Cardio7 Operation Manual

1) Location Of Electrode

Attach electrodes to the patient’s body to record an electrocardiogram of twelve


standard leads [ I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 ], as shown below.

Limb electrodes are located as follows.

 RL (N): Right Leg


 LL (F): Left Leg
 RA (R): Right Arm
 LA (L): Left Arm
 V1 (C1): Fourth Intercostal Space At The Right Border Of The Sternum
 V2 (C2): Fourth Intercostal Space At The Left Border Of Sternum
 V3 (C3): Midway Between Location V2 and V4
 V4 (C4): At The Mid-Clavicular Line In The Fifth Intercostals Space
 V5 (C5): At The Another Axillary Line On The Same Horizontal Level As V4
 V6 (C6): At The Anterior Axillary Line On The Same Horizontal Level As V4 and V5