You are on page 1of 13

CRITICAL CARE NEPHROLOGY

Central Venous Dialysis Catheter Dysfunction


Martine Leblanc, Jean-Yves Bosc, Emil P. Paganini,
and Bernard Canaud
Central venous catheter dysfunction is a limiting factor in regard to renal replacement therapy
efficiency and can thus influence patient morbidity. Early catheter dysfunction is frequently due to
mechanical problems such as inadequate positioning, kinking, or constriction, but early fibrin
deposition can develop soon after insertion. Delayed dysfunction usually results from thrombus
formation, either within the lumen, around the catheter ("fibrin sleeve"l, or in the host vein. Catheter
dysfunction is suspected clinically or documented by simple imaging studies. It is usually evident and
manifested by failure to aspirate blood from the lumen(sl, inadequate blood flow and/or high
resistance pressures during hemodialysis. However, a more subtle dysfunction may lead to a high
recirculation of dialyzed blood and be overlooked if dialysis adequacy is not monitored regularly. Local
instillation of a fibrinolytic agent is usually successful in restoring catheter patency. Central venous
dialysis catheters present intrinsic limitations consequent to their composition and design, whereas
extrinsic limitations result from site of insertion, blood properties and anatomic particularities of a
given individual. These characteristics largely determine overall catheter performances. Performance
parameters to consider include maximal consistently achievable blood flow rate, resistance to blood
flow indicated by arterial and venous pressures during hemodialysis, and blood recirculation rate.
Catheter longevity is an important consideration for cuffed catheters implanted for long-term use. The
tolerated blood recirculation within central venous dialysis catheters should be below 10% to 15%, and
is ideally between 3% to 7% in most clinical settings. Several recent studies confirm that short femoral
catheters recirculate significantly more than is desirable. Well functioning and non reversed internal
jugular and subclavian venous catheters have, in general, recirculation rates less than 5%. With regard
to various performance criteria, the TwinCath (Medcomp, Harleysville, PAl appears particularly
advantageous. In any case, a good catheter maintenance program is of critical importance for the
prevention and the early detection of catheter dysfunction.
© 1997 by the National Kidney Foundation, Inc.
Index Words: Blood flow; catheter; dialysis; recirculation; thrombosis.

C atheter malfunction is one of the most


frequent and frustrating events in hemo-
dialysis that can increase the morbidity of
Less frequently, early fibrin formation will
partially obstruct the lumen(s), reducing blood
flow availability, and causing a suction effect
renal failure patients. In addition to the associ- on the venous wall on aspiration.
ated inconvenience and complications result- Delayed or late dysfunction of central ve-
ing from the extra manipulations frequently nous dialysis catheters (arbitrarily defined here
required, the poor function of the dialysis as 15 days post-cannulation) has more com-
catheter can significantly compromise the effi- plex etiologies that generally relate to thrombo-
cacy of renal replacement therapy. The follow- genesis, either as an obstruction inside the
ing discussion provides some information catheter itself (thrombus that covers the inte-
about the causes of catheter malfunction, clues rior of the catheter lumen, thus reducing its
to catheter dysfunction, available diagnostic diameter and obstructing its distal and/or
tools, and potential therapeutic and preven- side holes), or as an external fibrin sleeve
tive modalities. surrounding the distal catheter and occluding
the side holes, 1 or finally, as a thrombosis of
the central host vein (described in the superior
Causes of Catheter Dysfunction vena cava, subclavian, internal jugular, brachio-
cephalic or femoral veins).2-6 Secondary dis-
The causes of dialysis catheter dysfunction
may vary depending on the time elapsed since
placement. Immediate or early malfunction From the Department of Nephrology, Lapeyronie Hospital,
usually results from mechanical problems in- Montpel/ier, France; and the Section of Dialysis and Extracorpo-
cluding malpositioning of the distal tip in the real Therapy, The Cleveland Clinic Foundation, Cleveland, OH.
Address correspondence to Bernard Canaud, MD, Division
vessel, catheter kinking in the subcutaneous of Nephrology, Lapeyronie Hospital, Montpel/ier 34295 France.
tunnel, or catheter constriction from a tight © 1997 by the National Kidney Foundation, Inc.
ligature; these causes can be easily corrected. 1073-4449/9710404-0010$3.00/0

Advances in Renal Replacement Therapy, Vol 4, No 4 (October), 1997: pp 377-389 377


378 Leblanc et al

lodging and kinking of the catheter can also 20% to 70% of cases using various diagnostic
occur with long-term use. modalities. 4,8 In addition to the uncommon
A previous study concerning acute renal finding of ipsilateral limb edema and dis-
failure patients showed that temporary (not tended veins, difficulty in cannulating the
tunneled) dual-lumen subclavian dialysis cath- vein, in achieving the desired blood flow
eters present early mechanical dysfunction during dialysis, and/ or in aspirating blood
more often than catheters inserted at other from the catheter all suggest that a thrombus
sites, and this led to access failure as early as has formed. These are indeed expected find-
the third hemodialysis session? A potential ings with poorly functioning catheters.
explanation comes from the anatomic features Whenever foreign material invades the
of the infraclavicular region that may cause bloodstream, circulating platelets tend to ad-
catheter kinking in the subcutaneous tunnel here and aggregate on its surface, subse-
near the clavicular bone. quently leading to the activation of the coagu-
lation cascade with the development of a fibrin
Catheter Thrombogenicity layer and a clot. There are a number of factors
involved in catheter thrombogenicity.
Catheter-related thrombosis remains a leading
cause of catheter dysfunction. Three broad
categories are recognized: (1) lumen throm- Catheter Properties
bus, (2) fibrin sleeve gradually developing Catheters are manufactured from standard
over the exterior surface of the catheter from polymers, including polyvinyl chloride, poly-
the site of insertion down to the tip; and (3) tetrafluoroethylene (Teflon), polyethylene,
mural thrombus along the venous wall prob- polyurethane, and silicone elastomer (Silastic).
ably after an endothelial injury. Figure 1 shows Most of these materials promote thrombus
a fibrin sleeve found tightly attached to a formation, and the procoagulant effect of cath-
removed dialysis catheter; the pathological eter material is most likely mediated by plate-
analysis showed the exclusive presence of lets and/ or activation of the contact phase. 9
fibrin. Because polyurethane and particularly sili-
These subtypes of thrombotic complica- cone elastomer have been associated with less
tions, alone or in combination, occur more thromboembolic complications, both are con-
frequently than suspected on clinical grounds. sidered to be the most suitable materials for
Early, partial, or complete thrombosis of femo- catheterization. 1o,1l
ral, internal jugular, or subclavian veins with Surface roughness and irregularities pro-
indwelling catheters have been detected in mote thrombogenicity because they serve as a
nidus for thrombus deposition; indeed, it has
been shown that fibrinogen is preferentially
adsorbed onto microscopic surface crevices. 12
Subtle changes in material manufacturing and
processing can also affect the blood response
and favor thrombogenesis.
Several surface treatments, such as heparin
immobilization onto the catheter surface, ion
implantation of protective substances offering
thrombus or infection-resistance, and protein
conditioning (with the purpose of inhibiting
fibrinogen adsorption or of specifically promot-
ing albumin adsorption which is inert in re-
gard to coagulation) have been attempted to
improve biocompatibility12-14; preliminary re-
Figure 1. A fibrin sleeve that was tightly attached sults appear promising15 and further clinical
to the external part of a removed Silastic dialysis evaluations are underway. However, heparin-
catheter. coated devices are costly and appear mostly
Critical Care Nephrology: Catheter Dysfunction 379

effective in the short-term because the coating contact between the catheter itself and the
tends to leach off or dissolve over time.!6 vessel wall; this factor is then probably not as
Catheter stiffness may be more important important for adults with dialysis catheters in
than its chemical and surface composition in large veins. Nonetheless, central venous dialy-
regard to thrombogenicity. Stiff, inflexible cath- sis catheters are located in low pressure ves-
eters may induce endothelial injury and vessel sels, with diameters that are very sensitive to
wall abrasion, whereas their softer counter- changes in intravascular volume. Local hemo-
parts are more likely to prevent damage to the dynamic factors are important to consider
endothelium. Compared to long and stiff poly- because they can be responsible for flow distur-
ethylene catheters, either heparin-impreg- bances associated with thrombus formation.
nated or not, soft pliable silicone elastomer To reduce these flow disturbances, the design
catheters have been shown to cause less throm- of the catheter should include a smooth sur-
bosis.!1 Nevertheless, some polyurethane cath- face and contours avoiding bends, bifurca-
eters that appear rigid at room temperature tions and sudden changes in diameter, thus
soften when located inside the body. When partially eliminating the resistance to fluid
considering all of these factors, polyurethane flow that promotes thrombogenesis. Finally,
appears to be a reasonable choice for short- the distribution of catheter holes at the distal
term catheterization, whereas Silastic is pre- end may confer greater or lesser stability
ferred for longer or "permanent" usey-2o In within the bulk flow of the vein, consequently
most centers, extremely compliant silicone influencing local hemodynamics.
catheters ordinarily require a surgical place-
ment and, consequently, are not readily avail-
able in the acute setting. Nevertheless, short- Expected Catheter Performance
term temporary silicone catheters that do not Versus Dysfunction
necessitate subcutaneous tunneling and that Limitations of dialysis catheters inserted into
are easy to insert at the bedside over a central veins depend in part upon their compo-
guidewire are now available (Hemo-Cath, sition, geometry, and location. The catheter
Medcomp, Harleysville, PA, and Uldall, Cook material itself affects compliance-elastance pa-
Critical Care, Bloomington, IN). On the other rameters, whereas the catheter length (any-
hand, percutaneous insertion of tunneled cath- where from 10 to 40 cm), its internal diameter
eters is possible at the bedside using the (0.5 to 2.0 mm), as well as its distal holes
Seldinger technique, with or without the aid of (number and distribution) influence the resis-
a pull-apart introducer (TwinCath, Medcomp, tance to blood flow. Poiseuille's law states that
Harleysville, PA, and PermCath, Quinton In- the pressure decrease or internal resistance
strument Co, Seattle, WA).18,21 within the catheter is primarily related to its
internal diameter and secondarily to its length,
Patient-Related Factors determining blood flow. Catheter length and
Besides catheter material and stiffness, throm- lumen diameter are the two key characteristics
bus formation involves interactions among to consider for optimizing resistance to flow.
several other factors: size of the host vessel
and state of its wall; local blood flow and Effective Blood Flow Rate
hemodynamics; and catheter residence time For optimal function, a central venous dialysis
within the vessel. In addition, certain blood catheter should be able to deliver an extracor-
characteristics can further contribute to the poreal blood flow consistently between 250
clotting tendency (eg, enhanced viscosity from and 400 mL / min during intermittent sessions,
a high hematocrit value, reactive thrombocyto- and ideally greater than 300 mL/min. This
sis, and fibrinogen release in inflammatory requisite is especially important for end-stage
states). renal disease patients with catheters in place
There exists an inverse relationship be- for long-term use. Unfortunately, most thin
tween the incidence of catheter-related throm- dual-lumen venous catheters cannot provide
bus formation and the diameter of cannulated consistently maximal blood flow rates greater
vessels. 12 In large central veins, there is less than 300 mL / min and this becomes a limiting
380 Leblanc et al

factor for any high-flux modality.22 Currently, 10% to 15% is expected, particularly at higher
the maximal blood flow rates achieved with flows. 22,24 In a recent survey of 27 patients
venovenous central dialysis catheters (with an dialyzed in Montpellier and having the Twin-
acceptable venous resistance) reach 400 to 450 Cath, various levels of blood flow rate as
mL / min, as provided by the TwinCath. This indicated by the dialysis machines were com-
catheter, for which a large experience is avail- pared to the measured blood flow rates using a
able in Europe, uses two free floating single 10 noninvasive method relying on ultrasound
Fr lumen catheters generally inserted within velocity dilution (Transonic Systems Inc, Ithaca,
the same vein and having two distinct subcuta- NY, and Meditor, France).2s,26 A mean differ-
neous tunnels of 8 to 10 cm. 2l ence of 7.5 ± 9.0% was found in a total of 114
An interesting study by Athirakul et aP3 measurements. Overall, the discrepancies did
reported the maximal achievable blood flow not appear significant from a clinical perspec-
with three different cuffed permanent cath- tive, although they tended to be greater with
eters (in 30 treatments per catheter); blood increasing blood flows (Fig 2).
flow reached 396 mL / min with TwinCath, 384
mL/min with PermCath, and a significantly Catheter Tip and Distal Holes
lower mean value of 320 mL / min with VasCath The position of the catheter tip may influence
SoftCell (Bard Instrument Co, Toronto, On- the blood flow achieved. Nevertheless, correct
tario). The maximal blood flow rate was lim- positioning of the tip into the distal innomi-
ited by the generated negative pressure (lower nate vein or proximal superior vena cava is
than -300 mm Hg). essential to minimize complications; the tip
On the other hand, if the treatment option is should lie above the superior vena cava-right
continuous or slow intermittent renal replace- atrial junction to avoid damage to the heart
ment therapy, as often used in critcally ill acute (perforation of the right chambers, valvular
renal failure patients, a blood flow of 200 to tear, tamponade). To reach this goal, the inser-
250 mL / min usually suffices, and most tempo- tion depth should be kept at 20 cm or less in
rary catheters available on the market are most adults. Although placing the catheter tip
suited to deliver such flows. However, it should lower, ie, into the right atrium, may reduce
be realized that the effective blood flow rate is malfunction and allow higher blood flow
not indicated accurately by most dialysis ma- rates,27 this practice is potentially dangerous
chines, and, in general, an overestimation of and not advocated. 28-30

500T----------------------------------------------------,

400 368±30
400

300

200

Figure 2. Blood flow rate


100 indicated by the blood
pump (_) compared with
the blood flow rate mea-
sured by the ultrasound ve-
o locity method (~). A total
of 114 measures were per-
Recorded Qb versus measured Qb (mUmin) formed.
Critical Care Nephrology: Catheter Dysfunction 381

Various catheter types vary significantly wall and possibly induction of hemolysis. At
when considering their distal end (Fig 3). blood flow rates of 350 to 400 mL/min, the
Numerous distal holes should better preserve positive pressure on the venous or efferent
blood flow and prevent any partial luminal catheter side should be less than 250 mmHg.
obstruction from causing an extreme resis- The ratio of the venous pressure to the blood
tance to blood flow. However, these lateral flow rate should usually be around 1:2.32
holes can be occluded by a fibrin sleeve devel- Increased pressures on both sides (more nega-
oping around the catheter. On theoretical tive arterial pressure and more positive ve-
grounds, it then appears advantageous to nous pressure) indicate a greater resistance to
ensure two free-floating lumens within the flow, strongly suggesting catheter dysfunc-
vessel for several reasons: the larger indi- tion. For example, with the TwinCath, a signifi-
vidual lumens and the more numerous holes cant relationship is observed between achieved
can provide higher blood flows at lower resis- blood flows and arterial and venous pressures
tive pressures; the longitudinal distance be- as monitored by the dialysis machine (Fig 4).
tween inflow and outflow sides can be rela-
tively longer than in dual-lumen single Blood Recirculation
catheters; and the lateral distance taking place Recirculation of dialyzed blood within the
between the two catheters could potentially catheter itself is another performance param-
reduce blood recirculation (as discussed be- eter to monitor since this phenomenon has a
low). negative impact on treatment efficiency. Unfor-
Pressure Recordings tunately, there is probably an obligatory mini-
mal blood recirculation rate in almost all dual-
Pressures recorded by the dialysis machine on lumen catheters (or double catheters when
both afferent and efferent blood sides provide inserted into the same central vein). To mini-
useful information and are also indicative of mize or abolish central catheter blood recircu-
catheter performance. 31 During hemodialysis, lation, either two different catheters would
the negative pressure generated by the blood have to be inserted in two distant veins or one
pump on the" arterial side" of the catheter to each into an artery and a vein, as for continu-
ensure blood inflow into the extracorporeal ous arteriovenous modalities performed in
circuit, can potentially induce vein collapse, critically ill patients. Obviously, such options
especially in conditions of hypovolemia. Ide- are not suitable in day-to-day practice. One
ally, the pressure on the arterial side should should also remember that cardiopulmonary
not be more negative than 300 mmHg to recirculation does not occur with venovenous
prevent vessel collapse with damage to its catheters because, by definition, this phenom-
enon relates exclusively to arteriovenous ac-
cesses. 33
e The information on blood recirculation rates
with indwelling catheters is relatively recent.
Twardowski et ap4 reported a negligible recir-
culation of 2% in long-term well-functioning
.I
r_-_-_-_-_-_--_-_-_-_-_-_~_~~~r~~--------~~
• r--""'""
B PermCath and VasCath (VasCath Inc, Missis-
sauga, Ontario) catheters implanted at the
Dual·lumen upper thorax level. Blood recirculation was
Double-catheter shown to depend in part on the site of catheter
C insertion. A few studies have now confirmed
.,..J ;J e ~t" 0
.......... '!.-+
I' that femoral catheters, particularly the shorter
0 ones, are associated with higher blood recircu-
Double cathe ter lation rates. 35-38 Kelber et ap5 showed in 17
patients having temporary catheters (Flexxi-
Figure 3. Different catheter types: (A) dual-lumen
with co-axial side holes and a distal hole; (B) con, VasCath Inc, Mississauga, Ontario) recir-
dual-lumen with distal holes; (C) twin or double culation rates of 4%, 2%, and 10% in internal
catheter with distal and side holes. jugular, subclavian, and 24-cm femoral vein
382 Leblanc et al

Arterial Pressure (mmHg) Venous Pressure (mmHg)


y = -O.68x + 14.39 r2 = 0.97 y = O.83x - 98.99 r2 = = 0.92
o ~~-----~,..,....,..., 350------------~~

300
250
200
150
100
50
O+-~--~--~~--;
o 100 200 300 400 500 o 100 200 300 400 500
Blood Flow (mL/min) Blood Flow (mL/min)
Figure 4. Arterial and venous pressures as monitored by the dialysis machine in relation to blood flow rate for
well-functioning TwinCaths.

catheters at a set blood flow rate of 250 cm) femoral catheters when compared to
mL/min. Those rates did not significantly 19.5-cm femoral or 13.5- and 19.5-cm subcla-
increase when the blood flow was increased vian catheters (Table 1; Fig 5).36 There was no
up to 400 mL / min. However, 15-cm femoral relationship between recirculation rates and
catheters had much higher recirculation rates arterial or venous resistances recorded during
(18 :::!: 7% and 38 :::!: 14% at respective blood dialysis.
flows of 250 and 400 mL / min). Kraus et al also obtained similar results. 37
In a study of 38 dual-lumen temporary They studied various temporary catheter
dialysis catheters (Quinton), recently inserted brands (silicone and polyurethane) inserted in
for acute dialytic support (mean delay of 2 femoral and subclavian veins for acute dialytic
days), and well-functioning, we also found support and reported that femoral catheters
much higher recirculation rates in short (13.5- recirculated significantly more at all blood

Table 1. Temporary Central Catheter Blood Recirculation Rates


Recirculation (% )
Qb Time AR VR
Mean SD SEM Range (mUmin) (h) (mmHg) (mmHg) URR(%) N
All 10.3 9.2 1.3 0-33.9 298::!:: 9 3.8::!:: 0.4 248::!:: 38 133::!:: 19 57.8::!:: 13.0 50
Femoral
All 16.1* 9.1 1.8 1.9-33.9 3.8 ::!::0.4 54.5::!:: 14.2 26
19.5 em 12.6 6.9 1.7 1.9-29.4 296::!:: 11 3.6 ::!:: 0.3 239::!:: 44 130::!:: 18 54.4::!:: 14.6 17
13.5 em 22.8t 9.1 3.0 9.3-33.9 300::!:: 6 3.9::!:: 0.2 259::!:: 24 136::!:: 13 54.8::!:: 14.3 9
Subclavian
All 4.1 3.6 0.7 0-12.9 3.9::!:: 0.4 62.5 ::!:: 10.9:\: 24
19.5 em 5.4 4.3 1.3 0-12.9 300::!:: 7 3.9::!:: 0.4 264::!:: 30 138::!:: 22 60.8::!:: 10.2 11
13.5 em 3.0 2.6 0.7 0-7.1 298::!:: 8 3.9 ::!:: 0.5 240::!:: 38 132::!:: 22 63.9::!:: 11.6 13
NOTE. AR, VR, Qb, and URR are expressed as means:!: SD.
Abbreviations: AR I VR, arterial and venous resistances; Qb, blood flow rate; URR, urea reduction ratio.
"Femoral versus subclavian catheter blood recirculations, P = .OOOL
t13.S cm versus 19.5 cm femoral catheter blood recirculations, P = .004.
:j:Subclavian versus femoral catheter urea reduction ratios, P = .03.
Reprinted with permission from Leblanc et al. 36
Critical Care Nephrology: Catheter Dysfunction 383

12

10
...
I/)

.S!
Q) 8
.c
cou
6
'0
...
Q)
.D 4
E
:::l
Z
2

0
0-3.0 3.1-6.0 6.1-10.0 10.1-15.0 15.1-20.0 20.1-25.0 25.1-30.0 30.1-35.0

Catheter blood recirculation rate (%)

Figure 5. Distribution of recirculation rates (50 measurements) for subclavian (E21) and for femoral temporary
dialysis catheters (lwl) well functioning and recently inserted. Reprinted with permission from Leblanc et aP6

flows tested. Mean blood recirculation rates We recently evaluated 26 well-functioning


for femoral catheters were, respectively 12.8 ± TwinCath catheters of variable residence time
13.2%,24.5 ± 18.0%, and 24.7 ± 19.0% at blood (24 hours to 24 months) for blood recirculation
flows of 200, 250, and 300 mL/min, whereas rates at increasing blood flow rates (200 to 400
for subclavian catheters recirculation was only mL/ min) using a different method. All these
4.2 ± 3.5% at a blood flow rate of 300 mL / min. cuffed and tunneled catheters were inserted
The explanation for this phenomenon may into the right internal jugular vein, except two
be anatomic; short (13- to 15-cm) femoral (one in the subclavian and one in the femoral
catheters probably do not reach the inferior vein). A total of 110 measures were performed
vena cava in most adult patients, and their tip with the ultrasound velocity dilution tech-
may thus remain in a smaller vein providing a nique (Transonic Systems Inc, and Meditor),
reduced bulk flow and a potential for higher thus avoiding the pitfalls of blood sampling.
blood recirculation. The results are presented in Fig 6. A large
Finally, Athirakul et aP3 compared three variability was observed in recirculation rates
types of cuffed permanent catheters inserted at the different blood flows tested, and the
mainly in the right jugular vein, and reported correlation between recirculation and blood
mean recirculation rates of 3.7% for the flow, although slightly positive, was rather
PermCath, 3.9% for the TwinCath, and 4% for weak. Figure 7 is an example of the informa-
the VasCath SoftCell. tion available on the bedside computer after
It is well known that reversing the lines of an injection of saline into the venous drip
the catheter (using the venous port for inflow chamber with this method.
and vice-versa) leads to much higher recircula-
tion rates and lower treatment efficacy34-36; it is
Consequences of Catheter Dysfunction
particularly important for the nursing staff to
be aware of this phenomenon. Central venous catheter dysfunction will com-
Despite the useful information they pro- promise dialysis efficiency and the overall
vide, the above studies could be criticized quality of renal replacement therapy- There-
because of the method used for recirculation fore, any significant dysfunction should be
measurement. All relied on blood-side solute detected and corrected as soon as possible to
concentrations, which are not necessarily reli- prevent further therapy inadequacy and fur-
able. 39 Nonetheless, they all used a similar ther complications (mainly vessel damage).
sampling technique, ie, low or stop-flow two- The well-recognized stenosis and thrombosis
needle method, and their results are in excel- of subclavian veins in dialysis patients has
lent agreement. 40-42 notably reduced the indications for subclavian
384 Leblanc et al

--
0~
40
35 y = 0,024x

-...
R2= 0,0279
Q)
(U 30

• •
-
c
.-
0
25

- 20 • • ••
('Q
:J
... • •
I•
0
15
• I

IIi
0
...
Q)
10

- I ••
"C
0 5
-m
0

0
0 50 100 150 200 250 300 350 400 450

Blood flow rate (mL/min)


Figure 6. Blood recirculation rates in TwinCath of various residence times at different blood flows, as
measured by the dilution ultrasound velocity method (110 measures).

catheter insertion,38,43 while internal jugular placement in acute situations not associated
cannulation, associated with a lesser incidence with a need for long-term dialytic support. 46,47
of thrombosis, is gaining more and more in If simple urea kinetic evaluation is used to
popularity.44,45 In fact, subclavian dialysis cath- monitor dialysis adequacy, one has to be care-
eters should be avoided except for short-term ful for the patient with catheter dysfunction.

Dialyzer Blood Flow (Ob) =295 mUmln Recirculation = 10%


1.20
Volt. .90
.60
.30
.oo+--____ L_C.~~ _ _ __ _ _ _ _ _ __ Figure 7. Example of the
information available on
the bedside computer when
-10 .00 10. 20. 30. 40. evaluating access recircula-
Time (sec)
tion with the Transonic He-
modialysis Monitor (Tran-
M.REe 3 19:06 Ob = 295 mUmln Recirculation =10% sonic Systems Inc, Ithaca,
M.REe 2 19:05 Ob = 337 mUmln
= 376 mUmln
Recirculation =13% NY, and Meditor, France).
M.REe 1 19:04 Ob Recirculation = 19%
Critical Care Nephrology: Catheter Dysfunction 385

The urea reduction ratio (URR) can be mislead- failure patients with a mean duration of use of
ing, being falsely higher with blood recircula- 218 days (ranging from 30 days to 8 years). A
tion, and if the sample is taken immediately total of 620 patients have been on high-
post-session. In our previous study, however, efficiency dialytic modalities with such cath-
URR was lower at high blood recirculation eters for more than 3 months with satisfactory
probably because there was a slight delay in results in terms of dialysis adequacy. The
urea sampling after the session, thus allowing actuarial TwinCath survival (unassisted pa-
the recirculation effect to dissipate. 36 tency) for this subgroup is shown in Figure 9.
For the same reason, using a urea sample In another series of 64 PermCath catheters in
drawn immediately post-dialysis to calculate 51 patients, the mean duration of use was
Kt/V may overestimate the effective dialysis 316 ± 337 days, leading to actuarial 1-, 2-, 3-,
dose in patients with significant blood recircu- and 4-year survival rates of 74%, 43%, 25%,
lation. More precise and sophisticated measure- and 12%,17 Another survey of 210 patients also
ments of blood recirculation and double-pool using PermCath catheters reported mean cath-
Kt/V calculations may help to detect access- eter longevity rates between 12 and 85 weeks. 49
related problems. As a broad guideline, the However, in this last study, flow and clotting
tolerated blood recirculation rate within cen- problems were frequent with respective inci-
tral venous dialysis catheter should be less dences of 46% and 24%.
than 10% to 15% in most settings, and ideally
between 3% to 7%, especially in situations
mandating an extremely efficient dialysis ses- Diagnostic Approach and
sion (low dialysis dose, high catabolism, intoxi- Therapeutic Options
cations). Catheter dysfunction is manifested clinically
As a permanent access for patients without when the desired blood flows cannot be
a functioning peripheral access, a good com- reached or when abnormal venous and/ or
promise has been to use the TwinCath, prefer- arterial resistance pressures are indicated on
ably implanted into the right internal jugular the dialysis machine. A complete catheter oc-
vein, and to keep a length difference of 3 to 5 clusion will generally be detected by the inabil-
cm between both tips (Fig 8).21,48 During the ity to aspirate blood from the catheter port(s)
last decade, 1,176 permanent TwinCath cath- with a syringe as performed routinely at the
eters have been implanted in 997 chronic renal beginning of the session. An attempt at flush-
ing an obstructed lumen might be helpful but
is not without risk and should not be a routine
procedure. One has to be aware of the small
risk for clot embolization into the vasculature
and the lung.l o However, clinical experience
shows that the consequences of this manipula-
tion are usually minimal. 50
A catheter obstruction or a complete occlu-
sion may be confirmed by local intracatheter
instillation of contrast media (catheterogram)
(Fig 10). To evaluate the host or proximal/
distal veins for possible thrombosis and/or
stenosis, a phlebography is the most accurate
diagnostic test, whereas Doppler ultrasonogra-
phy and 1311 fibrinogen scans are useful but
less reliable alternatives. 51 If catheter occlusion
is diagnosed, intracatheter instillation of a
fibrinolytic agent with a sufficient dwelling
Figure 8. TwinCath inserted into the right internal
jugular vein. Note the distance between the two time will restore patency in the majority of
catheters at the distal end and the subcutaneous cases.9,52-55 Urokinase is most widely used and
tunnel. has been associated with relatively good re-
386 Leblanc et al

Permanent TwinCath (used 3 months or more)

Actuarial Survival (%)


100 N =620
80

60

40

20 '.~''''''
.'~~~ ....
, ..
.
".
"

o L,------.---~--,_~--_.------.,.-'~-.->~-~-~)~-=--=--=-:--~ Figure 9. Actuarial sur-


o 20 40 60 80 100 120 vival of 620 TwinCaths in-
serted for 3 months or
Time (months) longer.

sults. Tissue plasminogen activator has also or without stent placement can be performed
been used successfully. Various proposed pro- with an encouraging initial success. However,
tocols are presented in Table 2. We recommend patency rates decrease over time because of
instillating at least 5,000 IV of urokinase in restenosis. 56,57 The management of central vein
each catheter lumen with a volume sufficient thrombosis should include the removal of the
to fill the lumen completely (to the tip) and catheter. In anecdotal cases, this has led to a
leave it in situ 1 hour. After flow restoration spontaneous resolution of the thrombus. Fi-
with a thrombolytic agent, many centers now nally, surgical thrombectomy may be neces-
advocate low dose oral anticoagulants to main- sary in cases of failure of less aggressive
tain catheter patency.9 approaches.
If subclavian stenosis is found, percutane- Pulmonary emboli and obstruction to blood
ous transvenous angioplasty of the vessel with return to the right side of the heart are the
or without fibrinolytic administration and with most lethal or feared complications; thus, pre-

Figure 10. Contrast media


injected into a TwinCath
(catheterogram). The lu-
men on the left is patent
while a distal occlusion is
seen on the right.
Critical Care Nephrology: Catheter Dysfunction 387

Table 2. Thrombolytic Agents to Reopen Clotted Hemodialysis Catheters


Investigator Agent Proposed Regimen
Athirakul and Schwab54 Urokinase 5,000 IU /mL For PermCath: 1 mL per branch plus 0.2 or 0.3 mL
of saline in proximal or distal ports, respectively.
Aspirate after 30 minutes; 0.3 mL of saline may
be instilled every 10 minutes (X3).
Shrivastara et al53 Urokinase 5,000 IU / mL Inject an adequate filling volume in each line for 1
(can be diluted 1:2 hour; if still occluded, a similar dose in both
with saline) ports can be injected and left in situ for 24 hours.
Glynn et al63 Urokinase diluted in 2.5 mL in occluded branch left for 3 hours then
saline aspirate and flush the line with heparin.
Lawson et al55 Urokinase 5,000 IU / mL 0.2 to 0.4 mL or volume sufficient to fill catheter
lumen. Attempt to aspirate every 5 minutes
X 15 minutes. If not patent after 1 hour, a second
similar instillation is recommended.
Powelson JA (personal Urokinase 5,000 IU /mL 1.2 mL in each port left in situ for 20 minutes. If not
communication) successful, systemic urokinase can be consid-
ered: 125,000 IU (one 250,000 IU vial is reconsti-
tuted with 5 mL sterile water and its total con-
tent is added to 150 mL of D5W) infused over 5
minutes in each port (250,000 IU over 10 min-
utes). Pretreat with Benadryl and continue with
intravenous heparin.
Paulsen et al52 TIssue plasminogen acti- For PermCath: volume sufficient to fill the lumen.
vator 2 mg/2 cm3 Dwell time over 60 minutes or up to next
dialysis.

vention and early detection are essential. If lems (avoiding catheter bending or kinking,
suspected, transesophagal ultrasonography is appropriate suturing, verifying immediate
useful to document a thrombus extending into flows, flushing the lumen[s] with saline fol-
the right atrium. Because it prevents an exten- lowed by heparin immediately after vessel
sion of a subclavian, internal jugular, innomi- insertion, localizing of the catheter tip and
nate, or superior vena cava thrombus, sys- position on x-ray). The creation of a subcutane-
temic anticoagulation is recommended for a ous tunnel allows a better stability of the
period of 6 weeks to 3 months. catheter for longer use and is more comfort-
In the absence of vessel thrombus and when able for patients; indeed tunneled catheters
the catheter lumen appears relatively patent, it have been called "permanent."17,48,49,59 The rela-
seems reasonable to replace a dysfunctional
tive decrease in infectious risk attributed to
catheter over a guidewire, although this prac-
subcutaneous tunnels remains controversial
tice is controversial from an infectious compli-
and is probably more in relation to the inser-
cation standpoint because it may increase the
tion technique and catheter care. 9,60,61
risk of bloodstream infection. 58 One has to
measure this inherent risk against the poten- Care of the catheter after its insertion is of
tial complications resulting from catheter inser- utmost importance, particularly with regard to
tion at a different site. infectious complications. 62 Dialysis catheters
must be reserved exclusively for hemodialysis
sessions and the exit site covered after the
Prevention of Catheter Dysfunction procedure. To maintain patency between treat-
Prevention of catheter malfunction involves ments, each lumen should be flushed with
considering the characteristics of the catheter saline and heparin should be left in situ. The
chosen, the site and technique of insertion, the heparin concentration is anywhere between
residence time of the catheter, and the fre- 5,000 to 10,000 IU / mL and the amount to
quency of previous catheterization( s) at a given infuse should be determined carefully to fill
site. Several precautions taken at the time of each catheter lumen totally and to reach its
implantation prevent some mechanical prob- distal end. The local heparin is usually re-
388 Leblanc et al

moved before tubing connection at the next the Quinton Permcath for vascular access. Nephrol
treatment. Dial Transplant 6:269-274, 1991
18. Canaud B, Leray-Moragues H, Garred LJ, et al: Perma-
nent central vein access. Semin Dial 9:397-400, 1996
19. Uldall PR, DeBruyne M, Besley M, et al: A new
References vascular access catheter for haemodialysis. Am J
Kidney Dis 3:270-277,1993
1. Hoshal VL, Ause RG, Hoskins PA: Fibrin sleeve
20. Mustata 5, Less P, Agraharar M, et al: Further experi-
formation on indwelling subclavian central venous
ence with a percutaneously inserted double-lumen
catheters. Arch Surg 102:353-358,1971
Silastic catheter for end-stage renal failure patients
2. Madden B, Huissoon Ap, Colhoun EN, et al: Superior
with refractory vascular access problems. ASAIO J
vena caval obstruction: A rare complication of Quin-
42:244-245,1996
ton double-lumen haemodialysis catheters. Nephrol
21. Canaud B, Beraud JJ, Joyeux H, et al: Internal jugular
Dial TranspI4:586-587, 1989
vein cannulation with two silicone rubber catheters: A
3. Domoto DT, Garvin PJ: Recurrent venous thrombosis
new and safe temporary vascular access for hemodialy-
associated with permanent internal jugular vein hemo-
sis. Thirty month's experience. Artif Organs 10:397-
dialysis catheters. Trans Am Soc Artif Intern Organs
403, 1986
33:140-143,1987
4. Trottier SJ, Veremakis C, O'Brien J, et al: Femoral deep 22. Collins AJ: High-flux, high-efficiency procedures, in
vein thrombosis associated with central venous cath- Henrich WL (ed): Principles and Practice of Dialysis.
eterization: Results from a prospective, randomized Baltimore, MD, Williams & Wilkins, 1994, pp 76-88
trial. Crit Care Med 23:52-59,1995 23. Athirakul K, Conlon P, Schwab 5: Cuffed central
5. Clark DD, Albina JE, Chazan JA: Subclavian vein venous hemodialysis catheters and adequacy of dialy-
stenosis and thrombosis: A potential serious complica- sis. J Am Soc NephroI7:1402-1403, 1996
tion in chronic hemodialysis patients. Am J Kidney 24. Schmidt DF, Schniepp BJ, Kurtz SB, et al: Inaccurate
Dis 15:265-268, 1990 blood flow rate during rapid hemodialysis. Am J
6. Spinowitz BS, Galler M, Golden RA, et al: Subclavian Kidney Dis 17:34-37, 1991
vein stenosis as a complication of subclavian catheter- 25. Krivitski NM: Theory and validation of access flow
ization for hemodialysis. Arch Intern Med 147:305- measurement by dilution technique during hemodialy-
307,1987 sis. Kidney Int 48:244-250, 1995
7. Paganini EP, Pudelski B, Bednarz D: Dialysis delivery 26. Krivitski NM: Novel method to measure access flow
in the ICU-Are patients receiving the prescribed during hemodialysis by ultrasound velocity dilution
dialysis dose? J Am Soc NephroI3:384, 1992 (abstr) technique. ASAIO J 41:M741-M745, 1995
8. Rahn NH, Rubin E, Koehler RE: Thrombophlebitis of 27. Feldman HI, Kobrin 5, Wasserstein A: Hemodialysis
the internal jugular vein: Noninvasive imaging. South vascular access morbidity. J Am Soc Nephrol 7:523-
Med J 77:1308-1312, 1984 535, 1996
9. Bander SJ, Schwab SJ: Central venous angioaccess for 28. Defalque RJ, Campbell C: Cardiac tamponade from
hemodialysis and its complications. Semin Dial 5:121- central venous catheter. Anesthesiology 50:249-252,
128,1992 1979
10. MacLean-Ross AH, Griffith CDM, Anderson JR, et al: 29. Food and Drug Administration: Precautions neces-
Thromboembolic complications with silicone elas- sary with central venous catheters. Washington, DC,
tomer subclavian catheters. J Parenter Enteral Nutr US Government Printing Office, 1989, p 15
6:61-63,1982 30. McGee WT, Ackerman BL, Rouben LR, et al: Accurate
11. Curelaru I, Gustavsson B, Hansson AH, et al: Material placement of central venous catheters: A prospective,
thrombogenicity in central venous catheterization. II. randomized, multicenter trial. Crit Care Med 21:1118-
A comparison between plain silicone elastomer and 1123,1993
plain polyethylene, long, antebrachial catheters. Acta
31. Stroud CC, Meyer SL, Bawkon MC, et al: Vascular
Anaesthesiol Scand 27:158-164,1983
access for extracorporeal circulation. Resistance in
12. Eberhart RC, Clagett CP: Catheter coatings, blood
double lumen cannulas. ASAIO J 37:M418-M419, 1991
flow, and biocompatibility. Semin Hematol 28:42-48,
32. Raja MR: Vascular access for hemodialysis, in Ing TS,
1991
13. Sioshansi P: New processes for surface treatment of Daugirdas JT, (eds): Handbook of Dialysis. Boston,
catheters. Artif Organs 18:266-271, 1994 MA, Little, Brown, 1994, pp 53-77
14. Nichols AB, Owen J, Grossman B, et al: Effect of 33. Schneditz D, Kaufman AM, Polaschegg H, et al:
heparin bonding on catheter-induced fibrin formation Cardiopulmonary recirculation during hemodialysis.
and platelet activation. Circulation 70:843-850, 1984 Kidney Int 42: 1450-1456, 1992
15. Krafte-Jacobs B, Sivit CJ, Mejia R, et al: Catheter- 34. Twardowski ZJ, Van Stone Jc, Jones ME, et al: Blood
related thrombosis in critically ill children: Compari- recirculation in intravenous catheters for hemodialy-
son of catheters with and without heparin bonding. J sis. J Am Soc NephroI3:1978-1981, 1993
Pediatr 126:50-54,1995 35. Kelber J, Delmez JA, Windus DW: Factors affecting
16. Vanholder R, Ringoir 5: Vascular access for hemodialy- delivery of high-efficiency dialysis using temporary
sis. Artif Organs 18:263-265, 1994 vascular access. Am J Kidney Dis 22:24-29, 1993
17. Gibson Sp, Mosquera D: Five years experience with 36. Leblanc M, Fedak 5, Mokris G, et al: Blood recircula-
Critical Care Nephrology: Catheter Dysfunction 389

tion in temporary central catheters for acute hemodi- 50. Anderson AJ, Krasnow SH, Boyer MW, et al: Hickman
alysis. Clin NephroI45:315-319, 1996 catheter clots: A common occurrence despite daily
37. Kraus M, McMcusky M, Clark W: Temporary hemodi- heparin flushing. Cancer Treat Rep 71 :651-653, 1987
alysis catheter recirculation studies: A comparison of 51. Caruana RJ, Raja RM, Zeit RM, et al: Thrombotic
design and site selection. Blood Purif 13:390, 1995 complications of indwelling central catheters used for
(abstr) chronic hemodialysis. Am J Kidney Dis 6:497-501,
38. Uldall R: Hemodialysis access-Part A: Temporary, in 1987
Jacobs C, Kjellstrand CM, Koch KM, et al (eds):
52. Paulsen D, Reisoether A, Aasen M, et al: Use of tissue
Replacement of Renal Function by Dialysis. Kluwer
plasminogen activator for reopening of clotted dialy-
Academic Publishers, Dordrecht, 1996, pp 277-292
sis catheters. Nephron 64:468-470,1993
39. Hester RL, Curry E, Bower J: The determination of
hemodialysis blood recirculation using blood urea 53. Shrivastava D, Lundin Ap, Dosunmu B, et al: Salvage
nitrogen measurements. Am J Kidney Dis 6:598-602, of clotted jugular vein hemodialysis catheters. Neph-
1992 ron 68:77-79, 1994
40. Pederson JA, Dunlay R, Williams C, et al: Two-needle 54. Athirakul K, Schwab SJ: Opening thrombosed silastic
calculation of recirculation compared with the stan- hemodialysis catheters. Semin Dial 9:289, 1996
dard three-needle method. Clin Nephrol 33:203-206, 55. Lawson M, Bottino JC, Hurtubise MR, et al: The use of
1990 urokinase to restore the patency of occluded central
41. Sherman RA: The measurement of dialysis access venous catheters. Am J Intr Ther Clin Nutr 1:29-32,
recirculation. Am J Kidney Dis 22:616-621,1993 1982
42. Sherman RA, Matera JJ, Novik L, et al: Recirculation 56. Beathard GA: Percutaneous transvenous angioplasty
reassessed: The impact of blood flow rate and the in the treatment of vascular access stenosis. Kidney Int
low-flow method reevaluated. Am J Kidney Dis 23:846-
42:1390-1397,1992
848,1994
57. Newman GE, Saeed M, Himmelstein S, et al: Total
43. Cimochowski GE, Worley E, Rutherford WE, et al:
central vein obstruction. Resolution with angioplasty
Superiority of the internal jugular over the subclavian
access for temporary dialysis. Nephron 54:154-161, and fibrinolysis. Kidney Int 39:761-764,1991
1990 58. Cobb DK, High KP, Sawyer RG, et al: A controlled trial
44. Schillinger F, Schillinger D, Montagnac R, et al: Post- of scheduled replacement of central venous and pul-
catheterisation vein stenosis in hemodialysis: Com- monary-artery catheters. N Engl J Med 327:1062-1068,
parative angiographic study of 50 subclavian and 50 1992
internal jugular access. Nephrol Dial Transplant 6:722- 59. Kinnaert P, Hooghe L, De Pauw L, et al: Use of the
724,1991 Hickman catheter as permanent vascular access for
45. Agraharkar M, Isaacson S, Mendelssohn D, et al: hemodialysis. Trans Am Soc Artif Intern Organs 36:104-
Percutaneously inserted silastic jugular hemodialysis 106,1990
catheters seldom cause jugular vein thrombosis. 60. Moss AH, Vasilakis C, Holley JL, et al: Use of a silicone
ASAIO J 41:169-172, 1995 dual-lumen catheter with a dacron cuff as a long-term
46. De Moor B, Vanholder R, Ringoir S: Subclavian vein vascular access for hemodialysis patients. Am J Kid-
hemodialysis catheters: Advantages and disadvan- ney Dis 16:211-215, 1990
tages. Artif Organs 18:293-297, 1994
61. De Meester J, Vanholder R, De Roose J, et al: Factors
47. Konner K: Subclavian haemodialysis access: Is it still
and complications affecting catheter and technique
justified in 1995? Nephrol Dial Transplant 10:1988-
survival with permanent single-lumen dialysis cath-
1991,1995
eters. Nephrol Dial Transplant 9:678-683, 1994
48. Canaud B, Cristol JP, Klouche K, et al: Silastic twin
catheters: A flexible solution for temporary or long- 62. Vanherweghem JL, Dhaene M, Goldman M, et al:
term vascular access problems in hemodialysis pa- Infections associated with subclavian dialysis cath-
tients, in Man NK, Botella J, Zucchelli P (eds): Blood eters: The key role of nurse training. Nephron 42:116-
Purification in Perspective. Cleveland, OH, ICAOT 119,1986
Press, 1992, pp 25-29 63. Glynn MFX, Langer B, Jeejeebhoy KN: Therapy for
49. Dunea G, Domenico L, Gunnerson P, et al: A survey of thrombotic occlusion of long-term intravenous alimen-
permanent double lumen catheters in hemodialysis tation catheters. J Parenter Enteral Nutr 4:387-390,
patients. ASAIO Trans 37:M276-277, 1991 1980

You might also like