ILLUMINANCE METER
USER’S MANUAL
V2.02
Foreword
Thank you for purchasing the EVERFINE Z-10 Illuminance Meter. This user’s
manual contains useful information involving the instrument’s functions and operating
procedures as well as precautions that should be noticed during operation. In order to
use the instrument correctly, please read the manual carefully first, then put it in a
right place for quick references.
Notes:
EVERFINE pursues a continuing improvement of the performance and functions
of its products, therefore, the contents of this manual may be changed without
prior notice.
Great effort has been made in preparation of this manual to ensure the accuracy of
its contents. If you have any questions or find any errors, please contact your
dealer or EVERFINE sales office.
If you have different understanding to this manual, please refer to the Technical
Service Department of EVERFINE.
Please check the instrument carefully when you unpack the box for the first time.
If the instrument and related accessories are missing or appear abnormal, please
contact the dealer immediately.
Copyright Statement
The copyright of this manual and the related information belong to EVERFINE,
and it is protected by the copyright law of Peoples Republic of China and other
relevant international treaties. Copying, modifying, spreading, excerpting, backing up
and translating the whole or part contents of this manual by any company or personnel
without the written permission of EVERFINE is prohibited. Otherwise it will be
treated as infringement and the infringer will assume law responsibility and all loss of
EVERFINE. Any infringement related above can be traced back to the responsible
user by the unique product number printed in the manual.
If EVERFINE has signed a written agreement with user and the contents in the
agreement are in conflict with above terms, the contents in the written agreement have
preferential force effect.
Safety Precautions
The following general safety precautions must be observed during all phase of
operations, including service and repair. If the instrument is used in the manner not
specified in this manual, the function of this instrument may be hindered.
1. Keep the instrument in dry environment.
2. If there has dust on the optical surface, use the wash-ear blower to clean it.
Touching the optical surface with hand or hard articles is strictly prohibited.
3. Comprehend the configuration, performance and the operation rules before
operating the instrument.
4. If the down right corner of LCD display" " during operation, it means the
power is short and needs to switch a battery. Turn off the instrument firstly and open
the battery cover to change.
5. Calibrate or verify the instrument annually.
Contents
Foreword........................................................................................................................ 1
Copyright Statement.......................................................................................................2
Safety Precautions.......................................................................................................... 3
Contents..........................................................................................................................4
Chapter 1 Overviews......................................................................................................5
Chapter 2 Specifications.................................................................................................6
Chapter 3 Instrument Instructions.................................................................................. 7
Chapter 4 Operation.................................................................................................. 10
4.1 Integrated mode...............................................................................................10
4.2 Separated mode............................................................................................... 15
4.3 Multi-point mode.............................................................................................16
Chapter 5 Common Trouble and Troubleshooting.......................................................19
Chapter 1 Overviews
Standard .
Chapter 2 Specifications
Measurement
±1% (Standard class) (according to JJG245-2005)
accuracy*
User Correction
CCF (Color Correction Factor)
Function
Communication
USB communication
interface
Dimension 176mm×71mm×30mm
software
* Please refer to JJG 245-2005 for the classification; and refer to JJG245-2005, CIE 69:1987, EN13032-1:2004 or
1. Detector component
2. LCD display
3. Buttons
1. Detector
2. Detector adapter
3. Connecting line
4. Host adapter
Chapter 4 Operation
4.1 Integrated mode
Connect the host ant the detector, as shown in fig4.1.
The host will read the ID code of the detector automatically after turning on and
the code will be displayed in the top right corner of LCD. As shown in fig4.2.
" " means the address of detector is "00" and " " means the
current illuminance is 123.45lx.
It requires to turn off the instrument if need to modify the ID code (please refer
to 4.3 for details). If ID code is modified during measurement, the host can't identify
and abnormal situation may arise. Please operate carefully.
Press key to enter illuminance difference mode after turning on. The
value of current illuminance minus the reference illuminance (range:
-999k~300klx) will be shown on the main area of LCD. The bottom of LCD
displays the ratio of measuring illuminance to reference illuminance
(0.001%~9999%).
After entering illuminance difference measurement mode, it needs to set
"reference illuminance", the procedures are as follows:
and .
so as to keep the current data and the sign "HOLD" will be shown in the LCD.
Release HOLD: Press "HOLD" button again to release hold state and the sign
"HOLD" disappears. The instrument re-enters illuminance measurement state.
the sign " " appears. The bottom of LCD shows the cumulation time of
integrating (0.001h~9999.9h).
After entering integral illuminance mode, it needs to set "upper limit of integrated
illuminance", the procedures are as follows:
a. press " " under integral illuminance mode to enter into the
corresponding setting interface.
keep the current data and the sign "HOLD" will be shown on LCD.
Release HOLD: Press "HOLD" button again to release hold state and the HOLD
signal disappear. The instrument re-enters integral illuminance measurement
state.
keep the current data and the sign "HOLD" will be shown on LCD.
Release HOLD: Press "HOLD" button again to release hold state and the HOLD
signal disappear. The instrument re-enters average integral illuminance
measurement state.
Note: it is not available to set integral time and integral illuminance at the
same time. The latter set is effective.
Press key under illuminance measurement mode. The LCD shows the
current illuminance which has multiplied CCF (range: 0.01~600k lx) and the sign
"CCF".
Press key to select a digit. One of the five digits flickering means that
Press HOLD key under illuminance measurement mode to enter hold state, and
4.1.8 Memory
Press "HOLD" key to enter hold state under illuminance measurement mode, and
then press to save the holding data.In the bottom of LCD, there will
show the memory number for 1 second. The number is between 00 and 30. All
data can be saved when the power is off.
b. Press" + " or " + "at the same time to check the ID switch
a. Set CCF for each detector respectively, no matter the instrument is under
CCF mode or not.
b. Under CCF mode, the instrument will display CCF value no matter the
instrument is under which measurement mode.
MULTI AVG display
a. Any detector presents failure (exceeding limit, communication error etc), the
instrument will not display value during the measurement.
b. The instrument will not display value under illuminance
difference/illuminance ratio, integral illuminance and automatic integral stopping
mode. Even the CCF is effective, the sign "CCF" will not be shown.
If the troubles can’t be solved, please connect with the Technical Service Department
of EVERFINE.