SE-9000
NE-8000 SF-3000
K-4500
E-5000
M-2000 KX-21
K-
CC-800 1000
CC-780
PL-100 F-820
CC-180 F-800
CC-120
CC-107
CC-1001
Main Menu
Introduction
.
Main Menu
.. small footprint
little sample volume
..
Main Menu
easy handling
automatic system-check during start-up
Main Menu
Display
. LCD-display with user-
friendly menu structure
. in case of error-messages
automatical switch into the
HELP-program
Main Menu
Main Menu
Main Menu
Stored Data
Main Menu
Specifications
18 parameters:
3 histograms:
.. clear LCD-display
HELP - program available
. Pre-diluted Mode 20 µl
. Cellpack
. Stromatolyser-WH
Only one reagent for Leukocyte- and
Hemoglobin measurement
Alarm in case of use of non-Sysmex reagents!
Cellpack 20 L 34 ml 580
Throughput
• 60 samples / hour
Cycle time
• 60 seconds / sample
.. 1 X bar
6 QC files for control blood
...
20 parameters
60 data points
L-J charts
. Eightcheck-3WP
Stored Data internal stored data for 240 samples (without histogram)
external stored data in K-DPS available
Main Menu
.
Main Menu
. 60 samples/ hour
for medium
throughputs
. little sample-
volume
paediatrics,
emergency labs
.
Main Menu
LCD-display
clear information
. automatic system
check
safety
. HELP-program
uncomplicated
trouble shooting
Main Menu
. hot keys
easy handling
.
fast access
flag-system
helpful support
. little maintenance
time saving
. maintenance intervals
are displayed work simplifying
.
Main Menu
. LCD-display
low radiation
no ageing
. non-toxic reagent
system environment
protection
Main Menu
Measurement Principles
orifice
sample beaker
vacuum
U=RxI
250 µl 500 µl
Diaphragm-
pumps
pulse height
pulse number
t1= 0,5 sec = t2= 0,5 sec = t3= 0,5 sec
pulse number t1 pulse number t2 pulse number t3
pulse number
t1= 0,5 sec t2= 0,5 sec t3= 0,5 sec
pulse number t1= =
pulse number t2 pulse number t3
time
1 2 3 4 5 6 7 8 9 10 11 12 13 14
pulse height
time
Main Menu
Pulse Diagram
(30 cells)
pulse height
1 2 3 4 5 6 7 8 9 1011 1213 14
30
Cumulative
20
Distribution Curve
10
Cumulative
20
Distribution Curve
10
1 2 3 4 5 6 7 8 9 1011 1213 14
4 1 0 0 0 1 2 3 4 5 4 3 2 1
Histogram
Histogram 2000
1500
0
0 6 12 18 24 30 36 42 48 54 60 66 72 78 84 90 96 102108114120
stop start
signal-pulses
noise-pulses
discriminator
signal-pulses
off on
Not to be changed by the user:
counter sample volume;
dilution ratio;
analysed volume;
diameter of the orifice
display:
The instrument automatically checks:
S pulse / volume counting time, condition of the orifice
during the analysis (noise-control)
Advantages:
no calibration of the counts,
pulse management digital automatical check of the orifice
SYSMEX EUROPE GMBH
38
KX-21
DC Detection Method
Training Relative Counting
sample aperture
Main Menu vacuum
reference-sample
for system -
calibration
signal-pulses
noise-pulses
discriminator
Rated value of
reference-sample signal-pulses
Counting of particles per time unit
Unsafety:
pulses /time
reference sample;
electrical strain condition of the orifice;
digital-analog changer vacuum
Adjustment of
the strain to pulse management digital
the rated value
of the ref.
sample pulse management analog
analog-digital changer
(digital-Voltmeter)
display:
SYSMEX EUROPE GMBH
39 digital Volt
(= rated value of the Ref.-sample)
KX-21
DC Detection Methods
Training Comparison
Main Menu
25-75 fl 200-250 fl
fixed at
2-6 fl 12 fl 12-30 fl
Main Menu
Main
Main Menu
Menu
15
10
0
New borns Babies Children Women Men
. MTAC Myristyltrimethylammoniumchloride
. LTAC Lauryltrimethylammoniumchloride
Hemoglobin-molecule
MTAC +
LTAC
RBC RBC
Hemoglobin-molecule
RBC RBC
Hemoglobin-molecule
O2
Methemoglobin-complex
Absorption STROMATOLYSER-WH
SULFOLYSER
. 2. reaction step:
Methemoglobin reacts with Potassium cyanide (KCN),
building the red and stable Cyan hemoglobin-complex with
an maximum in absorption at 546 nm.
Start
Main Menu
Hematocrit Determination
Main Menu
0,4
0,3
0,2
0,1
0
New borns Babies Children Women Men
Total volume VT
Centrifuge
VT
HCT (%) = V / VT x100
60
V SYSMEX EUROPE GMBH
KX-21
Hematocrit
Training Centrifugation
Main Menu
Transducer
Start-Sensor
Defined volume VT
Stop-Sensor
Ph = k x VEry
VEry = 1 / k x Ph
Ph Pulse height
k Constant
Ph VEry Volume of one RBC
Ph = k x VEry
VEry = 1 / k x Ph
V = VEry
V = 1 / k Ph
Ph Pulse height
k Constant
VEry Volume of one RBC
SYSMEX EUROPE GMBH
64
KX-21
Hematocrit
Training Cum. Pulse Height Detection Method
Main Menu
VT
Ph
Transducer Ph = k x VEry
HCT (%) = V / VT x 100
VEry = 1 / k x Ph
V= VEry = 1 / k Ph
Ph Pulse height
k Constant
HCT (%) = 1 / VTk Ph x
VEry Volume of one RBC
100
SYSMEX EUROPE GMBH
65
KX-21
Training
Main Menu
Erythrocyte Quotients
Biological Variation
Main Menu
60
40
-20
103/µl 103/µl g/dl x 106/µl % fl g/dl pg
-40
Main Menu
Sample Processing
Main Menu
Sample Processing
1,994 ml
1,0 ml WBC detection chamber
Stromatolyser-WH 1:500
(PDV)
Sample Rotor Valve
6 µl
HBG
photometer
1:500
50 µl whole blood
SYSMEX EUROPE GMBH
71
KX-21
Training
Main Menu
Sample Processing
Pre-diluted Mode
1,922 ml
1,0 ml WBC detection chamber
Stromatolyser-WH 1:1000
(PDV)
Sample Rotor Valve
78 µl
HBG
photometer
1:1000
Main Menu
Dilution Ratios
1. dilution
4 µl whole blood + 1,996 ml Cellpack
= 500 dilution factor
4 µl whole blood
2. dilution
40 µl 1. dil.. + 1,960 ml Cellpack
= 50 dilution factor
40 µl 1. dilution
Pre-dilution
20 µl whole blood + 500 µl Cellpack
= 26 dilution factor
20 µl whole blood
1. dilution
2,08 µl pre-dil. + 1997,92 µl Cellpack
= 962 dilution factor
2,08 µl pre-dilution
pre-diluted mode:
Pre-dilution
20 µl whole blood + 500 µl Cellpack
= 26 dilution factor
20 µl whole blood
1. dilution
78 µl pre-dil. + 2,922 ml Str-WH
= 38,5 dilution factor
78 µl pre-dil.
5.000 cells / µl
x 500 µl = 5.000 cells
500 (dilution ratio)
Pre-diluted mode:
5.000 cells / µl
x 500 µl = 2.500 cells
1.000 (dilution ratio)
RBC/PLT WBC
dilution factor counted dilution ratio counted
cells / µl cells / µl
whole blood mode
KX-21 25.000 50.000 500 5.000
K-1000/K-4500 25.000 50.000 250 8.000
pre-diluted mode
KX-21 25.000 50.000 1.000 2.500
K-1000/K-4500 25.000 50.000 1.250 1.600
Main Menu
Upper part
(fixed)
Middle part
(movable)
Lower part
(fixed)
Aspiration pipette
Upper part
(fixed)
Middle part
(movable)
Lower part
(fixed)
Aspiration pipette
Upper part
(fixed)
Middle part
(movable)
Lower part
(fixed)
Aspiration pipette
Diaphragm-Pump
50 µl whole blood
SYSMEX EUROPE GMBH
86
KX-21
Sample Rotor Valve
Training Sample Separation
Main Menu
6 µl (WBC) 4 µl (RBC)
mixing chamber
4 µl (RBC)
diluent-
Diaphragm pump
SYSMEX EUROPE GMBH
88
KX-21
Sample Rotor Valve
Training Aspiration for 2. RBC-dilution
Main Menu
mixing chamber
2. dilution
& Diaphragm pump
SYSMEX EUROPE GMBH
89
KX-21
Sample Rotor Valve
Training Dosage for 2. RBC-dilution
Main Menu
40 µl
RBC-transducer-
chamber
diluent-Diaphragm pump
SYSMEX EUROPE GMBH
91
KX-21
Sample Rotor Valve
Training WBC- and HGB- Sample-dilution
Main Menu
WBC-transducer
chamber
HGB-detector block
6µl (WBC)
diluent-Diagphragm pump
SYSMEX EUROPE GMBH
92
KX-21
Training
Main Menu
Histograms
RBC
250
RDW-SD 40,0 fl
PLT
40
PDW 13,1 fl
MPV 10,4 fl
SYSMEX EUROPE GMBH
94 P-LCR 28,1 %
KX-21
Histograms
Training Normal haematological Report
Interpretation
NO. 4
Main Menu DATUM: 9/10/95
15:11
MODE: whole blood
WBC
103/µl
5,8 x CBC - parameter
RBC 4,84
x106/µl
HGB 13,7 g/dl
HCT 42,0 %
MCV 86,8 fl
MCH 28,3 pg WBC
WBC
MCHC 32,6 g/dl
PLT 257
300
Leukocyte-histogram
x103/µl
LYMPH% 31,2 %
MXD% 6,8 % Lymphocytes in % and absolute
NEUT% 62,0 % Eo, Mono, Baso in % and absolute
LYMPH# 1,8 x103/µl Neutrophiles in % and absolute
MXD# 0,4 x103/µl
NEUT# 3,6 x103/µl
Erythrocyte-histogram
RBC
RBC
Erythrocyte distribution width
250
RDW-SD 40,0 fl
Thrombocyte-histogram
PLT Thrombocyte distribution width
Mean Platelet Volume
PLT
40 Platelets - Large Cell Ratio
PDW 13,1 fl
MPV 10,4 fl
SYSMEX EUROPE GMBH
95 P-LCR 28,1 %
KX-21
Histograms
Training Reference Ranges of the red CBC
Main Menu
MCV 85-95 fl
MCH 27-33 pg 1,68-2,05 fmol
MCHC 32-36 g/dl 19,9-22,4 mmol/l
Histograms
2 fl 25-30 fl 250 fl
s µ s
68,26 %
of all values
100 %
RDW-SD: 37 - 46 fl
20 %
SYSMEX EUROPE GMBH
98
KX-21
RBC- Histogram
Training Interpretation
Main Menu
Histograms
Base line
2 fl 250 fl
LD
RBC
PLT
LD
RBC
PLT
•Platelet aggregation
UD
RBC
UD
RBC
RBC
RBC
Possible cause:
• Transfusions, the patient’s and the donor’s RBCs most
probably have different sizes.
RBC
RBC
Turning Points
s µ s RDW-CV
68,26 % 11 - 16 %
of all values
100 % The erythrocyte distribution
width is used as an indicator for
anisocytosis.
20 %
RDW-SD
37 - 46 fl
SYSMEX EUROPE GMBH
108
KX-21
PLT- Histogram
Training Reference Ranges
Main Menu
• The distribution curve should lie between the discriminators and start
and end at the base line.
• The measurement area is between 2 and 30 fl.
PLT
PLT
PLT
Curve does not start at the base line.
Possible cause:
• Transfusions, the patient’s and the donor’s PLTs most
probably have different sizes.
PLT
PLT
• The distribution curve should lie between the discriminators and start
and end at the base line.
• The measurement area is between 2 and 30 fl.
Electrolyte-solution
Mitochondrion
Nucleus
Nucleolus
Cell membrane
Ribosome
Cytoplasm
after lysis
Lymphocytes
CAUTION:
WBC - results and all results flagged
“WL” should be checked
(counting chamber if needed)!
Possible cause:
• Thrombocyte aggregates, Pseudo-Thrombopenia,
sample coagulated
• elevated osmotic resistance
(erythrocytes are not lysed)
• Normoblasts / Erythroblasts SYSMEX EUROPE GMBH
• Cold Agglutinins
KX-21
WBC-Histogram
Training Flagging
Main Menu
CAUTION:
WBC - results and all results flagged
“WU” should be checked
(counting chamber if needed)!
LD T1 T2 UD ( fixed)
F = fraction
population 3 = F3
population 1 = F1 population 2 = F2
Main Menu
Maintenance
1. External control
2. Turn KX-21 on
Turn KX-21 with the power switch on (right instrument side).
3. An automatical rinse is performed.
Acceptable blanks:
WBC ± 0,3 x 103/µl
RBC ± 0,02 x 106/µl
HGB ± 0,1 g/dl
PLT ± 10 x 103/µl
4. KX-21 turns into READY.
After a pre-defined time period the system turns into Standby.
5. Perform the quality control.
Switching Off
Daily:
Clean transducer chambers and tubing (Shutdown)
Check trap chamber and empty if needed.
Weekly:
Clean tray below the Sample Rotor Valve.
As needed:
..Automatical rinsing sequence
Clean rinsing unit
.Clog removal
Program: ”Clog removal"
Main Menu
Main Menu
K-1000 KX-21
Measurement Principles DC Detection Method DC Detection Method
(WBC, RBC, PLT) (WBC, RBC, PLT)
SLS-Hb-Method (HGB) Cyanide-free Hb Method (HGB)
Cum. Pulse Height Detection Method Cum. Pulse Height Detection Method
Parameter 18 18
WBC, RBC, HGB, HCT WBC, RBC, HGB, HCT
MCV, MCH, MCHC, PLT MCV, MCH, MCHC, PLT
W-SCR, W-MCR, W-LCR W-SCR, W-MCR, W-LCR
W-SCC, W-MCC, W-LCC W-SCC, W-MCC, W-LCC
RDW-CV or RDW-SD RDW-CV or RDW-SD
PDW, MPV, P-LCR PDW, MPV, P-LCR
Throughput
Whole Blood Mode 80 samples / hour 60 samples / hour
Pre-diluted Mode 60 samples / hour 60 samples / hour
Cycle Time
Whole Blood Mode 45 seconds 60 seconds
Pre-diluted Mode 60 seconds 60 seconds
Sample Volume
Whole Blood Mode 100 µl 50 µl
Pre-diluted Mode 40 µl 20 µl
SYSMEX EUROPE GMBH
136
KX-21
Training
Comparison
Main Menu
K-1000 KX-21
Precision [ % ] Whole Blood Capillary Blood Whole Blood Capillary Blood
WBC 3,0 9,0 3,5 6,0
RBC 1,5 4,5 2,0 3,0
HGB 1,5 4,5 1,5 2,5
HCT 1,5 4,5 2,0 3,0
MCV 1,5 4,5 2,0 3,0
MCH 1,5 4,5 2,0 3,0
MCHC 1,5 4,5 2,0 3,0
PLT 4,0 12,0 6,0 9,0
Main Menu
K-1000 KX-21
Data Storage No 240 samples without histogram
Main Menu
Competition
Main Menu
Abbott CD 1600
Coulter AcT 8
Hoffmann-La Roche Cobas-
Minos STX
ABX Micros
Dr. Lange LC 90
Becton-DickensonQB II
SYSMEX EUROPE GMBH
140
KX-21
Competition
Training CD 1600
Main Menu
Competition
K-X-21 CD 1600
Manufacturer TOA MEDICAL ELECTRONICS Abbott
Measurement Principles DC Detection Method DC Detection Method
(WBC, RBC, PLT) (WBC, RBC, PLT)
SLS-Hb-Method (HGB) Cyan methemoglobin-Method (HGB)
Cum. Pulse Height Detection Method
Parameter 18 18
WBC, RBC, HGB, HCT WBC, RBC, HGB, HCT
MCV, MCH, MCHC, PLT MCV, MCH, MCHC, PLT
W-SCR, W-MCR, W-LCR GRAN (%, #), LYMPH (%,#), MID (%,#)
W-SCC, W-MCC, W-LCC
RDW-CV or RDW-SD RDW
PDW, MPV, P-LCR PDW, MPV, MPC
Throughput 60 samples / hour 60 samples / hour
Cycle Time 60 seconds 60 seconds
Sample Volume
Whole Blood Mode 50 µl 30 µl
Pre-diluted Mode 20 µl 40 µl
Competition
KX-21 CD 1600
Precision [ % ] Whole Blood Capillary Blood Range
WBC 3,5 6,0 2 4,6 - 10,2 x 103/µl
RBC 2,0 3,0 1 4,0 - 6,1 x 106/µl
HGB 1,5 2,5 1 12,2 - 18,1 g/dl
HCT 2,0 3,0
MCV 2,0 3,0
MCH 2,0 3,0
MCHC 2,0 3,0
PLT 6,0 9,0 5 142 - 424 x 103/µl
Competition
KX-21 CD 1600
Data Storage 240 samples without histogram 320 samples
Competition
K-X-21 AcT 8
Manufacturer TOA MEDICAL ELECTRONICS Coulter (USA)
Measurement Principles DC Detection Method Impedance Measurement
(WBC, RBC, PLT) (WBC, RBC, PLT)
SLS-Hb-Method (HGB) Cyan methemoglobin-Method (HGB)
Cum. Pulse Height Detection Method
Parameter 18 8
WBC, RBC, HGB, HCT WBC, RBC, HGB, HCT
MCV, MCH, MCHC, PLT MCV, MCH, MCHC, PLT
W-SCR, W-MCR, W-LCR
W-SCC, W-MCC, W-LCC
RDW-CV or RDW-SD
PDW, MPV, P-LCR
Throughput 60 samples / hour 55 samples / hour
Cycle Time 60 seconds 60 seconds
Sample Volume
Whole Blood Mode 50 µl 12 µl
Pre-diluted Mode 20 µl 20 µl
Competition
KX-21 AcT 8
Precision [ % ] Whole Blood Capillary Blood Range
WBC 3,5 6,0 3,0 6,0 - 15,0 x 103/µl
RBC 2,0 3,0 3,0 3,0 - 6,0 x 106/µl
HGB 1,5 2,5 2,0
HCT 2,0 3,0
MCV 2,0 3,0 3,0 80 - 100 fl
MCH 2,0 3,0
MCHC 2,0 3,0
PLT 6,0 9,0 7,0 200 - 500 x 103/µl
Competition
KX-21 AcT 8
Data Storage 240 samples without histogram ?
Competition
K-X-21 Cobas-Minos STX
Manufacturer TOA MEDICAL ELECTRONICS Hoffmann-La Roche AG
Measurement Principles DC Detection Method Impedance Measurement
(WBC, RBC, PLT) (WBC, RBC, PLT)
SLS-Hb-Method (HGB) Cyan methemoglobin-Method (HGB)
Cum. Pulse Height Detection Method Cum. Pulse height Detection Method
Parameter 18 18
WBC, RBC, HGB, HCT WBC, RBC, HGB, HCT
MCV, MCH, MCHC, PLT MCV, MCH, MCHC, PLT
W-SCR, W-MCR, W-LCR LY (%, #)
W-SCC, W-MCC, W-LCC MON / GRA (%,#)
RDW-CV or RDW-SD RDW
PDW, MPV, P-LCR MPV, PDW, PCT
Throughput 60 samples / hour 70 samples / hour
Cycle Time 60 seconds 60 seconds
Sample Volume
Whole Blood Mode 50 µl 25 µl
Pre-diluted Mode 20 µl
Competition
KX-21 Cobas-Minos STX
Precision [ % ] Whole Blood Capillary Blood Range
WBC 3,5 6,0 2 10 x 103/µl
RBC 2,0 3,0 2 5 x 106/µl
HGB 1,5 2,5 1 15 g/dl
HCT 2,0 3,0 2 45 %
MCV 2,0 3,0
MCH 2,0 3,0
MCHC 2,0 3,0
PLT 6,0 9,0 5 300 x 103/µl
Competition
KX-21 Cobas-Minos STX
Data Storage 240 samples without histogram 300 samples
Dimensions [ mm ] Main Unit with integrated Main Unit 430 x 370 x 400
Pneumatic Unit Printer 370 x 130 x 440
420 x 480 x 355 Monitor 320 x 330 x 320
Competition
K-X-21 Micros
Manufacturer TOA MEDICAL ELECTRONICS ABX
Measurement Principles DC Detection Method
(WBC, RBC, PLT)
SLS-Hb-Method (HGB)
Cum. Pulse Height Detection Method
Parameter 18 18
WBC, RBC, HGB, HCT WBC, RBC, HGB, HCT
MCV, MCH, MCHC, PLT MCV, MCH, MCHC, PLT
W-SCR, W-MCR, W-LCR W-SCR, W-MCR, W-LCR
W-SCC, W-MCC, W-LCC W-SCC, W-MCC, W-LCC
RDW-CV or RDW-SD RDW-CV or RDW-SD
PDW, MPV, P-LCR MPV, PDW, PCT
Throughput 60 samples / hour 40 samples / hour
Cycle Time 60 seconds 90 seconds
Sample Volume
Whole Blood Mode 50 µl 12 µl
Pre-diluted Mode 20 µl
Competition
KX-21 Micros
Precision [ % ] Whole Blood Capillary Blood
WBC 3,5 6,0
RBC 2,0 3,0
HGB 1,5 2,5
HCT 2,0 3,0
MCV 2,0 3,0
MCH 2,0 3,0
MCHC 2,0 3,0
PLT 6,0 9,0
Linearity [%]
(Whole Blood Mode)
WBC 3,0 10,0 - 99,9 x 103/µl
RBC 3,0 1,0 - 7,0 x106/µl
HGB 1,0 10,0 - 25,0 g/dl
HCT 3,0 33,4 - 60,0 %
PLT 5,0 200 - 999 x 103/µl
Competition
KX-21 Micros
Data Storage 240 samples without histogram option
Competition
KX-21 LC 90
Manufacturer TOA MEDICAL ELECTRONICS Dr. Lange
Measurement Principles DC Detection Method DC Detection Method
(WBC, RBC, PLT) (WBC, RBC, PLT)
SLS-Hb-Method (HGB) photometrical measurement of HGB
Cum. Pulse Height Detection Method at 540 nm wave length
Parameter 18 28
WBC, RBC, HGB, HCT WBC, RBC, HGB, HCT
MCV, MCH, MCHC, PLT MCV, MCH, MCHC, PLT
W-SCR, W-MCR, W-LCR LY (%,#), MO (%,#), GR (%,#)
W-SCC, W-MCC, W-LCC µRBC (%,#), macroRBC (%, #), EVB
RDW-CV or RDW-SD MPV, TKT, Mode PLT, Median PLT
PDW, MPV, P-LCR µPLT (%, #), macroPLT (%,#), PDW
Throughput 60 samples / hour 60 samples / hour
Cycle Time 60 seconds ?
Sample Volume
Whole Blood Mode 50 µl 30 µl
Pre-diluted Mode 20 µl 15 µl
Competition
KX-21 LC 90
Precision [ % ] Whole Blood Capillary Blood
WBC 3,5 6,0
RBC 2,0 3,0
HGB 1,5 2,5
HCT 2,0 3,0
MCV 2,0 3,0
MCH 2,0 3,0
MCHC 2,0 3,0
PLT 6,0 9,0
Linearity [%]
(Whole Blood Mode)
WBC 3,0 10,0 - 99,9 x 103/µl
RBC 3,0 1,0 - 7,0 x106/µl
HGB 1,0 10,0 - 25,0 g/dl
HCT 3,0 33,4 - 60,0 %
PLT 5,0 200 - 999 x 103/µl
Competition
KX-21 LC 90
Data Storage 240 samples without histogram 10.000 samples
Competition
K-X-21 QBC II
Manufacturer TOA MEDICAL ELECTRONICS Becton-Dickinson
Measurement Principles DC Detection Method Centrifugation and buffy-coat
(WBC, RBC, PLT) determination (fluorescence and
SLS-Hb-Method (HGB) absorption of differentiated cell-layers
Cum. Pulse Height Detection Method HGB-calculation from density factors
Parameter 18 9
WBC, RBC, HGB, HCT WBC, HGB, HCT
MCV, MCH, MCHC, PLT MCHC, PLT
W-SCR, W-MCR, W-LCR Gran (% or #)
W-SCC, W-MCC, W-LCC Lymph / Mono (% or #)
RDW-CV or RDW-SD Fibrinogen
PDW, MPV, P-LCR
Throughput 60 samples / hour depending on user
Cycle Time 60 seconds 10 minutes in total
Sample Volume
Whole Blood Mode 50 µl 111 µl
Pre-diluted Mode 20 µl 55 - 65 µl
Competition
KX-21 QBC II
Precision [ % ] Whole Blood Capillary Blood Whole Blood Capillary Blood
WBC 3,5 6,0 7,0 8,0
RBC 2,0 3,0
HGB 1,5 2,5 1,5 1,5
HCT 2,0 3,0 2,0 2,0
MCV 2,0 3,0
MCH 2,0 3,0
MCHC 2,0 3,0 2,5 2,0
PLT 6,0 9,0 9,0 13,5
Linearity [%]
(Whole Blood Mode)
WBC 3,0 10,0 - 99,9 x 103/µl
RBC 3,0 1,0 - 7,0 x106/µl
HGB 1,0 10,0 - 25,0 g/dl
HCT 3,0 33,4 - 60,0 %
PLT 5,0 200 - 999 x 103/µl
Competition
KX-21 QBC II
Data Storage 240 samples without histogram ?
Dimensions [ mm ] Main Unit with integrated Autoread Analyser 343 x 102 x 241
Pneumatic Unit Centrifuge 235 x 127 x 298
420 x 480 x 355
attention!