You are on page 1of 5

BTL CardioPoint Holter

RECORDER

BTL-08 HOLTER H100 H300 H600


RECORDER

ECG lead system (number of leads) 3 leads 3 or 7 leads 3.7 or 12 leads


Number of electrodes attached to the
5 electrodes 5 electrodes 5 or 10 electrodes
patient
1 - 7 days*
Max. record duration 1 - 2 days 1 - 7 days
*12 leads 1 - 4 days
Certified for use with children lighter
than 10 kg
Certified for ST segment evaluation

Patient activity monitoring N/A N/A 2 axes accelerometer


100 µs / dedicated circuit with 40000 Hz
Pacemaker detection N/A
detection function
Patient button / voice recording

Signal quality check On built-in display or in BTL CardioPoint via USB

Automatic start of recording 20 minutes after inserting the batteries

Automatic continuation of the record


after replacing the batteries

ECG frequency response 0.049 - 220 Hz (filters off)

ECG: Delta-Sigma A/D converter, 8 x 2000 Hz


Sampling frequency
Pacemaker: 2 axis detection by dedicated circuit with 40000 Hz function

Digital resolution 24 bits / 1.52 µV

Dynamic range ± 66 mV

Input signal range Dynamic range: ±33 mV, polarisation: ±393 mV, consistent voltage offset: ±5 V
5 lead cable: >100 dB
Common mode rejection ratio (CMRR)
10 lead cable: >94 dB

Noise level < 7 µV

Dimensions (including patient cable


102 x 62 x 24 mm
connector)
Operating conditions Temperatuture: 1°C - 55°C, humidity: 10 % - 95 %
Weight 106 g without batteries, 155 g including batteries
Battery type 2 x AA > 1.2 V
Lithium AA: cca 116 hours
Battery capacity / type Alkaline AA: cca 75 hours
NIMH 2500 mAh: cca 51 hours
Ingress protection IP22

BTL CARDIOPOINT HOLTER 55


SOFTWARE

BTL CARDIOPOINT HOLTER H100 H300 H600

SIGNAL OVERVIEW

Fully customizable interface


Interactive trends
Graphical user interface Preffered lead for analysis setting
Filter view settings
Selective view of channels

QRS MORPHOLOGY

Automatic QRS classification into templates (N, B, V, S, J, artifact)


Advanced manual QRS labeling
Template merge
Evaluation tools Group templates re-marking
ECG phragment view, 3 min interval view, Tiles view
QRS to template navigation
Manual measurement with auto snap to P, PQ, QRS, QT

Graphical split

Look for similar beat

RHYTHM AND ARRHYTHMIAS

Interactive HR trend
Detailed data table
Automatic rhythm detection
Evaluation tools
(SR, STachy, SBrady AF, V rhythm, AIVR, VT, SVT, SVER)
Prematurity beat setting
Parametrs limit setting

Interactive patient activity trend

Relief map

Waterfall

Poincare

Tachogram

ISCHEMIE EVALUATION

Interactive ST analysis, ST table, ST deviation trend


Evaluation tools
Parameters limit setting, anual measurements

ST level and slope trend per lead

ST relief map

56 BTL CARDIOPOINT HOLTER


BTL CARDIOPOINT-HOLTER H100 H300 H600

AUTONOMIC NERVOUS SYSTEM EVALUATION, HRV ANALYSIS

Evaluation tools Time and frequency analysis, circadian index, LF/HF graph

Trend graphs

Tachogram
2D and 3D power spectral density
graph - PSD
Poincare

PQ, QT ANALYSIS

Evaluation tools Table values, parameters limit settings, QTc method selection

Interactive PQ, QT and QTc trend

PQ histogram

Scattergram

Relief graph

PACEMAKER ANALYSIS

Automatic detection of pacemaker


pulses
Morphology segmentation
(Pv, Pa, Pd, Pf, ~f)
Interactive trend, histogram
and table data
Zone and rate settings

Pacemaker type selection

REPORT

Individual automatic report


Multiple template report options
Report options Automatic conclusion editor
Manual medical finding editor
Instant snapshot to report function
Export to hospital informational system HL7
standards DICOM
Export to file system PDF, SDF, EDF, BDF, XML, CSV formats (FAT32 file system)

BTL CARDIOPOINT HOLTER 57


BTL ABPM 61

You might also like