You are on page 1of 6

Specifications & more

HumaStar 80
>> System Specifications
>> Consumables
>> Accessories

System Overview Cat. No. 16880


Analyzer type Automatic analyzer
Reagent system Open (Human pre-installed)
Throughput (max. and typical) 80 tests/hr maximum
40 tests/hr typical
Operation modes Batch analyzer with STAT
Barcode reader None
ISE None
Reaction / reading System Micro Flow Cell 18 µl
Languages available English, French, Italian, Spanish
User interface External PC
HumaStar 80
Automated Clinical Chemistry Batch Analyzer

Tests Measuring modes Routine chemistry, Turbidimetry


Analysis modes End point
Differential
Fixed time
Kinetic
Calculation modes Factor
Multi Standard
Single calibrator
Calculation algorithms Spline, linear regression, square regression, polygonal
# of programmable tests Unlimited
# of programmable profils Up to 9
# of standards per test Up to 8

Sample / Control /
Calibrator Sample type Serum, plasma, urine, CSF
Sample vessels Sample cups 1.2 ml
Sample positions Total of 54 positions for samples, calibrators, controls
Sampling volume 2 to 200 µl/test (in increments of 1 µl steps)
Loading STAT
Pre-Dilution Yes
Post-Dilution Yes

Reagent Reagent positions 20 reagent positions (45 ml per position)


Reagent cooling None
Reagent 1 volume range 30 to 1000 µl in 1µl steps
Reagent 2 volume range 0 to 1000 µl in 1/16 µl steps
Reagent 3 volume range None

Reaction unit Reaction wells /


positions for incubation 12 rows with 12 wells each (REF 16885/54)
Reaction volume (minimum) 400 µl
Reaction volume (maximum) 1 ml
Cuvette washing system n.a. (flow cell)
Incubation temperature 37°C (+/-1°C in transfer arm, ± 2°C in cuvette, ± 0.2°C in flow cell)
Incubation time 1st: 21 to 9999 s; 2nd: 0 to 180 s

Liquid Handling Liquid transportation Reagent and sample: Single-syringe


Measurement: Continuous flow system with peristaltic pump
Liquid level sensors Resistance
Collusion detection for probe Yes
Reading Optical system Interference filters
Readings Mono- or bichromatic
Light source Halogen lamp (Osram, 12 V, 20 W)
Spectral range 320 to 690 nm
Wavelengths pre-installed 340, 405, 492, 505, 546, 578, 630 nm
Max. # of wavelengths installed 7
Wavelength error (accuracy) ± 2 nm
Wavelength bandwidth (precision) 8 ± 2 nm
Detector Silicon photodiode
Absorbance range (linearity) -0.200 to 2.500 O.D.
Resolution detector 0.0001 OD
Precision detector CV<1% at 2 OD

Data Processing Memory for Sample results, calibration, patient data, QC data, blank
Memory capacity Unlimited
Reports for Patient and result report, kinetic curves, QC and calibration
Quality control module Levey-Jennings, Shewart, Analytical limits control: blank, linearity, factor
Max. number of control levels Up to 3 levels per test
Test statistics None
Warnings Analytical limits, control, blank, linearity, factor
Printer Printer connected to external PC
LIS RS-232 serial port, bi-directional, proprietary protocol

General PC min. requirements External PC, Processor 2GHz; CD-ROM; Memory: 256 Mbytes RAM;
Hard disk capacity > 20 GBytes; R
S 232 and USB 2.0 port,
Operating system: Windows 98, XP or Windows 7 (software min. h1. 12E)
Physical dimensions (W x D x H) Instrument without any components: 72 x 73 x 45 cm
Space required for routine use: 105 x 93 x 85 cm
Packaging: 87 x 86 x 89 cm
Weight: Gross: 79.0 kg, net: 42.0 kg
Mains fuse 2x 5A (velocity: f)
Electrical requirements 230/110 VAC (± 15%) (autosense); 50/60 Hz; 350 VA
Environmental Operating: temperature 18...34 °C, humidity < 80 % rel. non condensing
Transport: temperature -10...50°C, humidity< 80% rel. non condensing
Wash / waste tank 0.5 l each with level sensor; external waste tank possible


HumaStar 80
Automated Clinical Chemistry Batch Analyzer

Scope of Supply Unit/Size Cat.No.


HumaStar 80 1 16880
Packlist
Settings CD 1 -------
CD Software 1 -------
CD Layout Files 1 -------
Unpacking Sheet 1 -------
Reagent Racks 2 16885/52
PC with Windows 7 1 18992P
Sample Tray (one in instrument) 1 16885/511
Reagent Bottles 45 ml & Caps 40 pcs. 1 16885/53
Reagent Rack
Black Reagent Bottles 45 ml & Caps 5 pcs. 1 -------
Cat.No. 16885/52
Reaction Wells Segment 50 pcs. 1 16885/54
Sample Cups of 1.2 ml 500pcs. 1 16885/55
Kodak Tubes for R2 of 2.5 ml 20 pcs. 1 16885/202
Pages of Reagent Label 2 -------
Tube for External Tank 1 m 1 -------
Tubekit, Wash- & Waste Bottle 1
16885/56 + /57
Wash Station for Needles 1 16885/59
Part of the Case (to put under the arm) 1 16885/66
Washstation
for Needles Cleaning Needle 1 -------
Cat.No.
16885/59 Halogen Lamp 12 V 20 W 1 16885/48
Fuses of 5 A 1 18750/14
Power Cable 1 18999
Serial Cable 1 -------
User Manual 1 16880/1

Obligatory Items Service Kits Cat.No


Need to be ordered with the first instrument:
New Starter Spare Part Kit HumaStar 80 from S/N 20586 (Sufficient for up to 5 instruments) 16880/501

Need to be ordered with each instrument:
New Maintenance Kit HumaStar 80 from S/N 20586 16880/401

Accessories
Monitor LCD 19 inch 1 17901M
HP Laser Printer for Human Instruments (USB + Parallel) 1 18993L
Monitor and printer are necessary for the operation of this instrument.
Human offers the items above.


Obligatory Items Accessories Unit/Size Cat.No.
UPS Power Supply Unit 1KVA / 230 V 1 18961
UPS - Use of an online UPS with this instrument is obligatory. If this item is to be purchased locally,
the specifications must be approved by HUMAN tech service prior to the HumaStar 80 delivery.

Consumables
Wash Additive 4 x 25 ml 18971

Wash Additive
Cat.No. 18971

Optional Items Consumables Cat.No


Kodak Tubes 2.5 ml 20 pcs. 16885/202
Reagent Bottle + Caps 40 pcs. 16885/53
Reaction Wells 50 pcs. (for 600 Tests) 16885/54
Sample Cups 1.2 ml 1000 pcs. 16885/55

Reaction Wells Segment Kodak Tubes for R2 of 2.5 ml Sample Cups 1.2 ml
Cat.No. 16885/54 Cat.No. 16885/202 Cat.No. 16885/55

Reagent Bottles + Caps


Cat.No. 16885/53
16880/2013-08 © 2013 Human GmbH

Legal statement
While Human is making every effort to include accurate and up-to-date information, we make

no representations or warranties, express or implied, as to the accuracy or completeness of the

information provided in this document and disclaim any liability for the use of it.