You are on page 1of 8

European Review for Medical and Pharmacological Sciences 2018; 22: 2191-2198

Titanium alloy vs. stainless steel miniscrews:


an in vivo split-mouth study
P. BOLLERO1, V. DI FAZIO2, C. PAVONI3,6, M. CORDARO4, P. COZZA5,7, R. LIONE3,6

1
Department of Systems Medicine, University of Rome “Tor Vergata”, Rome, Italy
2
Private practice, Rome, Italy
3
Department of Clinical Sciences and Translational Medicine, University of Rome “Tor Vergata”,
Rome, Italy
4
Department of Pediatric Dentistry, School of Dentistry, “Catholic University, of the Sacred Heart,
Gemelli Foundation, Rome, Italy
5
Department of Clinical Sciences and Translational Medicine, University of Rome “Tor Vergata”,
Rome, Italy
6
Department of Dentistry UNSBC, Tirana, Albany
7
Department of Dentistry UNSBC Tirana, Albany

Abstract. – OBJECTIVE: To compare in vi- Introduction


vo Titanium Alloy (TiA) with Stainless Steel
(SS) miniscrews Temporary Anchorage Devices The anchorage unit’s stability, defined as the
(TADs) using removal torque and Scanning Elec-
amount of allowed movement of the reactive unit,
tron Microscopic (SEM) analysis.
PATIENTS AND METHODS: 15 subjects (6 plays a very important role in the management
males and 9 females) who required maximum of orthodontic treatment. Controlled tooth move-
anchorage were recruited. For each patient, a TiA ments during space closure, without undesirable
TAD and a SS TAD with same length and width reciprocal movement in the anchorage unit are
were implanted following a randomized split- very difficult to achieve1. To reinforce anchor-
mouth study design. Retraction was carried out age, adjunctive extra-oral or intraoral appliances
with nickel-titanium spring ligated directly from
the anterior hooks of the archwire to the TADs to
such as transpalatal bar, nance holding arch, or
produce 90 to 100 g of force. When no further an- extra-oral traction, are usually necessary2. How-
chorage supplementation was needed, the TADs ever, these methods tend to cause discomfort,
were removed. The removal torque values were leading to a lower patient’s compliance 3.
registered with a digital screwdriver. After remov- The orthodontic miniscrews have been intro-
al, the TADs were collected in a fixed solution and duced as new devices for anchorage reinforce-
examined using SEM and X-ray microanalysis. ment4. Orthodontic miniscrews offer various ad-
RESULTS: All TADs remained intact, with a
100% success rate. There was no difference in
vantages to both the orthodontist and the patient:
removal torque between TiA and SS miniscrews simple insertion and removal, increased patient
(4.4 ± 1.3 N-cm and 5.1 ± 0.7 N-cm, respectively). comfort, and favorable cost-benefit ratio5,6.
All specimens’ loss of gloss with signs of bio- Miniscrews are also called “Temporary An-
logical contaminations resulted in a dull implant chorage Devices” (TADs)7,8 because they are de-
surface. SEM photomicrographs of TiA minis- signed to be removed after orthodontic treatment9.
crews showed predominantly blood cells while Miniscrews are generally made of Titanium Al-
SS miniscrews showed the precipitation of an
amorphous layer with low cellular component. loys (TiA) or surgical Stainless Steel (SS) that
There was no difference in spectroscopic anal- are both highly biocompatible materials10. TiA
ysis between TiA and SS miniscrews. allows for direct bone contact (osteo-integration)
CONCLUSIONS: TiA and SS miniscrews had between endosseous dental implants and the host
comparable removal torque values. SEM photo- bone11. Contrary, SS miniscrews tend to develop
micrographs showed no evidence of osteointe- a fibrous tissue interface between the screw and
gration with both TADs having similar biologi-
cal responses.
bone12,13. Despite of the differences between TiA
and SS, both materials provide relatively predict-
Key Words able clinical outcomes. They offer similar success
Titanium alloy miniscrew, Stainless steel miniscrew, in fulfilling the biomechanical requirement of
SEM analysis, Removal torque.
stability14-16. A recent meta-analysis by Papado-

Corresponding Author: Massimo Cordaro, MD; e-mail: massimo.cordaro@unicatt.it 2191


P. Bollero, V. Di Fazio, C. Pavoni, M. Cordaro, P. Cozza, R. Lione

poulos et al17 reported success rates of 87.7% for N-cm differences in removal torque between the
both TiA and SS. The main difference between two miniscrewers, with a standard deviation of
the two materials is that SS provides greater me- 0.4 N-cm22 (SigmaStat 3.5, Systat Software, Point
chanical characteristics18,19, when the insertion or Richmond, CA, USA). Participants were recruit-
removal torque exceeds the torsional strength20,21. ed at the Department of Orthodontics at the Uni-
Thus SS, because it is stronger than the traditional versity of Rome “Tor Vergata” (Rome, Italy). All
TiA, would reduce the risk of breakage22,23. The patients were in good general health with healthy
information gathered may help orthodontists to periodontium, generalized probing depths not ex-
understand the intricacies of success or failure of ceeding 3 mm, and no radiographic evidence of
miniscrews related to their material composition. periodontal bone loss. An experienced clinician
Gritsch et al24 in 2013 evaluated the use of imme- who assessed them as needing maximum an-
diate loaded TiA and SS screws in pigs growing chorage treated patients. This was defined as “no
model. They studied the specific bone response in mesial movement of the molars during the period
the vicinity of the devices (test-zone) compared to of anchorage supplementation”. No attempt was
a control zone. No significant differences between made to achieve distal molar movement since
the materials were found regarding the percentage this was not clinically required. Patients who re-
of “bone-to-implant contact” or the static and dy- quired functional appliance therapy, those who
namic bone parameters. However, 5% threshold had previous orthodontic treatment or extraction,
of “bone-to-implant contact” was obtained after hypodontia, craniofacial syndromes or cleft, an-
4 weeks with the SS devices, leading to increased tibiotic therapy within the past six months, and
survival rate values. anti-inflammatory drugs within the past month,
Brown et al22 in 2014 compared TiA minis- were excluded.
crews with identically sized SS implants, which All patients were fitted with Straight Wire
were inserted into the tibias of 12 rabbits. All Mirabella prescription (Sweden&Martina, Due
miniscrews demonstrated mechanical stability Carrare, Padoa, Italy) maxillary and mandibular
after 6 weeks with no mobility. No significant pre-adjusted edgewise appliances and the neces-
differences in micro damage or bone-to-implant sary extractions were performed. Before placing
contact were found between the tension and com- the TADs, the maxillary arch was aligned and lev-
pression sides of the implants. TiA and SS minis- eled by using 3 archwires (0.016 NiTi, 0.019x0.025
crews showed similar total micro damage burden NiTi, 0.019x0.025 posted Stainless Steel).
values when subjected to 100 g of loading. This For each patient, a TiA miniscrew (Spider
work showed a 100% success rate in rabbit tibias, Screw, Sarcedo, Vicenza, Italy) and a SS minis-
suggesting that SS can be used as TADs with at crew (Leone, Florence, Italy) were placed follow-
least the same efficacy of TiA miniscrews. ing a randomized split-mouth study design. All
To our knowledge, no study analyzed in vivo TADs had the same length (8 mm) and width (1.5
the biological response of the TiA and SS minis- mm). TiA TADs were self-drilling and inserted
crews with immediate orthodontic loading. In directly with a hand screwdriver without mak-
the present investigation, we compared TiA and ing a pilot hole with a hand piece. SS miniscrews
SS miniscrews by analyzing: 1) the insertion and required a pilot drill (1.2 mm in diameter). Both
removal torque; 2) the morphological, structural, types of miniscrews were implanted buccally in
and compositional alterations in used orthodontic the posterior alveolar crest.
miniscrews derived from Scanning Electron Mi- To avoid the risk of root proximity during the
croscopic (SEM). TADs placement, an accurate radiological ex-
amination was performed and surgical templates
were used. Surgical template was made with or-
Patients and Methods thodontic wire and acrylic resin in order to pre-
cisely locate the insertion point to avoid damage
Patients to the adjacent structures. The acrylic fits over
15 subjects (6 males and 9 females with a the occlusal surfaces of the teeth near the surgi-
mean age 16.2 ± 4.6), who required maxillary first cal site, and the wire was inserted in the acrylic
premolar extractions and canine distalization, and bent so that it was corresponded to the point
were recruited in the study. Before starting the of screw placement. The distance from this point
research, we estimated that a sample of 15 sub- to the adjacent anatomic structures was deter-
jects would give 80% power to detect at least 0.7 mined radiographically using the long-cone par-

2192
Titanium alloy vs. stainless steel miniscrews

allel technique. The TADs were placed under formed at many times the original magnification
local anesthesia, at the junction of the attached window, 200-second acquisition time, and 30-
gingiva with the reflected mucosa and mesial to 33% dead time. The quantitative analysis of the
the maxillary molars. The insertion torque for percentage of weight concentration of the probed
both types of TADs was assessed, despite TiA elements was performed by nonstandard analysis
miniscrews being self-drilling (while SS minis- and ZAF routines by using TEAM software (ver-
crews were not), and the difference in the shape sion 4.2.2, EDAX, Florence, Italy).
of the final portion of the tip. The mean value of This project was approved by the Ethical Com-
Implant Placement Torque (IPT) when tighten- mittee of the University of Rome “Tor Vergata”
ing the mini-implant into the bone was measured (Rome, Italy) (Protocol number: 134/15). Written
using a torque dynamometer (HTG-2N, IMA- informed consents were obtained from all patients
DA, Toyohashi, Japan). The miniscrews were or the parents of those under 18 years of age.
placed before any retraction force was applied
on the archwire with anterior hooks placed distal Statistical Analysis
to the lateral incisors. Retraction was carried out Statistical analysis was performed using SPSS
with nickel titanium spring ligated directly from (Statistical Package for the Social Sciences; SPSS,
the anterior hooks of the archwire to the TADs Version 12, Chicago, IL, USA). Mean values and
to produce 90 to 100 g of force. Anchorage sup- standard deviations were calculated for inser-
plementation was discontinued once the canines tion torque, removal torque, and periotest value
were in Class I and Overjet reduction complet- (PTV) testing. Any difference in the periotest
ed. At this point the operator judged that no fur- value (PTV) was detected by means of t-test for
ther anchorage supplementation was needed and independent samples. The maximum insertion
the TADs were removed. The mobility of each torque, removal torque, and EDS values were not
screw was measured by using Periotest (Siemens normally distributed (Kolmogorov-Smirnov test)
AG, Bensheim, Germany) before removing the with unequal of variance (Levene’s test). The Wil-
TADs. The following guidelines were used to coxon Signed Rank test was used for comparing
obtain a reproducible registration: (1) the cali- TiA and SS miniscrews. Results were considered
bration of the device was checked; (2) the hand- significant if p < 0.05.
piece of the periotest was held in the horizontal
position; (3) the handpiece was applied perpen-
dicularly to the screw, which was perpendicular Results
to the floor; (4) the tip of the handpiece was less
than 4 mm from the screw. No complication occurred operatively or
The maximal torque values required to loos- post-operatively and no infection was detected
en the miniscrews were registered with a digital in any of the patients throughout the study. All
torque dynamometer (HTG-2N, IMADA, Toyo- TADs remained intact with a 100% success rate.
hashi, Japan). After removal, miniscrews were The mean period of TADs function as an anchor-
immediately fixed in Karnovsky solution (para- age unit was 160.8 ± 23 days. The periotest value
formaldehyde and glutaraldehyde) and were col- recorded before removal of 8.8 for TiA TADs and
lected to evaluate the bone-miniscrew interface 9.1 for SS TADs, indicated mechanical stability
using SEM (FEI, Quanta 200, Florence, Italy) for both types of miniscrews with no statistical
analysis. differences (p > 0.05). There was no statistical
Each retrieved miniscrew was analyzed at difference in insertion and removal torque be-
three magnification of 85X, 200X, 600X focus- tween TiA and SS miniscrews. The mean inser-
ing on orthodontic head, transmucosal collar and tion torque for TiA was 3.2 ± 1.1 N-cm while for
bony portion. Greater magnifications of 3000- SS miniscrews was 2.9 ± 0.9 N-cm. The mean
10000X were necessary to detect bony islands. removal torque for TiA was 4.4 ± 1.3 N-cm while
Furthermore, the retrieved TADs were sub- for SS miniscrews was 5.1 ± 0.7 N-cm. Figure 1
jected to energy-dispersive X-ray microanalysis shows optical microscopic image of TiA and SS
to investigate their elemental composition with a miniscrews before and after their removal. Frac-
silicon- lithium energy dispersive microanalysis tured screw tips and stripped screw threads after
(EDS) detector (Sapphire CDU, EDAX, Mahwah, clinical use were not observed. All specimens’
NJ, USA) equipped with a super ultrathin beryl- loss of gloss with signs of biological contamina-
lium window. Elemental microanalysis was per- tions resulted in a dull implant surface.

2193
P. Bollero, V. Di Fazio, C. Pavoni, M. Cordaro, P. Cozza, R. Lione

At higher magnification, SEM photomicro- idence from dental implantology that bone qual-
graphs of TiA miniscrews showed predominantly ity and oral hygiene influence TADs prognosis.
blood cells covering most of the surface. SS minis- Randomized split-mouth study design minimized
crews showed the precipitation of an amorphous the effect of all the inter-individual variables such
layer with low cellular component (Figures 2,3,4). as age, sex, periodontal health, and duration of
Randomly organized osteo-integration islets on load, reducing the need of a larger sample. The
both types of miniscrew surfaces were found. trial showed a 100% success rate for both types
Finally, there was no statistically difference of miniscrews. No TADs fractured during place-
in spectroscopic analysis between TiA and SS ment, clinical use, or removal, suggesting that
miniscrews. Elements such as carbon, oxygen, TiA miniscrews can be used as temporary ortho-
and sodium, were found. EDS indicated the pres- dontic anchorage with at least the same efficacy
ence of Titanium (Ti), Aluminum (Al), and Vana- than SS miniscrews. Indeed, the high degree of
dium (V) as the main elements on the surfaces of osteo-integration required for dental implants is
all TiA miniscrews and it was the only statistical- not a requirement for orthodontic TADs to func-
ly significant difference (Table I). tion as anchorage devices 22,25.
Not the material, but insertion technique,
thread design, and dimension of TADs are the
Discussion essential factors affecting implant primary stabil-
ity26. A valid method to assess the primary stabil-
We aimed to compare in vivo, immediately ity of implants quantitatively is the measurement
loaded orthodontic TiA vs. SS TADs analyzing of insertion torques26-28. Motoyoshi et al28 exam-
the removal torque and the morphological, struc- ined the insertion placement torques of mini-im-
tural, and compositional alterations by means plants and recommended torque values of 5-10
of SEM. The split-mouth study design involved N-cm. Higher values may result in higher failure
for each subject the use of TADs of two differ- rates because of distinctive bone compression
ent materials presenting with the same length (8 with micro damages or may even cause mini-im-
mm) and diameter (1.5 mm). There is clinical ev- plant fracture29,30.

Figure 1. Optical microscopic


image of TiA and SS miniscrews
before (A and C) and after (B and
D) their removal. The retrieved
ones have lost of gloss with signs
A B C D
of biological contaminations.

2194
Titanium alloy vs. stainless steel miniscrews

A B C D

Figure 2. SEM photomicrographs of bony portion of TiA and SS miniscrews after clinical use. A, C, TiA miniscrew 85X. B,
D, SS miniscrew (200x).

A B

Figure 3. SEM photomicrographs of bony portion of TiA miniscrews after clinical use with red blood cells, (1200x).

Figure 4. SEM photomicrographs of bony portion of TiA and SS miniscrews after clinical use. A, TiA miniscrew 600X. B, SS
miniscrew (600x).

2195
P. Bollero, V. Di Fazio, C. Pavoni, M. Cordaro, P. Cozza, R. Lione

Table I. Chemical composition of the biological material on TiA and SS miniscrew surfaces as highlited by EDS analysis
(Wilcoxon Signed Rank Test, p<0.05).
Elements Weight % Error % Weight % Error % Difference p-value
SS SS TiA TiA

Carbon (C) 56.1 6.4 51.4 7,2 4.7 NS


Nitrogen (N) 2.0 33.8 0 0 2 NS
Oxygen (O) 32.8 9.9 34.6 9.7 -1.8 NS
Sodium (Na) 5.8 8.6 0 0 5.8 NS
Silicon (Si) 0.1 8.6 0.2 20.9 -0.1 NS
Phosphorus (P) 1.9 3.1 5.4 4 -3.5 NS
Sulphur (S) 0.4 4.7 0 0 0.4 NS
Calcium (Ca) 0 27.2 0 0 0 NS
Chromium (Cr) 0.2 8.1 0 0 0.2 NS
Manganese (Mn) 0 54.4 0 0 0 NS
Iron (Fe) 0.6 4.3 0.5 16.8 0.1 NS
Nickel (Ni) 0.1 12.1 0.2 30 -0.1 NS
Aluminium (Al) 0 0 0.3 9.8 -0.3 **
Strontium (Sr) 0 0 0.2 6 -0.2 NS
Titanium (Ti) 0 0 12.4 1.2 -12.4 **
Vanadium (V) 0 0 0.5 4.8 -0.5 **

NS: Not significant; **p<0.01.

To avoid the risk of root proximity during the the tibias of 12 rabbits reporting no differences in
TADs placement, an accurate radiological exam- removal torque between TiA and SS TADs (4.4 ±
ination was performed and surgical templates were 1.7 N-cm and 4.4 ± 2.3 N-cm, respectively). We
used. Kuronda et al31 suggested that the proximity did not evaluate the insertion torque since the two
of miniscrews to the adjacent tooth roots is one miniscrews presented different tip designs that
of the major risk factors for failure of screw an- might influence a greater torque resistance during
chorage. Considering that the width of periodon- the insertion of SS TADs. After clinical use, frac-
tal ligament is approximately 0.25 mm, Poggio et tured screw tips and stripped screw threads were
al32 reported that a minimum clearance of 1 mm not observed by means of optical microscope.
of alveolar bone around the screw is sufficient All retrieved TADs showed signs of biological
for periodontal health. No gingival inflammation contaminations on their surfaces (Figure 1). The
was observed in the treated subjects. According SEM analysis showed the presence of biological
to Miyawaki et al33, the risk of tissue inflamma- materials on the surfaces of retrieved TADs (Fig-
tion is lessened when the miniscrews are inserted ure 2). Particularly, at higher magnification, SEM
into keratinized attached gingiva avoiding areas photomicrographs of TiA TADs showed predom-
of frenulum, muscle tissues and mobile mucosa inantly blood cells covering most of the surface
(non-attached gingiva). In addition to histological (Figure 3). SS TADs showed the precipitation of
evaluation, a non-invasive method for determin- an amorphous layer with low cellular component
ing implant stability and the extent of osteo-in- (Figure 4). These results are consistent with those
tegration is the measurement of removal torque. reported by Albrektsson et al12 and Gotman et al13
The removal torque reflects the characteristics of who observed that SS tends to develop a fibrous
the implant-bone interface during and after their tissue interface between the screw and the bone.
usage34. Partial osteo-integration can increase Finally, the EDS microanalysis detected statisti-
the torque values and complicate the removal of cally significant differences only in the percentages
TADs22. In the present study, TiA and SS TADs of titanium, aluminum, and vanadium. The presence
resulted both not osteo-integrated. The interac- of iron was similar in both miniscrews. However,
tion between bone and mini-implant was mainly the results should be carefully interpreted. The iron
due to mechanical interlocking. These results are in the TiA TADs is a consequence of the presence
in agreement with those reported by previous re- of blood cells, while the iron in SS TADs represents
searches22,33,35. Brown et al22 compared TiA minis- the material of the specimens itself 34. A limit of our
crews with identically sized SS mini-implants investigation was the different shape of the final por-
(length: 6 mm; diameter: 1.6 mm), inserted into tion between Tia and SS TADs. The selected TADs

2196
Titanium alloy vs. stainless steel miniscrews

resulted the most similar available miniscrews for  11) Branemark PI, Hansson BO, A dell R, Breine U,
diameter, length and thread interval. However, all L indstrom J, Hallen O, Ohman A. Osseointegrated
implants in the treatment of the edentulous jaw.
the other limits, such as the difference in soft tissue Experience from a 10-year period. Scand J Plast
thickness, oral hygiene, periodontal health, age, sex, Reconstr Surg Suppl 1977; 16: 1-132.
and duration of load, have been overcome by the  12) A lbrektsson T, Hansson HA. An ultrastructural char-
split mouth study design. acterization of the interface between bone and
sputtered titanium or stainless steel surfaces.
Biomaterials 1986; 7: 201-205.
 13) Gotman I. Characteristics of metals used in im-
Conclusions plants. J Endourol 1997; 11: 383-389.
 14) Papadopoulos MA. Orthodontic treatment of class
TiA and SS miniscrews showed similar re- II malocclusion with miniscrew implants. Am J
moval torque values. SEM photomicrographs of Orthod Dentofacial Orthop 2008; 134: 604.e1-
604.e16.
TiA TADs demonstrated blood cells covering  15) Reynders R, Ronchi L, Bipat S. Mini-implants in or-
most of the surface. SEM photomicrographs of SS thodontics: a systematic review of the literature.
TADs showed the precipitation of an amorphous Am J Orthod Dentofacial Orthop 2009; 135: 564.
layer with low cellular component. No evidence e1-565.e19.
of osteo-integration was detected.  16) Sung SJ, Jang GW, Chun YS, Moon YS. Effective en-
masse retraction design with orthodontic mini-im-
plant anchorage: a finite element analysis. Am J
Orthod Dentofacial Orthop 2010; 137: 648-657.
Conflict of Interest:  17) Papadopoulus MA, Papageorgiou SN, Zogakis IP.
The Authors declare that they have no conflict of interests. Clinical effectiveness of orthodontic miniscrew
implants: a meta-analysis. J Dent Res 2011; 90:
969-976.
 18) Buschang PH, C arrillo R, Ozenbaugh B, Rossouw PE.
References Survey of AAO members on miniscrew usage. J
Clin Orthod 2008; 42: 513-518.
  1) M aino BG, M aino G, Mura P. Spider screw: skel-  19) Francioli D, Ruggiero G, Giorgetti R. Mechanical
etal anchorage system. Prog Orthod 2005; 6: properties evaluation of an orthodontic miniscrew
70-81. system for skeletal anchorage. Prog Orthod 2010;
  2) R ajcich MM, Sadowsky C. Efficacy of intra-arch 11: 98-104.
mechanics using differential moments for achiev-  20) A nsell RH, Scales JT. A study of some factors
ing anchorage control in extraction cases. Am J which affect the strength of screws and their
Orthod Dentofacial Orthop 1997; 112: 441-448. insertion and holding power in bone. J Biomech
  3) Fortini A, Cacciafesta V, Sfondrini MF, Cambi S, Lupoli M. 1968; 1: 279-302.
Clinical applications and efficiency of miniscrews for  21) Wilmes B, Panayotidis A, Drescher D. Fracture resis-
extradental anchorage. Orthodontics 2004; 1: 1-12. tance of orthodontic mini-implants: a biomechani-
  4) K anomi R. Mini-implant for orthodontic anchorage. cal in vitro study. Eur J Orthod 2011; 33: 396-401.
J Clin Orthod 1997; 31: 763-767.  22) Brown R, Sexton BE, Chu TM, K atona TM, Stewart
  5) Fritz U, Ehmer A, Diedrich P. Clinical suitability KT, Kyung H, Kyung H, L iue SS. Comparison of
of titanium miniscrews for orthodontic anchor- stainless steel and titanium alloy orthodontic
age-preliminary experiences. J Orofac Orthop miniscrew implants: a mechanical and histologic
2004; 65: 410-418. analysis. Am J Orthod Dentofacial Orthop 2014;
  6) K im TW, Baek SH, K im JW, Chang Y. Effects of 145: 496-504.
microgrooves on the success rate and soft tis-  23) C arano A, Lonardo P, Velo S, Incorvati C.
sue adaptation of orthodontic miniscrews. Angle Mechanical properties of three different com-
Orthod 2008; 78: 1057-1064. mercially available miniscrews for skeletal an-
  7) K im JW, Baek SH, K im TW, Chang YI. Comparison chorage. Prog Orthod 2005; 6: 82-97.
of stability between cylindrical and conical type
mini-implants. Mechanical and histological prop-
erties. Angle Orthod 2008; 78: 692-698.
  8) M ah J, Bergstrand F. Temporary anchorage de-
vices: a status report. J Clinic Orthod 2005; 39:
132-136.
  9) Ohnishi H, Yagi T, Yasuda Y, Takada K. A mini-im-
plant for orthodontic anchorage in a deep over-
bite case. Angle Orthod 2005; 75: 444-452.
 10) Cornelis MA, Scheffler NR, De Clerck HJ, Tulloch
JF, Behets CN. Systematic review of the experi-
mental use of temporary skeletal anchorage de-
vices in orthodontics. Am J Orthod Dentofacial
Orthop 2007; 131: 52-58.

2197
P. Bollero, V. Di Fazio, C. Pavoni, M. Cordaro, P. Cozza, R. Lione

 27) Friberg B, Sennerby L, Roos J, Johansson P, Strid  32) Poggio M, Incorvati C, Velo S, C arano A. “Safe
CG, L ekholm U. Evaluation of bone density using zones’’: a guide for miniscrew positioning in the
cutting resistance measurements and microra- maxillary and mandibular arch. Angle Orthod
dio-graphy: an in vitro study in pig ribs. Clin Oral 2006; 76: 191-197.
Implants Res 1995; 6: 164-171.  33) Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara
 28) Motoyoshi M, Hirabayashi M, Uemura M, Shimizu T, Takano -Yamamoto T. Factors associated with the
N. Recommended placement torque when tight- stability of titanium screws placed in the posterior
ening an orthodontic mini-implant. Clin Oral region for orthodontic anchorage. Am J Orthod
Implants Res 2006; 17: 109-114. Dentofacial Orthop 2003; 124: 373-378.
 29) Wilmes B, R ademacher C, Olthoff G, Drescher D.  34) Motoyoshi M, Uemura M, Ono A, Okazaki K, Shigeeda
Parameters affecting primary stability of orthodontic T, Shimizu N. Factors affecting the long-term sta-
mini-implants. J Orofac Orthop 2006; 67: 162-174. bility of orthodontic mini-implants. Am J Orthod
 30) Wawrzinek C, Sommer T, Fischer-Brandies H. Dentofacial Orthop 2010; 137: 588.e1-588.e5.
Microdamage in cortical bone due to the over-  35) Chen YJ, Chen YH, L in LD, Yao CCJ. Removal
tightening of orthodontic microscrews. J Orofac torque of miniscrews used for orthodontic an-
Orthop 2008; 69: 121-134. chorage-a preliminary report. Int J Oral Maxillofac
 31) Kuroda S, Sugawara Y, Deguchi T, Kyung HM, Takano - Implants 2006; 21: 283-289.
Yamamoto T. Clinical use of miniscrew implants as  34) Eliades T, Zinelis S, Papadopoulos MA, Eliades G.
orthodontic anchorage: success rates and post- Characterization of retrieved orthodontic minis-
operative discomfort. Am J Orthod Dentofacial crew implants. Am J Orthod Dentofacial Orthop
Orthop 2007; 131: 9-15. 2009; 135: 10.e1-10.e7.

2198

You might also like