You are on page 1of 2

Reagent Parameters for BA-88/88A Chemistry Analyzer

Reagent Total
Sample Measure Incubation Linearity Range
Pri. Sec. Mix Reagent Delay time Reagent
Tests Reac. Type Direction Unit Decimal Volume time temperature Calibration type
Wave Wave Ratio Volume (seconds) Blank
(µl) (seconds) (℃) Low Up
(R1:R2) (µl)

ALT Kinetic Decrease U/L 0.1 340 405 40 4:1 1000 60 180 37 / 4 1000 One-point linear/K factor
AST Kinetic Decrease U/L 0.1 340 405 40 4:1 1000 60 180 37 / 4 800 One-point linear/K factor

ALP Kinetic Increase U/L 0.1 405 546 16 4:1 1000 60 180 37 / 5 800 One-point linear/K factor
γ-GT Kinetic Increase U/L 0.1 405 546 80 4:1 1000 60 180 37 / 4 650 One-point linear/K factor

LDH Kinetic Increase U/L 0.1 340 670 30 4:1 900 60 180 37 / 4 1000 One-point linear/K factor
α-HBDH Kinetic Decrease U/L 0.1 340 546 30 4:1 900 60 180 37 / 10 1000 One-point linear/K factor
CK Kinetic Increase U/L 0.1 340 546 20 4:1 1000 180 180 37 / 5 1000 One-point linear/K factor

CK-MB Kinetic Increase U/L 0.1 340 670 40 4:1 1000 180 180 37 / 5 600 One-point linear/K factor

α-AMY Kinetic Increase U/L 0.1 405 578 20 4:1 1000 60 120 37 / 5 1500 One-point linear/K factor
TBA Kinetic Increase μmol/L 0.1 405 510 10 3:1 900 60 180 37 Yes 2 180 One-point linear
UREA Kinetic Decrease mmol/L 0.01 340 670 10 4:1 833 30 60 37 Yes 0.9 40 One-point linear

CREA-J Fixed Time Increase μmol/L 0.1 510 670 50 1:1 1000 30 120 37 Yes 9 2420 One-point linear
Glu-O Endpoint Increase mmol/L 0.01 510 670 10 4:1 1000 6 5 37 Yes 0.3 28 One-point linear
Glu-HK Endpoint Increase mmol/L 0.01 340 670 8 2:1 1000 6 5 37 Yes 0.6 33 One-point linear

UA Endpoint Increase μmol/L 0.1 546 670 15 4:1 900 6 5 37 Yes 20.8 1500 One-point linear

CREA-S Endpoint Increase μmol/L 0.1 546 670 25 3:1 1000 6 5 37 / 10 7000 One-point linear
T-Bil (V) Endpoint Decrease μmol/L 0.01 450 546 28 4:1 980 6 5 37 Yes 2 684 One-point linear
D-Bil (V) Endpoint Decrease μmol/L 0.01 450 546 28 4:1 980 6 5 37 Yes 1 430 One-point linear

T-Bil (D) Endpoint Increase μmol/L 0.01 546 630 100 Single 1000 6 5 37 Yes 1.7 600 One-point linear
D-Bil (D) Endpoint Increase μmol/L 0.01 546 630 100 Single 1000 6 5 37 Yes 1 260 One-point linear

T-Bil (D) II Endpoint Increase μmol/L 0.01 546 630 10 4:1 833 6 5 37 Yes 1.7 600 One-point linear
D-Bil (D) II Endpoint Increase μmol/L 0.01 546 670 80 4:1 1000 6 5 37 Yes 1 260 One-point linear
HDL-C Endpoint Increase mmol/L 0.01 578 670 10 3:1 1000 6 5 37 Yes 0.05 6 One-point linear

LDL-C Endpoint Increase mmol/L 0.01 578 670 10 3:1 1000 6 5 37 Yes 0.05 20 One-point linear

ApoA1 Endpoint Increase g/L 0.01 340 / 7.5 3:1 1000 6 5 37 Yes 0.2 2.3 SPLINE
ApoB Endpoint Increase g/L 0.01 340 / 7.5 3:1 1000 6 5 37 Yes 0.2 2.2 SPLINE
Reagent Total
Sample Measure Incubation Linearity Range
Pri. Sec. Mix Reagent Delay time Reagent
Tests Reac. Type Direction Unit Decimal Volume time temperature Calibration type
Wave Wave Ratio Volume (seconds) Blank
(µl) (seconds) (℃) Low Up
(R1:R2) (µl)

Lp(a) Endpoint Increase mg/L 0.01 340 / 50 10 : 1 916 6 5 37 Yes 40 1000 SPLINE
IgG Endpoint Increase g/L 0.01 670 / 10 2:1 1000 6 5 37 Yes 0.3 35 SPLINE

IgA Endpoint Increase g/L 0.01 340 / 10 2:1 1000 6 5 37 Yes 0.2 5.6 SPLINE
IgM Endpoint Increase g/L 0.01 340 / 12 4:1 1000 6 5 37 Yes 0.05 4.8 SPLINE

C3 Endpoint Increase g/L 0.01 340 / 10 2:1 1000 6 5 37 Yes 0.04 3.3 SPLINE
C4 Endpoint Increase g/L 0.001 340 / 10 2.7 : 1 916 6 5 37 Yes 0.015 0.8 SPLINE
CRP Endpoint Increase mg/L 0.1 340 / 60 4:1 882 6 5 37 Yes 2 150 SPLINE

PA Endpoint Increase mg/L 0.1 340 / 25 2.7 : 1 982 6 5 37 Yes 10 800 SPLINE

TG Endpoint Increase mmol/L 0.01 510 670 10 Single 1000 6 5 37 Yes 0.1 12.5 One-point linear
TC Endpoint Increase mmol/L 0.01 510 670 10 Single 1000 6 5 37 Yes 0.1 20 One-point linear
TP Endpoint Increase g/L 0.1 546 630 20 Single 1000 6 5 37 Yes 2 120 One-point linear

ALB Endpoint Increase g/L 0.1 578 510 10 Single 1000 6 5 37 Yes 3 60 One-point linear
Ca Endpoint Increase mmol/L 0.01 630 / 10 Single 1000 6 5 37 Yes 0.1 3.75 One-point linear
Mg Endpoint Increase mmol/L 0.01 510 630 10 Single 1000 6 5 37 Yes 0.04 2.05 One-point linear

P Endpoint Increase mmol/L 0.01 340 450 10 Single 1000 6 5 37 Yes 0.3 4 One-point linear

Note:
1. Second wavelength is based mainly on the elimination of instrument noise, stray light and sample interference;
2. For dual reagent parameters, according to the number of tests per day to determine the amount of mixed reagent;
3. The sample and reagent volume can be adjusted according to the ratio, but not less than the minimum aspiration volume.