You are on page 1of 167

PM135 Powermeter Series

PM135P/PM135E/PM135EH

Installation and Operation Manual

BG0518 Rev. A3

Downloaded from www.Manualslib.com manuals search engine


LIMITED WARRANTY
The manufacturer offers the customer a 24-month functional warranty on the
instrument for faulty workmanship or parts from date of dispatch from the
distributor. In all cases, this warranty is valid for 36 months from the date of
production. This warranty is on a return to factory basis.
The manufacturer does not accept liability for any damage caused by instrument
malfunction. The manufacturer accepts no responsibility for the suitability of the
instrument to the application for which it was purchased.
Failure to install, set up or operate the instrument according to the instructions
herein will void the warranty.
Only a duly authorized representative of the manufacturer may open your
instrument. The unit should only be opened in a fully anti-static environment.
Failure to do so may damage the electronic components and will void the warranty.
The greatest care has been taken to manufacture and calibrate your instrument.
However, these instructions do not cover all possible contingencies that may arise
during installation, operation or maintenance, and all details and variations of this
equipment are not covered by these instructions.
For additional information regarding installation, operation or maintenance of this
instrument, contact the manufacturer or your local representative or distributor.

WARNING
Read the instructions in this manual before performing installation and take
note of the following precautions:

Ensure that all incoming AC power and other power sources are
turned OFF before performing any work on the instrument. Protect
the measurement AC Inputs voltage (V1, V2, V3) with 2A external
overcurrent protection device and the power supply source inputs
with 5A external overcurrent protection device, located close to the
equipment.

Before connecting the instrument to the power source, check the


labels on the back of the instrument to ensure that your instrument
is equipped with the appropriate power supply voltage, input
voltages and currents. Failure to do so may result in serious or even
fatal injury and/or equipment damage.

Under no circumstances should the instrument be connected to


a power source if it is damaged.

To prevent potential fire or shock hazard, do not expose the


instrument to rain or moisture.

The secondary of an external current transformer must never be


allowed to be open circuit when the primary is energized. An open
circuit can cause high voltages, possibly resulting in equipment
damage, fire and even serious or fatal injury. Ensure that the current
transformer wiring is secured using an external strain relief to
reduce mechanical strain on the screw terminals, if necessary.

Only qualified personnel familiar with the instrument and its


associated electrical equipment must perform setup procedures.

Do not open the instrument under any circumstances when it is


connected to a power source.

2 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


Do not use the instrument for primary protection functions where
failure of the device can cause fire, injury or death. The instrument
can only be used for secondary protection if needed.

Read this manual thoroughly before connecting the device to the


current carrying circuits. During operation of the device, hazardous
voltages are present on input terminals. Failure to observe
precautions can result in serious or even fatal injury or damage to
equipment.

 This equipment does not require cleaning for proper operation

All trademarks are property of their respective owners.


Copyright  2007-2012

PM135 Powermeter Series 3

Downloaded from www.Manualslib.com manuals search engine


Quick Start Guide
This section can be used by a licensed electrician to install and perform basic
PM135 setup. For more detailed PM135 setup and use instructions, see the
following chapters in this manual.
This quick start guide will assist you to have the unit running for the first
time.
During the operation of the meter, hazardous voltages are present in
the input terminals. Failure to observe precautions can result in
serious or even fatal injury or damage to equipment.
For complete and accurate in-depth instructions, refer to the following
chapters in this manual.

1. Installing the PM135


Mounting the PM135 Unit
To mount the PM135:
1. Position the PM135 unit in the square or round
cutout. If two PM135 are positioned side by side,
take care of proper interval between them.
2. Attach the PM135 unit using washers and nuts. Make
sure that the unit is securely attached into the wall
or cabinet fixture.

Mounting the PM135 (Square or Round Cut-out)

4 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


Mounting two PM135 side by side

Connecting the PM135 Unit


To connect the PM135:
1. Ensure that all incoming power sources are OFF.
2. Check that you have the appropriate power supply.
3. Connect to the external CT by passing the external
CT wire through the meter CT core. Observe the
arrow that indicates the current direction.
4. In case of a retrofit application where each external
CT ends with two wires:
 Pass one wire through the meter CT core.
 Connect the wire to one of the meter termination
screws.
 Connect the second wire from the external CT to
the termination screw.
5. Connect the measured voltage inputs
6. Connect COM1 – RS-485 communication port
7. Connect the Power Supply inputs using 1.5
mm2/14AWG-dedicated wires.

PM135 Powermeter Series 5

Downloaded from www.Manualslib.com manuals search engine


Common Wiring Mode: 4LL3 or 4Ln3

CT wiring

6 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


To connect an Option module:
1. Assemble the module on the meter.
2. Power the PM135 unit on.

RS-485 Terminals

Module connector cover

Remove Module Connector cover


before assembling module

Assembling a module

To operate the PM135:


1. Perform device diagnostics.
2. Configure the device through the PM135 unit front
panel display.

PM135 Powermeter Series 7

Downloaded from www.Manualslib.com manuals search engine


2. Configuring the PM135 remotely
1. Install the PAS application software on your PC.
2. Configure the PAS database for your meter.
3. Configure the PAS communications settings.
4. Upgrade the meter firmware if a new version is
available.
5. Set up the meter using the PAS application software.
6. Configure your security settings through the meter
security setup.
7. Configure your communication protocol settings.
8. Configure Billing/TOU registers.

At this stage, the PM135 should be ready for operation.

8 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


Table of Contents
Chapter 1 General Information ..........................................................................12
1.1 Features.......................................................................................................................12
1.2 Available Options .......................................................................................................14
Digital I/O ............................................................................................................ 14
Analog Output ...................................................................................................... 15
Additional Communication Port – COM2 ................................................................... 15
TOU - Battery-Operated Clock Unit.......................................................................... 15
1.3 Customized Options ..................................................................................................15
Device Resolution.................................................................................................. 16
Energy Rollover .................................................................................................... 16
Display Options..................................................................................................... 16
1.4 Measured Parameters ..............................................................................................16
Chapter 2 Installation ..........................................................................................19
2.1 Site Requirements .....................................................................................................19
2.2 Package Contents......................................................................................................19
2.3 Mechanical Installation.............................................................................................19
Panel Mounting ..................................................................................................... 20
DIN Rail Mounting ................................................................................................. 20
2.4 Electrical Installation.................................................................................................22
Typical Installation ................................................................................................ 22
Terminals ............................................................................................................. 23
Power Source Connection....................................................................................... 23
Voltage Input connection ....................................................................................... 24
Current Input Connection....................................................................................... 24
Wiring Diagrams ................................................................................................... 25
2.5 I/O Connections ........................................................................................................30
4DI/2DO Module ................................................................................................... 30
8 DI module ......................................................................................................... 32
12DI/4RO Module ................................................................................................. 32
4AO Module - Analog Outputs................................................................................. 34
TOU module – RTC and 4 Digital Inputs ................................................................... 34
2.6 Communications Connections.................................................................................36
COM1 RS-485 Connection ...................................................................................... 37
ETH module – COM2 Ethernet Connection................................................................ 37
PRO module – COM2 PROFIBUS Connection............................................................. 38
RS-232/422-485 module – COM2 Connection .......................................................... 39
GPRS/GSM modem module – COM2 GPRS connection .............................................. 40
Chapter 3 Using Front Display ............................................................................41
3.1 Display Operations ....................................................................................................41
Navigation Buttons ................................................................................................ 42
Navigating in Data Display Mode............................................................................. 43
Status Indicators................................................................................................... 43
Display Features ................................................................................................... 44
3.2 Data Displays..............................................................................................................46
TEST Mode Data Display ........................................................................................ 46
Billing Period Data Displays .................................................................................... 47
Measurements Maximum Demand Data Display ....................................................... 50
Measurements Data Display ................................................................................... 51

PM135 Powermeter Series 9

Downloaded from www.Manualslib.com manuals search engine


Measurements Minimum/Maximum Data Display ...................................................... 54
Measurements Energy Data Display ........................................................................ 55
Harmonics Display ................................................................................................ 56
Waveform Display ................................................................................................. 57
Phasor Display ...................................................................................................... 57
Digital I/O ............................................................................................................ 57
Device Control Display ........................................................................................... 58
Basic Settings Display ............................................................................................ 59
Device Info Display ............................................................................................... 59
Load Bar Graph .................................................................................................... 59
Energy Pulse LED .................................................................................................. 60
Port Activity LEDs.................................................................................................. 60
3.3 Device Setup...............................................................................................................61
Entering the Password ........................................................................................... 61
Viewing and Changing Setup Options ...................................................................... 61
Chapter 4 Using PAS Software ...........................................................................63
4.1 Installing PAS .............................................................................................................63
4.2 Creating a New Site for your Meter ........................................................................63
4.3 Setting up Communications ....................................................................................64
Communicating through a Serial Port ...................................................................... 65
Communicating through the Internet ...................................................................... 66
4.4 Setting Up the Meter .................................................................................................67
Preparing Setups for the Meter ............................................................................... 67
Downloading Setups to the Meter ........................................................................... 68
Uploading Setups from the Meter ............................................................................ 69
4.5 Authorization..............................................................................................................69
Chapter 5 Configuring the PM135......................................................................71
5.1 Configuring Communications..................................................................................71
Setting Up Serial Communication Ports.................................................................... 71
Setting Up Ethernet............................................................................................... 73
Setting Up GPRS Network ...................................................................................... 74
Setting Up eXpertPower Client ................................................................................ 76
Setting Up TCP Notification Client ............................................................................ 77
5.2 General Meter Setup .................................................................................................79
Basic Meter Setup ................................................................................................. 79
Transformer Correction .......................................................................................... 81
Device Options ..................................................................................................... 82
Configuring Digital Inputs....................................................................................... 83
Configuring Relay Outputs...................................................................................... 85
Configuring Analog Outputs.................................................................................... 86
Configuring Counters ............................................................................................. 88
Configuring Alarm/Control Setpoints........................................................................ 89
Configuring the Display .......................................................................................... 93
Updating the meter Clock....................................................................................... 94
Local Time Settings ............................................................................................... 94
5.3 Configuring Meter Security ......................................................................................96
5.4 Configuring Billing/TOU ...........................................................................................99
Configuring Billing/Tariff Registers........................................................................... 99
Configuring the Daily Tariff Schedule ..................................................................... 101
Configuring the Season Tariff Schedule.................................................................. 101
5.5 Configuring Recorders ........................................................................................... 104
Configuring Meter Memory ................................................................................... 104
Configuring the Event Recorder ............................................................................ 106

10 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


Configuring the Data Recorder .............................................................................. 106
5.6 Configuring Communication Protocols............................................................... 110
Configuring Modbus............................................................................................. 110
Configuring DNP3................................................................................................ 111
Chapter 6 Device Control and Upgrading....................................................... 114
6.1 Resetting Accumulators, Maximum Values and Files ...................................... 114
6.2 Viewing and Clearing Device Diagnostics .......................................................... 116
6.3 Viewing Communication Status and Statistics.................................................. 117
6.4 Remote Relay Control ............................................................................................ 117
6.5 Upgrading Device Firmware................................................................................. 118
Chapter 7 Monitoring Meters ........................................................................... 121
7.1 Viewing Real-time Data......................................................................................... 121
7.2 Viewing Real-time Min/Max Log ......................................................................... 123
7.3 Viewing Real-time Waveforms ............................................................................ 124
7.4 Viewing Real-time Harmonic Spectrum............................................................. 126
Chapter 8 Retrieving and Storing Files........................................................... 129
8.1 Uploading Files on Demand.................................................................................. 129
8.2 Using the Upload Scheduler.................................................................................. 130
8.3 Viewing Files On-line.............................................................................................. 132
8.4 Exporting Files......................................................................................................... 132
Exporting Files in COMTRADE and PQDIF Formats................................................... 132
Exporting Files in Excel Format ............................................................................. 134
8.5 Archiving Files ......................................................................................................... 134
Chapter 9 Viewing Files .................................................................................... 136
9.1 Operations with Files ............................................................................................. 136
9.2 Viewing Options...................................................................................................... 136
Customizing Views .............................................................................................. 136
Working with Tables ............................................................................................ 137
Working with Graphic Windows............................................................................. 138
9.3 Viewing the Event Log ........................................................................................... 139
9.4 Viewing the Data Log............................................................................................. 141
Appendix A Technical Specifications .............................................................. 142
Appendix B Analog Output Parameters ......................................................... 149
Appendix C Setpoint Triggers and Actions .................................................... 151
Appendix D Parameters for Data Monitoring and Logging ......................... 154
Appendix E Billing/TOU Profile Log File ......................................................... 163
Appendix F Data Scales .................................................................................... 165
Appendix G Device Diagnostic Codes ............................................................. 166
Index ...................................................................................................................... 167

Designations used throughout the manual:


E - available in the PM135E
EH - available in the PM135EH

PM135 Powermeter Series 11

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 1 General Information Features

Chapter 1 General Information

The PM135 is a compact, multi-function, three-phase AC Powermeter


specially designed to meet the requirements of users ranging from
electrical panel builders to substation operators.

The PM135 comprises of three types of models:


 PM135P: the basic model which offers standard voltage, current,
power and frequency measurements, and control capabilities. A
special amp-demand version can be ordered with a simplified
display layout especially suitable for current measurements.
 PM135E: offers all the features of the basic model energy
measurements and data logging.
 PM135EH: offers all the features of the PM135E with harmonic
analysis capabilities.
The PM135 units include:
 A 3.5" Monochrome Graphic LCD display enabling easy reading of
local meters, and can be provided in different languages1.
 A standard RS-485 communication port and a second optional
Ethernet, PROFIBUS or RS-232/RS-422/RS-485 port. These ports
allow local and remote automatic meter readings and setup
through the supplemental communication or user data acquisition
software.
 Different communication options for remote communications with
the meter. These options enable LAN and Internet communication
with the unit.
All models are suitable for mounting on both 4-inch round and 9292mm
square cutouts.

1.1 Features
Multifunctional 3-phase Power Meter
 3 voltage inputs and 3 current transformer-isolated AC inputs for
direct connection to power line or via potential and current
transformers

1
Contact your local distributor for the desired Display languages support.

12 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 1 General Information Features

 True RMS, volts, amps, power, power factor, neutral current,


voltage and current unbalance, frequency
 Ampere/Volt demand meter
 25/50/60/400 Hz measurement capabilities

Billing/TOU Energy Meter (PM135E and PM135EH)


 Class 0.5S IEC 62053-22 four-quadrant active and reactive
energy polyphase static meter
 Three-phase total and per phase energy measurements; active,
reactive and apparent energy counters
 Time-of-Use, 4 totalization and tariff energy/demand registers x 8
tariffs, 4 seasons x 4 types of days, 8 tariff changes per day,
 One–time easy programmable tariff calendar schedule
 Automatic daily energy and maximum demand profile log for total
and tariff registers

Harmonic Analyzer (PM135EH)


 Voltage and current THD, current TDD and K-Factor, up to 40th
order harmonic
 Voltage and current harmonic spectrum and angles

Real-time Waveform Capture


 Real-time “scope mode” waveform monitoring capability
 Simultaneous 6-channel one-cycle waveform capture at a rate of
64 samples per cycle

Programmable Logical Controller


 Embedded programmable controller
 16 control setpoints; programmable thresholds and delays
 Relay output control (see Available Options)
 1-cycle response time

Event and Data Recording (PM135E and PM135EH)


 Non-volatile memory for long-term event and data recording
 Event recorder for logging internal diagnostic events and setup
changes
 Two data recorders; programmable data logs on a periodic basis;
automatic daily energy and maximum demand profile log

Digital I/O
 Optional four, eight or 12 digital inputs with 1-ms scan time;
automatic recording of last five digital input change events with
timestamps (see the PM135 Modbus Reference Guide)
 Optional two relay outputs with 1-cycle update time; unlatched,
latched, pulse and KYZ operation; energy pulses

Display
 3.5 inch Monochrome Graphic LCD display with 240 x 128 dots
resolution, adjustable update time backlit and user defined
brightness setting
 Auto-scroll option with adjustable page exposition time; auto-
return to a default page

PM135 Powermeter Series 13

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 1 General Information Available Options

 LED bar graph showing percent load with respect to user-


definable nominal load current

Real-time Clock
 Internal clock with 20-second retention time
 Optional battery-operated clock unit (see Available Options)

Communications
 Standard 2-wire RS-485 communication port; MODBUS RTU,
DNP3, and SATEC ASCII communication protocols
 Optional second communication port (see Available Options);
MODBUS RTU, MODBUS/TCP, DNP3, DNP3/TCP, SATEC ASCII and
PROFIBUS DP communication protocols
 eXpertPower client for communicating with the SATEC
proprietary eXpertPower Internet services (with the Ethernet
module or with the RS-232 module using an external GPRS
modem, see Setting Up eXpertPower Client)
 TCP notification client for communicating with a remote
MODBUS/TCP server on events or periodically on a time basis
(with the Ethernet module or with the RS-232 module using an
external GPRS modem, see Setting Up TCP Notification Client)

Meter Security
 Password security for protecting meter setups and accumulated
data from unauthorized changes

Upgradeable Firmware
 Easy upgrading device firmware through a serial or Ethernet port.

Software Support
 PAS – free meter configuration and data acquisition tool
 eXpertPowerTM – SATEC proprietary Internet services

1.2 Available Options


The PM135 can be provided with an optional expansion module from the
following list:
 Digital I/O
 Analog outputs
 TOU - Battery-operated clock unit
 Ethernet communication port
 PROFIBUS DP communication port
 RS-232/RS-422/RS-485 communication port
 GPRS communication port

Digital I/O
The PM135 digital I/O expansion module provides:

4DI/2DO module
 4 dry contact digital inputs (DI) for monitoring external contacts
and receiving pulses from energy, water, and gas meters
 Programmable de-bounce time; 1-ms scan time.

14 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 1 General Information Customized Options

 2 electro-mechanical or solid-state relay outputs (RO) for alarms


and controls, and for output energy pulses; unlatched, latched
and pulse operations, failsafe operation for alarm notifications;
programmable pulse width; direct remote relay control through
communications; 1-cycle update time.

8DI module
 8 dry contact digital inputs (DI) for monitoring external contacts and
receiving pulses from energy, water, and gas meters

 Programmable de-bounce time; 1-ms scan time.

12DI/4DO module
 12 dry contact digital inputs (DI) for monitoring external contacts
and receiving pulses from energy, water, and gas meters
 Programmable de-bounce time; 1-ms scan time.
 4 electro-mechanical relay outputs (RO) for alarms and controls,
and for output energy pulses; unlatched, latched and pulse
operations, failsafe operation for alarm notifications;
programmable pulse width; direct remote relay control through
communications; 1-cycle update time.

Analog Output
The PM135 analog output (AO) expansion module provides:
 4 optically isolated analog outputs with an internal power supply;
 Options for 0-20mA, 4-20mA, 0-1mA, and  1mA output; 1-cycle
update time.

Additional Communication Port – COM2


A second COM2 communication port can be ordered as an expansion
module. COM2 options available:
 Ethernet 10/100BaseT port; MODBUS/TCP , DNP3/TCP and IEC
60870-5-104 communications protocols
 PROFIBUS DP port
 RS-232/RS-422/RS-485 port; MODBUS RTU, DNP3, SATEC ASCII
and IEC 60870-5-101 communication protocols;
 GPRS communications port

TOU - Battery-Operated Clock Unit


The TOU module provides:
 A precise clock with battery backup; 6-year clock retention time
 4 dry contact digital inputs (DI) for monitoring external contacts
and receiving pulses from energy, water and gas meters;
programmable de-bounce time; 1-ms scan time.

1.3 Customized Options


Presentation of data on the front display and via communications can be
customized to best suit the user application.

PM135 Powermeter Series 15

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 1 General Information Measured Parameters

Device Resolution
A low or high-resolution option can be selected for the presentation of
voltage, current, and power for use in high and low power applications.
See Measurement Units for more information.

Energy Rollover
The energy rollover limit can be changed in the meter to provide 4-digit to
9-digit energy resolution. See Device Options in Chapter 5 for details. The
meter display is capable of showing full 9-digit energy counters using two
LED windows.

Display Options
Different display options are available for customization to be used in dark
or non-safe locations, or in places that are hardly accessible for
observation. See Configuring the Display in Chapter 5 for more
information.

1.4 Measured Parameters


Table 1: Measured and Displayed Parameters
Parameter Display Comm. Analog Pulse Alarm
1-cycle Real-time Measurements
RMS Voltage per phase    
RMS Current per phase    
kW per phase   
kvar per phase   
kVA per phase   
Power Factor per phase   
Total kW    
Total kvar    
Total kVA    
Frequency    
Neutral Current    
Total Power Factor    
Voltage & Current unbalance   
1-sec Average Measurements
RMS Voltage per phase    
RMS Current per phase    
kW per phase   
kvar per phase   
kVA per phase   
Power Factor per phase   
Total kW    
Total kvar    
Total kVA    
Total Power Factor    
Frequency    
Neutral Current    
Voltage & Current unbalance   
Amps & Volt Demands
Ampere & Volt Demand per phase   

16 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 1 General Information Measured Parameters

Parameter Display Comm. Analog Pulse Alarm


Ampere Maximum Demand per phase   
Voltage Maximum Demand per phase   
Power Demands E, EH
kW Accumulated Demand Import & Export   
kvar Accumulated Demand Import & Export   
kVA Accumulated Demand   
kW Demand Import & Export  
kvar Demand Import & Export  
kVA Demand  
kW Sliding Demand Import & Export  
kvar Sliding Demand Import & Export  
kVA Sliding Demand  
kW Predicted Demand Import & Export  
kvar Predicted Demand Import & Export  
kVA Predicted Demand  
kW Maximum Demand Import  
kW Maximum Demand Export  
kvar Maximum Demand Import  
kvar Maximum Demand Export  
kVA Maximum Demand  
Total Energy E, EH
Total kWh Import & Export   
Total kvarh Import & Export   
Total kvarh Net 
Total kVAh   
Energy per Phase E, EH
kWh Import per phase  
kvarh Import per phase  
kVAh per phase  
TOU Registers E, EH
4 TOU energy registers (kWh and kvarh  
import & export, kVAh, 4 pulse sources)
4 TOU maximum demand registers  
8 tariffs, 4 seasons x 4 types of day   
Harmonic Measurements EH
Voltage THD per phase   
Current THD per phase   
Current TDD per phase   
K-factor per phase   
Voltage harmonics per phase up to order 40  
Current harmonics per phase up to order 40  
Voltage harmonic angles up to order 40  
Current harmonic angles up to order 40  
Fundamental Component EH
Voltage and Current per phase  
kW, PF per phase  
kvar, KVA per phase  
Total kW, PF  
Total kvar, KVA  
Min/Max Logging
Min/Max A, V, total kW, kvar, kVA, PF  

PM135 Powermeter Series 17

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 1 General Information Measured Parameters

Parameter Display Comm. Analog Pulse Alarm


Min/Max Frequency, Neutral current  
Phase Rotation   
Voltage and Current Phase Angles  
Day and Time  
Pulse Counters   
Digital Inputs (optional)   
Relay Outputs (optional)   
Remote Relay Control (optional) 
Alarm Triggers/Setpoints  
Self-diagnostics  

18 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Site Requirements

Chapter 2 Installation
This chapter discusses the following types of physical installations for the
PM135 Powermeter:

 Mechanical Installation
 Electrical Installation
 I/O Connections
 COM Port Connections.

2.1 Site Requirements


 Environmental conditions: as specified in Technical Specifications
in Appendix A
 Electrical requirements: as specified in Technical Specifications in
Appendix A
See Technical Specifications in Appendix A for more details

2.2 Package Contents


The PM135 Powermeter package contains the following items:

 PM135 Powermeter unit


 Technical Documentation CD
 Optional accessories (depending on the options ordered, if any)
 Cables

2.3 Mechanical Installation


Refer to the figures provided in this section to properly perform the
mechanical installation.

Figure ‎2-1. Dimensions

PM135 Powermeter Series 19

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Mechanical Installation

Panel Mounting
To mount the meter in cutout (ANSI 4" round or DIN 92x92mm
square cutout):
1. Position the meter in the cutout.
2. Affix the meter using washers and
nuts. (Add short text on Panel
Mounting, a heading should always
have text)

Figure ‎2-2. Mounting

DIN Rail Mounting


The PM135 can be mounted on a 35-mm DIN rail.

Figure ‎2-3. Dimensions

20 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

Figure ‎2-4. DIN Rail Mounting

Figure ‎2-5 PM135 with 12DI/4RO module

PM135 Powermeter Series 21

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

2.4 Electrical Installation

The equipment installation shall conform to the following


instructions:
a) a switch or circuit-breaker shall be included in the building
installation;
b) It shall be in close proximity to the equipment and within
easy reach of the OPERATOR;
c) It shall be marked as the disconnecting device for the
equipment.
Before installing, ensure that all incoming power sources
are shut OFF. Failure to observe this practice can result in
serious or even fatal injury and damage to equipment.

Typical Installation

Figure ‎2-6. Typical Installation Diagram

22 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

Terminals

Figure ‎2-7. Terminals -Rear View

Power Source Connection


The equipment installation shall conform to the following
instructions:
a) a switch or circuit-breaker shall be included in the building
installation;
b) It shall be in close proximity to the equipment and within
easy reach of the OPERATOR;
c) It shall be marked as the disconnecting device for the
equipment.
Before installing, ensure that all incoming power sources
are shut OFF. Failure to observe this practice can result in
serious or even fatal injury and damage to equipment.

The power source can be a dedicated fuse, or a monitored voltage if it is


within the instrument power supply range.
To connect an AC power supply:
1. Connect the Line wire to terminal
L/+.
2. Connect the Neutral wire to terminal
N/-.
To connect to a DC power supply:
1. Connect the positive wire to terminal
L/+
2. Connect the negative wire to
terminal N/-.

PM135 Powermeter Series 23

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

Voltage Input connection


The equipment installation shall conform to the following
instructions:
a) a switch or circuit-breaker shall be included in the building
installation;
b) It shall be in close proximity to the equipment and within
easy reach of the OPERATOR;
c) It shall be marked as the disconnecting device for the
equipment.
Before installing, ensure that all incoming power sources
are shut OFF. Failure to observe this practice can result in
serious or even fatal injury and damage to equipment.

690V Inputs (Standard)


690V Input (Standard)
400

690

400

690 690 400

Voltages
V1 V2 V3 VN
690V inputs are usually used with direct connection. Use any of the seven
wiring configurations shown in Figures 2-8 through 2-15.

120V Inputs (Option


120V U)
Input (Option U)
120

120

120

120 120 120

Voltages
V1 V2 V3 VN

120V inputs usually imply use of a potential transformer (PT). The PT


requires use of any of the four wiring configurations shown in Figures 2-7
through 2-10.

Current Input Connection


The PM135 series provide two different CT connections:
 Using internal CT, the PM135 does not have current terminals
 Using external CT (HACS – High Accuracy SATEC Current Sensor),
the PM135 provides current terminals
To connect to the external CT, pass the external CT wire through the
meter CT core, see Figure ‎2-8 for details and observe the arrow that
indicates the current direction.

24 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

In case of a retrofit application where each external CT ends with two


wires:
1. Pass one wire through the meter CT
core.
2. Connect the wire to one of the
meter termination screws.
3. Connect the second wire from the
external CT to the termination screw
to close the loop.

Figure ‎2-8. Current Input Connection

Wiring Diagrams
For AC input ratings, see Technical Specifications in Appendix A for more
details.
Table 2 presents the available wiring configurations in the meter. For
more details, see Basic Meter Setup in Chapter 5.

Table 2: Wiring Configurations

Wiring Configuration Setup Code Figure


3-wire 2-element Delta direct connection using 2 CTs 3dir2 2-8
4-wire 3-element Wye direct connection using 3 CTs 4Ln3 or 4LL3 2-9
4-wire 3-element Wye connection using 3 PTs, 3 CTs 4Ln3 or 4LL3 2-10
3-wire 2-element Open Delta connection using 2 PTs, 2 CTs 3OP2 2-11
4-wire 2½-element Wye connection using 2 PTs, 3 CTs 3Ln3 or 3LL3 2-12
3-wire 2½-element Open Delta connection using 2 PTs, 3 CTs 3OP3 2-13
4-wire 3-element Delta direct connection using 3 CTs 4Ln3 or 4LL3 2-14
3-wire 2½-element Broken Delta connection using 2 PTs, 3 CTs 3bLn3 or 3bLL3 2-15

PM135 Powermeter Series 25

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

Figure ‎2-9 3-Wire 2-Element Delta Direct Connection Using 2 CTs (Wiring Mode = 3dir2)

Figure ‎2-10 4-Wire Wye 3-Element Direct Connection Using 3 CTs (Wiring Mode = 4LL3 or 4Ln3)

26 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

Figure ‎2-11 4-Wire Wye 3-Element Connection Using 3 PTs, 3 CTs (Wiring Mode = 4LL3 or 4Ln3)

Figure ‎2-12 3-Wire 2-Element Open Delta Connection Using 2 PTs, 2 CTs (Wiring Mode = 3OP2)

PM135 Powermeter Series 27

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

This configuration provides accurate power measurements only if the voltages are balanced.

Figure ‎2-13 4-Wire Wye 2½-Element Connection Using 2 PTs, 3 CTs (Wiring Mode = 3LL3 or 3Ln3)

Figure ‎2-14 3-Wire 2½-Element Open Delta Connection Using 2 PTs, 3 CTs (Wiring Mode = 3OP3)

28 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Electrical Installation

Figure ‎2-15 4-Wire 3-Element Delta Direct Connection Using 3 CTs (Wiring Mode = 4LL3 or 4Ln3)

Figure ‎2-16 3-Wire 2½-Element Broken Delta Connection Using 2 PTs, 3 CTs (Wiring Mode = 3bLn3 or
3bLL3)

PM135 Powermeter Series 29

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation I/O Connections

2.5 I/O Connections


Before I/O Module installation ensure that all incoming
power sources are shut OFF. Failure to observe this practice
can result in serious or even fatal injury and damage to
equipment.

RS-485 Terminals

Module connector cover

Remove Module Connector cover


before assembling module

Figure ‎2-17 Module Connector Cover – Before Module Assembly

For I/O ratings, see Technical Specifications in Appendix A.

4DI/2DO Module

Figure ‎2-18 4DI/2DO Module Assembly

30 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation I/O Connections

Relay Outputs
There are two relay outputs provided for energy pulsing, alarms, or
remote control.

Figure ‎2-19 Relay Output Connection

Digital Inputs
Four optically isolated status inputs are provided for status monitoring,
pulse counting, external power demand period, and time synchronization.

Figure ‎2-20 Digital Input Connection

PM135 Powermeter Series 31

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation I/O Connections

8 DI module
Eight optically isolated status inputs are provided for status monitoring,
pulse counting, external power demand period, and time synchronization

Figure ‎2-21 8 DI - Digital Input Connection

12DI/4RO Module
The 12DI/4RO module can be equipped with optional communication port
COM2 – ETHERNET or RS-422/485.

Figure ‎2-22 12DI/4RO Module

Before I/O Module installation ensure that all incoming power


sources are shut OFF. Failure to observe this practice can
result in serious or even fatal injury and damage to equipment.

Relay Outputs
There are four electro-mechanic relay outputs provided for energy
pulsing, alarms, or remote control.

32 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation I/O Connections

Figure ‎2-23 Relay Output Connection

Digital Inputs
12 optically isolated status inputs are provided for status monitoring,
pulse counting, external power demand period, and time synchronization.

Figure ‎2-24 12 Digital Input Connection

PM135 Powermeter Series 33

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation I/O Connections

4AO Module - Analog Outputs


The 4AO module has four optically isolated analog outputs with an internal
power supply and current output options of 0-20 mA and 4-20 mA
(current loop load of up to 500 Ohm), 0-1 mA and ±1 mA (2mA 100%
overload, current loop load of up to 5 kOhm).

Figure ‎2-25: Analog Output Connection

It is recommended to connect unused Analog output


channels to Common terminal.

 The 4AO module TERMINAL is for use only with equipment


which has no live parts which are ACCESSIBLE
 The RATING of the insulation of the external equipment for
use with the 4AO module, shall comply according to
Installation Category III for insulation to be suitable for
SINGLE FAULT CONDITION
  The external equipment TERMINAL connection type is
normally terminal block for wire size 14 AWG (up to 1.5
mm2)
 The type of equipment that might be connected to the
TERMINAL is:
 Programmable Logic Controller for automation – PLC
 Digital or Analog meter

TOU module – RTC and 4 Digital Inputs


The TOU provides a battery-operated real time clock (RTC) with four
optically isolated inputs for status monitoring, time synchronization, pulse
counting, and external power demand period.

34 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation I/O Connections

Digital Inputs

Figure ‎2-26: TOU Digital Input Connection

TOU – Battery Replacement

WARNING!
Only qualified personnel familiar with the instrument and its
associated electrical equipment must perform the RTC battery
backup replacement.
To replace the CR1632 RTC battery:
1. Remove the TOU module from the
PM135 compartment
2. Open the TOU MODULE case by
applying a flat screwdriver at three
snap-in slit (1, 2 and 3), as shown
in Figure ‎2-27.
3. Remove the old battery by lifting up
the battery holder retractable tab.
4. Place the new CR1632 battery into
the battery holder while holding up
the battery holder retractable tab in
such a way that the (+) battery pole
is toward the battery holder, as
shown in Figure ‎2-27.

PM135 Powermeter Series 35

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Communications Connections

Figure ‎2-27: TOU RTC Battery Replacement

2.6 Communications Connections

Before installing the Communication Module, ensure that all


incoming power sources are shut OFF. Failure to observe this
practice can result in serious or even fatal injury and damage
to equipment.

Several communication options are available for the PM135:


 COM1: RS-485
 COM2, on optional communication module:
Ethernet 10/100BaseT
PROFIBUS DP
GPRS
RS-232 or RS-422/485
RF modem (ISM frequency)
A connection to the Ethernet connector is made through a cable adaptor
provided with your meter.
A full description of the communication protocols is found in the PM135
protocol guides that come with your meter.

 The 12DI/4RO module can be equipped with optional communication

36 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Communications Connections

port COM2 – ETHERNET or RS-422/485

COM1 RS-485 Connection

Figure ‎2-28: COM1 RS-485 2-Wire Connection

The connector is removable with three captured-wire terminals.

ETH module – COM2 Ethernet Connection

Figure ‎2-29: COM2 Ethernet Connection

PM135 Powermeter Series 37

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Communications Connections

 The ETH module TERMINAL is for use only with equipment


which has no live parts which are ACCESSIBLE
 The RATING of the insulation of the external equipment
for use with the ETH module, shall comply according to
Installation Category III for insulation to be suitable for
SINGLE FAULT CONDITION
  The external equipment TERMINAL connection type is
RJ-45
 The type of equipment that might be connected to the
TERMINAL is:
 Personal Computer – PC or LAPTOP
 10/100Base-T LAN HUB and/or Switch

PRO module – COM2 PROFIBUS Connection

Figure ‎2-30: COM2 PROFIBUS Connection

 The PRO module TERMINAL is for use only with equipment


which has no live parts which are ACCESSIBLE
 The RATING of the insulation of the external equipment for
use with the PRO module, shall comply according to
Installation Category III for insulation to be suitable for
 SINGLE FAULT CONDITION
 The external equipment TERMINAL connection type is DB9
 The type of equipment that might be connected to the
TERMINAL is:
 Programmable Logic Controller for automation – PLC

38 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Communications Connections

RS-232/422-485 module – COM2 Connection

Figure ‎2-31: COM2 RS-232 connection

Figure ‎2-32: COM2 RS-422/485 and 12DI/4RO-RS-422/485 modules connection

PM135 Powermeter Series 39

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 2 Installation Communications Connections

 The RS-232/422-485 module TERMINALS are for use only


with equipment which has no live parts which are
ACCESSIBLE
 The RATING of the insulation of the external equipment for
use with The RS-232/422-485 module, shall comply
according to Installation Category III for insulation to be
suitable for SINGLE FAULT CONDITION

 The external equipment TERMINAL connection type is
normally terminal block for wire size 14 AWG (up to 1.5
mm²) – RS-422/485 port and DB9 male-to-female cable
more than 22 AWG (0.3mm²)
 The type of equipment that might be connected to the
TERMINAL is:
 Personal Computer – PC or LAPTOP

GPRS/GSM modem module – COM2 GPRS connection


A GSM/GPRS modem module can be connected to the meter COM2 port to
provide communications with the remote MODBUS/TCP server via a
wireless GPRS network.

Figure ‎2-33: GSM/GPRS modem Module

The GSM/GPRS SIM must not have any incoming voice call.
The customer must require from the Service Provider for DATA
services only
 The GPRS modem module can be equipped with two different
antennas: internal Antenna for installation into plastic closet or
no-metallic environment. For metallic installation use external
antenna

See Setting Up GPRS Network in Chapter 5 for information on configuring


GPRS communications in your meter.

40 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Chapter 3 Using Front Display


This chapter provides PM135 Power meter series front panel information
and operating procedures.

Figure ‎3-1: PM135 Unit

3.1 Display Operations


The PM135 has a high-contrast graphical LCD display with backlight for
local data read outs, meter setup and servicing.
The display operates in two modes:
 Multi-page data display mode with Auto-Scroll
feature allows you to scroll through display screens
and pages to view various billing, instrumentation
and status data.
 Programming mode allows you to enter menu-driven
device setups for inspecting and changing factory
set meter parameters, or resetting maximum
demands, counters and device diagnostics
messages.
The display is normally updated once per second.

PM135 Powermeter Series 41

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Navigation Buttons
The PM135 is provided with five navigation buttons as described
below:
Button Operation

SELECT/ENTER

TAB – browse from submenu to next submenu or


move to required digit setup

PAGE DOWN– scrolling DOWN

PAGE UP – scrolling UP

ESCAPE

The SELECT/ENTER button - function changes depending


on the display mode:
 While navigating to submenus, pressing the
SELECT/ENTER button selects the highlighted line
menu
 In "Basic Setup" or "Device Setup" menus, while
changing device parameters pressing the
SELECT/ENTER button stores the parameter
changes into the device

The TAB button - function changes depending on the


current display:
 In display data mode – monitoring, pressing the
TAB button moves from current data display to next
data display
 For instance, in "Basic Setup" selecting "CT Primary
Current A", pressing the TAB button moves to next
digit to be set

The UP/DOWN scroll buttons - function changes


depending on the current display:
 In display menu mode – pressing the UP/DOWN
scroll buttons, navigates between each
function/parameter in menus screens
 In display data mode – pressing the UP/DOWN
scroll buttons, navigates from current display data
screen to next display data screen

The ESC/MENU button - function changes depending on the


current display:
 In display menu mode – while in sub-menu,
pressing the ESC/MENU button, moves to upper
menu screen
 In display data mode – pressing the ESC/MENU
button, moves to Main Menu screen

42 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Navigating in Data Display Mode


The PM135 provides multiple pages data displays. Your present location is
indicated upper bar as shown in the following picture. See Data Displays
for the full displays list.

Display Sub-menu Sub-menu Parameter display


name

Use UP/DOWN scroll buttons to scroll through data displays.

Status Indicators
Graphical icons on the bottom status bar give immediate meter status
indication and show the present tariff rate.
The present date and time are indicated at left on the status bar. The date
order can be changed according to local rules via the "Real Time Clock"
device setup menu.

Voltage phase
presence indicators

Present date and time No load/Power flow


direction indicator

Operating mode Present tariff rate indicator

Operating Mode Indicator


The mode indicator gives information on the load presence and shows the
direction of active power.

Indicator Icon Description


Normal mode
Test mode

Tariff Rate Indicator


The tariff rate indicator (T1 through T8) shows the currently active tariff
rate.

No Load/Power Flow Direction Indicator


The power flow indicator gives information on the load presence and
shows the direction of active power.

Indicator Description
Icon
No load.

PM135 Powermeter Series 43

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Direct active power flow – delivered active energy.


Reversed active power flow – received active energy.

Phase Presence Indicators


Phase presence indicators “123” show the status of either V1-V2-V3
phase-to-neutral voltages in line-to-neutral wiring modes, or V12-V23-
V31 phase-to-phase voltages in line-to-line wiring modes.

Indicator Description
Icon
123 All voltages are present and above the voltage dip threshold.
Blinking phase indicator – the phase voltage is below the
1 3 defined voltage dip threshold. Possibly indicates an incorrect
meter nominal voltage setting (see Basic Meter Setup in
123 Chapter 5).
Dashed phase indicator - the phase voltage is either missing or
1–3 below the voltage interruption threshold.

If the phase voltage is below the defined voltage dip threshold, its
corresponding phase indicator is blinking.
If the phase voltage is either missing or below the voltage interruption
threshold, the phase indicator is replaced with a dash

Display Features
While energizing the device the display will show an init screen followed
by phase-voltages measurements as described below:

The PM135 display has a number of programmable features that can be


disabled, enabled and adjusted via the meter Display Setup (see Display
Setup in Chapter 5).
Backlight
A short press on any button while the display backlight is off switches the
backlight on.
The backlight stays on as long as you selected in the display setup and
then dims to conserve power. The backlight time is factory set to 1 minute
and can be programmed from 1 to 10 minutes. You can temporarily set
the backlight to continuous operation if you need to work in dark for more
time.
Contrast
The contrast can be adjusted via the meter Display Setup (see Display Setup
in Chapter 5).
Auto-Return
If the Auto-Return feature is enabled and no button is pressed for the
programmable Auto-Return interval (1 to 30 minutes for data displays;

44 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

fixed at 5 minutes for setup menus), the display automatically returns to


the default page from any other data display or a setup menu.
If the Auto-Scroll feature is enabled, the display immediately enters the
auto scroll sequence.
Auto-Scroll
If the Auto-Scroll feature is enabled, the data display automatically scrolls
through all pages of all data displays that are included into the
programmable auto-scroll sequence. The scroll interval is adjustable in the
range of 2 to 30 seconds. The scroll sequence may include all or only
selected displays.
The display automatically enters auto scrolling if no button is pressed for
the Auto-Return interval when the Auto-Return feature is enabled or in 1
minute if this feature is disabled. In the last case, the scroll sequence is
restored from the point where it was interrupted.
To stop auto scrolling, press briefly any button if the backlight is on; else
press briefly any button twice since the first press only sets the backlight
on and does not affect auto scrolling.
Auto-Scroll is not operational in TEST mode.

Measurement Units
The PM135 has a selectable resolution for volts, amps and powers
presented on the front display and via communications. See Device
Options in Chapter 5 for information on selecting the data resolution in
the PM135 .
Low Resolution Option
Currents are displayed in whole amperes below 10,000 A, and in kilo
amperes above 10,000 A.
Measurement units for voltage and power depend on the voltage
connection scheme:
 For direct wiring (PT=1) or wiring via PT with the PT ratio up to
and including 4.0, voltages are displayed in volts, and power in
kilowatts.
 For the PT ratio above 4.0, voltages are displayed in kilovolts with
three decimal places, and power in megawatts with three decimal
places.
High Resolution Option
Currents are displayed in amperes with up to two decimal places below
10,000 A, and in kilo amperes above 10,000 A.
Measurement units for voltage and power depend on the voltage
connection scheme:
 When direct wiring is used (PT=1), voltages are displayed in volts
with one decimal place, and power in kilowatts with three decimal
places.
 When wiring via PT is used with the PT ratio up to and including
4.0, voltages are displayed in volts, and power in whole kilowatts.
 For the PT ratio above 4.0, voltages are displayed in kilovolts with
three decimal places, and power in megawatts with three decimal
places.
The small round “Kilo” and “Mega” LEDs light up showing the appropriate
measurement units for a displayed page.

PM135 Powermeter Series 45

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Primary and Secondary Volts


Volts can be displayed in primary (default) or secondary units. The volts
display mode can be changed through the display setup (see Configuring
the Display).

Phase Power Readings


In configurations with the neutral wire, in addition to total three-phase
powers, the meter can show per-phase power readings. By default, they
are disabled. See Configuring the Display on how to enable per-phase
power readings in your meter.

Fundamental Component
The PM135EH can display total displacement power factor and active
power for the fundamental component if it is enabled through the display
setup (see Configuring the Display).
When phase power readings are allowed, the PM135EH also displays per-
phase displacement power factor and active power for the fundamental
component.

3.2 Data Displays


The PM135 has 8 multi-page data displays listed in the following table.

Data Display Display Label Display Contents


Menu
Sequence
1 Measurements Present 7 Measurements Data Displays:
V, I, P, S, Q, PF, Max. DMD, MIN/MAX,
Energy and Billing/TOU
2 Harmonics V/I individual harmonics and Harmonics
Spectrum
3 Waveforms Vn/In Waveforms with THD value
4 Phasor Voltage and current phasors diagram
5 Digital I/O Status of counters, Digital IN and relays
out
6 Device Control Diagnostics and Alarms displays
7 Basic Settings Device configuration display
8 Device Info Device model, identification, firmwares
version, COM1,COM2

TEST Mode Data Display


The TEST data display is shown in TEST mode in place of the billing period
data displays.

TEST mode display:


test LED pulse rate in secondary Wh/imp;
test kWh and kvarh energy readings in primary units
with an extended 0.001 kWh resolution.
TEST mode setup menu is enabled in Device
Setup/Device Options menu.

46 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Billing Period Data Displays


The PM135 provides billing period data displays for energy and general
purpose volume data as m³, cf or Ccf calculated using Digital Input for
water and/or gas meter application.
Only registers you selected in the billing/TOU register setup and tariff
rates listed in the TOU daily profiles are included (see Configuring
Billing/Tariff Registers and Configuring the Daily Tariff Schedule in
Chapter 5).

The following example demonstrates the present billing period displays for
two configured billing registers (kWh imported and kvarh imported) and
for three active tariff rates. The actual register contents in your
installation may be different depending on your selection of register
sources.

Each billing period display lists all total and tariff energy, maximum
demand and cumulative maximum demand registers for all configured
billing registers and all active tariffs. Use the UP/DOWN button to scroll to
the desired Billing/TOU period register display

Register 1 - total readings:


Total Import Active energy data.
Billing period according to TOU predefined profile
(Reg1 for TOU actve energy and Reg2 for TOU
reactive energy are predefined TOU/Register factory
setup, can be changed by user, see Configuring
Billing/Tariff Registers and Configuring the Daily
Tariff Schedule in Chapter 5)

Register 1 – tariff 1 readings:


Total Import Active energy data.

Register 1 – tariff 2 readings:


Total Import Active energy data.

PM135 Powermeter Series 47

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Register 1 – tariff 3 readings:


Total Import Active energy data.

Register 2 - total readings:


Total Export Active energy data.

Register 2 – tariff 1 readings:


Total Export Active energy data.

Register 2 – tariff 2 readings:


Total Export Active energy data.

Register 2 – tariff 3 readings:


Total Export Active energy data.

Register 3 - total readings:


Total Import Reactive energy data

48 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Register 3 – tariff 1 readings:


Total Import Reactive energy data.

Register 3 – tariff 2 readings:


Total Import Reactive energy data.

Register 3 – tariff 3 readings:


Total Import Reactive energy data.

Register 4 - total readings:


Total Export Reactive energy data

Register 4 – tariff 1 readings:


Total Export Reactive energy data.

Register 4 – tariff 2 readings:


Total Export Reactive energy data.

PM135 Powermeter Series 49

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Register 4 – tariff 3 readings:


Total Export Reactive energy data.

Measurements Maximum Demand Data Display


Maximum demand displays show measurements maximum demands (not
billing maximum demands) for powers, voltages, currents and total
harmonics. Each quantity is displayed with the date and time of the last
update. Use the UP/DOWN button to scroll to the desired Max. DMD data
display

Import kW maximum demand


Import (inductive) kvar maximum demand
Import kVA maximum demand

Export kW maximum demand


Export (inductive) kvar maximum demand
Export kVA maximum demand

V1-V3 maximum demand


Indicate V12-V31 voltage in line-to-line
configurations.

I1-I3 and In maximum demand

50 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Measurements Data Display


Measurements data represent general instrumentation data you can use
while installation and inspecting the meter. Use phase angles displays to
check the order of phases when connecting wires to the meter terminals.
Use the UP/DOWN button to scroll to the desired data display

Phase-to-neutral voltages. Only displayed in 4-wire


configurations with a neutral.

Phase-to-phase voltages

Phase-to-neutral voltages maximum demand with


time stamp

Phase voltage/current angle

Line frequency

PM135 Powermeter Series 51

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Voltage unbalance, %

Phase voltage THD


Indicate phase-to-phase voltage THD in line-to-line
configurations

Phase and neutral currents

Phase and neutral currents maximum demand with


time stamp

Current unbalance, %

Phase Current Total Demand Distorsion, %

52 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Phase Current THD, %

Total powers and power factor

Total powers maximum demand export

Total powers maximum demand import

Phase 1 powers and power factor

Phase 2 powers and power factor

PM135 Powermeter Series 53

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Phase 3 powers and power factor

Measurements Minimum/Maximum Data Display


Minimum/Maximum displays show measurements minimum/maximum for
powers, voltages, currents, power factor and frequency. Each quantity is
displayed with the date and time of the last update. Use the UP/DOWN
button to scroll to the desired Min/Max data display

Phase currents minimum/maximum values with


time stamp.

Neutral current minimum/maximum values with


time stamp.

Phase-to-neutral voltages minimum/maximum


values with time stamp.

Phase-to-phase voltages minimum/maximum values


with time stamp.

54 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Total powers minimum/maximum values with time


stamp.

Total power factor minimum/maximum value with


time stamp.

Line frequency minimum/maximum value with time


stamp.

Measurements Energy Data Display


Energy displays show measurements total import/export for energy and
phases energy. Use the UP/DOWN button to scroll to the desired energy
data display

Total import energy.

Total export energy.

PM135 Powermeter Series 55

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Phase 1 import energy.

Phase 2 import energy.

Phase 3 import energy.

Harmonics Display
Harmonics display shows individual harmonics distortion for phase
voltages and current, and phase voltage/current harmonic spectrum
graphs. Use the UP/DOWN button to scroll to the specific harmonic
number or voltage and current channels harmonic spectrum.

Individual harmonics phase voltages and currents,


%

Per-phase current harmonic spectrum I1-I3

56 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Waveform Display
The waveform display shows per-phase voltage and current waveforms
and V/I values + THD. Use the UP/DOWN button to scroll through the
phases.

Per-phase voltage and current waveforms

Phasor Display
The Phasor display shows a three-phase network Phasor diagram. All
phase angles are given relatively to the V1 channel.

Three-phase voltage and current Phasor diagram

Digital I/O
The Digital I/O display shows three sub-menus: "Digital Inputs" display to
show status of each digital input, "Pulse/Event Counters" display to count
external pulses or setpoint events, or as time counters to count setpoint
operation time, "Relay Outputs" display to show status of each relay
output. Use the UP/DOWN button to move from one display to another.

Digital Inputs status

external pulses or setpoint events counter, or as


time counters to count setpoint operation time

PM135 Powermeter Series 57

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Relay outputs status

Device Control Display


The device control display shows two sub-menus: "Diagnostics" and
Alarms". The diagnostics display shows device diagnostic messages
recorded as a result of the meter self-test diagnostics during start-up and
operation, the alarm display shows a list of operated alarm setpoints
along with the alarm trigger labels if there are alarms recorded during
meter operation.

Device Control sub-menu, use the UP/DOWN button


to select whether Diagnostics or Alarms displays

List of diagnostics messages

List of alarms messages

If there are diagnostic messages, the i diagnostic green led on the


device panel flashes until you clear the device diagnostics. Some of the
diagnostics events are cleared automatically as the event source
disappears. See Device Diagnostic Codes in Appendix H for a full list of
diagnostic messages and their meanings. See Clearing Device Diagnostics
for information on how to clear the device diagnostics from the display
and via PAS.
The diagnostic Led indication can be disabled or enabled via the Display
Setup menu.

58 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Basic Settings Display


The basic settings display shows basic device settings that can be required
for immediate inspecting while meter testing and at the time of
installation. Use the UP/DOWN button to scroll through the settings

Device wiring mode (see Basic Meter Setup for full


list of wiring modes), external potential transformer
ratio and the nominal device voltage.
Primary ratings of the external current transformers
(main and auxiliary current inputs) and the nominal
device frequency
Power demand period (number of blocks x block
demand period), and voltage and ampere demand
periods.

Device Info Display


The device info display provides different service information that may be
required for meter identification and inspection, like product/module and
firmware information, I/O module type, communication settings, and so
on. Use the UP/DOWN button to scroll through the device info.

Meter identification info

Meter communication info

Load Bar Graph


The load bar graph displays the amount, in percent (40% to 110%), of
the present current load with respect to user-defined nominal load
current. The reference nominal current can be set up in amps through the
display setup (see Configuring the Display). If it is set to 0 (default), the
current load is referenced to the specified CT primary current.

PM135 Powermeter Series 59

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Display Operations

Energy Pulse LED


The PM135E and PM135EH have a red “Energy Pulse” LED. It flashes at a
constant rate when a load is applied to the meter.
There are two modes of LED operation:
 NORMAL mode: the LED pulses indicate imported Wh at a rate
of 1,000 pulses per kWh
 TEST mode: the LED pulses indicate either imported Wh, or
imported (inductive) varh at a rate of 10,000 pulses per
kWh/kvarh
The energy test mode can be enabled through the Device Options setup.
When in test mode, the energy and demand accumulators do not account
for consumed energy.

Port Activity LEDs


The meter has two green LEDs “RX” and “TX”, which indicate activity on
the COM1 communication port. The LEDs flash when the port is receiving
or transmitting data.

60 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Device Setup

3.3 Device Setup


The PM135 setup is menu-driven. The device provides 12 menus that
allow local accessing a limited number of meter setups and control
functions listed in the following table. Access to particular menus is
granted depending on the password you entered if enabled.

Menu Label Menu Function

Reset Reset of engineering maximum demands,


device diagnostics, meter and battery
operation time counters and failure counters
Real Rime Clock RTC clock setup
Display Setup Display setup
Basic Setup Basic device setup
Device options Device options setup
COM1 Setup COM1 serial port setup
Local Settings Local settings
Setpoint Setup
Password Setup Meter passwords setup

Entering the Password


The Setup Change menu can be secured by a four-digit user password.
You can change the password and enable password protection through the
Access Control menu (see Configuring Meter Security). The meter is
primarily shipped with the password preset to 0 and password protection
disabled.
If password protection is enabled, you are prompted for a password when
entering the setup change menu.

To enter the password:


1. Select the desired digit field using the TAB button
2. Select the desired digit using the
UP/DOWN button
3. Press ENTER to confirm the
password.
If the password you entered is correct, you are moved to the Main menu,
otherwise you return back to the Device Setup menu.

PM135 Powermeter Series 61

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 3 Using Front Display Device Setup

Viewing and Changing Setup Options


Once you entered a correct password you are moved to the Device Setup
menu.
The Device Setup menu consists of sub-menus list.
To select a desired menu entry from the menu list:
Device Setup
Reset
Use the UP/DOWN button - to scroll through the menu list to
Real Time Clock
Display Setup the desired menu entry
Basic Setup
Device Options
COM1 Setup
COM2 Setup
Press the SELECT/ENTER button - to enter the selected submenu.
Parameters that are represented by values can be changed in two ways:
Device Setup
Reset
Real Time Clock
Display Setup  By pressing the SELECT/ENTER button - , a new sub-menu
Basic Setup
Device Options appears by presenting possible values to be selected
COM1 Setup
COM2 Setup

 By pressing the TAB button - , the highlight cursor moves to


Display Setup the actual value to be changed
Auto Scroll 5
Auto Return 5
Backlight 1
Diagnostics LED Disabled Use the UP/DOWN button - to scroll through the desired value,
Phase Powers Disabled
Fundamental Powers Disabled
Contrast 15
then press the SELECT/ENTER button - to store the selected value

Display Setup Display Setup


Auto Scroll Auto Scroll Disabled
Disabled Auto Return 5 Note: While being in the Device Setup operation mode, the PM135 display
2
3
Backlight
Diagnostics LED
1
Disabled will return to normal operation, i.e. move to data display, during idle
4
5
Phase Powers
Fundamental Powers
Disabled
Disabled operation of one minute
6 Contrast 15

62 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Installing PAS

Chapter 4 Using PAS Software


The support PAS software is a configuration and data acquisition tool that
allows you to configure all of the PM135 features, monitor your meters
on-line, retrieve recorded files and view reports. PAS can communicate
with your PM135 via a serial port and via the Ethernet.
This chapter gives information on how to install and run PAS on your
computer, and how to prepare information for your meter using PAS.
See Chapter 5 Configuring the PM135 for instructions on how to configure
particular features in your meter. Refer to Chapters 7 and 8 for
instructions on retrieving data from the meters and viewing reports.

4.1 Installing PAS


You need PAS V1.4 Build 4 or higher to take an advantage of the meter
data logging options.
To install PAS on your PC:
1. Insert the installation CD supplied
with your meter into CD drive.
2. Open My Computer on your
Desktop.
3. Click on your CD drive icon, select
the PAS directory, and then double
click on Setup (shown as an
Application type file).
4. Follow InstallShield® Wizard
instructions on the screen.
PAS is installed by default to the C:\Pas folder.
When installation is complete, the PAS icon appears on your Desktop.
Double click on the PAS icon to run PAS.
For general information on how to work with PAS, see the “PAS Getting
Started” guide supplied on the installation CD.

4.2 Creating a New Site for your Meter


PAS keeps all communication and configuration data for your meter in a
configuration database called a site database. During configuration, store
all setup data to the site database so that PAS recognizes device
properties regardless of whether the meter is online or offline.
To communicate with the meters, create a separate site database for each
device.
To create a new database for your meter:
1. Select Configuration from the Tools menu.

PM135 Powermeter Series 63

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Setting up Communications

Figure ‎4-1: Configuration Dialog Box – Instrument Setup Tab

2. Click the Sites button on the right-hand-side.

3. From the Look in box, select the directory where a


new database will be stored. By default, it is the Sites
directory.
4. Type a site name for your device in the File name
box, click New, and then click OK.
5. On the Instrument Setup tab, select PM135 in the
Model box. PAS automatically selects the appropriate
instrument options for your meter.
6. Select a correct CT secondary current (5A or 1A) for
your meter.
7. If you wish to add any comments for your meter, type
the comments in the Comment box.

4.3 Setting up Communications


You can communicate with the PM135 via a PC RS-232 serial port or
through the Internet.
To configure communications with the PM135:

64 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Setting up Communications

1. Select Configuration from the Tools menu. Under the


Communication group on the Instrument Setup tab,
select the type of connection for your device.
2. Set the device communication address you assigned to
the PM135 port. When communicating via the Ethernet,
the PM135 responds to any address you select.
3. In the Sampling Rate box, select a rate at which PAS
updates data on your screen when you continuously poll
the device in the PAS Data Monitor.
The communication protocol and port settings must match the settings
you made in your meter.

Communicating through a Serial Port


Select Serial Port/Modem Site on the Configuration tab, and then
click on the Connection tab to configure your serial port settings.
Configuring a Serial Port
1. On the Connection tab, select a COM port from the Device
box, and then click Configure.

Figure ‎4-2: Serial Port Setup Dialog Box

2. Specify the baud rate and data format for the port. Choose
the same baud rate and data format as you have set in the
meter, and then click OK.
The factory settings for the local PM135 RS-232 and RS-422/485 ports
are 9600 baud, 8 bits with no parity.
Selecting the Communications Protocol
1. On the Connection tab, click
Protocol.

PM135 Powermeter Series 65

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Setting up Communications

Figure ‎4-3: Protocol Setup Dialog Box

2. In the Protocol box, select the


same communications protocol as
you have set in your meter.
3. In the Response Timeout box,
define the maximum time that PAS
should wait for the meter response
before announcing a failure.
4. In the Break Timeout box, define
the maximum line idle time that PAS
should wait after receiving the last
message character before closing a
connection with the Modbus RTU or
DNP3 protocol. It does not affect
ASCII communications. Note that
this time is added to the message
transfer time, and excessive
increasing it may slow down
communications. If you frequently
receive the “Communication error”
message, try to increase Break
Timeout.
5. In the Retries box, define the
number of attempts that PAS should
use to receive a response from the
meter in the event the
communication fails, before
announcing a communication
failure.

Communicating through the Internet


If you are communicating through the Ethernet port, define the IP address
of your meter on the network.
To configure the meter IP address:
1. On the Instrument Setup tab, select
Internet Site.
2. Click on the Connection tab.
3. Click on the IP address and type in
the IP address of your meter. The
default IP address preset in the
meter at the factory is
192.168.0.203.
4. In the Protocol box, select the
communications protocol for the TCP
port. The meter provides
Modbus/TCP connections on TCP
port 502 and DNP3/TCP connections
on port 20000. The host port is set
automatically as you select the
protocol. Select Modbus RTU/TCP
for Modbus/TCP or DNP3 for
DNP3/TCP.
5. In the Wait for answer box, adjust
the time that PAS waits for a

66 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Setting Up the Meter

connection before announcing an


error.

Figure ‎4-4: Configuration Dialog Box – Connection Tab

6. In the Retries box, specify the


number of retries PAS will use to
receive a response from the meter if
communications fail.
7. Click OK.

4.4 Setting Up the Meter

Preparing Setups for the Meter


PAS allows you to prepare setup data for your meter off-line without the
need to have it connected to your PC.
To prepare a setup for your meter:
1. Select the device site from the list
box on the PAS toolbar.
2. Select the desired setup group from
the Meter Setup menu. Click on
the tab with the setup you want to
create or modify.
3. Fill in the boxes with the desired
configuration data for your meter.
4. Click the Save as… button to store
the data to the meter site database.
5. Click OK.

Always set up and store the Basic Setup data to the site
database first. PAS uses this data as a reference when
arranging other meter setups.

To save your setup to another site database:


1. Click the Save as… button.

PM135 Powermeter Series 67

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Setting Up the Meter

2. Select the target database from the


file pane.
3. Click OK.
You can also reuse a setup from another site by copying it to your present
site database.
To copy a setup from another site’s database:
1. Click Open.
2. Select the desired source site
database.
3. Click OK. The opened setup is
copied to your dialog window.
4. Click the Save as… button.
5. Select the target database from the
file pane.
6. Click OK.
To copy all setups from one site database to another site's
database:
1. In the list box on the toolbar, select
a source device site from which you
wish to copy setups.
2. Select Copy to... from the Meter
Setup menu.
3. Select the target site database to
which you wish to copy setups, and
click OK.

Downloading Setups to the Meter


You can update each setup in your meter one at a time or download all
setups together from the site database.
Individual Download
To update a particular setup in your meter:
1. Check the On-line button on the
PAS toolbar
2. Select a meter site from the list box
on the toolbar.
3. Select the desired setup group from
the Meter Setup menu. Click on
the setup tab you want to download
to the meter. As the setup dialog
box opens, PAS retrieves and
displays the present meter setup
data.
4. If you wish to download a setup
saved in the site database, click
Open, and then click OK, or fill in
the boxes with the desired
configuration data for your device.
5. Click Send.

68 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Authorization

Batch Download
To download all setups to your device at once:
1. Check the On-line button on the
PAS toolbar
2. Select the device site from the list
box on the toolbar.
3. Select Download Setups from the
Meter Setup menu.

Uploading Setups from the Meter


Individual Upload
To get a particular setup from your device:
1. Check the On-line button on the
PAS toolbar.
2. Select a meter site from the list box
on the toolbar, and then select the
desired setup group from the Meter
Setup menu.
3. Click on the tab of the setup you
want to read from the meter. As the
dialog box opens, PAS retrieves and
displays the present setup data from
the meter. Click Receive if you wish
to retrieve the meter setup once
again.
4. To store the setup to the meter site
database, click Save As, and then
click OK.
Batch Upload
To upload all setups from the device to the site database at once:
1. Check the On-line button on the
toolbar.
2. Select the device site from the list
box on the toolbar.
3. Select Upload Setups from the
Meter Setup menu.

4.5 Authorization
If communications with your meter is secured, you are prompted for the
password when you send new setup data to the meter.

Figure ‎4-5: Authorization Dialog Box

PM135 Powermeter Series 69

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 4 Using PAS Software Authorization

Enter the password and click OK.


If your authorization was successful, you are not prompted for the
password again until you close the dialog box.
See Configuring Meter Security in Chapter 5 for more information on the
meter password security.

70 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

Chapter 5 Configuring the PM135


This chapter describes how to configure different options in your meter
from the front panel display or via PAS.

5.1 Configuring Communications

Setting Up Serial Communication Ports


Using the Front Display
Select COM1 Setup or COM2 Setup from the Device Setup menu. See
Viewing and Changing Setup Options in Chapter 3 for information on
configuring parameters via the front display.
Device Setup
Reset
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup

See below for available communication options.

Using PAS
Select Communications Setup from the Meter Setup menu, and then
click on the Serial Ports Setup tab. In the Port box, select the desired
device port.

PM135 Powermeter Series 71

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

Figure ‎5-1: Communication Setup Dialog Box – Serial Ports Setup Tab

See Table 3 below for available communication options.

Table 3: COM Port Options


Parameter Options Default Description
Protocol MODBUS ASCII MODBUS The communications
SATEC ASCII RTU protocol for the port
MODBUS RTU Not changeable on COM2
DNP3 with the PROFIBUS
IEC 60870-5 module
PROFIBUS DP (COM2)
Interface COM1: 485 = RS-485 RS-485 Communication interface.
COM2: Not changeable on COM2
232 = RS-232 with the Ethernet and
485 = RS-485 PROFIBUS modules
422 = RS-422
Eth = Ethernet
PRO = PROFIBUS DP
GPRS = GPRS modem
Device address SATEC ASCII: 0-99 1 Device network address
MODBUS: 1-247 (126 for
DNP3: 0-65532 PROFIBUS)
PROFIBUS: 0-126
Baud rate 300 bps-115.2 kbps 9600 bps The port baud rate
Data format 7E, 8N, 8E 8N Data format and parity.
7E data format should not
be used with the MODBUS
RTU and DNP3 protocols
Response delay 0-1000 ms 5 ms The minimum time after
the last request character
is received to start the
transmission.

NOTES:
1. The meter provides the permanent
MODBUS TCP server on port 502.

72 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

2. Selecting the DNP3 protocol


launches the second DNP3 TCP
server in addition to the MODBUS
server allowing simultaneous
connections on both ports. Selecting
the MODBUS protocol disables the
DNP3 TCP server.
3. When you change the COM2
settings through the Ethernet port,
the device port restarts so
communications will be temporarily
lost. You may need to wait some
additional time until PAS restores a
connection with your device.

Setting Up Ethernet
Using the Front Display
Select Network Setup from the Device Setup menu. This menu entry
appears only if the optional Ethernet module is plugged into the
meter. It allows you to set up the meter network address and the default
gateway.
See Viewing and Changing Setup Options in Chapter 3 for information on
configuring parameters via the front display.
Device Setup
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup
Network Setup

Using PAS
Select Communications Setup from the Meter Setup menu, and then
click on the Network Setup tab.

PM135 Powermeter Series 73

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

Figure ‎5-2: Communication Setup Dialog Box – Network Setup Tab

The table below lists available network options.


Table 4: Ethernet Setup Options
Parameter Options Default
Device IP Address 192.168.0.203
Network Subnet Mask 255.255.255.0
Network Default Gateway 192.168.0.1
TCP Service Port 502 = Modbus/TCP 502
20000 = DNP3/TCP

NOTES
1. The meter provides the permanent
MODBUS TCP server on port 502.
2. Selecting the DNP3 TCP service port
launches the second DNP3 TCP
server allowing simultaneous
connections on both TCP ports.
Selecting the MODBUS TCP port
disables the DNP3 TCP server.

The TCP service port can also be changed trough the COM2
serial port setup. Changing the communication protocol for the
port automatically changes the TCP port for the Ethernet.

3. When you change the device


network settings through the
Ethernet port, the device port
restarts so communication will be
temporarily lost. You may need to
wait some additional time until PAS
restores a connection with your
device.

Setting Up GPRS Network


The PM135 can provide wireless GPRS communications with the remote
MODBUS/TCP server via the GPRS modem module. See Connecting a

74 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

GSM/GPRS modem in Chapter 2 on how to connect a modem to your


meter.

Using the Front Display


Select GPRS interface from the COM2 Setup menu. This menu entry
appears only if the optional GPRS module is plugged into the
meter. It allows you to set up the meter network address and the default
gateway.
See Viewing and Changing Setup Options in Chapter 3 for information on
configuring parameters via the front display
Device Setup
Reset
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup

COM2 Setup
Protocol Modbus RTU
Interface GPRS
Device Address 1

Using PAS
To set up GPRS communications:
1. Select Communications Setup
from the Meter Setup menu, and
then click on the GPRS Setup tab.

Figure ‎5-3: Communication Setup Dialog Box – GPRS Setup Tab

PM135 Powermeter Series 75

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

The following table lists available GPRS options.

Table 5: GPRS Setup Options


Parameter Default Description
Access Point Name (APN) internetg The mobile network APN name
User name Username (if required)
Password Password (if required)

2. Configure your mobile network APN,


username and password. Consult
your network operator regarding
proper network settings. Leave the
username and password fields blank
if network authorization is not
required.
3. Send your GPRS settings to the
meter.
4. Select the GPRS interface in the COM2 port setup (see Setting Up
Serial Communication Ports).
5. Configure your eXpertPower client
(see Setting Up eXpertPower Client)
or/and TCP Notification client (see
Setting Up TCP Notification Client)
for communicating with a remote
server.
You can check the status of the GPRS communications from the front
panel via the Status Display or via the Device Control dialog in PAS (see
Viewing Communication Status and Statistics).

Setting Up eXpertPower Client


The PM135 has an embedded eXpertPowerTM client that provides
communications with the eXpertPowerTM server – the SATEC proprietary
Internet services. Connections to the eXpertPowerTM server are handled
on a periodic basis.
To set up communications with the eXpertPowerTM server, select
Communication Setup from the Meter Setup menu, and then click on
the ExpertPower Client Setup tab.

76 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

Figure ‎5-4: eXpertPower Client Setup Tab

The following table lists available options. Refer to your eXpertPower


service provider for the correct network settings.

Table 6: eXpertPower Client Setup Options


Parameter Options Default Description
XPW Server IP Address 207.232.60.18 The IP address of the
eXpertPower server
XPW Server Port 0-65535 5001 The TCP service port of the
eXpertPower server
XPW Client Enabled NO, YES NO Enables operations of the
eXpertPower client
Time to Next Session, 1-99999 The time remaining to the
min next connection session

NOTES
6. Do not enable the eXpertPower client in your meter if you do not use
the eXpertPowerTM service.
7. Do not change the connection time
setting. It is for information only.
The eXpertPower server updates it
automatically.

Setting Up TCP Notification Client


The TCP notification client can establish connections with a remote
MODBUS/TCP server and send notification messages either on events, or
periodically on a time basis.
To set up communications with a remote TCP Notification server, select
Communication Setup from the Meter Setup menu, and then click on
the TCP Notification Client Setup tab.

PM135 Powermeter Series 77

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communications

Figure ‎5-5: TCP Notification Client Setup Tab

The following table lists available client options.

Table 7: TCP Notification Client Setup Options


Parameter Options Default Description
Client Enabled NO, YES NO Enables operations of the
notification client
Server IP Address 192.168.0.3 The IP address of the notification
server
Server Port 0-65535 502 The TCP service port of the
notification server
Message Exchange 0-65535 1000 The start address of a block
Address of 16 MODBUS registers for
receiving notification messages

Connections with a remote server are triggered via programmable


setpoints. To send event notifications to a server, configure a setpoint to
respond to desired triggers or to periodic time events and put the
"Notification" action to the setpoint action list (see Configuring
Alarm/Control Setpoints).
See the PM135 Modbus Reference guide for more information on operation
of the notification client and the notification message structure.

78 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

5.2 General Meter Setup

Basic Meter Setup


This section describes how to configure the PM135 for your particular
environment and application.
Before operating your meter, provide the device with basic information
about your electrical network.

Using the Front Display


Select the Basic Setup entry from the Device Setup menu. See Viewing
and Changing Setup Options in Chapter 3 for information on configuring
parameters via the front display.
Device Setup
Reset
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup

Using PAS
Select General Setup from the Meter Setup menu. See the table below
for the Basic Setup options.

Table 8: Basic Setup Options


Parameter Options Default Description
Basic Configuration
Wiring Mode See Table 9 4Ln3 The wiring connection of the device
PT Ratio 1.0-6500.0 1.0 The phase potential transformer’s
primary to secondary ratio
PT Ratio Multiplier ×1, ×10 ×1 PT Ratio multiplication factor. Used
in extra high voltage networks to
accommodate the PT ratio for
500 kV and higher networks.
CT Primary Current 1-50000 A 5A The primary rating of the phase
current transformer
Nominal Frequency 50,60,25,400 Hz 60 Hz The nominal line frequency
Maximum Demand 0-50000 A 0 The maximum demand load
Load Current current (0 = CT primary current)
Power block demand 1, 2, 3, 5, 10, 15, 15 min The length of the demand period
period E, EH 20, 30, 60 min, for power demand calculations. If
E=external sync the external synchronization is
selected, a pulse front on the
digital input DI1 denotes the start
of the demand interval.
The number of 1-15 1 The number of blocks to be
blocks in the sliding averaged for sliding window
demand window E, EH demands

PM135 Powermeter Series 79

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Parameter Options Default Description


Volt/Ampere 0-1800 sec 900 sec The length of the demand period
Demand Period for ampere and volt demand
calculations

1. Always specify the wiring mode and transformer ratings prior to


setting up setpoints and analog outputs.
2. The maximum value for the product of the phase CT primary
current and PT ratio is 57,500,000. If the product is greater, power
readings are zeroed.

Figure ‎5-6: General Setup Dialog Box – Basic Setup Tab

Table 9 lists the available wiring modes.

Table 9: Wiring Modes


Wiring Mode Description
3OP2 3-wire Open Delta using 2 CTs (2 element)
4LN3 4-wire Wye using 3 PTs (3 element), line-to-neutral voltage readings
3DIR2 3-wire Delta Direct Connection using 2 CTs (2 element)
4LL3 4-wire Wye using 3 PTs (3 element), line-to-line voltage readings
3OP3 3-wire Open Delta using 3 CTs (2½ element)
3LN3 4-wire Wye using 2 PTs (2½ element), line-to-neutral voltage
readings
3LL3 4-wire Wye using 2 PTs (2½ element), line-to-line voltage readings
3BLN3 3-wire Broken Delta using 2 PTs, 3 CTs (2½ element), line-to-neutral
voltage readings
3BLL3 3-wire Broken Delta using 2 PTs, 3 CTs (2½ element), line-to-line
voltage readings

 In 4LN3, 3LN3 and 3BLN3 wiring modes, Min/Max volts, volt


demands and voltage harmonics represent line-to-neutral

80 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

voltages; otherwise, they will be line-to-line voltages.

Transformer Correction
While using external CT meter option, transformer correction allows you
to compensate ratio and phase angle inaccuracies of the user voltage and
current instrument transformers.
The following table lists available options.
Label Parameter Options Default Description
Ratio V1-V3 V1-V3 voltage 0.700 to 1.300 1.000 The ratio of the true
transformer transformer ratio to the
ratio correction marked ratio.
factor
Angle V1-V3 V1-V3 -600 to 600 0 The phase
transformer displacement, in
phase angle minutes, between the
error, minutes primary and secondary
values. The phase angle
of a voltage transformer
is positive when the
secondary value leads
the primary value.
Ratio I1-I3 I1-I3 current 0.700 to 1.300 1.000 The ratio of the true
transformer transformer ratio to the
ratio correction marked ratio.
factor
Angle I1-I3 I1-I3 -600 to 600 0 The phase
transformer displacement, in
phase angle minutes, between the
error, minutes primary and secondary
values. The phase angle
of a current transformer
is positive when the
secondary value leads
the primary value.

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Transformer Correction tab.

PM135 Powermeter Series 81

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Device Options
The Device Options setup allows changing user-configurable device
options or putting the meter into energy test mode.

Using the Front Display


Select Device Options from the Device Setup menu. See Viewing and
Changing Setup Options in Chapter 3 for information on configuring
parameters via the front display.
Device Setup
Reset
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Device Options tab.
Table 10 lists available device options.

Table 10: User-configurable Device Options


Parameter Options Default Description
Power Calculation Reac = using reactive S=f(P, Q) The method used for
Mode power S=f(P, Q), calculating reactive and
Nact = using non-active apparent powers (see Power
power Q=f(S, P) Calculation Modes below)
Energy Roll Value 10000 kWh 10000000 The value at which energy
E, EH 100000 kWh counters roll over to zero
1000000 kWh
10000000 kWh
100000000 kWh
1000000000 kWh
Phase Energy Disabled Disabled Enables phase energy
Calculation E, EH Enabled calculations
Energy Test Mode Disabled Disabled Setting this option puts the
E, EH Wh pulses meter into the energy test
varh pulses mode (see Energy Pulse LED
in Chapter 3)
Starting Voltage 1.5-5.0% 1.5% The device starting voltage in
percent of FS (120V or 400V)
Device Resolution Low Low The voltage, current and
High power resolution on the front
display (see Measurement Units
in Chapter 3) and in
communications (see
communication guides)
Volts Scale, V 60-828 V 144 V The maximum voltage scale
allowed, in secondary volts.
See Data Scales in Appendix F

82 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Parameter Options Default Description


Amps Scale, A 1.0-10.0 A 2  CT The maximum current scale
secondary allowed, in secondary amps.
See Data Scales in Appendix F

Figure ‎5-7: General Setup Dialog Box – Device Options Tab

Power Calculation Modes


The power calculation mode option allows you to change the method for
calculating reactive and apparent powers in presence of high harmonics.
The options work as follows:
 When the reactive power calculation mode is selected, active and
reactive powers are measured directly and apparent power is
calculated as:

S  P2  Q2

This mode is recommended for


electrical networks with low harmonic
distortion, commonly with THD < 5%
for volts, and THD < 10% for currents.
In networks with high harmonics, the
second method is preferable.
 When the non-active power calculation mode is selected, active
power is measured directly, apparent power is taken as product S
= V x I, where V and I are the RMS volts and amps, and reactive
power (called non-active power) is calculated as:

N  S2  P2

Configuring Digital Inputs


The PM135 can be provided with four to twelve (depend on DI/O module
type) digital inputs that can be linked to control setpoints to give an
indication on input status change (see Configuring Alarm/Control

PM135 Powermeter Series 83

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Setpoints), or can be linked to general pulse counters to count incoming


pulses (see Configuring Counters). They can also be linked to the
Billing/TOU registers to count pulses from external wattmeters or gas and
water meters.

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Digital Inputs tab.

Figure ‎5-8: General Setup Dialog Box – Digital Inputs Dialog Box

The available options are shown in Table 11.

Table 11: Digital Input Options


Parameter Options Default Description
Pulse Input PULSE MODE PULSE In pulse mode, either leading, or
Mode KYZ MODE MODE trailing edge of the input pulse is
recognized as an event. In KYZ
mode, both leading and trailing
edges of the input pulse are
recognized as separate events.
Pulse Polarity NORMAL (N.O.), NORMAL For the normal polarity, the open
INVERTING (N.C.) to closed transition is considered
a pulse. For the inverting polarity,
the closed to open transition is
considered a pulse.
It has no meaning in KYZ mode
where both transitions are used.
Debounce 1-100 ms 10 ms The amount of time while the
Time state of the digital input should
not change to be recognized as a
new state. Too low debounce time
could produce multiple events on
the input change.

The debounce time is applied the same for all digital inputs. If you change
the debounce time for a digital input, the same debounce time is
automatically assigned to the others.

84 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Configuring Relay Outputs


The PM135 can be provided with two to four (depend on DI/O module
type) optional relay outputs. Each relay can be operated either locally
from the alarm/control setpoints in response to an event or by a remote
command sent through communications. It can also be linked to an
internal pulse source to produce energy pulses.

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Relay Outputs tab.

Figure ‎5-9: General Setup Dialog Box – Relay Outputs Tab

The available relay outputs options are shown in Table 12.

Table 12: Relay Output Options


Parameter Options Default Description

PM135 Powermeter Series 85

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Parameter Options Default Description


Operation UNLATCHED UNLATCHED Unlatched mode: the relay goes into its
mode LATCHED active state when the control setpoint is in
PULSE active (operated) state, and returns into
KYZ pulse its non-active state when the setpoint is
released.
Latched mode: the relay goes into its
active state when the control setpoint goes
into active state and remains in the active
state until it is returned into its non-active
state by a remote command.
Pulse mode (normal pulse): the relay goes
into its active state for the specified time,
goes into non-active state for the specified
time and remains in the non-active state.
KYZ mode (transition pulse): the relay
generates transition pulses. The relay
changes its output state upon each
command and remains in this state until
the next command.
Polarity NORMAL (N.O.) NORMAL Normal polarity: the relay is normally de-
INVERTING (N.C.) energized in its non-active state and is
energized in its active (operated) state.
Inverting polarity: the relay is normally
energized in its non-active state and is de-
energized in its active (operated) state. It
is called failsafe relay operation.
Pulse width 20-1000 ms 100 ms The actual pulse width is a multiple of the
1-cycle time rounded to the nearest bigger
value.
The pause time between pulses is equal to
the pulse width.
Pulse source NONE NONE Links a pulse relay to the internal energy
kWh IMP pulse source. The relay must be set into
kWh EXP either pulse, or KYZ mode.
kvarh IMP
kvarh EXP
kvarh TOT
AkVAh TOT
Pulse rate, 0.1-1000.0 1.0 Defines the pulse weight in kWh units per
kWh/Pulse kWh/Pulse pulse

Generating Energy Pulses through Relay Outputs


To generate energy pulses through a relay output:
1. Set a relay to either pulse, or KYZ
mode, and then select a polarity
(active pulse edge) for energy
pulses and a pulse width.
2. Select a source energy accumulator
and the pulse rate for your output.
3. Send your new setup to the meter.

Configuring Analog Outputs


The meter can be ordered with two optional analog outputs with options
for 0-1mA, ±1mA, 0-20mA and 4-20mA current outputs.

86 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Analog Outputs tab.

Figure ‎5-10: General Setup Dialog Box – Analog Outputs Tab

The available analog output options are described in Table 13.


Table 13: Analog Output Options
Option Range Description
AO type 0-1mA The analog output type. When connected to the
±1mA meter, shows the actual AO type received from
0-20mA the device. When working off-line, select the
4-20mA analog output option corresponding to your
analog module.
Output See Appendix B Selects the measured parameter to be
parameter transmitted through the analog output channel.
Zero scale Defines the low engineering scale (in primary
units) for the analog output corresponding to a
lowest (zero) output current (0 or 4 mA)
Full scale Defines the high engineering scale (in primary
units) for the analog output corresponding to a
highest output current (1 or 20 mA)

When you select an output parameter for the analog output channel, the
default engineering scales are set automatically. They correspond to the
maximum available scales. If the parameter actually covers a lower range,
you can change the scales to provide a better resolution on an analog
output.
Scaling Non-Directional Analog Outputs
For non-directional analog outputs with a 0-1mA, 0-20mA or 4-20mA
current option, you can change both zero and full engineering scales for
any parameter. The engineering scale need not be symmetrical.

PM135 Powermeter Series 87

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Scaling Directional Power Factor


The engineering scale for the signed power factor emulates analog power
factor meters.
The power factor scale is -0 to +0 and is symmetrical with regard to
1.000 (-1.000  +1.000). The negative power factor is scaled as -1.000
minus the measured value, and non-negative power factor is scaled as
+1.000 minus the measured value. To define the entire power factor
range from -0 to +0, the default scales are specified as -0.000 to 0.000.
Scaling ±1mA Analog Outputs
Programming engineering scales for directional ±1mA analog outputs
depends on whether the output parameter represents unsigned (as volts
and amps) or signed (as powers and power factor) values.
For an unsigned output value, you can change both zero and full
engineering scales.
For a signed (directional) value, you should only provide the engineering
scale for the +1 mA output current.
The engineering scale for the 0 mA output current is always equal to zero
for all values except the signed power factor, for which it is set to 1.000
(see Scaling Directional Power Factor above).
The meter does not allow access to the low scale setting if the parameter
is directional. Whenever the sign of the output parameter is changed to
negative, the meter automatically uses the full engineering scale setting
for +1 mA with a negative sign.
Scaling Analog Outputs for 0-2 mA and ±2 mA
The 0-1mA and ±1mA current outputs provide a 100% overload, and
actually output currents up to 2 mA and ±2mA whenever the output value
exceeds the engineering scale you set for the 1 mA or ±1mA.
The output scales for 0-1 mA and ±1 mA analog outputs are programmed
for 0 mA and +1 mA regardless of the required output current range.
To use the entire output range of 2 mA or ±2 mA, set the analog output
scales as follows:
 0-2 mA: set the 1 mA scale to ½ of the required full scale output
for uni-directional parameters, and set the 0 mA scale to the
negative full scale and the 1 mA scale to zero for bi-directional
parameters.
 ±2 mA: set the 1 mA scale to ½ of the required full-scale output
for both uni-directional and bi-directional parameters.
For example, to provide the 0 to 2 mA output current range for Volts
measured by the meter in the range of 0 to 120V, set the 1 mA scale to
60V; then the 120V reading will be scaled to 2 mA.

Configuring Counters
The PM135 has four six-digit general counters that can count pulses
delivered through the device digital inputs with a programmable scale
factor. Each counter can also be incremented in response to any internal
or external event, checked and cleared through the Control Setpoints.

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Pulse/Event Counters tab.
Table 14 lists available counter options.

88 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Table 14: Counter Options


Option Range Default Description
Pulse Input None, None Links a digital input
DIGITAL INPUT #1 - #12 to the counter
Multiplier 1-9999 1 The value added to
the counter when a
pulse is detected on
the pulse source
input
Counter Value Displays the present
counter contents

Figure ‎5-11: General Setup Dialog Box – Pulse/Event Counters

You can preset a counter to a required value or clear it without affecting


the counter setup.
To preset or clear a counter:
1. Click the Online button on the PAS
toolbar before accessing the setup
dialog box.
2. Type in the required value into the Counter Value field.
3. Click Send

Configuring Alarm/Control Setpoints


The PM135 has an embedded logical controller that can perform different
actions in response to user-defined internal and external events. Unlike a
PLC, the meter uses a simplified programming technique based on
setpoints that allows the user to define a logical expression based on
measured analog and digital values that produce a required action.
The meter provides 16 control setpoints with programmable operate and
release delays. Each setpoint evaluates a logical expression with one
trigger argument. Whenever an expression is evaluated as “true”, the
setpoint performs a programmable action that can send a command to the
output relay, or increment a counter.
The logical controller provides very fast response to events. The scan time
for all setpoints is 1 cycle time (16.6 ms at 60Hz and 20 ms at 50/400
Hz).

PM135 Powermeter Series 89

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Using the Front Display


Select Setpoint Setup menu from the Device Setup menu to enter the
Device Setup
Real Time Clock

setup menu. See Viewing and Changing Setup Options in Chapter 3 for
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup
information on configuring parameters via the front display.
Setpoint Setup

Use the UP/DOWN buttons to scroll to the required setpoint.


To select a setpoint parameter:
1. Press SELECT/ENTER button to
select required setpoint.
2. Use the UP/DOWN buttons to scroll
to the required parameter.
To change the parameter value:
1. Press SELECT/ENTER button to
enter the parameter options menu.
2. Use the UP/DOWN buttons to
select the required value.
3. Press SELECT/ENTER button to
confirm the new setting
To store your new setpoint settings after you configured all
setpoint parameters:
1. SELECT Save Setup function from
Setpoint Setup menu and press
SELECT/ENTER button to enter
Save Setup menu
You are returned to the upper window
and can select another setpoint or exit
the menu.
2. SELECT Save and press
SELECT/ENTER button to Save
Setpoint parameters.

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Control/Alarm Setpoints tab.
The following table lists the available setpoint options.
Table 15: Setpoint Options
Option Range Description
Trigger parameter See Appendix C The trigger parameter that is used as an
argument in the logical expression
Operate limit The threshold (in primary units) at which the
conditional expression would be evaluated to
true. Not applicable for digital triggers.

90 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Option Range Description


Release limit The threshold (in primary units) at which the
conditional expression would be evaluated to
false. Defines the hysteresis for analog
triggers. Not applicable for digital triggers.
Operate delay 0.1-999.9 sec The time delay before operation when the
operate conditions are fulfilled
Release delay 0.1-999.9 sec The time delay before release when the
release conditions are fulfilled
Action See Appendix C The action performed when the setpoint
expression is evaluated to true (the setpoint
is in operated state)

Figure ‎5-12: General Setup Dialog Box – Control/Alarm Setpoints Tab

Using Numeric Triggers


For numeric (analog) triggers, you can specify two thresholds for each
trigger to provide hysteresis (dead band) for setpoint operations.
The Operate Limit defines the operating threshold, and the second
Release Limit defines the release threshold for the trigger. The trigger
thresholds are specified in primary units.
If you do not want to use hysteresis for the trigger, set the Release Limit
to the same value as the Operate Limit.
Using Binary Triggers
Binary (digital) triggers like digital inputs and relays are tested for
ON/CLOSED or OFF/OPEN status.
In the PM135, the binary events are level-sensitive events. An event is
asserted all the time while the corresponding condition exists.

Delaying Setpoint Operations


Two optional delays can be added to each setpoint to extend monitoring
triggers for a longer time before reaching a decision on whether the
expected event occurred or not. When a delay is specified, the logical
controller changes the setpoint status only if all conditions are asserted for
a period of at least as long as the delay time.

PM135 Powermeter Series 91

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Using Setpoint Events and Actions


When a setpoint status changes, i.e., a setpoint event is either asserted or
de-asserted, the following happens in your meter:
 The new setpoint status is logged to the setpoint status register
that can be monitored through communications from the SCADA
system or from a programmable controller in order to give an
indication on the expected event.
 The operated setpoint status is latched to the setpoint alarm latch
register that can be inspected through communications and via
the display (see Status Display in Chapter 3). The register holds
the last setpoint alarm status until it is explicitly cleared through
communications or via the display.
 A programmable action is performed on setpoint status transition
when a setpoint event is asserted.
Generally, setpoint actions are performed independently for each setpoint
and can be repeated a number of times for the same target. The
exceptions are relay operations that are shared for each target relay
between all setpoints using an OR scheme.
A relay output is operated when one of the setpoints linked to the relay is
activated and remains operated until all of these setpoints are released
(except for latched relays that require a separate release command to be
deactivated).
Using Time Counters
Any of the general counters can be used to count the setpoint operation
time. If you select the TIME CNT n action for a setpoint, the target
counter measures the time while the setpoint is in the operated state. The
counter resolution is 0.1 hour. See Status Display in Chapter 3 on how to
examine the counters via the front display.

92 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Configuring the Display


This setup allows configuring the meter display.

Using the Front Display


Select the Display Setup entry from the Device Setup menu. See
Viewing and Changing Setup Options in Chapter 3 for information on
configuring parameters via the front display.
Device Setup
Reset
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup

Table 16: Display Setup Options


Display Parameter Options Default Description
Auto Scroll Auto Scroll interval None, None Defines the scroll interval
2-15 sec for the main data display
or disables auto scroll
Auto Return Auto Return to Data diS = disabled, Enabled Enables automatic return
Display screen En = Enabled to the main display if no
buttons are pressed for 5
minutes
Backlight Display Backlight Continuous, 1- 1 Defines the backlight
10 min. duration to be "ON" in idle
operation
Diagnostics LED Diagnostics indication Disabled, Disabled Enables the diagnostic
LED Enabled LED
Phase Powers Phase Powers display Disabled, Disabled Disables or enables phase
mode Enabled powers in the main
display
Fundamental Fundamental component Disabled, Disabled Disables or enables
Powers display mode Enabled fundamental values in the
main display
Contrast Display Contrast 1-25 15 Sets LCD contrast
Load Bar Scale, A Reference load current 0-10,000A 0 0 Defines the nominal
for LED bar graph (0 = CT load
primary (100%) level for the bar
current) graph display

PM135 Powermeter Series 93

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Updating the meter Clock


This setup allows you to setup your local date and time

Using Front Display


Select the Real Time Clock entry from the Device Setup menu. See
Viewing and Changing Setup Options in Chapter 3 for information on
configuring parameters via the front display.
Device Setup
Reset
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
COM2 Setup

Using PAS
Ensure that the On-line button on the PAS toolbar is checked, and then
select RTC from the Monitor menu or click on the PAS toolbar Real-
Time Clock Button.
The RTC dialog box displays the current PC date and time and the time in
your meter.

Figure ‎5-13: Real Time Clock Dialog Box – Time and Date setup

To synchronize the meter clock with the PC clock, click Set.

Local Time Settings


This setup allows you to specify your time zone, daylight saving time, and
clock synchronization options.

Using the Front Display


Select the Local Settings entry from the Device Setup menu. See
Viewing and Changing Setup Options in Chapter 3 for information on
configuring parameters via the front display.

94 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 General Meter Setup

Device Setup
Reset
Real Time Clock
Display Setup
Basic Setup
Device Options
COM1 Setup
Local Settings

Local Settings
Date Order DD/MM/YY
Country Israel
Daylight Saving Time (DST) Enabled
DST Start Month March
DST Start Week of Month Last
DST Start Weekday Friday
DST Start Hour 2

Local Settings
DST Start Weekday Friday
DST Start Hour 2
DST End Month September
DST End Week of Month Last
DST End Weekday Sunday
DST End Hour 2
Time Synchronization Input None

Using PAS
Select General Setup from the Meter Setup menu, and then click on the
Local Settings tab.

Figure ‎5-14: General Setup Dialog Box – Local Settings Tab

The available options are described in Table 17.

PM135 Powermeter Series 95

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Meter Security

Table 17: Local Time Options


Display Label Parameter Options Default Description
Country Country Default, or country Default Defines calendar setting.
name The default setting stands
for the U.S.A.
Daylight Saving Time Daylight Saving Time Disabled, Enabled Disabled When DST is disabled, the
RTC operates in standard
time only. When enabled,
the meter automatically
updates the time at the pre-
defined DST switch dates.
DST Start Month DST Start Month Jan-Dec March The date when Daylight
DST Start Week of Month DST Start Week Last, 1-4 Last Saving Time begins.
DST Start Weekday DST Start Weekday Monday-Sunday Sunday
DST Start Hour DST Start Hour 1-6 2 The hour when Daylight
Saving Time begins.
DST End Month DST End Month Jan-Dec March The date when Daylight
DST End Week of Month DST End Week Last, 1-4 Last Saving Time Ends.
DST End Weekday DST End Weekday Monday-Sunday Sunday
DST End Hour DST End Hour 1-6 2 The hour when Daylight
Saving Time begins.
Time Synchronization Time Synchronization Input None None The external port receiving
Input DI1 PPM – DI12 PPM the time synchronization
pulses

Daylight Saving Time


When the daylight saving time is enabled, the meter automatically
advances the device clock by one hour when daylight saving time begins
and puts the clock back one hour when it ends. The default daylight
saving time change points are preset for the U.S.A.
The daylight saving time option is disabled in the PM135 by default. If the
daylight saving time option is disabled, you need to manually adjust the
device clock for daylight saving time.
Time Synchronization Pulses
External time synchronization pulses can be delivered through one of the
digital inputs.
If a digital input is selected as the time synchronization source, the edge
of an external pulse adjusts the device clock at the nearest whole minute.
The time accuracy could be affected by the debounce time of the digital
input, and by the operation delay of the external relay.

5.3 Configuring Meter Security


This setup allows changing the user password and enabling or disabling
password protection.
The password in your meter is preset to 0 at the factory, and password
protection is disabled.

Using the Front Display


Select the Password Setup entry from the Device Setup menu. See
Viewing and Changing Setup Options in Chapter 3 for information on
configuring parameters via the front display.

96 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Meter Security
Device Setup
Display Setup
Basic Setup
Device Options To change the password:
COM1 Setup
Local Settings 1. Select the Password Setup entry
Setpoint Setup from the Device Setup menu using
Password Setup
the UP/ DOWN buttons.
2. Press the ENTER button to enter
the "Password Setup" menu.
3. Select "Password Value" with the UP
/DOWN and ENTER buttons to
access the password value field.
4. select the desired digit using the
UP/DOWN button and move to next
digit by pressing the TAB button

5. Press ENTER button when desired


password value is achieved that
returns to "Password Value" field

The new password is effective for both the display and


communication ports.

To enable or disable password protection:


1. Select "Password Enabled" field
using the UP /DOWN buttons.
2. Press the ENTER button to enter
the "OFF/ON Password Enabled"
display.
3. Use the UP/DOWN buttons to
select the option:
ON enables password protection and
OFF disables password protection.
4. Press ENTER to confirm your new
setting.
5. Press ESC to exit the menu.

When password protection is enabled, you are not able


to change the device settings through the display or
communications unless you provide a correct password.
If you cannot provide a proper password, contact your
local distributor for the appropriate password to override
password protection.

PM135 Powermeter Series 97

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Meter Security

Using PAS
Ensure that the On-line button on the PAS toolbar is checked, select
Administration from the Monitor menu, and then select Change
Password -> Password 1.

Figure ‎5-15: Password Setup Dialog Box

To change the password:


1. Type in a new 4-digit password
2. Repeat the password in the
Confirm new password box
3. Check Enable password
protection to enable password
checking
4. Click Send.

98 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Billing/TOU

5.4 Configuring Billing/TOU


The TOU battery-backed clock unit is highly recommended in case
of using time-scheduled tariff rates, otherwise a long power
outage may cause the meter clock to lose time so your tariff
counters would not comply with the calendar schedule.

Billing Energy Registers


The PM135E/EH has 4 fully programmable billing energy registers that
can be linked to any internal energy source or to an external pulse source
that delivers pulses through the meter digital inputs.
Any energy register can provide either a single-tariff energy accumulation
or be individually linked to the TOU system providing both total and multi-
tariff energy billing.
Tariff Rates
The meter tariff structure supports 8 different tariff rates using an
arbitrary tariff schedule. A total of 4 types of days and 4 seasons are
supported with up to eight tariff changes per day.
Maximum Demand Registers
Any of billing energy registers can be individually linked to the maximum
demand register providing the same demand tariff structure as you
selected for energy registers.
Recording Billing Data and Load Profiling
The PM135E/EH can provide automatic recording of the daily energy and
maximum demand profile to a data log file. Maximum demand profiling
can be individually configured for every register.
See Factory Preset Data Log Files in Chapter 5 and Billing Profile Log File
in Appendix E for more information on the file layout and contents.
To configure the billing registers and the tariff system in your
meter:
1. Link the billing registers to the
respective energy sources.
2. Configure the options for the
registers to whether the only
totalization or both total and tariff
registers would be used, and
whether daily profiling should be
enabled for the energy usage and
maximum demand registers.
3. Configure the daily tariff schedule
using the TOU daily profiles for all
types of days and seasons.
4. Configure the season tariff schedule
using the TOU calendar.

Configuring Billing/Tariff Registers


To configure the billing/TOU registers in your meter:
1. Select Energy/TOU from the
Meter Setup menu.

PM135 Powermeter Series 99

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Billing/TOU

Figure ‎5-16: Energy/TOU Setup Dialog Box – Billing/TOU Registers tab

2. Configure the register options


according to the valid parameters
shown in Table 18.

Table 18: Billing/TOU Register Options


Parameter Options Default Description
Billing/TOU Registers
TOU Unchecked Unchecked Links tariff registers to the selected energy
Checked source
Use Profile Unchecked Checked Enables recording energy registers in a daily
Checked billing profile file (both total and tariff
registers if TOU is enabled).
Dmd Profile Unchecked Unchecked Enables recording maximum demand
Checked registers in a daily billing profile file (both
total and tariff registers if TOU is enabled)
Sum Profile Unchecked Checked Enables recording total (summary) registers
Checked in a daily billing profile file.
Units kWh, kvarh, kVAh, None The register measurement units. When a
m3, CF (cubic register is linked to an internal energy
foot), CCF source, it is set automatically. When an
(hundred cubic external pulse source is used, the user can
feet) select a measurement unit for the register.
Register Source List
Source Input None None Links an energy source to the register
kWh Import
kWh Export
kvarh Import
kvarh Export
kVAh,
DI1-DI12
Multiplier 0.001 to 100.000 1.000 The multiplication factor for the energy
source. Unchangeable for internal energy
sources.
Target Reg#1- Reg#4 None Defines the target billing register for the
energy source. It is set automatically.

100 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Billing/TOU

Configuring the Daily Tariff Schedule


To configure your daily tariff schedule, select Energy/TOU from the
Meter Setup menu, and then click on the TOU Daily Profiles tab.

Figure ‎5-17: Energy/TOU Setup Dialog Box – TOU Daily Profiles Tab

The daily profile setup allows you to specify the daily tariff change points
with a 15-minute resolution for 4 seasons using 4 different daily schedules
for each season.
To configure your daily profiles:
1. Select the desired season and day
type.
2. Select the start time for each tariff
change point and the corresponding
active tariff number.
3. Repeat the setup for all active
profiles.
The first tariff change point is fixed at 00:00 hours, and the last tariff
change you specified will be in use until 00:00 hours on the next day.
NOTE
The billing daily profile log file is automatically configured for the number
of active tariffs you defined in the meter TOU daily profiles.

Configuring the Season Tariff Schedule


To configure your season tariff schedule, select Energy/TOU from the
Meter Setup menu, and then click on the TOU Calendar tab.

PM135 Powermeter Series 101

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Billing/TOU

Figure ‎5-18: TOU Calendar Setup Dialog Box – TOU Calendar Tab

The meter TOU calendar allows you to configure any tariff schedule based
on any possible utility regulation. The calendar provides 32 entries that
allow you to specify profiles for working days and holidays through all
seasons in any order that is convenient for you, based on simple intuitive
rules.
To configure your season tariff schedule:
1. In the Season/Period box, select
the season, and in the Day Type
box, select a day type for this
calendar entry.
2. Define the time interval when this
daily tariff schedule is effective,
based on the start and end
weekdays and, for a multi-season
schedule, on the start and end
month for the selected season. It
does not matter which order of
weekdays or months you select: the
meter recognizes the correct order.
3. For exception days like weekends
and designated holidays, define a
specific day either by a month and a
month day, or by selecting a month,
a week and a weekday within the
month.

There are no limitations on how to build your schedule. A common


recommendation is to use minimum time constraints and only when it is
needed to avoid ambiguity. You need not to define month days if a daily
schedule is effective all days of the month, or to define the start and end
months if it is effective through all the year. If you wish to define a
specific period within a month using the start and end days, put this entry
before allocating the remaining days to another daily schedule without
specified month days, so it would be checked first for a match.

102 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Billing/TOU

The above picture shows a typical single-season tariff schedule with two
daily tariff profiles configured for working days, and weekends and the
designated U.S.A. holidays.

PM135 Powermeter Series 103

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Recorders

5.5 Configuring Recorders


The PM135E/EH has a 125-KByte onboard non-volatile memory for data
and event recording. The memory is fully configurable and can be freely
partitioned between log files.
The meter provides memory for a total of 3 log files:
Event log
Two data logs
The two data log files are pre-configured at the factory for recording a 15-
minute energy and demand profile and for the daily billing energy data
profile. If you wish to change the factory settings, follow the guidelines in
the next section.

Configuring Meter Memory


To view the present memory settings, select Memory/Log from the
Meter Setup menu, and then click on the Log Memory tab.

Figure ‎5-19: Log Setup Dialog Box – Log Memory Tab

The following table describes file options.

Option Range Description


Type Wrap-around Wrap-around: recording continues over the oldest records.
Non-wrap Non-wrap: recording is stopped until the file is cleared.
Size The size of memory allocated to the file.
Sections/Channels 0-8 The numbers of sections in a multi-section profile data log
file
Num. of Records 0-65535 Allocates the file memory for predefined number of records
Record size The size of the file record for a single channel or a single
section. It is set automatically depending on the file and on
the number of parameters in the data records
Parameters 0-9 The number of parameters in a single data log record

To change the file properties or to create a new file:


1. Double click on the file you want to
change.

104 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Recorders

2. Select desired parameters for your


log.
3. Click OK.
For your reference, the record size and the number of records available
for your file are reported in the dialog box.
To delete an existing file partition:
1. Click on Delete.
2. Click OK.

NOTES
1. Memory is allocated for a file
statically when you set up your files
and will not change unless you re-
organize files.
2. The meter automatically performs
de-fragmentation of the memory
each time you re-organize your files.
This prevents possible leakage of
memory caused by fragmentation. It
may take a couple of seconds.
For more information on configuring specific files, see the following
sections.
The following table can help you calculate an estimated file size when
planning your memory allocation.

File Record Size, Bytes File Size, Bytes


Event Log 16 Record size  Number of records
Conventional data Log 12 + 4  Number of parameters Record size  Number of records
Billing/TOU daily 12 + 4  (Number of season tariffs + 1 Record size  Number of billing registers
profile log for the TOU summary/total register) ( 2 for the maximum demand profile) x
Number of records

The factory pre-set file configuration is shown in the following table.

No. File Size, Channels Number of Number Factory-set Configuration


Bytes Records of Events
1 Event log 3200 200 200 200 last events
2 Data log #1 103680 4320 4320 15-min data profile for 15 days
17 Data log #16 8640 4 90 90 Daily billing/TOU profile for 90
days, 4 registers, totals + 3 tariffs

PM135 Powermeter Series 105

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Recorders

Configuring the Event Recorder


To configure the Event log file:
3. Double click on the Event Log file partition with the left mouse
button.

4. Select a desired file type for your file.


5. Select the maximum number of
records you want to be recorded in
the file.
6. Click OK, then send your new setup
to the meter or save to the device
database.
By default, the Event recorder stores all events related to configuration
changes, resets, and device diagnostics.

Configuring the Data Recorder


Conventional Data Log Files
The Data recorder is programmable for recording up to 9 data parameters
per record in each of the conventional data log files. The list of
parameters to be recorded to a data log is configurable individually for
each file.
To create a new data log file or re-configure an existing file:
1. Double click on the file partition with
the left mouse button.

2. Select a partition type for your file.


3. Select the number of parameters
you want to be recorded in the file
records.

106 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Recorders

4. Select the maximum number of


records you want to be recorded in
the file.
5. Click OK, and then send your new
setup to the meter, or save to the
device database.
To define the contents of the file:
1. Highlight the data log file row with
the left mouse button, and then
click on the Setup Recorder
button, or click on the Data
Recorder tab and select the
corresponding log number.

Figure ‎5-20: Log Setup Dialog Box – Data Recorder Tab

2. Configure the list of parameters to


be recorded in a log file. You are not
allowed to select more parameters
than you defined when configuring
your file. Refer to Appendix D for a
list of available parameters.
3. For your convenience, PAS follows
your selection and helps you
configure a series of the neighboring
parameters: when you open the
Group box for the next parameter,
PAS highlights the same group as in
your previous selection; if you select
this group again, PAS automatically
updates the Parameter box with
the following parameter in the
group.
4. Add the name for your data log file
in the Name box. It will appear in
the data log reports.
5. Save your new setup to the device
database, and send it to the meter.

PM135 Powermeter Series 107

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Recorders

Billing/TOU Daily Profile Log File


Data log #16 is configurable to store TOU daily profile log records on a
daily basis.
The file is organized as a multi-section file that has a separate section of
the same structure for each billing energy and maximum demand register.
The number of sections is taken automatically from the Billing/TOU Registers
setup (see Configuring Billing/Tariff Registers). If the maximum demand
profiling is used, then the number of sections in the file will be twice the
number of the allocated billing registers.
To configure a daily profile log file:
1. Configure your Billing/TOU registers and tariff schedule in
the meter (see Configuring Billing/Tariff Registers) first.
2. Double click on the Data Log#16 partition with the left
mouse button.

3. Select the TOU Daily Profile file type.


4. Select the number of season tariffs in your TOU schedule.
Add one additional parameter if you selected to record the
Summary (TOU total) registers as well.
5. Select the maximum number of records you want to be
recorded in the file assuming that a new record will be
added once a day.
6. Click OK and send your setup to the meter or save to the
meter database.

Factory Preset Data Log Files


Conventional Data Log #1
Data log #1 is factory preset for 15-min periodic recording of the standard
energy and demand quantities. You can freely change the list of recorded
parameters and the file update rate.
The default list of parameters is shown in the following table.

No. Parameter
1 kWh import
2 kWh export
3 kW import sliding demand
4 V1 demand
5 V2 demand
6 V3 demand

Periodic recording data is triggered by Setpoint #1 that is linked to the


meter clock. To change the periodic rate at which data is recorded,

108 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Recorders

change the time interval for the MINUTE INTERVAL trigger in Setpoint #1
(see Configuring Alarm/Control Setpoints).
Billing/TOU Profile Data Log #16
Data log #16 is pre-configured for daily billing energy and maximum
demand recording for the last 90 days. It is automatically updated once a
day.
See Billing Profile Log File in Appendix E for the file record structure.

PM135 Powermeter Series 109

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communication Protocols

5.6 Configuring Communication Protocols


This section describes how to customize protocol options for use with your
application software.

Configuring Modbus
Modbus Point Mapping
The PM135 provides 120 user assignable registers at addresses 0 to 119.
You can re-map any register available in the meter to any assignable
register so that registers found at different locations may be accessed
with a single request by re-mapping them to adjacent addresses.
Initially these registers are reserved and none of them points to an actual
data register. To build your Modbus register map:
1. Select Protocol Setup from the
Meter Setup menu, and click on
the Modbus Registers tab.
2. Click on the Default button to cause
the assignable registers to reference
the actual default meter register
6656 (0 through 119 are not
allowable register addresses for re-
mapping).

Figure ‎5-21: Protocol Setup Dialog Box – Modbus Registers Tab

3. Type in the actual addresses you


want to read from or write to via the
assignable registers. Refer to the
PM135 Modbus Reference Guide for
a list of the available registers. Note
that 32-bit Modbus registers should
always start at an even register
address.
4. Click Send to download your setup
to the meter.

110 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communication Protocols

Changing 32-bit Register Format


The PM135 allows you to read 32-bit Modbus analog registers, energy
counters and binary counters either in integer format, or in IEEE single
precision floating point format.
The 32-bit Modbus registers are factory-set to integer format. To change
the register format:
1. Select Protocol Setup from the
Meter Setup menu, and click on
the Modbus Registers tab.
2. Change the 32-bit register format in
the Modbus Options pane.
3. Click Send to download your setup
to the meter.

Configuring DNP3
Refer to the PM135 DNP3 Reference guide for information on the DNP3
protocol implementation and a list of the available data points.

DNP Options
Select Protocol Setup from the Meter Setup menu and click on the
DNP Options tab.

Figure ‎5-22: Protocol Setup Dialog Box – DNP Options Tab

The following table describes available options. Refer to the DNP3 Data
Object Library document available from the DNP User’s Group on the
DNP3 object types.

Table 19: DNP Options


Parameter Options Default Description
Binary Inputs (BI)

PM135 Powermeter Series 111

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communication Protocols

Parameter Options Default Description


Binary Input Object Single-bit Single-bit The default BI object variation for
With Status requests with qualifier code 06 when no
specific variation is requested
Analog Inputs (AI)
Analog Input Object 32-bit 16-bit-Flag The default AI object variation for
32-bit–Flag requests with qualifier code 06 when no
16-bit specific variation is requested
16-bit–Flag
Binary Counters (BC)
Binary Counter Object 32-bit+Flag 32-bit-Flag The default BC object variation for
32-bit–Flag requests with qualifier code 06 when no
16-bit+Flag specific variation is requested
16-bit–Flag
DNP General Options
16-bit AI Scaling Disabled Enabled Allows scaling 16-bit analog input
Enabled objects (see description below)
16-bit BC Scaling x1, x10, x1 Allows scaling 16-bit binary counter
x100, x1000 objects (see description below)
2
SBO Timeout 2-30 sec 10 Defines the Select Before Operate (SBO)
timeout when using the Control-Relay-
Output-Block object
3
Time Sync Period 0-86400 sec 86400 Defines the time interval between
periodic time synchronization requests
Multi Fragment 50-500 ms 50 Defines the time interval between
Interval fragments of the response message
when it is fragmented

Scaling 16-bit AI objects


Scaling 16-bit AI objects allows accommodating native 32-bit analog input
readings to 16-bit object format; otherwise it may cause an over-range
error if the full-range value exceeds a 16-bit point limit.
Scaling is enabled by default. It is not related to points that are read using
32-bit AI objects.
Refer to the PM135 DNP3 Reference Guide for information on the data
point scales and a reverse conversion that should be applied to the
received scaled values.
Scaling 16-bit Binary Counters
Scaling 16-bit Binary Counters allows changing a counter unit in powers of
10 to accommodate a 32-bit counter value to 16-bit BC object format.
If the scaling unit is greater than 1, the counter value is reported being
divided by the selected scaling unit from 10 to 1000. To get the actual
value, multiply the counter reading by the scaling unit.

Configuring DNP Class 0 Responses


The most common method of getting static object information from the
meter via DNP is to issue a read Class 0 request. The PM135 allows you to
configure the Class 0 response by assigning ranges of points to be polled
via Class 0 requests.

2
The Select Before Operate command causes the meter to start a timer. The
following Operate command must be sent before the specified timeout value
expires.
3
The meter requests time synchronization by bit 4 in the first octet of the
internal indication word being set to 1 when the time interval specified by the
Time Sync Period elapses. The master should synchronize the time in the meter
by sending the Time and Date object to clear this bit. The meter does not send
time synchronization requests if the Time Sync Period is set to 0.

112 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 5 Configuring the PM135 Configuring Communication Protocols

To view or build a DNP Class 0 response message:


1. Select Protocol Setup from the
Meter Setup menu and click on the
DNP Class 0 Points tab.
2. Select the object and variation type
for a point range.
3. Specify the start point index and the
number of points in the range. Refer
to the PM135 DNP3 Reference Guide
for available data points.
4. Repeat these steps for all point
ranges you want to be included into
the Class 0 response.
5. Click Send to download your setup
to the meter.
The factory-set Class 0 point ranges are shown in the picture below.

Figure ‎5-23: Protocol Setup Dialog Box – DNP Class 0 Points Tab

PM135 Powermeter Series 113

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 6 Device Control and Upgrading Resetting Accumulators, Maximum Values and Files

Chapter 6 Device Control and Upgrading


This section describes operations on the meter you can perform from the
front display or via PAS. To access device control options from PAS, you
should have your meter online.

6.1 Resetting Accumulators, Maximum Values and Files


Using the Front Display
Select the RESET entry from the Device Setup menu. See Viewing and
Changing Setup Options in Chapter 3 for information on configuring
parameters via the front display.
To reset the desired values:
1. Select the desired parameter to be
cleared using the UP/DOWN button
then press the ENTER button.
2. Press the ENTER button for
approximately 5 sec until the "Do"
notification is changed to "Done"
Table 20 shows the reset options available from the front display.

Table 20: Front Display Reset Options


Display Label Description
Volt/Ampere Demands Clears ampere and volt maximum demands
Power Demands Clears power maximum demands
All Demands Clears all maximum demands
Min/Max Clears Min/Max log
Energy Clears all total energies
Counters Clears all counters
Counter 1 Clears counter #1
Counter 2 Clears counter #2
Counter 3 Clears counter #3
Counter 4 Clears counter #4
Diagnostics Clears device diagnostics

Using PAS
Ensure that the On-line button on the PAS toolbar is checked, and then
select Reset from the Monitor menu.

114 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 6 Device Control and Upgrading Resetting Accumulators, Maximum Values and Files

Figure ‎6-1: Reset Dialog

To reset the desired values or files:


1. Click on the corresponding button,
and then confirm your command.
2. If an entry has more than one
target, you are allowed to select
targets to reset.
3. Check the corresponding boxes, and
then click OK.

Figure ‎6-2: Reset Maximum Demands Dialog Box

PM135 Powermeter Series 115

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 6 Device Control and Upgrading Viewing and Clearing Device Diagnostics

6.2 Viewing and Clearing Device Diagnostics


Using the Front Display
See Diagnostics Display in Chapter 3 on how to view and clear device
diagnostics from the front display.

Using PAS
Ensure that the On-line button on the PAS toolbar is checked, select
Device Control from the Monitor menu, and then click on the Device
Diagnostics tab.
See Device Diagnostic Codes in Appendix G for the list of diagnostic codes
and their meaning.

Figure ‎6-3: Device Control Dialog Box – Device Diagnostics Tab

To clear the device diagnostics events, click on Clear.

116 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 6 Device Control and Upgrading Viewing Communication Status and Statistics

6.3 Viewing Communication Status and Statistics


Ensure that the On-line button on the PAS toolbar is checked, select
Device Control from the Monitor menu, and then click on the
Communications tab.

Figure ‎6-4: Device Control Dialog Box – Communications Tab

This window indicates the present GPRS communication status (see


Setting Up GPRS Network in Chapter 5) and connection statistics of the
TCP clients (see Setting Up eXpertPower Client and Setting Up TCP
Notification Client in Chapter 5).
To clear the communication counters, click on Clear.
You can also clear the communications counters via the PAS Reset dialog
(see Resetting Accumulators, Maximum Values and Files).

6.4 Remote Relay Control


You can use PAS to send a remote command to your meter to operate any
relay output or release a latched relay, except of the relays linked to an
internal pulse source. These relays are blocked for operating from outside of
the meter.
To access the relay control dialog, ensure that the On-line button on the PAS
toolbar is checked, select Device Control from the Monitor menu, and then
click on the Remote Relay Control tab.

PM135 Powermeter Series 117

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 6 Device Control and Upgrading Upgrading Device Firmware

Figure ‎6-5: Device Control Dialog Box – Remote Relay Control Tab

To send a remote command to a relay:


1. Select a desired command in the
Relay Command box for a relay:
OPERATE – to operate a relay
RELEASE – to remove your remote
command, or to release a latched
relay
2. Click Send.

6.5 Upgrading Device Firmware


Your meter has upgradeable firmware. If you need to upgrade your
device, download a new firmware file to the meter through PAS.
Firmware can be downloaded via the Modbus RTU or Modbus/TCP protocol
through any communication port.
To download a new firmware file to your device:
1. Ensure that the communication port
you are connected through to the
meter operates in Modbus mode.
2. If you are connected to the meter
through a serial interface, it is
recommended to set the port baud
rate to 115,200 bps. See Setting Up
Serial Communication Ports on how
to remotely change the protocol and
baud rate in your meter.
3. Ensure that the On-line button on
the PAS toolbar is checked, and
then select Flash Downloader
from the Monitor menu and
confirm downloading.
4. Point to the firmware upgrade file
for your meter, click Open, and then
confirm upgrading the meter.

118 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 6 Device Control and Upgrading Upgrading Device Firmware

5. You are asked for the password


regardless of the password
protection setting in your meter.
Type the meter password, and click
OK. If you did not change the
password in the meter, enter the
default password 0.

6. Observe the PM135 Display screen


with "Flash Loader Running…"
notification as below

7. Wait until PAS completes upgrading


your device. It takes about 3-4
minutes at 115,200 bps to download
the file to the meter.

8. After upgrading firmware is


completed, the meter restarts, so
communications can be temporarily

PM135 Powermeter Series 119

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 6 Device Control and Upgrading Upgrading Device Firmware

lost. You may need to wait a short


duration until PAS restores a
connection with your device.

120 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Data

Chapter 7 Monitoring Meters


7.1 Viewing Real-time Data
Real-time data can be continuously retrieved from your devices and
updated on the screen at the rate you defined in the Instrument Setup.
To get real-time data from your meter:
9. Ensure that the On-line button on the PAS toolbar is checked.
10. Select the device site from the list
box on the PAS toolbar.
11. Point to RT Data Monitor on the
Monitor menu, and then select a
data set you want to view.

Polling Devices
Click on the “Poll” or “Continuous poll” button to poll the meter once
or continuously.

Click on the Stop button to stop continuous polling.

The following picture shows a typical data


monitor window.

Figure ‎7-1: RT Data Monitor Window

You can open as many monitor windows as you wish, either for different
sites, or for the same site using different data sets. An open data monitor
window is linked to the current site and does not change if you select
another site in the site list.
You can view acquired data in a tabular form or in a graphical form as a
data trend.

Organizing Data Sets


PAS supports 33 programmable data sets with up to 40 data parameters.
Set #0 is intended for simple meters, which have a limited number of
parameters, and is not recommended for the use with the PM135 . To re-
organize data sets, select RT Data Sets from the Monitor menu or click
on the button on the local toolbar.

PM135 Powermeter Series 121

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Data

Some data sets are preset for your convenience and others are empty.
You can freely modify data sets.
See Appendix D for a list of data available in your meter.

Polling Options
To change the polling options, click on the Data Monitor window with the
right mouse button and select Options.
If you check Do not stop on errors, polling is resumed automatically
when a communication error occurs, otherwise polling stops until you
restart it manually.

Viewing a Data Table


Changing the Data View
PAS displays data in either a single record or multi-record view. To change
the view, click on the Data Monitor window with the right mouse button
and select either Wrap to see a single record, or UnWrap to go to the
multi-record view.
Adjusting the Number of Rows in a Multi-Record View
Click the window with the right mouse button, select Options, adjust the
number of records you want to see in the window, and then click OK.
When the number of retrieved records exceeds the number of rows in the
window, the window scrolls up so that older records are erased.
See Working with Tables in Chapter 9 for more information on working
with tables.

Viewing Data Trend


To view a data trend, click on the button on the local toolbar.

To change the time range for your graph, click on the button on the
local toolbar, and then select the desired date and time range.
See Working with Graphic Windows in Chapter 9 for more information on
working with graphs.

Saving Data to a File


To save retrieved data to a file for later analysis, click on the Save button
, select an existing database or type the name for a new database,
and then click Save.

122 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Min/Max Log

To avoid confusion, do not store data files into the Sites directory where
site databases are located.

Printing Data
To check the report, as it will look when printed, select Print Preview
from the File menu.

To print retrieved data, click on the button on the PAS toolbar, select
a printer, and then click OK.

Real-time Data Logging


PAS allows you to store data records to a database automatically at the
time it updates data on the screen.
To setup the real-time logging options:
1. Open the Data Monitor window.
2. Click on the RT Logging On/Off
button on the local toolbar, or
select RT Logging Options from
the Monitor menu.
3. Select a database, or type the name
for a new database and select a
directory where you want to save it.
4. Select the number of tables, and the
number of records in each table you
want recorded.
5. Adjust the file update rate for
automatic recording. It must be a
multiple of the sampling rate that
you defined in the Instrument Setup
dialog.
6. Click Save.
When you run real-time data polling, PAS automatically saves retrieved
records to a database at the rate you specified.

The RT Logging On/Off button on the toolbar should be checked all


the time. You can suspend logging by un-checking the button, and then
resume logging by checking it again.

7.2 Viewing Real-time Min/Max Log


To retrieve the real-time Min/Max log data from your meter:
1. Select the device site from the list
box on the PAS toolbar.
2. Point to RT Min/Max Log on the
Monitor menu, and then select a
data set you want to view.

3. Ensure that the On-line button


on the PAS toolbar is checked.

4. Click on the Poll button .

PAS supports 9 programmable data sets that you can organize as you
wish. To build your data sets, select MinMax Data Sets from the
Monitor menu or click on the button on the local toolbar.

PM135 Powermeter Series 123

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Waveforms

See Working with Tables in Chapter 9 for more information on working


with tables.

7.3 Viewing Real-time Waveforms


To retrieve real-time waveforms from your meter:

1. Ensure that the On-line button


on the PAS toolbar is checked.
2. Select the device site from the list
box on the toolbar.
3. Select RT Waveform Monitor from
the Monitor menu or click on the
button on the PAS toolbar.

Use the Poll button for a single-step poll or the Continuous poll
button for continuous polling.

To stop continuous polling, click on the Stop button .

The meter provides simultaneous capture of six one-cycle voltage and


current AC waveforms at a rate of 64 samples per cycle. To give you a
more representative picture, PAS extends the waveforms across the
window up to eight cycles by repeating the captured waveforms.
To select the channels you want to view, click with the right mouse button
on the waveform window, select Channels..., check the channels for the
phase you want displayed, and then click OK.
See Working with Graphic Windows in Chapter 9 for more information on
working with waveforms.
Retrieved waveforms can be displayed in different views as overlapped or
non-overlapped waveforms, as RMS cycle-by-cycle plot, or as a harmonic
spectrum chart or table.

Viewing a Waveform Graph


When you open a new file, PAS shows you a waveform graph with non-
overlapped waveforms as shown in the picture above.

Click on the button on the local toolbar to see overlapped waveforms.

Click on the button for non-overlapped waveforms.

124 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Waveforms

Figure ‎7-2: RT Waveform Monitor Window

Viewing a Frequency Plot


Click on the button to view a cycle-by-cycle frequency plot for the
sampled voltage waveforms.

Viewing a Harmonic Spectrum


Click on the button to view a spectrum chart for the selected
waveform channel. PAS provides voltage, current, active power and
reactive power spectrum charts. See Viewing Real-time Harmonic
Spectrum for more information on viewing options.

Viewing Phasor Diagrams


The phasor diagrams show you relative magnitudes and angles of the
three-phase voltage and current fundamental component. All angles are
shown relative to the reference voltage channel.
To change the reference channel, click on the waveform window with the
right mouse button, select Options..., click on the Phasor tab, check the
channel you want to make a reference channel, and then click OK.

Viewing Symmetrical Components


Waveform views have an additional pane at the right where PAS displays
the symmetrical components for voltages and currents, calculated for the
point indicated by the left marker line.
To enable or disable the symmetrical components, click on the waveform
window with the right mouse button, select Options..., check or uncheck
the Symmetrical components box on the Channels tab, and then click
OK.

Viewing Phase-to-phase Voltages


PAS can transform phase-to-neutral voltage waveforms in configurations
with a neutral into phase-to-phase waveforms allowing you to view the
waveshape, angle relationships and harmonics of the phase-to-phase
voltages.
Click on the button on the waveform window toolbar. Click the button
once again to return to phase-to-neutral waveforms.

PM135 Powermeter Series 125

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Harmonic Spectrum

7.4 Viewing Real-time Harmonic Spectrum


To retrieve real-time harmonic spectrum from your meter:

4. Ensure that the On-line button on the PAS toolbar is checked.

5. Select the device site from the list


box on the toolbar.
6. Select RT Harmonic Monitor from
the Monitor menu or click on the
button on the PAS toolbar.

Click on the “Poll” or “Continuous poll” button to poll the meter once
or continuously. Click on the Stop button to stop continuous polling.

Figure ‎7-3: RT Harmonic Monitor – Spectrum Chart

PAS retrieves harmonic spectrum for V1-V3 and I1-I3 channels.


Harmonics can be displayed as a spectrum chart for a selected channel or
in a table. PAS can also synthesize waveforms based on the harmonic
spectrum to let you view a shape of the voltage and current waveforms in
your network.

Viewing a Spectrum Chart


Click on the button to view a spectrum chart for the selected channel.
To change a channel, click on the window with the right mouse button,
select Channels..., check the channel you want displayed, and then click
OK. PAS provides voltage, current, active power and reactive power
spectrum charts.
PAS can give you indication on whether harmonic levels in the sampled
waveforms exceed compliance limits defined by the power quality
standards or local regulations.

To review or change harmonic limits:

126 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Harmonic Spectrum

1. Click on the spectrum window with


the right mouse button and select
Limits….
2. Select a known harmonics standard,
or select Custom and specify your
own harmonic limits.
3. Check the Enabled box to visualize
harmonic faults on the spectrum
graph and in harmonic tables.
Harmonics that exceed selected compliance levels are colored in red on
the graph and in the tables.

Figure ‎7-4: Harmonic Limits

Viewing a Spectrum Table


Click on the button on the local toolbar to display the harmonics
spectrum in a tabular view for a selected phase or for all phases together.
The spectrum table shows voltage, current, active power and reactive
power harmonic components both in percent of the fundamental and in
natural units, and harmonic phase angles.
To change a phase, click on the window with the right mouse button,
select Options..., check the phase you want displayed, and then click OK.

PM135 Powermeter Series 127

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 7 Monitoring Meters Viewing Real-time Harmonic Spectrum

Figure ‎7-5: RT Harmonic Monitor – Spectrum Table

Viewing Synthesized Waveforms


To view the synthesize waveforms based on the sampled harmonic spectrum,
click on the button on the local toolbar to view non-overlapped voltage
and current waveforms, or click on the button to view them overlapped.

PAS shows a pair of 4-cycle voltage and current synthesized AC waveforms


for a single phase.
To select the channels you want to view, click with the right mouse button on
the waveform window, select “Channels...”, check the channels for the phase
you want displayed, and then click OK.

Figure ‎7-6: RT Harmonic Monitor – Synthesized Waveforms

128 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 8 Retrieving and Storing Files Uploading Files on Demand

Chapter 8 Retrieving and Storing Files


PAS allows you to retrieve recorded events and data from your meters
and to save them to files on your PC in the Microsoft Access database
format.
Historical data can be uploaded on demand any time you need it, or
periodically through the Upload Scheduler that retrieves data
automatically on a predefined schedule, for example, daily, weekly or
monthly.
If you do not change the destination database location, new data is added
to the same database so you can store long-term data profiles in one
database regardless of the upload schedule you selected.

8.1 Uploading Files on Demand


To retrieve the log files from your meter:

1. Ensure that the On-line button


on the PAS toolbar is checked.
2. Select a device site from the list box on the PAS toolbar.
3. Select Upload Logs from the Logs
menu.

4. Select a database, or type the name


for a new database, and select a
directory where you want to save it.
5. Click on the Select Logs button and
check boxes for logs you want to be
retrieved from the meter.
6. If you wish to retrieve data starting
with a known date, check the
“From” box and select the start
date for retrieving data.
7. If you wish to retrieve data recorded
before a known date, check the “To”
box and select the last date for
retrieving data.
8. Click OK.

PM135 Powermeter Series 129

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 8 Retrieving and Storing Files Using the Upload Scheduler

8.2 Using the Upload Scheduler


To setup the Upload Scheduler:
1. Select Upload Scheduler from the
Logs menu.

Figure ‎8-1: Upload Scheduler Setup Dialog Box

2. Click Add Site, point to the site


database for which you want to
organize the schedule, and then
click OK.
3. Click Browse and select a database
for storing retrieved data, or type
the name for a new database, select
a directory where you want to save
it, and then click OK.
4. Click Configure or double click on
the site row.

130 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 8 Retrieving and Storing Files Using the Upload Scheduler

5. Select a daily, weekly or monthly


schedule, and adjust the start time.
If you wish to upload data
periodically in predefined intervals,
click on Periodic and define the
time period in hours and minutes.
6. Select the number of attempts to
upload data in the event of
temporary communication problems
or unavailability of your device, and
the delay between attempts in
minutes and seconds.
7. If you wish to use the schedule to
synchronize the device clock with
your PC, check the RTC
Synchronization Enable box. If
your device is password protected
by a communications password,
type in the password you set in the
device to allow PAS to update the
clock.
8. Click on the Select Logs button,
check the boxes for logs you want to
upload on a schedule, and then click
OK.
9. Check the Enabled box at left to
activate a schedule for the device.
10. Click Close to store your schedule.

To keep the Upload Scheduler running, the On-line button on the PAS
toolbar must be checked all the time. If you uncheck it, the scheduler
stops operations. This does not cause loss of data, since the scheduler will
resume operations when you check this button again.

Suspending the Scheduler


To suspend the Upload Scheduler, check the Suspend Scheduler box at
right. To activate the Upload Scheduler, leave this box unchecked.

PM135 Powermeter Series 131

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 8 Retrieving and Storing Files Viewing Files On-line

Running the Scheduler on Demand


You can run the scheduler at any time outside the schedule by checking
the Start Now box at right. This is a one-time action. After uploading is
completed, the Upload Scheduler un-checks this box automatically.

Reviewing Upload Problems


When the Upload Scheduler fails to retrieve data from the device, or some
data is missing, or another problem occurs, it puts an error message to
the log file. To review this file, select System Log from the View menu.

8.3 Viewing Files On-line


Sometimes, it is useful to review a particular piece of historical data on-
line at the time you expect new events to appear in the log. PAS allows
you to retrieve historical data from a particular log without storing it to a
file. The data appears only in the window on your screen. You can save it
manually to the database.

To view the log data on-line, check the On-line button on the PAS
toolbar, select the log you want to retrieve in the Logs menu, and then
click on the Poll button . Only new log records are retrieved from the
device. If you want to review the entire log from the beginning, click on
the Restore log button , and then click on the Poll button .

NOTE
When reading multi-section profile data, only the first section is available
for reading online.
See Chapter 9 Viewing Files for information on using different log views.

8.4 Exporting Files

Exporting Files in COMTRADE and PQDIF Formats


The COMTRADE and PQDIF file converters allow you to convert saved real-
time waveforms into COMTRADE or PQDIF file format, and data log tables
– into PQDIF format.

Manual Converting
To manually convert your waveforms or a data log into COMTRADE or
PQDIF format:

1. Click on the Export button on the


PAS toolbar.

132 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 8 Retrieving and Storing Files Exporting Files

2. Select the database and a data log


table you want to export, and then
click Open.

3. Select a folder where you want to


store your exported files, type a file
name that identifies your files,
select a file output format, and then
click on the Save button.
4. The PQDIF files are commonly
recorded in compressed format. If
you do not want your files to be
compressed, uncheck the
Compress box before saving the
file.
In COMTRADE format, each waveform event is recorded into a separate
file.
PQDIF file names are followed by a timestamp of the first event recorded
to the file, and may look like follows:
12KVSUB_20040928T133038.pqd.

Automatic Converting
PAS allows you to automatically convert data logs into PQDIF format at
the time you upload data from your devices via the Upload Scheduler.
To automatically convert your data log tables into PQDIF format:
1. Open the Upload Scheduler.
2. Highlight a desired device site with
the left mouse button, and then
click on the Export button.
3. Check the Enabled box for a data
log or a waveform log table you
want to automatically convert at the
upload time.
4. Highlight the Record to… row for
the selected table and click on the
Browse button.
5. Select a folder where you want to
store converted files, type in the
converted file’s name, select a

PM135 Powermeter Series 133

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 8 Retrieving and Storing Files Archiving Files

desired output file format, and then


click on Save.
6. Repeat the same for all tables you
wish to be converted.
7. Click OK.

Exporting Files in Excel Format


PAS allows you to convert data tables into the Microsoft Excel workbook
format, either manually, or automatically while retrieving data from your
meters via the Upload Scheduler.
To store files in Excel format, follow instructions in the previous section
and select Excel Workbook as the output file format.
The first row of the Excel table lists data names (see Appendix D) and the
second row provides data codes, which identify recorded data points (see
Modbus communications guide for data codes) that may be useful for
automated table processing.
Each table row is provided with the device identifier that you can define in
the meter database (see Creating a New Site for your Meter).

8.5 Archiving Files


Microsoft Access databases tend to grow fast. Databases above 0.5
Gigabytes can drastically slow down file operations.
To avoid enormous growing files, you can either periodically change the
target database, or use the Upload Scheduler’s file archiver to
automatically move older data to archives.
The Upload Scheduler archives files upon a weekly, monthly or yearly
schedule. When archiving data, a new database is created to where older
data from your present database with the expired archiving date is
moved.
An archive file keeps the original database name to which the date of the
oldest database record is added, so you can easily identify your archives
and work with them as you work with a regular database.
To provide a schedule for archiving files:
1. When defining a schedule for
uploading files from your meter,
click on Configure or double click
on the site row.
2. Click Auto Archive.

3. Check the Enable box and select a


periodic schedule for archiving your
files for this site.
4. Click OK.

134 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 8 Retrieving and Storing Files Archiving Files

To avoid archiving partially updated data, archiving is performed in a day


after expiring a scheduled period and not before 2 hours a.m.

PM135 Powermeter Series 135

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 9 Viewing Files Operations with Files

Chapter 9 Viewing Files


9.1 Operations with Files
Files you read from the meters are stored in one or in a number of tables
in the meter database. Sections of multi-section files like energy load
profiles are stored in multiple tables – each file section in a separate
database table.

Opening a Database Table


To open a database table:

1. Click on the Open button on


the PAS toolbar, or select Open...
from the File menu.
2. Select Access Database (*.mdb)
in the Files of type box; select a
directory where your files are
located, and point to the file you
wish to open.
3. Select a desired table on the right
pane, and then click Open, or
double click on the table name.
Names of the last 16 files you opened are stored in the File menu, so you
can select them directly from the menu.

Saving Data to a File


To save data from the open database table to a file:

1. Click on the Save button , and


select a directory where you want
your file to be stored.
2. Select a database or type the name
for a new database.
3. Click Save.
To avoid confusion, do not store data files into the Sites directory where
site databases are located.

9.2 Viewing Options

Customizing Views
Changing Date Order
To change the way PAS displays the date:
1. Select Options from the Tools
menu and click on the Preferences
tab.
2. Select the preferred date order.
3. Click OK.

136 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 9 Viewing Files Viewing Options

Selecting Timestamp Format


The timestamp is normally recorded and displayed on the screen at a 1-
ms resolution. If you have an application that does not support this
format, you may instruct PAS to drop milliseconds.
To change the way PAS records and displays the timestamp:
1. Select Options from the Tools
menu and click on the Preferences
tab.
2. Select the preferred timestamp
format.
3. Click OK.

Working with Tables


Selecting Font and Grid
To change the table font or a type of the grid lines:
1. Click with right mouse button on the
table, select Options and click on
the Table tab.
2. Select the font type and size and
how you wish the table grid to be
shown.
3. Click OK.

Selecting Primary and Secondary Units


Voltages and currents can be displayed in primary or secondary units.
To change units, click on the table with the right mouse button, select
Options, select the desired units for voltages and currents, and then click
OK.

Copying a Table
To copy the entire table, or its part, into the Clipboard or into another
application such as Microsoft Excel or Word:
1. Click on the data window with the
right mouse button and choose
Select All, or click on the upper-left
corner of the table (where the “No.”
label is commonly displayed).
2. Click with the right mouse button on
the window again and choose Copy,
or click on the Copy button on
the PAS toolbar.
3. Run the application to which you
want to copy data, position the
cursor at the correct place.

4. Click the Paste button on the


application's toolbar or select Paste
from the Edit menu.
When copying, table columns are separated by a tab character.

PM135 Powermeter Series 137

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 9 Viewing Files Viewing Options

Printing a Table
To check how your document appears on a printed page, select Print
Preview from the File menu.

To print a table to a printer, click on the print button on the toolbar,


select a printer and click OK.

Working with Graphic Windows


Selecting Channels
To select the channels you want to view on the screen, click on the graph
window with the right mouse button, select Channels..., check the
channels you want displayed, and then click OK.
Checkboxes for channels that are not available in the present view are
dimmed.

Selecting Primary and Secondary Units


Voltages and currents can be displayed in primary or secondary units.
To change units, click on the table with the right mouse button, select
Options, select the desired units for voltages and currents, and then click
OK.

Selecting the Time Axis


In waveform views, the horizontal axis can be displayed either in absolute
time with date and time stamps, or in milliseconds relatively to the
beginning of a waveform.
To change the time units, click on the waveform window with the right
mouse button, select Options..., click on the Axes tab, select the desired
units, and then click OK.

Selecting Line Styles and Colors


Channel waveforms can be displayed using different colors and line styles.
To change the colors or line styles, click on the graph window with the
right mouse button, select Options..., click on the Display tab, adjust
colors and styles, and then click OK.

Selecting Grid and Frame Colors


Click on the graph window with the right mouse button, select Options...,
and click on the Display tab
To change the color or style of the grid lines, click on the Grid line on the
left pane, and then select the color and style for the grid. To disable the
grid, uncheck the Grid Visible box.
To change the window frame color to white, check the White Frame box
at right.

Using Marker Lines


The waveform and trend windows have two blue dashed marker lines. The
left marker indicates the starting position and the right marker indicates
the end position for calculating the average and peak values.
The minimum distance between the two markers is exactly one cycle.

To change the marker position, click on the button, or click on the


window with the right mouse button and select Set Marker, and then
click on the point where you want to put the marker.

138 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 9 Viewing Files Viewing the Event Log

You can also drag both markers with the mouse, or use the right and left
arrow keys on your keyboard to change the marker position. Click on the
graph pane to allow the keyboard to get your input before using the
keyboard.

Delta Measurements
To measure the distance between two waveform or trend points, click on
the Delta button , then click on the first point, and then click on the
second point.
The first reference point is still frozen until you uncheck and check the
Delta button again, while the second point can be placed anywhere within
the graph line by clicking on the graph to the left or right from the
reference point.
To disable delta measurements, click on the Delta button once again.

Using a Zoom
You can use a horizontal and, for waveforms, also a vertical, zoom to
change size of your graph.
Use the buttons on your local toolbar to zoom in and zoom
out. One click gives you a 100-percent horizontal or 50-percent vertical
zoom. Two buttons representing magnifying glasses give you a
proportional zoom in both directions.

Copying a Graph
To copy a graph, or its part, into the Clipboard or into another application
such as Microsoft Excel or Word:
1. Click on the graph window with the
right mouse button and choose
Copy All, or Copy Waveform.
Some windows may have additional
options.
2. Position the cursor at the place
where you whish to copy the graph.

3. Click the Paste button on the


application's toolbar or select Paste
from the Edit menu.

Printing a Graph
To check how the graph appears on a printed page, select Print Preview
from the File menu.

To print a graph to a printer, click on the Print button on the PAS


toolbar, select a printer and click OK.

9.3 Viewing the Event Log


The Event log contains time-tagged events related to configuration
changes, resets and device diagnostics.
The Event log is displayed in a tabular view, one event per row. Use the
scroll bar to view the entire log contents.

PM135 Powermeter Series 139

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 9 Viewing Files Viewing the Event Log

Figure ‎9-1: Event Log Window

See Working with Tables for more information on viewing options.

Filtering and Sorting Events


You can use filtering to find and work with a subset of events that meet
the criteria you specify.

Click on the Filter button , or click on the report window with the right
mouse button and select Filter.... Check the causes of events you want to
display, and then click OK. PAS temporary hides rows you do not want
displayed.
To change the default sorting order based on the date and time, click on
the Sort button , or click on the report window with the right mouse
button and select Sort..., check the desired sort order, and then click
OK.

140 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Chapter 9 Viewing Files Viewing the Data Log

9.4 Viewing the Data Log


Data log files can be displayed in a tabular view, one data record per row,
or in a graphical view as a data trend graph.

Figure ‎9-2: Data Log Window

Viewing Data Trend


To view data in a graphical form, click on the Data Trend button on
the local toolbar.
To change the time range for your graph, click on the Time Range button
on the local toolbar, and then select the desired date and time range.

PM135 Powermeter Series 141

Downloaded from www.Manualslib.com manuals search engine


‎Appendix A Technical Specifications

Appendix A Technical Specifications


A.1 Environmental Conditions
Operating temperature: -30°C to 60°C (-22°F to 140°F)
Storage temperature: -30°C to 85°C (-40°F to 185°F)
Humidity: 0 to 95% non-condensing

A.2 Construction
Dimensions see Figure 2-1
Weight: 0.70 kg (1.54 lb.)

Materials
Case enclosure: plastic PC/ABS blend
Front panel: plastic PC
PCB: FR4 (UL94-V0)
Terminals: PBT (UL94-V0)
Connectors-Plug-in type: Polyamide PA6.6 (UL94-V0)
Packaging case: Carton and Stratocell® (Polyethylene Foam) brackets
Labels: Polyester film (UL94-V0)

A.3 Power Supply


120/230 VAC-DC Option:
Rated input: 85-265VAC 50/60/400 Hz, 88-290VDC, Burden 9VA
Isolation:
Input to ground: 2500 VAC

12 VDC Option:

Rated input: 9.5-18 VDC, Burden 4VA


Isolation: 1500VDC

24/48 VDC Option:

Rated input: 18.5-58 VDC, Burden 4VA


Isolation: 1500VDC
Wire size: up to 12 AWG (up to 3.5 mm2)

142 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix A Technical Specifications

A.4 Input Ratings


Voltage Inputs – Installation category III
Operating range: 690VAC line-to-line, 400VAC line-to-neutral
Direct input and input via PT (up to 828VAC line-to-line, up to 480VAC line-
to-neutral)
Input impedance: 1000 k
Burden for 400V: < 0.4 VA
Burden for 120V: < 0.04 VA
Over-voltage withstands: 1000 VAC continuous, 2000 VAC for 1 second
Wire size: up to 12 AWG (up to 3.5mm2)
Current Inputs (via CT) – Installation category III
Wire size: 12 AWG (up to 3.5 mm2)
Galvanic isolation: 3500 VAC
5A secondary (standard)
Operating range: continuous 10A RMS
Burden: < 0.2 VA @ In=5A (with 12AWG wire and 1 m long)
Overload withstand:
15A RMS continuous, 300A RMS for 1 second (with 12AWG section wire)
1A secondary (option)
Operating range: continuous 2A RMS
Burden: < 0.02 VA @ In=1A (with 12AWG wire and 1 m long)
Overload withstand:
3A RMS continuous, 80A RMS for 1 second (with 12AWG section wire)
40mA secondary – External Solid or Split core CT (option)
External CT Operating range: continuous 10, 100-1200A RMS
Burden: < 0.02 VA @ nominal current
Sampling Rate measurement
128 samples/cycle

A.5 Optional Relay Outputs


Electromechanical relay - DRY contact, option (4DI/2DO or 12DI/4DO
Optional module)
2 or 4 relays rated at 5A/250 VAC; 5A/30 VDC, 1 contact (SPST Form A)
Galvanic isolation:
Between contacts and coil: 3000 VAC 1 min
Between open contacts: 750 VAC
Operate time: 10 ms max.
Release time: 5 ms max.
Update time: 1 cycle
Wire size: 14 AWG (up to 1.5 mm2)

PM135 Powermeter Series 143

Downloaded from www.Manualslib.com manuals search engine


‎Appendix A Technical Specifications

Solid State relay option (4DI/2DO only)


2 relays rated at 0.15A/250 V AC/DC, 1 contact (SPST Form A)
Galvanic isolation: 3750 VAC 1 min
Operate time: 1 ms max.
Release time: 0.25 ms max.
Update time: 1 cycle
Connector type: removable, 4 pins.
Wire size: 14 AWG (up to 1.5 mm2)

A.6 Optional Digital Inputs


4, 8 or 12 Digital Inputs (4DI/2DO, 8DI or 12DI/4DO Optional module)
Dry Contacts, internally wetted @ 24VDC
Wet contact @ 250VDC (12DI/4DO only)
Sensitivity: Open @ input resistance >100 k , Closed @ Input
resistance < 100 
Galvanic isolation: 3750 VAC 1 min
Internal power supply: 24VDC
Scan time: 1 ms
Connector type: removable, 5 pins.
Wire size: 14 AWG (up to 1.5 mm2)

A.7 Optional Analog Outputs


4 Analog Outputs optically isolated (AO Optional module)
Ranges (upon order):
1 mA, maximum load 5 k (100% overload)

0-20 mA, maximum load 510 

4-20 mA, maximum load 510 

0-1 mA, maximum load 5 k  (100% overload)


Isolation: 2500 VAC 1 min
Power supply: internal
Accuracy: 0.5% FS
Update time: 1 cycle
Connector type: removable, 5 pins.
Wire size: 14 AWG (up to 1.5 mm2)

A.8 Communication Ports


COM1
RS-485 optically isolated port
Isolation: 3000 VAC 1 min
Baud rate: up to 115.2 kbps.
Supported protocols: MODBUS RTU, DNP3, SATEC ASCII, IEC 60870-5-101

144 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix A Technical Specifications

Connector type: removable, 3 pins.


Wire size: up to 14 AWG (up to 1.5 mm2).
COM2 (Optional module)

Ethernet Port
Transformer-isolated 10/100BaseT Ethernet port
Supported protocols: MODBUS/TCP (Port 502), DNP3/TCP (Port 20000), IEC
60870-5-104
Number of simultaneous connections: 4 (2 MODBUS/TCP + 2 DNP3/TCP).
Connector type: RJ45 modular.
PROFIBUS DP (IEC 61158)
RS-485 optically isolated PROFIBUS interface.
Connector type: removable, 5 pins.
Baud rate: 9600 bit/s – 12 Mbit/s (auto detection).
32 bytes input, 32 bytes output.
Supported protocols: PROFIBUS
RS-232/422-485 Port
RS-232 or RS-422/485 optically isolated port
Isolation: 3000 VAC 1 min
Baud rate: up to 115.2 kbps.
Supported protocols: MODBUS RTU, DNP3, and SATEC ASCII, IEC 60870-5-
101
Connector type: removable, 5 pins for RS-422/485 and DB9 for RS-232.
Wire size: up to 14 AWG (up to 1.5 mm2).

A.9 Real-time Clock


Standard Meter Clock
Non-backed clock
Accuracy: typical error 1 minute per month @ 25C
Typical clock retention time: 30 seconds
TOU Module Meter Clock
Battery-backed clock
Accuracy: typical error 5 seconds per month @ 25C (±2.5ppm)
Typical clock retention time: 36 months

A.10 Display
3.5" LCD Monochrome Display, 240 x 128 dots resolution
Tri-color LED load bar graph (40-110%)
COM1 RX/TX activity LED
Diagnostics indication LED
kWh/kvarh Pulse LED (for E and EH models)
Keypad: 5 push buttons

PM135 Powermeter Series 145

Downloaded from www.Manualslib.com manuals search engine


‎Appendix A Technical Specifications

A.11 Standards Compliance


Accuracy:
Complies IEC62053-22, class 0.5S
Meets ANSI C12.20 –1998, class 10 0.5%
Electromagnetic Immunity:
Comply with IEC 61000-6-2:
IEC 61000-4-2 level 3: Electrostatic Discharge
IEC 61000-4-3 level 3: Radiated Electromagnetic RF Fields
IEC 61000-4-4 level 3: Electric Fast Transient
IEC 61000-4-5 level 3: Surge
IEC 61000-4-6 level 3: Conducted Radio Frequency
IEC 61000-4-8: Power Frequency Magnetic Field
Meets ANSI/IEEE C37.90.1: Fast Transient SWC
Electromagnetic Emission:
Comply with IEC 61000-6-4: Radiated/Conducted class A
Comply with IEC CISPR 22: Radiated/Conducted class A
Safety/Construction:
Meets IEC 61010-1: 2006
AC and Impulse Insulation:
Comply with IEC 62052-11:
2500 VAC during 1 minute
6KV/500Ω @ 1.2/50 μs impulse

146 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix A Technical Specifications

A.12 Measurement Specifications


Table 21: Measurement Specifications Parameters

Parameter Full Scale @ Input Accuracy Range


Range % % FS Conditions
Reading
Voltage 120VxPT @ 120V 0.2 0.02 10% to 120% FS 0 to 1,150,000 V
400VxPT @ 690V Starting voltage
1.5-5.0% FS (selectable)
Line current CT 0.2 0.02 For In = 5A 0 to 50,000 A
1% to 200% FS Starting current 0.1% FS
For In = 1A
5% to 200% FS
Active power 0.36PTCT @ 120V 0.3 0.02 |PF|  0.51 -10,000,000 kW to
1.2PTCT @ 690V +10,000,000 kW
Reactive power 0.36PTCT@ 120V 0.3 0.04 |PF|  0.91 -10,000,000 kvar to
1.2PTCT @ 690V +10,000,000 kvar
Apparent power 0.36PTCT @ 120V 0.3 0.02 |PF|  0.51 0 to 10,000,000 kVA
1.2PTCT @ 690V
Power factor 1.000 0.2 |PF|  0.5, -0.999 to +1.000
I  2% FSI
Frequency 50 Hz 0.02 VL-N > 25V 15 Hz to 70 Hz
60 Hz 0.04 15 Hz to 70 Hz
25 Hz 15 Hz to 70 Hz
400 Hz 320 Hz to 480 Hz
Total Harmonic 999.9 1.5 0.2 THD  1%, 0 to 999.9
Distortion, THD V  10% FSV and
V (I), %Vf (%If) VL-N > 25V,
I  10% FSI
Total Demand 100 1.5 TDD  1%, 0 to 100
Distortion, TDD, I  10% FSI,
% VL-N > 25V
Active energy Class 0.5S under conditions as per 0 to 999,999,999 kWh
Import & Export IEC 62053-22:2003
Reactive energy Class 0.5S under conditions as per 0 to 999,999,999 kvarh
Import & Export IEC 62053-22:2003, |PF|  0.9
Apparent energy Class 0.5S under conditions as per 0 to 999,999,999 kVAh
IEC 62053-22:2003

PT - external potential transformer ratio


CT - primary current rating of the external current transformer
FSV - voltage full scale
FSI - current full scale

Vf - fundamental voltage

If - fundamental current

NOTES
1. Accuracy is expressed as  (percentage of reading +
percentage of full scale)  1 digit. This does not
include inaccuracies introduced by the user's
potential and current transformers. Accuracy
calculated at 1second average.

1
@ 80% to 120% of voltage FS, 1% to 200% of current FS and frequency 50/60 Hz

PM135 Powermeter Series 147

Downloaded from www.Manualslib.com manuals search engine


‎Appendix A Technical Specifications

2. Specifications assume: voltage and current


waveforms with THD  5% for kvar, kVA and PF, and
reference operating temperature 20C - 26C.
3. Measurement error is typically less than the
maximum error indicated.

148 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix B Analog Output Parameters

Appendix B Analog Output Parameters


The following table lists parameters that can be provided on the meter’s
analog outputs.

Table 22: Analog Output Parameters


Display Code Designation Description
none NONE None (output disabled)
1-Cycle Phase Values
rt.U1 V1/12 RT 1 V1/V12 Voltage
rt.U2 V2/23 RT 1 V2/V23 Voltage
rt.U3 V3/31 RT 1 V3/V31 Voltage
rt.U12 V12 RT V12 Voltage
rt.U23 V23 RT V23 Voltage
rt.U31 V31 RT V31 Voltage
rt.C1 I1 RT I1 Current
rt.C2 I2 RT I2 Current
rt.C3 I3 RT I3 Current
1-Cycle Total Values
rt. P kW RT Total kW
rt. q kvar RT Total kvar
rt. S kVA RT Total kVA
rt. PF PF RT Total PF
r.PF.LG PF LAG RT Total PF Lag
r.PF.Ld PF LEAD RT Total PF Lead
1-Cycle Auxiliary Values
r.nEU.C In RT In Current
rt. Fr FREQ RT Frequency
1-Sec Phase Values
Ar.U1 V1/12 AVR 1 V1/V12 Voltage
Ar.U2 V2/23 AVR 1 V2/V23 Voltage
Ar.U3 V3/31 AVR 1 V3/V31 Voltage
Ar.U12 V12 AVR V12 Voltage
Ar.U23 V23 AVR V23 Voltage
Ar.U31 V31 AVR V31 Voltage
Ar.C1 I1 AVR I1 Current
Ar.C2 I2 AVR I2 Current
Ar.C3 I3 AVR I3 Current
1-Sec Total Values
Ar. P kW AVR Total kW
Ar. q kvar AVR Total kvar
Ar. S kVA AVR Total kVA
Ar. PF PF AVR Total PF
A.PF.LG PF LAG AVR Total PF Lag
A.PF.Ld PF LEAD AVR Total PF Lead
1-Sec Auxiliary Values
A.nEU.C In AVR In Current
Ar. Fr FREQ AVR Frequency
Demands E, EH
Acd.P.i kW IMP ACD Accumulated kW import demand
Acd.P.E kW EXP ACD Accumulated kW export demand

PM135 Powermeter Series 149

Downloaded from www.Manualslib.com manuals search engine


‎Appendix B Analog Output Parameters

Display Code Designation Description


Acd.q.i kvar IMP ACD Accumulated kvar import demand
Acd.q.E kvar EXP ACD Accumulated kvar export demand
Acd.S kVA ACD Accumulated kVA demand

1 In 4LN3, 3LN3 and 3BLN3 wiring modes, the voltages will be line-to-neutral; for
any other wiring mode, they will be line-to-line voltages.

150 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix C Setpoint Triggers and Actions

Appendix C Setpoint Triggers and


Actions
Table 23: Setpoint Triggers
Display Code Designation Description
None NONE None (condition is not active)
Status Inputs
DI1 On STAT INP #1 ON Status input #1 ON
DI2 On STAT INP #2 ON Status input #2 ON
DI3 On STAT INP #3 ON Status input #3 ON
DI4 On STAT INP #4 ON Status input #4 ON
DI5 On STAT INP #5 ON Status input #5 ON
DI6 On STAT INP #6 ON Status input #6 ON
DI7 On STAT INP #7 ON Status input #7 ON
DI8 On STAT INP #8 ON Status input #8 ON
DI9 On STAT INP #9 ON Status input #9 ON
DI10 On STAT INP #10 ON Status input #10 ON
DI11 On STAT INP #11 ON Status input #11 ON
DI12 On STAT INP #12 ON Status input #12 ON
DI1 OFF STAT INP #1 OFF Status input #1 OFF
DI2 OFF STAT INP #2 OFF Status input #2 OFF
DI3 OFF STAT INP #3 OFF Status input #3 OFF
DI4 OFF STAT INP #4 OFF Status input #4 OFF
DI5 OFF STAT INP #5 OFF Status input #5 OFF
DI6 OFF STAT INP #6 OFF Status input #6 OFF
DI7 OFF STAT INP #7 OFF Status input #7 OFF
DI8 OFF STAT INP #8 OFF Status input #8 OFF
DI9 OFF STAT INP #9 OFF Status input #9 OFF
DI10 OFF STAT INP #10 OFF Status input #10 OFF
DI11 OFF STAT INP #11 OFF Status input #11 OFF
DI12 OFF STAT INP #12 OFF Status input #12 OFF
Relays
RO1 ON RELAY #1 ON Relay #1 ON
RO2 ON RELAY #2 ON Relay #2 ON
RO3 ON RELAY #3 ON Relay #3 ON
RO4 ON RELAY #4 ON Relay #4 ON
RO1 OFF RELAY #1 OFF Relay #1 OFF
RO2 OFF RELAY #2 OFF Relay #2 OFF
RO3 OFF RELAY #3 OFF Relay #3 OFF
RO4 OFF RELAY #4 OFF Relay #4 OFF
Phase Reversal
Pos Phase Reversal POS PHASE Positive phase rotation reversal
REVERSAL
Neg Phase Reversal NEG PHASE Negative phase rotation reversal
REVERSAL
Low/High 1-Cycle Values on any Phase
High Volt RT HI VOLT RT 1 High voltage
Low Volt RT LO VOLT RT 1 Low voltage
High Amps RT HI AMPS RT High current
Low Amps RT LO AMPS RT Low current

PM135 Powermeter Series 151

Downloaded from www.Manualslib.com manuals search engine


‎Appendix C Setpoint Triggers and Actions

Display Code Designation Description


High Volt THD RT HI V THD 2 High voltage THD
High Current THD RT HI I THD 2 High current THD
High KF RT HI KF RT High K-Factor
High Current TDD RT HI I TDD High current TDD
1-Cycle Auxiliary Values
High Freq RT HI FREQ RT High frequency
Low Freq RT LO FREQ RT Low frequency
High Volt Unb% RT HI V UNB% RT 1 High voltage unbalance
High Curr Unb% RT HI I UNB% RT High current unbalance
1-Sec Phase Values
High I1 Avr HI I1 AVR High I1 current
High I2 Avr HI I2 AVR High I2 current
High I3 Avr HI I3 AVR High I3 current
Low I1 Avr LO I1 AVR Low I1 current
Low I2 Avr LO I2 AVR Low I2 current
Low I3 Avr LO I3 AVR Low I3 current
1-Sec Values on any Phase
High Volt Avr HI VOLT AVR 1 High voltage
Low Volt Avr LO VOLT AVR 1 Low voltage
High Amps Avr HI AMPS AVR High current
Low Amps Avr LO AMPS AVR Low current
1-Sec Total Values
High kW Imp Avr HI kW IMP AVR High total kW import
High kW Exp Avr HI kW EXP AVR High total kW export
High kvar Imp Avr HI kvar IMP AVR High total kvar import
High kvar Exp Avr HI kvar EXP AVR High total kvar export
High kVA Avr HI kVA AVR High total kVA
Low PF Lag Avr HI PF LAG AVR Low total PF Lag
Low PF Lead Avr HI PF LEAD AVR Low total PF Lead
1-Sec Auxiliary Values
High In Avr HI In AVR High neutral current
High Freq Avr HI FREQ RT High frequency
Low Freq Avr LO FREQ RT Low frequency
Demands
High V1/12 Dmd HI V1/12 DMD 1 High V1/V12 Volt demand
High V2/23 Dmd HI V2/23 DMD 1 High V2/V23 Volt demand
High V3/31 Dmd HI V3/31 DMD 1 High V3/V31 Volt demand
High I1 Dmd HI I1 DMD High I1 Ampere demand
High I2 Dmd HI I2 DMD High I2 Ampere demand
High I3 Dmd HI I3 DMD High I3 Ampere demand
High kW Imp BD HI kW IMP BD High block kW import demand
High kVA BD HI kVA BD High block kVA demand
High kW Imp SD HI kW IMP SD High sliding window kW import demand
High kVA SD HI kVA SD High sliding window kVA demand
High kW Imp Acc Dmd HI kW IMP ACD High accumulated kW import demand
High kVA Imp Acc Dmd HI kVA ACD High accumulated kVA demand
High kW Imp Prd Dmd HI kW IMP PRD High predicted kW import demand
High kVA Imp Prd Dmd HI kVA PRD High predicted kVA demand

152 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix C Setpoint Triggers and Actions

Display Code Designation Description


Time and Date Parameters
Day of Week DAY OF WEEK Day of week
Year YEAR Year
Monh MONTH Month
Day of Month DAY OF MONTH Day of month
Hours HOURS Hours
Minutes MINUTES Minutes
Seconds SECONDS Seconds
Minute Interval MINUTE INTERVAL Minute interval: 1-5, 10, 15, 20, 30, 60 min
1 In 4LN3, 3LN3 and 3BLN3 wiring modes, the voltages will be line-to-neutral; for
any other wiring mode, they will be line-to-line voltages.

Table 24: Setpoint Actions


Display Code Designation Description
None NONE None (no action)
Relay 1 ON OPERATE RELAY #1 Operate relay RO1
Relay 2 ON OPERATE RELAY #2 Operate relay RO2
Relay 3 ON OPERATE RELAY #3 Operate relay RO3
Relay 4 ON OPERATE RELAY #4 Operate relay RO4
Relay 1 OFF RELEASE RELAY #1 Release latched relay RO1
Relay 2 OFF RELEASE RELAY #2 Release latched relay RO2
Relay 3 OFF RELEASE RELAY #3 Release latched relay RO3
Relay 4 OFF RELEASE RELAY #4 Release latched relay RO4
Increment counter 1 INC CNT #1 Increment counter #1
Increment counter 2 INC CNT #2 Increment counter #2
Increment counter 3 INC CNT #3 Increment counter #3
Increment counter 4 INC CNT #4 Increment counter #4
Time counter 1 TIME CNT #1 Count operation time using counter #1
Time counter 2 TIME CNT #2 Count operation time using counter #2
Time counter 3 TIME CNT #3 Count operation time using counter #3
Time counter 4 TIME CNT #4 Count operation time using counter #4
Notification NOTIFICATION Send a notification message
Data Log 1 DATA LOG #1 Record data to Data Log #1

PM135 Powermeter Series 153

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Appendix D Parameters for Data


Monitoring and Logging
The following table lists parameters measured by the meter that are
available for monitoring through communications and for recording to a
data log file. The left column shows data abbreviations used in PAS.
Parameter groups are highlighted in bold.

Table 25: Data Monitoring and Logging Parameters


Designation Description
NONE None (stub, read as zero)
DIGITAL INPUTS Digital Inputs
DI1:16 Digital Inputs Status DI1:DI12
RELAYS Relays
RO1:16 Relay Status RO1:RO4
COUNTERS Pulse Counters
COUNTER 1 Counter #1
COUNTER 2 Counter #2
COUNTER 3 Counter #3
COUNTER 4 Counter #4
RT PHASE 1-Cycle Phase Values
V1 V1/V12 Voltage 1
V2 V2/V23 Voltage 1
V3 V3/V31 Voltage 1
I1 I1 Current
I2 I2 Current
I3 I3 Current
kW L1 kW L1
kW L2 kW L2
kW L3 kW L3
kvar L1 kvar L1
kvar L2 kvar L2
kvar L3 kvar L3
kVA L1 kVA L1
kVA L2 kVA L2
kVA L3 kVA L3
PF L1 Power factor L1
PF L2 Power factor L2
PF L3 Power factor L3
V1 THD V1/V12 Voltage THD 1
V2 THD V2/V23 Voltage THD 1
V3 THD V3/V31 Voltage THD 1
I1 THD I1 Current THD
I2 THD I2 Current THD
I3 THD I3 Current THD
I1 KF I1 K-Factor
I2 KF I2 K-Factor
I3 KF I3 K-Factor
I1 TDD I1 Current TDD
I2 TDD I2 Current TDD

154 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Designation Description
I3 TDD I3 Current TDD
V12 V12 Voltage
V23 V23 Voltage
V31 V31 Voltage
RT TOTAL 1-Cycle Total Values
kW Total kW
kvar Total kvar
kVA Total kVA
PF Total PF
PF LAG Total PF lag
PF LEAD Total PF lead
kW IMP Total kW import
kW EXP Total kW export
kvar IMP Total kvar import
kvar EXP Total kvar export
V AVG 3-phase average L-N/L-L voltage
V LL AVG 3-phase average L-L voltage
I AVG 3-phase average current
RT AUX 1-Cycle Auxiliary Values
In In (neutral) Current
FREQ Frequency
V UNB% Voltage unbalance 2
I UNB% Current unbalance 2
AVR PHASE 1-Second Phase Values
V1 V1/V12 Voltage
V2 V2/V23 Voltage
V3 V3/V31 Voltage
I1 I1 Current
I2 I2 Current
I3 I3 Current
kW L1 kW L1
kW L2 kW L2
kW L3 kW L3
kvar L1 kvar L1
kvar L2 kvar L2
kvar L3 kvar L3
kVA L1 kVA L1
kVA L2 kVA L2
kVA L3 kVA L3
PF L1 Power factor L1
PF L2 Power factor L2
PF L3 Power factor L3
V1 THD V1/V12 Voltage THD 1
V2 THD V2/V23 Voltage THD 1
V3 THD V3/V31 Voltage THD 1
I1 THD I1 Current THD
I2 THD I2 Current THD
I3 THD I3 Current THD
I1 KF I1 K-Factor
I2 KF I2 K-Factor

PM135 Powermeter Series 155

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Designation Description
I3 KF I3 K-Factor
I1 TDD I1 Current TDD
I2 TDD I2 Current TDD
I3 TDD I3 Current TDD
V12 V12 Voltage
V23 V23 Voltage
V31 V31 Voltage
AVR TOTAL 1-Second Total Values
kW Total kW
kvar Total kvar
kVA Total kVA
PF Total PF
PF LAG Total PF lag
PF LEAD Total PF lead
kW IMP Total kW import
kW EXP Total kW export
kvar IMP Total kvar import
kvar EXP Total kvar export
V AVG 3-phase average L-N/L-L voltage 1
V LL AVG 3-phase average L-L voltage
I AVG 3-phase average current
AVR AUX 1-Second Auxiliary Values
In In (neutral) Current
FREQ Frequency
V UNB% Voltage unbalance 2
I UNB% Current unbalance 2
PHASORS Phasors
V1 Mag V1/V12 Voltage magnitude 1
V2 Mag V2/V23 Voltage magnitude 1
V3 Mag V3/V31 Voltage magnitude 1
I1 Mag I1 Current magnitude
I2 Mag I2 Current magnitude
I3 Mag I3 Current magnitude
V1 Ang V1/V12 Voltage angle 1
V2 Ang V2/V23 Voltage angle 1
V3 Ang V3/V31 Voltage angle 1
I1 Ang I1 Current angle
I2 Ang I2 Current angle
I3 Ang I3 Current angle
DEMANDS Present Demands (Power Demands E, EH)
V1 DMD V1/V12 Volt demand 1
V2 DMD V2/V23 Volt demand 1
V3 DMD V3/V31 Volt demand 1
I1 DMD I1 Ampere demand
I2 DMD I2 Ampere demand
I3 DMD I3 Ampere demand
kW IMP BD kW import block demand
kvar IMP BD kvar import block demand
kVA BD kVA block demand
kW IMP SD kW import sliding window demand

156 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Designation Description
kvar IMP SD kvar import sliding window demand
kVA SD kVA sliding window demand
kW IMP ACD kW import accumulated demand
kvar IMP ACD kvar import accumulated demand
kVA ACD kVA accumulated demand
kW IMP PRD kW import predicted sliding window demand
kvar IMP PRD kvar import predicted sliding window demand
kVA PRD kVA predicted sliding window demand
PF IMP@kVA DMD PF (import) at Maximum kVA sliding window demand
kW EXP BD kW export block demand
kvar EXP BD kvar export block demand
kW EXP SD kW export sliding window demand
kvar EXP SD kvar export sliding window demand
kW EXP ACD kW export accumulated demand
kvar EXP ACD kvar export accumulated demand
kW EXP PRD kW export predicted sliding window demand
kvar EXP PRD kvar export predicted sliding window demand
In DMD In (neutral) current demand
SUMM ACC DMD Billing Summary (Total) Accumulated Demands E, EH
REG1 ACD Register #1 accumulated demand
REG2 ACD Register #2 accumulated demand
REG3 ACD Register #3 accumulated demand
REG4 ACD Register #4 accumulated demand
SUMM BLK DMD Billing Summary (Total) Block Demands E, EH
REG1 BD Register #1 block demand
REG2 BD Register #2 block demand
REG3 BD Register #3 block demand
REG4 BD Register #4 block demand
SUMM SW DMD Billing Summary (Total) Sliding Demands E, EH
REG1 SD Register #1 sliding demand
REG2 SD Register #2 sliding demand
REG3 SD Register #3 sliding demand
REG4 SD Register #4 sliding demand
ENERGY Total Energy E, EH
kWh IMPORT kWh import
kWh EXPORT kWh export
kvarh IMPORT kvarh import
kvarh EXPORT kvarh export
kVAh TOTAL kVAh total
SUMMARY REGS Billing Summary (Total) Energy Registers E, EH
SUM REG1 Summary energy register #1
SUM REG2 Summary energy register #2
SUM REG3 Summary energy register #3
SUM REG4 Summary energy register #4
PHASE ENERGY Phase Energy E, EH
kWh IMP L1 kWh import L1
kWh IMP L2 kWh import L2
kWh IMP L3 kWh import L3
kvarh IMP L1 kvarh import L1
kvarh IMP L2 kvarh import L2
kvarh IMP L3 kvarh import L3

PM135 Powermeter Series 157

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Designation Description
kVAh L1 kVAh total L1
kVAh L2 kVAh total L2
kVAh L3 kVAh total L3
%HD V1 V1/V12 Harmonic Distortions EH 1
V1 %HD01 H01 Harmonic distortion
V1 %HD02 H02 Harmonic distortion
… ...
V1 %HD40 H40 Harmonic distortion
%HD V2 V2/V23 Harmonic Distortions EH 1
V2 %HD01 H01 Harmonic distortion
V2 %HD02 H02 Harmonic distortion
… ...
V2 %HD40 H40 Harmonic distortion
%HD V3 V3/V31 Harmonic Distortions EH 1
V3 %HD01 H01 Harmonic distortion
V3 %HD02 H02 Harmonic distortion
… ...
V3 %HD40 H40 Harmonic distortion
%HD I1 I1 Harmonic Distortions EH
I1 %HD01 H01 Harmonic distortion
I1 %HD02 H02 Harmonic distortion
… ...
I1 %HD40 H40 Harmonic distortion
%HD I2 I2 Harmonic Distortions EH
I2 %HD01 H01 Harmonic distortion
I2 %HD02 H02 Harmonic distortion
… ...
I2 %HD40 H40 Harmonic distortion
%HD I3 I3 Harmonic Distortions EH
I3 %HD01 H01 Harmonic distortion
I3 %HD02 H02 Harmonic distortion
… ...
I3 %HD40 H40 Harmonic distortion
ANG V1 V1/V12 Harmonic Angles EH 1
V1 H01 ANG H01 Harmonic angle
V1 H02 ANG H02 Harmonic angle
… ...
V1 H40 ANG H40 Harmonic angle
ANG V2 V2/V23 Harmonic Angles EH 1
V2 H01 ANG H01 Harmonic angle
V2 H02 ANG H02 Harmonic angle
… ...
V2 H40 ANG H40 Harmonic angle
ANG V3 V3/V31 Harmonic Angles EH 1
V3 H01 ANG H01 Harmonic angle
V3 H02 ANG H02 Harmonic angle
… ...
V3 H40 ANG H40 Harmonic angle
ANG I1 I1 Harmonic Angles EH
I1 H01 ANG H01 Harmonic angle

158 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Designation Description
I1 H02 ANG H02 Harmonic angle
… ...
I1 H40 ANG H40 Harmonic angle
ANG I2 I2 Harmonic Angles EH
I2 H01 ANG H01 Harmonic angle
I2 H02 ANG H02 Harmonic angle
… ...
I2 H40 ANG H40 Harmonic angle
ANG I3 I3 Harmonic Angles EH
I3 H01 ANG H01 Harmonic angle
I3 H02 ANG H02 Harmonic angle
… ...
I3 H40 ANG H40 Harmonic angle
H1 PHASE Fundamental (H01) Phase Values
V1 H01 V1/V12 Voltage 1
V2 H01 V2/V23 Voltage 1
V3 H01 V3/V31 Voltage 1
I1 H01 I1 Current
I2 H01 I2 Current
I3 H01 I3 Current
kW L1 H01 kW L1
kW L2 H01 kW L2
kW L3 H01 kW L3
kvar L1 H01 kvar L1
kvar L2 H01 kvar L2
kvar L3 H01 kvar L3
kVA L1 H01 kVA L1
kVA L2 H01 kVA L2
kVA L3 H01 kVA L3
PF L1 H01 Power factor L1
PF L2 H01 Power factor L2
PF L3 H01 Power factor L3
HRM TOT POW Fundamental and Harmonic Total Power Values
kW H01 Total fundamental kW
kvar H01 Total fundamental kvar
kVA H01 Total fundamental kVA
PF H01 Total fundamental PF
MIN PHASE Minimum 1-Cycle Phase Values
V1 MIN V1/V12 Voltage 1
V2 MIN V2/V23 Voltage 1
V3 MIN V3/V31 Voltage 1
I1 MIN I1 Current
I2 MIN I2 Current
I3 MIN I3 Current
MIN TOTAL Minimum 1-Cycle Total Values
kW MIN Total kW
kvar MIN Total kvar
kVA MIN Total kVA
PF MIN Total PF

PM135 Powermeter Series 159

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Designation Description
MIN AUX Minimum 1-Cycle Auxiliary Values
In MIN In Current
FREQ MIN Frequency
MAX PHASE Maximum 1-Cycle Phase Values
V1 MAX V1/V12 Voltage 1
V2 MAX V2/V23 Voltage 1
V3 MAX V3/V31 Voltage 1
I1 MAX I1 Current
I2 MAX I2 Current
I3 MAX I3 Current
MAX TOTAL Maximum 1-Cycle Total Values
kW MAX Total kW
kvar MAX Total kvar
kVA MAX Total kVA
PF MAX Total PF
MAX AUX Maximum 1-Cycle Auxiliary Values
In MAX In Current
FREQ MAX Frequency
MAX DMD Maximum Demands (Power Demands E, EH)
V1 DMD MAX V1/V12 Maximum volt demand 1
V2 DMD MAX V2/V23 Maximum volt demand 1
V3 DMD MAX V3/V31 Maximum volt demand 1
I1 DMD MAX I1 Maximum ampere demand
I2 DMD MAX I2 Maximum ampere demand
I3 DMD MAX I3 Maximum ampere demand
kW IMP SD MAX Maximum kW import sliding window demand
kW EXP SD MAX Maximum kvar import sliding window demand
kvar IMP SD MAX Maximum kW export sliding window demand
kvar EXP SD MAX Maximum kvar export sliding window demand
kVA SD MAX Maximum kVA sliding window demand
In DMD MAX In (neutral) current maximum demand
MAX SUMMARY DMD Billing Summary (Total) Maximum Demands E, EH
REG1 MD Summary register #1 maximum demand
REG2 MD Summary register #2 maximum demand
REG3 MD Summary register #3 maximum demand
REG4 MD Summary register #4 maximum demand
AO RAW Raw Analog Outputs (A/D Units)
AO1 Analog output AO1
AO2 Analog output AO2
AO3 Analog output AO3
AO4 Analog output AO4
TOU PRMS TOU Parameters E, EH
ACTIVE TARIFF Active TOU tariff
ACTIVE PROFILE Active TOU profile
TOU REG1 Billing TOU Energy Register #1 E, EH
REG1 TRF1 Tariff #1 register
REG1 TRF2 Tariff #2 register
… …
REG1 TRF8 Tariff #8 register

160 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Designation Description
TOU REG2 Billing TOU Energy Register #2 E, EH
REG2 TRF1 Tariff #1 register
REG2 TRF2 Tariff #2 register
… …
REG2 TRF8 Tariff #8 register
TOU REG3 Billing TOU Energy Register #3 E, EH
REG3 TRF1 Tariff #1 register
REG3 TRF2 Tariff #2 register
… …
REG3 TRF8 Tariff #8 register
TOU REG4 Billing TOU Energy Register #4 E, EH
REG4 TRF1 Tariff #1 register
REG4 TRF2 Tariff #2 register
… …
REG4 TRF8 Tariff #8 register
TOU MAX DMD REG1 Billing TOU Maximum Demand Register #1 E, EH
REG1 TRF1 MD Tariff #1 maximum demand
REG1 TRF2 MD Tariff #2 maximum demand
… …
REG1 TRF8 MD Tariff #8 maximum demand
TOU MAX DMD REG2 Billing TOU Maximum Demand Register #2 E, EH
REG2 TRF1 MD Tariff #1 maximum demand
REG2 TRF2 MD Tariff #2 maximum demand
… …
REG2 TRF8 MD Tariff #8 maximum demand
TOU MAX DMD REG3 Billing TOU Maximum Demand Register #3 E, EH
REG3 TRF1 MD Tariff #1 maximum demand
REG3 TRF2 MD Tariff #2 maximum demand
… …
REG3 TRF8 MD Tariff #8 maximum demand
TOU MAX DMD REG4 Billing TOU Maximum Demand Register #4 E, EH
REG4 TRF1 MD Tariff #1 maximum demand
REG4 TRF2 MD Tariff #2 maximum demand
… …
REG4 TRF8 MD Tariff #8 maximum demand

1 In 4LN3, 3LN3 and 3BLN3 wiring modes, the voltages will be line-to
neutral; for any other wiring mode, they will be line-to-line voltages.
2 The value is calculated as a relation of the maximum deviation of
phase values from a 3-phase average value to a 3-phase average.
NOTE
Designations of some enginering demands and billing energy and demand
registers are shown using a short name notation available in PAS V1.4. By
default, PAS uses long names compatible with older versions of PAS. You
can select a desired notation from the Tools/Options/Preferences tab.
PAS does not allow to store data in files using different data names. If you
have a file uploaded with a previous version of PAS using long data
names, either continue using long data names, or store data in a new file.
See table below for a list of parameters with short and long names.

PM135 Powermeter Series 161

Downloaded from www.Manualslib.com manuals search engine


‎Appendix D Parameters for Data Monitoring and Logging

Short Data Long Data Name Description


Name
kW IMP ACD kW IMP ACC DMD Accumulated demand
kW IMP PRD kW IMP PRD DMD Predicted sliding window demand
PF IMP@kVA MD PF IMP@kVA MXDMD PF (import) at maximum kVA demand
REG1 ACD SUM REG1 ACC DMD Billing summary (total) register accumulated demand
REG1 BD SUM REG1 BLK DMD Billing summary (total) register block demand
REG1 SD SUM REG1 SW DMD Billing summary (total) register sliding demand
REG1 SUM REG1 Billing summary (total) energy register
REG1 MD SUM REG1 DMD MAX Billing summary (total) register maximum demand
REG1 TRF1 TOU REG1 TRF1 Billing tariff energy register
REG1 TRF1 MD DMD1 TRF1 MAX Billing tariff register maximum demand
TRF1 SEASON TRF1 Generic billing tariff energy register
TRF1 MD SEASON TRF1 Generic billing tariff register maximum demand

162 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix E Billing/TOU Profile Log File

Appendix E Billing/TOU Profile Log File


The following table shows the record structure for the daily billing data
profile log file.
The second column shows data abbreviations used in the PAS data log
reports. Data log file sections are highlighted in bold.

Table 26: Daily Billing/TOU Profile Data Log (Data Log #16)

Field No. Designation Description


Energy Register #1
1 REG1 Summary (total) energy reading
2 TRF1 Tariff #1 energy reading
3 TRF2 Tariff #2 energy reading
4 TRF3 Tariff #3 energy reading
5 TRF4 Tariff #4 energy reading
6 TRF5 Tariff #5 energy reading
7 TRF6 Tariff #6 energy reading
8 TRF7 Tariff #7 energy reading
9 TRF8 Tariff #8 energy reading

Energy Register #4
1 REG4 Summary (total) energy reading
2 TRF1 Tariff #1 energy reading
3 TRF2 Tariff #2 energy reading
4 TRF3 Tariff #3 energy reading
5 TRF4 Tariff #4 energy reading
6 TRF5 Tariff #5 energy reading
7 TRF6 Tariff #6 energy reading
8 TRF7 Tariff #7 energy reading
9 TRF8 Tariff #8 energy reading
Daily Maximum Demand Register #1
1 REG1 MD Summary (total) max. demand reading
2 TRF1 MD Tariff #1 max. demand reading
3 TRF2 MD Tariff #2 max. demand reading
4 TRF3 MD Tariff #3 max. demand reading
5 TRF4 MD Tariff #4 max. demand reading
6 TRF5 MD Tariff #5 max. demand reading
7 TRF6 MD Tariff #6 max. demand reading
8 TRF7 MD Tariff #7 max. demand reading
9 TRF8 MD Tariff #8 max. demand reading

Daily Maximum Demand Register #4
1 REG4 MD Summary (total) max. demand reading
2 TRF1 MD Tariff #1 max. demand reading
3 TRF2 MD Tariff #2 max. demand reading
4 TRF3 MD Tariff #3 max. demand reading
5 TRF4 MD Tariff #4 max. demand reading
6 TRF5 MD Tariff #5 max. demand reading
7 TRF6 MD Tariff #6 max. demand reading
8 TRF7 MD Tariff #7 max. demand reading
9 TRF8 MD Tariff #8 max. demand reading

PM135 Powermeter Series 163

Downloaded from www.Manualslib.com manuals search engine


‎Appendix E Billing/TOU Profile Log File

The number of parameters in each section is automatically configured


depending on the number of actual tariffs you defined in the TOU Daily
Profiles.

164 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Appendix F Data Scales

Appendix F Data Scales


The maximum values for volts, amps and power in the PM135 setup and
in communications are limited by the voltage and current scale settings.
See Device Options in Chapter 4 on how to change the voltage and
current scales in your meter.
The following table shows the meter data scales.

Table 27: Data Scales Values


Scale Conditions Range
Maximum voltage All configurations Voltage scale  PT Ratio, V 1
(V max)
Maximum current All configurations Current scale  CT Ratio, A 2, 3
(I max)
Maximum Power 4 Wiring 4LN3, 3LN3, 3BLN3 V max  I max  3, W
Wiring 4LL3, 3LL3, 3BLL3, V max  I max  2, W
3OP2, 3OP3, 3DIR2
Maximum frequency 25, 50 or 60 Hz 100 Hz
400Hz 500 Hz

1 The default voltage scale is 144V. The recommended voltage scale is


120V+20% = 144V for using with external PT’s, and 690V+20% = 828V
for a direct connection to power line.
2 CT Ratio = CT primary current/CT secondary current

3 The default current scale is 2  CT secondary (2.0A with 1A secondary


and 10.0A with 5A secondary).
4 Maximum power is rounded to whole kilowatts. With PT=1.0, it is
limited to 9,999,000 W.

PM135 Powermeter Series 165

Downloaded from www.Manualslib.com manuals search engine


‎Appendix G Device Diagnostic Codes

Appendix G Device Diagnostic Codes


Table 28: Device Diagnostic Codes
Diagnostic Diagnostic Description Reason
Display Message
RAM/DATA fault RAM/DATA Fault Memory/Data fault Hardware failure
HW watchdog HW Watchdog Hardware watchdog Hardware failure
reset Reset reset
CPU exception CPU Exception CPU exception Hardware failure
Run-time error Run-time Error Run-time software Hardware failure
error
SW watchdog SW Watchdog Software watchdog Hardware failure
reset Reset timeout
Power down/Up Power Down Power Down/Up Normal power-up sequence
External restart Device Reset Warm restart External restart via
communications or by
firmware upgrade
Configuration Configuration Configuration reset Corrupted setup data has
reset Reset been replaced with the
default configuration
RTC fault RTC Fault RTC fault The clock time has been lost
EEPROM fault EEPROM Fault EEPROM fault Hardware failure

See Diagnostics Display in Chapter 3 for more information on the PM135


built-in diagnostics.

166 PM135 Powermeter Series

Downloaded from www.Manualslib.com manuals search engine


‎Index

Index
A
L
Accuracy, 149
Active energy, 149 Low Resolution Option, 46
Active power, 149
analog outputs, 15, 34, 81, 88, M
89, 151
Apparent energy, 149 Maximum Demand Load Current,
Apparent power, 149 80
Modbus, 112

B
N
billing, 101
Broken Delta, 26, 30, 81 notification, 78

C O

communication, 66, 72 Open Delta, 26, 28, 29, 81


counters, 89
CT Primary Current, 80 P
PAS, 64
D password, 62, 71
Data recorder, 108 Phase Energy, 83
daylight saving, 97 Power Source, 24
Delta, 26 Profibus, 12, 14, 15, 36, 39, 73,
demand, 18, 81 147
diagnostic, 118 PT Ratio, 80, 167
digital inputs, 14, 15, 16, 85, 86, Pulse LED, 61
89, 93, 98, 101
display, 42, 46, 94 R
DNP3, 113
Reactive energy, 149
Reactive power, 149
E relay outputs, 14, 15, 31, 33, 86
Ethernet, 12, 14, 15, 36, 38, 64, reset, 116
66, 67, 73, 74, 75, 76, 147
Event recorder, 108 S
eXpertPower, 77
setpoints, 91
Starting Voltage, 83
F
factory defaults, 107 T
test mode, 61, 83
G Total Demand Distortion, 149
GPRS, 14, 16, 41, 73, 76, 77, Total Harmonic Distortion, 149
119 TOU, 101
TOU module, 16, 35

H
W
High Resolution Option, 46
Wiring Mode, 6, 27, 28, 29, 30,
80, 81
I Wye, 26, 27, 28, 29, 81
inputs, 18

PM135 Powermeter Series 167

Downloaded from www.Manualslib.com manuals search engine

You might also like