This technical note provides information for installing and operating Vaisala sampling cells
DMT242SC, DMT242SC2, DSC74, DSC74B, DSC74C, DMCOIL.
All Vaisala Sample Cells use G1/2" The sampling cells are compatible
ISO parallel threads for the probe with the following Vaisala meters
connection point. and transmitters:
Sampling cell models To fit 6 mm tubing to the connectors, screw is opened and closed with
an adapter such as Swagelok® a screwdriver. The leak screw
Reducer SS-6M0-R-4 (not supplied should be 1/2 turn open. This can
DMT242SC & SC025NPT Sampling by Vaisala) can be used. be verified by first closing the leak
Cell - basic sample cell screw, then turning 1/2 turn to
DMT242SC2 is the suitable choice open it. Then the air coming out
The DMT242SC and SC025NPT are in e.g. plastics drying systems, will make a hardly audible little
a basic sample cell with only the where the measurement is made by fizz and it can be lightly felt when
main sampling cell body. The inlet tapping off the dryer system and putting a hand in front of the air
and outlet are female threaded bringing a small air stream to the stream.
sample connectors. DMT242SC sensor. The swagelok connectors
have inlet G3/8", outlet G1/4" ISO, of DMT242SC2 easily connect to a The DSC74 comes with a quick
and the SC025NPT has 1/4"NPT cooling coil or tubing providing the connector that fits to the industry
inlet and 1/4"NPT outlet. essential function of cooling the dry standard compressed air line
air to ambient temperature before it connectors. This allows for easy
The DMT242SC and SC025NPT are reaches the sensor. installation and detachment of
suitable for users requiring only a the dewpoint transmitter without
sampling cell to fit the probe into 2 having to shut down the process.
and doing further assembly (piping Alternative ways to connect are
into inlet & outlet, valves, possible through the two different thread
flow meter) by themselves. adapters (G3/8" to G1/2" and
G3/8" to G1/4" ISO) that are
supplied with each DSC74 unit.
2
1 1
2
3
1
DSC74B
DSC74C
• Reducing Adapter (thread adapter), G3/8" – G1/4" 1 = Gas goes in. Also the coil can be used here.
2 = Gas comes out
• diffusion coil (for measurements in atmospheric 3 = Coil
pressure) 4 = Valve
Chamber
for DMP74 Valve
Measurement
in the process Chamber Removable
Configuring for measurements at pressure for DMP74 Valve flowlimiter
Sample gas
inlet DSC74B
Diffusion
DSC74C coil
DMCOIL operation