Specification: M800

SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming New District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax:+86-755-26431232
Email: info@szcomen.com
Web: www.comen.com

2021 V 1.0
Syringe Pump
M800

Physical Characteristics
Dimensions: 290mm×146.5mm×77.5mm(L*W*H) Interface
Weight ≤1.55 kg (Including battery) AC power input
Screen Size: 3.5” Capacitive touch screen TFT LCD Multipurpose Jack that connects to an
with Auto-Brightness Adjustment
expansion cable
Classification: Type CF, Defibrillation-proof, Class I
Dynamic Pressure System (DPS)
Operating Conditions
Display: DPS indicator displayed as pie chart with
Temperature: 5~40°C
dynamic pressure segment limits for
Humidity: 15%~95%
occlusion alarm
Power
Selectable patient Neonatal, Pediatric, or Adult use
requirement: 100-240V~, 50/60Hz
types:
Battery Type: Rechargeable Lithium-ion battery
Pressure Levels: P1-P12 selectable depending on patient
Battery Capacity: 2200mAh
type
Battery Recharging No more than 6h to full charge when
Neonatal: P1-P5
Time: powered off
150 mmHg~450mmHg
Battery Up to 10 hours of continuous working
Pediatric: P4-P9
performance: (at 5ml/h)
375 mmHg~750 mmHg
Working modes: Rate mode, Time mode, Weight mode,
Adult: P5-P12
Intermittent mode, First dose mode,
450 mmHg~975 mmHg
Sequential mode, Micro mode, Ramp
Ergonomics
mode, Dose Time mode
Space-saving: Stackable and allows docking with other
Indicators
units of M800 or ME900 in any desired
Hardware Battery status indicator: Charging,
combination onto an MX8900
Indicators: Battery in use, Battery low/depleted
Workstation.
AC Power indicator
Each bracket allows slotting of 1~4
Alarm Light bar indicator
pumps.
On-Screen Dynamic Pressure System (DPS)
Stackable height: 1~3 brackets to form a
Indicators:
tower.
Infusion related information:
Maximum allowed interconnected
Selected drug, Mode, Left Vol, Volume,
towers: 2 towers (equivalent of 3+3=6
Rest Time, Injection Rate, Alarm status,
brackets, or 12+12=24 total pumps).
Battery status, Syringe brand and size
Mounting Allows for mounting through a pole
Running(Infusion status) bar
Solutions: clamp for either stationary use on a bed
rail or portability through the use of a
trolley (TC800). Weight mode setting range
Portability: Accommodates a carrying handle that Parameters: The patient's Weight, Drug type
supports up to 4 times the weight of 12 selection, Drug Concentration, Dose
pumps, creating an easy to pick up and rate, Volume to be infused (VTBI) can
carry workstation. For stacked pumps, a be set and the Injection rate is
trolley solution is available. automatically calculated according to
the parameters set above.
Modes: Weight range: 0.10~500.00kg; Step: 0.01kg
Rate mode setting range Volume range: 0.10~9999.99ml; Step: 0.01ml
Parameters: Drug type selection, Time, Dose rate and Drug concentration 0.001~9999.999ml; Step: 0.001ml;
Volume to be infused (VTBI) can be set setting range: Unit: g/ml, mg/ml, ug/ml, ng/ml, U/ml,
to automatically calculate the remaining IU/ml, KU/ml
parameter Dose rate setting 0.001~999.999; Step: 0.001; Unit:
2/3ml 0.10ml/h~90.00 ml/h range: g/kg/h, mg/kg/h, ug/kg/h, g/kg/min,
Step: 0.01 ml/h mg/kg/min, ug/kg/min
5ml 0.10ml/h~150.00 ml/h
Step: 0.01 ml/h Intermittent mode setting range
10ml 0.1ml/h~300.0ml/h Parameters: Drug type selection, Intermittent
Step:0.1ml/h Volume (Intmt. Vol.), Intermittent Rate
20ml 0.1ml/h~600.0ml/h (Intmt. Rate), Keep Rate, Interval and
Step:0.1ml/h Volume to be infused (VTBI) can be
30ml 0.1ml/h~1200.0ml/h set. Intermittent Mode can be set to
Step:0.1ml/h auto-loop.
50/60ml 0.1ml/h~2000.0ml/h 2/3ml 0.10ml/h~90.00 ml/h
Step:0.1ml/h Step: 0.01 ml/h
Volume Range: 0.10~9999.99ml; Step: 0.01ml 5ml 0.10ml/h~150.00ml/h
Injection Accuracy: 0.10<Rate<1.00ml/h, Accuracy≤ ±5% Step: 0.01ml/h
Rate>1.00ml/h, Accuracy≤ ±1.8% 10ml 0.10ml/h~300.00ml/h
Mechanical ≤ ±1% Step: 0.01ml/h
Precision: 20ml 0.10ml/h~600.00ml/h
Unit selection ml/h, g/h, mg/h, ug/h, ng/h, g/kg/h, Step: 0.01ml/h
30ml 0.10ml/h~1200.00ml/h
mg/kg/h, ug/kg/h, ng/kg/h, g/kg/min,
Step: 0.01ml/h
mg/kg/min, ug/kg/min, ng/kg/min , 50/60ml 0.10ml/h~2000.00ml/h
U/kg/min, U/kg/h, KU/kg/min, KU/kg/h, Step: 0.01ml/h;
IU/kg/min, IU/kg/h Volume range 0.10~9999.99ml; Step: 0.01ml
Interval time range: 00:00:01~99:59:59(h: m: s);
Step: 1s
Time mode setting range
Keep Rate range 0.001~999.999; Step: 0.001; Unit:
Parameters: Drug type selection, Time, Dose rate
g/kg/h, mg/kg/h, ug/kg/h, g/kg/min,
and Volume to be infused (VTBI) can
mg/kg/min, ug/kg/min
be set to automatically calculate the
remaining parameter
Ramp mode setting range:
Time range: 00:00:01~99:59:59(h: m: s); Step: 1s
Parameters: Drug type selection, Total Time, Up
Volume range: 0.10~9999.99ml; Step: 0.01ml
Time, Down Time and Volume to be
infused (VTBI) can be set to and Volume to be infused (VTBI) can
automatically calculate the Steady be set to automatically calculate the
Rate. The rates below refer to the remaining parameter
values in which the automatically Volume range: 0.10~1000.00ml; Step: 0.01ml
calculated steady rate may not exceed. Rate setting range: 0.10~100.00ml/h; Step: 0.01ml;
2/3ml 0.10ml/h~90.00ml/h Time range: 00:00:01~99:59:59(h: m: s); Step: 1s
Step: 0.01ml/h
5ml 0.10ml/h~150.00ml/h
Step: 0.01ml/h First Dose mode setting range:
10ml 0.10ml/h~300.00ml/h Parameters: Drug type selection, Main time, Main
Step: 0.01ml/h rate, Main volume, First time, First
20ml 0.10ml/h~600.00ml/h rate and First volume can be set.
Step: 0.01ml/h 2/3ml 0.10ml/h~90.00ml/h
30ml 0.10ml/h~1200.00ml/h Step: 0.01ml/h
Step: 0.01ml/h 5ml 0.10ml/h~150.00ml/h
50/60ml 0.10ml/h~2000.00ml/h Step: 0.01ml/h
Step: 0.01ml/h; 10ml 0.10ml/h~300.00ml/h
Volume range: 0.10~9999.99ml; Step: 0.01ml Step: 0.01ml/h
Time (Total time, Up 20ml 0.10ml/h~600.00ml/h
time, Down time) Step: 0.01ml/h
range: 00:00:01~99:59:59(h: m: s); Step: 1s 30ml 0.10ml/h~1200.00ml/h
Step: 0.01ml/h
Sequential mode setting range: 50/60ml 0.10ml/h~2000.00ml/h
Parameters: Drug type selection, Time, Dose rate Step: 0.01ml/h;
and Volume to be infused (VTBI) can Volume range: 0.10~9999.99ml; Step: 0.01ml
be set for each sequence to Time range: 00:00:01~99:59:59(h: m: s); Step: 1s
automatically calculate the remaining
parameter and display a total VTBI. Dose Time mode setting range:
Sequences: Up to 10 sequences per pump. Parameters: Drug type selection, Drug
2/3ml 0.10ml/h~90.00ml/h Concentration, Time and Volume to be
Step: 0.01ml/h infused (VTBI) can be set to
5ml 0.10ml/h~150.00ml/h automatically calculate the Injection
Step: 0.01ml/h rate. The rates below refer to the
10ml 0.10ml/h~300.00ml/h values in which the automatically
Step: 0.01ml/h calculated Injection rate may not
20ml 0.10ml/h~600.00ml/h exceed.
Step: 0.01ml/h 2/3ml 0.10ml/h~90.00ml/h
30ml 0.10ml/h~1200.00ml/h Step: 0.01ml/h
Step: 0.01ml/h 5ml 0.10ml/h~150.00ml/h
50/60ml 0.10ml/h~2000.00ml/h Step: 0.01ml/h
Step: 0.01ml/h; 10ml 0.10ml/h~300.00ml/h
Volume range: 0.10~9999.99ml; Step: 0.01ml Step: 0.01ml/h
Time range: 00:00:01~99:59:59(h: m: s); Step: 1s 20ml 0.10ml/h~600.00ml/h
Step: 0.01ml/h
Micro mode setting range 30ml 0.10ml/h~1200.00ml/h
Parameters: Drug type selection, Time, Dose rate Step: 0.01ml/h
50/60ml 0.10ml/h~2000.00ml/h
Step: 0.01ml/h; Drug Library and Information Management:
Volume range: 0.10~9999.99ml; Step: 0.01ml Drug Library: Up to 5000 drug types can be
Drug concentration 0.001~9999.999ml; Step: 0.001ml; Unit:
registered. (Default 60 types)
setting range: g/ml, mg/ml, ug/ml, ng/ml, U/ml, IU/ml,
Customisation: Manual input of drug names and
KU/ml
Time range: 00:00:01~99:59:59(h: m: s); Step: 1s parameters along with the soft and
hard limits is supported.
Bolus injection setting range: Patient Information A patient information management
Bolus Rate: Management: system is inbuilt that allows for patient
2/3ml 0.10ml/h~90.00 ml/h information input and recording.
Step: 0.01 ml/h (Default: 90.00ml/h) Patient Type: Adult, Neonatal, Paediatric selectable
5ml 0.10ml/h~150.00 ml/h Other information: Name, Patient ID, Bed No., DOB,
Step: 0.01 ml/h (Default: 150.00ml/h) Height, Weight, Age, Sex, Blood type,
10ml 0.1ml/h~300.0ml/h BSA, BMA, Doctor in Charge
Step: 0.01ml/h (Default: 300.00ml/h)
20ml 0.1ml/h~600.0ml/h Alarms:
Step: 0.01ml/h (Default: 400.00ml/h) Alarm pressure level: P1 to P12
30ml 0.1ml/h~1200.0ml/h Alarm volume: Max: 65db (At 1m distance)
Step: 0.01ml/h (Default: 600.00ml/h) Min: 45db (At 1m distance)
50/60ml 0.1ml/h~2000.0ml/h Occlusion Alarm: Range: 150~975mmHg (P1-P12)
Step: 0.01ml/h (Default: 1500.00ml/h) Accuracy: ±15% or ±75mmHg,
whichever is greater
Bolus Preset Volume: Default setting: 525mmHg (P6)
2/3ml 0.10 ml ~2.00 ml Unit: mmHg, kPa, bar, psi
Step: 0.01 ml Supported Alarm Occlusion alarm, Completion alarm,
5ml 0.10 ml~5.00 ml Types: KVO Completion alarm, Purging
Step: 0.01 ml completion alarm, Purging near
10ml 0.1 ml~10.00 ml completion alarm, Syringe detection
Step: 0.01 ml alarm, System error alarm, Idle alarm,
20ml 0.1 ml~20.00 ml Low battery alarm, Depleted battery
Step: 0.01 ml alarm, Disconnected battery alarm,
30ml 0.1 ml~30.00 ml/h Abnormal pressure alarm, Near
Step: 0.01 ml completion alarm, Disconnected power
50/60ml 0.1 ml~50.00 ml supply alarm, Communication/network
Step: 0.01ml error alarm, Piston column abnormality
alarm, Syringe clamp abnormality
Injection Accuracy alarm, Syringe bracket abnormality
Rate=0.10~1.00ml/h ≤ ±5% alarm, Control board error alarm, Main
Rate>1.00ml/h ≤ ±1.8% motor speed abnormality alarm, Main
Mechanical ≤±1% (The percentage error motor directional abnormality alarm,
precision: between the specified motion Main motor slave drive directional
distance and the actual motion abnormality alarm, Main motor slave
distance in the full stroke of the driver speed abnormality alarm, Slave
syringe pump) driver 24V power supply abnormality
alarm, Slave driver communications Open) range: minutes, then an alarm will sound)
error alarm. Step:0.01ml/h
Features Infusion history: Up to 2000 records. Records any
Startup guidance and Syringe installation procedure system changes, alarms and actions
syringe installation: displayed upon device startup in the taken.
event where a syringe is not installed. Night mode Automatic adjustment of volume and
Syringe size is automatically detected screen brightness
and syringe brand selection will be Purging: Manual and automatic selectable
displayed upon syringe installation. Software Update: Via USB port
Syringe selection With auto syringe identification No operation time 1~5min (step: 1min)
parameters: function. settings:
Brands selectable upon syringe Continuous Titration: Allows editing of injection rate for
installation: Shuangge, Jierui, B-D ongoing infusions without the need to
(2/3ml, 5ml, 10ml, 20ml, 30ml, stop any currently ongoing infusions.
50/60ml) Others: Automatic screen lock, Standby mode,
Supports up to 3 user-defined custom Occlusion sensor, Patient data input,
syringe brands. Automatic pressure release, Syringe
Ingress Protection: IP33 calibration function, Startup syringe
KVO (Keep Vein 0.10~5.00ml/h; (KVO will last for 30 installation guidance.
*Notice: Specifications subject to change without prior
notice. All rights reserved by COMEN