You are on page 1of 22

Nipro Trading 

 (Shanghai) Co., LTD.
Reduce the labor burden
 Automatic priming & reinfusion
 IC card & Network system
 Flexibility
 Hygiene and disinfection
 User friendly & Safety
Automatic priming & reinfusion*
Automatic function by saline.
HDF model uses saline or dialysate.

 Automatic priming
 Automatic reinfusion
 Automatic drainage
IC card & Network system
 IC card system*
Patient treatment setting is 
written into NCU‐18 via IC 
card which is written to each 
card from PC beforehand.
 Network system*
NCU‐18 outputs data to PC via 
network cable.
Flexibility
 HD double needle
 HD single needle
 HDF double needle; pre or post dilution*
 HDF single needle; pre or post dilution*
 HF; pre or post dilution*
 SQUF (Sequential UF)
 Dialysate flow; 100ml/min to 800ml/min
 2 central concentration connectors*
 Advanced accurate Kt/V calculation "Dose finder"
 Bicarbonate and sodium profiling
 UF profiling
Kt/V calculation "Dose finder"
 Data input for Female;
 Size[cm]
 Weight[kg]

 Data input for Male
 Size[cm]
 Weight[kg]
 Age[years]

 8 patients can be registered
Bicarbonate and sodium profiling
UF profile

 7 patterns can be stored
 A pro and B pro can be used at a same time
 SQUF is available during UF profile
Hygiene and disinfection
 Arterial and venous drip chamber level control*
 ETRF filter for dialysate, CF‐609N*
 Leak check function for Online HDF filter CF‐609N*
 Bicarbonate cartridge function*
 Heat disinfection
 IHR (Integrated Hot Rinse)
 Automatic disinfection programs
 Sheath‐type dialysis concentration suction
ETRF filter, CF‐609N
Bicarbonate cartridge function
 cartridge change in all phases of 
operation .
Automatic disinfection programs

 Citric acid hot: 37 minutes
 Peracetic acid: 39 minutes
 S‐hypochlorite: 40 minutes
 Low disinfectant consumption  (< 70 ml citric acid)
 Weekly disinfection program
 Disinfection history
User friendly & Safety
 First aid button
 NIBP*
 300mm width slim body
 Turnable 10.4” inch screen
 English, Spanish, and Chinese (simplified Chinese, 
traditional Chinese) are available on display.
 Power failure backup
 Start up test
 Closed circuit leak test during treatment
 Stand by mode (Qd:100 ml/min )
NIBP

• Automatic Blood pressure control 
linked to:
* Bypass
* UFR
* Blood flow
Product Range 1
 Basic model
 Single Pump
 Dialysate flow; 100ml/min to 800ml/min
 Heat disinfection
 Kt/V calculation "Dose finder“
 Bicarbonate and sodium profiling
 UF profiling
 Sheath‐type dialysis concentration suction
 Power failure backup
 Second disinfectant port
Product  Range 2
 Online model
 Basic model function
 Substitution pump
 CF1 ,CF2 unit, and CF leak check unit (Including CF‐609N)
 Drain port
Option 1
 NIBP
 Remote SW for NIBP
 CF1 unit
 CF2 unit
 CF leak check unit
 Bicarbonate Powder Assy (Factory option)
 Central supply system (Two A solution or A and B solution)
Option 2
 Heat Exchanger Assy
 Auto priming and Auto blood return (Factory option)
 External network
 Drain port
 Nurse call
 IC card (Factory option)
 Solution tank tray for dialysate
 Solution tank tray for disinfectant
Comparison 1
Model NCU-18 4008S Artis Dialog+
Dimensions 300(W)*460(D)*1480(H) 390(W)*340(D)*1330(H) 660(W)*690(D)*1552(H) 510(W)*637(D)*1678(H)

Weight approx:100Kg approx:80kg approx:130kg approx:85kg

Display 10.4 inch touch display 10 inch touch display Touch display 15 inch touch display

Hydraulic control Piston pump

Dialysate flow 100 to 800ml/min 0,300,500,800ml/min 300 to 800ml/min 300 to 800ml/min


range

Max Blood flow 400/600 ml/min 600/600ml/min /500ml/min /600ml/min


(ID6.35/8.00)
Comparison 2
Model NCU-18 4008S Artis Dialog+
Heat
O (with IHR) O O O
disinfection

B-powder O O O O

Online HDF O O O O

Central supply
O O O O
system
BTS panel
O O O O
mount
Kt/V
O X X O
(calculation)
Comparison 3
Model NCU-18 4008S Artis Dialog+
Kt/V (on-line) X O O O

BVM X O O O

Sheath type
O O O O
AB port
External
O O O O
network
Fluid level
O O O O
adjustment
Patient card
O X O O
system
Comparison 4
Model NCU-18 4008S Artis Dialog+
Auto
O X X X
Reinfusion

Integrated BP X X O X

Double Pump
X X X X
SN

Language eng,chi,spa eng eng eng,chi

Regulation FSC,CE,CCC CE,CCC CE CE,CCC


Thank you for your attention!

Optimized treatment for patient & medical staff

2012.12

You might also like