0007 BACK-UP MEMORY VERIFY FAULT

Detection Always
Control section only
Detecting Checksum error occurred in the backup memory.
condition Discrepancy was found by inversion check.

Operation Lamp Sound BP stop HP stop Clamp close Bypass UF stop Heater OFF Hyd. line stop
Red blinks Beep ○ ○ Venous line ○ ○ ○ ○
Reset
Turn OFF and ON power supply.
condition
Note Stops all operations in present process.

Possible cause Analyzing/Measures


Abnormal data in Battery Back-UP RAM on the Analyzing: Check whether Calib. Data is normal value.
Control board See Maintenance Manual 14.3(Setting Items on Each Screen)

Measures: Initializing the "Parameter" and "Calibration" data.


See Maintenance Manual 16.1
NOTE: Before operate the Initializing, please take memo for all data.
And reimput the data to the mahcine after initialized.
NOTE: Check whether "Message history" or "Maintenace data" value
are normal value. If you found any abnormal value, please Clear them.

Due to the damaging of harness between Conductivity boAnalyzing: ・Check whether harness between Conductivity board and
and Mother board.   Mother board is not damaged.
See following photo
photo.

・The damaged point touches the heat sink of the Pump 1 driver IC
Measures: Taping the harness for prevent the short circuit again.

Abnomal power supply to the Control PCB. Analyzing: ・Check whether harness between Power supply and Mother
  board is connected properly.
See Maintenance Manual 11.5 (Brock diagram)
・Check no loosing the connection between Mother board and
Control PCB.
Measures: Connect the harness properly

Malfunction of control board Measures: Replace control board.


See Maintenance Manual 13.4

Malfunction of power supply unit Measures: Replace power supply unit.


See Maintenance Manual 12.1