0006 LCD FAULT

Detection Always
Control section only
Detecting Communication error has occurred with display section (LCD).
condition ・Transmission retry error has occurred (3 times at intervals of 3 sec.).

Operation Lamp Sound BP stop HP stop Clamp close Bypass UF stop Heater OFF Hyd. line stop
Red blinks Beep ○ ○ Venous line ○ ○ ○ ○
Reset Turn OFF and ON power supply.
condition
Note Stops all operations in present process.

Possible cause Analyzing/Measures


Malfunction or freeze of CPU Analyzing: Turn off the machine completely and turn on the
machine. Check 1 and 2.
※If you cannot turn off the I/O sw, turn off the power
breaker and turn on the machine.
1. Check cooling fan works.
2. Check the environment around the machine
whether the equipment which generate the
electromagnetic noise.
Measures: 1. Clean the air filter. If cooling fan does not work,
change it with new one.
2. Take the equipment which generate the noise
from the machine.
After reboot the machine, check the setting including
the treatment.
※Even this alarm appears after the reboot,
do the following check.

Harness between mother PCB and LCD display PCB Measures: Turn off the machine and reconnect the harness.
is disconnected. See Maintenance Manual 11.5(wiring diagram)

Harness between mother PCB and LCD display PCB Measures: Replace the harness.
is damaged. See Maintenance Manual 11.5(wiring diagram)

Power to LCD display PCB is not normal. Analyzing: Check the connection of power supply unit, mother
PCB and LCD display PCB.
See Maintenance Manual 11.5(wiring diagram)
Measures: Reconnect the harness.
Malfunction of LCD display board Measures: Replace LCD display board
See Maintenance Manual 13.3.

Malfunction of mother board Measures: Replace mother board.


See Maintenance Manual 13.7.

Malfunction of power supply unit Measures: Replace power supply unit.


See Maintenance Manual 12.1.