You are on page 1of 6

Specifications & more

HumaCount 30TS
> System Specifications
> Scope of Supply
> Obligatory Items
> Optional Items

System Overview Cat.No. 16420/30


Analyzer type Automatic hematology analyzer
Reagent system Closed
# of Parameters 22 parameter, 3-part differential of WBC
Throughput 30 samples / hour
Barcode reader External 2D barcode reader (optional)
Reaction / reading system Impedance method (1 chamber)
LED 540nm for HGB measurement
Languages available English, French, German, Indonesian, Italian,
Portuguese, Russian, Spanish, Czech, Polish
User interface Color touch screen LCD
high-contrast, backlit, 320 x 240 -dots
HumaCount30TS
Automatic Hematology System

System Specifications
Tests Parameters WBC, LYM, MON, GRA, LYM%, MON%, GRA%, HGB, RBC, HCT, MCV, MCH,
MCHC, RDWcv, RDWsd, PLT, PCT, MPV, PDWcv, PDWsd, P-LCC*, P-LCR*
Calibration 1-, 2-, or 3-measurement automatic
Manual (factors) calibration of WBC, HGB, RBC, PLT, MCV (or HCT), RDW, MPV
Independent calibration of pre-diluted mode.
Veterinary mode Yes
Self check Yes
Flags/Messages Interpretative messages, (diagnostic flags)
Warning flags

Sample Sample type Whole blood, diluted blood


Sample vessels Open tube system with automatic sample rotor
Standard EDTA tubes
Sample positions 1
Sampling volume 26 µl of whole blood or in pre-diluted mode
Dilution Yes, internal

Reagent Reagent type Diluent


Lyse
Cleaner

Liquid Handling Liquid transportation Dilutor pumps, vacuum


Liquid level sensors Yes, internal

Reading Light source Green LED for HGB


Wavelengths pre-installed 540 nm
Detector Light to frequency converter
Aperture diameter 70μm (RBC/PLT/WBC)

* research only

HumaCount 30TS Seite 2 von 6


Data Processing Memory for QC, records including flags and histograms
Memory capacity 10,000
Reports for Result, QC, self check, calibration
Quality control module 16- and 64-day Levey-Jennings charts
Separate QC database (6 level)
Max. number of control levels 6
Warnings Range, linearity, errors
Printer Built-in thermal printer 58 mm wide roll paper, full report with histograms
External all HP printers via USB
LIS Special protocol on USB 2.0 port

General Physical dimensions (W x D x H) Instrument without any components: 32 x 26 x 36.5 cm


Space required for routine use: 52 x 46 x 50 cm
Packaging: 50 x 41 x 59 cm
Weight: Gross: 12.5 kg, net: 12 kg
Electrical requirements External auto-ranging power unit for 100...120 or 200...240 Vac, 50–60Hz, 60 VA
Environmental Operating: temperature: 15…30°C, humidity: < 80 % non-condensing
Storage: temperature: 15…30°C, humidity: < 80 % non-condensing
Altitude: up to 3,500 m (with special software)
Wash / waste tank Waste Container 20 L

Characteristics

Parameter: Accuracy Reproducibility Carry-over sample Test range Unit


(CV) to sample
WBC 3% 2.7% <1% 4.0-20.0 103/μl
RBC 3% 1.7% <1% 4.0-15.0 106/μl 
HCT 3% 3% <1% 25.0-50.0 %
MCV 2% 1.7% N/A 5-9 fl
HGB 2% 2% <1% 9-16 g/dl
PLT 5% 6% <3% or <20 200-900 103/μl


Seite 3 von 6 HumaCount 30TS
HumaCount 30TS
Automatic Hematology System

Scope of Supply Unit/Size Cat.No.


HumaCount 30TS 1 90050
Packlist
External Power Supply and Cable 1 16420/168
Power Safety Socket 1 18969
Adapter EU to USA 1 18967
Adapter EU to UK 1 18968
Reagent Tubing Kit 1 16420/324
1 x Diluent Tube (green)
1 x Lyse Tube (yellow)
Reagent Tubing Kit 1 x Cleaner Tube (blue)
2 x Waste Tube (red)
1 x Cleaning Tube Kit
4 x Caps for Reagent Containers
Thermal Printer Paper (2 rolls) 1 16420/621
Tube Adapter I 1 16420/180
Tube Adapter II 1 16420/181
Tube Adapter III 1 16420/323
Tube for Vacuum Pump 1 ……….
User Manual 1 16420/601
Waste Container 20 L + Box 1 ……….

Obligatory Items Service Kits Cat.No


Need to be ordered with the first instrument:
HumaCount 30TS Starter Spare Part Kit (Sufficient for up to 5 instruments) 16420/200

Need to be ordered with each instrument:


Lubrication Kit (for HumaCount, HumaCount Plus, HumaCount 30TS & 60TS, HumaCount 5L) 16420/12

System Reagents Cat.No


HC-Diluent (20 L) 17400/10
HC-Lyse CF (1 L) (Cyanid-free reagent) 17400/20
HC-Cleaner (1 L) 17400/30

HumaCount 30TS Seite 4 von 6


Obligatory Items
Controls (Strongly recommended) Cat.No
HC-CONTROL (3 x 2.5 ml) 17400/40
As with all other blood control products on the market, the hematology controls have a limited shelf life.
As a result, controls are available only on the basis of a standing order in accordance with a fixed delivery
schedule. Controls are delivered as separate shipment under cooling conditions.

HC-CONTROL (Level 1, 2, 3)

Accessories Cat.No
UPS - Power Supply Unit 1 KVA / 230 VAC 18961
Use of an online UPS with this instrument is obligatory. If this item is purchased locally, the specifications
must be approved by HUMAN technical service prior to the HumaCount 30TS delivery.

Optional Items Consumables Cat.No


Thermal Printer Paper (57 mm) 5 pcs. 18144/5

Accessories Cat.No
Upgrade Kit to close HC30TS & HC60TS 16430/331
HumaPower System 230 Volt (rechargeable battery 12 Volt) 18965/12
Human SolarCharge (solar panel) 18965/14
2D Barcode Reader 16430/11
Tube Adapter I 16420/180
Tube Adapter II 16420/181
Tube Adapter III 16420/323

HumaPower System Human SolarCharge 2D Barcode Reader

Tube Adapter I Tube Adapter II Tube Adapter III


Seite 5 von 6 HumaCount 30TS
16420/30/2015-07 © 2015 Human

Legal statement
While Human is making every effort to include accurate and up-to-date information, we make = this item is classified as dangerous goods

no representations or warranties, express or implied, as to the accuracy or completeness of the

information provided in this document and disclaim any liability for the use of it.

You might also like