You are on page 1of 5

The Journal of Arthroplasty Vol. 24 No. 6 Suppl.

1 2009

Analysis of Wear in Retrieved


Mobile and Fixed Bearing Knee Inserts
Gerard A. Engh, MD, Rebecca L. Zimmerman, MS,
Nancy L. Parks, MS, and C. Anderson Engh, MD

Abstract: One purported advantage for mobile bearing (MB) knee implants is reduced
polyethylene wear. Twenty-three retrieved Low Contact Stress (LCS) MB and 31 Anatomic
Modular Knee (AMK) fixed bearing tibial polyethylene inserts, both Enduron (Depuy, Warsaw,
IN) and sterilized by gas plasma, were graded for top and backside wear using the Hood grading
system. For articular surface wear, there was no statistical difference between pitting and
scratching, but burnishing was twice as much for MB inserts (P = .003). For backside scores, there
was a minimal amount of pitting for both, but the scratching score was twice (P = .000) and the
burnishing score was 3 times greater for MB inserts (P = .000). At more than 5 years in situ, the
linear wear measurements were similar for the 2 groups (mean penetration was 0.329 mm for MB
and 0.320 mm for fixed bearing). Keywords: mobile bearing, fixed bearing, knee implants.
© 2009 Published by Elsevier Inc.

Polyethylene wear is a major limitation of the long-term results support this hypothesis with revision rates for
success of total knee arthroplasty (TKA) [1,2]. Although aseptic loosening between 0% and 0.6% [3-5]. Lastly,
new sterilization techniques adopted in the mid 1990s the mobility of the insert offers the advantage of
have improved the performance of ultrahigh molecular postoperative self-alignment. The ability of the bearing
weight polyethylene, long-term wear is still a concern. to align itself allows the contact surface area between
Even with tibial bearings that were sterilized with the femoral component and tibial bearing to be
conventional current methods (methods other than maximized [6].
gamma-air), polyethylene wear still accounts for most In contrast, traditional fixed bearing (FB) knees have
revisions at the Anderson Clinic that were revised after 2 less conforming geometry. The decreased level of con-
years in situ. straint between the femoral component articulating
Mobile bearing (MB) technology was introduced more geometry and tibial bearing allows the soft tissue to direct
than 25 years ago as a means to reduce knee implant movement, which will theoretically allow the knee to
wear. By decoupling the joint and adding mobility move anatomically. This reduction in constraint will also
between the tibial insert and the tibial tray, the lower the torsional and shear forces on the bone-implant
articulating tibial geometry can more highly conform interface. This design, like the MB technology, allows
to the femoral condylar geometry. This increase in more postoperative self-alignment than more congruent
conformity leads to a decrease in contact stress and FB implants and reduces the likelihood of mechanical
theoretically a reduction in polyethylene wear. By loosening. Therefore, the major differences we see
allowing this rotation, it also reduces the torsional and between the 2 designs are the level of femoral compo-
shear forces seen at the bone-implant interface. Clinical nent/tibial bearing conformity and the introduction of a
second articulating surface with MB implants.
Retrieval studies are useful in providing the efficacy of
From the Anderson Orthopaedic Research Institute, Alexandria, Virginia.
Submitted January 14, 2009; accepted March 15, 2009. designs in vivo. Previous studies have investigated the
Benefits or funds were received in partial or total support of the wear patterns on both MB tibial inserts and FB tibial
research material described in this article. These benefits or support inserts; however, a direct comparison in the literature has
were received from the following sources: Royalties for the AMK Knee
System from Depuy Orthopaedics, INOVA Health System, and a not been made between FB and MB tibial inserts
cooperative agreement that was awarded and administered by the U. machined from the same polyethylene [7-10]. Many of
S. Army Medical Research and Materiel Command (USAMRMC), and these studies were also performed on tibial bearings
the Telemedicine and Advanced Technology Research Center (TATRC),
Under Contract Number W81XWH-05-2-0079. sterilized by gamma radiation, and exposed to air during
Reprint requests: Rebecca L Zimmerman, MS, Anderson Orthopae- storage [7,10]. This method of sterilization has been
dic Institute, 2501 Parkers Lane Suite 200, Alexandria, VA 22306. proven to result in oxidation, which prematurely fatigues
© 2009 Published by Elsevier Inc.
0883-5403/09/2406-0007$36.00/0 the polyethylene and causes a loss in mechanical
doi:10.1016/j.arth.2009.03.010 properties and greater wear.

28
Knee Mobile and Fixed Bearing Knee Inserts  Engh et al 29

In this study, Low Contact Stress (LCS) rotating patient, a loose cemented tibia in 1 patient, a femoral
platform inserts (Depuy, Warsaw, IN) and Anatomic fracture in 1 patient, and a failed patella in 1 patient.
Modular Knee (AMK) FB inserts (Depuy) retrieved at Three FB inserts retrieved and analyzed were from
revision TKA were analyzed to evaluate whether the revision TKAs, and 3 were from primary TKAs.
increased conformity and addition of an articulating Linear wear was calculated by measuring the most
surface affected the in vivo wear of the polyethylene deformed or thinned area on the medial and lateral
inserts. Both sets of inserts were manufactured from articulating surface on each tibial insert and subtracting
Enduron polyethylene (Depuy) and sterilized with gas this value from the original thickness of the insert. The
plasma. We hypothesized that retrieved MB tibial inserts original thickness was obtained by subtracting the
would have less surface wear and less linear penetration thickness of the tibial tray from the thickness labeled on
than their less conforming FB counterparts. the insert. The deviation from the original thickness
represented the final linear penetration value. Manufac-
Materials and Methods turing tolerances of 0.254 mm were neglected when
Twenty-three MB and 31 FB tibial polyethylene making our measurements.
inserts all retrieved at the time of revision were
analyzed. The MB retrieved specimens were obtained Results
from 8 men and 15 women. The average age at revision The most common form of articular surface wear
surgery was 65 ± 13 years, and the average length of identified was burnishing, followed by scratching and
implantation was 29 ± 31 months (range, 1-102 months). pitting. Surface deformation and abrasion were very
The average patient weight was 186 ± 36 lb. The reasons minimal on the articulating surface of both FB and MB
for revision included instability/loosening (61%), infec- tibial inserts. The average articulating surface deforma-
tion (17%), osteolysis (4%), and other (17%). tion score was 0.9 ± 1.2 for the MB inserts and 0.8 ± 1.3
The FB retrieved specimens were obtained from 17 for the FB inserts. The average articulating surface
men and 14 women. The average age at revision surgery abrasion score was 0.6 ± 0.9 for the MB inserts and 0.5
was 67 ± 11 years, and the average length of implantation ± 1.2 for the FB inserts. There was no delamination,
was 32 ± 33 months (range, 1-141 months). The average embedded cement debris, or metal debris observed on the
patient weight was 218 ± 38 lb. The reasons for revision articulating surface for any of the tibial inserts, FB or MB.
included instability/loosening (42%), infection (32%), For the 3 modes of wear that were the most prevalent,
osteolysis (6%), other (19%). there was no statistical difference between MB and FB
All tibial inserts were inspected visually for the type and pitting and scratching, but there was twice as much
severity of wear using a technique developed by Hood et burnishing for the MB inserts (P = .003) (Fig. 1).
al [11]. All inserts were inspected by 2 independent Table 1 contains the total wear scores for the different
observers for evidence of surface deformation, pitting, zones that were scored on the articulating surface of the
embedded cement debris, scratching, burnishing, abra- insert (medial anterior, medial posterior, lateral anterior,
sion, delamination, and metal debris. Wear was classified and lateral posterior). Severe asymmetric wear was not
from 0 to 3, in which 0 indicates no wear, and 3 indicates observed in either the MB or FB inserts.
severe wear. Wear on both the articulating surface and There was a minimal amount of pitting on the backside
backside surface was assessed with the same technique. for both MB and FB inserts, but the average scratching
The thicknesses of 6 MB and 6 FB tibial polyethylene score was twice (P = .000), and the average burnishing
inserts in situ 55 to 148 months were measured using score was 3 times greater for MB inserts (P = .000) (Fig.
digital calipers. The MB inserts were retrieved from 5 2). Surface deformation, embedded cement debris, and
women and 1 man. The average age at revision surgery
was 65 ± 11 years, and the average length of implantation
was 76.6 ± 17.2 months (range, 55-102 months). The
average patient weight was 186 ± 30 lb. Four inserts were
10 mm, and 2 were 15 mm thick. Preoperative diagnosis
included a loose cemented tibia in 2 patients, infection in
2 patients, a loose cemented femur in 1 patient, and
osteolysis/polywear in 1 patient. All MB inserts retrieved
and analyzed were primary TKAs.
The FB inserts were retrieved from 2 women and 4
men. The average age at revision surgery was 67 ± 11
years, and the average length of implantation was 87.7 ±
30.6 months (range, 66-141 months). The average
patient weight was 218 ± 54 lb. One insert was 12 mm,
4 were 14 mm, and 1 was 16 mm. Preoperative diagnosis
included infection in 2 patients, osteolysis/polywear in 1 Fig. 1. The MB vs FB articular surface scores and P values.
30 The Journal of Arthroplasty Vol. 24 No. 6 Suppl. 1 September 2009

Table 1. Total Articular Surface Wear Scores for Individual Table 2. Average Linear Wear Rate for the Medial and Lateral
Zones Compartments of MB and FB Tibial Inserts
Medial Medial Lateral Lateral Linear Wear
Anterior Posterior Anterior Posterior Linear Wear (mm) Rate (mm/y)
AMK Medial compartment
Average 2.8 2.5 2.3 2.1 MB 0.407 ± 0.394 (range, 0.263 to 1.207) 0.061 ± 0.041
Standard 1.6 1.6 1.4 1.6 FB 0.381 ± 0.554 (range, −0.157 to 1.357) 0.054 ± 0.083
deviation P value for MB vs FB medial .699
LCS compartment measurements
Average 3.8 3.6 3.7 3.4 Lateral compartment
Standard 1.8 1.7 1.6 1.7 MB 0.252 ± 0.359 (range, 0.057 to 0.177) 0.035 ± 0.040
deviation FB .259 ± .285 (range, −0.183 to 0.730) 0.029 ± 0.029
P value for MB vs FB lateral .937
compartment measurements
abrasion were very minimal on the backside of both FB
and MB tibial inserts. The average backside surface
deformation score was 0.5 ± 0.7 for the MB inserts and pared to the FB knee. The LCS knee allows a very large
0.4 ± 1.9 for the FB inserts. The average backside area of contact between both the medial and lateral
embedded cement debris score for the FB inserts was femoral condyles and the tibial polyethylene insert
0.1 ± 0.3, and the average backside abrasion score was throughout a wide range of motion. The overall increase
0.1 ± 0.3. There was no delamination or metal debris in conformity has been documented to decrease contact
observed for either group of tibial inserts. stress; however, as evidenced in this study, it also
Table 2 contains the average linear penetration values increases the contact between the femoral component
(mm) and linear wear rates (mm/y) for both the medial and tibial component resulting in a larger percentage of
and lateral compartments of the FB and MB knees. There burnishing wear [6,12,13]. The increased contact area
were no statistically significant differences between either results in a dominant wear mechanism that is similar to
group. total hip arthroplasty.
All MB tibial inserts were retrieved from the patient's Wear on the backside of FB tibial inserts has been
primary TKA. The FB knees were retrieved from 3 knees recognized as a potential source of polyethylene debris,
that were primary TKAs and 3 knees that were revision even with the tibial insert firmly locked to the tray [14,15].
TKAs. The linear measurements were further grouped for Therefore, the addition of a second articulating surface for
the FB TKAs into primary and revision measurements. MB knees raises concerns. This retrieval study did confirm
The linear measurements for the medial compartment in that on the backside of MB inserts, pitting, scratching, and
the primary FB TKAs was 0.340 ± 0.306 mm, and for the burnishing were significantly greater than their FB
lateral compartment was 0.404 ± 0.313 mm. The linear counterparts. The pitting and scratching observed were
measurements for the medial compartment in the most likely a result of third body debris. The scratches
revision FB TKAs was 0.421 ± 0.818 mm, and for the observed on the backside of the MB inserts were
lateral compartment 0.114 ± .346 mm. concentric and were aligned with rotational movement.
Prior studies have attributed these concentric rings on the
Discussion backside of rotating platform tibial inserts to third body
The MB knee was designed to increase conformity and debris trapped between the tibial tray and tibial insert
consequently decrease polyethylene wear when com- [10]. When compared to FB tibial inserts with improved
locking mechanisms, it appears that the introduction of
the second articulating surface creates more visual wear.
Ho et al [7] performed a similar study comparing FB and
MB knees; however, the tibial inserts in their study were
sterilized by gamma-air methods. They grouped the wear
modes into low-grade and high-grade wear, where low-
grade wear consisted of burnishing, abrasion, and surface
deformation, and high-grade wear consisted of pitting,
scratching, and delamination. In accordance with their
study, we found that MB knees increased low-grade wear;
however, unlike their study, we did not find a significant
increase in high-grade wear for FB knees. They reported
that 22.2% of the FB tibial inserts they evaluated had
severe wear through from delamination, whereas no
delamination was present in our study for either FB or MB
inserts. We also observed during this study that the gas
Fig. 2. The MB vs FB backside surface scores and P values. plasma–sterilized inserts we analyzed have very low wear
Knee Mobile and Fixed Bearing Knee Inserts  Engh et al 31

rates. Prior studies have reported linear wear rates of revision rates for osteolysis. Garcia et al [8] found in
0.127 and 0.350 mm/y [16,17]. We observed wear rates of their study of retrieved rotating platform knees that
approximately 0.05 to 0.06 mm/y for the medial greater levels of damage were present on both the inferior
compartment and approximately 0.03 mm/y for the and superior surface of the tibial inserts in knees with
lateral compartment of both fixed and MB knees. These greater radiographic radiolucency scores and mechani-
wear rates are in agreement with Kops and Swarts [9], cally loose components. Minoda et al [19] collected the
who reported a penetration rate of 0.054 mm/y for the synovial fluid 1 year after implantation from 22 TKA
superior surface of LCS knees. They also analyzed patients with well-functioning prostheses. These were
polyethylene tibial inserts sterilized with current meth- conventionally sterilized MB and FB knees. They did not
ods. This is encouraging information that regardless of find a statistically significant difference between the
design, improvements in sterilization techniques have number of particles, particle size, or particle shape. In
made a large contribution to reducing polyethylene wear. contrast, Huang et al [20] reported their observations for
Although visually the LCS knee appears to have more the revision rate of MB and FB knees that were revised
articular surface wear than the AMK knee, this may be for osteolysis. Osteolysis was identified intraoperatively in
deceptive. A visual grading scale, such as the Hood scale, 16 (47%) of 34 knees in the MB group and in 6 (13%) of
limits the user to a 2-dimensional assessment of the 46 knees in the FB group. In this study, there was a
surface. The extent of damage from burnishing wear, in similar revision rate for osteolysis between the 2 groups.
particular, may be underestimated using this 2-d grading There are several imitations to this study. First is the
system. Burnishing wear occurs from the articulation of subset of inserts. The inserts analyzed were retrieved in
the metal femoral component polishing the surface of the conjunction with some form of failure; therefore, they
polyethylene tibial component. In the Hood grading scale, may not represent the larger population of well-function-
burnishing is scored by the percentage of the surface that ing knees. In addition, only a specific implant from a
is burnished, ranging between 1 (mild) and 3 (severe). single manufacturer was analyzed to represent FB and
The important factor that is not accounted for using this MB knees. The findings may not be applicable to other
2-d system is that polishing of the surface coupled with inserts of different designs or by different manufacturers.
cold flow of the polyethylene will eventually deform the Lastly, the study does not further expound on the reason
surface, meaning more material will be lost, but the score for failure and how it affects the wear pattern. The
will remain the same. For instance, the LCS knee is more different types of failure observed in this study are
conforming than the AMK knee; therefore, there is more markedly different and could affect the type of wear that
contact between the femoral component and the tibial was observed.
insert. Hypothetically, this contact may result in a large The information presented in this study warrants further
percentage of surface being polished or burnished, which investigation into the volume of material lost from the 2
using a visual grading scale would warrant a score of 3. implant designs. Visually, the benefits of reduced wear
This polishing may be very shallow and may not result in with MB tibial inserts was not realized in our study, and
a large amount of material removal, but it is termed linear penetration was similar for the 2 groups.
“severe.” On the other hand, the AMK knee may have a
smaller area that has been burnished. This area of
burnishing may be very deep, but because it covers less
References
1. Lonner JH, Siliski JM, Scott RD. Prodromes of failure in
than 50% of the surface, it is considered “moderate.” In total knee arthroplasty. J Arthroplasty 1999;14:488.
terms of actual material lost in our hypothetical example, 2. Sharkey PF, Hozack WJ, Rothman RH, et al. Insall Award
the AMK may have more material removed, but with our Paper: why are total knee arthroplasties failing today? Clin
visual grading scale, the LCS would be given the severe Orthop Relat Res 2002;404:7.
score. Because of this phenomenon, the factor that is 3. Callaghan JJ, O'Rourke M, Iossi M, et al. Cemented
more important when assessing wear damage is volu- rotating-platform total knee replacement: a concise fol-
metric loss of material. Further investigation into the low-up, at a minimum of fifteen years, of a previous report.
volumetric loss of material is currently being done at JBJS Am 2005;87:1995.
our laboratory. 4. Buechel F. Long-term followup after mobile-bearing total
Preliminary results from our volumetric measurements knee replacement. CORR 2002;404:40.
5. Buechel F. Mobile-bearing knee arthroplasty: rotation is
have shown that surface deformation caused by burnish-
our salvation! JOA 2004;19(4 Suppl 1):27.
ing produces a larger volume of material loss than wear 6. Stukenborg-Colsman C, Ostermeier S, Hurschler C, et al.
modes, which are visually more unappealing, such as Tibiofemoral contact stress after total knee arthroplasty:
pitting and scratching. This is a concern because comparison of fixed and mobile-bearing inlay designs. Acta
burnishing wear generates smaller particles, which have Orthop Scand 2002;73:638.
been shown to produce a greater cellular response than 7. Ho F, Ma H, Liau J, et al. Mobile-bearing knees reduce
larger particles [18]. Previous studies have been con- rotational asymmetric wear. CORR 2007;462:143.
ducted on the in vivo generation of particulate debris for 8. Garcia R, Kraay M, Messerschmitt P, et al. Analysis of
the MB and FB prostheses and subsequently their retrieved ultra-high-molecular-weight polyethylene tibial
32 The Journal of Arthroplasty Vol. 24 No. 6 Suppl. 1 September 2009

components from rotating-platform total knee arthroplasty. 15. Conditt M, Ismaily S, Alexander J, et al. Backside wear of
JOA 2009;24:131. modular ultra-high molecular weight polyethylene tibial
9. Kops A, Swarts E. Quantification of polyethylene degradation inserts. JBJS 2004;86-A:1031.
in mobile bearing knees: a retrieval analysis of the anterior- 16. Lavernia C, Sierra R, Hungerford D, et al. Activity level and
posterior-glide (APG) and rotating platform (RP) low contact wear in total knee arthroplasty: a study of autopsy retrieved
stress (LCS) knee. Acta Orthopaedica 2007;78:364. specimens. JOA 2001;16:446.
10. Atwood S, Currier J, Mayor M, et al. Clinical wear 17. Benjamin J, Szivek J, Dersam G, et al. Linear and
measurement on low contact stress rotating platform knee volumetric wear of tibial inserts in posterior cruciate-
bearings. JOA 2008;23:431. retaining knee arthroplasties. CORR 2001;392:131.
11. Hood R, Wright T, Burstein A. Retrieval analysis of total knee 18. Fisher J, McEwen H, Tipper J, et al. Wear, debris, and
prostheses: a method and its application to 48 total condylar biologic activity of cross-linked polyethylene in the knee.
prostheses. J of Biomedical Materials Research 1983;17:829. Benefits and potential concerns. CORR 2004;428:114.
12. McNamara JL. A comparison of contact pressures in tibial 19. Minoda Y, Kobayashi A, Iwaki H, et al. Characteristics of
and patellar total knee components before and after service polyethylene wear particles isolated from synovial fluid after
in vivo. CORR 1994;299:104. mobile-bearing and posterior-stabilized total knee arthro-
13. D' Lima DD. Polyethylene contact stresses, articular con- plasties. J Biomed Mater Res B Appl Biomater 2004;71:1.
gruity, and knee alignment. CORR 2001;392:232. 20. Huang C, Ma H, Liau JJ, et al. Osteolysis in failed total
14. Conditt M, Thompson M, Usrey M, et al. Backside wear of knee arthroplasty: a comparison of mobile-bearing and
polyethylene tibial inserts: mechanism and magnitude of fixed-bearing knees. J Bone Joint Surg Am 2002;84-A:
material loss. JBJS 2005;87-A:326. 2224.

You might also like