Professional Documents
Culture Documents
EN Calibration Source M390S - Manual PDF
EN Calibration Source M390S - Manual PDF
Service Centers
LumaSense Technologies, Inc. LumaSense Technologies GmbH
North America Other Than North America
Sales & Service Sales & Support
Santa Clara, CA, USA Frankfurt, Germany
Ph: +1 800 631 0176 Ph: +49 (0) 69 97373 0
Ph: +1 408 727 1600 Fax: +49 (0) 69 97373 167
Fax: +1 408 727 1677
E-mail info@LumaSenseinc.com
support@LumaSenseinc.com
eusupport@LumaSenseinc.com
supportdk@LumaSenseinc.com
Website http://www.LumaSenseinc.com
Part No 511-0001-01- EN
Revision A
August 2014
Contents
1 General Information...................................................................................................... 5
1.1 Information about the user manual..................................................................... 5
1.1.1 Legend ................................................................................................................. 5
1.2 Safety ..................................................................................................................... 5
1.2.1 Additional M390 Safety Notifications ................................................................. 5
1.3 Limit of Liability and Warranty ............................................................................ 6
1.4 Unpacking the Instrument .................................................................................... 6
1.5 Service Request, Repair, or Support ..................................................................... 6
1.6 Shipments to LumaSense for Repair .................................................................... 7
1.7 Transport, Packing, Storage .................................................................................. 7
1.8 Disposal / decommissioning ................................................................................. 7
2 Introduction ................................................................................................................... 9
2.1 Description............................................................................................................. 9
2.2 Technical Data ..................................................................................................... 10
3 Installation................................................................................................................... 11
3.1 Installing the M390 ............................................................................................. 11
3.2 Water and Purge Gas Supply Requirements ...................................................... 11
3.2.1 Water ................................................................................................................. 11
3.2.2 Purge Gas........................................................................................................... 11
4 Operation ..................................................................................................................... 13
4.1 Starting Up........................................................................................................... 13
4.1.1 Front Panel Push Buttons .................................................................................. 13
4.2 Changing the Setpoint ........................................................................................ 13
4.2.1 Other Controller Modes .................................................................................... 14
4.3 Shutting Down .................................................................................................... 14
4.4 Changing Temperature Units .............................................................................. 14
4.5 Setting Serial Communications .......................................................................... 15
4.5.1 RS485 Option ..................................................................................................... 16
4.6 PID Tuning Constants .......................................................................................... 17
4.7 Controller Calibration ......................................................................................... 17
4.8 Switch Comm Mode from Ei Bisync to Modbus ................................................. 19
5 Maintenance ................................................................................................................ 21
5.1 Cavity Life ............................................................................................................ 21
5.2 Safety Interlocks.................................................................................................. 21
5.3 Periodic Inspections ............................................................................................ 21
5.3.1 Pyrometer Alignment Check ............................................................................. 21
5.4 Furnace Core Disassembly .................................................................................. 22
5.5 Fuse Locations ..................................................................................................... 23
5.6 Cleaning ............................................................................................................... 24
Index .................................................................................................................................. 25
Warnings and Cautions: The general warnings and cautions symbol signifies the
potential for bodily harm or damage to equipment.
1.2 Safety
This manual provides important information on safely installing and operating the
M390S blackbody. Several sections of this manual provide safety warnings to avert danger.
These safety warnings are specified with a warning symbol. You must read and understand the
contents of this manual before operating the instrument even if you have used similar
instruments or have already been trained by the manufacturer.
It is also important to continually pay attention to all labels and markings on the instrument and
to keep the labels and markings in a permanent readable condition.
7. Save all packing materials, including the carrier’s identification codes, until you have
inspected all components and find that there is no obvious or hidden damage.
Note: Before shipment, each instrument is assembled, calibrated, and tested at the
LumaSense Factory. If you note any damage or suspect damage, immediately contact
the carrier and LumaSense Technologies, Inc.
2.1 Description
The M390 series are high temperature blackbody calibrators that can produce a very high
temperature, high emissivity targets, and at the same time stabilize at the required temperature
within a few minutes of switch-on. The cavity is a graphite tube target which is available with a
16mm (.625") diameter, or optionally 25mm (1.00") or 38mm (1.5”) aperture with an effective
emissivity of 0.995. Target is sensed by a rapid response LumaSense infrared fiber optic
thermometer which drives a PID controller to regulate the target temperature precisely to the
desired value. The controller is mounted on the calibration source and remote set point
programming may be achieved via an RS422/485 communication port.
Blackbody Exit
Aperture
Alarms
Flow Meter
Main Power
Note: The Blackbody Cavity (or tubular heating element) is normally shipped in place.
Note: If the element or quartz tube is damaged or requires replacement, please see
the maintenance section for cavity replacement procedure.
Warning: DO NOT USE NITROGEN (N2) FOR PURGE GAS. AT HIGH TEMPERATURES,
CYNIDE GAS COULD BE GENERATED.
4.1 Starting Up
To start up the M390S:
1. Turn on the power switch/breaker and allow 20 seconds for the system to self test.
Warning: Do not connect or disconnect while power is on. Do not attempt to operate
without cables connected.
12. Adjust the set point to the desired temperature and press auto/manual button.
4.1.1 Front Panel Push Buttons
PAGE Engages access to controller parameters. Do not alter any
configurations.
Blackbody Setpoint
Eurotherm controller
M390 Unit 2
(Address 2)
Note: A CD containing a sample program is included with your order. For more details
on communications, refer to the electronic communication manual also found on the
CD provided.
Note: You may download Eurotherm iTools Software from this web site:
http://www.eurotherm.com/products/controllers/itools/
With this software and some practice, you will see this GUI interface screen. To permit
iTools to talk to your blackbody, you may have to set the controller comm. port to
“Modbus” mode. Procedure to change communications to Modbus are shown in
Section 4.8.
1 2 3 4
1. Press and Hold 1 until ACCESS appears. If Access Level and Level 1 appears, skip to
step 3.
2. Press 4 until Level 3 appears.
3. Press 4 until 3 appears. Wait and "pass" appears.
4. Press 1 until Lin16 appears.
5. Press 2. until display shows the Output # that corresponds to the calibration point
shown in the listing on the previous page.
6. Use UP/DWN buttons (3 and 4) to dial in the measured radiance temperature from a
freezing point transfer device or other radiance standard. Adjust Out# only!
7. Press 2 to view/change more correction points.
8. When finished with all calibrations and to lock the controller, press 1 until ACCESS
appears.
9. Press 3 until Level 1 appears.
Example: In place of Silver freezing point correction at 961.78 °C, you prefer to use 1000 °C.
1. Using the above steps, set INPUT 9 equal to 1000.0. Set OUTPUT 9 equal to 1000.0.
2. Allow blackbody to control and stabilize at 1000.0 °C.
3. With your radiance standard, you measure 1001.2 °C.
4. Use the above procedure to dial in 1001.2 in the OUTPUT 9 location.
5. Wait for blackbody to re-stabilize at 1000.0 and verify radiance temperature.
All corrections between calibration points are performed using straight-line interpolation.
1 2 3 4
1. Attach RS232 cable to computer using the comm. cable that came with the kit (straight
DB9 cable).
2. Open iTools program on PC.
3. Press SCAN button on upper bar of program.
4. Select YES. If connected properly, you should see hard disk activity when SCAN
channel 1 is on.
5. After the program ‘sees’ the controller (a picture of a controller appears at lower left of
screen).
6. Follow directions in Section 3.2.5 to access the controller radiance temperature LUT
table.
7. Reset the communications mode to EIBISYNCH using the steps above except replace
Modbus with EIBISYNCH in step 6 above.
Note: The distance between electrode ends should be at least 1/4" less than the quartz
tube.
8. Ensure that a firm electro-mechanical connection is made within the tapered joints. Tap
gently with a rubber mallet to ensure firm joints.
9. Turn on water and check all connections.
10. See Section 4.1 to operate the unit*.
* Keep in mind that the temperature should be gradually increased by 100 °C every
15-20 minutes for the initial cavity baking. A temporary smell of sulfur may be noticed
during this time.
Rear
Electrodes Slide this electrode out to Electrode tension
repair/replace cavity. adjustment
Push buttons 13
C Pyrometer Alignment Check 21
Cavity life 21
Cleaning 24 R
Controller Calibration 17
Repair 6, 7
Controller Modes 14
RS485 16
D
S
Disposal 7
Safety 5
Safety Interlocks 21
E Serial Communications 15
Eurotherm iTools Software 17 Service Request 6
Setpoint 13
Shutting Down 14
F Starting Up 13
Furnace Core Disassembly 22 Storage 7
Fuse Locations 23 Support 6
G T
General Information 5 Technical Data 10
Temperature Units 14
Transport 7
I
Inspections 21 U
Installation 11
Unpacking the Instrument 6
L
W
Legend 5
Liability 6 Warranty 6
Look-up Table 18
M
Maintenance 21
Modbus 19
O
Operation 13
P
Packing 7
PID Tuning Constants 17
Purge Gas Supply Requirements 11