Laboratories in conformance with
standards set forth by UL 2601-1, CAN/CSA-C22.2 No. 601.1-M-90, and IEC 601-2-32.
It has been found to comply with these standards and, therefore, bears the above
“Recognized Component” symbol for UL and UL-C.

UL File No. E181750

Progeny, Inc. is registered to ISO9001, EN46001, and the Medical Device Directive
93/42/EEC Annex II by SGS Yarsley, Ltd.

November 2005 (1–2) Linear IV Collimator


1.0 INTRODUCTION
This manual contains information for the assembly, installation, adjustment, testing and
maintenance of the LINEAR IV radiographic/fluoroscopic collimator manufactured by
Progeny, Inc.

1.1 YOU HAVE LEGAL OBLIGATIONS

The manufacturers of beam limiting devices are required to provide instructions for the
assembly, installation, adjustment and testing adequate to assure compliance with
applicable provisions of DHHS Performance Standards 21 CFR Sub-Chapter J, Part
1020.

Those who assemble or service beam limiting devices must follow the instructions of the
original manufacturer and process the FD-2579 Assemblers Report where applicable.

You assume responsibility for compliance of this product if you fail to follow the original
manufacturer’s instructions or modify any component which affects radiation safety.

The FDA (BRH) requires that manufacturers must include a specific requirement that
the assembler perform all applicable tests at the time of installation. A thorough
explanation of the equipment required and step-by-step instructions must be provided
by the manufacturer. The instructions include a requirement to record key data to
demonstrate at a later time that all tests were performed and that the equipment was left
in full compliance with the standards.

As an assembler, you must perform these tests for the applicable requirements at the
time of installation and following any repairs which could alter the performance.

A Compliance Data Log is provided in this manual to record the results of the tests.

1.2 BACKGROUND

An X-ray collimator functions as an apparatus for regulating the cross-sectional size and
shape of a beam of radiation which emerges from an X-ray tube.

The source of radiation is virtually a point-source and, due to the tube housing design,
emerges from the port as a solid diverging cone of radiation. The finite angle of the
anode surface limits the X-ray beam on the anode side (heel-effect) forming a ‘D”
shaped X-ray field, limiting the useful coverage.

In “collimating” a beam to a given size and shape, a geared pair of lead shutters are
moved symmetrically into the beam to absorb the unwanted portion of the emerging
beam. A second geared pair of shutters are positioned at right angles to the first pair,
and again are moved symmetrically into the beam. In this manner, a continuously
variable square/rectangular beam is formed.

November 2005 ( 1 -3 ) Linear IV Collimator


The landing area of the beam will contain a radiographic image receptor located in a
plane perpendicular to the beam at pre-determined distances from the radiation source
(focal spot).

The size and shape of the image receptor will determine the maximum useful cross-
sectional size and shape of the beam in the plane of the image receptor. The source-to-
image receptor distance (SID) determines the actual shutter opening required to
regulate the beam size and shape in the plane of the image receptor.

“Positive” beam-limiting (PBL) devices incorporate means to prevent X-ray production


until the beam limitation meets the applicable provisions of the Performance Standards.

Automatic PBL devices incorporate motors that regulate the shutter opening as
required. Manual PBL devices require the operator to adjust the shutters by the use of
knobs, levers, etc.

The primary objective of the electronic logic circuitry is to limit the beam to the size of
the image receptor and to provide other standardized operations consistent with the
DHHS Performance Standards 21 CFR Sub-Chapter J. This is accomplished by
electrically measuring the size of the SID involved. The resultant signal is then
compared to a signal which represents the collimator shutter opening to form a means
of limiting the beam.

The second objective is to provide convenience features, such as a status indicator light
to aid and guide the operator in the use of the collimator, particularly with a manual PBL
device.

1.3 LINEAR IV STANDARD FEATURES


The Linear IV collimation system from Progeny includes all features required for
diagnostic excellence, as well as to minimize installation and diagnostic down times:

• Full manual operation for table-top radiography.


• Automatic PBL operation upon insertion of a cassette into the bucky tray.
• Manual field size reduction after automatic PBL cycle.
• Optional remote shutter control capability for fluoroscopic or radiographic operation. Allows
selection of up to three image intensifier sizes and all other features accessible at the front
panel.
• Spotfilm capability for fluoro/spotfilm operation.
• Square or rectangular pattern continuously variable from 17” x 17” at 36” SID (or 43 cm x 43
cm at 90 cm SID) to fully closed.
• Size sensing capability for all metric and inch size cassettes.
• Compatible with urological tables that are equipped with image intensifier systems.
• Optional cone track available for accessories.
• Swivel mount for angulated positioning.

November 2005 (1–4) Linear IV Collimator


• Tape measure provided for accurate SID positioning.
• SID monitor provided for continuous SID vertical operation. Second SID monitor can be
incorporated for differential SID operation.
• Bright bucky centering light-line.
• Digital display of SID and field size in English or metric units, selectable during calibration.
• Alpha numeric display of operational modes; to indicate ready, manual, and exposure hold
conditions and the cause of the condition; to indicate RAD, FLUORO or SPOTFILM modes;
and to display diagnostic, calibration, and error messages.
• Power assisted manual drive of the shutters or manually controlled knobs for size
adjustment.
• AUTOCAL, built in smart software allows collimator calibration to be done solely at the front
panel. No access to the logic board is necessary to complete calibration, no potentiometers
to adjust, and interface to accessories such as SID monitors and Bucky trays are very
adaptable allowing for almost any to be used in conjunction with the collimator. 1
• Calibration information is stored in non-volatile memory with write protection. Thus power
failures and most all electrical noise will not affect unit operation or calibration parameters.

1 Maximum interface voltage not to exceed 2.5 Vdc.

November 2005 (1-5) Linear IV Collimator


1.4 SPECIFICATIONS - LINEAR IV
Operation: Fully automatic (PBL) within 2 seconds, and manual.

SID's for Automatic Continuously variable from 36" (90 cm) through 72" (180 cm)
Operation: vertical, and 36", 40", 48" and 72" (90cm, 100 cm, 120 cm, and
180 cm) horizontal.

SID Indicators: Digital display of SID, selectable in Metric or English. Display is


blank if valid SID is not confirmed.

Field Size Indicators: Digital display of X-ray field size at image receptor for the
particular SID. Display is blanked when a valid field size cannot
be determined.

Radiation Shielding: Rated at 150 kV. Less than 50 mR/Hr at one meter.

Film Coverage: Continuously variable from 17" x 17" (43 cm x 43 cm) to 5" x 5"
(13 cm x 13 cm) in automatic mode at all listed distances.
Manual range from 17" x 17" (43 cm x 43 cm) to fully closed.
28° coverage - 14° from central ray.

Light Field: More than 160 LUX (15 footcandles) with a minimum edge
contrast ratio of 4:1 at one meter. Controlled by internal timer
with quick turn-off capability. Timer turns off lamp after 25 + 10
seconds.

Accuracy: 2% of SID in use.

X-Ray Field Accuracy: Within 2% of SID in use in length and width (sum less than 4%).

SID Digital Display Within 1% of measured SID.


Accuracy :

Field Size Display Corresponds to X-ray field size within 1% of the SID in use.
Accuracy:

Bucky Light Line: Bright center line extending from center of field to withdrawn tray.

Stereographic/ Available at power supply. Accepts voltages of 5, 15, and 24 VAC or


Tomographic PBL switched ground.
By-Pass:

MirrorAdjustment: Allows for coarse or fine (thumbscrew) adjustment of mirror


position.

November 2005 (1-6) Linear IV Collimator


Inherent Filtration: 2.0 mm (min.), aluminum equivalent at 70 kV and above. Low
filtration mirror/bracket available from Progeny (1.0 mm min AL).

Power Requirement: 115-230 VAC, 50/60 Hz, 3 AMP, 3 Wire, 1 Phase

Light Field Lamp: Type DZE, 25 VAC, 150 WATT (GE).


Radiographic Mode Switched ground.
Selector Input:

Image Intensified Mode Isolated from "C" contacts rated at 2 Amps at 120 VAC resistive
Control: load.

Exposure Interlock: Isolated from "C" contacts rated at 2 Amps at 120 VAC resistive
load.

Fuse - Input Supply (1) 3 Amp MDL

- Lamp Power (1) 8 Amp MDL

- Motor Supply (1) 3 Amp MDL

Weight - Collimator 19.3 lbs (8.8 kg)

- Power Supply 27.5 lbs (12.5 kg)

1.5 REMOTE CONTROL UNIT INTERFACE

The Linear IV collimator system has the capability of remote control operation by using
the built-in remote interface at the collimator head. The remote functions capable are
the same as those availableat the front panel of the collimator and include: light field
control, manual shutter control, automatic size preset, and image intensifier selection
control.

A full function hand-held remote with the above functionality is available from Progeny.

Other remote units can be used if they can be interfaced to the built-in remote port.
CONTACT PROGENY, INC.FOR INFORMATION REGARDING THE REMOTE
CONTROL INTERFACE.

1.6 RADIATION AND MECHANICAL/ELECTRICAL WARNING


(from NEMA Standards Publication/No. XR8-1979)

Radiation Warning for Diagnostic X-Ray Systems


X-rays are dangerous to both operator and others in the vicinity unless established,
safe, exposure procedures are strictly observed.

November 2005 (1-7) Linear IV Collimator


The useful and scattered beams can produce serious, genetic or potentially fatal bodily
injuries to any persons in the surrounding area if used by an unskilled operator.
Adequate precautions must always be taken to avoid exposure to the useful beam, as
well as to leakage radiation from within the source housing or to scattered radiation
resulting from the passage of radiation through matter.
Those authorized to operate, test, participate in or supervise the operation of the
equipment must be thoroughly familiar, and comply completely with the currently
established safe exposure factors and procedures described in publications such as
Sub-Chapter J of Title 21 of the Code of Federal Regulations, “Diagnostic X-Ray
Systems and their Major Components,” and the National Council on Radiation Protection
(NCRP) No. 33, “Medical X-Ray and Gamma-Ray Protection for Energies up to 10 MeV-
Equipment Design and Use,” as revised or replaced in the future.

Failure to observe these warnings may cause serious, genetic or potentially fatal bodily
injuries to the operator or those in the area.

Mechanical-Electrical Warning for Diagnostic X-Ray Systems


All of the moveable assemblies and parts of X-ray equipment should be operated with
care.

Only properly trained and qualified personnel should be permitted access to any internal
parts. Live electrical terminals are deadly; be sure line disconnect switches are opened
and other appropriate precautions are taken before opening access doors, removing
enclosure panels, or attaching accessories.

Do not remove the flexible high tension cables from the X-ray tube housing, or high
tension generator, or the access covers from the generator until the main and auxiliary
power supplies have been disconnected.

When disconnecting high voltage cables, they must be grounded immediately in order to
dissipate any electrical charge that may remain on the cables or the tube.

Failure to comply with the foregoing may result in serious or potentially fatal bodily
injuries to the operator or those in the area.

November 2005 (1-8) Linear IV Collimator


1.7 REMOTE CONTROL UNIT

The Linear IV collimator system is available with a factory installed remote control unit.
The remote functions available are to allow the user to implement field size changes
using the hand held unit. The unit provides functions which are the same as those
available at the front panel of the collimator. These functions include: light field control,
manual shutter control, automatic size pretest, and image intensifier selection control.

If required to route the remote cable through small openings or system conduit hosing, it
is possible to remove the cable from the hand held unit and reconnect it after the cable
is routed. To implement this, first ensure system power is off, then remove the 4 screws
from the bottom of the hand held unit and separate it into two pieces. Carefully remove
the cable with its connector from the mating header on the printed circuit board inside.
The cable may now be routed, with caution to protect the connector. The connector can
be reinstalled and the unit reassembled.

A diagram of the remote control panel is provided:

EUREKA

MEM MEM
ONE TWO

LAST MAG
SIZE SIZE

FIELD

SIZE

LAMP

November 2005 (1-9) Linear IV Collimator


1.8 COMPATIBILITY

The Linear IV collimator is compatible and can be adapted for use with X-ray
tube/housing assemblies that meet all of the following factors:
1. Focal Distance of X-Ray Tube:

The focal spot to collimator mounting flange distance must be 2.44 inches, +/-
0.031 inches (1/32”). Four (4) spacers are supplies for adaptation:

1 - 1/4 inch (6.35 mm) spacer

3 - 1/16 inch (4.23 mm) spacer

Use any of the above combination to achieve the requirements.

2. Leakage Radiation:

Maximum leakage radiation from the X-ray tube/housing assembly must not
exceed 50 mR/hr at 100 cm (40 inches) at specified leakage technical factors.
This collimator is compatible with all x-ray tube housing assemblies having
leakage technique factors of 150 kV and 4 mA.

3. Inherent Filtration and Half-Value Layer:

The Eureka Linear™ IV collimator has a minimum value of 2.0 mm aluminum


equivalence at 70 kV. This value plus any tube inherent filtration plus any added
filtration must meet the minimum requirements of 21 CFR Sub-Chapter J, Part
1020.30 (m)(1) Table 1 on beam quality (e.g. minimum HVL at 100 kV must be
2.7 mm Al.

A low filtration mirror bracket (1.0 mm AL equivalent) is available from Progeny


for use with X-ray tubes with high inherent filtration.

4. Application:

The intended application is for general purpose radiographic, fluoroscopic,


tomographic, and chest applications. Maximum tube rating must be 150 kV or
less.

5. Installation:

Must be made with supplied hardware, including mounting flange, spacers (as
required), and four (4) 1/4” x 20 or four M6 x16 bolts equally spaced on a 3.625”
(9.2 cm) diameter bolt center.

November 2005 ( 1 - 10 ) Linear IV Collimator


1.9 MAINTENANCE

The Collimator system must be properly maintained to assure both compliance with FDA
regulations and useful life.
Preventive maintenance is to be performed once every twelve months. This includes
inspection and lubrication of the collimator mechanism. The collimator mounting ring
and locking screw (70-10036 and 26-00752) should be examined to ensure secure
mounting of the collimator. ONLY PROGENY P/N 26-00752 COLLAR LOCKING
SCREW SHOULD BE USED.

If collimator does not spin freely and smoothly, remove it from tube. Clean all surfaces
of collar and mounting flanges. Apply lubriplate or similar grease. Mount collimator onto
tube. If shutters seem too loose or too tight when turning knobs, remove the front bezel
cover and adjust the knob tension bar as required. This can be done using an allen key
to turn screws in bar.

Checkout should also occur if any of the following conditions occur:

• Lamp replacement
• Electronic component failure
• When collimator is removed from tube/housing assembly
• When collimator has been subjected to external damage

Refer to Section 5.0 for collimator CHECK-OUT procedure.

1.10 EUROPEAN REPRESENTATIVE

Progeny, Inc. has contracted with the following company to act as a European
Authorized Representative relative to the requirements of EN46001 and the Medical
Device Directive:

CE Partner 4U
Esdoornlaah 13
3951DB Maarn
The Netherlands

Phone: +31.343.442.524
Fax: +31.343.442.162

European customers should direct any customer complaints or requests for product
technical files to CE Partner 4U.

November 2005 ( 1 – 11 ) Linear IV Collimator


NO TEXT

November 2005 ( 1 – 12 ) Linear IV Collimator


SECTION 2.0

INSTALLATION

November 2005 (2-1) Linear IV Collimator


INTERCONNECT AND COLLIMATOR
POWER SUPPLY UNIT

MANUAL MOUNTING HARDWARE OPTIONAL


AND SPACERS CONTINUOUS
SID MONITOR

FIGURE 2.1 - COMPONENT IDENTIFICATION

November 2005 (2-2) Linear IV Collimator


2.0 INSTALLATION
2.1 UNPACKING
Carefully unpack the equipment and check for damage incurred during shipment. Any
damage should be referred to the agency that delivered the product.

2.2 EQUIPMENT SUPPLIED

Refer to Figure 2-1 for component identification

• Linear IV Collimator
• Spacers and mounting hardware
• Interconnect and power supply unit
• Packet containing Instruction Manual
• Continuous SID Monitor

2.3 COLLIMATOR MOUNTING

1. Determine the collimator mounting surface to focal spot distance from the data
supplied with the X-ray tube (do not rely on an inscribed mark on the tube
housing).

Note: The collimator will not perform properly unless the focal spot to upper
swivel ring distance is 2-7/16” (2.44 inches, 62 mm) +/- 1/32” (.031 inches, 1 mm).
Be sure to include any permanent mounting plates in the focal spot to port boss
distance stated in the tube manufacturer’s data.

Note: The Linear IV is designed to be used with a lead diaphragm or cone in the
plastic port of the X-Ray tube. If it is found that lead diaphragms or cones require
removal or modification, consult the factory.

2. Determine the total thickness of the supplied spacer(s) that must be added to the
collimator mounting surface to obtain a focal spot to collimator mounting flange
distance of 2-7/16” (2.44 inches, 62 mm) +/- 1/32” (.031 inches, 1 mm). Refer
to Figure 2.2. Table 1 in the Appendix provides spacer information for some X-
ray tube housings.

3. Remove the upper swivel ring from the collimator by removing the 6-32 socket
head cap screw and opening the clamp ring.

NOTE: Two different lengths of screws are provided in the cloth bag containing
the spacers. Determine the correct length of screw to use, taking into account
the collimator spacing requirements and/or peculiarities of the tube housing port
boss.

November 2005 (2-3) Linear IV Collimator


DIMENSIONS
FIGURE 2.2

4. Clean the screws and housing port boss with alcohol and if necessary, remove
any debris which may be present in the tube housing mounting holes.

5. SECURELY fasten the upper mounting flange and spacers to the collimator
mounting surface. As a precaution, a medium strength thread locking
compound, such as Loctite #242, should be applied to the screws before
fastening the collimator mounting ring to the tube housing. The use of NYLOK
or other vibration resistant screws is also recommended.

Verify that the collimator mounting screws engage the housing by at least five (5)
threads when used with any required collimator spacer plate(s).

November 2005 (2-4) Linear IV Collimator


6. Carefully support the collimator in place and re-attach the clamping ring. The
hinge of the clamping ring must line up with the pin in the lower mounting ring.
Apply Loctite to the 6-32 socket head screw holding the clamping ring and
securely fasten together.

7. Inspect the fit of the collimator and tube housing. Grasp and attempt to move
the collimator and then the tube housing assembly while inspecting for loose
joints or gaps between the tube/collimator assembly as well as other tube
mounting areas.

WARNING! FAILURE TO ADHERE TO THE ABOVE GUIDELINES MAY RESULT IN


LOOSENING, DAMAGED SCREWS, OR MOUNT FAILURE WHICH CAN RESULT IN
HEAVY COMPONENTS FALLING DURING USE. INCIDENTS OF LOOSE SYSTEM
COMPONENTS SHOULD BE REPORTED IMMEDIATELY TO X-RAY SERVICE FOR
REPAIR.

2.4 POWER SUPPLY CHASSIS MOUNTING

The power supply chassis is a NEMA enclosure intended to be mounted on a wall or in


an equipment cabinet. There are knock-outs on the sides and bottom of the enclosure
for cable entry. Follow all local wiring codes and locate the enclosure in an area that will
permit.

• proper cable bend radii


• sufficient space for convection cooling
• clearance for enclosure door opening

The external connections to the system at the power supply chassis include:

• AC power input
• tube select input
• spotfilm alignment switch
• table image receptor input
• wall image receptor input
• horizontal SID input
• vertical SID input (discrete, continuous, or differential-continuous)
• table tilt monitor
• generator exposure interlock
• TOMO/STEREO by-pass input
• image intensifier select outputs

Refer to Figure 2.3 for power supply chassis dimensions.

NOTE: See the instruction manual that came with the optional Eureka remote
φ
control for installation φ system has the optional Eureka remote
instructions if your
control.

November 2005 ( 2 -5 ) Linear IV Collimator


11.000”
(27.94 cm)

φ φ
10.000”
(25.4 cm)

12.000”
(30.48 cm)
10.000”
(25.4 cm)

4.25”
EXISTING .281” DIA. 10.8 cm
MOUNTING HOLES
FIGURE 2.3

2.5 INTERCONNECT WIRING

CAUTION: YOU WILL BE WIRING 120 VAC OR 220 VAC INTO THE POWER
SUPPLY CHASSIS. BE SURE THAT THE X-RAY GENERATOR IS OFF BEFORE
PROCEEDING.

1. 120 VAC INPUT or 220 VAC INPUT

Connect the three wire cable supplied to the VAC source as follows:

Black - Hot
Wire - Neutral
Green - Ground

The power supply has a power line matching transformer capable of accepting
120 VAC or 220 VAC depending on power source available.

Remove the protective sheet metal cover.

Measure all AC power source with a RMS type voltmeter and record reading.

November 2005 (2-6) Linear IV Collimator


Connect the power source to the transformer tap closest to the power source
voltage read. The taps are numbered as follows for:
115 VOLT OPERATION:

----- Parallel -----

125-V L1 = 4-8, L2 = 1-5


115-V L1 = 3-7, L2 = 1-5
105-V L1 = 2-6, L2 = 1-5

230 VOLT OPERATION:

----- Series -----

250-V L1 = 8, L2 = 1, Connect 4-5


230-V L1 = 7, L2 = 1, Connect 3-5
210-V L1 = 6, L2 = 1, Connect 2-5

The voltage selection switch inside the power supply chasis must be set to
match the incoming line voltage.

November 2005 (2–7) Linear IV Collimator


FIGURE 2.4 - Wiring Block Diagram

November 2005 (2-8) Linear IV Collimator


2. EXPOSURE INTERLOCK

Connect the supplied cable indicated on figure 2.5 and to the exposure interlock
circuit of the generator. The collimator EXPOSURE HOLD is a set of normally
open contacts that remain open in the HOLD condition. The contacts are FORM
C with normally closed contacts available.

TS6

1 RED

4 BLK } TO X-RAY GENERATOR


EXPOSURE HOLD CIRCUIT

5 WHT

FIGURE 2.5

3. TOMO/STEREO INPUT

If the X-Ray system is equipped with either TOMO or STEREO shift capability, it is
necessary to connect a signal to the collimator TOMO/STEREO by-pass input.
This circuit accommodates a variety of signals. Determine the signal level available
from the generator installation manual. Connect as indicated in figure 2.6.

8 24 V

9 15V
AC OR DC
10 5V

7 COM

TOMO/STEREO BYPASS

FIGURE 2.6

November 2005 (2-9) Linear IV Collimator


4. TABLE IMAGE RECEPTOR INPUT*

The Linear collimator systems are designed to operate with input characteristics
as listed below.

Refer to the Cassette Tray Installation Manual and the appropriate Bucky Manual
to assure proper operation of these devices.

Connect the table Image receptor input as indicated in Figure 2.7.

FIGURE 2.7

ELECTRICAL SPECIFICATIONS

• Impedance - 1000 OHMS, + 5% total resistance


• Resistance Linearity - within 1.5%
• Output Voltage vs. Cassette Size with +2.5 volts applied between SUPPLY and COMMON
Terminals is shown in the center column on the following page.

*Spotfilm input on systems equipped with this capability.

November 2005 ( 2 - 10 ) Linear IV Collimator


The following table is for reference only since absolute calibration is not necessary for
correct collimator operation. The response need only be linear over the operation
range of the bucky. AUTOCAL software recognizes and compensates for absolute
error.
CASSETTE/FILM DC OUTPUT
SIZE VOLTAGE
(2.5V INPUT)
INCH CM (metric)
5 0.15
6 0.32
7 0.52
18 0.54
8 0.70
9 0.88
24 0.96
10 1.08
11 1.25
30 1.38
12 1.45
35 1.76
14 1.80
43 2.34
17 2.35

5. AUXILIARY WALL CASSETTE TRAY INSTALLATION

Connect the inputs to the power supply chassis as indicated in figure 2.7 when
an auxiliary wall cassette or size-sensing wall cassette holder is installed in the
system.

6. CONTINUOUS VERTICAL SID MONITOR INSTALLATION

The Linear IV collimator is designed to operate with a continuous SID input as


defined in figure 2.8. If the monitor is built into the equipment, refer to the
manufacturer’s equipment installation manual. Refer to Section 7.0 of this
manual for more information or if the Progeny Continuous SID monitor is to be
installed.

Any SID monitor that meets the following criteria is compatible to the Linear IV
collimator.

Input Impedance < 1 kΩ

Response is linear over the range of operation of the SID monitor.

For accurate sizing, the following conditions must be met: A 10 turn


potentiometer at a 1:1 ratio of movement must measure 14.4 ohms/inch; and a
4:1 ratio must measure 3.6 ohms/inch of movement. A ratio greater than 4:1 is
not recommended.
November 2005 ( 2 - 11 ) Linear IV Collimator
Remove jumper between TS4-2 and TS4-3. Also remove jumper between TS5-
15 and TS5-16. Make electrical connections as shown in Figure 2.8. Adjust
potentiometer to read 450-500 ohms with the tube at 40” SID. At typical voltage
would be 2.5 VDC.
TS4

72”
BLK
2 VERTICAL SID MONITOR
VERTICAL
SID RED 3

36”

Figure 2.8

7. DIFFERENTIAL-CONTINUOUS SID MONITOR INSTALLATION (OPTION)

The Linear IV collimator can accommodate a second SID monitor as described


in Section 7.0. This SID monitor is installed in an adjustable height table to allow
continuous SID operation regardless of table top position. Remove jumper
between TS4-4 and TS4-5. Also remove jumper between TS5-15 and TS5-16.
See Figure 2.9 for electrical connection. A typical 1:1 ratio would be 0.5 VDC
when table is in fully raised position.

TS4

BLK
4
DIFFERENTIAL VERTICAL
5 SID MONITOR
RED

Figure 2.9

8. If the system does not use continuous vertical SID monitoring, it is necessary to
install a switch to be activated at the 40” vertical SID position. The switch must
be installed to activate within + 0.2 inches of the 40 inch SID position. Connect
the normally open contacts to TS5-15 and TS5-16.

If the collimator is permanently mounted in the 40” vertical position, a jumper


may be installed between TS5-15 and TS5-16.

NOTE: If NO continuous SID monitor is installed, place a jumper between


TS4-2 and TS4-3 and also between TS4-4 and TS4-5.
November 2005 ( 2 - 12 ) Linear IV Collimator
9. HORIZONTAL SID INSTALLATION

A switched signal is required at the horizontal SID positions to be used. Only


one switch signal is to be present at any one time. Position each of the switches
to close within + 0.2 inches of their respective SID’s.

Connect the switches to the power supply chassis as indicated in Figure 2.10a
for Non-Tilting tables or Figure 2.10b for Tilting tables.

36” 40” 48” 72”


HORIZ. SID
SWITCHES

R W G O B

° ° ° °
E H R R L
D T N N K

POWER 12 17 18 19 20
CHASSIS TS5

FOR:
NON-TILT TABLE SYSTEMS HORIZONTAL SID SWITCH CKT.

Figure 2.10 (a)

36" 40" 48" 72"

TABLE TREND SID SWITCHES

TABLE UPRIGHT SID SWITCHES

WALL L OR R SID SWITCHES

14 13 12 17 18 19 20 TS5

FOR: HORIZONTAL SID SWITCH CKT.


TILT TABLE SYSTEMS

Figure 2.10 (b)


November 2005 ( 2 - 13 ) Linear IV Collimator
10. TABLE TILT MONITOR

It is necessary to mount the table tilt monitor to transfer angular position


information to the collimator if a tilting table is installed as part of the X-Ray
system. Proper orientation of the monitor is shown in figure 2.11.

Figure 2.11

Remove the jumper between TS6-13 and TS6-12. Connect the table tilt monitor
as indicated in Figure 2.12.

°
T=0
° ° ° GRN

° °
T = +90
BLU

T = -90

° °
BLK WHT RED TAN

13 14 15 12

NOTE: FOR NON-TILT TABLE SYSTEMS, JUMPER TS6-13 TO TS6-12

Figure 2.12

Systems that can tilt and still have the image receptor aligned with the X-ray
source usually have a signal that indicates whether they are aligned or not. If
aligned, exposure should be allowed and SID should be valid. Correct operation
can be achieved by paralleling the table tilt circuitry for the collimator with such a
signal. See Figure 2.13 for connection.
November 2005 ( 2 - 14 ) Linear IV Collimator

POWER SUPPLY CHASSIS

Figure 2.13

11. SPOTFILM ALIGNMENT SWITCH INPUT

Systems that utilize a spotfilm device can use this input to indicate an invalid
condition (e.g. spotfilm misalignment). Remove the factory installed jumper
between TS5-21 and TS5-22 and connect a normally open switch to TS5-21 and
TS5-22 as shown in Figure 2.14. This switch should open whenever an invalid
condition occurs.

21
REMOVE JUMPER
CONNECT N.O. SWITCH
22

SPOT FILM ALIGNMENT SWITCH


Figure 2.14
November 2005 ( 2 - 15 ) Linear IV Collimator
12. RADIOGRAPHIC/FLUOROSCOPIC MODE SWITCH

This input is used to indicate to the collimator that a radiographic (PBL) or a


fluoroscopic mode of operation is requested. Refer to Figure 2.15. A customer
supplied single-pole single throw switch should be installed between TS5-10 and
TS5-11 for combination operation. This switch should be closed when the
system is in radiographic mode, otherwise it should be open.

FLUORO
° RAD
° °
11

10
RADIOGRAPHIC/FLUOROSCOPIC
MODE WIRING

Figure 2.15

NOTE: THE RADIOGRAPHIC MODE MAY ONLY BE ESTABLISHED WHEN


THE X-RAY GENERATOR IS CONFIGURED TO POSITIVELY PREVENT
FLUOROSCOPIC PROCEDURES WHILE ALLOWING CONVENTIONAL
RADIOGRAPHIC TECHNIQUES.

Fluoroscopic/spotfilm only systems must have the factory installed jumper


between TS5-10 and TS5-11 removed.

Radiographic (PBL) systems must have the factory installed jumper between
TS5-10 and TS5-11 connected.

13. FLUORO/SPOTFILM SELECT SWITCH

Special procedure fluoroscopic systems which employ a spotfilm device can use
a set of isolated contacts connected to TS5-8 and TS5-9 as shown in Figure
2.16. This switch indicates to the collimator that spotfilm mode has been
selected. The switch should be configured to be open during fluoroscopic
procedures and closed when a spotfilm is requested. This function is normally
provided by the spotfilm device.
November 2005 ( 2 - 16 ) Linear IV Collimator

FLUORO
°
9
° ° SPOT
FILM
8
P/O SPOT FILM DEVICE

FLUORO/SPOT FILM SWITCH WIRING

Figure 2.16

14. IMAGE INTENSIFIER MODE CONTROL WIRING

This system can be used with any single mode (NORM-II 1), dual mode (MAG 1-
II 2), or tri-mode (Mag2-II 3) image tube.

When using an image intensifier with selectable modes the image intensifier
image diameter must be controlled by the collimator.

NOTE: When using an image intensifier with selectable modes, all other
mode selection controls elsewhere in the system must be disabled or
otherwise prevented from being used by the operator.

A customer supplied 3 conductor cable of appropriate gauge should be


connected to TS5 as shown in Figure 2.17.
TS5

MAG 1
3
° ° “NORMAL”
2 • (NO CONNECTION)

° °MAG 2
1

IMAGE INTENSIFIER MODE CONTROL WIRING


Figure 2.17
November 2005 ( 2 - 17 ) Linear IV Collimator
15. TUBE SELECT INPUT

An indicator must be provided to indicate which tube is in use when the same X-
ray generator can control more than one tube. A PBL OVER-RIDE condition will
be displayed on the collimator with the de-selected tube.

Remove the factory installed jumper between TS1-8 and TS1-9 and connect an
isolated set of contacts to TS1-8 and TS1-9 as shown in figure 2.18. The
contacts should close when the X-ray tube for this particular collimator has been
selected at the generator control.

NOTE: Opening the TUBE SELECT switch will create a PBL OVER-RIDE
condition. Exposure hold from the collimator will not be invoked for that
particular tube circuit.

WARNING! HAZARDOUS AC VOLTAGES MAY STILL BE PRESENT IN THE


POWER SUPPLY CHASSIS WHEN THE TUBE SELECT INPUT HAS BEEN
OPENED.

9
REMOVE JUMPER AND
CONNECT SWITCH TO CLOSE
8 WHEN COLLIMATOR SELECTED

“TUBE SELECT”
INDICATOR WIRING

Figure 2.18
November 2005 ( 2 - 18 ) Linear IV Collimator
16. Some models of the Linear IV Collimator have the capability of accommodating a
remote Field Lamp Switch. The Field Lamp Switch can be remotely activated by
connecting a switch to the power supply as follows. Connect one wire of the switch
to DGND, TS4-4. Connect the other wire of the switch to the white wire with the
red faston tab in the power supply. Refer to Figure 2.19. The switch will close a 5
volt logic sense input to ground, sinking no more than 50 mA DC.

TS-1

DGND
5

To Remote Lamp Switch

Figure 2.19
Remote Lamp Switch Connection

November 2005 ( 2 - 19 ) Linear IV Collimator


NO TEXT

November 2005 ( 2 - 20 ) Linear IV Collimator


SECTION 3

ALIGNMENT PROCEDURES

November 2005 (3-1 ) Linear IV Collimator


3.0 ALIGNMENT PROCEDURES
It is necessary to confirm that the light field accurately represents the X-ray field. The
X-ray field (image) can be compared to the light field by establishing a defined light field
and exposing a film to produce a density of 1.0.

The X-ray field should be determined by exposing a film to a density of 1.0 on the
developed image and observing the points at which the density is just visibly increased
above the base fog background of the film.

In a similar manner, the light field edges should be determined by observing the light
field on a white background. Comparisons may be made by observing the points at
which the light field is just visibly increased over the background illumination and
comparing this with the X-ray field (and to the tolerance marks on the pattern).

The Performance Standards 1020.30 (b) (22) and (45) define the edges of the light field
as the locus of points at which the illumination is one-fourth of the maximum and the
edges of the X-ray field as the locus of points at which the exposure rate is one-fourth of
the maximum.

PLEASE NOTE: Eureka collimators are 100% tested and calibrated at the factory.
Our intention is that you will only have to confirm operation.

3.1 LIGHT FIELD / X-RAY CONGRUENCE TEST

1. Place the X-ray source-to-table distance at 40” (100 cm) SID and lock in place.

2. Turn collimator power on and place into PBL OVERRIDE by jumpering TS5-4
and TS5-5 together.

3. Locate a cassette on the tabletop and accurately center the cassette to the light
field. Mark the position of the cassette on the tabletop.

4. Manually reduce the size of the light/X-ray field to the next smaller film size.

5. Identify the light field edges and carefully mark the edges by placing metal
markers as illustrated below in Figure 3.1.

6. Expose the film to a density of 1.0 and develop.

7. Carefully identify the X-ray field edges and measure the difference between the
X-ray field and edges and the light field edges.

8. The sum of the long axis difference (X1 + X2) shall not exceed 2 percent of the
SID, and the sum of the cross axis difference shall not exceed 2 percent of the
SID.

November 2005 (3-2 ) Linear IV Collimator


9. If adjustment is necessary, continue with Section 3.2. If no adjustment is
necessary skip to section 3.3.

METAL X1
MARKER

X-RAY
FIELD

Y1 Y2

LIGHT
FIELD

X1 + X2 MUST BE LESS THAN 2% OF THE SID


Y1 + Y2 MUST BE LESS THAN 2% OF THE SID

Figure 3.1 - Light Field Congruence

November 2005 (3-3 ) Linear IV Collimator


Cover Removed

Mirror Retractor

Lamp/Lightfield Adjustment Longitudinal Adjustment

CROSS ALIGNMENT LONGITUDINAL ALIGNMENT

1. Adjust two #6-32 screws for 1. Slightly loosen the two #6-32
light field alignment. screws.
2. (Option) Add Loctite to set in 2. Position lamp bracket laterally for
position. field alignment, left to right.
3. Tighten the two #6-32 screws.

Figure 3.2 - Light Field Adjustment

WARNING! THE LAMP AND HEAT DEFLECTORS MAY BE HOT ENOUGH TO CAUSE
SEVERE BURNS. DO NOT TOUCH ANY OBJECT IN THE LAMP AREA
WITH BARE SKIN.

WARNING! THE INTENSITY OF LIGHT OUTPUT IS SUFFICIENT TO TEMPORARILY


IMPAIR YOUR VISION IF ALLOWED TO ENTER THE EYES DIRECTLY .
MAINTAIN A POSITION IN WHICH YOU CAN SEE NEITHER THE
FILAMENT WHEN IT IS OFF, NOR ALLOW LIGHT TO DIRECTLY ENTER
YOUR FIELD OF VISION WHEN IT IS ON.

November 2005 (3-4 ) Linear IV Collimator


3.2 FIELD PROJECTION LAMP AND MIRROR ADJUSTMENT

This procedure must be performed when the field projection lamp is altered from it’s
original position or replaced.

This procedure must also be performed if the original mirror angle has been altered and
if any edge of the developed X-ray image is outside of the tolerance marks as defined in
Section 3.1.

Steps 1 through 9 should be carefully reviewed or repeated prior to a lamp or mirror


adjustment attempt. This is particularly important if only a single testing indicates a
failure to meet the requirements defined in Step 9.

The collimator position and the developed X-ray film must remain undisturbed from the
position defined in Steps 1 through 9 of Section 3.1.

Remove the rear cover and the lamp housing heat shield.

WARNING! THE LAMP AND HEAT DEFLECTORS MAY BE HOT ENOUGH TO


CAUSE SEVERE BURNS. DO NOT TOUCH ANY OBJECT IN THE LAMP AREA
WITH BARE SKIN.

WARNING! THE INTENSITY OF THE LIGHT OUTPUT IS SUFFICIENT TO


TEMPORARILY IMPAIR YOUR VISION IF ALLOWED TO ENTER THE EYES
DIRECTLY. MAINTAIN A POSITION IN WHICH YOU CAN SEE NEITHER THE
FILAMENT WHEN IT IS OFF, NOR ALLOW LIGHT TO DIRECTLY ENTER YOUR
FIELD OF VISION WHEN IT IS ON.

1. If the developed X-ray image (steps 1 through 9 in section 3.1) is off-center in the
longitudinal direction, loosen the two screws securing the lamp housing. See
Figure 3.2.

2. Use a pair of long nose pliers to move the lamp housing slightly until the light
field has shifted to a position that is centered to the developed X-ray image in
the longitudinal direction. Tighten the two screws securing the lamp bracket.

3. If the developed X-ray image (steps 1 through 9 in section 3.1) is in error in the
cross-table direction, adjust the angle of the mirror (using the adjustment screw)
until the light field has shifted to a position that is centered to the developed X-
ray image.

4. If the X-ray image is smaller than the light field, loosen the screws securing the
lamp socket and use a pair of long nose pliers to move the socket further away
from the center of the collimator until light field/X-ray field congruency has been
achieved.

November 2005 (3-5 ) Linear IV Collimator


5. If the X-ray image is larger than the light field, loosen the screws securing the
lamp socket and use a pair of long nose pliers to move the socket closer to the
center of the collimator until light field/X-ray field congruency has been achieved.

6. Repeat steps 1 through 9 in section 3.1 to confirm the results of the above
adjustment.

7. Tighten the lamp bracket screws and replace the rear cover.

3.3 CROSS HAIR WINDOW ADJUSTMENT

This procedure is to be performed if the cross hair shadows are not centered to the light
field (Reference Figure 4.1).

1. Remove the spacer and the entire collimator enclosure.

2. Loosen the screws securing the plastic window.

3. Move the plastic window to align and center the cross hair pattern to the
light field (center lines on the test pattern).

4. Tighten the screws and reassemble the collimator covers.

3.4 BUCKY CENTERING LIGHT-LINE ADJUSTMENT

If the collimator is equipped with a Bucky Light-Line Prism:


This procedure is to be performed if the centering light-line is not centered to the
correctly adjusted light-field. If it is centered replace the collimator enclosure and
skip to section 4.0.

WARNING: THE LAMP AND HEAT DETECTORS MAY BE HOT ENOUGH TO CAUSE
SEVERE BURNS. DO NOT TOUCH ANY OBJECT IN THE LAMP AREA WITH BARE
SKIN.

WARNING: THE INTENSITY OF THE LIGHT OUTPUT IS SUFFICIENT TO


TEMPORARILY IMPAIR YOUR VISION IF ALLOWED TO ENTER THE EYES
DIRECTLY. MAINTAIN A POSITION IN WHICH YOU CAN SEE NEITHER THE
FILAMENT WHEN IT IS OFF, NOR ALLOW LIGHT TO DIRECTLY ENTER YOUR
FIELD OF VISION WHEN IT IS ON.

1. If the centering light-line is off-center to the correctly centered light-field or


exhibits a rainbow of colors along one edge, loosen the two screws securing the
prism/slit bracket See Figure 3.2.

2. Use a pair of long-nose pliers to move the bracket as required to center the light-
line to the correctly adjusted light-field.

November 2005 ( 3 -6 ) Linear IV Collimator


NOTE: IN ORDER TO AVOID THE RAINBOW OF COLORS ALONG THE EDGES,
MAINTAIN THE PRISM IN A POSITION THAT IS CENTERED TO THE BRIGHT
LIGHT-LINE THAT IS OBSERVED ON THE BRACKET AT THE BASE OF THE
PRISM.

3. Tighten the screws and replace the collimator covers.

If the collimator is equipped with a Bucky Light-Line Laser:

WARNING! THE LAMP AND HEAT DEFLECTORS MAY BE HOT ENOUGH TO CAUSE
SEVERE BURNS. DO NOT TOUCH ANY OBJECT IN THE LAMP AREA WITH
BARE SKIN.

WARNING! THE INTENSITY OF LIGHT OUTPUT IS SUFFICIENT TO TEMPORARILY


IMPAIR YOUR VISION IF ALLOWED TO ENTER THE EYES DIRECTLY.
MAINTAIN A POSITION IN WHICH YOU CAN SEE NEITHER THE FILAMENT
WHEN IT IS OFF, NOR ALLOW LIGHT TO DIRECTLY ENTER YOUR FIELD
OF VISION WHEN IT IS ON.

CAUTION! This unit utilizes a low power (2.5 x 10ˉ³ W 650 nm) Class II laser to
produce an alignment beam. DO NOT STARE DIRECTLY INTO
BEAM OR VIEW WITH OPTICAL INSTRUMENTS.

Laser Adjustment

1.) Verify alignment of the field light to the X-ray beam.

2.) Remove the covers from the collimator.


3.) Open the long shutters to the maximum setting.

4.) Close the long shutter to project a .25 to .50 wide beam onto the work surface.

5.) Use a “rigid projection board” (piece of cardboard ~ 12” x 18”) to sweep the activated laser/field
light beam from the work surface up to the bottom of the collimator.

6.) Note the laser line to field light position and alignment at the extremes of the projection board
movement.

7.) Axial angular adjustment: Indicated by an angular mis-alignment of the laser line to the field light
projection. Loosen the 4-40 rotation stop screw and rotate the laser module until the laser line is
parallel and in alignment with the field light projection.

8.) Lateral (side to side) adjustment: Loosen the bracket laser mounting screws and adjust the laser
side to side in the slots provided. Tighten the mounting screws.

9.) Longitudinal angular adjustment: Indicated by the laser line projection moving from one side to
the other of the field light projection. Loosen the module mount screws. Move laser module to
appropriate direction.

10.) Reinstall the covers on the collimator.


November 2005 ( 3 –7 ) Linear IV Collimator
Figure 3.3
Laser Bucky line Projections

Figure 3.4
Reference View of Laser Parts
November 2005 ( 3 –8 ) Linear IV Collimator

(NO TEXT)

November 2005 ( 3 -9 ) Linear IV Collimator


SECTION 4

CALIBRATION

November 2005 (4-1 ) Linear IV Collimator


4.0 CALIBRATION
Your new Linear IV collimator is 100% tested and calibrated at the factory. Re-
calibration on-site is necessary to maximize the collimator’s response to your existing
peripheral equipment and provide for the correct operation of the system installed.
Calibration time has been significantly shortened with the introduction of AUTOCAL.
Calibration is done solely at the front panel and does not require any access to printed
circuit boards during the process.

The LINEAR IV can be calibrated to operate with either English or metric cassettes and
SID's in the radiographic mode. At this time, fluoroscopic operation is only in English
units. Contact the factory if you desire metric units in fluoroscopic operation.

The LINEAR IV is also equipped with write-protected non volatile memory to save
calibration and memory preset information without risk of loss due to power interruptions
and/or electrical interference.

The calibration default is a fixed SID system, with wall bucky and table bucky, and
fluoroscopic/spotfilm capability with a 9” image intensifier. Re-calibration enabled (W1
on).

NOTE: EXPOSURE HOLD IS ENABLED DURING CALIBRATION. IF IT IS


REQUIRED THAT EXPOSURES BE MADE (OPTIONAL STEPS) THEN THE
EXPOSURE HOLD CIRCUIT MUST BE DISCONNECTED FROM THE X-RAY
GENERATOR.

The calibration procedure assumes that installation of all equipment utilized with the
collimator is complete and correct. Section 5.0 will contain checkout procedures that
can be used to verify the collimator is working correctly and provide troubleshooting
information in the event it is not.

NOTE: THE LINEAR IV HAS A HARDWARE CALIBRATION INTERLOCK FEATURE


WHICH DISABLES THE CALIBRATION FUNCTION BUTTON ON THE FRONT
PANEL TO PREVENT UNAUTHORIZED OR ACCIDENTAL CALIBRATION. IF THIS
FEATURE IS ENGAGED THE CAL BUTTON WILL NOT CREATE ANY RESPONSE.
TO BYPASS THE INTERLOCK, THE COLLIMATOR ENCLOSURE MUST BE
REMOVED AND JUMPER W1 ON THE CPU BOARD BE INSTALLED.

NOTE: THE LINEAR IV ALSO HAS A SOFTWARE CALIBRATION INTERLOCK


FEATURE TO PREVENT ACCIDENTAL CALIBRATION. A SECURITY CODE MUST
BE ENTERED TO ACCESS CALIBRATION FEATURES.

Equipment required
1. 14” X 17” (35 cm x 43 cm) FILM Cassette
2. 8” X 10” (18cm x 24 cm) FILM Cassette

November 2005 (4-2 ) Linear IV Collimator


4.1 CALIBRATION PROCEDURE

INSTRUCTION BUTTONS DISPLAY


1. Press the CAL button on the front
panel.
CAL
SERVICE MODE?
The message “Service Mode?”
should appear on the display. If not
see section 4.2

2. Press the ENT/SET key within 4


ENT
seconds to confirm the request, or
else you will be returned to the SET
previous state of operation.
ENTER CODE _______
Upon confirmation the message
“ENTER CODE _______” will be
displayed.

3. Enter the security code needed to


access the calibration function. This
must be done within 10 seconds or
else you will be returned to your
previous state of operation.
ENTER CODE ______*
As you enter the code, * will replace LAST ENTER CODE _____**
the _ on the display for each code SIZE ENTER CODE ____***
entry. ENTER CODE ___****

The security code is the following: MAG


LAST SIZE, MAG SIZE, ENT/SET, SIZE
CAL followed by ENT/SET to enter
the code.
ENT
SET

CAL

ENT
SET

November 2005 (4-3 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

If the incorrect code is entered the


display will show “INVALID CODE”, INVALID CODE
“ACCESS DENIED” and you will be ACCESS DENIED
returned to the previous state of
operation.

If the correct code is entered the CAL


CALIBRATE?
display will show “CALIBRATE?”.
If it does not continue to press CAL
until it does.

If you do not wish to continue at this EXIT TO SYSTEM?


CAL
point, press CAL until “EXIT TO
SYSTEM?” is displayed and press
ENT/SET. You will be returned to
previous state of operation.

NOTE: IF YOU ENTER CALIBRATION IN STEP 4, YOU ARE REQUIRED TO


COMPLETE ALL FOLLOWING STEPS. YOU WILL NOT BE ALLOWED TO EXIT
PREMATURELY.

4. Press ENT/SET to enter calibration. AUTO-CALIBRATION


ENT
The display will show “AUTO-
CALIBRATION” for 4 seconds and SET
then it will display “MODE
ENGLISH?” MODE: ENGLISH?

5. Press Cross Close to display MODE: METRIC?


"MODE METRIC?" FIELD

Press Cross Open to display FIELD MODE: ENGLISH?


"MODE: ENGLISH?"

ENT
Press ENT/SET to execute the SET
OK
displayed response.

6. The LIMITS CAL sets the mechan- DO LIMITS CAL?


ical limits of the collimator shutters
into memory to prevent the collimator
from sizing to an unachievable
size and protect the drive motors.

November 2005 (4-4) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY
ENT
Press ENT/SET to start the LIMITS OK
CAL. SET

Press Cross Close to display “SKIP SKIP THIS STEP?


LIMITS CAL?”. FIELD

Press Cross Open to display “DO FIELD DO LIMITS CAL?


LIMITS CAL?”

ENT
Press ENT/SET to execute the OK
SET
displayed response.

If you decide to skip the LIMITS


CAL proceed to step 9.

7. LIMITS CAL STEP 1


ENT
The display will instruct you to Open shutters to
SET
“Open shutters”, “to maximum field”, maximum field and
“and press ENT”. Follow the press ENT.
instructions to continue.
OK

8. LIMITS CAL STEP 2


ENT
The display will instruct you to SET Close shutters to
“Close shutters”, “to minimum field”, minimum field and
“and press ENT”. Follow the press ENT
instructions to continue. OK

9. LIMITS CAL is now complete, the DO VERTICAL CAL?


display now shows “DO VERTICAL
CAL?”

10. VERTICAL CAL calibrates the


collimator for vertical operation, fixed
or continuous SID and table bucky.
This step is required for correct
vertical operation.

November 2005 (4-5 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY
ENT
Press ENT/SET to start OK
VERTICAL CAL. SET

Press Cross Close to display FIELD SKIP THIS STEP?


“SKIP VERT CAL?”

Press Cross Open to display FIELD DO VERTICAL CAL?


“DO VERTICAL CAL?”

ENT
Press ENT/SET to execute the OK
SET
displayed response.

If you decide to skip VERTICAL


CAL, skip to step 19.

11. VERTICAL CAL STEP 1

Set vertical SID to 40” (100 cm) ENT Set SID to 40” (100 cm)
and press ENT/SET. SET Vert
OK
Set pattern on the x-ray.
ENT Set field to 14” x 14" (35
Set field size to 14” X 14” (35 cm x SET cm x 35 cm)
35 cm) square using the light field OK
and press ENT/SET.
ENT Insert Cassette 14" x 17"
Insert a 14” X 17” (35 cm x 43 cm) SET (35 cm x 43 cm)
film cassette into the table bucky
tray, close the tray, and press
ENT/SET. OK
Either orientation is permitted.
ENT Set Field to 5” x 5" (13 cm
Set field size to 5” X 5” (13 cm x SET x 13 cm)
13 cm) square using the light field OK
and press ENT/SET.

Insert an 8” X 10” film cassette ENT Insert Cassette 8" x 10"


into the table bucky tray, close SET (18cm x 24cm)
the tray, and press ENT/SET.
Either orientation is permitted. OK

November 2005 (4-6 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

12. VERTICAL CAL STEP 2

Use the provided template to size DO NOT MOVE SID.


to the markings of 14” X 14” @ 72” SET FIELD TO 14" x
SID with actual SID set at 40” 14" (35 cm x 35 cm) @
(35 cm x 35 cm @ 180 cm with 72" (180 cm) PER
actual SID at 100 cm) TEMPLATE.

Press ENT/SET to continue. ENT OK


SET

Use the provided template to size to DO NOT MOVE SID.


the markings of 5" x 5" @ 72" SID SET FIELD TO 5" x 5"
with actual SID set at 40" (13 cm x (13 cm x 13 cm) @ 72"
13 cm @180 cm SID with actual SID (180 cm) PER
set at 100 cm). TEMPLATE

Press ENT/SET to continue. ENT OK


SET

13. VERTICAL CAL STEP 3

You will now choose whether you FIXED SID SYS?


have a fixed SID system or a
continuous SID system to the table.

Press Cross Close to choose a FIELD CONT. SID SYS?


continuous SID system.

Press Cross Open to choose a FIELD FIXED SID SYS?


fixed SID system.

Press ENT/SET to confirm the ENT OK


displayed response. SET

If you have a fixed SID system


skip to step 18.

November 2005 (4-7 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

14. VERTICAL CAL STEP 4

The following steps will calibrate SET SID TO 72” (180 cm)
the collimator for use with or
continuous SID monitors. SET SID TO 56” (140 cm)
or
SET SID TO 48” (120 cm)
Press the Cross Open to toggle FIELD or
through selections of target SID’s SET SID TO 36” (92 cm)
for calibration - choose the SID
furthest away from 40” (100 cm)
allowable by the tube support ENT
system, move to that SID, and SET
press ENT/SET. OK

15. VERTICAL CAL STEP 5

You will now choose whether SYS-SINGLE SID?


your system is equipped for
continuous SID (tube moveable
vertically) or differential (tube and
table moveable vertically).

If your system has installed a FIELD


SYS-DIFF SID?
second SID monitor for an
adjustable height table-top,
Press Cross Close.

If your system is equipped only SYS-SINGLE SID?


FIELD
with a single SID monitor, press
Cross Open.

Press ENT/SET to confirm ENT OK


the selection. SET

If you have selected a single


SID system skip to step 18.

16. VERTICAL CAL STEP 6


ENT
Set the adjustable table to the SET TBL FULL UP
SET
full up position and press
ENT/SET. OK

November 2005 (4-8 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

17. VERTICAL CAL STEP 7

Press Cross Open to toggle through FIELD SET TBL DOWN TO:
selections of target table heights 8" (20 cm)
for calibration - choose the lowest or
height from the selections, move to 6" (15 cm)
that setting, and press ENT/SET. or
4" (10 cm)
ENT or
SET 2" (5 cm)
OK
END CONT SID CAL

18. VERTICAL CAL STEP 8

You will now be asked to input the TBL SID OFFSET?


distance in inches from the image
receptor (bucky tray) to the table top.
This value is used to provide an
estimate size and SID for table top
operations.

Press LONG OPEN to increment (Offset distance is


offset by 1" (2.54 cm). FIELD displayed on SID
display).

Press LONG CLOSE to decrement


Offset by 1" (2.54 cm). FIELD

Press CROSS OPEN to increment FIELD


offset by 0.1" (.25 cm).

Press CROSS CLOSE to decrement


FIELD
Offset by 0.1" (.25 cm).

November 2005 (4-9 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

The size is indicated on the


SID display.

Press ENT/SET to store distance ENT OK


offset. SET

The collimator is now calculating WORKING…


the vertical compensation and
calibration parameters for your
system. The process should
take less than 5 seconds.

19. HORIZONTAL CAL STEP 1

Next you will be asked if you want to HORIZONTAL CAL?


perform HORIZONTAL Calibration.
HORIZONTAL Calibration is
Necessary if your system will operate
with a wall bucky or if the collimator
can tilt +90 degrees or –90 degrees.

Press Cross Close to skip FIELD SKIP THIS STEP?


HORIZONTAL CAL.

Press Cross Open to perform FIELD


HORIZONTAL CAL?
HORIZONTAL CAL.

Press ENT/SET to confirm selection. ENT OK


SET HORIZONTAL CAL

If you choose to skip HORIZONTAL


CAL proceed to step 22, else
continue.

Tilt the collimator to the appropriate ENT SET COLL TO WALL


wall where the wall bucky is located SET
and press ENT/SET. OK

November 2005 ( 4 - 10 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

If the tilt is identified incorrectly, a ENT INCORRECT TILT


message indicating incorrect tilt SET
will be displayed. Correct the tilt SET COLL TO WALL
and press ENT/SET. OK

Set horizontal SID to 40” (100 cm) ENT SET SID TO 40"
and press ENT/SET. SET (100 cm)

Set field size to 14” X 14” (35 cm ENT SET FIELD TO 14" x
x 35 cm) square using the light field SET 14" (35 cm x 35 cm)
and press ENT/SET. OK

Insert a 14" x 17" (35 cm x 43 cm) ENT Insert Cassette 14" x


film cassette into the wall bucky, SET 17" (35 cm x 43 cm)
close the tray, and press ENT/SET. OK

Set field size to 5" x 5" (13 cm x 13 ENT Set Field to 5" x 5" (13
cm) square using the light field and SET cm x 13 cm)
press ENT/SET. OK

Insert an 8" x 10" (18 cm x 24 cm) ENT Insert Cassette 8" x 10"
film cassette into the wall bucky, SET (18 cm x 24 cm)
close the tray, and press ENT/SET. OK

20. HORIZONTAL CAL STEP 2

Use the template provided to size the DO NOT MOVE SID


markings 14" x 14" @ 72" with SID SET FIELD TO 14" x
actually set at 40" (35 cm x 35 cm 14" (35 cm x 35 cm) @
with SID actually set to 100 cm). 72" (180 cm) PER
TEMPLATE.

Press ENT/SET to continue. ENT OK


SET

November 2005 ( 4 - 11 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

Use the template provided to size to DO NOT MOVE SID.


the markings 5" x 5" @ 72" with SID SET FIELD TO 5" X 5"
actually set at 40" (13 cm x 13 cm (13 cm X 13 cm ) @ 72"
@180 cm with SID actually set to (180 cm ) PER
100 cm). TEMPLATE.

Press ENT/Set to continue. ENT OK


SET

21. HORIZONTAL CAL STEP 3

The collimator is now calculating the WORKING…


horizontal compensation and
calibration parameters for your
system. This process should take
less than 5 seconds.

22. VERTICAL CAL STEP 9

Next you will be asked if you want to DO FLUORO CAL?


Perform fluoroscopic calibration.
FLUORO CAL is necessary if your
System has fluoroscopic capability
or you intend to use fluoroscopic
functions.

Press Cross Close to skip FLUORO SKIP THIS STEP


FIELD
CAL.

Press Cross Open to perform DO FLUORO CAL?


FIELD
FLUORO CAL.

Press ENT/SET to confirm your ENT OK


selection. SET

If you choose to skip FLUORO CAL


Proceed to step 29.

November 2005 ( 4 - 12 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

23. FLUORO CAL STEP 1 FLUORO CAL

FLUORO CAL allows you to set-up FLUORO CAL


three II magnification sizes
automatically or manually. The
automatic sizing allows you to
choose from a variety of preset II
sizes based on your current
calibration parameters. Manual
sizing allows three independent sizes
to be set for your II sizes.

First you will be asked to input the II FILM OFFSET?


distance in inches from the image
receptor (image intensifier) to the film
plane of the table bucky. This value
is used to provide field size
information with respect to the
image intensifier when it is the
selected image receptor.

Press CROSS OPEN to increment FIELD (Offset distance is


offset by 0.1". displayed on SID
display)

Press CROSS CLOSE to decrement FIELD


offset by 0.1".

Press LONG OPEN to increment


offset by 1". FIELD

Press LONG CLOSE to decrement


offset by 1". FIELD

This size is indicated on the SID


display.

Press ENT/SET to confirm your ENT MANUAL FLUORO CAL


selection. SET

November 2005 ( 4 - 13 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

Press Cross Close to choose AUTO FLUORO CAL


automatic FLUORO CAL. FIELD

Press Cross Open to choose manual FIELD MANUAL FLUORO CAL


FLUORO CAL.

Press ENT/SET to confirm your ENT OK


selection. SET

If you choose automatic FLUORO


CAL skip to step 27.

24. FLUORO CAL STEP 2

Set field size using the light field to ENT SET FLD TO NORM
the desired NORMal II size and press SET
ENT/SET. OK

25. FLUORO CAL STEP 3

Set field size using the light field to ENT Set Fld to MAG1
the desired MAG1 size and press SET
ENT/SET. OK

26. FLUORO CAL STEP 4

Set field size using the light field to ENT Set Fld to MAG2
the desired MAG2 size and press SET
ENT/SET. OK

27. FLUORO CAL STEP 5

For automatic FLUORO CAL you will II MAX SIZE = 9"


be asked to select the largest II size
available.

The sizes for MAG1 and MAG2 will


automatically be determined.

November 2005 ( 4 - 14 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

Press Cross Open to toggle through


choices and then press ENT/SET. FIELD

ENT
SET

MAX II MAG1 MAG2


6 6 6 II MAX SIZE = 6" or
9 6 4.5 II MAX SIZE = 9" or
12 9 6 II MAX SIZE = 12" or
14 12 9 II MAX SIZE = 14" or
16 14 12 II MAX SIZE = 16" or

OK

28. FLUORO CAL STEP 6

FLUORO CAL is complete. END FLUORO CAL

29. Next you will be asked to select SHOW CONT. SID?


whether you would like to display the
continuous SID on the front panel or
if you would like to blank it.

Some X-ray systems have built in


SID indication and it may be
preferable to blank one of the
displays to avoid confusion.

Hiding the display does not affect


the calculation of SID not automatic
operation.

Press Cross Close to blank the FIELD HIDE CONT. SID?


continuous SID display.

Press Cross Open to activate the SHOW CONT. SID?


continuous SID display. FIELD

Press ENT/SET to confirm selection. ENT OK


SET

November 2005 ( 4 - 15 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

30. Next you will be asked to select SHOW HORIZ. SID?


whether you would like to display the
horizontal SID on the front panel or if
you would like to blank it.

Some X-ray systems have built in


SID indication and it may be
preferable to blank one of the
displays to avoid confusion.

Hiding the display does not affect


the calculation of SID not automatic
operation.

Press Cross Close to blank the FIELD HIDE HORIZ. SID?


horizontal SID display.

Press Cross Open to activate the FIELD SHOW HORIZ. SID?


horizontal SID display.

Press ENT/SET to confirm selection. ENT OK


SET

31. Next the collimator will save and STORING DATA


store in permanent memory all of
your configuration items, including
any memory preset field sizes. This
should take less than 15 seconds.

END OF CALIBRATION END OF CALIBRATION

There will be a three second delay and then you will be put back into a runtime mode
in which normal collimator operation will resume.

4.2 DISABLING HARDWARE CALIBRATION INTERLOCK

If you were not required to come to this section from section 4.1, please skip to the next
section.

The following procedure defines the process of removing the hardware calibration
interlock feature of the Linear IV.

November 2005 ( 4 - 16 ) Linear IV Collimator


1. Remove the entire collimator enclosure.

2. Locate W1 on the CPU board and verify the jumper is installed. If W1 is installed
and calibration is still not accessible, return for service. Otherwise install jumper
W1.

3. Re-assemble collimator enclosure.

4.3 ENABLING HARDWARE CALIBRATION INTERLOCK

If desired, a hardware calibration interlock can be employed to prevent unauthorized or


accidental calibrations. This will prevent calibration to be performed unless entry to the
collimator head is established. This is an additional safeguard to the software code
interlock feature.

This section is optional.

1. Remove the entire collimator enclosure.

2. Locate W1 on the CPU board and remove the jumper on W1.

3. Test interlock by pressing the CAL button. If no action occurs the interlock is
established.

4. Re-assemble collimator enclosure.

November 2005 ( 4 – 17 ) Linear IV Collimator


(NO TEXT)

November 2005 ( 4 - 18 ) Linear IV Collimator


SECTION 5

COLLIMATOR
FUNCTION CHECK

November 2005 (5-1 ) Linear IV Collimator


5.0 COLLIMATOR FUNCTION CHECK
This section describes a number of tests which can be performed to verify operation of
the collimator and attached peripheral components.

The first part of the test is semi-automated and does not require knowledge of SID and
or cassette size. Its' purpose is to identify if the drive and feedback mechanisms and
circuits are functioning properly. The test consists of the collimator sizing through a
number of preset field sizes for different SID’s for confirmation by the installer. Data can
also be recorded on an attached table to verify linearity and accuracy of the collimator.

The rest of the tests will confirm operation of some of the peripheral components that
are attached to the collimator system.

Since not all sections of this procedure may apply to a particular situation, it is important
that the assembler thoroughly review this section along with Sections 2.0, 3.0, and 4.0
before starting. Also, different systems require different compliance requirements.
Section 6.0, “COMPLIANCE VERIFICATION” has been written to provide the assembler
with the information needed to assure compliance with the types of systems into which
this collimator may be installed. It is the responsibility of the assembler to determine
which compliance requirements are applicable.

YOU HAVE LEGAL OBLIGATIONS:


PRIOR TO RELEASE OF THE SYSTEM TO THE USER, THE RESULTS OF EACH
STEP MUST BE DEFINED.

ENTER THE APPROPRIATE DATA IN THE SPACES PROVIDED IN THE


COMPLIANCE DATA LOG AND RETAIN FOR YOUR RECORDS AS PROOF THAT
THESE TESTS WERE SUCCESSFULLY PERFORMED.

EQUIPMENT REQUIRED:

A. The X-ray tube support and the table must include angulation indicators in order
to comply with part 1020.31 (d)(2)(i) and 1020.21 (e)(1)(ii). These indicators are
to be used for the following tests.

B. The X-ray tube support device also must include SID indicating means in order
to comply with 1020.31 (e)(1)(i). This indicator will be used in these tests.

C. Measuring Tape (ruler). This is to be used as a backup for the SID indicating
means and as an operational range measurement.

D. Linear IV test patterns 70-09021 and 70-09022 supplied with this manual.

November 2005 (5-2 ) Linear IV Collimator


5.1 COLLIMATOR FUNCTION CHECK PROCEDURES

INSTRUCTION BUTTONS DISPLAY

1. Press the CAL button on the front CAL


panel.

The message “Service Mode?” SERVICE MODE?


should appear on the display.
If not see section 4.2.
ENT
2. Press the ENT/SET key within 4 SET
seconds to confirm the request,
else you will be returned to the
previous state of operation.

Upon confirmation the message


“ENTER CODE ______” will be ENTER CODE
displayed. _______

3. Enter the security code needed to


access the calibration function.
This must be done within 10
seconds or else you will be
returned to your previous state
of operation.

As you enter the code, * will ENTER CODE ___*


replace the _ on the display for ENTER CODE __**
each code entry. ENTER CODE _***
ENTER CODE ****
This security code is the following: LAST
SIZE
LAST SIZE, MAG SIZE, ENT/SET,
CAL followed by ENT/SET to enter
the code. MAG
SIZE
If the incorrect code is entered the INVALID CODE
display will show “INVALID CODE”, ACCESS DENIED
ENT
“ACCESS DENIED” and you will
SET
be returned to the previous state
of operation.
CAL

ENT
SET

November 2005 (5-3) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

If the correct code is entered the CALIBRATE?


display will show “CALIBRATE?”. CAL
Press the CAL button until
“VERIFICATION?” appears. VERIFICATION?

If you do not wish to continue CAL EXIT TO SYSTEM?


at this point, press CAL until
“EXIT TO SYSTEM?” is displayed
and press ENT/SET. You will be
returned to the previous state
of operation.
ENT
Press ENT/SET to enter the COLLIMATOR CHECK
SET
verification mode “COLLIMATOR
CHECK” will appear for three
seconds and then 14” FS @ 40” (35
cm FS @ 100 cm) will size to the 14” FS @ 40” SID or
parameters displayed. 35 cm FS @ 100 cm

4. Verify the size to be 14” X 14” at


40” SID ( 35 cm x 35 cm at 100
SID). Using the supplied template
verify that the illuminated size falls
within the tolerance marks.

Measure the illuminated field


size and record into table 5.1
or make an exposure and record
the size of the exposed area of
the film in table 5.1.

5. Press ENT/SET when finished. ENT 5” FS @ 40” SID or


The display will indicate 5” FS @ 40” SET
13 cm FS @ 100 cm SID
SID (12 cm FS @ 100 cm SID)and
the collimator will size to the
parameters displayed.

6. Verify the size to be 5” X 5” @ 40”


SID (12 cm x 12 cm @ 100 cm
SID). Using the template verify that
the illuminated size falls within the
tolerance marks.

November 2005 ( 5-4) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY
Measure the illuminated field
size and record into table 5.1
or make an exposure and
record the size of the exposed
area of film in table 5.1.

7. Calculate the error and


percent error in table 5.1.

Complete the rest of the table


and verify that the collimator is
operating in a linear fashion as
prescribed by the limits noted
on the table.
NOTE: IF THE COLLIMATOR FAILS THE LINEARITY TEST, YOU SHOULD FIRST RE-
CALIBRATE THE SYSTEM AS DESCRIBED IN SECTION 4. IF AFTER RE-
CALIBRATION THE COLLIMATOR CONSISTENTLY FAILS THE LINEARITY TEST,
RETURN THE COLLIMATOR HEAD FOR SERVICE.

8. Press ENT/SET when finished. ENT 14” FS @ 72” SID


The display will now indicate 14” 35 cm FS @ 180 cm
SET
FS @ 72” SID (35 cm FS @ 180
cm SID) and the collimator will size
to the parameters displayed.

9. Verify the size to be what is 5” FS @ 72” SID or


indicated on the front panel 13 cm FS @ 180 cm SID
display . Using the template
verify that the illuminated size 14” FS @ 36” SID or
falls within the tolerance marks or 35 cm FS @ 92 cm SID
make an exposure and verify
the exposure size is within the 5” FS @ 36” SID or
allowable error of 2% SID. 13 cm FS @ 92 cm SID

14” FS @ 48” SID or


35 cm FS @ 120 cm SID

5” FS @ 48” SID or
13 cm FS @ 120 cm

10. Press ENT/SET when finished ENT


to proceed to the next SET
test size.

November 2005 (5-5 ) Linear IV Collimator


INSTRUCTION BUTTONS DISPLAY

11. Repeat steps 8 - 10 for the EXIT TO SYSTEM?


remaining test sizes until the
message “EXIT TO SYSTEM?”
is displayed.

Press ENT/SET to return to ENT


run-time operation. The collimator SET
function check is complete.

5.2 HORIZONTAL COLLIMATOR FUNCTION CHECK

Repeat section 5.1 with the collimator in the horizontal position to verify collimator
operation in the horizontal position.

5.3 PREPARATION FOR FUNCTION CHECK PROCEDURES

Access to the power chassis PCB (Master PCB) may be necessary for the following
procedures. Make sure the collimator is in a run-time condition with a cassette installed
in the table cassette holder. The collimator should not be in calibration, verification, nor
cycle modes during these procedures.

5.4 COLLIMATOR TILT SWITCH CHECKOUT

Position the collimator at 0° horizontal. Angulate the collimator to a 0° beam-down


position. Slowly angulate the collimator +20° toward a beam left position and back to a
-20° beam-right position.

Observe the message MANUAL-TILT on the collimator front panel, observe the field
size indicators “___ ____”, and observe the SID display blanks. This should occur
whenever the collimator is tilted more than 11° in either direction from vertical. If this
does not occur, the tilt switch assembly is probably defective. It must be replaced as a
unit. Refer to the Replaceable Parts List in Section 8.

5.5 TABLE TILT SWITCH CHECKOUT (TILTING TABLES ONLY)

With a cassette installed in the table cassette holder, angulate the collimator to 0°
beam-down. Angulate the table to 0° horizontal and slowly tilt to +20° and then to
-20°. A READY-RAD message must be displayed with the table at 0° and remain on for
+11° and -11° from horizontal.

Position the table at +90° upright (usually actual of +85°) and angulate the collimator to
a +90° beam-left position. Slowly tilt the table 20° down from the maximum upright
position (+70°). The MANUAL-TILT message must be displayed at a position 11° from
+90° (+79° and remain on for tilt to 70°.

November 2005 ( 5-6 ) Linear IV Collimator


5.6 CONTINUOUS SID MONITOR CHECKOUT

Make sure the collimator is in run-time operation, in radiographic mode, with READY-
RAD displayed on the front panel.

Adjust the position of the head to 40” (100 cm) SID and verify that the SID display on
the front panel indicates 40” (100 cm) SID within the prescribed tolerances in the
specifications section in the beginning of the book.

Adjust the position of the head to numerous other SID’s and verify the display indicates
the correct SID within the prescribed limits each time.

(OPTION) If the system is equipped with an adjustable height table top and a second
SID monitor is installed. Verify SID and display by also moving the table top to
numerous table heights.

Above 72.0” (180 cm) SID and below 36.0” (92 cm) SID a message of EXP HLD - SID
LIM will be displayed to indicate that SID has passed the working limits of the collimator.
The SID display should blank.

5.7 EXTERNAL SID SIGNAL TEST (HORIZONTAL)

Angulate the collimator to 90 degrees at the wall mounted cassette holder. Insert a
cassette into the tray and verify that the collimator is in run-time operation with READY-
RAD displayed on the front panel.

Move the collimator horizontally until the distance from the source (focal spot) to the
image receptor (film in cassette) is at 72” (180 cm) SID. Slowly move the collimator to a
greater and then to a lesser SID while measuring the actual SID.

72.0 (180.0) will be displayed on the collimator front panel SID display at a measured
72” (180 cm) SID and must switch off at a maximum of 72.75” (182 cm) and at a
minimum of 71.25" (178 cm) SID.

Move the collimator horizontally to a 48” (120) SID, and slowly move the collimator to a
greater and then to a lesser SID.

48.0 (120.0) will be displayed on the collimator front panel SID display at a measured
48” (120 cm) SID and must switch off at a maximum of 48.48” (50 cm) and at a
minimum of 47.52” (46 cm) SID.

Move the collimator horizontally to a 40” (120 cm) SID, and slowly move the collimator to
a greater and then to a lesser SID.

40.0 (100.0) will be displayed on the collimator front panel SID display at a measured
40” (100 cm) SID and must switch off at a maximum of 40.40” (102 cm) and at a
minimum of 39.60” (98 cm) SID.

November 2005 ( 5-7 ) Linear IV Collimator


Move the collimator horizontally to a 36” (92 cm) SID, and slowly move the collimator to
a greater and then to a lesser SID.

36.0 (92.0) will be displayed on the collimator front panel SID display at a measured 36”
(92 cm) SID and must switch off at a maximum of 36.36” (94 cm) and at a minimum of
35.64” (90) SID.

5.8 TILTING TABLE RECEPTOR SID SIGNAL CHECKOUT


(FOR TILTING TABLES ONLY)

Angulate the table to a full upright position. Angulate the collimator beam-left to aim the
X-Ray beam at the upright table cassette tray. Place a cassette in the cassette tray and
verify that the collimator is in a run-time mode with READY-RAD displayed on the
collimator front panel.

Move the collimator horizontally to a 48” (120 cm) SID, and slowly move the collimator to
a greater and then to a lesser SID.

48.0 (120.0) will be displayed on the collimator front panel SID display at a measured
48” (120 cm) SID and must switch off at a maximum of 48.48” (122 cm) and at a
minimum of 47.52” (124 cm) SID.

Move the collimator horizontally to a 40” (100 cm) SID, and slowly move the collimator to
a greater and then to a lesser SID.

40.0 (100.0) will be displayed on the collimator front panel SID display at a measured
40” (100 cm) SID and must switch off at a maximum of 40.40” (102 cm) and at a
minimum of 39.60” (98) SID.

Move the collimator horizontally to a 36” (92 cm) SID, and slowly move the collimator to
a greater and then to a lesser SID.

36.0 (92.0) will be displayed on the collimator front panel SID display at a measured 36”
(92 cm) SID and must switch off at a maximum of 36.36” (94 cm) and at a minimum of
35.64” (90 cm) SID.

5.9 EXTERNAL RADIOGRAPHIC IR SIGNALS CHECKOUT


(TILTING OR NON-TILTING TABLES)

Angulate the collimator to a 0° beam-down direction with a measured SID of 40” (100
cm) measured to the bucky. For tilting tables, angulate the table to a 0° horizontal. The
MANUAL-NO CASS message should be displayed on the collimator front panel.

1. Locate a cassette in the table cassette tray.

A. EXPOSURE HOLD should be displayed while the collimator rapidly


adjusts to the desired size and stops.

November 2005 ( 5-8 ) Linear IV Collimator


B. READY-RAD should then display, the current SID and the current field
size will be indicated on the front panel.

C. The displayed field size should be the same as the cassette size inserted.

2. Do not remove the cassette at this time. Angulate the collimator to properly aim
at the wall cassette holder with horizontal SID at about 50” (127 cm).

NOTE: Although the cassette was left in the table cassette tray, it should not
produce an IR TRUE signal, thus MANUAL - SID FIX? should be displayed
indicating that the horizontal SID is not at a fixed position.

3. Fully insert a cassette into the wall cassette holder and move the collimator
horizontally to each of the available SID’s (i.e. 36”, 40”, 48”, and 72” or 92cm,
100cm, 120 cm, and 180 cm).

A. The correct SID value should be indicated.

B. The collimator should rapidly adjust to the desired size and stop. The
EXPOSURE HOLD message will change to a READY -RAD message
when the collimator stops.

C. The field size indicated on the display should be the same as the
cassette size inserted in the wall cassette holder.

4. Remove the cassette from the table cassette tray. The removal of this cassette
should not produce a response.

5.10 EXTERNAL FLUOROSCOPIC IR SIGNALS CHECKOUT


(TILTING OR NON-TILTING TABLES)

1. Angulate the collimator to a 0° beam-down direction with a measured SID OF


40” measured to the input screen of the image intensifier. Select
FLUORO/SPOTFILM mode. For tilting tables, angulate the table to 0°
horizontal. The collimator should adjust to the previously selected image
intensifier magnification size and READY-FLUORO should be displayed on the
collimator front panel.

2. Press the MAG SIZE button on the front panel or optional remote control.

A. The collimator should rapidly adjust and stop.

B. The light field size at the plane of the image receptor should be
approximately the same size as preset for NORM during calibration.

B. READY-FLUOR NORM should be displayed on the collimator front panel.

November 2005 ( 5-9 ) Linear IV Collimator


3. Press the MAG SIZE button on the front panel or optional remote control again.

A. The collimator should rapidly adjust and stop.

B. The light field size at the plane of the image receptor should be
approximately the same size as preset for MAG1 during calibration.

C. READY-FLUOR MAG1 should be displayed on the collimator front panel.

4. Press the MAG SIZE button on the front panel or optional remote control again.

A. The collimator should rapidly adjust and stop.

B. The light field size at the plane of the image receptor should be
approximately the same size as preset for MAG2 during calibration.

C. READY-FLUOR MAG2 should be displayed on the collimator front panel.

5. Increase the size of the light field by means of the front panel or optional remote
control power assist shutter control button.

The light field size should not enlarge beyond the currently selected image
intensifier magnification size.

5.11 TILTING TABLE EXTERNAL RADIOGRAPHIC IR SIGNAL CHECKOUT

1. Select RAD mode. Tilt the table to it’s full upright position. Up to +79° and
without a cassette in the table bucky, the message MANUAL-TILT should be
displayed on the collimator front panel.

2. Angulate the collimator to the +90° beam-left position. The message MANUAL-
NO CASS should be displayed on the collimator front panel.

3. Insert a cassette into the cassette tray and fully insert the tray into the table.
Adjust the SID to a position near 46" (117 cm). The message EXP HLD-LOCK
SID will be displayed on the collimator front panel.

4. Move the collimator horizontally to each of the available SID’s (i.e. 36”, 40”, and
48” or 92 cm, 100 cm or 120 cm).

A. The corresponding SID should be displayed on the SID display of the


collimator front panel.

B. The collimator should rapidly adjust and stop while the message
EXPOSURE HOLD changes to READY-RAD on the collimator front
panel.
C. The field size indicated should be approximately the same size as the
inserted cassette.

November 2005 ( 5-10 ) Linear IV Collimator


D. Repeat steps 1-4 of this section with the table and collimator tilted to -90° beam-
right position.

5.12 SPOTFILM IR SIGNALS CHECKOUT

1. Locate a cassette in the table cassette tray and fully insert the tray into the table.
Select FLUORO/SPOTFILM mode.

A. The collimator shutters adjust to the film size.

B. The message READY-SPOTFILM should be displayed on the collimator


front panel.

NOTE: Pressing the LAST SIZE button on the collimator front panel or the
optional remote control will cause the collimator to size to the last
fluoroscopic size that was in use.

5.13 STEREO/TOMO BYPASS SIGNAL CHECKOUT

1. Angulate the collimator to 0° beam-down direction with a measured SID of 40”


measured to the bucky. For tilting tables, adjust the table to 0° horizontal. The
message MANUAL-NO CASS should be displayed on the collimator front panel.

2. Locate a cassette in the table cassette tray and fully insert the tray into the table.

A. The message EXPOSURE HOLD will be displayed while the collimator


rapidly adjusts and stops. Then READY-RAD will be displayed.

B. The indicated size should be approximately the same size as the inserted
cassette size.

3. Create a STEREO/TOMO condition by placing jumpers between TS6-7 to


ground, and between TS6-10 and +5VDC. The message STEREO/TOMO
should be displayed on the collimator front panel and the collimator should be in
MANUAL mode.

4. Remove the jumpers. The collimator should return to the state described in 2.

5.14 NOTES ON CASSETTE TRAY ADJUSTMENT

The table cassette tray and the wall cassette holder do not need to be adjusted to be
synchronous with each other. The collimator retains separate calibration parameters for
both image receptors.

November 2005 ( 5-11 ) Linear IV Collimator


The collimator knows which tray it is shooting at when installed properly, therefore it is
unnecessary for the table and wall cassette tray potentiometers to be set exactly the
same.

The collimator has been calibrated at the factory around a cassette tray that is calibrated
for an 11” (12-1/8” outside) or 28 cm ( 31 cm outside) cassette. The 11” (28 cm)
dimension represents exactly one half of the 1 kilo-ohm cassette tray potentiometer.
Consult the tray manufacturer’s manual if adjustment is necessary.

NOTE: THE LINEAR IV’S AUTOCAL FEATURE “LEARNS” THE SETTINGS OF


YOUR TRAY, THEREFORE MINOR DEVIATIONS FROM THE CHART BELOW
ARE ACCEPTABLE AND ADJUSTMENT OF THE TRAY UNNECESSARY.
THE ONLY REQUIRED CONDITION IS THAT THERE IS A LINEAR
RELATIONSHIP BETWEEN CASSETTE SIZE AND OHMIC OUTPUT.

Cassette Tray Ohms Between Ts-13 or 14


English Metric and TS4-12
5" 13 cm 60 Ohms
7" 18 cm 207
8" 20 cm 280
10" 24 cm 427
11" 27 cm 500
12" 30 cm 573
14" 35 cm 720
17" 43 cm 940

1 2 3 4 1 5 2 3 8 4 6 7

5 6 7 8
CROSS LONG
° ° CASSETTE
PRESENT
SWITCH

November 2005 ( 5-12 ) Linear IV Collimator


DESIRED SIZE ACTUAL SIZE ERROR=DESIRED %ERROR = ERROR X 100
DESIRED SIZE
SIZE-ACTUAL SIZE

LONG 14” (35 cm) ERROR1=

CROSS 14” (35 cm) ERROR2=

LONG 5” (13 cm) ERROR3=

CROSS 5” (13 cm) ERROR4=

NOTE: Keep appropriate sign in all calculations

collimator operation is linear if

Error 1 - Error 3 < 1%

AND

Error 2 - Error 4 < 1%

ERROR 1-ERROR 3 < 1% PASS/FAIL

< 1%
ERROR 2 - ERROR 4 PASS/FAIL

TABLE 5.1

November 2005 ( 5-13 ) Linear IV Collimator


DESIRED SIZE ACTUAL SIZE ERROR=DESIRED %ERROR = ERROR X 100
DESIRED SIZE
SIZE-ACTUAL SIZE

LONG 14” (35 cm) ERROR1=

CROSS 14” (35 cm) ERROR2=

LONG 5” (13 cm) ERROR3=

CROSS 5” (13 cm) ERROR4=

NOTE: Keep appropriate sign in all calculations

collimator operation is linear if

Error 1 - Error 3 < 1%

AND

Error 2 - Error 4 < 1%

ERROR 1-ERROR 3 < 1% PASS/FAIL

< 1%
ERROR 2 - ERROR 4 PASS/FAIL

TABLE 5.1

November 2005 ( 5-14 ) Linear IV Collimator


SECTION 6

COMPLIANCE
VERIFICATION

November 2005 ( 6-1 ) Linear IV Collimator


Figure 6.1

LEFT SIDE MOUNTING HOLES FOR COMPLIANCE TESTING


Figure 6.2

November 2005 ( 6-2 ) Linear IV Collimator


6.0 COMPLIANCE VERIFICATION
It is necessary for the assembler to verify compliance. A series of tests, when
performed at the time of installation, will indicate compliance with 21 CFR, Sub-Chapter
J, Part 1020, Performance Standards.

The following tests are from NEMA Standards Publication, No. XR-8-1979 (Test
Methods for Diagnostic X-Ray Machines for Use During Initial Installation).

For each compliance item, there may be a variety of test methods described. Which
method is used will depend on the tester’s experience, availability of equipment, time, or
special requirements of the Eureka Linear Collimator. Any reference to tolerances on
compliance items are referenced directly from 21 CFR, Sub-Chapter J, Regulations.
They do not take into account inaccuracies brought about by the test equipment,
instrumentation, or the human element. These factors must be considered when these
tests are performed and the compliance of the equipment is being determined.

6.1 VERIFICATION OF TESTS TO BE PERFORMED

Test Procedure or Requirement Applicable Paragraph

1. Visual Definition of X-Ray Light Field XR8/2.14


2. Intensity of Light Field Illumination XR8/2.15
3. Minimum Field Size XR8/2.16
4. X-Ray Field/Receptor Center Alignment XR8/2.17
5. Indication of Field Size XR8/2.18
6. Positive Beam Limitation (PBL) XR8/2.19
7. X-Ray Field Limitation and Alignment XR8/2.20
8. Return to PBL with Image Receptor Change XR8/2.21
9. Beam Quality XR8/2.09
10. Fluoro X-Ray Field Limitation XR8/2.24
11. Fluoro Source to Skin Distance (SSD) XR8/2.25
12. Primary Barrier - X-Ray Lockout XR8/2.27
13. Alignment of X-Ray Edges (Spotfilm) XR8/2.29
14. X-Ray/IR Center Alignment (Spotfilm) XR8/2.30

RECORD THE RESULTS ON THE RECORD SHEET SUPPLIED AT THE END OF


THIS SECTION

Radiation Warning for Diagnostic X-Ray Systems

X-rays are dangerous for both the operator and others in the vicinity unless established
safe exposure procedures are strictly observed.

November 2005 ( 6-3 ) Linear IV Collimator


The useful and scattered beams can produce serious, genetic or potentially fatal bodily
injuries to any persons in the surrounding area if used by an unskilled operator.
Adequate precautions must always be taken to avoid exposure to the useful beam, as
well as leakage radiation from within the source housing or to scattered radiation
resulting from the passage of radiation through matter.

Those authorized to operate, test, participate in or supervise the operation of the


equipment must be thoroughly familiar and comply completely with the currently
established safe exposure factors and procedures described in publications such as
Sub-Chapter J of Title 21 of the Code of Federal Regulations, “Diagnostic X-Ray
Systems and their Major Components,” and the National Council on Radiation Protection
(NCRP) No. 33, “Medical X-Ray and Gamma-Ray Protection for Energies up to 10 Me
V-Equipment Design and Use,” as revised or replaced in the future.

Failure to observe these warnings may cause serious, genetic or potentially fatal bodily
injuries to the operator or those in the area.
Dotted line represents perimeter
of light field and corresponds to
Edge of the X- outer edge of metal strips
Ray field

L1

W2

W1

L2 = 0
Only in this example

Edge of Direct Print Paper

Figure 6.3

November 2005 ( 6-4 ) Linear IV Collimator


XR 8-2.14 VISUAL DEFINITION (RADIOGRAPHIC) OF X-RAY LIGHT FIELD
REQUIREMENT - Means shall be provided for visually defining the perimeter of the X-
ray field. The total misalignment of the edges of the visually defined field with the
respective edges of the X-ray field along either the length or width of the visually defined
field shall not exceed 2 percent of the distance from the source to the center of the
visually defined field when the surface upon which it appears is perpendicular to the axis
of the X-ray beam (see 21 CFR 1020.31 (d) (2)).

.01 METHOD 1 - FDA COMPLIANCE TEST METHOD

A. EQUIPMENT REQUIRED

1. BRH/FDA compliance test stand (including slide assembly). See Figure


6.1.

2. Four metal marker strips.

3. Plastic cassette, loaded with direct-print paper or film.

B. PROCEDURE

1. Attach the spacer, positioned out of the primary beam to the test stand.
Center the stand on the table. Center the source over the stand, assure
by the means provided that the axis of the X-ray beam is perpendicular to
the plane of the image receptor, and bring the beam-limiting device down
into firm contact with the spacer. Select the MANUAL mode of operation
(there must not be a cassette in the cassette holder).

2. Insert the slide assembly, grid side up, into slot 6 of the test stand and
the focal spot assembly into slot 1 (Figure 6-2). Place a cassette loaded
with direct print paper or film into the slide assembly.

3. Adjust the collimator so that no part of the light-field intersects any portion
of the top of the test stand. (Further collimation to a light field of less
than 15 by 20 centimeters (6 by 8 in) on the slide assembly grid may be
desirable to assure that the X-ray field will be fully contained on the direct
print paper for film in the slide assembly).

4. Position the outer edge of each metal strip to correspond with each side
of the light-field. One end of the metal strip shall extend to the center line
of the respective grid arm.

5. Select proper technique factors and make an exposure (may require


several exposures to obtain 1 R to the direct print paper).

6. Develop the direct print paper or film.

November 2005 ( 6-5 ) Linear IV Collimator


C. VERIFICATION OF COMPLIANCE

For determination of misalignment, compare the edges of the X-ray field to the
edges of the light-field as defined by the outer edges of the metal strips. On
each side of the rectangular fields, measure the separation between the X-ray
field and the outside edge on the image of the respective metal strip. Sum these
measured separations for opposite sides of the X-ray field to yield a total
misalignment in the length and width dimensions. Record the length
misalignment and width misalignment, both without regard to sign (see
Paragraph D on following page and Figure 6-3).

D. CALCULATIONS

Calculate the source to image distance (SID) per the following formula (to slot 6)
as the indicated source-to-table top distance minus 4.7 centimeters (1.85 in) and
record. Calculate 2 percent of this SID and record. Both the length and width
misalignment must be less than 2 percent of SID (to slot 6).

2.5 = X
S X + 13.95

2.5X + (2.5) 13.95 = XS

(2.5) 13.95 = XS - 2.5X See Fig. 6-4

34.875 = X (S - 2.5)

X = 34.875
S - 2.5

The misalignments are calculated as follows:

Length misalignment = L1 + L2 ≤ 2% SID


Width misalignment = W 1 + W2 ≤ 2% SID

Calculate 2% of the measured SID. Each of the misalignments, length or width,


must be less than or equal to 2% of the measured SID for compliance.
NEMA Standards 5-15-79

November 2005 ( 6-6 ) Linear IV Collimator


FOCAL SPOT

BRASS
FOCAL SPOT ASSEMBLY STRIPS X
IN SLOT 1

2.5”
TEST STAND
13.95”
SLIDE ASSEMBLY IN
SLOT 6

TABLETOP

DETERMINATION OF SID

Figure 6.4
DOTTED LINE COINCIDES WITH THE
OUTHER EDGE OF THE METAL
STRIPS AND IS THE PERIMETER OF
THE LIGHT FIELD

METAL STRIPS

Figure 6.5

November 2005 ( 6-7 ) Linear IV Collimator


.02 METHOD II - METAL MARKER METHOD

A. GENERAL

The actual versus indicated source-to-image receptor distance (SID) test must
be performed prior to attempting this test.

B. EQUIPMENT

1. Plastic cassette with direct-print paper or film.

2. Radio-opaque markers*.

* Each marker is approximately 1/32 inch galvanized sheet metal having the dimensions
of 1.5 by 1.5 inches.

C. PROCEDURE

1. Adjust the source assembly and the beam-limiting device so that they are
approximately centered over the table and perpendicular to the table top.
Then position the beam-limiting device to the SID previously determined
and record the indicated value.

2. Insert the cassette and turn on the light-field.** Adjust the beam-limiting
device to the next size smaller than the cassette size being used.

** Make a note to record the field size indicated on the dial of the beam-limiting device for
the SID being used.

3. Position the outer edge of each metal marker on the table top to
correspond with each side of the light-field (Fig. 6-5).

4. Select the appropriate technique factors and make an exposure.

5. Develop film or direct-print paper.

D. VERIFICATION OF COMPLIANCE

For determination of misalignment, compare the edges of the X-ray field to the
edges of the light-field as defined by the outer edges of the metal strips. On
each side of the rectangular fields, measure the separation between the X-ray
field and the outside edge of the image of the respective metal strip. Sum these
measured separations for opposite sides of the X-ray field to yield a total
misalignment in the length and width dimensions. Record the length
misalignment and width misalignment, both without regard to sign (see Par. E
below and Fig. 6-3).

November 2005 ( 6-8 ) Linear IV Collimator


E. CALCULATIONS

Calculate the SID per the following formula (to slot 6) as the indicated source-to-
tabletop distance minus 4.7 centimeters (1.85 in) and record. Calculate 2
percent of this SID and record. Both the length and width misalignment must be
less than 2 percent of SID (to slot 6).

2.5 = X
S X + 13.95

2.5X + (2.5) 13.95 = XS

(2.5) 13.95 = XS - 2.5X See Fig. 6-4

34.875 = X (S - 2.5)

X = 34.875
S - 2.5

The misalignments are calculated as follows:

Length misalignment = L1 + L2 ≤ 2% SID


Width misalignment = W 1 + W2 ≤ 2% SID

Calculate 2% of the measured SID. Each of the misalignments, length or width,


must be less than or equal to 2% of the measured SID for compliance.
NEMA Standards 5-15-79
.03 METHOD III - ALTERNATE TEST STAND METHOD

A. GENERAL

1. The image of the radiation field on the film must be of uniform density
with sharply defined edges.

2. The graduated template is utilized to minimize the amount of error


introduced into the measurement of the X-ray field size.

3. The actual versus indicated SID must be determined prior to performing


this test.

B. EQUIPMENT

1. Manufacturer’s recommended test stand.

2. Cassettes and film.

3. Graduated template.

November 2005 ( 6-9 ) Linear IV Collimator


C. PROCEDURE

1. Align the tube unit and image receptor and set the SID with the normal
operating aids (detents, scales, lights, etc.)

2. Load cassette and insert into image receptor.

3. Close shutters to a size smaller than that of the cassette placed into the
image receptor.

4. Position the test stand in accordance with the manufacturer’s instructions.

5. Energize the field light and record or define the position of the four light
field edges as shown on the graduated template or position four metal
markers so that the outer edge of each metal marker corresponds to an
edge on each side of the light-field or both.

6. Select proper technique factors, make an exposure, and develop film.

D. VERIFICATION OF COMPLIANCE

1. Calculate 2 percent of the actual SID and record.

2. Compare the edges of the X-ray field to the edges of the light field as
defined by the outer edges of the metal markers or by the graduated
scale.

3. Measure the distance between the edges of the two fields for each side
of the rectangular fields (see Fig. 6.3).

4. Arithmetically sum the misalignment of opposite sides, regardless of sign,


of the rectangles, to yield misalignment in each of the two directions.

Length misalignment = L1 + L2 ≤ 2% SID


Width misalignment = W 1 + W2 ≤ 2% SID

Both the length and width misalignment must be less than 2 percent SID
as calculated in Step 1.
NEMA Standards 5-15-79

XR 8-2.15 INTENSITY OF LIGHT FIELD ILLUMINATION


REQUIREMENT - When a light localizer is used to define the X-ray field, it shall provide
an average illumination of not less than 160 lux (15 footcandles) at 100 centimeters or at
the maximum SID, whichever is less. The average illumination shall be based on
measurements in the approximate center of each quadrant of the light field (See 32
CFR 1020.31 (d) (2) (ii) ).

November 2005 ( 6-10 ) Linear IV Collimator


.01 METHOD 1 - DIRECT TEST

A. GENERAL

1. Make certain that all surfaces in the light path are clean.

2. Reduce ambient light level as much as is feasible.

B. EQUIPMENT

Photometer capable of measuring 160 lux (15 footcandles).

C. PROCEDURE

1. Place the photometer on the tabletop and set the diagnostic source assembly so that the
sensing area of the photometer is at 100 centimeters or the maximum SID, whichever is less.

2. Open the beam-limiting device to assure that each quadrant of the light field is larger than the
sensing area of the photometer.

3. Refer to the manufacturer’s instructions for proper use of the photometer.

4. Turn on the light localizer.

5. At or near the center of a light field quadrant, determine the illuminance by subtracting the
ambient light level from the corresponding light level as measured when the light localizer is
energized. Do not move the photometer between measurements.

6. Repeat the procedure for the remaining three quadrants.

7. Determine the average illuminance of the four light field quadrants.

8. Record the model number, serial number, and the date of calibration of test instrument.

D. VERIFICATION OF COMPLIANCE

Verify that the average illumination is not less than 160 lux (15 footcandles).
NEMA Standards 5-15-79

.02 METHOD II - INDIRECT TEST

A. GENERAL
1. This indirect test is feasible after the correlation between light output and
voltage is made; the manufacturer then specifies a voltage to be
measured or adjusted, or both.
2.
November 2005 ( 6-11 ) Linear IV Collimator
2. Make certain that all surfaces in the light path are clean and unobstructed.

B. EQUIPMENT
Digital voltmeter.

C. PROCEDURE

1. Remove trim covers to gain access to the lamp socket.

2. Verify that the specified lamp is in the socket.

3. With the light-field energized, measure the voltage across the lamp socket terminals.

4. Record the voltage measured.

5. Record the model number, serial number and calibration date of the digital voltmeter.

D. VERIFICATION OF COMPLIANCE

The voltage recorded shall be within the tolerances specified by the manufacturer.

NOTE: THE AC VOLTAGE AT THE LAMP SOCKET MUST NOT BE LESS


THAN 19.5 VAC RMS

XR 8-2.16 MINIMUM FIELD SIZE


REQUIREMENT - Minimum field size at 100 centimeters (radiographic) or the
maximum SID (fluoroscopic) shall be less than or equal to 5 by 5 centimeters (see 21
CFR 1020.31 (d) (1), 1020.31 (e) (2), and 1020.32 (b) (2) ).

.01 METHOD 1 - FILM METHOD

A. GENERAL

1. The following test is to be used for radiographic, fluoroscopic, and spotfilm devices.

2. This procedure need not be performed if it is apparent by visual means that the beam-limiting
device can be adjusted to a size less than 5 by 5 centimeters at the specified SID.

November 2005 ( 6-12 ) Linear IV Collimator


B. EQUIPMENT

1. Cassette.

2. X-ray film or direct-print paper.

C. PROCEDURE 1 - RADIOGRAPHIC AND SPOTFILM DEVICES


1. Adjust the maximum SID obtainable (spotfilm devices or 100 centimeter radiographic devices).

2. Adjust the beam-limiting device to the smallest field size obtainable.

3. Load cassette and set proper technique factors.

4. Make an exposure.

5. Process film.

D. PROCEDURE 2 - FLUOROSCOPIC

1. Set fluoroscopic system for maximum SID and lock into position.

2. Remove all compression cones from the beam.

3. With the X-ray beam off, attach beam attenuator to the input surface of the image receptor.

4. Attach cassette to the bottom of the attenuator.

5. Close shutters as far as possible.

6. Set technique factors to assure proper exposure of the film.

7. Make exposure.

8. Process film.

E. VERIFICATION OF COMPLIANCE

Measure the X-ray field produced on film and verify that the field size is less than
or equal to 5 by 5 centimeters.
NEMA Standards 5-15-79

November 2005 (6-13 ) Linear IV Collimator


.02 METHOD II - VISUAL METHOD

A. GENERAL

1. The following test is to be used for radiographic, fluoroscopic, and spotfilm devices.

2. This procedure need not be performed if it is apparent by visual means that the beam-limiting
device can be adjusted to a size less than 5 by 5 centimeters at the specified SID.
3. These tests attempt to minimize radiation exposures as well as film processing by utilizing light-
field visual display, or image display, or visual mechanical movement.

B. EQUIPMENT

Beam attenuator

C. PROCEDURE 1 - RADIOGRAPHIC

1. Set the tube unit at a distance of 100 centimeters from the image receptor.

2. Pull out the bucky tray and place an unloaded cassette into the tray. Do not reinsert bucky tray.

3. Position source assembly over film cassette in the extended bucky tray.

4. Actuate the light field and close shutters to minimum size. If the light is completely blocked
from the image receptor, no further test is required.

5. Measure the light field size on the cassette of the extended bucky tray.

D. VERIFICATION OF COMPLIANCE

Measured field size must be less than or equal to 5 by 5 centimeters.

E. PROCEDURE 2 - SPOTFILM/FLUOROSCOPIC

1. Set SID for maximum obtainable and lock in place.

2. Position beam attenuator to intercept entire X-ray beam.

3. Set appropriate technique factors (both spotfilm and fluoroscopic).

4. Close shutters to smallest size obtainable and make an exposure.

November 2005 ( 6-14 ) Linear IV Collimator


5. Verify on viewing device that there is no visual indication of radiation. If
any radiation field is discernible, Method I must be utilized to determine
minimum field size.

FVERIFICATION OF COMPLIANCE

Verify that no visual indication of the radiation field is discernible.

XR 8-2.17 X-RAY FIELD/RECEPTOR CENTER ALIGNMENT


REQUIREMENT - Means shall be provided to align the center of the X-ray field with
respect to the image receptor to within 2 percent of the SID (See 21 CFR 1020.31 (e)
(1) ).
A. GENERAL

1. All exposures taken during this test must have a uniform film density of approximately 1.0.

2. Actual versus indicated SID must be determined prior to performing this test.

B. EQUIPMENT

Radiographic cassette loaded with film (8 by 10 inches).

C. PROCEDURE

1. Load cassette with film and place into the bucky tray.

2. Assure the X-ray beam is perpendicular to the image receptor and centered over the bucky
tray.

3. Set the SID to the value determined in the actual versus indicated SID test.

4. Reduce the X-ray field to approximately 6 by 8 inches.

5. Make an exposure and develop the film.

To determine as accurately as possible the corners of the image


recorded on the film, locate two points on each of the four sides of the
image. Through the two points on each side draw a straight line. These
four lines, when extended, intersect making a rectangle which is a close
approximation of the actual X-ray field. Draw a diagonal across the
image to determine the center of the X-ray image.

November 2005 ( 6-15 ) Linear IV Collimator


7. To determine the center of the X-ray film draw diagonals across the film
(the point where these two lines cross is the center of the film), or fold the
film into quarters (the point where the two folds cross is the center of the
film).

8. The distance from the film center mark to the image center mark is
measured and recorded as the linear displacement or misalignment of
the centers of the X-ray field and the image receptor.

D. VERIFICATION OF COMPLIANCE

Verify that this distance is less than or equal to 2 percent of the SID.
NEMA Standards 5-15-79

XR 8-2.18 INDICATION OF X-RAY FIELD SIZE


REQUIREMENT - Means shall be provided on the beam-limiting device to indicate
field size in the image receptor plane to within 2 percent of the SID (see 21 CFR
1020.31 (e) (1) ).

A. GENERAL

The actual versus indicated SID test must be performed prior to beginning this
test.

B. EQUIPMENT

A 24 by 30 centimeter or a 10 by 12 inch cassette with film.

C. PROCEDURE

1. Set the SID to the value determined in the actual versus indicated SID
test.

2. Center the film cassette in the cassette tray and insert into position.

3. Adjust the field size to 15 by 15 centimeters or 8 by 8 inches by means of


the numerical indicators on the beam-limiting device.

4. Make an exposure and develop film.

5. Measure and record the length and width dimensions of the image.

D. VERIFICATION OF COMPLIANCE

November 2005 ( 6-16 ) Linear IV Collimator


The deviation of any of the recorded dimensions must not exceed 2 percent of
the SID in Step 1.

NEMA Standards 5-15-79

XR 8-2.19 POSITIVE BEAM LIMITATION (PBL)


REQUIREMENT

Means shall be provided for positive-beam limitation (PBL) which will, at the SID for
which the device is designed, either cause automatic adjustment of the X-ray field in the
plane of the image receptor to the image receptor size within five seconds after insertion
of the image receptor or, if adjustment is accomplished automatically in a time interval
greater than 5 seconds or is manual, will prevent production of X-rays until such
adjustment is completed. At SID’s at which the device is not intended to operate, the (e)
(2) ).
A. GENERAL

The PBL requirement must be met if both the beam axis and table angulation are
within plus or minus 10 degrees of the horizontal or vertical and the film is used
in the cassette tray.

B. EQUIPMENT

Large size cassette.

C. PROCEDURE

1. Set the source assembly to a SID where the PBL system is intended to
operate.

2. Place the largest film cassette with which the system is intended to
operate into the cassette tray; do not insert cassette tray at this time.

3. Turn on the light localizer and adjust the beam-limiting device to the
smallest obtainable field size.

4. Insert the cassette tray and measure the time elapsed from the insertion
of the cassette tray with the cassette inserted to the adjustment of the
X-ray field to the image receptor size.

5. The adjustment must be accomplished within 5 seconds.

November 2005 ( 6-17 ) Linear IV Collimator


6. If the adjustment is not accomplished within 5 seconds or the beam-
limiting device is of the manual type, select low-range values of the tube
potential and tube current and attempt to make an exposure. The
production of X-rays must be prevented until the PBL adjustment is
completed.

7. Move the source assembly to a SID where the PBL system is not intended to operate (see
manufacturer’s specifications) and attempt to make an exposure. Exposures must not be
possible.
NEMA Standards 5-15-79

XR 8-2.20 X-RAY FIELD LIMITATION AND ALIGNMENT


REQUIREMENT

The X-ray field size in the plane of the image receptor, whether automatically or
manually adjusted, shall be such that neither the length nor the width of the X-ray field
differs from that of the image receptor by greater than 3 percent of the SID and that the
sum of the length and width differences without regard to sign be no greater than 4
percent of the SID, when the equipment indicates that the beam axis is perpendicular to
the plane of the image receptor (see 21 CFR 1020.31 (e) (2) (ii) ).

.01 METHOD 1 - FDA/CDRH TEST STAND METHOD

A. EQUIPMENT

1. FDA/CDRH compliance test stand with accessories

2. Slide assembly.

3. Plastic cassette containing a sheet of direct-print paper or X-ray film.

4. Ruler.

5. Cassette (preferably 8 to 10 inches or smaller).

B. PROCEDURE

1. Using the means provided, align the source assembly so that the beam
axis is perpendicular to the image receptor.

2. Place the test stand on the table.

3. Position the spacer so that it does not intersect the primary beam and
secure with the pushbutton connectors.

November 2005 ( 6-18 ) Linear IV Collimator


4. Center the source assembly over the test stand using the means
provided, e.g. the light-field used to define the X-ray field.

5. Bring the source assembly down into firm contact with the spacer.

6. Center the cassette tray with the source assembly using the means
provided, e.g. bucky light.

7. Insert the plastic cassette into the slide assembly. Then insert the slide
assembly into slot 5 (see Figure 6-2).

8. Center the film cassette in the cassette tray and insert into position. If the
positive-beam limitation will not operate at this SID, raise the source
assembly and lock in position at the first operable SID.

9. Make an exposure. Develop the image. Measure and record the length
and width dimensions of the image.

10. Calculate the field size correction factor as the SID/A where:
a. SID is the indicated source-to-image receptor distance, and

b. A is the indicated source-to-tabletop distance less 7.7 inches.


Multiply each of the measured dimensions by the correction
factor.

X-ray field length at


under-table image receptor = SID x (X-Ray field length at slot 5)
A

X-ray field width at


under-table image receptor = SID x (X-Ray field width at slot 5)
A
Determine the difference without regard to sign between the
corrected length and width dimensions and the corresponding
cassette film size dimensions (8 by 10, 5 by 7, etc.). Each of
these differences must be less than 3 percent of the SID, and the
sum of these differences must be less than 4 percent of the SID.

NEMA Standards 5-15-79

November 2005 ( 6-19 ) Linear IV Collimator


.02 METHOD II - ALTERNATE TEST STAND METHOD

A. GENERAL

Prior to performing this test, the magnification factor must be determined in


accordance with the X-ray/light field alignment test - Method III.

B. EQUIPMENT

1. Manufacturer’s recommended test stand.

2. Cassette with film.

C. PROCEDURE

1. Align the tube unit and image receptor and set SID to the value
determined in the actual versus indicated SID test.

2. Insert empty 8 by 10 inch cassette into bucky tray.


3. Position test stand in accordance with manufacturer’s instructions.

4. Load a second cassette and place in the designated position.

5. Select the proper technique factors, make an exposure, and develop film.

6. Measure the length and width of the X-ray image on the film.

7. Multiply each measurement by the magnification factor previously


determined.

D. VERIFICATION OF COMPLIANCE

Verify that the X-ray field size in the plane of the image receptor does not differ
from that of the image receptor by greater than 3 percent of the SID and that the
sum of the length and width differences without regard to sign is not greater than
4 percent of the SID.

NEMA Standards 5-15-79

.03 METHOD III - CASSETTE METHOD

A. GENERAL

This procedure can be used only when a capability is provided for overriding
positive-beam limitation.

November 2005 ( 6-20 ) Linear IV Collimator


B. EQUIPMENT

1. Large cassette with film.

2. Small cassette, empty.

C. PROCEDURE

1. Insert empty smaller cassette into bucky tray.

2. Switch system to the override mode.

3. Remove the smaller cassette and insert the loaded large cassette.

4. Select the proper technique factors, make an exposure, and develop film.

5. Measure the length and width of the X-ray image on the film.

D. VERIFICATION OF COMPLIANCE
Verify that the X-ray field size in the plane of the image receptor does not differ
from that of the image receptor (smaller cassette) by greater than 3 percent of
the SID and that the sum of the length and width differences without regard to
sign is not greater than 4 percent of the SID.

NEMA Standards 5-15-79

XR 8-2.21 RETURN TO POSITIVE-BEAM LIMITATION (PBL) WITH


IMAGE RECEPTOR CHANGE

REQUIREMENT

Return to positive-beam limitation (PBL) shall occur with a change in image receptor
size (see 21 CFR 1020.31 (e) (2) (iii) ).

A. EQUIPMENT

Medium size cassette.

B. PROCEDURE

1. Select positive-beam limitation mode.

November 2005 ( 6-21 ) Linear IV Collimator


2. Insert the medium size cassette into the bucky tray and record the field
size indication.

3. Collimate down to a field size smaller than the cassette.

4. Remove and reinsert the cassette.

C. VERIFICATION OF COMPLIANCE

Verify that the system has returned to positive-beam limitation. Record the field
size indication and verify that it is equal to the previously recorded field size.

NEMA Standards 5-15-79

XR 8-2.09 BEAM QUALITY (HALF-VALUE LAYER (HVL))


REQUIREMENT

The minimum beam quality requirements listed in Table 6-1 shall be met (see 21 CFR
1020.30 (m) ).
.01 METHOD I - VISUAL DETERMINATION OF HALF-VALUE LAYER (HVL)

A. GENERAL

The above HVL requirement will be considered to have been met if it can be
demonstrated that the aluminum equivalent of the total filtration in the primary
beam is not less than that shown in Table 6-2.

B. EQUIPMENT

None required

November 2005 ( 6-22) Linear IV Collimator


Table 6-1 MINIMUM BEAM QUALITY REQUIREMENTS

Kvp Range Measured kVp HVL (mmAl*)

Below 50.................. 30 0.3


40 0.4
49 0.5
50 to 70.................... 50 1.2
60 1.3
70 1.5
Above 70.................. 71 2.1
80 2.3
90 2.5
100 2.7
110 3.0
120 3.2
130 3.5
140 3.8
150 4.1
* Type 1100 aluminum alloy as given in Aluminum Association
Publication No. ASD-1, Aluminum Standards and Data
TABLE 6-2 ALUMINUM EQUIVALENT OF PRIMARY BEAM TOTAL FILTRATION

Operating Voltage Total Filtration


(kVp) (mm Al Equivalent)

Below 50.................. 0.5

50 - 70 1.5

Above 70 2.5

C. PROCEDURE

Visually inspect the system and determine the aluminum equivalence of the total
filtration in the primary beam. This includes the inherent filtration of the X-Ray
tube, X-Ray tube housing, beam-limiting device, and any additional filtration that
may have been added in the useful beam (in fluoroscopic systems the tabletop
system is included as part of the added filtration).

November 2005 ( 6-23 ) Linear IV Collimator


D. VERIFICATION OF COMPLIANCE

The aluminum equivalence of the total filtration must be equal to or greater than
the amount specified in Table 6-2.

NEMA Standards 5-15-79

.02 METHOD II - STANDARD ABSORBER METHOD

A. GENERAL

This test is to be used when the surveyor cannot remove or see the total filtration
equivalence.

The HVL determinations obtained from the following procedures are to be


compared with those illustrated in Table 6-1. The HVL in millimeters of
aluminum of the system being tested must be greater than or equal to the values
shown in Table 6-1.

B. EQUIPMENT
1. Radiation detector

2. Standard absorber with equivalent filtration of 2.5 millimeters of


aluminum.

C. PROCEDURE

1. With the detection device positioned horizontally, an exposure is made at


a preselected technique factor of 80 kV and appropriate mA and time.
The reading of the radiation output is recorded.

2. Position a total of 2.5 ,millimeters of aluminum at the port of the beam-


limiting device and repeat the exposure using the same technique
factors. Record the radiation output.

For X-ray units operating at low kVp (less than 50) and for
mammography units, it will be necessary to use an aluminum absorber of
0.6 millimeters at 49 kVp.

D. VERIFICATION OF COMPLIANCE

Verify that the radiation output in Step 2 is greater than or equal to 50 percent of
the radiation output in Step 1.

NEMA Standards 5-15-79

November 2005 ( 6-24 ) Linear IV Collimator


.03 METHOD III - FDA/CDRH COMPLIANCE TEST

A. GENERAL

The HVL determinations obtained from the following procedures are to be


compared with those illustrated in Table 6-1. The HVL in millimeters of
aluminum of the system being tested must be greater than or equal to the values
shown in Table 6-1.

B. EQUIPMENT

1. BRH/FDA compliance test stand with accessories (Figure 6.1).

2. Survey meter adapted for use with stand and ion chamber.

3. Several sheets of aluminum, each having a thickness of 0.5 or 1.0 millimeter.

C. PROCEDURE

1. Attach the spacer, positioned out of the primary beam, to the test stand. Center the stand on
the table. Center the source over the stand and bring the beam-limiting device down into firm
contact with the spacer. Select the MANUAL mode of operation (there must not be a cassette
in the cassette tray). Insert the beam-defining assembly in slot 1 of the stand with the leaded
side up (see Figure 6.2). Adjust the beam-limiting device so that the X-ray field slightly exceeds
the aperture of the beam-defining assembly. Mount the ion chamber at position B with the
chamber facing upward. Connect the chamber and meter with the cable provided. Select a
tube potential that is commonly used and is in the highest kV range of the X-ray system.

2. With no added filtration in the beam, make an exposure and record the
reading. For all diagnostic X-ray equipment, use Table 6-3 to determine
increments of filtration required to perform the half-value layer procedure.
Make an exposure and record the reading for each total thickness.

November 2005 ( 6-25 ) Linear IV Collimator


TABLE 6-3 HIGHEST DESIGN OPERATING RANGE

TOTAL ADDED FILTRATION (mm Al)


Below 50 kVp 50-70 kV Above 70 kVp

0.5 1.0 1.5

1.0 1.5 2.5

1.5 2.5 3.5

2.0 3.5 4.5

The recorded data is plotted on semi-log graph paper (Examples A and B, Fig.
6.6) and the half-value layer is read directly from the graph.

D. VERIFICATION OF COMPLIANCE

Verify that the half-value layer of the useful beam for a given X-Ray tube
potential is not less than the values shown in Table 6-1.
November 2005 ( 6-26 ) Linear IV Collimator
TABLE 6-4 HALF-VALUE LAYERS AS A FUNCTION OF FILTRATION AND TUBE
POTENTIAL FOR DIAGNOSTIC UNITS*

PEAK POTENTIAL (kV)


Total
Filtration 30 40 50 60 70 80 90 100 110 120
mm Al Typical Half-Value Layers (mm Al)

0.5 0.36 0.47 0.58 0.67 0.76 0.84 0.92 1.00 1.08 1.16
1.0 0.55 0.78 0.95 1.08 1.21 1.33 1.46 1.58 1.70 1.82
1.5 0.78 1.04 1.25 1.42 1.59 1.75 1.90 2.08 2.25 2.42
2.0 0.92 1.22 1.49 1.70 1.90 2.10 2.28 2.48 2.70 2.90
2.5 1.02 1.38 1.69 1.95 2.16 2.37 
2.58 
2.82 
3.06 
3.30 

3.0 .... 1.49 1.87 2.16 2.40 2.62 2.86 3.12 3.38 3.65
3.5 .... 1.58 2.00 2.34 2.60 2.86 3.12 3.40 3.68 3.95

* For full-wave rectified potential

Recommended minimum HVL for radiographic units

 Recommended minimum HVL for fluoroscopes


EXAMPLE A EXAMPLE B

Figure 6.6

November 2005 ( 6-27 ) Linear IV Collimator

XR 8-2.24 FLUOROSCOPIC X-RAY FIELD LIMITATION

REQUIREMENT

A. NONIMAGE-INTENSIFIED FLUOROSCOPY

The X-Ray field by non image-intensified fluoroscopic equipment shall not


extend beyond the entire visible area of the image receptor. This requirement
applies to field size during fluoroscopic procedures.

B. IMAGE-INTENSIFIED FLUOROSCOPY

1. During fluoroscopic procedures, neither the length nor the width of the X-
Ray field in the plane of the image receptor shall exceed the visible area
of the image receptor by more than 3 percent of the source-to-image
distance (SID). The sum of the excess length and the excess width shall
be no greater than 4 percent of the SID.

2. Compliance shall be determined with the beam axis perpendicular to the


image receptor. For rectangular X-Ray fields used with circular image
receptors, the error in alignment shall be determined along the length and
width dimensions of the X-Ray field which pass through the center of the
visible area of the image receptor. (See 21CFR 1020.32 (b).)
.01 METHOD I - FDA/CDRH TEST STAND METHOD

A. EQUIPMENT

1. FDA/CDRH compliance test stand and accessories.

2. Brass or copper attenuation block.

3. Plastic cassette and direct-print paper or redipack film cassette.

B. PROCEDURE

1. Position all movable grids and compression cones out of the path of the
primary beam. If necessary, level the table and retract the spotfilm
carriage. If the system utilizes a television monitor, turn it on and allow
time for warm up.

2. Place the test stand in the approximate center of the table.

3. Insert the slide assembly, grid side down, into slot 5 (See Fig. 6.2.)

4. Place attenuator block on top of slide assembly.

November 2005 ( 6-28 ) Linear IV Collimator


5. Fully open the beam-limiting device.

6. Center the image receptor over the test stand by observing the image of
the grid on the image receptor.

7. Bring the imaging assembly down into firm contact with the top of the
stand.

8. If using a dual-field image intensifier, e.g., 6-and 9-inch capability, select


the mode of largest image magnification, e.g., 6-inch mode in this case.

9. Adjust the controls to obtain the best quality image.

10. Insert a plastic cassette containing a sheet of direct-print paper or


redipack film cassette into the slide assembly and make an exposure.

11. At the image output, count the units from the center of the grid to each
edge of the image as shown by lines A, B, C and D of Fig. 6.7.

Note that line A passes between grid numbers 1 and 4, B passes


between numbers 1 and 2, etc. Record the measurements in order.

12. Develop the film and outline the perimeter of the image.
13. Count the units from the center of the grid to each edge as you did with
A, B, C, and D above. Record these as A’, B’, C’ and D’.

2 1

C A

3 4

Figure 6.7
FLUOROSCOPIC X-RAY FIELD LIMITATION TEST STAND METHOD

November 2005 ( 6-29 ) Linear IV Collimator


C. VERIFICATION OF COMPLIANCE

For compliance of non image-intensified systems, A shall be less than or equal


to A’, B shall be less than or equal to B’, C shall be less than or equal to C’, and
D shall be less than or equal to D’.

For image-intensified systems, determine the source-to-slot-5 distance by adding


18.5 centimeters (7.3 in) to the minimum source-to-skin distance (SSD)
determined by the SSD test procedure using the FDA/CDRH test stand.
Calculate and record 3 percent and 4 percent of this distance. Determine the
difference, without regard to sign of A and A’, B and B’, C and C’, and D and D’.
Sum the differences. They must be less than or equal to 4 percent of the
source-to-slot-5 distance. In addition, the sum of A and A’ plus C and C’
differences and the sum of the B and B’ plus the D and D’ differences must each
be less than or equal to 3 percent of the source-to-slot-5 distance.

.02 METHOD II - DIRECT METHOD

A. EQUIPMENT

1. Lead foil phantom or copper attenuator.

2. Direct-print paper.
3. One by one-inch steel marker.

4. Four additional metal markers (each marker is approximately 1/32-inch


galvanized sheet metal having the dimensions of 1.5 by 0.5 inches).

B. PROCEDURE - NON IMAGE-INTENSIFIED FLUOROSCOPE

1. Place the image receptor as close as possible to the tabletop.

2. Open the shutters to the maximum and activate the fluoroscopic tube.

3. Slowly increase the distance between the tabletop and the image
receptor. The unit is in compliance if, at the greatest extension of the
image receptor, there is an unilluminated border all round the visible area
of the image receptor. No further tests need to be made.

4. If the illuminated area expands to the edges of the visible area of the
image receptor, lower the image receptor until there is a small
unilluminated border.

5. Close the shutters to produce approximately a 2-inch wide unilluminated


border. Resume increasing image-receptor-to-tabletop distance. The
unit is not in compliance if the illuminated area continues to expand.

November 2005 ( 6-30 ) Linear IV Collimator


6. If the illuminated area does not expand, then at the maximum SID, slowly
open the shutters to the maximum. The unit is in compliance if the
illuminated area does not expand.

C. PROCEDURE - IMAGE INTENSIFIED FLUOROSCOPE


(Over-the-table Tube, Variable Collimator)

This procedure can be used only if the distance from the input phosphor to the
tabletop does not change with SID and if the SID is variable.

1. Raise the tube housing to its greatest SID and, if the beam-limiting device
is not automatic, close the shutters to their fullest extent.

2. Activate the tube and slowly increase the field size while observing the
monitor. If at the maximum field size it is obvious that the image of the
beam-limiting device is apparent at the edge of the image receptor, no
further test need be made. (See Fig. 6.8.)

SHUTTERS
Figure 6.8

November 2005 ( 6-31 ) Linear IV Collimator


3. If there is no unilluminated area at any of the edges of the image
receptor, lower the tube head until the bottom of the beam-limiting device
is at its minimal distance from the tabletop (or 12 inches, whichever is
greater). Record this distance.

4. Place a lead foil phantom or copper attenuator on the tabletop and center
a loaded cassette in the light-field (Fig. 6.9)

5. Adjust the light-field to give a 5-by 7 inch field on the cassette and mark
the edges with metal markers.

6. Attach a 1-by 1-inch steel marker to the center of the bottom of the
beam-limiting device. (Exact centering is not necessary.)

7. Make an appropriate exposure.

8. Develop the direct-print paper or film and check light and X-ray beam
alignment.

9. Measure length of marker image.

10. By similar triangles, determine the source-to-tabletop distance.


11. This procedure can continue only if the unit is aligned to radiographic
requirements (plus or minus 2 percent SID). Raise the tube head to
maximum SID.

12. Mark the edges of the visible area by placing two radiopaque markers
within the light-field and parallel to each other and to the edges of the
light-field.

13. While viewing the markers through the image intensifier, activate the tube
on and off, adjusting the parallel markers between exposures until the
outer edges of the markers are no longer visible (Fig. 6.10)

14. Again, turn on the light-field and measure the distance from the edges of
the light-field to the parallel markers.

15. To determine the maximum source-to-tabletop distance, subtract the


beam-limiting-device-to-tabletop distance previously determined from the
source-to-tabletop distance as previously calculated and add this result to
the measured distance from the bottom of the beam-limiting device to the
tabletop. Calculate the percent of the SID of the X-ray field that is in
excess of the visible area.

16. Place the markers at right angles to the original position and repeat the
last three steps.
NEMA Standard 5-15-1979

November 2005 ( 6-32) Linear IV Collimator


D. PROCEDURE - IMAGE INTENSIFIED FLUOROSCOPE (Under-the-table Tube,
Variable Collimator)

NOTE: This procedure is not included since it does not apply to the Linear IV
Collimator System.

E. VERIFICATION OF COMPLIANCE

For non image-intensified fluoroscopy, verify that the X-Ray field does not extend
beyond the visible area of the image receptor.

For image-intensified fluoroscopy, verify that neither the length or width


misalignment exceeds the visible area of the image receptor by more than 3
percent of the SID. The sum of the excess length and the excess width shall be
no greater than 4 percent of the SID.

NEMA Standard 5-15-1979

OVER-THE-TABLE TUBE
SOURCE

BLD
RADIOPAQUE
1” X 1” STEEL MARKER MARKERS
BEAM DPP CASSETTE
ATTENUATOR

TABLETOP

IMAGE RECEPTOR ASSEMBLY

Figure 6.9

November 2005 ( 6-33 ) Linear IV Collimator

VISIBLE AREA

MARKERS

Figure 6.10

XR 8-2.25 FLUOROSCOPIC SOURCE-TO-SKIN DISTANCE (SSD)

REQUIREMENT - The source-to-skin distance (SSD) shall be not less than 38


centimeters (15 in) on stationary fluoroscopes. (See 21 CFR 1020.32 (f).)
A. GENERAL

If the SSD can be measured directly, no further testing is necessary.

B. EQUIPMENT

1. FDA/CDRH compliance test stand with accessories (Fig. 6.1).

2. Plastic cassette and direct-print paper or redipack film cassette.

3. Brass or copper attenuation block.

C. PROCEDURE

1. Position all movable grids and compression cones out of the primary
beam. If necessary, return the table to the horizontal position and raise
the spotfilm carriage. If the system utilizes a television monitor, turn it on
and allow time for warm up.

2. Place the test stand in the approximate center of the table.

3. Insert the slide assembly, grid side down, at slot 3 (See Fig. 6.2.).

4. Place attenuation block on the top of slide assembly.

November 2005 ( 6-34 ) Linear IV Collimator


5. Fully open the beam-limiting device.

6. Center the image receptor over the test stand by observing the image of
the grid on the image receptor.

7. Bring the image assembly down into firm contact with the spacer bar on
the top of the stand.

8. Slide the focal spot assembly, brass side up, into slot 7.

9. Insert the plastic cassette or the redipack film cassette into the slide
assembly.

10. Make an exposure assuring proper density on the film or paper.

11. Remove the cassette and develop the direct-print paper or film.

12. Measure the minimum separation of the outside edges of the focal spot
strips (see Fig. 6.13).

D. CALCULATIONS

Calculate the minimum SSD.

Minimum SSD (in inches)=


_____28.25 in2_______ -0.4 in
(Image size in inches - 2.5)

Minimum SSD (in centimeters) =

____182.3 cm2______ -1.0 cm


(Image size in cm - 6.4)

E. VERIFICATION OF COMPLIANCE

Verify that the SSD is not less than 38 centimeters (15 in).

NEMA Standard 5-15-1979

FILM IMAGE OF OUTSIDE


SEPARATION OF
6.4 CM (2.5”) BRASS STRIPS
November 2005 ( 6-35 ) Linear IV Collimator

FILM

28.7 CM
11.3”

FOCAL SPOT ASSEMBLY


X-RAY BEAM

FOCAL SPOT

EQUATION:
X = 182.3 cm2
(IMAGE SIZE IN CENTIMETERS – 6.4 cm

EQUATION:
X = 28.25”
(IMAGE SIZE IN INCHES – 2.5”)

MINIMUM SSD IN (CENTIMETERS) = X – 1.0 cm


MINIMUM SSD (IN INCHES) = X – 0.4”

Figure 6.11 SPOT FILM GEOMETRY

November 2005 ( 6-36 ) Linear IV Collimator


XR 8-2.27 PRIMARY BARRIER - X-RAY LOCKOUT

REQUIREMENT - For equipment manufactured after February 25, 1978, when the
angle between the image receptor and the beam axis is variable, means shall be
provided to indicate when the axis of the X-ray beam is perpendicular to the plane of the
image receptor.

The fluoroscopic tube shall not produce X-rays unless the barrier is in position to
intercept the entire useful beam. (See 21 CFR 1020, 32 (a) and 1020.32 (b) (2) (iii).)

A. VERIFICATION OF COMPLIANCE

Move the primary barrier off the detents position and verify that X-rays cannot be
produced by the fluoroscopic X-ray tube. Verify that, if the beam angulation is
variable, a means is provided to indicate that the beam axis is perpendicular to
the plane of the image receptor.
NEMA Standards 5-15-1979

SPOTFILM RADIOGRAPHIC TESTS

XR 8-2.29 ALIGNMENT OF X-RAY FIELD EDGES TO SELECTED


PORTIONS OF THE IMAGE RECEPTOR
REQUIREMENT - The misalignment of the X-ray field with that portion of the image
receptor selected on the spotfilm device, along either the length or the width dimension,
shall not exceed 3 percent of the source-to-image-receptor distance (SID), and the total
misalignment without regard to sign along both dimensions shall not exceed 4 percent of
the SID. (See 21 CFR 1020.31 (g).)

A. GENERAL

Technique factors must be selected for a film density of approximately 1.0.

B. EQUIPMENT

1. Manufacturer’s recommended test stand.

2. Graduated template.

3. Two cassettes.

4. Film.

C. PROCEDURE

November 2005 ( 6-37 ) Linear IV Collimator


1. Verify by the means provided that the axis of the X-ray beam is
perpendicular to the plane of the image receptor (applicable only where
variable beam angulation is provided.)

2. Load the cassette with film and place in the spotfilm device.

3. Position test stand in accordance with manufacturer’s instructions.

4. Load a second cassette with film and place in the designated position on
the test stand. This film must be large enough to accept a one-on-one
format.

5. Place graduated template in a position between the X-ray source and the
two cassettes.

6. Select proper techniques factors and a one-on-one spotfilm format.

7. Make an exposure and develop films.

8. Repeat steps 1 through 7 for each available spotfilm format.

D. CALCULATIONS

1. Calculate the magnification factor by measuring the distance between the


same two points on the graduated template image on each of the two
films. The two points chosen must be as far apart as possible. Divide
the smaller measurement into the larger measurement to determine the
magnification factor.

2. If the source-to-test stand film distance is known, calculate the actual SID
by multiplying the magnification factor by the known source-to-test film
distance.

3. If the source-to-test stand film distance is not known, measure the actual
distance from the graduated template position to the test stand film plane
(distance Z, see Fig. 6.14).

4. Calculate the source-to-template distance (distance X) using the following


formula:

A = X or X = AZ
B X+Z B-A

where:

A = two points on a graduated template.


B = magnification of the same two points on
the graduated template used in A.

November 2005 ( 6-38 ) Linear IV Collimator


5. Calculate the source-to-test stand film distance by adding distance X and
distance Z. Multiply the source-to-test stand film distance by the
magnification factor determined in step 1 to determine the actual SID.

6. On the test stand film measure and record distances a, b, c, and d from
the center of the graduated template to each respective edge of the
radiographic image (see Fig. 6.15).

7. Multiply each distance a through d by the magnification factor and record


the products as a’, b’, c’, and d’. Note the a’, b’, c’, and d’ are the
respective distances they would have been if they had been recorded at
the spotfilm plane.

8. On the spotfilm image measure distances A, B, C, and D from the center


of the graduated template image to each respective edge of the selected
portion of the spotfilm (see Fig. 6.16). Record distances A, B, C, and D
as measured.

9. Calculate the length differences and width differences without regard to


sign as follows:

Length difference = B -b’ + D - d’

Width difference = A-a’ + C - c’

Record the sum of the length and width differences.


E. VERIFICATION OF COMPLIANCE

1. Calculate 3 and 4 percent of the actual SID and record.

The individual length and width misalignment, regardless of sign, must be


less than or equal to 3 percent of the SID as calculated in step 1. The
summation of the length and width misalignments, regardless of sign,
must be less than or equal to 4 percent of the SID as calculated in step 1.

NEMA Standard 5-15-1979

XR 8-2.30 ALIGNMENT OF CENTER OF X-RAY FIELD TO CENTER OF


SELECTED PORTION OF IMAGE RECEPTOR

REQUIREMENT - The center of the X-ray field in the plane of the film shall be aligned
with the center of the selected portion of the film to within 2 percent of the source-to-
image distance (SID). (See 21 CFR 1020.31 (g) (4).)

A. GENERAL

November 2005 ( 6-39 ) Linear IV Collimator


1. All exposures taken during this test should have a uniform film density of
approximately 1.0.

2. The actual SID must be determined prior to performing this test.

B. EQUIPMENT

Radiographic cassette and film.

C. PROCEDURE

1. Load cassette with film and place in the spotfilm device.

2. Select a one-on-one spotfilm format.

3. Assure that the SID is set to the value previously determined in the actual
SID determination test.

4. Reduce the X-ray field size to a size slightly smaller than the size of the
film area to be exposed.

5. Make an exposure and develop film.

6. Repeat steps 1 through 5 for each available spotfilm format.

7. To determine the centers of the X-ray films draw diagonals across the film
(the point where these two lines cross is the center of the film).
8. To determine the corners of the image recorded on the film , locate two
points on each side of the four sides of the image. Through the two
points on each side draw a straight line. These four lines when extended
intersect making a rectangle which is a close approximation of the actual
X-ray field. Draw diagonals across the rectangular figure determined.
The point at which the two diagonals cross is assumed to be the center of
the X-ray field.

9. The distance from the film center mark is measured and recorded as the
linear displacement or misalignment of the centers of the image receptor
and the X-ray field.

D. VERIFICATION OF COMPLIANCE

Verify that this distance is less than or equal to 2 percent of the SID.

NEMA Standard 5-15-1979

FOCAL SPOT

NovemberGRADUATED
2005 TEMPLATE ( 6-40) Linear IV Collimator
POSITION

X
FOCAL SPOT ASSEMBLY
IN SLOT 1

TEST STAND

TEST STAND PLANE

TABLETOP B

Y
IMAGE RECEPTOR
FILM PLANE

UNDERTABLE
CASSETTE
C

Figure 6.12
DETERMINATION OF SID

November 2005 B( 6-41 ) Linear IV Collimator

2 1

C A

3 4

Figure 6.13
B

2 1

C A

3 4
D

Figure 6.14

November 2005 ( 6-42 ) Linear IV Collimator

NO TEXT
November 2005 ( 6-43 ) Linear IV Collimator
RECORD SHEET

This sheet is to be used by the assembler to assure that all points of compliance are covered.
It will also serve as a maintenance log.
HOSPITAL ______________________________________ ROOM # ___________________
DATE OF INSTALLATION _________________________ ASSEMBLER _______________
_______________________________________________ _______________

Applicable Installation
Requirement Paragraph Date Date Date Date Date

1. Visual definition of XR8/2.14


X-Ray light field

2. Intensity of light-field XR8/2.15

3. Min. field size XR8/2.16

4. X-Ray field/receptor XR8/2.17


center alignment
5. Indication of field size XR8/2.18

6. Positive beam XR8/2.19


limitation

7. X-Ray field limitation XR8/2.20


& alignment

8. Return to PBL XR8/2.21

9. Half value layer XR8/2.09

10. Fluoroscopic X-Ray XR8/2.24


field limitation

11. Fluoroscopic Source XR8/2.25


to Skin Distance (SSD)

12. Primary barrier - XR8/2.27


X-Ray lockout

November 2005 ( 6-44 ) Linear IV Collimator

Applicable Installation
Requirement Paragraph Date Date Date Date Date

13. Alignment of X-Ray XR8/2.29


edges to selected
portions of the
image receptor.

14. Alignment of center XR8/2.30


of X-Ray field to
center of selected
portion of image
receptor

15. Remote control


inspection/cleaning

16. Collimator mounting Section 2


inspection 2.3

17. Cassette tray


inspection/cleaning
18. Electrical cable
inspection

19. SID Monitor Sections


Inspection 7.2 & 7.3

INITIALS:

NOTES

November 2005 ( 6-45 ) Linear IV Collimator

NO TEXT
November 2005 ( 6-46 ) Linear IV Collimator

SECTION 7
CONTINUOUS SID
ADJUSTMENT

November 2005 ( 7-1 ) Linear IV Collimator


7.0 CONTINUOUS SID ALIGNMENT PROCEDURE

7.1 GENERAL

The purpose of the continuous SID monitor is to provide an output voltage (VSID)
proportional to the SID. The Eureka SID monitor provides a nominal 14.4 OHMS per
inch resistance change. If the SID monitor is attached to follow the X-ray tube SID
excursion directly, this would then be a 1:1 ratio system. Most integrated tubestand/flat
bucky table systems are 1:1 ratio systems. However, if the monitor is attached to an
overtable tube carriage, it will probably be attached to a telescopic member that moves
at a ratio of 2:1, 3:1, or 4:1 with respect to the X-ray tube movement. Hence, most
ceiling suspended tube carrier systems are either 3:1 or 4:1 ratio systems.

The SID interface to the Linear IV has been greatly simplified to accommodate a range
of different SID monitors and/or different SID ratios. The Linear IV during calibration will
“learn” the rate of tube excursion for the SID monitor attached, thus making it
unnecessary to set switches to indicate ratios and making it easier to modify SID circuits
to accommodate ratios which are slightly different than their whole number values.

The SID circuit provides a source current of 5mA to the SID monitor. The Eureka SID
monitor is a 1 kOhm pot that will thus produce a maximum of 5V, and a minimum of 0V.
Any SID monitor can be used as long as the voltage range is linear with respect to SID
and does not exceed 5V.

7.2 SID MONITOR INSTALLATION GUIDELINES

CAUTION: To prevent damage to the mechanism which would require


replacement of the SID monitor, observe the following guidelines:

Install the SID monitor so that the measuring cable, cable clamp, and SID monitor
mechanism will remain unobstructed in all conditions of use.

If the Overhead Table System (OTS) allows the SID to be varied by more than 66
inches (170 cm), positive mechanical stops must be used to limit the range to prevent
SID cable breakage and/or mechanical damage.

Do not arbitrarily pull the measuring cable during handling and installation or allow the
measuring cable to recoil unrestricted into the monitor.

Do not clamp, kink or knot the measuring cable, which would distort the cable within its
working range.

November 2005 ( 7-2 ) Linear IV Collimator


Route the SID monitor interconnection cable directly to the Linear IV power supply unit.
Verify that the cable is free of kinks or knots and that it will not come in contact with any
other moving parts of the system. Provide adequate strain relief where required.

OPTION - For adjustable height table systems using the differential SID option, follow
the above guidelines for the second SID monitor installed in the adjustable height table.

7.3 SID MONITOR POTENTIOMETER ADJUSTMENT

With power to the Linear IV collimator system off, disconnect the red and black wires of
the SID monitor interconnection cable from terminal strip locations TS4-2 and TS4-3 in
the Linear IV power supply enclosure.

Place a digital ohm meter to the red and black wire. Vary the SID of the OTS and
observe that as SID decreases, the SID potentiometer resistance increases. If a
decrease in SID does not produce an increase in resistance, then remove the Buss wire
on the SID potentiometer from the red wire and connect it to the black wire.

Set the SID of the OTS to 72 inches (180 cm) or its design operating range (whichever
is less). Adjust the SID monitor cable so that the SID monitor resistance will be
approximately 40 ohms when properly attached to the cable. Typical good signal to
noise ratio setting at 40” (100 cm) SID would be 450 ohms or 2.5 VDC.
NOTE: It is not critical that the specified resistance above be set exactly, as the
Linear IV will “learn” the setting and compensate for it. It is critical that the SID
monitor move freely within its operating range and provide a voltage linearly
proportional to SID.

NOTE: If the specified resistance is not obtainable, it may be necessary to


relocate the cable clamp, attach additional cable, or readjust the SID
potentiometer by loosening the set screw attaching the flexible coupling and pot
shaft, then re-tightening the set screw when finished adjusting.

Reconnect the SID monitor interconnection cable as shown in Figure 2-8. Make sure all
cable clamps are properly tightened.

November 2005 ( 7-3 ) Linear IV Collimator


7.4 (OPTION) DIFFERENTIAL (TABLE) SID MONITOR POTENTIOMETER
VERIFICATION

NOTE: This section should only be used if the X-ray system under installation
employs an adjustable height table top and a second SID monitor is installed in
the table for differential SID sensing.

With the power to the Linear IV collimator system off, disconnect the red and black wires
of the SID monitor interconnection cable from terminal strip locations TS4-4 and TS4-5
in the Linear IV power supply enclosure.

Place a digital ohm meter to the red and black wire. Vary the SID of the Table and
observe that as SID increases, the SID potentiometer increases. If an increase in SID
does not produce an increase in resistance for the table SID monitor only, then remove
the Buss wire on the SID potentiometer from the black wire and connect it to the red
wire.

Set the table to its full up position. Adjust the adjustable table SID monitor cable so that
the SID monitor resistance will be approximately 10 to 20 ohms when properly attached
to the cable. A typical on a 1:1 ratio is 0.5 VDC when table is in full up position.
NOTE: It is not critical that the specified resistance above be set exactly, as the
Linear IV will “learn” the setting and compensate for it. It is critical that the SID
monitor move freely within its operating range and provide a voltage linearly
proportional to SID.

NOTE: If the specified resistance is not obtainable, it may be necessary to


relocate the cable clamp, attach additional cable, or readjust the SID
potentiometer by loosening the set screw attaching the flexible coupling and pot
shaft, then re-tightening the set screw when finished adjusting.

Reconnect the SID monitor interconnection cable as shown in figure 2.9. Make sure all
cable clamps are properly tightened.

7.5 SID CIRCUIT VERIFICATION

The SID circuit of the Linear IV will provide continuous SID indication from 36” to 72” (92
cm to 180 cm) with the correct (compatible) SID monitor installed. Optionally the Linear
IV can provide differential SID (adjustable height tables) within the same range of
operation with the second (table) SID monitor installed.

November 2005 ( 7-4 ) Linear IV Collimator

The SID circuit on the master board (70-08320) provides two fixed constant current
sources for each SID (tube and table) monitor installed. The current from these sources
is nominally 5mA, thus any SID monitor that provides a linear voltage within its operating
range and does not exceed 1kohm at its output within its operating range is deemed
compatible with the circuit.

The constant current sources can be verified using the following procedure.

1. With power applied to the Linear IV power supply, connect a digital DC ammeter
to TP1 (mA1+) and TP2 (ma1-) test points on the Linear IV master board
(70-08320).

2. Remove jumper W1 and verify a reading of 5mA + 0.25mA.

3. Reconnect jumper W1.

4. Connect the digital ammeter to TP3 (mA2+) and TP4 (mA2-) test points on the
Linear IV master board (70-08320).

5. Remove jumper W2 and verify a reading of 5mA + 0.25 mA.


6. Reconnect jumper W2.

NOTE: No other adjustments are necessary as the SID monitors rate and offset
are determined by software during the auto-calibration process. There are no
potentiometers to adjust.

November 2005 ( 7-5 ) Linear IV Collimator

NO TEXT
November 2005 ( 7-6 ) Linear IV Collimator

SECTION 8
OPERATING
INSTRUCTIONS

November 2005 ( 8-1 ) Linear IV Collimator


FIGURE 8.1
LINEAR IV FRONT PANEL

November 2005 ( 8-2 ) Linear IV Collimator

8.0 OPERATING INSTRUCTIONS


8.1 GENERAL INSTRUCTIONS

Operation of the Linear IV collimator is accomplished through the front panel of the
collimator. The front panel contains an alphanumeric display which provides collimator
system status and error messages. Membrane switch buttons are used to control the
field size, enter the calibration mode, store memorized field sizes, recall stored field
sizes, and toggle between image intensifier sizes in fluoroscopic operation. The front
panel also has knobs for manual control of field size as well as a button to illuminate the
field lamp. A digital LED display indicates current SID and field size. The next sections
provide details of the front panel, its messages and typical operating sequences.

8.2 FRONT PANEL DESCRIPTION

1. Alpha-numeric display for collimator mode information and instructions. Used in


run-time operation, calibration and other diagnostic modes.

2. Cross field size display indicates the current cross-table field size in inches or
centimeters at the currently selected image receptor.

3. Long field size display indicates the current long-table field size in inches or
centimeters at the currently selected image receptor.

4. SID display indicates current SID in inches or centimeters.

5. CAL button – used to initiate a calibration, and/or select the current service
mode.

6. MEM ONE button – Used to store the current size into memory, or to change the
field size to the currently stored size in MEM ONE.

To store the current size, press and hold the MEM ONE button until theindicator
(7) turns off, then release the button and size will be stored.

To move the stored size, press and release the MEM ONE button until the
indicator (7) will light and the shutters will move to the stored present size.

7. MEM ONE indicator – Illuminates if MEM ONE has been selected.

8. LAST SIZE button – Moves the shutters to the previous shutter size prior to the
current PBL cycle. No movement will occur during the current PBL cycle if the
shutters are moved manually as this becomes the new LAST SIZE for the next
PBL cycle.

9. LAST SIZE indicator – Illuminates if LAST SIZE was chosen.

November 2005 ( 8-3 ) Linear IV Collimator


10. ENT/SET button – Used in service mode to acknowledge questions and
instructions.
Changes the SID display (4) and field size displays (2) and (3) during normal
run-time operation to indicate (for approximately 8 seconds) the vertical SID to
the table top and a table top field size estimate based on calibration information.

11. MEM TWO button – Used to store the current size memory, or to change the
field size to the currently stored size in MEM TWO.

To store the current size, press and hold the MEM TWO button until the indicator
(12) turns off, then release the button and size will be stored.

To move the stored size, press and release the MEM TWO button until the
indicator (12) will light and the shutters will move to the stored preset size.

12. MEM TWO indicator – Illuminates if MEM TWO has been selected.

13. MAG SIZE button – Toggles through three preset II size selections in
fluoroscopic operation.

14. MAG SIZE indicator – Illuminates if any one of the preset II size have been
selected.

15. Cross shutter manual control knob.

16. Long shutter manual control knob.

17. Cross Open button – Power assisted drive for the cross-table shutter to increase
in size.

18. Long Open button – Power assisted drive for the long-table shutter to increase in
size.

19. Cross Close button – Power assisted drive for the cross-table shutter to
decrease in size.

20. Long Close button – Power assisted drive for the long-table shutter to decrease
in size.

21. Lamp switch button – Illuminates the internal projection lamp to activate the
timed light field. If pressed again, it will turn off the internal lamp and thus act as
a quick turn-off if the full time cycle is not required.

22. SID Tape Measure – Correct measurement made to the end of the ring.

November 2005 ( 8-4 ) Linear IV Collimator


8.3 LIST OF MESSAGES
General Messages
1. LINEAR IV SYSTEM Start up message.

2. REVISION (number) Displays current software revision indicated by


number.

Run-time Messages

3. READY-RAD Collimator is ready in radiographic mode

4. (X) READY FLUOR NORM Collimator is ready in fluoroscopic mode with


NORMal II size selected. X indicates F for Full
Film mode, P for Preset mode.

5. (X) READY-FLUOR MAG1 Collimator is ready in fluoroscopic mode with


MAG 1 II size selected. X indicates F for Full
Film mode, P for Preset mode.

6. (X) READY-FLUOR MAG2 Collimator is ready in fluoroscopic mode with


MAG 2 II size selected. X indicates F for Full
Film mode, P for Preset mode.

7. (X) RDY-SPOTFILM Collimator is ready in Spotfilm/fluoroscopic mode.


X indicates F for Full Film mode, P for Preset
mode.

8. MANUAL-TILT Collimator is in manual mode due to tile (table or


collimator) which is 10 degrees or more from
vertical or horizontal and no cassette is present
in bucky.

9. MAN-WRONG WALL Collimator is in manual when tilted horizontal


opposite to the selected wall with the film bucky.

10. MAN-NO CASS Collimator is in manual operation in radiographic


mode when no cassette is present in the
appropriate bucky.

11. MAN-SID NOT SET Collimator is in manual operation because


discrete SID operation is expected, but no
discrete SID switches have been set, with no
cassette present.

12. STEREO/TOMO Collimator is in manual operation because a


STEREO/TOMO request has been sensed.

November 2005 ( 8-5 ) Linear IV Collimator


13. PBL OVER-RIDE Collimator is in manual operation because the
PBL OVER-RIDE switch has been set.
14. EXP HLD-NOT VERT Collimator is preventing exposure in fluoroscopic
mode if the tile is not vertical.

NOTE: TILT SWITCH INPUTS CAN BE PARALLELED BY A SIGNAL INDICATING


THAT THE IMAGE RECEPTOR AND X-RAY SOURCE ARE ALIGNED AND
THUS "READY" OPERATION CAN CONTINUE AT ANY ANGLE OF TILT
AS LONG AS THE ABOVE CONDITION EXISTS.

15. EXP HLD-LOCK SID Collimator is preventing exposure because the


discrete vertical SID switch has not been set or a
cassette is present without an appropriate SID
signal.

16. EXP HLD-SID LIM Collimator is preventing exposure because of the


current detected SID is beyond the limits of 36"
or 72" (92 cm or 180 cm) with a cassette present.

17. EXPOSURE HOLD Collimator is preventing exposure in radiographic


mode because the shutters are moving to their
requested size.

18. TBL TOP ESTIMATE Indicates that the user has requested a table-top
SID and field size estimate to the table-top
display.

19. SYSTEM FAILURE Indicates the system override (same as PBL-


OVERRIDE) key has been set during fluoro
operation.

Service Mode Messages

Various service mode messages exist. They are explained in Sections 4 and 5 in the
manual.

8.4 RADIOGRAPHIC (AUTOMATIC) OPERATION

1. Place X-ray generator in radiographic mode.

2. Insert cassette into tray and close tray.

3. Collimator will switch from MANUAL to EXPOSURE HOLD while the shutters are
moving, then to READY-RAD when the shutters are correctly positioned.

4. Press "LAMP" button to activate light field and bucky centering light line.

November 2005 ( 8-6 ) Linear IV Collimator


5. Properly center the bucky/cassette in the light field. This can be done by either
moving the bucky/cassette or the tube/collimator if possible. The bucky
centering light line should be lined up with the center of the tray handle to ensure
centering of the film cassette.

6. An exposure may now be taken. You may reduce the X-ray field size if desired
by either using the manual shutter control knobs or the long and cross open and
close buttons on the front panel. Please note that automatic PBL operation will
not allow the field size to be larger than the cassette size in the tray.

7. Any time a new cassette is put in the tray, the shutters will automatically size to
that cassette. Pressing the LAST SIZE button will move the shutters to the size
held prior to the new cassette being loaded.

8. The MEM ONE and MEM TWO buttons can be used to store the current field
size. Press and hold the MEM button (#6 or #11 in Figure 8.1) until the indicator
light (#7 or #12) turns off, then release the button and the size will be stored. To
later move to the stored size, press and release the MEM button and the shutters
will move to the stored size.

9. Please note the collimator will not move the shutters to a size larger than the
current cassette. For example, if a 10 x 10 is stored in MEM ONE while a 17 x
17 cassette is in place and later an 8 x 8 cassette is inserted, pressing MEM
ONE will have no effect. If one dimension is smaller than the new cassette size,
the shutters will move to that size. In the previous example, if 10 x 7 is stored in
MEM ONE, once an 8 x 8 cassette is inserted, pressing MEM ONE will move the
shutters to 8 x 7. However, 10 x 7 is still maintained as MEM ONE if a larger
cassette is later inserted.

10. The collimator is designed to operate at a fixed vertical SID (typically 40 inches
or 100 cm) unless it is a continuous or differential SID system. If the collimator is
not at the fixed vertical SID, the display will read either:

a. EXP HLD-LOCK SID if there is a cassette present but an improper SID


for automatic PBL operation, or

b. MAN-SID NOT SET if there is no cassette present indicating manual


table top exposure is allowed but the SID is unknown.

11. If part of the automatic PBL (positive beam limiting) system is not functioning or
a special procedure is required, the PBL system can be disabled by turning the
PBL override switch located at the collimator power supply. The collimator will
allow any shutter size to be selected, even if it is larger than the cassette size.
The collimator functions all remain the same except that the MEM ONE, MEM
TWO, and LAST SIZE buttons are disabled. When the collimator is returned to
normal automatic operation, the shutter size will remain unchanged unless the
current size is larger than the cassette. In this case, the shutters will size to the
cassette.

November 2005 ( 8-7 ) Linear IV Collimator


8.5 RADIOGRAPHIC (TABLE TOP/MANUAL) OPERATION
1. Place the X-ray generator in radiographic mode.
2. This display should read MAN-NO CASS with " - - x - - " for field size.

3. Place the film cassette on the table top. Press the LAMP button to activate the
lamp field. Position the cassette in the lamp field and manually adjust the
shutters and SID for desired technique.

4. Press the ENT/SET button to display table top estimates of field size and SID.

5. MEM ONE, MEM TWO, and LAST SIZE buttons perform no function in this
mode.

6. If the vertical SID is not the fixed SID (typically 40 inches or 100 cm), the display
will read MAN-SID NOT SET allowing table top exposure at an unknown SID.

8.6 CONTINUOUS AND DIFFERENTIAL SID OPERATION

1. The Linear IV collimator is capable of vertical continuous (tubestand movable)


and differential (both table and tubestand moveable) operation. Continuous SID
operation requires one SID monitor while differential SID operation requires two
SID monitors.

2. Operation sequences remain the same as in Section 8.4 with differences only in
what is displayed for SID and field size.

3. As the table or tubestand is moved, the SID display is updated to reflect the
current SID. If a cassette is loaded in the tray, the message area will display SID
MOVING while the shutters adjust to the new SID size. Once the shutters stop
moving, the display will revert to READY-RAD. If there is no cassette present
while the table or tubestand is moved, the SID display will be updated but the
field size will only indicate " - - x - -" and the message area MAN-NO CASS
until a cassette is inserted in the tray.

4. Please note that if you have adjusted the field size smaller than the cassette size
and move the table or tubestand to a different SID, the collimator will resize to
the full cassette size once movement is stopped. However, the MEM buttons
can be used to store and recall desired field sizes.

5. The MEM and LAST SIZE buttons function operate in the same manner as in
Section 8.4. In addition, the field size is maintained irregardless of SID. That is,
if 15 x 15 is stored in MEM ONE at 40 SID and the SID is changed to 48,
recalling MEM ONE will move the shutters to 15 x 15 at 48. If a new cassette
has not been loaded, this can also be easily accomplished by pressing the LAST
SIZE button. This will recall the last adjusted field size before the SID was
changed.

November 2005 ( 8-8 ) Linear IV Collimator


6. SID movement in the automatic mode can then be treated just like an insertion of
a new cassette in the tray. The field size is adjusted to the cassette in place and
any "coned down" size is placed in LAST SIZE for recall.
8.7 HORIZONTAL (WALL) OPERATION

1. The Linear IV is capable of operating in an automatic mode at four fixed


horizontal SID's (36", 40", 48" and 72" or 92 cm, 100 cm, 120 cm, and 180 cm).
Please refer to X-ray system information about the availability of these four
locations in your equipment.

2. The Linear IV operation horizontally is identical to that described in section 8.4.


Please note that continuous or differential operation is only possible vertically.

3. The current SID is displayed on the collimator as the collimator moves to


different horizontal SIDs. When the collimator is moved to a new horizontal SID,
the shutters automatically resize to the cassette currently in place (if any). The
MEM or LAST SIZE buttons function as in 8.4 to recall field sizes used at a
previous SID.

4. If the collimator is pointed to a wall device, but is not at one of the four allowed
horizontal SIDs, one of two messages will be displayed. If there is a cassette in
the wall holder, the display will read EXP HOLD-LOCK SID, indicating an invalid
SID position for automatic operation. If there is no cassette in the wallholder, the
display will indicate MAN-SID NOT SET which allows for non-bucky horizontal
exposures.

5. If the display reads MANUAL-TILT, the collimator is not pointing directly at the
wallholder (+/- 10%).

8.8 FLUOROSCOPIC OPERATION

1. Place the X-ray generator in fluoroscopic mode.

2. The message area should read (F) RDY-FLR NORM. The (F) indicates the
collimator is in the Full Film spotfilm mode (see Section 8.9); (P) would indicate
Preset spotfilm mode (Section 8.9). RDY-FLR indicates the collimator is ready
for fluoroscopic operation. NORM indicates the image intensifier (I.I.) set up as
NORM is selected.

3. Press the MAG SIZE button to toggle through the three preset I.I. size
selections. Each time a new I.I. is selected, the collimator will size to the correct
shutter position. You can verify the size of an I.I. selected by either 1) pressing
the LAST SIZE button if you have not adjusted the shutters since changing I.I.
size, or 2) pressing the ENT/SET button which toggles between Full Film and
Preset modes but will also display the I.I. size in the field size display.

November 2005 ( 8-9) Linear IV Collimator


4. You can reduce the X-ray field size at the front panel or by the use of an optional
remote control. The new field size is not displayed. To return to the full I.I. size,
press the MAG SIZE button once. Please note that if you've adjusted the field
size smaller then the I.I., the MAG SIZE button will reset to the full size of the
current I.I., not change I.I. sizes. If the current I.I. is fully sized, pressing the
MAG SIZE button will toggle I.I. sizes. The I.I. is fully sized if the field size
display reads " - - x - -". Pressing LAST SIZE when " - - x - -" is displayed will
show the I.I. size.

5. The MEM ONE and MEM TWO buttons function the same as in Section 8.4.
The sizes in memory are maintained even after switching I.I. sizes. Please note
that the shutters will never move larger than the current I.I. size. Unlike
radiographic operation, selecting MEM location if larger than I.I. will replace MEM
location with current I.I. size.

8.9 SPOTFILM OPERATION

1. While in the FLUORO mode, the Linear IV will enter the SPOTFILM mode when
the cassette tray is advanced into the exposure area. See your system
information for details on spotfilm operation.

2. The spotfilm operation of collimator will depend on which mode has been
selected: FULL FILM or PRESET. The PRESET mode allows the collimator to
keep a particular shutter size (as long as it is less than the cassette size) when
switching from fluoroscopic to spotfilm applications. Pressing the ENT/SET
button in FLUORO will toggle between PRESET and FULL FILM modes.

3. Indication of the current FULL FILM or PRESET operation during FLUORO or


SPOTFILM modes is indicated in two ways. First after depressing ENT/SET,
FULL FILM or PRESET will be indicated on the alphanumeric display on the front
panel for approximately two seconds. Afterwards a (F) or a (P) will precede the
message on that display to indicate FULL FILM or PRESET respectively.

4. FULL FILM OPERATION

When FULL FILM is selected the collimator can still be coned down to sizes
smaller than the maximum allowed by the current image receptor in use. But,
when switching from FLUORO to SPOTFILM modes, the collimator will always
initially cycle the shutters to the full cassette size in use. To return to the
previous coned down size, the user must depress the LAST SIZE button.

Returning to FLUORO, the collimator will resize to maximum II size for the
current magnification selected.

Selecting FULL FILM while in PRESET SPOTFILM mode will cause the
collimator to resize to the full cassette size. Pressing LAST SIZE will allow the
user to return to the previous coned down size.

November 2005 ( 8-10) Linear IV Collimator


Selecting FULL FILM while in PRESET FLUORO mode has no effect on shutter
position but does change the mode for the next SPOTFILM. MAG SIZE can be
used to open the shutters to the maximum II size for currently selected
magnification.

5. PRESET OPERATION

When PRESET is selected and SPOTFILM mode is entered from FLUORO, the
collimator will slightly readjust it's position in SPOTFILM mode to obtain the
same size it had just prior in FLUORO mode if it is within the limits of the
cassette size, without first sizing to the full cassette size. If a shutter is larger
than a cassette dimension, the shutter will close to the cassette size.

When PRESET is selected and FLUORO mode is entered from SPOTFILM, the
collimator will slightly readjust it's position in FLUORO mode to obtain the same
size it had just prior in SPOTFILM mode if it is within the limits of the maximum II
size for the current magnification, without first sizing to the maximum II size.
Again, a shutter which is beyond the limit will close to the II size.

Selecting PRESET while in a FULL FILM SPOTFILM mode has no effect on


shutter position, but it does set the current size as the new PRESET size.
Pressing LAST SIZE in this case will have no effect, as the current size is also
the LAST SIZE.

Selecting PRESET while in FULL FILM FLUORO mode has no effect on shutter
position but does change the mode for the next spotfilm application and engages
the size display to indicate numerically.

6. MAG SIZE BUTTON

Operation of MAG SIZE has not changed. It will always cause the shutter to
move to the maximum II size for that magnification and toggle the magnification
of I.I. between three inches.

8.10 TILTING COLLIMATOR AND TILTING TABLE OPERATION

1. Logic contained in the Linear IV makes sure that the collimator is pointed toward
the desired image receptor.

2. If the collimator is angled more than 10°, the collimator will display MANUAL-
TILT and blank the SID and field size display (regardless of whether a cassette
is present). Continuing to angle the collimator to 90° will lead to an error
message of either MAN-NO CASS or MAN-WRONG WALL. MAN-NO CASS
indicates the wall is an acceptable image receptor, but no cassette is present.
MAN-WRONG WALL indicates the collimator is pointing 180° from the calibrated
wall device. If there is a second wall device, only manual operation is allowed.

November 2005 ( 8-11) Linear IV Collimator


3. If the display reads MAN-SID NOT SET or EXP HOLD-LOCK SID is displayed
when pointing at a wall holder, the collimator is not located at a defined
horizontal SID (see Section 8.7).
4. If a table is tilted while the collimator is pointing down vertically, the display will
read MANUAL TILT. If a table is tilted to 90° in either direction and the collimator
is pointed at the table, the following error messages are possible:

a. EXP HOLD-LOCK SID: If the collimator is not at a defined horizontal


SID.

b. MAN-NO CASS: If there is no cassette in the table.

5. If the table can tilt 90° and be in the same direction as a wall holder, the
collimator logic looks at the table. That is, if a table is at 90° but without a
cassette, the display will read MAN-NO CASS irregardless if a wall holder in the
same direction has a cassette present.

8.11 STEREO/TOMO OPERATION

1. Stereo/Tomographic operation is enabled by a signal sent from the X-ray system


to the Linear IV collimator. Please refer to your specific system documentation
for stereo/tomo details.

2. When STEREO/TOMO is enabled, the collimator is placed in a completely


manual mode of operation. Field size and SID displays are blanked and the
collimator shutter size opening must be set by using the light field.

November 2005 ( 8-12) Linear IV Collimator


SECTION 9

REPLACEMENT
PARTS LIST

November 2005 ( 9-1 ) Linear IV Collimator


REPLACEMENT PARTS LIST
LINEAR IV

NOTE: THE ITEMS LISTED BELOW ARE THE MOST COMMONLY REQUESTED ITEMS.
FOR EUREKA PART NUMBERS OF ITEMS NOT LISTED HERE, PLEASE REFER TO THE
BILLS OF MATERIAL IN SECTION 11.

COLLIMATOR

EUREKA P/N DESCRIPTION


70-11201 Swivel Mounting Ring – Tube Side
70-11008 Swivel Mounting Ring – Collimator Side
70-10036 Swivel Ring Collar
26-00854 Thumb Screw
70-11089 Window - Cross Hair

70-04752 Knob - Front Panel


70-11039 Lamp Switch Pushbutton Assembly
70-04571 Lamp - Light Field - DZE 24 VAC, 150 W (mfg. before December 2003)
70-04300 Lamp – Light Field – FCS 24 VAC, 150 W (mfg. after November 2003)
70-04572 Socket – Lamp (mfg. before December 2003)
70-04299 Socket – Lamp (mfg. after November 2003)
70-01901 Current Limit Resistor

70-11129 Cover - Left Side


70-11130 Cover - Right Side
70-11026 Cover - Front Bezel
70-11127 Cover - Top
70-11028 Cover - Rear

70-03051 Triac - 15 A Light Timer


70-10096 Prism - Centering Line
70-10147 Tilt Switch
70-20265 Cone Track Kit (Option)
70-11200 Dual Cone Track (for use with dosimeters)
70-20164 Low Filtration Mirror/Bracket Assembly
70-20024 Mirror/Bracket Assembly
70-20318 Handheld Remote
27-30092 Motor, Linear IV Stepper
70-11075 Tape Measure

November 2005 ( 9-2 ) Linear IV Collimator

EUREKA P/N DESCRIPTION


70-08167 EEPROM - Linear IV
70-08314 CPU Board Assembly
70-08316 Driver Board Assembly
70-08318 Display Board Assembly
70-08153 Front Panel Touchscreen
70-04280 5 Volt Logic Switching Power Supply

POWER SUPPLY CHASSIS


70-06018 Transformer - Power 27/19 VAC
70-04032 EMI Filter
70-04605 Fuse - 3A Slo Blo - Power
70-04607 Fuse - 8A Slo Blo - Lamp
70-08320 Linear IV Master Board Assembly
70-04001 Relay - 24 VDC

SID MONITOR
70-00652 Potentiometer, 1K, 10 Turn
70-04241 Switch, Snap Action
70-10558 Cable - 70 inch

November 2005 ( 9-3 ) Linear IV Collimator


NO TEXT

November 2005 (9-4 ) Linear IV Collimator


SECTION 10

APPENDIX

November 2005 ( 10-1 ) Linear IV Collimator

DEFINITIONS
SID Source to Image receptor Distance

VSID Voltage representing SID

VCSID Voltage representing Continuous SID

SID TRUE Signal representing the Operating SID Range

XF X-ray Field

VXFC Voltage at the Collimator Feedback Potentiometer Wiper repesenting the


X-Ray Field in the Cross Dimension

VXFL Voltage at the Collimator Feedback Potentiometer Wiper representing the


X-Ray Field in the Long Dimension

IR Image Receptor (Cassette Tray)

VIRC Voltage from the Cassette Sensing Element representing the Image
Receptor size in the Cross Dimension

VIRL Voltage from the Cassette Sensing Element representing the Image
Receptor size in the Long Dimension

IR TRUE Voltage representing the Presence of a Cassette

VCPC Voltage applied to the Collimator Potentiometers in the Cross Position

VCPL Voltage applied to the Collimator Potentiometers in the Long Position

VXFDC Voltage representing the X-Ray Field Desired in the Cross Dimension

VXFDL Voltage representing the X-Ray Field Desired in the Long Dimension

November 2005 ( 10-2 ) Linear IV Collimator


TABLE 10.1
The following list is intended to help the installer determine mounting information only, and does
not imply compatibility. See Section 1-7 for compatability information.

FOCAL SPOT DISTANCE TO


TO COLLIMATOR
MANUFACTURER TUBE HOUSING PORT MOUNTING MOUNTING FLANGE
Eureka Emerald Series 2 - 1/16” 3/8” (.375)
Diamond Series 2 - 1/16” 3/8” (.375)
Sapphire Series 2 - 1/16” 1/4” (.250)
Eimac B100 2 - 17/64” 11/64” (.171)
B150 2 - 11/64” 17/64” (.265)
General Electric Maxiray 100 2 - 5/16” 1/8” (.125)
HRT, MX75 2 - 1/16” 3/8” (.375)
Picker/Dunlee DU - 140 2 - 1/16” 3/8” (.375)
DU - 200 2 - 1/16” 3/8” (.375)
DU - 300 2 - 9/32” 5/32” (.156)
PX - 400 2 - 5/16” 1/8” (.125)
PX - 1300 2 - 3/16” 8/32” (.252)
PX - 1400 2 - 5/16” 1/8” (.125)
Machlett DX40 Series 2 - 1/16” 3/8” (.375)
DX50 Series 2 - 3/16” 1/4” (.250)
DX60 Series 2 - 5/16” 1/8” (.125)
DX70 Series 2 - 5/16” 1/8” (.125)

November 2005 (10-3) Linear IV Collimator

LINEAR IV TILT SWITCH WIRING (70-10147)


GRN

BLU
GRN J6-4
BRN

WHT J6-1

RED
J6-2
BLK
J6-3

FIGURE 10.1

LINEAR IV FEEDBACK POTS WIRING (70-00655)

BLK J6-12
1
CROSS
POT

WHT J6-11
2

VIO J6-10
3

J6-9
1

J6-8
LONG POT 2

3
J6-7

LINEAR IV MOTOR WIRING (27-30092)


RED FIGURE 10.2
J2-10

November 2005 ( 10-4 ) Linear IV Collimator


PHASE A

CROSS
MOTOR BRN
J2-9
BLK
J2-12

PHASE B

GRN
J2-11

RED
J2-3

LONG
MOTOR

PHASE A

BRN
J2-4
BLK
J2-5

PHASE B

J2-6
GRN

FIGURE 10.3
November 2005 ( 10-5 ) Linear IV Collimator
LINEAR IV COLLIMATOR
PAUSCH URO TABLE INTERFACE
1. Product consists of a single printed circuit card (PCB) to be a maximum dimension of 6”
x 4” rectangle. The PCB has a threaded standoff in each corner extending 1 inch below
the solder side for mounting in the customer’s interface unit.

2. The PCB has a screw down wire terminal strip connector for making all interface
connections (the same type of terminal strip as used in the Linear IV Master PCB).

3. The required interface connections are as follows:

FROM PAUSCH CASSETTE TRAY: TO PCB TERMINAL STRIP (J1)


/TRAY IN J1-1
/17x14 J1-2
/14x17 J1-3

FROM COLLIMATOR MASTER PCB POWER SUPPLY:

Coll. Ref. Source (TS4-15) J1-7


Coll. Ref. Return (TS4-12) J1-8
Coll. +15V (TS1- 3) J1-5
Coll. GND (TS1- 2) J1-6
Coll. Long Ref (TS4-13) J1-9
Coll. Cross Ref (TS4-14) J1-10

4. The Optional Interface connections are as follows:

Coll. Exp Hold 1 J1-11


Coll. Exp Hold 2 J1-12

RAD/FLUORO Interlock 1 J1-14


RAD/FLUORO Interlock 2 J1-15

November 2005 (10-6 ) Linear IV Collimator


SECTION 11

THEORY OF OPERATION

November 2005 ( 11-1 ) Linear IV Collimator


Linear IV Collimator
Figure 11.1

November 2005 ( 11-2 ) Linear IV Collimator


11.1 MECHANICAL OPERATION

Eureka Linear IV collimators contain two major sets of shutters, long and cross, which
define the absolute X-ray field size. There are also two secondary sets of shutters
which protrude into the port of the X-ray tube to help reduce the effects of off-focus
radiation. The secondary cross shutters are mechanically linked to the primary cross
shutters and follow the action of the primary shutters to provide additional reduction of
off-focus radiation with smaller X-ray field sizes.

Both shutter mechanisms are driven by bipolar stepper motors and the front panel
knobs. Spring loaded gears are used to prevent backlash for accurate, repeatable
shutter positioning.

All Eureka Linear Series collimators have a swivel mount configuration. Detents are
located at 90 degree increments. The collimator may be oriented to any position for
achieving proper X-ray field to cassette alignment for table-top or non-bucky operation.

The shutter mechanism has been precisely aligned with respect to the mounting flange
at the factory. Therefore, the necessity for field alignment of the central ray has been
virtually eliminated.

The shutters may be positioned manually with the remote shutter controls on the
optional remote control unit, or the controls on the front panel.

11.2 ELECTRONIC OPERATION

The collimator electronics reside on a number of printed circuit boards:

- CPU Logic PCB


- Master Board
- Driver Board
- Display Board
- Front Panel
- Five Volt Power Supply

For simpler schematic interpretation, each board has been divided into logical
subsections.

The operating system is represented by the block diagrams of Figure 10.2 a and b.
Refer to section 9 for definitions.

November 2005 ( 11-3 ) Linear IV Collimator


(OPTION)

REMOTE COLLIMATOR
CONTROL HEAD

HORIZONTAL TABLE
SID – BUCKY
SWITCHES

VERTICAL POWER WALL


SID SUPPLY BUCKY
MONITORS CHASSIS

IMAGE
INTENSIFIER
CONTROL

Figure 11-2a - Collimator Block Diagram - General

November 2005 ( 11-4 ) Linear IV Collimator


Figure 11-2b - Collimator Block Diagram – Logic

November 2005 ( 11-5 ) Linear IV Collimator


11.3 POWER SUPPLIES

All power supplies are self contained: a transformer with 110, 115, 120, 220, 230, etc.
VAC primary windings supplies AC power for the projection lamp (27 VAC) and for the
stepper motors (28 VDC). Please note the proper incoming voltage (110V or 220V) has
to be selected on the switch on the power supply.

A separate logic supply provides 5 VDC, and +15 VDC for the PC boards.

11.4 CONTINUOUS SID CIRCUITS

The constant current sources associated with Q2 and Q3 on the Linear IV master board
(page 5, 70-08319) provide 5mA nominally to the connected SID monitors. The voltage
developed across the SID monitor potentiometer therefore varies proportionally with
changes in SID. The voltages VSID1 and VSID2 are fed into the A/D Converter (U30,
70-08313). The software interprets the incoming voltages and based on calibration
information, determines the correct SID in inches. This ensures that the collimator
shutters open or close to maintain constant frame size for a given format at all SID’s in
the range of the table vertical carriage.

NOTE: VSID2 is used only if the system has an adjustable height table and a
second SID monitor is installed in the table and connected to the collimator
(CSID2).

11.5 IMAGE RECEPTOR SIZE CIRCUIT

The voltages representing image receptor size (VIRL, VIRC) are fed into differential
amplifiers U11a and U11b (70-08319). These amplifiers are used to provide floating,
balanced inputs for the outboard potentiometers. This circuitry minimizes noise induced
in the sometimes long interconnecting cables between the potentiometers and logic
circuitry.

These voltages are then fed into the A/D converter (U30, 70-08313). The software then
interprets these voltages based on calibration information to determine the correct
image receptor size.

11.6 COLLIMATOR FEEDBACK CIRCUIT

The voltages representing the position of the shutters is fed into the A/D converter (U30)
through the unity gain follower amplifiers of U28a and U28b (70-08313). The software
then interprets these voltages based on calibration information and all required inputs
(including SID, IR size) as a position (angle) which produces an associated frame size.

11.7 A/D CONVERTER CIRCUIT

The A/D converter (U30, 70-08313, sheet 4) accepts the analog signals described in

November 2005 ( 11-6) Linear IV Collimator


10.4 - 10.6. An internal multiplexor samples the given inputs and stores the results in an
internal FIFO at a preconfigured rate. An interrupt is generated by the A/D converter
when the FIFO is full which causes the CPU to store the current values of the analog
inputs.

This technique requires a minimum of overhead from the CPU and allows for a fast,
accurate means of always recording the current analog input values for use by the
software when required.

11.8 CPU CORE SECTION (Schematic 70-08313 sheet 1, Board 70-08314)

U1 is the microprocessor used in the system. It is a Zilog Z180 running at 10 MHz. It


has a 16 bit address space buffered by U4 and U5, and an 8 bit data bus buffered by
U3.

The logic associated with U7 provides the required signals to determine I/O and Memory
reads and writes.

U8 is a 8K X 8 (6264) static ram used for storing volatile data and the parameter stack.

U6 is a 32K X 8 (28C256) EEPROM and contains the run-time code and is used to store
the nonvolatile parameters determined during calibration. The EEPROM is Hardware
and Software write protected to provide the maximum security against losing data from
false writes which may be encountered during power surges or from noise spikes.

11.9 CPU DISPLAY/KEY BOARD SECTION (70-08313 sheet 2)

U11 provides the correct operational clock speed for U10 the 8279 keyboard display
controller.

U10, U16 and part of U23 supply the required scan and return signals for keyboard
recognition from the front panel and an optional remote keypad.

U10, U9 and the associated drive circuits provide the required display signals to
illuminate the front panel, display board (70-08317), and the optional remote keypad
control.

11.10 CPU INPUT/OUTPUT SECTION (70-08313 sheet 3)

U20 the 74LS138 is a 3 to 8 decoder used to develop the appropriate chip enable
signals used throughout the CPU board.

U21 is an octal latch and is used to maintain signals concerned with discrete SID
switches, Radiographic or Fluoroscopic mode, and Stereo/TOMO mode. The software
checks the latch when the information is required.

November 2005 ( 11-7) Linear IV Collimator


U19 an 8255 I/O Controller along with the associated buffer circuits provide for the
discrete input/output requirements of the collimator system. The chip is divided into 3
ports. Port A Monitors the Tilt switch positions, and Spotfilm true signals. Port B
provides the outputs for control of the Stepper Motors through the driver board (70-
08315), and Port C outputs current tilt information and the signal required for Image
intensifier relays and the Exposure hold Relay.

The 8255 is configured in software and is controlled by the software.

11.11 MASTER BOARD (Schematic 70-08319, Board 70-08320)

Sheet 1 contains the schematic for power supply wiring and the STEREO/TOMO
interconnect circuit. The schematic is relatively simple and straight-forward so no
further explanation will be provided.

Sheet 2 contains the circuit schematics for the 28VDC supply for the stepper motors,
and the drive circuitry for the relays associated with PBL OVER_RIDE, EXPOSURE
HOLD, and IMAGE INTENSIFIER selection. Again the circuits here are simple and
straight forward and require no additional explanation.

Sheet 3 contains the circuit schematics for discrete SID determination. U7 (74139) and
U4 (74148) are used to respectively decode and code the current discrete SID switch
status used by the CPU and software. The following logic tables describe the signal
dependencies.

Sheet 4 contains the Image Receptor reference and measurement circuits. U11 is
described in section 10.5. The circuits associated with Q1 and U9 create a power
amplifier circuit which provides a constant voltage to the Image Receptor
potentiometers. U10 and U3 compare the currents associated with a cassette in and
out of the image receptor to provide a signal which tells the CPU whether a cassette is
present in the selected bucky. Relay K5 is controlled by the CPU and connects the
desired bucky tray to the reference and output signals.

Sheet 5 contains the SID circuits described in section 11.4.

TILT INFO B TILT INFO A SID SID SID VERT


INFO C INFO B INFO A IN
VERT OUT 0 0 72”SID 0 1 1 X
TABLE L 0 1 48” SID 0 1 0 X
TABLE R 1 0 40” SID 0 0 1 X
WALL L/R 1 1 36” SID 0 0 0 X
VERT IN X X X 1
NO SID 1 0 0 X

Sheet 6 contains the circuit for Radiographic or Fluoroscopic mode selection.


Stereo/Tomographic indication, and table tilt indication. These circuits again are
relatively simple and no further explanation will be given.

November 2005 ( 11-8 ) Linear IV Collimator


11.12 DRIVER BOARD (Schematic 70-08315, Board 70-08316)

Sheet 1 contains the circuit for illuminating and turning of the internal projection lamp.
The logic of U6 and U8 trigger or reset the timer U1 based on current timer status. The
timer-in-turn triggers the optically isolated triac which turns on/off the lamp.

If the lamp is off and the lamp switch is pressed the lamp will turn on until it times out
(approximately 22 seconds) or until it is turned off by pressing the lamp switch again.

Sheet 2 contains the drive and control circuitry associated with the stepper motor drive.
U9 and U7 provide two time bases for manual and automatic motor speeds. The speed
selected is determined through the logic of U2. Automatic speed is used during
automatic sizing cycles of the collimator. A slower manual speed is introduced for
power assisted manual control of the shutter position using the front panel controls or
the optional remote keypad controls.

U3 and U10 provide the logic interface for power H-bridges U11 and U4. Together they
form a current chopper drive for the stepper motors of the cross and long shutters.
Each shutter is controlled by an enable and direction signal fed from the CPU and runs
at the speed determined by the logic of U2 fed from the CPU.

11.13 DISPLAY BOARD (Schematic 70-08317, Board 70-08318)

Sheet 1 contains the schematic of the alpha-numeric display section of the display
board. DS1-DS4 are “smart” alpha-numeric displays which display the ASCII input on
the data bus into the appropriate ASCII character on the selected character of the
display.

The logic of U2 selects the display and the character of the display from information
from the CPU.

U1 and it’s associated circuit provide a dimmer circuit for the displays to maintain a
lower current draw and provide for optimum illumination.

Sheet 2 contains the schematic for the field size and SID numeric displays. The circuit
is simple and consists of driver chips and seven segment numeric displays. The
displays are driven in a multiplexed fashion controlled by the 8279 on the CPU board.
U3 selects which is to be illuminated, U4 contains the information on the segments to be
illuminated.

11.14 FRONT PANEL 70-08152

The front panel consists of matrix of metallic dome keypad switches which are
interpreted by the 8279 on the CPU board for their respective function. There are a
number of indicators on the front panel controlled by the CPU’s 8279 used to indicated
functional status. Section 8.0, Operating Instructions, provides detailed interface
instructions.

November 2005 ( 11-9 ) Linear IV Collimator


11.15 INTERCONNECT CABLE 70-08166

Figure 11.3 describes the specific signal and wire information on the cable.

November 2005 ( 11-10 ) Linear IV Collimator


FIGURE 11.3
INTERCONNECT CABLE (70-08166)

POS SIGNAL NAME WIRE PWR SUPPY HEAD CPU


1 SID INFO C RED/WHT/BLK TS2-1 T J8 – 16
2 SID INFO B RED/BLK/GRN TS2-2 S J8 – 17
3 SID INFO A RED/WHT/GRN TS2-3 R J8 – 18
4 VIN ORG/GRN TS2-4 U J8 – 15
5 TILT INFO B WHT/BLU/RED TS2-5 P J8 – 19
6 TILT INFO A WHT/BLK/RED TS2-6 N J8 – 20
7 NC TS2-7
8 +5 VDC RED TS2-8 c J2 – 18
RED/BLK d J2 – 17
RED/GRN d J2 – 16
RED/WHT e J2 – 15
WHT/BLK e J2 – 14
9 DGND BLK TS2-9 b J2 – 19
BLK/RED f J2 – 13
BLU/RED f J2 – 12
BLK/WHT h J2 – 11
ORG/RED/GRN h J2 – 10
BLK/GRN/WHT j J2 – 9
BLK/BLU/WHT j
BLK/GRN/RED j
10 PGND BLK/RED/WHT TS2-10 k J2 – 8
WHT/RED/GRN k J2 – 7
BLK/WHT/ORG l J2 – 6
WHT/RED/ORG l J2 – 5
11 28 VDC ORG TS2-11 m J2 – 4
GRN/ORG/BLK m J2 – 3
BLU/WHT/ORG n J2 – 2
GRN/WHT/BLK n J2 – 1
12 0 VAC BLU/RED/GRN TS2-12 t Jumper
GRN/WHT/BLU t to u
13 0 VAC GRN/BLK TS2-13 u To Triac
BLU u GND
14 27 VAC WHT/BLU/BLK TS2-14 r Jumper
GRN r to S
15 27 VAC ORG/RED TS2-15 s TO
BLU/WHT s LAMP
16 EHR ORG/WHT/BLK TS3-1 NN J8 – 1
17 M2R BLK/ORG/RED TS3-2 MM J8 – 2
18 M1R GRN/RED/ORG TS3-3 LL J8 – 3
19 N.C. TS3-4

November 2005 ( 11-11 ) Linear IV Collimator


FIGURE 11.3 CONTINUED
INTERCONNECT CABLE (70-08166)

POS SIGNAL NAME WIRE PWR SUPPY HEAD CPU


20 T0 ORG/WHT/BLU TS3 – 5 X J8 – 12
21 T + 90 ORG/BLK/GRN TS3 – 6 W J8 – 13
22 T – 90 ORG/BLK TS3 – 7 V J8 – 14
23 SFT BLU/BLK/WHT TS3 – 8 z J8 – 9
24 R/F ORG/RED/BLU TS3 – 9 AA J8 – 8
25 IRT RED/WHT/BLU TS3 – 10 CC J8 – 7
26 VIRL GRN/WHT TS3 – 11 EE J8 – 5
27 VIRC WHT/BLK/GRN TS3 – 12 DD J8 – 6
28 VSID2 BLU/BLK TS3 – 13 a J8 – 11
29 VSID1 WHT/RED TS3 – 14 w J8 – 10
30 ST BLU/RED/ORG TS3 – 15 JJ J8 – 4
31 IR ALIGN WHT N.C. x To Fast On

November 2005 ( 11-12 ) Linear IV Collimator


SECTION 12

SCHEMATICS AND BILLS


OF MATERIAL

November 2005 ( 12-1 ) Linear IV Collimator


Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08314


Rev. E ECN: P0818
-ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.
NO. SIZE NO. CLS Part No. Designator
REF 0 70-08318 Schematic, CPU Board
0001 1 70-06303 CPU PCB Fabrication Dwg
0002 1 70-03152 Z180 MPU Z80180 U1
0003 1 70-03121 IC 8 Pin Dip Watchdog Timer DS1232 U2
0004 1 70-03150 IC 20 Pin Dipo Octal Bus/Line Tranceiver 74LS245 U3
0005 2 70-03151 IC 20 Pin Dip Octal Buffer & Line Driver 74LS244 U4, U5
0006 1 70-08167 Programmed EEPROM Rev. 2.04 28C256 U6
0007 1 70-03142 IC 14 Pin Dip Quad 2 – Input or Gate 74LS32 U7
0008 1 70-03157 8K x 8 RAM 6264 U8
0009 1 70-03139 IC 24 Pin Dip, 14-16 Decoder/Demultiplexer 74L154 U9
0010 1 70-03118 PRO Keyboard Display Controller 8279 U10
0011 1 70-03153 D-Type Flip Flop 14 Pin 74LS74 U11
0012 4 70-03138 Hex Inverting Buffer 14 Pin 7406 U12, 13, 15, 24
0013 3 70-03119 Hex Buffer/Hi Voltage 14 Pin 7407 U22, 23, 26
0014 1 70-03087 8 Channel Source Buffer 18 Pin UDN2585 U17
0015 2 70-03086 8 Channel Current Sink Buffer 18 Pin UDN2595 U18, 25
0016 1 70-03140 Programmable I/O 40 Pin 8255 U19
0017 2 70-03128 Decoder/Demultiplexer 16 Pin 74LS138 U16, 20
0018 1 70-03124 Octal D – Type Latch 20 Pin 74LS373 U21
0019 3 70-03351 Dual Op-Amp 8 Pin LM358 U27-29
0020 1 70-03148 12 Bit A/D Data Acq System 44 Pin LM12458 U30

SHEET 1 OF 3
ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08314


Rev. E ECN: P0818
NO. SIZE NO. CLS Part No. Designator
0022 33 70-02010 Capacitor, .1 MFD 50 V C1-C30, C32, C34,
C41
0024 9 70-02511 Capacitor, Tant. 10 MF, 35V-10% C31,C33,C35,C36,
C37,C46,C49,C52,
C53
0025 2 70-02031 Capacitor, MICA, 22PF 300V C38,C39
0026 1 70-02510 Capacitor, Tant. 100 MF, 10 V C40
0027 2 70-02506 Cap. Solid Tant. 2.2UF 35 WVDC C47,C48
0028 2 70-03012 Rectifier 1 Amp 100 V CR2, CR3
0029 1 70-03071 Suppressor, Transient +5V CR1
0030 1 70-03302 3 Term Adj. Regulator LM317 VR1
0031 1 70-03154 Crystal 10 MHZ Y1
0032 1 70-04253 Switch, Pushbutton SW1
0033 2 70-00700-11 Resistor, Metal Film 237 Ohms 1% R1, R2
0034 1 70-00314 Resistor, 4.99K Ohms-1/4W 1% R3
0035 3 70-00042 Resistor, 1K Ohms-1/4W 5% R4,R7,R8
0036 1 70-00069 Resistor, Carbon Film, 20K Ohms 1/4W 5% R5
0037 8 70-00510 Resistor, Array, 10 Pin sip, 10K Ohms RP1-RP4
RP7-RP10
0038 3 70-00511 Resistor, Array, 6 Pin sip, 10K Ohms RP5,RP11,RP12
0039 1 70-00518 Resistor Network, 6 Pin, 1K Ohms RP6
0041 4 70-04726 Header 20 Pin J1, J2, J3, J8
0042 1 70-04727 Conn., Header 20 Pin IDC Rt. Ang. J4
0043 1 70-04526 Header, 34 Pin J5
SHEET 2 OF 3

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.


Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08314


Rev. E ECN: P0818
NO. SIZE NO. CLS Part No. Designator
0044 1 70-04731 Conn., HDR 12 Pin Mini PV Rt. Ang. J6
0045 1 70-04730 Conn., Header 6 Pin Mini PV Rt. Ang. J7
0046 1 70-04282 Socket, IC 68 Pin PLCC to PGA U1
0047 1 70-04279 Socket, IC 44 Pin PLCC to PGA U30
0048 2 70-04220 Socket, Mach. 40 Pin Dip U10,U19
0049 2 70-04219 Socket, Mach. 28 Pin Dip U6, U8
0050 1 70-04283 Socket, Open Frame 24 Pin U9
0051 4 70-04221 Socket, Mach. 20 Pin Dip U3,U4,U5,U21
0052 3 70-04281 IC Open Frame Collet Socket 18 Pin U17,U18,U25
0053 2 70-04217 IC Socket Mach. 16 Pin U16,U20
0054 9 70-04216 Socket, Mach. 14 Pin Dip U7,U11,U12,U13,
U15,U22,U23,U24,
U26
0055 4 70-04222 Socket, Mach. 8 Pin Dip U2,U27,U28,U29
0056 1 70-04806 Two Circuit Shunt W1
0057 1 70-04758 2 Pin Header W1
0058 2 70-00064 Resistor, 10K Ohms, 1/4W, 5% R6,R9
0059 2 70-02012 Capacitor, Ceramic 100V C46,C49

SHEET 3 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08314-1


Rev. D ECN: P0818
ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.
NO. SIZE NO. CLS Part No. Designator
REF 0 70-08318 Schematic, CPU Board
0001 1 70-06303 CPU PCB Fabrication Dwg
0002 1 70-03152 Z180 MPU Z80180 U1
0003 1 70-03121 IC 8 Pin Dip Watchdog Timer DS1232 U2
0004 1 70-03150 IC 20 Pin Dipo Octal Bus/Line Tranceiver 74LS245 U3
0005 2 70-03151 IC 20 Pin Dip Octal Buffer & Line Driver 74LS244 U4, U5
0006 1 70-08389 Programmed EEPROM Rev. 2.041 28C256 U6
0007 1 70-03142 IC 14 Pin Dip Quad 2 – Input or Gate 74LS32 U7
0008 1 70-03157 8K x 8 RAM 6264 U8
0009 1 70-03139 IC 24 Pin Dip, 14-16 Decoder/Demultiplexer 74L154 U9
0010 1 70-03118 PRO Keyboard Display Controller 8279 U10
0011 1 70-03153 D-Type Flip Flop 14 Pin 74LS74 U11
0012 4 70-03138 Hex Inverting Buffer 14 Pin 7406 U12, 13, 15, 24
0013 3 70-03119 Hex Buffer/Hi Voltage 14 Pin 7407 U22, 23, 26
0014 1 70-03087 8 Channel Source Buffer 18 Pin UDN2585 U17
0015 2 70-03086 8 Channel Current Sink Buffer 18 Pin UDN2595 U18, 25
0016 1 70-03140 Programmable I/O 40 Pin 8255 U19
0017 2 70-03128 Decoder/Demultiplexer 16 Pin 74LS138 U16, 20
0018 1 70-03124 Octal D – Type Latch 20 Pin 74LS373 U21
0019 3 70-03351 Dual Op-Amp 8 Pin LM358 U27-29
0020 1 70-03148 12 Bit A/D Data Acq System 44 Pin LM12458 U30

SHEET 1 OF 3
ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08314-1


Rev. D ECN: P0818
NO. SIZE NO. CLS Part No. Designator
0022 33 70-02010 Capacitor, .1 MFD 50 V C1-C30, C32, C34,
C41
0024 9 70-02511 Capacitor, Tant. 10 MF, 35V-10% C31,C33,C35,C36,
C37,C46,C49,C52,
C53
0025 2 70-02031 Capacitor, MICA, 22PF 300V C38,C39
0026 1 70-02510 Capacitor, Tant. 100 MF, 10 V C40
0027 2 70-02506 Cap. Solid Tant. 2.2UF 35 WVDC C47,C48
0028 2 70-03012 Rectifier 1 Amp 100 V CR2, CR3
0029 1 70-03071 Suppressor, Transient +5V CR1
0030 1 70-03302 3 Term Adj. Regulator LM317 VR1
0031 1 70-03154 Crystal 10 MHZ Y1
0032 1 70-04253 Switch, Pushbutton SW1
0033 2 70-00700-11 Resistor, Metal Film 237 Ohms 1% R1, R2
0034 1 70-00314 Resistor, 4.99K Ohms-1/4W 1% R3
0035 3 70-00042 Resistor, 1K Ohms-1/4W 5% R4,R7,R8
0036 1 70-00069 Resistor, Carbon Film, 20K Ohms 1/4W 5% R5
0037 8 70-00510 Resistor, Array, 10 Pin sip, 10K Ohms RP1-RP4
RP7-RP10
0038 3 70-00511 Resistor, Array, 6 Pin sip, 10K Ohms RP5,RP11,RP12
0039 1 70-00518 Resistor Network, 6 Pin, 1K Ohms RP6
0041 4 70-04726 Header 20 Pin J1, J2, J3, J8
0042 1 70-04727 Conn., Header 20 Pin IDC Rt. Ang. J4
0043 1 70-04526 Header, 34 Pin J5
SHEET 2 OF 3

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.


Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08314-1


Rev. D ECN: P0818
NO. SIZE NO. CLS Part No. Designator
0044 1 70-04731 Conn., HDR 12 Pin Mini PV Rt. Ang. J6
0045 1 70-04730 Conn., Header 6 Pin Mini PV Rt. Ang. J7
0046 1 70-04282 Socket, IC 68 Pin PLCC to PGA U1
0047 1 70-04279 Socket, IC 44 Pin PLCC to PGA U30
0048 2 70-04220 Socket, Mach. 40 Pin Dip U10,U19
0049 2 70-04219 Socket, Mach. 28 Pin Dip U6, U8
0050 1 70-04283 Socket, Open Frame 24 Pin U9
0051 4 70-04221 Socket, Mach. 20 Pin Dip U3,U4,U5,U21
0052 3 70-04281 IC Open Frame Collet Socket 18 Pin U17,U18,U25
0053 2 70-04217 IC Socket Mach. 16 Pin U16,U20
0054 9 70-04216 Socket, Mach. 14 Pin Dip U7,U11,U12,U13,
U15,U22,U23,U24,
U26
0055 4 70-04222 Socket, Mach. 8 Pin Dip U2,U27,U28,U29
0056 1 70-04806 Two Circuit Shunt W1
0057 1 70-04758 2 Pin Header W1
0058 2 70-00064 Resistor, 10K Ohms, 1/4W, 5% R6,R9
0059 2 70-02012 Capacitor, Ceramic 100V C46,C49

SHEET 3 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08316


Rev. G ECN P0719

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.


NO. SIZE NO. CLS Part No. Designator
0001 1 B 70-06304 Linear IV Driver Board Fabrication
0002 1 A 70-03030 Bridge Rectifier DF02 CR2
CR3-CR10, CR12-CR19
0003 16 A 70-03034 Ultrafast Rectifier MUR105
0004 6 A 70-02509 Capacitor, Solid Tantalum 35 WVDC 2.2 UF C27-C32
C2, C3, C4, C11, C13, C15,
0005 11 A 70-02010 Capacitor, CER 50V 0.1 UF C16, C17, C22, C24, C33
0006 2 A 70-02524 Capacitor, ELEC 50V 100 UF C5, C23
0007 2 A 70-02525 Capacitor, Polyester Film 63V 0.22 UF C6, C25
0008 7 A 70-02012 Capacitor, CER 0.01 UF C7, C10, C12, C14,
C18, C19, C20
0009 3 A 70-02511 Capacitor, TANT 35 VDC 10 UF C8, C9, C34
0010 1 A 70-02032 Capacitor, CER .003 UF C21
0011 1 A 70-02523 Capacitor, ELEC 50 VDC 47 UF C26
0012 1 A 70-02011 Capacitor, CER 50 VDC 0.47 UF C35
0013 1 B 70-04288 Connector, Header 20 Pin J1
0014 1 B 70-04289 Connector, Header 12 Pin Straight J2
0015 2 A 70-00511 Resistor Array 6 Pin sip 10 K RP1, RP4
0016 1 A 70-00518 Resistor Array 6 Pin sip 1K RP2
SHEET 1 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08316


Rev. G ECN P0719

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.


NO. SIZE NO. CLS Part No. Designator
0017 1 A 70-00501 Resistor Array 8 Pin sip 680 RP3
0018 1 A 70-00018 Resistor, Carbon Film 1/4 W +/- 5% 100 R1
0019 1 A 70-00110 Resistor, Carbon Film 1/4 W +/- 5% 2.2 M R2
0020 1 A 70-00085 Resistor, Carbon Film 1/4 W +/- 5% 91 K R3
0021 3 A 70-00079 Resistor, Carbon Film 1/4 W +/- 5% 51 K R4, R9, R15
0022 1 A 70-00056 Resistor, Carbon Film 1/4 W +/- 5% 4.7 K R5
0023 4 A 70-00325 Resistor, Carbon Film 1/4 W +/- 5% 1.0 R6, R7, R13, R14
0024 1 A 70-00100 Resistor, Carbon Film 1/4 W +/- 5% 470 K R8
0025 1 A 70-00070 Resistor, Carbon Film 1/4 W +/- 5% 22 K R10
0026 1 A 70-00035 Resistor, Carbon Film 1/4 W +/- 5% 510 R11
0027 1 A 70-00030 Resistor, Carbon Film 1/4 W +/- 5% 330 R12
0028 1 A 70-00421 Resistor, Wirewound 3 W +/- 5% 47 R16
0029 3 A 70-03353 Timer 555 8 Pin Dip LM555 U1, U7, U9
0030 1 A 70-03136 Quad 2-Input NAND Gate 14 Pin Dip 7400 U2
0031 2 A 70-03084 Stepper Motor Controller L297 U3, U10
0032 2 A 70-03035 Dual Full - Bridge Driver L298 U4, U11
0033 1 A 70-03043 Opto Coupler/Isolator MOC3011 U5
0034 1 A 70-03160 I.C. Schmitt Trigger 74HC132 U6
0035 1 A 70-03104 Dual D Flip-Flop 14013 U8
SHEET 2 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08316


Rev. G ECN P0719

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.


NO. SIZE NO. CLS Part No. Designator
0036 1 A 70-03303 Voltage Regulator, +5V 7805 U12
0037 3 A 70-03094 Opto Isolator MCT62 U13, U14, U15
0038 2 A 70-06524 Jumper, Buss Bar, 24 AWG JMP W1, W2
0039 1 A 70-00108 Resistor, Carbon Film 1/4 W +/- 5% 1M R17
0041 3 A 70-11166 Nut, 6-32 “Thin”
0042 1 B 70-04778 Heat Sink U12
0043 6 70-04222 Socket U1, U7, U9, U13-15
0044 3 70-04216 Socket U2, U6, U8
0045 1 70-04506 Socket U5
0046 2 70-04221 Socket U3, U10
0047 3 A 26-00830 Screw, #6-32 x ¼" Phillips Pan Head
REF 0 B 70-08315 Schematic, Linear IV Driver Board
REF 0 B 70-08316 Assembly, Linear IV Driver Board

SHEET 3 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08318


Rev. B ECN: P0914

QTY DWG PART ITEM DESCRIPTION Vendor Ref. Design


ITEM SIZE NO. CLS Part No.
.Ref 0 70-08317 Schematic, Display PCB
0001 1 70-06305 Linear IV Display PCB Fabrication
0002 1 70-04733 Right Angle Header J1
0003 1 70-03353 Timer 555 pin dip LM555 U1
0004 1 70-03136 Quad 2 – Input Nand gate, 14 pin dip 74LS00 U2
0005 1 70-03087 8 Channel Source Buffer, 18 pin dip UDN2585 U3
0006 1 70-03086 8 Channel Current Sink Buffer, 18 pin dip UDN2595 U4
0007 4 70-03088 5 x 7 Dot Matrix: Alpha Numeric Display 2416 DS1-4
0008 8 70-03089 .56” Seven Segment Display HDSP-5601 DS5-12
0009 1 70-03064 Green LED Light Bar 2855 DS13
R1
0010 1 70-00064 Resistor, Carbon Film, 10 K ohms, + 5%
0011 1 70-00089 Resistor, Carbon Film, 150 K ohms, + 5% R2
0012 2 70-00042 Resistor, Carbon Film, 1 K ohms, + 5% R3, 5
0013 1 70-00086 Resistor, Carbon Film, 100 K ohms, + 5% R4
0014 8 70-00324 Resistor, Carbon Film 10, + 5% R6-13
0015 8 70-02010 Capacitor Ceramic, .1 uF, 50V C2-7, 10, 11
0016 1 70-02033 Capacitor Ceramic, 150 pF, .50V C8
0017 1 70-02511 Capacitor Tant, 10 uF, 35V C1
0018 1 70-03071 Transorb MPTE-5 CR1
0019 2 70-03001 Diode 1N914 CR2, 3
0023 4 70-04754 Socket Sip 10 pin DS5-8
0025 1 70-04501 Socket 8 pin DS13
0026 1 70-00014 Resistor Carbon Film, 68 ohm, + 5% R14
SHEET 1 OF 2
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08318


Rev. B ECN: P0914

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref. Design


NO. SIZE NO. CLS Part No.
0027 1 70-04220 Socket, 40 pin machined DS9
0028 8 70-04586 Socket, Terminal Strip DS1-4

SHEET 2 OF 2
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08320


Rev. B ECN: P0929

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref. Design


NO. SIZE NO. CLS Part No.
.Ref 0 70-08319 Schematic, Master Board
0001 1 70-06306 Master PCB Fabrication
0002 2 70-03201 Transistor – Arrays ULN20004U1, 5
0003 1 70-03041 Opto-coupler MCT2 U2
0004 2 70-03119 Hex Buffer 7407 U3, 15
0005 1 70-03158 Encoder 8 to 3 line 74LS148 U4
0006 1 70-03138 Hex Inverter 7406 U6
0007 1 70-03159 Decoder 2 to 4 line 74139 U7
0008 1 70-03087 Source Driver UDN2585 U8
0009 4 70-03351 Dual Op Amp LM358 U9, 12-14
0010 1 70-03352 Dual Comparator LM393 U10
0011 1 70-03355 Dual Op Amp LF442 U11
0012 1 70-00056 4.7 K, 5% Resistor R1
0013 7 70-00064 10 K, 5% Resistor R2-4, 21, 34, 43, 54
0014 1 70-00220 390 ohms, ½ w Resistor R6
0015 5 70-00018 100 ohms, 5% Resistor R7, 8, 10, 52, 63
0016 1 70-00045 1.5 K, 5% Resistor R9
R11, 23, 24, 29, 33, 64, 65
0017 7 70-00042 1K, 5% Resistor
0018 9 70-00010 47 ohms, 5% Resistor R12-20
0019 3 70-00048 2.2 K, 5% Resistor R22, 42, 53
0020 3 70-00314 4.99 K, 5% Resistor R25, 30, 31
0021 2 70-00700- 2.00 K, MF Resistor R26, 28
0022 1 70-00108 1M, 5% Resistor R27
SHEET 1 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08320


Rev. B ECN: P0929

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref. Design


NO. SIZE NO. CLS Part No.
0023 14 70-00700-7 100 K, MF Resistor R32, 35-41, 47, 49, 50,
56, 60, 61
0024 2 70-00700-11 237 MF Resistor R44, 55
0025 6 70-00700-6 33.2 K, MF Resistor R45, 48, 51, 57, 58, 62

0026 2 70-00700-5 10.0 K, MF Resistor R46, 59


0027 1 70-00515 1K, 8 pin sip Bussed RP1
0028 1 70-00517 390 ohms, 8 pin sip Bussed RP2
0029 1 70-00516 10 K, 8 pin sip Bussed RP3
0030 1 70-02516 Elec Cap 2200 uF C1
0031 1 70-02510 Tant Cap 100 uF C3
0032 3 70-02511 Tant Cap 10 uF C4, 5, 15
0033 1 70-02523 Elec Cap 47 uF C6
0034 18 70-02010 Cer Cap .1 uF C7-13, 16-23, 26-28
0035 1 70-02514 Tant Cap 4.7 uF C14
0036 2 70-02509 Tant Cap 2.2 uf C24, 25
0037 2 70-03032 Bridge Rect. CR1, 9
0038 5 70-03012 Diode IN4002 CR2-5, 15
0039 1 70-03071 5V Rect. MPTE5 CR6
0040 2 70-03069 15V Rect. MPTE15 CR7, 8
0041 3 70-03301 LM113 CR10, 16, 18
0042 4 70-03001 Diode 1N4148 CR11-14
0043 2 70-03021 Zener IN750 CR17, 19
0044 2 70-04016 Relay Type 603 K1, 4
0045 2 70-04015 Relay Type 683 K2, 3
SHEET 2 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08320


Rev. B ECN: P0929

ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref. Design


NO. SIZE NO. CLS Part No.
0046 1 70-04001 Relay 24v 4PDT K5
0047 1 70-04002 Relay Socket K5
0048 1 70-04003 Retaining Clip K5
0049 1 70-04605 Fuse 3A F1
0050 2 70-04635 Fuse Holder F2
0051 1 70-03402 Trans. 2N2222 Q1
0052 2 70-03401 Trans. 2N2905 Q2, 3
0053 1 70-04521 22 Position Header TS5
0054 5 70-04535 15 Position Header TS1-4, 6
0055 1 70-04748 6 Pin Header J1
0056 1 70-04506 6 Pin Socket U2
0057 4 70-04217 16 Pin Socket U1, 4, 5, 7
0058 1 70-04281 18 Pin Socket U8
0059 3 70-04216 14 Pin Socket U3, 6, 15
0060 6 70-04222 8 Pin Socket U9-14
0061 2 70-04758 2 Pin Header W1, 2
0062 2 70-04806 Two Circuit Shunt W1, 2
0063 4 70-04595 Jack, Test TP1-4
0064 4 70-04779 Spacer, Insulating .250 Dia.
0065 2 70-04780 Spacer, Insulating .375 Dia.

SHEET 3 OF 3
Eureka Progeny Inc.• 1407 Barclay Blvd. •Buffalo Grove, IL 60089

BILL OF MATERIAL Bill of Material No. 70-08355


Rev. B
ITEM QTY DWG PART ITEM DESCRIPTION Vendor Ref.
NO. SIZE NO. CLS Part No. Designator
.Ref 0 B 70-08354 Schematic, Pausch URO Interface PCB
0001 2 A 70-02524 Capacitor, Aluminum 63 VDC 100 uF C1, C2
0002 1 A 70-02511 Capacitor, Tantalum 35VDC 10% 10 mF C3
0003 4 A 70-03012 Rectifier, 1 Amp 1N4006 D1 - D4
0004 1 A 70-04521 Header, 22 Position Terminal Strip J1
0005 2 B 70-04014 Relay, DPST Miniature Sealed, 12 VDC DS2E-M-DC2 K1, K4
0006 2 A 70-04006 Relay, 4PDT 12 VDC DS4E K2, K3
0007 4 B 70-00654-3 Potentiometer, 25 Turn 5K R1 - R4
0008 1 A 70-04201 Dip Switch, 4 SPST CST-206-4 SW1
0010 1 A 70-04778 Heat Sink U1
0011 1 B 70-06315 Fabrication, Pausch URO Interface PCB
0012 1 A 70-03305 12V 3 Term. Pos. LM340-12 7812

SHEET 1 OF 1
REVISIONS

REV DESCRIPTION DATE APP'D

SW1 CR1 SW5 CR5 A RELEASED PER PRN4080


MAG_SIZE 1N914 CROSS_OPEN 1N914
SW2 CR2 SW6 CR6
LAST_SIZE 1N914 CROSS_CLOSE 1N914
SW3 CR3 SW7 CR7 SW9 CR9
PRESET_1 1N914 LONG_OPEN 1N914 CAL 1N914
SW4 CR4 SW8 CR8 SW10 CR10
PRESET_2 1N914 LONG_CLOSE 1N914 ENTER/SET 1N914
20
KEYBD_ROW_0 J1-20
19
KEYBD_ROW_1 J1-19
18
KEYBD_ROW_2 J1-18
17
KEYBD_SCAN_0 J1-17
16
KEYBD_SCAN_1 J1-16
15
KEYBD_SCAN_2 J1-15
14
KEYBD_SCAN_3 J1-14
SW11 CR11
1
LAMP' J1-1
LAMP
1N914
3
LAMP_RTN J1-3
LAMP_ON
2
LAMP_ON J1-2
D1 180 5%

R1
LAMP_ON

D7 180 5%
4
LAMP_ON' J1-4
13
DSP_DIGIT_1 J1-13
12
DSP_DIGIT_8 J1-12

D2 D3 D4 D5 D6
MEM2 MEM1 LAST_SIZE MAG_SIZE "X" NOTES

1) LINEAR IV FRONT PANEL ASSEMBLY 70-08153

R2 R3 R4 R5 R6
10 10 10 10 10
5% 5% 5% 5% 5%

11
DSP_SEG_A J1-11
10
DSP_SEG_B J1-10
9
DSP_SEG_C J1-9
8
DSP_SEG_D J1-8
7
DSP_SEG_H J1-7

ORIGINAL PRODUCT

PROGENY INCORPORATED
APPROVALS DATE

DRN
KT 06/28/93
LINEAR IV
THIS DRAWING IS THE PROPERTY OF AND
CHK
FRONT PANEL SCHEMATIC
CONTAINS INFORMATION PROPRIETARY TO APP SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR B 70-08152
DISCLOSED WITHOUT THE PRIOR WRITTEN
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME: FPANEL.DSN SHEET 1 OF 1
REVISIONS

RP2

RP3

RP4
+5V +5V +5V +5V REV DESCRIPTION DATE APP'D
1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10
CPU RESET U2 A RELEASED PRN4080 (SAME AS REV 3) 10/13/93 K. Torchalski

DATA0
DATA1
DATA2
DATA3
DATA4
DATA5
DATA6
DATA7
ADDR12
ADDR13
ADDR10
ADDR11

ADDR14
ADDR15
ADDR3
ADDR4

ADDR7
ADDR8
ADDR1

ADDR9
1 8

ADDR0

ADDR2

ADDR5
ADDR6
PBRST VCC

SW1
B C42-45,50,51 Removed, ECN3444 1/6/94 K. Torchalski
2 TD ST 7

3 TOL RST
5 C Program with ver 2.02 EPROM A. Krema

4 GND RST
6 D Add C54 - C57 9/6/02 A. Krema
DS1232 +5V
DGND U5 U6
ST' ADDR14 1 19 DATA7
1 20 A14 D7
G1 VCC ADDR1326 18 DATA6
19 10 A13 D6
G2 GND ADDR12 2 A12 D5
17 DATA5
DGND 2 18 ADDR15 DGND ADDR1123 16 DATA4
1A1 1Y1 A11 D4
4 16 ADDR14 ADDR1021 15 DATA3 22
6
1A2 1Y2
14 ADDR13 ADDR9 24
A10 D3
13 DATA2 21
J5-22 ADDR3

8
1A3 1Y3
12 ADDR12 ADDR8 25
A9 D2
12 DATA1 19
J5-21 ADDR2

11
1A4 1Y4
9 ADDR11 ADDR7 3
A8 D1
11 DATA0 20
J5-19 ADDR1

13
2A1 2Y1
7 ADDR10 ADDR6 4
A7 D0 J5-20 ADDR0
2A2 2Y2 A6
15 5 ADDR9 ADDR5 5 ADDR0 - ADDR15
2A3 2Y3 A5
17 2A4 2Y4
3 ADDR8 ADDR4 6 A4
74LS244 ADDR3 7 A3 RST
+5V +5V ADDR2 8 A2 +5V
ADDR1 9 13
U1 U4 ADDR0 10
A1 J3-13 RST'
A0
34 VCC A19
35 1 G1 VCC
20 RST'
19 G2 GND
10 VCC
28
32 DGND DGND ADDR1520
A17 2 18 ADDR7 CE
DATA0 - DATA7
1 31 1A1 1Y1 22
VSS A16 4 16 ADDR6 OE
2 30 1A2 1Y2 27 14
VSS A15 6 14 ADDR5 WE GND
18 29 1A3 1Y3 29
20pF 20pF 36
VSS A14
28
8 1A4 1Y4
12 ADDR4 28C256 28
J5-29 DATA6
C39 MET Y1 MET C38 VSS A13
27
11 2A1 2Y1
9 ADDR3 DGND
27
J5-28 DATA5
A12
26
13 2A2 2Y2
7 ADDR2 U8 26
J5-27 DATA4
DGND
3
A11
25
15 2A3 2Y3
5 ADDR1
ADDR12 2 25
J5-26 DATA3
XTAL A10
24
17 2A4 2Y4
3 ADDR0
ADDR1123
A12
24
J5-25 DATA2
A9
23 74LS244 ADDR1021
A11
19 DATA7 23
J5-24 DATA1
10MHz
A8
22 ADDR9 24
A10 D7
18 DATA6
J5-23 DATA0
A7 A9 D6
4 EXTAL A6
21 +5V ADDR8 25 A8 D5
17 DATA5
20 ADDR7 3 16 DATA4
A5
19 U3 ADDR6 4
A7 D4
15 DATA3
NOTES
A4 A6 D3
17 ADDR5 5 13 DATA2
A3 19 20 A5 D2
16 G VCC ADDR4 6 12 DATA1
A2
15
1 DIR GND
10
ADDR3 7
A4 D1
11 DATA0
1. 70-06303 LINEAR IV CPU PCB
A1 A3 D0
8 RESET A0
14 ADDR2 8 A2 70-08314 LINEAR IV CPU ASSEMBLY
DGND ADDR1 9
+5V DGND A1 +5V
RP1

D7
44 2 A1 B1
18 DATA7 ADDR0 10 A0
D6
43 3 A2 B2
17 DATA6
52 TEST D5
42 4 A3 B3
16 DATA5 ADDR1526 CS2 VCC
28
1 2 3 4 5 6 7 8 9 10 13 41 5 15 DATA4
ST D4 A4 B4
6 BUSACK D3
40 6 A5 B5
14 DATA3 22 OE
7 BUSREQ D2
39 7 A6 B6
13 DATA2 27 WE GND
14
45 RTS0 D1
38 8 A7 B7
12 DATA1 20 CS1
46 CTS0 D0
37 9 A8 B8
11 DATA0 DGND
6264
47 DCD0 74LS245 DGND
50 CKA0 0
68 CLK_OUT
51 TXA1 MREQ
63
53 66
RXA1 WR 9 12
54 67
CKA1 RD 8 11
55 TXS IORQ
62
10
U7B 13
U7C I/O_WR'
56 +28V
RXS 74LS32 74LS32 18
57 CKS A18/TOUT
33 +5V J5-18 I/O_WR'
58 DREQ1 R3 4
59 6
60
TEND1
10 5
U7A
HALT INT0 4.99K MF_1%

14
61 RFSH INT1
11 74LS32
5 WAIT INT2
12 R4 1
64 E 3
65 MI RXA0
49 1K 5% 2
U7D I/O_RD'
9 NMI TXA0
48 74LS32
PLCC R5
Z80180

7
ADINT'
+5V
KEYINT' 20K 5%
14 DGND DGND
+5V J2-14
15
+5V J2-15 ORIGINAL PRODUCT - LINEAR IV

+5V J2-16 16
C1
0.1
C3
0.1
C5
0.1
C7
0.1
PROGENY INCORPORATED
17
+5V J2-17 CER CER CER CER
18
CR1 C37 APPROVALS DATE
J2-18
LINEAR IV CPU BOARD
+5V
+

10
10 DRN
DGND J2-10 MPT-5 TANT
C2 C4 C6 C8
K. Torchalski 06/28/93
DGND J2-11 11 0.1
CER
0.1
CER
0.1
CER
0.1
CER
THIS DRAWING IS THE PROPERTY OF AND CHK
K. Torchalski 10/13/93
CORE SECTION
12 CONTAINS INFORMATION PROPRIETARY TO
DGND J2-12 APP SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING K. Torchalski 10/13/93
13
J2-13
B 70-08313
DGND
SHALL NOT BE REPRODUCED, USED OR
19
DGND J2-19 DISCLOSED WITHOUT THE PRIOR WRITTEN
DGND
CONSENT OF PROGENY INCORPORATED.
SCALE FILENAME:CPU.DSN SHEET 1 OF 4
+5V

RP5
1
2
3
4
5
6
20
J4-20 K_ROW0_FP
1
J1-1 K_ROW0_R
19
J4-19 K_ROW1_FP
2
J1-2 K_ROW1_R
18
J4-18 K_ROW2_FP
3
+5V J1-3 K_ROW2_R
+5V
U16
U10 16 VCC Y0
15 17
J4-17 K_SCAN0_FP
14
40 38 Y1
VCC RL0 8 13 16
RL1
39 GND Y2
12
J4-16 K_SCAN1_FP
1 Y3
RL2 11 15
20 VSS RL3
2
1
Y4
10
J4-15 K_SCAN2_FP
5 A Y5
RL4 2 9 14
RL5
6
3
B Y6
7
J4-14 K_SCAN3_FP
DATA0 - DATA7 DGND RL6
7 C Y7
8 3 4 4
RL7 6 G1 J1-4 K_SCAN0_R
+5V 4 7407
35 G2A
DATA7 19
SL3
34
5 G2B U23D 5 6 5
DATA6 18
DB7 SL2
33 74LS138 J1-5 K_SCAN1_R
14 DB6 SL1 7407
U13A DATA5 17 DB5 SL0
32
U23C
DATA4 16 DGND
DB4 9 8 6
KEYINT' 2 1 DATA3 15 DB3 SHIFT
36 J1-6 K_SCAN2_R
DATA2 14 37 7407
7406
DATA1 13
DB2 CNTL/STB
23
+5V
U23B U23A
DB1 BD 11 10 7
7 DATA0 12 DB0 U9 J1-7 K_SCAN3_R
OUTA3
24 7407
DGND 4 25 24 1 16
10
IRQ OUTA2
26
VCC O0
2 15
J5-16 DSP_DIG0_DB
I/O_RD'
11
RD OUTA1
27
O1
3 14
J5-15 DSP_DIG1_DB
I/O_WR'
3
WR OUTA0
28
O2
4 13
J5-14 DSP_DIG2_DB
+5V
ADDR0 21
CLK OUTB3
29 20
O3
5 12
J5-13 DSP_DIG3_DB

9
A0 OUTB2
30 21
D O4
6 11
J5-12 DSP_DIG4_DB

22
RST OUTB1
31 22
C O5
7 10
J5-11 DSP_DIG5_DB
CS OUTB0
23
B O6
8 9
J5-10 DSP_DIG6_DB
14 7 8279A-5
A O7
9 1 18 12
J5-9 DSP_DIG7_DB
O8
10 2
UDN2585A
17 13
J4-12 DSP_DIG8_FP
VCC GND DGND O9
11 3
U17 16
J4-13 DSP_DIG1_FP
2 5 O10
+5V
D Q 3 4 19 G2 O11
13 13 12 4 15
18 14 5 14
U11B
74LS74
7406 G1 O12
15
7407 +5V
6 13
U13F O13
16
U26F 7 12
CLK_OUT 3 CK Q 6 O14
12 GND O15
17 14 8 11
CL PRE
1 7406 2 9 VS GND
10
74LS154 3 4
5
U12 6
1 4 DGND DGND 8
9 8
9
J1-8 DSP_DIG9_R
11 10
10
J1-9 DSP_SEGA_R
13 GND
12
11
J1-10 DSP_SEGB_R
7
J1-11 DSP_SEGC_R
ADDR0 - ADDR15 7 J5-7 DSP_SEGB_DB
8
J5-8 DSP_SEGA_DB
RST U13C DGND 1 UDN2595A
18 11
J4-11 DSP_SEGA_FP
5 6 2 17 10
CS6' 3
U18 16 9
J4-10 DSP_SEGB_FP
7406 U13D 4 15 8
J4-9 DSP_SEGC_FP
9 8
5 14 7
J4-8 DSP_SEGD_FP
U13E +5V 6 13
J4-7 DSP_SEGH_FP
7406
11 10
7 12
7406 U13B 8 11
13 12 9 VS GND
10
14
7406
1 2 DGND 6
3
7406
4 5
J5-6 DSP_SEGC_DB

5
U15 6 4
J5-5 DSP_SEGD_DB
9 8 3
J5-4 DSP_SEGE_DB
11 10 2
J5-3 DSP_SEGF_DB
13 12 1
J5-2 DSP_SEGG_DB
+5V GND J5-1 DSP_SEGH_DB
7

DGND

C9 C10 C11 C12 C13 C14 C15 C16 C17 C18 PROGENY INCORPORATED
0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1
CER CER CER CER CER CER CER CER CER CER

LINEAR IV CPU BOARD


DGND
THIS DRAWING IS THE PROPERTY OF AND DISPLAY/KEYBOARD SECTION
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
DISCLOSED WITHOUT THE PRIOR WRITTEN
B 70-08313
CONSENT OF PROGENY INCORPORATED
SCALE FILENAME: CPU.DSN SHEET 2 OF 4
+5V

1
10K 2%

10K 2%

10K 2%

10K 2%

10K 2%

10K 2%

10K 2%

RP10
+5V

RP9
+5V

RP8
20 11 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10

RP7-3

RP7-4

RP7-5

RP7-6

RP7-7

RP7-8

RP7-9
VCC LE

10
10 1 1 2 3 4 5 6 7 8 9 10

9
9 GND OC
TO DRIVER BOARD - +5V J3-9
33 14
J5-33
+5V DGND
DATA0 2 3 2 1 18
TO DISPLAY BOARD -
+5V J5-34 34
DATA1 5
Q1 D1
4 4
7407
3 17
J8-18 SID_INFO_A
Q2 D2 U22 J8-17 SID_INFO_B

F000H
LAMP_RTN J1-13 13 DATA2 6 7 6 5 16
DATA3 9
Q3 D3
8 8 9 15
J8-16 SID_INFO_C
LAMP_ON_RTNJ1-15 15
DATA4 12
Q4 D4
13 10 11 8
J8-15 VERT_IN
3 DATA5 15
Q5 D5
14 12 13 4
J8-8 RAD /FL'
LAMP_RTN J4-3 Q6 D6 GND J8-4 ST/TOMO
DATA6 16 17
LAMP_ON_RTNJ4-4 4 Q7 D7 7
DATA7 19 Q8 D8
18 U23F
TO DRIVER BOARD - DGND J3-10 10 74HC373

31
U21 DGND 12 13 7
J8-7 IR_TRUE
DGND J5-31
7407
TO DISPLAY BOARD _
DGND J5-32 32 +5V W1 14
U23E
DGND 2 1 9
DGND U19 J8-9 SF_TRUE
7407
26 VCC PA7
37
DATA0 - DATA7 38 7 14
PA6
DATA7 27 39 2 1 13
DATA6 28
D7 PA5
40 4
7406
3 14
J8-13 T=+90'
DATA5 29
D6 PA4
1 6
U24 5 12
J8-14 T=-90'
DATA4 30
D5 PA3
2 8 9 1
J8-12 T=0'
DATA3 31
D4 PA2
3 10 11 2
J6-1 C=+90'
DATA2 32
D3 PA1
4 12 13 3
J6-2 C=-90'
DATA1 33
D2 PA0 GND J6-3 C=0'
D1 10 1 18 7 4
DATA0 34 D0
PC7
11 2
UDN2595A
17
J6-4 TILT_C_COM
PC6
12 3
U25 16
PC5
I/O_RD' 5 RD PC4
13 4 15 DGND
5 14 DGND
17 19
I/O_WR' 36 WR
PC3
16 +5V 6 13
20
J8-19 TILT_INFO_B
PC2
15
7 12
3
J8-20 TILT_INFO_A
6 CS
PC1
14
8 11
2
J8-3 MAG1_RLY
PC0 9 VS GND
10
1
J8-2 MAG2_RLY
PB7
25 J8-1 EXP_HLD_RLY
ADDR0 - ADDR15 ADDR1 8 24
A1 PB6 DGND
PB5
23 ST'
ADDR0 9 22 14

RP11
A0 PB4
21 +5V
PB3
RST 35 20
RST PB2 1 2
PB1
19

1
2
3
4
5
6
7 GND PB0
18 7407

7
U26E
8255A-5 18
17
J3-18 CRS_ENABLE
TRIAC_0VAC J7-1 1 DGND
DGND
16
J3-17 CRS_DIR
2
U26D 15
J3-16 LNG_ENABLE
TRIAC_0VAC J7-2 3 4
14
J3-15 LNG_DIR
5 7407 U26C J3-14 AUTO/MAN
TRIAC_GATEJ7-5
5 6
TRIAC_GATEJ7-6 6
+5V 7407 U26B
GATEJ3-12 12 9 8

11
7407 U26A
0VAC J3-11 U20 11 10
16 VCC Y0
15 7407
14 17
LAMP_ON J3-20 20
8
Y1
13
J5-17 CS1'
GND Y2 CS2'
12 CS3'
19 Y3
LAMP_SW' J3-19 DGND Y4
11 CS4'
ADDR12 1 10 CS5'
1 A Y5
LAMP' J4-1 ADDR13 2 B Y6
9 CS6'
ADDR14 3 7
12 C Y7
LAMP_SW' J1-12
ADDR15 6
LAMP_ON J1-14 14 G1
4 G2A
RP6

5
LAMP_ON J4-2 2 G2B
74LS138
1
2
3
4
5
6

DGND

+5V

PROGENY INCORPORATED
DGND

C19
0.1
C20
0.1
C21
0.1
C22
0.1
C23
0.1
C24
0.1
C25
0.1
C26
0.1
LINEAR IV CPU BOARD
CER CER CER CER CER CER CER CER
THIS DRAWING IS THE PROPERTY OF AND INPUT/OUTPUT SECTION
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
DGND PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
DISCLOSED WITHOUT THE PRIOR WRITTEN
B 70-08313
CONSENT OF PROGENY INCORPORATED.
SCALE FILENAME: CPU.DSN SHEET 3 OF 4
+28V
1
+28V J2-1 1
J3-1 +28V
2
+28V J2-2 2
J3-2 +28V
3
+28V J2-3 3
J3-3 +28V
4
+28V J2-4 4
J3-4 +28V
5 5
PGND J2-5 CR3 J3-5 DGND
6 6
PGND J2-6 +5V J3-6 DGND
7 7
PGND J2-7 +5V
1N4004 J3-7 DGND
8 8
PGND J2-8 PGND VR1 J3-8 DGND
LM317T
1 2 7
AGND J2-9 9 VIN VOUT J6-7 VCP (LONG)

1
ADJ
12
RP7-1 RP7-2 C34 3 J6-12 VCP (CROSS)
10K 10K CR2 R1 C36

+
AGND 0.1
237 10
+5V CER 1N4004 10
MF_1% TANT J6-10 VCP_RET (CROSS)

3
U30 J6-8 VCP_RET(LONG)
22 ALE VA+
1
30 12 AGND
BW VD+
C35 AGND

+
C30 C31 C32 C33

+
18
17
D15
0.1 10 0.1 10 R2 10
16
D14
CER TANT CER TANT C57 C56 237
TANT
D13
15 33 0.1 0.01 MF_1%
D12 GND
14 CER CER
13
D11
42
C40 C41 AGND
D10 VREF-
11 43 100 AGND 0.1
D9 VREF+

+
DATA0 - DATA7 10 44 TANT CER
DATA7 9
D8 VREFOUT AGND R6 AGND
D7 11
DATA6 8 D6 IN7
41
R7 J8-11 VSID2
DATA5 7 D5 C46 10K 5%
6
DATA4 6 D4
+ J8-6 VIRC
DATA3 5 40 DGND 10 TANT
1K 5%
DATA2 4
D3 IN6
U29B
D2 6 AGND
DATA1 3
D1
- +5V C47
DATA0 2 39 7 +
D0 IN5
U27B
5 6 2.2 TANT
ADDR0 - ADDR4 + -
LM358 AGND
ADDR428 A4 IN4
38 7 8
ADDR327 A3 5
U27A 2
ADDR226 A2
+ -
LM358
ADDR125 37 1
ADDR024
A1 IN3 R8
A0 3 5
LM358A
+ J8-5 VIRL
19 36 1K 5%
I/O_RD' RD IN2 4
I/O_WR' 20 WR U28B C48 +
CS5' 21 CS
6
- 2.2 TANT
ADINT' 31 35 AGND
INT IN1 7 AGND
32 DMARQ
29 +5V 5 11
23
SYNC
34
C51 LM358
+ J6-11 VXFC
CLK_OUT CLK IN0 C52

+
LM12458
OPEN 10
8
C42 U28A
TANT
2
-
OPEN 1 AGND AGND
RP12-1 +
3 8
J6-8 VXFL
1 2 LM358A
C50
10K 4 C53

+
C43 OPEN
+5V
10 C49
AGND TANT +
OPEN
10 TANT
AGND 8
RP12-2 U29A AGND
AGND
1 3 2
-
10K 1
C44 R9
3 10
LM358A
+ J8-10 VSID
10K 5%
OPEN 4 C54 C55
RP12-3 0.1 0.01

PROGENY INCORPORATED
+5V 1 4 AGND CER CER
10K
C45
AGND
OPEN
C27 C28 C29
0.1 0.1 0.1 1
RP12-4
5
LINEAR IV CPU BOARD
CER CER CER
10K
THIS DRAWING IS THE PROPERTY OF AND ANALOG/POWER SECTION
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
AGND PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
DISCLOSED WITHOUT THE PRIOR WRITTEN
B 70-08313
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME: CPU.DSN SHEET 4 OF 4
REVISIONS

REV DESCRIPTION DATE APP'D

A RELEASED PRN4080 (SAME AS REV 3) 10/13/93 K Torchalski

B ECR 3878 - RP2 was 10K 6/8/94 K Torchalski

C ECR P0002 - Enable, Lamp

+5V +5V D ADDED ISOLATION TO DRIVE 5/24/96 C. HARDY


+5V

+5V +5V
R15 C13 C15
51K
5%
0.1 0.1
CER CER
DGND DGND

C2 R2
2.2M
0.1

14
5%

4
12 CER
R VCC 11

14
6 S Q
1 13
19 1 4 8
J1-19 3 3
U8B 74LS00 9
CP 4013
2 U6A DGND 8
5 2 4 RESET VCC 10
D Q 7
74LS00 6 2 DISCH
C14 C9 TRIG

7
U6D GND 5 74LS00
+

0.01 10 U6B
U1

7
CER TANT DGND 74LS00 C8

+
U6C LM555 10
DGND
TANT
DGND

THRESH 6 +5V
3 OUT
DGND CTL 5 DGND
GND
1
C7
0.01

10
CER
R2
8 13
20 S2 Q2
J1-20 U8A
11 CP2
DGND DGND 4013

9 D2 Q2
12
R12
330
11
J1-11 5%

1
4 1 NOTES
+
C12 1. 70-06304 LINEAR IV DRIVER FABRICATION
12
R1 4 3 0.01 U5 70-08316 LINEAR IV DRIVER ASSEMBLY
J1-12 DF02M
CER
100 5% CR2
MOC3011
-
6 2
2
+5V

DGND

9
J1-9
ORIGINAL PRODUCT - LINEAR IV
J1-10 10
PROGENY INCORPORATED
APPROVALS DATE
DGND
DRN
K Torchalski 06/28/93
LINEAR IV DRIVER BOARD
THIS DRAWING IS THE PROPERTY OF AND CHK
K Torchalski 10/13/93 LAMP SWITCH SECTION
CONTAINS INFORMATION PROPRIETARY TO
APP SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING K Torchalski 10/13/93
SHALL NOT BE REPRODUCED, USED OR
B 70-08315
DISCLOSED WITHOUT THE PRIOR WRITTEN
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME:DRIVER.DSN SHEET 1 OF 2
+28V +5V_ISOL +5V_ISOL
1
J1-1 +5V_ISOL
+5V_ISOL +28V

J1-2 2 U12 NOTE: W1 AND W2 DETERMINE CHOPPER


R16 7805
ACTION, DO NOT JUMPER BOTH
R10
3 1 3
J1-3 VIN VOUT
47 5W

1
GND
4 22K 5% C24

MUR105

MUR105

MUR105

MUR105
J1-4 C26 C35 C33 R17 C25 C23

2%

2%

CR12

CR13

CR14

CR15
RP1-5

RP4-1

RP4-2
+

+
2
5 47 1 1 1M C22 W1 C21
0.1
0.1 100
J1-5 10k CER

7.5K

7.5K
ELEC CER CER 5% 0.1 0 CER ELEC
2%
6 CER
J1-6 3.3nF CER

12

16

4
2

6
J1-7 7
RP3-1 W2 VS OSC PGND VSS VS
PGND PGND
11 4 5 2 12
8
1 2 1 MCT62 7 CONTROL A IN1 O1 J2-12
J1-8 C27 680 C34 OPEN
1 6 7
PGND U13A + SYNC B IN2 CROSS PHASE 1
18 0.1 CER 2 8
CROSS ENABLE J1-18 10 3 HOME C
7 10 IN3 O2
3 11
J2-11
RP3-2 TANT
1 3 4 MCT62 6 18 9 12 13 10
PGND CLOCK D IN4 O3 J2-10
C28 680
U13B 10 5 6
ENABLE INH1 ENA CROSS PHASE 2
17 0.1 CER 3 5
CROSS DIR J1-17 17 8 11 14 9
+5V RP3-5 CW/ CCW INH2 ENB O4 J2-9
1 6 1 MCT62 7

5
+5V_ISOL 20 RESET

MUR105

MUR105

MUR105

MUR105
2%
2%
C31 680

CR16

CR17

CR18

CR19
RP2-2

RP2-4
U15A 15
RESET' J1-13 13 0.1 CER 2 8 RP4-3 VREF

1K
1K
4 1
RP3-3 19 HALF/ FULL
1 4 1 MCT62 7 7.5K 2% U11 ALT CROSS PHASE COM
8
L297 GND S2 S1 SA SB GND L298 J2-8

1
C29 680
U14A U10

13

14

15
+28V

8
LONG ENABLE J1-16 16 0.1 CER 2 8 RP4-4
5 1 PGND
RP3-4 R14 PGND
1 5 4 MCT62 6 7.5K 2%
7
C30 680
U14B R13 1.0 5% J2-7
PGND CROSS PHASE COMMON
15 0.1 CER 3 5
LONG DIR J1-15 1.0 5%

+5V_ISOL +5V_ISOL +5V_ISOL +5V_ISOL +28V


+5V_ISOL +5V_ISOL +5V_ISOL +5V_ISOL PGND
PGND

C17 C16
0.1
0.1 C4 C11
CER PGND PGND

MUR105

MUR105

MUR105

MUR105
CER C6 C5

CR3

CR4

CR5

CR6
+
R8 R3 0.1 CER 0.1 CER
0.1 100
4 8 470K 4 8 91K CER ELEC
5% 5%

12

16

4
RESET VCC RESET VCC
7 7 VS OSC
PGND
VSS VS
PGND PGND
2 TRIG
DISCH 2 TRIG
DISCH
11 4 5 2 6
AUTO CONTROL A IN1 O1 J2-6
1

MANUAL R4
2%

5%

2%

CLOCK
RP1-1

RP1-2

CLOCK 1 SYNC B
6 7 IN2 LONG PHASE 1
U9 U7 RATE 51K
R9

RATE
LM555 LM555 5% 3 7 10 3 5
J2-5
10K

51K

10K

HOME C IN3 O2

18 9 12 13 4
CLOCK D IN4 O3 J2-4
2

THRESH 6 THRESH 6
10 5 6 LONG PHASE 2
3 OUT 3 OUT
ENABLE INH1 ENA

CTL 5 CTL 5
17 8 11 14 3
GND GND CW/ CCW INH2 ENB O4 J2-3
1 C20 C19 1 C18 C10

MUR105

MUR105

MUR105

MUR105
CR10
2

4
20

CR7

CR8

CR9
RESET

2%
2%
0.01 0.01 0.01 0.01

RP2-1

RP2-3
CER CER CER CER 15 VREF

1K
1K
19 HALF/ FULL
PGND
PGND PGND PGND PGND PGND U4 ALT LONG PHASE COM 2
+5V_ISOL L297 GND S2 S1 SA SB GND L298 J2-2

1
+5V +5V_ISOL
+5V_ISOL U3 +28V

13

14

15
2

8
PGND
R7
PGND
C3
1

1
J2-1
7

RP1-3 0.1 R5 R6 1.0 5%


LONG PHASE COMMON
C32
RP3-6

10K CER PGND


4.7K
0.1 2% 1.0 5%
5%
680

PROGENY INCORPORATED
CER
4 DGND
PGND
14

6
4

U2C
1

5
1
3
74LS00
2
U2D
9
8 74LS00 R11 LINEAR IV DRIVER BOARD
U2B
7

10 510

U15B 12 74LS00
5%
THIS DRAWING IS THE PROPERTY OF AND MOTOR DRIVE SECTION
11 PGND
4 MCT62 6 13
U2A CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING

AUTO/MAN J1-14 14 3 5
74LS00

PGND
SHALL NOT BE REPRODUCED, USED OR
DISCLOSED WITHOUT THE PRIOR WRITTEN
B 70-08315
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME: DRIVER.DSN SHEET 2 OF 2
PGND
REVISIONS
+5V +5V
REV DESCRIPTION DATE APP'D

A RELEASED PRN4080 (SAME AS REV 2)


+5V J1-34 34 C6
33 CR1 C1
0.1 R3
+5V J1-33 CER

+
4 8 1K
MPT5 10
32 5%
DIGITAL GND J1-32 TANT

31 RESET VCC
DIGITAL GND J1-31 DGND
2 DISCH
7
TRIG

R1 U1
DGND
10K LM555
5%

29
R4 CR2 R5
DATA_6 J1-29 THRESH 6
3 OUT 100K 5% 1N914 1K 5%
28
DATA_5 J1-28 CTL 5
GND
27
DATA_4 J1-27 1
26
DATA_3 J1-26 C8 R2 CR3
25
DATA_2 J1-25 150pf 150K 5% 1N914
24 CER
DATA_1 J1-24 DGND

23
DATA_0 J1-23
+5V +5V +5V +5V

DATA0
DATA1
DATA2
DATA3
DATA4
DATA5
DATA6

DATA0
DATA1
DATA2
DATA3
DATA4
DATA5
DATA6

DATA0
DATA1
DATA2
DATA3
DATA4
DATA5
DATA6

DATA0
DATA1
DATA2
DATA3
DATA4
DATA5
DATA6
11 C2 C3 C4 C5
12
13
14
17
16
15
18

11
12
13
14
17
16
15
18

11
12
13
14
17
16
15
18

11
12
13
14
17
16
15
18
9

9
0.1 0.1 0.1 0.1

VCC

VCC
VCC

VCC
D0
D1

D4
D5
D6

D0

D3
D4
D5

D0
D1
D2

D5
D6

D2
D3
D4
D2
D3

D1
D2

D6

D3
D4

D0
D1

D5
D6
BL

BL

BL

BL
CER CER CER CER

DS2
DL2416

DL2416

DL2416
DL2416
DS1

DS3

DS4
LEFT LEFT MIDDLE RIGHT MIDDLE RIGHT

+5V +5V +5V +5V

GND

GND
GND

GND
CUE

CUE
CUE

CUE
CLR

CE2

CLR

CE1

CE1
CE2
CE1
CE2

CLR

CE1

CE2

CLR
WR

WR

WR
WR
CU

CU

CU

CU
A1
A0

A1
A0

A1
A0

A1
A0
10

10

10

10
5
3
4

7
8

1
2
6

5
3
4

7
8

1
2
6

5
3
4

7
8

1
2
6

5
3
4

7
8

1
2
6
DGND DGND DGND DGND DGND DGND DGND DGND
19
ADDR_1 J1-19
20
ADDR_0 J1-20
WR' J1-18 18 NOTES
21 4
ADDR_2 J1-21 6 12
1. 70-06305 LINEAR IV DISPLAY FABRICATION
5
U2C 11 70-08318 LINEAR IV DISPLAY ASSEMBLY
13
U2A
74LS00
74LS00

9
8
10
U2B

74LS00
22
ADDR_3 J1-22

+5V

C7 ORIGINAL PRODUCT - LINEAR IV

0.1 PROGENY INCORPORATED


14

CER
1
17 3
DGND APPROVALS DATE
J1-17 U2D
LINEAR IV DISPLAY BOARD
CS'
2
DRN
74LS00
KT 01/05/93
ALPHA SECTION
7

THIS DRAWING IS THE PROPERTY OF AND CHK


CONTAINS INFORMATION PROPRIETARY TO
DGND APP SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
B 70-08317
DISCLOSED WITHOUT THE PRIOR WRITTEN
CONSENT OFPROGENY INCORPORATED. SCALE FILE NAME:DISPLAY.DSN SHEET 1 OF 2
3 7 7 3 3 7 7 3

A1

A1
A1

A1
A

A
A

A
8 6 6 8 8 6 6 8

A2

A2
A2

A2
B

B
B

B
4 4 4 4

C
C

C
2 2 2 2

D
D

D
+5V
16
J1-16
15
J1-15 1 1 1 1

E
E

E
14
J1-14 1 18
13 UDN2585A
J1-13 2
U3 17
12 3 16 9 9 9 9

CASSETTE RIGHT MIDDLE


J1-12 R14

F
F

CASSETTE LEFT MIDDLE


4 15

SEVEN SEGMENT DISPLAY

SEVEN SEGMENT DISPLAY


SEVEN SEGMENT DISPLAY

SEVEN SEGMENT DISPLAY


11
J1-11 5 14 68
10 6 13 5%

SID LEFT MIDDLE


SID RIGHT MIDDLE
J1-10 7 12 10 10 10 10

G
G

G
J1-9 9 8 11 DS13
+5V 9 10
VS GND 2 1
5 5 5 5

DP

DP
DGND

DP

DP
3 4

C10 DS11 DS10 DS7 DS6 6 5


0.1 3 7 7 3 3 7 7 3

A1

A1
A1

A1
7 8

A
A

A
CER

DGND 8 6 6 8 8 6 6 8
A2

A2
A2

A2
B

B
B

B
DGND

4 4 4 4

C
C

C
2 2 2 2

D
D

D
1 1 1 1

E
E

E
9 9 9 9
F

F
F

F
SEVEN SEGMENT DISPLAY

SEVEN SEGMENT DISPLAY


SEVEN SEGMENT DISPLAY

SEVEN SEGMENT DISPLAY


CASSETTE RIGHT

CASSETTE LEFT
10 10 10 10
G

G
G

G
SID RIGHT

SID LEFT
8
J1-8 5 5 5 5
DP

DP
DP

DP
7
J1-7
J1-6 6
R6 DS12 DS9 DS8 DS5
1 18
5 UDN2595A
J1-5 2
U4 17 10 5% R7
4 3 16
J1-4 4 15
J1-3 3
5 14
R8 10 5%

2 6 13
J1-2 7 12
10 5% R9
1 8 11
J1-1 +5V 9 10
R10 10 5%
VS GND
10 5% R11
R12 10 5%

C11
0.1
10 5% R13 PROGENY INCORPORATED
10 5%
CER
DGND

DGND LINEAR IV DISPLAY BOARD


THIS DRAWING IS THE PROPERTY OF AND NUMERIC SECTION
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
B 70-08317
DISCLOSED WITHOUT THE PRIOR WRITTEN
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME:DISPLAY.DSN SHEET 2 OF 2
REVISIONS
+28V
28V POWER SECTION REV DESCRIPTION DATE APP'D
F1 CR1
A RELEASED PRN4080 (SAME AS REV 3) 10/13/93 KT
12 10
19VAC TS1-12 TS1-10 28V OUT
B K2,K3 CHANGED PINOUT 2/19/95
FROM 3A AGC

TRANSFORMER
13
R1 C1

+
OVAC TS1-13 4.7K 2200uF
5% ELECT

+28V +28V

11
TS1-11 PGND OUT

1
K1
3
4 PGND
5 9
CR2 2 TS1-9
PBL OVERIDE SECTION 1N4002 TUBE SELECT SWITCH

8
4
TS1-8
TS5-4
11
KEY SW
+15V TS2-11 28V_SW

5
TS5-5 MOTOR
+28V
POWER

TO CPU
PGND R2 K2
10K 4
1 10
5% CR3 3 TS2-10 PGND
MAG RELAY CIRCUIT 1N4002 5

9
6
8
3 6 11 2
MAG1_RLY TS3-3 G
7 PGND
ULN2004A
+15V
U1C

8
PGND
+28V 3
TS5-3 MAG1
R3
10K K3
5% 4 2
1
3
TS5-2 MAG COM II-OUT
CR4 5
1N4002
FROM CPU

6
DIODE 8 1
MAG2_RLY TS3-2 2 1 16 2
7
TS5-1 MAG2

ULN2004A

U1B
+15V
3
NOTES:
+28V TS6-3
EXP_HLD_COM 1. DGND, AGND PGND TO BE TIED
4
TS6-4 TOGETHER AT 1 POINT.
1
K4
3
R4 4 5
2. 70-08320 ASSEMBLY.
EXPOSURE 10K CR5 5
TS6-5 EXP_HLD_NC
70-06306 PCB FABRICATION.
5% 2
HOLD CIRCUIT 1N4002
1
TS6-1
EXP_HLD_NO
1 2 15
EXP_HLD_RLY TS3-1 2
TS6-2
ULN2004A

U1A

ORIGINAL PRODUCT - LINEAR IV

PROGENY INCORPORATED
APPROVALS DATE

DRN KB, KT 4/5/93


LINEAR IV MASTER BOARD
THIS DRAWING IS THE PROPERTY OF AND
CHK KT 10/13/93 28V PWR. RELAY SECTION.
CONTAINS INFORMATION PROPRIETARY TO APP KT 10/13/93 SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR B 70-08319
DISCLOSED WITHOUT THE PRIOR WRITTEN
CONSENT OF PROGENY INCORPORATED. SCALE FILE NAME:MASTER.DSN SHEET 1 OF 6
27VAC
TRANSFORMER INTERCONNECT

15 15
27VAC TS1-15 TS2-15 27VAC

TRANSFORMER
14

FROM
TS2-14 27VAC TO LAMP CIRCUIT
OR
14 13
0VAC TS1-14 TS2-13 OVAC
AUX.CIRCUIT

12
TS2-12 OVAC
+5V
POWER SUPPLY INTERCONNECT.

2 6
J1-2 TS1-6 +5V OUT
+5V IN
3 8
J1-3 CR6 C3 TS2-8 +5V OUT

+
100uF
MPTE5 9
TANT TS2-9 DGND

4 5
DGND J1-4 TS1-5 DGND

+15V
DGND
FROM
LOGIC 1 3
SUPPLY
+15V IN J1-1 TS1-3 +15V OUT

CR7 C4

+
10uF
MPTE15 TANT

5 2
AGND J1-5 TS1-2 AGND

CR8 C5

+
AGND
10uF
MPTE15 TANT

6 1
-15V IN J1-6 TS1-1 -15V OUT

-15V
+15V
STEREO TOMO CIRCUIT
U2
R6 R7 R10 MCT2
8 1 5
24VAC TS6-8
390 .5W 100 5% 100 5%

C6 2 4

+
47uF 6
ELECT
ST/TOMO CIRCUITRY
TO GENERATOR

15VAC TS6-9 9 CR9

10
5VAC TS6-10

7
COM TS6-7

U1D
15 14 3
TO CPU ST/TOMO TS3-15
ULN2004A
PROGENY INCORPORATED
+15V
R9
1.5K
R8
TS5-6 6
5%
LINEAR IV MASTER BOARD
100 5%
S/T SWICH
THIS DRAWING IS THE PROPERTY OF AND POWER/STEREO TOMO CIRCUIT
7
TS5-7 CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING

AGND
DGND SHALL NOT BE REPRODUCED, USED OR
B 70-08319
DISCLOSED WITHOUT THE PRIOR WRITTEN
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME:MASTER.DSN SHEET 2 OF 6
+5V

COLLIMATOR TILT, TABLE TILT,


+5V
OUTPUT CIRCUIT,

1
+5V +5V
DISCRETE HSID CIRCUIT.

RP1-1

RP1-2

RP1-3

RP1-4

RP1-5
1K 1K 1K 1K 1K
R11 C9
2% 2% 2% 2% 2%
C7 C8 1K
5% 0.1uF CER

6
0.1uF CER
0.1uF CER DGND

9
U4

16
14
R12
DGND DGND 15 VCC 10 11 6 20
3 2 1
EO I0
11 G TS5-20 72" SID
SID INFO A TS2-3 14
I1
12 ULN2004A
47 5%
7407 GS I2
U3E I3
13 U5C U5B

8
U3D 7 9 A0
74LS148
I4
1 R13
2 16 1 19
2 4 3 DGND 7
I5
3
TS5-19 48" SID
TO CPU

SID INFO B TS2-2 A1 I6


4 U5A ULN2004A
R14
47 5%
7407 U3C 6
I7
5 15 2 18
1 6 5
A2
GND
EI TS5-18 40" SID
SID INFO C TS2-1 ULN2004A U5D 47 5%
R15
7407

8
14 3 17
DGND TS5-17 36" SID
U6F U5E 47 5%
DGND
ULN2004A
R16
4 4 3 13 4 16
VERT IN TS2-4 TS5-16 VERT IN
7406 ULN2004A
47 5%

6
2%

2%

2%

2%

2%
RP2-1

RP2-2

RP2-3

RP2-4

RP2-5
390

390

390

390

390
1

1
DGND

+5V

C10
R17
0.1uF CER 15
TS5-15 VERT OUT
16

47 5%
TILT INFO A TS2-6 6 2 1A
VCC
1Y0
4 DGND 1 UDN2585A
18 R18
TO CPU

2 17 14
3 5 3
U8 16
TS5-14 TABLE L
1B 1Y1 47 5%
U7A 4 15 R19
5 1 6 5 14 13
TILT INFO B TS2-5 1G 1Y2 +5V
6 13
TS5-13 TABLE R
74139A 47 5%
1Y3
7 7 12 R20
GND 8 11 12
9 10
TS5-12 WALL L/R
VS GND 47 5%
8

+15V
C12 DGND
DGND
0.1uF
CER
+5V

C11 R21
DGND

10K
0.1uF CER
5%
14
DGND

1 2 4 13 TO RELAY K5 PIN 4
ON SHEET 4.
7406 ULN2004A

7 U6A U1E
PROGENY INCORPORATED
DGND
LINEAR IV MASTER BOARD
THIS DRAWING IS THE PROPERTY OF AND TILT/HSID CIRCUIT
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
DISCLOSED WITHOUT THE PRIOR WRITTEN
B 70-08319
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME:MASTER.DSN SHEET 3 OF 6
+15V
IR REFERENCE SECTION
+5V

+15V
+15V
R25 R29
C13 R23 4.99K
1K
MF
1K 5%
5%
0.1uF CER
R22
2.2K R26 R27
5% 1.2V 8 AGND
2K MF 1M 5%

3
2
- R24 +15V
1 2 Q1
3 1K 5% 2N2222A
+
LM358A
C16

1
4
U9A 0.1uF CER
CR10 C14
+

8
LM113 4.7uF
AGND 6 AGND
TANT -
7 9 8 10
U10A TS3-10 IR TRUE TO CPU
5 7407
+
R30 LM393A U3B

4
4.99K MF R33 R28 AGND
AGND
R31 1K
2K
5%
4.99K MF 8
MF +28V TS4-8 WALL VIRC

14
K5 TS4-14 TABLE VIRC

AGND 1 15
16 7
AGND
14
TS4-7 WALL VIRL
13
C15 R34 CR15 11 13
TS4-13 TABLE VIRL
+15V 12

+
10K
10uF 1N4002 9
5%
TANT 10 9
C17 R32 8
TS4-9 WALL SOURCE

100K 7
0.1uF CER 5 21
5%
FROM U1 PIN 13 4 6
TS5-21
AGND SHEET 3 SF ALIGN
8 AGND
AGND R35 22
TS5-22
R64 U11B -
2
11 1 100K 5%
VIRL TS3-11 AGND 15
TS4-15 TABLE SOURCE
1K 5% 3
LF442A
+ CR11 CR12 AGND

4 IN4148 1N4148 6
TS4-6 WALL RTN
R36
C18 12
100K 5% TS4-12 TABLE RTN
0.1uF CER
TO CPU

R37 10
TS4-10 WALL SHIELD
100K
-15V AGND 5%
11
TS4-11 TABLE SHIELD
R41 R38
100K 5% 100K 5%
AGND

U11A CR13 CR14 AGND


6
R65 - 1N4148 1N4148
12 7
VIRC TS3-12 R39
1K 5% 5
PROGENY INCORPORATED
+
LF442B 100K 5%

R40
100K
5%
LINEAR IV MASTER BOARD
THIS DRAWING IS THE PROPERTY OF AND IR REF SECTION
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
AGND PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
DISCLOSED WITHOUT THE PRIOR WRITTEN
B 70-08319
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME:MASTER.DSN SHEET 4 OF 6
+15V
SID CIRCUIT 1
R44 R45 R46

+15V 237 MF 33.2K MF 10K MF R49


+15V
+15V
+15V 100K MF
+15V

C19 C20
C22
R43 0.1uF CER
0.1uF CER 8
10K 0.1uF CER
CR16 AGND 5% 2
AGND
8
LM113 8 - R50 AGND
1 2
CR17 U13A

1
2 -
- 100K MF
3 1 14
1 2 Q2 R47
+
LM358A
U14A TS3-14 VSID1
2N2905 3
3 1N750 100K 4 +
+ LM358A
LM358A MF R48 C21

3
4
4
U12A 33.2K R51 C23
0.1uF CER
MF 33.2K
R42 AGND MF 0.1uF CER
AGND
2.2K
AGND
5%

-15V -15V
AGND
W1 AGND
C24 3
TS4-3 CSID1

+
AGND
2.2uF
TANT

R52
2
TS4-2 CSID1 RTN
MA1+ MA1- 100 5%
TP1 TP2
SID CIRCUIT 2 AGND
+15V
AGND

R55
R58 R59 R60
+15V 237 MF
+15V 33.2K MF 10K MF 100K MF

R54 6
- R61
CR18 10K 7 6
LM113 5% -
5 100K MF 7 13
R56
+
LM358
TS3-13 VSID2
CR19
1

6 5
- +
7 2
Q3 100K U13B LM358
2N2905
MF R57 U14B
5
+ 1N750 33.2K R62
LM358 MF 33.2K
3

U12B MF

R53
2.2K
AGND
5% W2 AGND
C25 5
TS4-5 CSID2
+

2.2uF
TANT
AGND R63
4
AGND TS4-4 CSID2 RTN
MA2+ MA2- 100 5%
TP3 TP4
AGND

PROGENY INCORPORATED
LINEAR IV MASTER BOARD
THIS DRAWING IS THE PROPERTY OF AND SID CIRCUITS 1 2
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
B 70-08319
DISCLOSED WITHOUT THE PRIOR WRITTEN
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME:MASTER.DSN SHEET 5 OF 6
+5V

1
2%

2%

2%

2%

2%
RP3-1

RP3-2

RP3-3

RP3-4

RP3-5
10K

10K

10K

10K

10K
2

6
11 5 6 9
TS5-11 TS3-9 RAD/FL
FLUORO/RAD 7406
SWITCH
10 U6C
TS5-10
DGND
9 9 8 8
TS5-9 TS3-8 SF TRUE
SF TRUE 7406
SWITCH +5V
8 U6D
TS5-8
C26
DGND 14
0.1uF CER TO CPU
DGND
15 1 2 7
T=-90' TS6-15 TS3-7 T=-90'
7407

7
U15E

DGND
14 3 4 6
T=+90' TS6-14 TS3-6 T=+90'
7407
TABLE U15D
TILT
SWITCH

13 5 6 5
T=0' TS6-13 TS3-5 T=0'
7407

12
U15C
TILT COM TS6-12

11
SHIELD TS6-11

DGND

PROGENY INCORPORATED
LINEAR IV MASTER BOARD
THIS DRAWING IS THE PROPERTY OF AND EXT INPUT SID INTR CIRCUIT
CONTAINS INFORMATION PROPRIETARY TO
SIZE ITEM CLASS DRAWING NO.
PROGENY INCORPORATED. THIS DRAWING
SHALL NOT BE REPRODUCED, USED OR
DISCLOSED WITHOUT THE PRIOR WRITTEN
B 70-08319
CONSENT OF PROGENY INCORPORATED. SCALE FILENAME:MASTER.DSN SHEET 6 OF 6

You might also like