You are on page 1of 1

HOSPITAL : ALLIED CARE EXPERTS MEDICAL CENTER - VALENZUELA

ADDRESS : MACARTHUR HIGHWAY, VALENZUELA, METRO MANILA


DATE PERFORMED : Thursday,5 January 2023
BIOMED PERFORMED : JAYSON ANTONIO, ECE / JO RAIRAM VALIDO
CONTACT PERSON : SIR JAYJAY
ASSESSMENT AND CHECKING OF LIFE SUPPORT EQUIPMENTS AND MECHANICAL HOSPITAL USE.

BRAND MODEL SERIAL NUMBER PROBLEM DEPARTMENT


AMSCO SURGICAL TABLE 0416881045 SIDE SELECTOR FOR REPAIR AND LUBRICATION OPERATING ROOM 5
HYDRAULIC FLUID LEAK
RATTLE SOUND
NO BRAND DUAL OR LIGHT - 2 HALOGEN BUSTED (SMALL DOME) OPERATING ROOM 5
2 HALOGEN BUSTED (BIG DOME)
WHEN INSTALLING NEW HALOGEN, IT DOENS'T LASTS LONGER
NO BRAND DUAL OR LIGHT - 2 HALOGEN BUSTED (SMALL DOME) OPERATING ROOM 4
2 HALOGEN BUSTED (BIG DOME)
WHEN INSTALLING NEW HALOGEN, IT DOENS'T LASTS LONGER
NO BRAND DUAL OR LIGHT - 1 HALOGEN BUSTED (SMALL DOME) OPERATING ROOM 1
2 HALOGEN BUSTED (BIG DOME)
WHEN INSTALLING NEW HALOGEN, IT DOENS'T LASTS LONGER
TIANJIN SCI-TECH DEV ANESTHESIA MACHINE 1404009 MISSING LOCK FOR PRESSURE RELIEF VALVE OPERATING ROOM 4

TERUFUSION SYRINGE PUMP TE-331 1311000361 LOOSE CLAMP, UNABLE TO SECURE SYRINGE BODY 4TH FLOOR
TERUMO INFUSION PUMP 1403010022 BUBBLE SENSOR FAULTY/DEFECTIVE 4TH FLOOR
TERUMO INFUSION PUMP 1306010110 AIR IN LINE ERROR 4TH FLOOR
TERUMO INFUSION PUMP 1312012225 AIR IN LINE ERROR 4TH FLOOR
BIOCARE INFUSION PUMP IP 12-B 70302100 DEFECTIVE DISPLAY 4TH FLOOR
BIOCARE INFUSION PUMP IP 12-B XA20180305B028 DEFECTIVE DISPLAY 4TH FLOOR
ZOLL DEFIB M2 SERIES BA18L000062 BATTERY CALIBRATION REQUIRED NURSE STATION
DEFLATED BATTERY, NO VOLTAGE READING, STERNUM LOCK
ZOLL DEFIB M2 SERIES BA181000081 EMERGENCY ROOM
BROKEN
ZOLL DEFIB M2 SERIES - DEFLATED BATTERY, NO VOLTAGE READING EMERGENCY ROOM
NO BRAND AUTO CLEAVE ? L - DISALIGN DOOR CHASSIS OPERATING ROOM
LEAK AT PRESSURE VALVES
NOT HEATING
IMPROVISED WATER LEVEL INDICATOR
NO BRAND AUTO CLEAVE 50 L 14L-1186 OVER TEMPERATURE ERROR MICROBIOLOGY
OVERHEATING
MORTARA ECG MACHINE 119180001180 BATTERY REPLACEMENT EMERGENCY ROOM
MORTARA ECG MACHINE 114340226897 BATTERY REPLACEMENT EMERGENCY ROOM
FOR REPAIR OF LCD KEYPADS
MORTARA ECG MACHINE 114340226891 BATTERY REPLACEMENT EMERGENCY ROOM
FOR REPAIR OF LCD KEYPADS