Professional Documents
Culture Documents
Hematology Calibrator
u(y)
Figure 1 Metrological traceability chain of hematology analyzer’s quantity
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 3, 16 pages in total
In Figure 1:
p.1: set of fit for purpose primary measurement procedure, the methods used for blood cell
analysis mainly include counting, spectrophotometer and length.
p.3: primary measurement procedures defining the measurand. Measured include the cell con-
centration of white blood cells, the cell concentration of red blood cells, the hemoglobin concen-
tration, the hematocrit, and the cell concentration of platelets in normal human fresh blood.
m.4: Manufacturer’s working calibrators. The reference measurement laboratory selects fresh
whole blood from healthy people as the experimental sample for the reference measurement
method, assigns it to the manufacturer's working calibrator.
p.5: Manufacturer’s standing measurement procedure. Hematology analyzers calibrated with the
manufacturer's working calibrator can be used as a standard machine after passing the verifica-
tion of performance tests, etc. The standard machine is routinely maintained in strict accordance
with the requirements, and the reliability of its results is ensured according to the standard ma-
chine management specifications.
m.5: End-user IVD MD calibrators End-user IVD medical device calibrators produced through a
qualified manufacturing process for the calibration of end-user instruments. The reference value
of the end-user in vitro diagnostic medical device calibrator is determined repeatedly by the
standard machine according to the corresponding assignment process, and the average value is
taken as the reference value.
p.6: End-user IVD MD. The assigned end-user IVD medical device calibrators can be used for
the calibration of end-of-line hematology analyzers. Calibrated hematology analyzers test rou-
tine samples with traceable results.
m.6: Human sample with result.
The specific traceability transmission chains involved in the blood cell reference measurement
procedure are as follows:
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 4, 16 pages in total
u(y)
Figure 2 Metrological traceability chain of WBC
WBC count refers to counting the number of white blood cells contained in a unit volume of
blood. The measurand is the amount of white blood cells in human fresh blood, and the type of
quantity is cell concentration and the unit of white blood cell count is /L. The matrix in the white
blood cell count is all components of fresh blood except white blood cells.
After consulting relevant international scientific/medical organizations (JCTLM, ICSH,
CLSI/NCCLS, etc.) and authoritative institutions (WHO, BCR, IRMM, NIST, etc.), WBC only has
the reference measurement procedure recommended by ICSH, and there is no reference mate-
rial. Therefore, we choose this reference measurement program as the reference measurement
program for leukocyte count. According to the ISO 17511:2020 document clause 5.3, the
traceability chain and calibration hierarchy of WBC is shown in Figure 2. p.3 is the primary
measurement procedures defining the measurand (WBC). m.4 is manufacturer’s working cali-
brators. Normal human fresh blood assigned by the reference method (p.3) is selected as the
manufacturer’s working calibrator. p.5 is manufacturer’s standing measurement procedure which
calibrated with the manufacturer's working calibrator. m.5 is end-user IVD MD calibrators.
End-user IVD MD calibrators assigned through a qualified manufacturing process for manufac-
turer’s standing measurement procedure. p.6 is End-user IVD MD. The assigned end-user IVD
medical device calibrators can be used for the calibration of end-of-line hematology analyzers.
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 5, 16 pages in total
m.6.Human sample
%ru(y)c with result
u(y)
Figure 3 Metrological traceability chain of RBC
RBC count refers to counting the number of red blood cells contained in a unit volume of blood.
The measurand is the amount of red blood cells in human fresh blood, and the type of quantity is
cell concentration and the unit of white blood cell count is /L. The matrix in the red blood cell
count is all components of fresh blood except red blood cells.
After consulting relevant international scientific/medical organizations (JCTLM, ICSH,
CLSI/NCCLS, etc.) and authoritative institutions (WHO, BCR, IRMM, NIST, etc.), RBC only has
the reference measurement procedure recommended by ICSH (Reference method for the
enumeration of eryth-rocytes and leucocytes. Clinical and Laboratory Hematology, 16:131-138,
1994), and there is no reference material. Therefore, we choose this reference measurement
program as the reference measurement program for leukocyte count. According to the ISO
17511:2020 document clause 5.3, the traceability chain and calibration hierarchy of RBC is
shown in Figure 3. p.3 is the primary measurement procedures defining the measurand (RBC).
m.4 is manufacturer’s working calibrators. Normal human fresh blood assigned by the reference
method (p.3) is selected as the manufacturer’s working calibrator. p.5 is manufacturer’s standing
measurement procedure which calibrated with the manufacturer's working calibrator. m.5 is
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 6, 16 pages in total
end-user IVD MD calibrators. End-user IVD MD calibrators assigned through a qualified manu-
facturing process for manufacturer’s standing measurement procedure. p.6 is End-user IVD MD.
The assigned end-user IVD medical device calibrators can be used for the calibration of
end-of-line hematology analyzers. m.6 is human sample with result.
2.3 Determination of Hemoglobin Concentration (HGB)
Material Measurement procedure
m.5.End-user IVD
%rucalb
MD calibrators
The hemoglobin concentration refers to the amount of hemoglobin contained in a unit volume (L)
of blood. The measurand is the hemoglobin concentration in human fresh blood, and the type of
quantity is mass concentration, and the hemoglobin concentration is expressed in g/L. The ma-
trix in the determination of hemoglobin is all components in fresh blood except hemoglobin in red
blood cells.
After consulting relevant international scientific/medical organizations (JCTLM, ICSH, CLSI/
NCCLS, etc.) and authoritative institutions (WHO, BCR, IRMM, NIST, etc.), HGB has the refer-
ence measurement procedure recommended by CLSI/NCCLS (standard H15-A3) and ICSH
(recommendations for reference method for hemoglobinometry in human blood (ICSH standard
1995) and specifications for international hemoglobincyanide standard (4th edition)). We se-
lected the latest version of the standard as the reference measurement procedure (standard
H15-A3). Hemoglobin has two reference materials, BCR 522 produced by European Commis-
sion and code: 98/708 produced by NIBSC. They are not primary reference materials, and their
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 7, 16 pages in total
main function is used as a trueness control materials to test the accuracy of the instrument.
There is no difference in performance between them, but the reference material code 98/708 is
cheaper and easier to obtain, so we choose it as the reference material. According to the ISO
17511:2020 document clause 5.3, the traceability chain and calibration hierarchy of HGB is
shown in Figure 4. p.3 is the primary measurement procedures defining the measurand (HGB).
m.4 is manufacturer’s working calibrators. Normal human fresh blood assigned by the reference
method (p.3) is selected as the manufacturer’s working calibrator. p.5 is manufacturer’s standing
measurement procedure which calibrated with the manufacturer's working calibrator. m.5 is
end-user IVD MD calibrators. End-user IVD MD calibrators assigned through a qualified manu-
facturing process for manufacturer’s standing measurement procedure. p.6 is End-user IVD MD.
The assigned end-user IVD medical device calibrators can be used for the calibration of
end-of-line hematology analyzers. m.6 is human sample with result.
2.4 Hematocrit Determination (HCT)
Material Measurement procedure
Hematocrit refers to the percentage of red blood cells in the total blood volume. It reflects the
ratio of red blood cells to plasma. The results of hematocrit determination are expressed as
volume fractions, not involving matrix effects, and the unit is L/L (%).
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 8, 16 pages in total
u(y)
Figure 6 Metrological traceability chain of PLT
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 9, 16 pages in total
Platelet count refers to the number of platelets contained in a unit volume of blood. The type of
amount is cell concentration and the unit of platelet count is /L. The matrix in platelet count is all
components in fresh blood except platelets. After consulting relevant international scien-
tific/medical organizations (JCTLM, ICSH, CLSI/NCCLS, etc.) and authoritative institutions
(WHO, BCR, IRMM, NIST, etc.), PLT has the reference measurement procedure recommended
by WHO (Recommended Methods for the Visual Determination of White Blood Cell Count and Platelet
Count.) and ICSH (Platelet Counting by the RBC/Platelet Ratio Method. Am J Clin Pathol.
2001.115(3):460-4.),and there is no reference material. Both of them are selected for method-
ological comparison. According to the ISO 17511:2020 document clause 5.3, the traceability
chain and calibration hierarchy of RBC is shown in Figure 6. p.3 is the primary measurement
procedures defining the measurand (PLT). The principle of these two reference measurement
procedures is counting: 1) According to the reference method recommended by WHO, dilute
fresh blood samples with 1% ammonium oxalate, and manually count platelets under a micro-
scope. 2) According to the reference method recommended by ICSH, the platelets were fluo-
rescently labeled, measured by flow cytometry, and the ratio of red blood cells to platelets was
counted, and then calculated according to the formula to count the number of platelets in fresh
blood. m.4 is manufacturer’s working calibrators. Normal human fresh blood assigned by the
reference method (p.3) is selected as the manufacturer’s working calibrator. p.5 is manufactur-
er’s standing measurement procedure which calibrated with the manufacturer's working cali-
brator. m.5 is end-user IVD MD calibrators. End-user IVD MD calibrators assigned through a
qualified manufacturing process for manufacturer’s standing measurement procedure. p.6 is
End-user IVD MD. The assigned end-user IVD medical device calibrators can be used for the
calibration of end-of-line hematology analyzers. m.6 is human sample with result.
Sample No. Reference value WBC RBC HGB HCT MCV PLT
Assigned value 1 of
7.40 5.65 155 48.5 85.8 234
working calibrator
2022122701
Relative expanded un-
2.7 1.9 1.9 0.6 1.9 5.8
certainty/%
Assigned value 2 of
6.49 4.59 142 42.1 91.7 215
working calibrator
2022122702
Relative expanded un-
2.7 1.9 1.9 0.6 1.9 5.8
certainty/%
3.2.3Commutability of Calibrator
Commutability is one of the important characteristics of the calibrator. Whether it has commuta-
bility means whether it has a matrix effect, that is, the performance of the calibrator should be
consistent with the performance of the sample to be tested. Use the above-mentioned hema-
tology analyzer for testing, and use the product calibrator whose value is initially assigned as the
standard for calibration. Select 20 fresh venous blood samples anticoagulated with EDTAK2.
When selecting the sample, it should be noted that its values should be distributed as evenly as
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 12, 16 pages in total
possible within the full measurement range of the hematology analyzer. Test each venous blood
sample within 4 hours after collection respectively with reference method (working calibrator)
and the hematology analyzer that is calibrated with the product calibrator. Each sample was
tested 3 times on the blood cell analyzer, and the average value was taken. Three calibrators
were randomly selected and tested by the same method, and the calibrators were randomly in-
terspersed in the new blood samples. Record the corresponding results, perform the correlation
analysis, calculate the correlation coefficient, slope, intercept, and establish linear regression
equation to prove the commutability of manufacturer product calibrator. The result of the hema-
tology analyzer calibrated by product calibrator should be correlated with the result of the he-
matology analyzer calibrated by working calibrator, and the correlation coefficient r≥0.975 (or
equivalent to r2≥0.95), then the value assigned by the product calibrator can be regarded as re-
liable.
Procedure 1: Manufacturer's standing measurement procedure
Procedure 2: End-user IVD MD
Pro- Pro- Pro- Pro- Pro- Pro- Pro- Pro- Pro- Pro-
Sam-
cedure cedure cedure cedure cedure cedure cedure cedure cedure cedure
ple No.
1 2 1 2 1 2 1 2 1 2
WBC WBC RBC RBC HGB HGB MCV MCV PLT PLT
2 7.23 7.11 2.76 2.83 109 110 97.1 99.3 217 219
3 5.8 5.68 4.27 4.32 159 161 90.5 89.4 205 206
4 5.76 5.84 2.03 2.05 144 144 104.6 104.8 226 225
5 10.37 10.37 4.72 4.82 127 127 96.5 96.5 232 226
cali-
bra- 9.33 9.35 4.77 4.75 135 137 84.7 84.7 252 257
tor1
7 11.83 11.63 4.64 4.49 175 175 81.7 82.8 388 402
8 6.91 6.91 5.11 5.13 141 141 75.7 74.4 258 262
9 5.96 5.81 4.59 4.54 142 142 68.7 68.5 257 262
10 7.25 7.41 4.62 4.45 132 130 70.9 70.3 296 297
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 13, 16 pages in total
Pro- Pro- Pro- Pro- Pro- Pro- Pro- Pro- Pro- Pro-
Sam-
cedure cedure cedure cedure cedure cedure cedure cedure cedure cedure
ple No.
1 2 1 2 1 2 1 2 1 2
WBC WBC RBC RBC HGB HGB MCV MCV PLT PLT
12 5.7 5.63 3.84 3.86 133 133 91.2 92.3 239 235
cali-
bra- 9.36 9.44 4.78 4.72 136 137 84.4 84.2 255 254
tor2
14 4.56 4.39 4.82 4.79 128 128 68.5 69.3 204 211
15 2.05 2.03 3.32 3.29 133 133 93.6 94.7 169 165
17 7.2 7.07 4.07 4.04 136 136 103.5 102.9 198 199
18 5.13 5.14 5.07 5.10 122 122 104 103.8 428 445
cali-
bra- 9.37 9.35 4.69 4.70 136 137 84.1 84.1 253 251
tor3
19 5.4 5.33 4.18 4.20 149 148 112.9 110.6 249 250
20 5.46 5.48 4.19 4.17 122 123 113.5 113.8 217 219
Stand
Correlation coefficient r2≥0.95
ard
Judgm
Qualified Judgment Qualified Judgment Qualified
ent
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 14, 16 pages in total
2
𝑢𝑢𝑟𝑟𝑟𝑟𝑟𝑟,𝑐𝑐 = �𝑢𝑢𝑟𝑟𝑟𝑟𝑟𝑟,𝑐𝑐ℎ𝑎𝑎𝑎𝑎 + 𝑢𝑢2𝑟𝑟𝑟𝑟𝑟𝑟,𝑏𝑏𝑏𝑏 + 𝑢𝑢2𝑟𝑟𝑟𝑟𝑟𝑟,𝑠𝑠
When coverage factor k=2, its relative expanded uncertainty Urel=K×urel,c=2×urel,c (K=2 means the
confidence level is 95%)
See table 5:
Assigned value of
9.33 4.70 134 84.5 256
the calibrator
Relative expanded
uncertainty (K=2) 3.5 3.1 2.5 2.4 6.7
Urel (%)
the arithmetic mean of the concentration values measured by the conventional measure-
ment procedure as the Y axis to draw the scatter diagram and perform linear regression
analysis.
2. Draw a regression line and a 95% prediction interval line on the scattergram. Observe the
positions of the results on the scattergram. If they fall within the 95% prediction interval line,
then it indicates that it has the commutability; otherwise, it has no commutability, indicating
that there is a certain difference in matrix effect between the product calibrator and the
clinical sample. The final assigned values of product calibrator need to be adjusted through
methodology.
The evaluation scattergram of the commutability of the product calibrator, the 95% prediction
interval line and the regression line are shown in the figure below.
File No.: RMLAB-WI-T-020
Technical File
Version: Version 1.0
Traceability Research Report
Page: Page 16, 16 pages in total