Specification: ME 900

SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax:+86-755-26431232
Email: info@szcomen.com
Web: www.comen.com

2021 V 1.0
Infusion Pump
ME 900

Physical Characteristics
Dimensions: 216mm×135mm×77mm (L*W*H) Interface
Weight ≤1.5kg (Including battery) AC power interface
Screen Size: 3.5” capacitive touch screen TFT LCD Multipurpose Jack that connects to
with Auto-Brightness Adjustment
expansion cable
Classification: Type CF, Defibrillation-proof, Class I
Drop sensor Port
Operating Conditions
Dynamic Pressure System (DPS)
Temperature: 5~40°C
Display: DPS indicator displayed as pie chart with
Humidity: 15%~95% (non-condensing)
dynamic pressure segment limits for
Power
occlusion alarm
requirement: 100-240V~, 50/60Hz±1Hz
Selectable patient Neonatal, Pediatric, or Adult use
Battery Type: Rechargeable Lithium-ion battery
types:
Battery Capacity: 2200mAh
Pressure Levels: P1-P12 selectable depending on patient
Battery Charging Not more than 4 hours to full charge
type
Time: when powered off
Neonatal: P1-P5
Battery Up to 9 hours of continuous working
150 mmHg~450mmHg
performance: (at 25ml/h)
Pediatric: P4-P9
Working modes: Rate Mode, Time Mode, Drip Mode,
375 mmHg~750 mmHg
Weight Mode, Sequential mode, First
Adult: P5-P12
dose mode, Ramp mode, Micro mode,
450 mmHg~975 mmHg
Intermittent Mode, Dose Time Mode
Ergonomics
Infusion system: Peristaltic
Space-saving: Stackable and allows docking with other
Indicators
units of M800 or ME900 in any desired
Hardware Battery status indicator: Charging,
combination onto an MX8900
Indicators: Battery in use, Battery low/depleted
Workstation.
AC Power indicator
Each bracket allows slotting of 1~4
Alarm Light bar indicator
pumps.
On-Screen Dynamic Pressure System (DPS)
Stackable height: 1~3 brackets to form a
Indicators:
tower.
Infusion related information:
Maximum allowed interconnected
Selected drug, Mode, Left Vol, Volume,
towers: 2 towers (equivalent of 3+3=6
Rest Time, Injection Rate, Alarm status,
brackets, or 12+12=24 total pumps).
Battery status, Syringe brand and size
Mounting Allows for mounting through a pole
Running(Infusion status) bar
Solutions: clamp for either stationary use on a bed ≥20d/ml infusion set: 1~133d/min
rail or portability through the use of a The maximum drip rate can be
trolley (TC800). calculated according to the maximum
Portability: Accommodates a carrying handle that flow rate of 400.00ml/h and the drop
supports up to 4 times the weight of 12 factor of the infusion set;
pumps, creating an easy to pick up and Drip rate (d/min) = Flow rate (ml/h)
carry workstation. For stacked pumps, a ×Drop factor (d/ml)/ 60; Minimum
trolley solution is available. amount: 1; unit: d/min.
VTBI range: 0.00~9999.99ml
Modes: Time range: 00:00:01~99:59:59 (h:m:s)
Rate Mode setting ranges:
Parameters: Drug type selection, Rate, Time and Weight Mode setting ranges:
Volume To Be Infused (VTBI) can be set Parameters: Drug type selection, Drug Concentration,
to automatically calculate the remaining Patient Weight, Dose Rate and Volume
parameter. To Be Infused (VTBI) can be set to
Rate range: 0.10~2000.00ml/h, Step:0.1ml/h; automatically calculate the infusion rate.
Infusion Accuracy: ≤±4.5% (For specified Rate range: 0.10~2000.00ml/h, Step:0.1ml/h;
and calibrated infusion sets, accuracy of Infusion Accuracy: ≤±4.5% (For specified
infusion: ≤±3%) and calibrated infusion sets, accuracy of
VTBI range: 0.00~9999.99ml infusion: ≤±3%)
Time range: 00:00:01~99:59:59 (h:m:s) VTBI range: 0.00~9999.99ml
Time Mode setting ranges: Time range: 00:00:01~99:59:59 (h:m:s)
Parameters: Drug type selection, Rate, Time and
Volume To Be Infused (VTBI) can be set Sequential Mode setting ranges:
to automatically calculate the remaining Parameters: Drug type selection, Time, Dose rate and
parameter. Volume to be infused (VTBI) can be set
Rate range: 0.10~2000.00ml/h, Step:0.1ml/h; for each sequence to automatically
Infusion Accuracy: ≤±4.5% (For specified calculate the remaining parameter and
and calibrated infusion sets, accuracy of display a total VTBI.
infusion: ≤±3%) Sequences: Up to 10 sequences per pump.
VTBI range: 0.00~9999.99ml VTBI range: 0.10~9999.99ml; Step: 0.01ml. The total
Time range: 00:00:01~99:59:59 (h:m:s) sum of VTBI for all sequences cannot
exceed 9999.99ml.
Drip Mode setting range: Rate range: 0.10~2000.00ml/h; Step: 0.01ml/h
Parameters: Drug type selection, Infusion set drop Time range: 00:00:01~99:59:59(h: m: s); Step: 1s
factor selection, Drip rate, Time and
Volume To Be Infused (VTBI) can be set. First Dose mode setting range:
Drip rate or time will automatically be Parameters: Drug type selection, Main time, Main
calculated according to VTBI and Drop rate, Main volume, First time, First
factor once either parameter has been rate and First volume can be set.
set. Rate range: 0.10ml/h~2000.00ml/h
Infusion Set Drop 10~60drops/ml Step: 0.01ml/h
Factor range: Time range: 00:00:01~99:59:59(h: m: s); Step: 1s
Drip Rate range: ≥10d/ml infusion set: 1~66/min Volume range: 0.10~9999.99ml; Step: 0.01ml
≥15d/ml infusion set: 1~100d/min
Ramp mode setting range: 0.001~999.999 (editable). when the unit
Parameters: Drug type selection, Total Time, Up of VTBI is ml, the rate unit is ml/h, with a
Time, Down Time and Volume to be range of 0.10~2000.00ml/h.
infused (VTBI) can be set to Volume range: 0.10~9999.99ml; Step: 0.01ml
automatically calculate the Steady Drug concentration 0.001~9999.999ml; Step: 0.001ml; Unit:
Rate. The rate range below refers to setting range: g/ml, mg/ml, ug/ml, ng/ml, U/ml, IU/ml,
the values in which the automatically KU/ml
calculated steady rate may not exceed. Time range: 00:00:01~99:59:59(h: m: s); Step: 1s
Rate range: 0.10ml/h~2000.00ml/h
Step: 0.01ml/h Bolus and Purging ranges:
Volume range: 0.10~9999.99ml; Step: 0.01ml Rate range: 0.10~2000.00ml/h; Step: 0.01ml/h
Time (Total time, 00:00:01~99:59:59(h: m: s); Step: 1s (Default: 800.00ml/h)
Up time, Down If the up time and down time are not Volume range: 0.10~99.99ml
time) range: input, it defaults to 0s.
Micro Mode setting range: Infusion Accuracy:
Parameters: Drug type selection, Rate, Time and Compliance: IEC/EN 60601-2-24
Volume To Be Infused (VTBI) can be set Infusion accuracy: ≤±4.5%
to automatically calculate the remaining (For specified and calibrated infusion
parameter. sets, accuracy of infusion: ≤±3%)
Rate range: 0.10~100.00ml/h Bolus accuracy: ≤±5% or ≤0.02ml, whichever value is
Step: 0.01ml/h larger
Volume range: 0.10~1000.00ml Drip accuracy: ≤±10%
Step: 0.01ml
Time range: 00:00:01~99:59:59 (h:m:s) Drug Library and Information Management:
Intermittent Mode setting range: Drug Library: Up to 5000 drug types can be
Parameters: Drug type selection, Intmt.Vol
registered. (Default 60 types)
(Intermittent Volume), Intmt. Rate
Customisation: Manual input of drug names and
(Intermittent Rate), Keep Rate (Basal
Rate), Interval and Volume To Be parameters along with the soft and
Infused (VTBI) can be set. hard limits is supported.
Intmt. Rate range: 0.10ml/h~2000.00ml/h Patient Information A patient information management
Step: 0.01ml/h Management: system is inbuilt that allows for patient
Volume range: 0.10~9999.99ml; Step: 0.01ml information input and recording.
Keep Rate range: 0.10~5.00ml/h; Step: 0.01ml/h Patient Type: Adult, Neonatal, Paediatric selectable
Time range: 00:00:01~99:59:59 (h:m:s) Other information: Name, Patient ID, Bed No., DOB,
Height, Weight, Age, Sex, Blood type,
BSA, BMA, Doctor in Charge
Dose Time Mode setting range:
Parameters: Drug type selection, Drug Concentration, Alarms:
Time and Volume to be infused (VTBI) Compliance: IEC60601-1-8
can be set to automatically calculate the Alarm level: P1 to P12
Infusion rate. The rates below refer to Alarm volume: Max: 65db (At 1m distance)
the values in which the automatically Min: 45db (At 1m distance)
calculated Infusion rate may not exceed. Occlusion Alarm: Range: 150~975mmHg (P1-P12)
Rate range: When the unit of VTBI is other than ml, Accuracy: ±15% or ±75mmHg,
the rate is dose rate, and the range is
whichever is greater (Keep Vein Open): minutes, then an alarm will sound)
Default setting: 525mmHg (P6) Step:0.01ml/h
Unit: mmHg, kPa, bar, psi Infusion history: Up to 2000 records. Records any system
Supported Alarm High Level Alarms: changes, alarms and actions taken.
Types: Upstream Occlusion alarm, Downstream Supports record log export through USB.
Occlusion alarm, Upstream Pressure Night mode Automatic adjustment of volume and
Sensor Error alarm, Downstream screen brightness
Pressure Sensor Error alarm, VTBI Purging: Manual and automatic selectable
Complete alarm, KVO Complete alarm, Software Update: Via USB port
Bubble alarm, Cumulative Bubble alarm, No operation time 1~5min (step: 1min)
Tube off (Infusion set disconnected) settings:
alarm, Drip Error alarm Bubble Detection: Single bubble detection allows selection
Low Level Alarms: of 6 different sizes of bubble. (Bigger
System Error alarm, No Action alarm, size equates to lower bubble sensitivity)
Low Battery alarm, Battery Depleted An inbuilt Accumulated Bubble volume
alarm, Battery Disconnected alarm, function sounds an alarm in the case
Abnormal System alarm, Nearly where accumulated bubble volume
Completed(Near Completion) alarm, exceeds the input threshold even
Standby Time End alarm, Pump Code without activation of single bubble
Conflict alarm, Motor Connection Failed alarm.
alarm, Communications Error alarm, Automatic Flow Algorithm that compensates for reduced
Empty Bottle(Infusion container empty) Compensation: flow rate in specific infusion sets after a
alarm, Tube(Infusion set) Error alarm, period of time of usage
Drip Rate Error alarm, and Abnormal Continuous Allows editing of infusion rate for
Pressure alarm. Titration: ongoing infusions without the need to
Features stop any currently ongoing infusions.
Startup guidance Infusion set installation procedure Others: Infusion line pressure sensor built into
and Infusion set displayed upon device startup in the the infusion set bracket (One upstream
installation: event where an infusion set is not sensor and One downstream sensor),
installed. Automatic screen lock, Bubble sensor,
Infusion set brand selection will be Automatic Pressure Release, Infusion
displayed upon syringe installation. Set Calibration function, Drop sensor,
Syringe selection Brands selectable upon syringe Startup Infusion set installation
parameters: installation: BaoAn,B.Braun, guidance.
(2/3ml, 5ml, 10ml, 20ml, 30ml, 50/60ml) Disposable Infusion 10drops/ml, 15drops/ml, 20 drops/ml,
Supports up to 3 user-defined custom set specification: 60 drops/ml. Manual input allowed,
Infusion Set Brands. with rate limits based on input
Ingress Protection: IP33 specification.
KVO Rate range 0.10~5.00ml/h; (KVO will last for 30
*Notice: Specifications subject to change without prior
notice. All rights reserved by COMEN