WBC/Baso and LMNE Principles
 Contents:
MDSS : distribution
WBC/Baso circuit
WBC/Baso dilution
LMNE dilution
LMNE transfer dilution
LMNE injection
WBC formula
Absorbance measurement
Alarms

3
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
MDSS : Distribution

Sampling Distribution Mixing

4
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
MDSS : Distribution

WBC/BASO

Blood 10µL
RBC/PLT Basolyse II 2mL
Final Dil. : 1/200
1st Dil. 42,5µL
LMNE Diluent 2.5mL
Blood 25µL Final Dil. : 1/10000
Eosinofix 1mL
1stdilution
Diluent 1mL
HGB Final Dil.: 1/80
Rinsing
Blood 10µL
Blood 3µL
Diluent 2.1mL
Blood 5µL
Lyse 0.4mL
Final Dil. : 1/250

5
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC/Baso circuit

WBC/Baso
chamber

6
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC/baso dilution
The electronic measurement
principle is the same as the
RBC/PLT.
• An aliquot of 10 µl of blood Sampling needle
is pourred into the WBC/Baso
chamber with a synchronized
flow of 2 ml of Basolyse II.
• Reaction temperature : 35°Celsius

• Incubation time : 9 sec

The Basolyse II lyses RBC, lyses PLT,


lyses WBC which release their nucleus, P R L M N E B
excepted for the basophils population.

7
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC counting
WBC / Baso
Chamber
Vacuum WBC pulses

Count 1 (5 seconds) Count 2 (5 seconds)

2 counts of 5
seconds

𝐖𝐁𝐂 =
Aperture (𝑾𝑩𝑪𝟏+𝑾𝑩𝑪𝟐)
𝟐
×𝑲wbc×𝒌
WBC count (80µm)
Kwbc = WBC calibration coefficient
Impedance K = coïncidence coefficient
principle
8
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC Formula
Then, using WBC/Baso chamber, trought an aperture of 80µm, count is done
and volumetric pulses are obtained. These pulses are then displayed on a
volumetric histogram :

Number

Volume

WBC nuclei Baso

If we get a total of 300 pulses, with 297 between 1 & 2 thresholds,


and 3 between 2 & 3 threshold, we have 3/300 = 1% of basophils.

9
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
LMNE Dilution
• The blood aliquot (25µl) is poured
into a synchronised tangential flow
of 1ml of Eosinofix into the LMNE Sampling needle
chamber

• Incubation time : during 12


secondes at 35° Celsius.

During incubation, the Eosinofix


1 lyses the RBC,
2 gives a specific black coloration
to Eosinophils
3 protect the others WBC volume.
P R L M N E B
Then 1 ml of diluent is poured
into LMNE chamber, dilution is 1/80

10
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
LMNE transfer dilution principle

LMNE chamber

LMNE injection syringes


11
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
LMNE injection principle

Flowcell

Back to LMNE chamber

1st sheath with


diluent
2nd sheath with diluent

LMNE syringes
12
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC Formula

Volumetric absorbance
measurement measurement

The percentage of
the white blood cell
populations is
obtained by flow
2nd sheath with diluent
cytometry.

Injection speed : 12 km/h


1st sheath with diluent
Injected volume : 72 μL

13
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC Formula
•Constant speed, it is possible
with the transfer time to check
that the volumetric pulse and transfer time
the absorbance pulse belong to 200 µs
the same cell (correlation
notion)

• In case of large quantity of absorbance


WBC (ex: hyperleukocytosis, volume
leukemia), the system will
take into account the
volumetric and absorbance
informations of the current
cell prior to perform the
measurement for a next one

14
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Absorbance measurement
Crossing light
Diffracted light

 Emitted light
Transmitted light Absorbed light

Absorbance All of the cell characteristics


Emitted light will be taken into account.
Transmitted light

15
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC Formula
Cell Aspect Volume Lost light
(Absorbance + diffraction)

Lymphocytes Small A little

Monocytes Big A little +

Neutrophiles Medium Medium

Eosinophiles Bigger due to the Big (Reagent Action)


Reagent action

16
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
WBC Formula
WBC localisation on matrix
Eosinophils

Neutrophils

Lymphocytes
Large immature cells

Atypical Lymphocytes Monocytes

17
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Raw Values

The raw counts are printed at the bottom of the Report printout.
Avalaible on printout only

18
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Raw Values

If activated, the raw counts are printed at the bottom of the Report printout.

19
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Raw Values (CBC)

WBC = (WBC1+WBC2)/2 x Kwc x k


Example :
Kwc = calibration coefficient of WBC
WBC = 6,6 103/mm3
K = coïncidence coefficient

20
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Raw Values (DIFF)

LIC = RN + RM + C127

DIFF# = LYM + MON +


NEU + EOS
DIFF#
CO % ≈ x 100
RA #
21
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Alarms
WBC / Baso Alarms :
L1 alarm can be triggered out either on DIF or CBC mode.
-> WBC result will be flagged with a « ! » (suspicion point)

Values for L1
if 3 % or 200 #

Suspected abnormalities:
PLT aggregates
NRBCs
Small lymphocytes

22
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Alarms
5 DIFF Alarms and others populations :

2% 80#

3% 100#

5% 55#

23
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Alarms

1.8% 60#

0.5% 40#

0.7% 60#

24
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Alarms
► Separation alarms:

2% 120#

1.2% 100#

1.1% 30#

25
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Alarms

►Other populations:
+ Channel 127

3% 0.3#

2.5% 0.25#

26
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Alarms
NO:
Lymphocytes – Monocytes – Neutrophils - Eosinophils
Platelets aggregates NE:
NRBC’s Young eosinophils
Erythrocyte membrane Hypersegmented
Large number of platelets neutrophils
Pollution Imature cells

LN: LIC:
Neutrophil destruction (old blood) Large monocytes
Contamination Large neutrophils
Stroma Myelocytes..
Platelet aggregates

NL: RN:
Slighty neutrophils Large neutrophils
Segmented lymphocytes Imature cells
Activated lymphocytes (metamyelocytes,
Hyposegmented lymphocytes promyelocytes..)

RM:
LL / LL1 Large monocytes
Small lymphocytes Myelocytes,
Platelet aggregates Promyelocytes
NRBC’s Large blasts
Resistant RBC
MN:
ALY: Hyperbaosophilic
Large lymphocytes monocytes
Plasmocytes Young neutrophils
27
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Alarms

Other alarms:

• LMNE+; LMNE- : if the difference in + or in – between the WBC/Baso channel


and the LMNE channel are higher than the defined conditions.
-> WBC result will be flagged with a « ! » (suspicion point)

• BASO+ : if the Baso% found in the WBC/Baso is greater than 50 %


-> WBC result will be flagged with a « ! » (suspicion point)
and BASO result will not be reported « --.-- ».

• CO : if correlation between resistive and optical measurements in the LMNE


channel is less than 50 %
-> LMNE results will not be reported « --.-- »

28
12O P60 5DIFF BASO LMNE EN Ind A
Restricted Diffusion
© 2007 HORIBA, Ltd. All rights reserved.
Any questions ?