I. General
1. Specifications
2. Configuration
II. Assembly / Disassembly
1. Main Body
2. Doppler Probe
III. Theory of Operation
1. Doppler Probe Block
2. Main Block
IV. Troubleshooting
I. General
1. Specifications
Ultrasound Frequency :
- Fetal Doppler: 2MHz
- Vascular Doppler: 4MHz, 5MHz, 8MHz
Intensity : <10mW/cm2
Heart Rate Counting Range : 50~240bpm
FHR Accuracy : ±2% of range
Battery Type : 1.5V x 2(AA Type)
Battery Life : About 360minutes (Continuously use)
PC Interface : Sound Card
2. Configuration
• Hl-bebe (Fetal Doppler)
Main Body
Operator's
Manual
Probes
^
(4, 5, 8MHz)
II. Assembly / Disassembly
1. Main Body
hooks
...
TX Signal
Doppler Demodulation
Sensor Doppler
Signal
2. Main Block
LCD
1L
Power
IV. Troubleshooting
Power Trouble