Professional Documents
Culture Documents
1.3 Configuration and Expansion to the system ................................... 1 1.5 FUNCTION ................................................................................... 10
1.3.1 Variation ................................................................................... 1 1.5.1 Start Up ..................................................................................... 10
1.3.2 Configuration............................................................................ 1 1.5.2 Calibration function................................................................. 11
1.3.3 Reagents.................................................................................. 1 1.5.3 Quality Control Function ......................................................... 12
1.3.4 Control Blood ........................................................................... 1 1.5.4 Online QC Function (optional for SNCS) ............................. 12
1.3.5 Expansion to the system, and interface ................................... 2 1.5.5 Discrete Analysis Function ..................................................... 12
1.4 Performance Characteristics .......................................................... 3 1.5.6 Analysis Order Registration Function (Work Load List Function)
........................................................................................................... 12
1.4.1 Intended Use............................................................................ 3
1.5.7 Sampler Unit Function(When Sampler Unit OPSU-11 is
1.4.2 Measurement Principle ............................................................ 3
installed.)............................................................................. 13
1.4.3 Sample Aspiration Volume and requird volume .......................... 4
1.5.8 Manual Mode Measurement Function .................................... 15
1.4.4 Analysis Mode and Throughput ............................................... 4
1.5.9 Capillary Mode Measurement Function.................................. 16
1.4.5 Measurement, analysis Parameters ........................................ 4
1.5.10 Abnormal Sample Monitoring Function ................................... 16
1.4.6 Analysis and Display Range .................................................... 6
1.5.11 Abnormal Data Monitoring Function ........................................ 16
1.4.7 Reproducibility ......................................................................... 6
1.5.12 Analysis Result Storage Function............................................ 17
1.4.8 Accuracy .................................................................................. 6
1.5.13 Stored Data (Analysis Result) Processing Function................ 17
1.4.9 Linearity ...................................................................................... 6
1.5.14 Stored Data Processing Function by Patient ID ...................... 18
1.4.10 Dilution Linearity ....................................................................... 6
1.5.15 Patient Information Control Function....................................... 19
1.4.11 Carryover .................................................................................. 6
1.5.16 Function corresponding to XT pro (Standard Function, equal to
1.4.12 Difference between Manual Mode and Sampler Mode............. 7 XT-1800i)................................................................................. 19
1.4.13 Stability ..................................................................................... 7 1.5.17 Available Languages ............................................................... 19
1.4.14 Within-a-Day Stability after Collection of Blood ........................ 7 1.5.18 Shut Down Function................................................................ 20
1.3.1 Variation
1.1 OUTLINE
Table 2: Dilution ratio, required sample volume, required reagent volume 1.4.4 Analysis Mode and Throughput
(Capillary Mode)
Required: The below analysis modes are available.
Whole Blood 20 uL (1) Sample Aspiration Mode (XS-1000i, XS-1000iC)
Dilution 120 uL Manual Mode
Dilution Ratio 7 times Capillary Mode
Total volume of coordinated sample 140 uL
Auto sampler Mode
Sample Reagen Measured
Dilution Volume t sample (2) Sample Aspiration Mode (XS-800i)
Channel μL Reagent
Ratio volume volume(W.bloo Manual Open Mode
(W.Blood) μL d convert) μL
STROMATOLYSER-4DL 1000
Capillary Mode
WBC/DIFF 137 7.9 0.7
STROMATOLYSER-4DS 30
(3) Discrete Mode and Throughput
RBC/PLT 1563 1.29 CELLPACK 1000 0.0066
Model Analysis mode Discrete Throughput
CELLPACK 1000 - XS-1000i Manual CBC Approx. 60
HGB 2340 0.64
SULFOLYSER 500 - XS-1000iC (Open/closed) samples/hour
CBC+DIFF Approx. 60
samples/hour
1.4.3 Sample Aspiration Volume and requird volume Capillary CBC Approx. 49
(Open) samples/hour
Same volme in CBC or CBC+DIFF mode. CBC+DIFF Approx. 49
samples/hour
Model Analysis mode Aspirated volume Required sample Sampler CBC 20 samples
volume /approx.23
XS-1000i Manual/W.Blood 20uL Standard sample minutes
XS-1000iC tube:500uL CBC+DIFF 20 samples
/approx.23
Micro-sample
minutes
tube:90uL
XS-800i Manual CBC Approx. 60
Manual/Capillary 67uL (7 times Micro-sample (Open/ samples/hour
dilution) tube:140uL
CBC+DIFF Approx. 60
Sampler/W.Blood 20uL Standard sample samples/hour
tube : 1mL
Capillary CBC Approx. 55
XS-800i Manual/W.Blood 20uL - (Open) samples/hour
Manual/Capillary 67 uL (7 times - CBC+DIFF Approx. 55
dilution) samples/hour
Research
W.Blood Capillary
N.America
Screen
Parameter Remarks
For
DIFF
DIFF
CBC
CBC
White Blood Cell Count (WBC) WBC Histogram is indicated
v - v - v v
CBC for research.
Parameters Red Blood Cell Count (RBC) v - v - v v
Hemoglobin (HGB) v - v - v v
Hematocrit (HCT) v - v - v v
Mean Corpuscular Volume (MCV) v - v - v v
Mean Corpuscular Hemoglobin (MCH) v - v - v v
Mean Corpuscular Hemoglobin
v - v - v v
Concentration (MCHC)
Platelet Count (PLT) v - v - v v
RBC Distribution Width (RDW-SD,
v - - - v v
RDW-CV)
Platelet Distribution Width (PDW) v - - - - v
Mean Platelet Volume (MPV) v - - - v v
Platelet -Large Cell Ratio (P-LCR) v - - - - v
Plateletcrit (PCT) v - - - - v
Neutrophil Percent (NEUT%) - v - v v v
DIFF Lymphocyte Percent (LYMPH%) - v - v v v
Parameters Monocyte Percent (MONO%) - v - v v v
Eosinophil Percent (EO%) - v - v v v
Basophil Percent (BASO%) - v - v v v
Neutrophil Count (NEUT#) - v - v v v
Lymphocyte Count (LYMPH#) - v - v v v
Monocyte Count (MONO#) - v - v v v
Eosinophil Count (EO#) - v - v v v
Basophil Count (BASO#) - v - v v v
Research
W.Blood Capillary
Screen
N.America
Parameter Remarks
CBC
DIFF
CBC
DIFF
For
Immature Granulocyte Percent (IG%) - v - v v v Option
DIFF Immature Granulocyte Count (IG#) - v - v v v Option
Parameters - - - - - v When DIFF is analyzed.
Other %
Other # - - - - - v When DIFF is analyzed.
* In the Research screen, result can be displayed as Non-Reportable, without being masked.
1.4.12 Difference between Manual Mode and Sampler Mode 1.4.14 Within-a-Day Stability after Collection of Blood
None. (Using the same Aspiration system.) Sample should be stored from 18 to 26°C or cold place (from 2 to
8°C).
1.4.13 Stability (1) Blood Cell Classification (NEUT%, LYMPH%, MONO%, EO%,
BASO%)
The stability obtained by standard analysis method is shown below. Fluctuation ratio of WBC 5 classification values for normal
However, aging of sample itself is not considered. person's sample passed over from immediately after collection of
blood to 36 and 48 hours should be within the range shown in
(1) Stability related to Temperature Appendix 1-3 at 95% or more of probability till 36 hours, at 90% or
When fresh normal blood or control blood is used, the data more probability till 48hours.
fluctuation obtained by standard analysis method should be Fluctuation range of WBC count obtained from fresh blood should
within the range shown in Appendix 1-3. Variance range of be within the range shown in Appendix 1-3 at 95% or more
temperature should be within from 15 to 30°C both reagent and probability.
instrument.
Fresh normal blood should be used within 12 hours after (2) HCT, MCV
collection. Fluctuation ratio of HCT and MCV for healthy person's sample
Any change in the sample should be subtracted from change passed over from immediately after collection of blood to 8 and
ratio. 24 hours (24 hours and 48 hours for XS-1000iC) should be within
the range shown in Appendix 1-3 at 95% or more of probability.
(2) Within-a-Day Stability When the sample stored in cool place is used, it's required to
The data fluctuation obtained by standard analysis method using back to room temperature. Analysis time needs 2 times or more,
control blood should be within the range shown in Appendix 1-3. and it's needed to adopt average.
(1) Required Reagent volume for sample analysis Table 7: Required Reagent volume at starting up
Background check Background check
Table 5: Required Reagent volume for one sample analysis once twice
CBC
Discrete Mode CBC Total volume Approx.167mL Approx.38.5mL
+DIFF
Total volume Approx.34.5mL Approx.38.5mL CELLPACK Approx.162mL Approx.36mL
STROMATOLYSER-4DL Approx.4mL Approx.2mL
CELLPACK Approx.32mL Approx.32mL
STROMATOLYSER-4DS Approx.0.09mL Approx.0.03mL
STROMATOLYSER-4DL 2mL 2mL
SULFOLYSER Approx.1.0mL Approx.0.5mL
STROMATOLYSER-4DS - 0.03mL
SULFOLYSER Approx.約0.5mL Approx.0.5mL
(4) Required Reagent volume at shutdown
2) Required Reagent volume for auto rinse
Table 8: Required Reagent volume at shutdown
CELLPACK Approx. 95 17 mL
Table 6: Required Reagent volume for auto rinse
Background check Background (5) Required Reagent volume for resume from sleep
once check twice
Total volume Approx. 77mL Approx.38.5mL XS-1000i: 72 mL
CELLPACK Approx. 72mL Approx.36mL XS-800i: 74mL
STROMATOLYSER-4DL Approx. 4mL 2mL
STROMATOLYSER-4DS Approx. 0.06mL 0.03mL
SULFOLYSER Approx. 1.0mL Approx.0.5mL
* If the analysis is not performed more than 12 hours, the auto
rinse sequence will be performed when resuming from sleep
mode.
(1) Quality Control Method and Quality Control File Dairy QC data is sent to compile center through communication line,
and customer can refer to the result
Table 11: Quality Control Method and Quality Control File
Point File
Name Analysis Mode level
Number Number 1.5.5 Discrete Analysis Function
Control 300
Control Material 1, 2, 3 Refer to “1.4.3 Analysis Mode and Throughput”.
X Contro Material /Lot
l or 300 20
L-J OTHER1 (Human Blood) -
/Lot 1.5.6 Analysis Order Registration Function (Work Load List
Control
300 Function)
OTHER2 (Human Blood) -
/Lot
XM Xbar-M (Human Blood) - 300 1 (1) Registration Number
Control 1,000 samples of analysis information can be registered.
12 - 15mm
40mm
error occurs. A
Barcode
a) X M Limit Error
label
b) ID Read Error (Rack ID Read Error function is not employed.)
c) Abnormally Low Value
21mm
d) Abnormal Value (This does not mean Patient Limit but for
Sampler Stop Condition.)
e) Abnormal Sample Aspiration (Barcode should be positoned within the A area.)
f) Expired Control Blood
g) No Registered Control Blood
Barcode label specification should be as below.
(8) Input of Sample ID, Rack ID, and Rack Position
a) Manual Input
b
d
It's possible to input Sample ID, Rack ID, and Rack Position by
keyboard of IPU before analysis is started. 41624269960424
b) Input from Barcode Reader
If built-in type barcode reader is installed, it's possible to set or a c a
cancel reading sample/Rack ID from bar code affixed on the
e
sample collection tube and sample rack.
c) Auto Increment (Barcode label size)
If bar code is not used, sample ID and rack ID are
automatically incremented. <Bardoce label specificaiton>
d) The position of Barcode label appixed Margin (dimensiona) 5mm and more
In the analysis of sampler mode, the barcode label Bar height (dimensionb) 20mm and more
positonshould be as blew. Effective area (dimensionc) 40mm and below
width (dimensiond) 30mm and more
length (dimensione) 58mm and below
Narrow range ― 0.19mm and more
(1) Blood Collection Tube and Sample Screen and detect the abnormal morphology flags using histogram,
a) Blood Collection tube scattergram and counted data. Positive or Negative is judged by this
Same as “1.5.8 Manual Mode Measurement Function”. function.
b) Sample
Diluted sample which whole blood is diluted 7 times.
1.5.11 Abnormal Data Monitoring Function
(2) Input of Sample ID
Same as “1.5.8 Manual Mode Measurement Function”. (1) Abnormal Data (Patient Limit)
The Normal range of each parameter (Upper and Lower limits) can
(3) Input of Order be set and monitored.
Same as “1.5.7 Sampler Unit Function”. If the patient information program (optional) is installed, normal
range for each condition of age and gender can be set.
(4) Display of Sample ID and Order
Same as “1.5.8 Manual Mode Measurement Function”.
(2) Linearity Error
(5) Measurement Start Operation Analysis data is monitored whether it is within the linearity assured
Same as “1.5.8 Manual Mode Measurement Function”. range.
(3) Input/Edit Japanese, English, German, Italian, French, Spanish, Chinese, Greek,
Input or edit operation is possible from IPU. It is possible to input Russian and Portuguese are available.
the information with the Patient ID by transfering from Host
Computer.
(4) Display
It is possible to display on the IPU screen.
Rinsing of instrument is performed, and the instrument will become Alarm sounds when the error occurs.
the status that power of Main Unit can be OFF.
(1) Shut Down starts (1) Alarm Sound
a) Selection of Shut down Mode When error occurs on Main Unit, alarm sounds from Main Unit.
After shut down mode is selected, it proceeds in the next step.
(Can be canceled) (2) Stop of Alarm Sound
b) Starting Shut Down Alarm sound from Main Unit can be stopped by operating IPU.
However, if IPU does not communicate with Main Unit (e.g Power
(2) Shut Down Operation of IPU is OFF), power of Main Unit should be turned OFF.
Rinsing of manual whole blood line, CP whole blood line, and
measurement line are performed. (3) Alarm Setting
(3) Output Alarm sound (tone) on the Main Unit can be set (selected).
Message that shows power of Main Unit can be turned OFF is
displayed on the IPU screen. Re-start of the Main Unit is
possible.
The following maintenance functions are available. Alarm sounds when waste tank is full.
(2) IPU
Fuse (depends on the PC specification.)
Table16: Maintenance items and time to do Table 18: Spare pars for replacement
Items Time to do Condition Item Included Qty
Shut down Once/day Every 100 analysis Main Unit Fuse 1 set
Piercer Replacement Once/30000 CP Every 30000 CP Tubings Some
analysis analysis
Pump Replacement Once/30000 analysis Every 30000 (2) Tools
(Pneumatic Unit) analysis
Flow Cell Rinse As required Table 19: Tool
RBC Detector Clog As required Item Included Qty Memo
Removal Screwdriver (+) 1
Flow Cell Bubble As required Transducer Brush 1 RBC Detector Clog Removal
Removal
Reagent As required
Replacement
Waste Container As required
Replacement
Fuse Replacement As required
60 dB and below during operation (excluding the accidental noise) 1.9.1 Environmental Requirements
45 db and below (at Ready status)
(1) Ambient Temperature: 15 to 30°C (The reagent temperature
should also be within this range.)
1.8 Dimension and Weight
(2) Relative Humidity: 30 - 85%
Table 20: Dimension and Weight
XS-1000i, Dimensions (W x D x H) (mm) Weight
(3) Atmospheric Pressure: 9kPa – 106kPa
XS-1000iC
Main Unit 320x 413x 403 Approx. 24 kg (4) Installation Condition: Avoid installation in a place where the
Data Processing instrument may be exposed to the direct sunlight, dust, vibration,
Depends on the PC specification. or acid.
Unit
RBC (RDW-
Distribution SD) - - 3.0% and less Not measured - - -
Width
RBC (RDW-
Distribution CV) - - 3.0% and less Not measured - - -
Width
Platelet (PDW)
Distribution - - 10.0% and less Not measured - - -
Width
Mean (MPV)
Platelet - - 4.0% and less Not measured - - -
Volume
Platelet (P-LCR
Large Cell ) - - 18.0% and less Not measured - - -
Ratio
Plateletcrit (PCT)
- - 6.0% and less Not measured - - -
Value
Reproducibility Accuracy
Parameters Analysis Display Range W. Blood Mode Capillary Mode Condition
W.Blood Mode Capillary Mode
Range
Neutrophil 30.0NEUT% and
1) r=0.90 and more
Percent more, 1) r=0.70 and more
(NEUT%) - - 8.0% and less 16.0% and less 2 2)±3.0 NEUT% and
WBC 40×10 /μL and 2)±3.0 NEUT% and less
less
more
Lymphocyte 15.0LYMPH% and
1) r=0.90 and more 1) r=0.70 and more
Percent more,
(LYMPH%) - - 8.0% and less 16.0% and less 2 2)±3.0 LYMPH% 2)±3.0 LYMPH% and
WBC 40×10 /μL and
and less less
more
Monocyte Percent 5.0MONO% and
1) r=0.75 and more
more, 1) r=0.6 and more
(MONO%) - - 20.0% and less 40.0% and less 2 2)±2.0 MONO%
WBC 40×10 /μL and 2)±2.0 MONO% and less
and less
more
2
Eosinophil - - 25.0% and less, or 40.0% and less WBC 40×10 /μL and 1) r=0.80 and more 1) r=0.60 and more
Percent (EO%) ±1.5EO% and less more 2)±1.0 EO% and 2)±1.0 EO%以内
less
2
Basophil Percent - - 40.0% and less, or 50.0% and less, or WBC 40×10 /μL and 1) r=0.50 and more 1) r=0.50 and more
(BASO%) ±1.0BASO% and ±1.5BASO% and less more 2)±1.0 BASO% and 2)±1.0 BASO% and less
less less
2
Neutrophil Count - - 8.0% and less 16.0% and less 12.0×10 /μL and - -
(NEUT#)
more
2
Lymphocyte - - 8.0% and less 16.0% and less 6.0×10 /μL and - -
(LYMPH#) more
Count
2
Monocyte Count - - 20.0% and less 40.0% and less 2.0×10 /μL and - -
(MONO#)
more
Eosinophil Count - - 25.0% and less, or 40.0% and less, or - - -
2 2
(EO#) ±1.2×10 /μL and ±1.2×10 /μL and less
less
Basophil Count - - 40.0% and less, or 50.0% and less, or - - -
2 2
(BASO#) ±0.6×10 /μL and ±0.6×10 /μL and less
less
±2% or ±7% or
Hemoglobin Concentration (HGB)
±0.2g/dL and less ±0.7g/dL and less
0.0-25.0g/dL -
±3% or ±6% or
Hematocrit (HCT)
±1 HCT% and less ±2.0HCT% and less
0.0-60.0HCT% -
Variation of RBC 4
Mean Corpuscular RBC Approx.450×10 /μL,
(MCV) concentration should be - 4 -
(erythrocyte) volume 200<RBC<700×10 /μL
±2fL and less
Mean Corpuscular
(MCH) - - - -
Hemoglobin
Mean Corpuscular
(MCHC) - - - -
Hemoglobin Concentration
4 4
±5% or ±1×10 /μL and 0-100.0×10 /μL, 2
4 1. ±5% or ±1×10 /μL 4
less ±10% or Note, RBC is approx.400×10 /μ, 1. 0-200.0×10 /μL
Platelet Count (PLT)
RBC influence should be
4
±2.0×10 /μL and less 4
and PLT is approx.30.0×10 /μL,
and less 4
2. 200.1-500×10 /μL
4 4 2. ±16%
±2×10 /μL and less 200<RBC<700×10 /μL
RBC Distribution Width (RDW-SD) - - - -
RBC Distribution Width (RDW-CV) - - - -
Platelet Distribution Width (PDW) - - - -
Mean Platelet Volume (MPV) - - - -
Platelet Large Cell Ratio (P-LCR) - - - -
Plateletcrit Value (PCT) - - - -